Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Similar documents
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Author. Abstract. Introduction

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

MassHunter TOF/QTOF Users Meeting

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Application Note. Authors. Abstract. Introduction. Environmental

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

MassHunter Software Overview

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Making Sense of Differences in LCMS Data: Integrated Tools

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Agilent All Ions MS/MS

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Agilent MassHunter Profinder: Solving the Challenge of Isotopologue Extraction for Qualitative Flux Analysis

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

USP<467> residual solvents

Highly Sensitive and Rugged GC/MS/MS Tool

Application. Gas Chromatography February Introduction

Agilent TOF Screening & Impurity Profiling Julie Cichelli, PhD LC/MS Small Molecule Workshop Dec 6, 2012

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

GC Analysis of Polybrominated Flame Retardants Application

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Agilent MassHunter Quantitative Data Analysis

Determination of Volatile Substances Proof of Food Adulteration

Agilent G2350A Atomic Emission Detector (AED)

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

EGASUS. 4D GCxGC TOFMS. Delivering the Right Results

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Multi-residue analysis of pesticides by GC-HRMS

Agilent MassHunter Quantitative Data Analysis

Identifying Disinfection Byproducts in Treated Water

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Thermal Desorption Technical Support

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

GC-CI-MS analysis of TMS derivatives

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Supplementary Materials: Model-based peak alignment of metabolomic profiling from comprehensive two-dimensional gas chromatography mass spectrometry

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Workflow Using Hyphenated Mass Spectrometry Techniques

Expectations for GC-MS Lab

U.S. EPA Method 8270 for multicomponent analyte determination

ENVIRONMENTAL analysis

Understanding Your Spectra Module. Agilent OpenLAB CDS ChemStation Edition

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

Agilent G3212 GC-APCI Source

Application. Gas Chromatography March 1998

Low level quantification of NDMA and non-targeted contaminants screening in drinking water using GC Orbitrap mass spectrometry

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

NCFS Sample Preparation and Instrumental Parameters

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Transcription:

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting Application Note Authors Angela Smith Henry and Bruce Quimby Agilent Technologies, Inc. Wilmington, DE Introduction Laboratories conducting water analysis increasingly need to screen samples for a large number of compounds prior to performing a full quantitative analysis. The qualitative analysis of extracted water samples with GC/MSD provides the ability to understand what is in the sample at approximate levels. Existing qualitative screening workflows depend on manual screening, which is extremely time-consuming and highly dependent on the analyst s skill. Manual screening processes can also lead to overlooked or misassigned compounds, potentially due to complex matrices. Additionally, the discrepancy between analysts results and any bias associated with compound identification can lead to a significant amount of time exhausted on data analysis. Typically, a compound list for manual review is approximately 50 compounds. Since each compound is reviewed and identified by its retention time (RT), mass spectrum, and target and qualifier ion ratios, greatly increasing the number of compounds to review per single sample would multiply the complexities already faced. For example, if an analyst were to screen a sample for 1,000 possible compounds, it could take up to 18 hours to review that single sample. The Agilent SureTarget GC/MSD Water Pollutants Screener provides a straight forward and easy analysis workflow for the qualitative screening of water samples. Not only will the Agilent SureTarget analysis workflow allow for the fast screening of a large, wide-reaching compound list, but it will also remove bias and inconsistencies in compound identification. The SureTarget GC/MSD Water Pollutants Screener is preconfigured with the optimal hardware, consumables, software, and analytical methods to allow for the fast implementation of screening methods for pollutants in water.

Experimental Water samples were collected from two sources: unfiltered tap water in Lehigh County, Pennsylvania, USA, and effluents from the Wilmington, DE, USA wastewater treatment plant. From the Wilmington wastewater plant, three samples were drawn: Primary effluent Sedimentation stage Secondary effluent Biological content degradation Final effluent Final filtration and disinfection Sample preparation Each water sample, including the blanks, were extracted through liquid-liquid extraction (LLE). Three milliliters of dichloromethane (DCM) were added to a 30 ml water sample, shaken for 5 minutes, and the DCM layer was extracted and deposited into 2-mL autosampler vials for analysis. Reference standards Method 8270 Semi-Volatile by Capillary GC/MS mixtures (Mixes 1, 2, 3, 4A, 4B, 5, and 6) were purchased from AccuStandard and diluted to concentrations ranging between 100 ppb and 10 ppm in DCM (86 compounds in total). One microliter of the Reference Gas Oil mix (RGO) was spiked into each of the diluted AccuStandard samples to provide a complex matrix. Instrumentation All analyses were run on an Agilent 7890B Gas Chromatograph (GC) coupled with an Agilent 5977B InertPlus Mass Spectrometer Detector (MSD). A CO 2 -cooled multimode inlet (MMI) was used to temperature program the inlet, which provided additional separation between the DCM solvent peak and some early-eluting semivolatile compounds. Table 1 provides the GC and MSD method parameters for data acquisition. Data analysis All samples were analyzed using the SureTarget workflow with deconvolution in Agilent MassHunter WorkStation Software Quantitative Analysis Version B.08.00 for GC/MS. The results of the SureTarget workflow were then compared to the results of the (same) samples analyzed in MSD ChemStation Data Analysis program with Deconvolution Reporting Software (DRS), which uses AMDIS (deconvolution software developed by NIST) as the deconvolution tool. Table 1. GC Conditions Column Agilent HP-5MS UI, 30 m 0.25 mm, 0.25 µm Injection Solvent 1.0 µl cold splitless with CO 2 cooled MMI Dichloromethane Inlet temperature program Rate ( C/min) Temp ( C) Hold time (min) Initial 20 0.05 Rate 720 300 0 Carrier gas and mode RTL Compound and time Helium in constant pressure mode* Fluorene at 15.577 minutes Oven temperature program Rate ( C/min) Temp ( C) Hold time (min) Initial 40 2 Rate 10 300 8 MS transfer line temperature 280 C MS Conditions Solvent delay Scan acquisition range Tune GC and MSD Parameters for Data Acquisition 1.82 minutes 35 550 amu Etune.u Source temperature 250 C Quad temperatures 150 C * Final pressure dependent on RT-Locking procedure The SureTarget workflow is enabled in MassHunter Quantitative Analysis upon installation of the water screener feature disc. When the SureTarget analysis method is used, the deconvolution algorithm will remove the background and interfering mass fragments in the raw mass spectra from an identified compound s mass fragments (Figures 1 and 2). Figure 1A displays a total ion chromatogram (TIC) with an overlaid extracted ion chromatogram (EIC) for 1-naphthylamine in a time window of 14.5 to 15.5 minutes with the 1-naphthylamine peak located at 14.92 minutes. Figure 1B shows the complex raw mass spectrum under the peak selected at 14.92 minutes; the deconvolution algorithm cleans this complex mass spectrum of interferences. Figure 2 illustrates the result of the deconvolution process, which is a clean, deconvoluted mass spectrum for 1-naphthylamine at 14.92 minutes. The deconvoluted spectrum was compared, in a head-to-tail fashion, with the reference library spectrum of that compound. Next, the software automatically compared the deconvoluted mass spectra to the reference library, and generated a library match score (LMS). This LMS is based on the similarity between the deconvoluted mass spectrum and the library mass 2

Counts 10 4 5 4 1 23 A 13.8 14.0 14.2 14.4 14.6 14.8 15.0 15.2 15.4 15.6 15.8 16.0 16.2 16.4 16.6 10 4 Acquisition time (min) B 55.1 69.1 83.1 97.1 148.1 1.6 105.1 119.1 1.2 41.1 0.8 133.1 159.1 173.1 187.1 204.2 0.4 0 0 20 40 60 80 100 120 140 160 180 200 220 240 Acquisition time (min) Counts Figure 1. 10 3 1.0 0.8 0.6 0.4 0.2 0-0.2-0.4-0.6-0.8-1 Counts Figure 2. TIC (A) and raw spectrum (B) of 1-naphthylamine (from AccuStandard mixtures) in reference gas oil matrix. Deconvoluted compound mass spectrum 39 51 Reference library spectrum 30 40 50 58 63 72 89 60 Head-to-tail comparison of the deconvoluted mass spectrum (top) and reference library spectrum (bottom) for 1-naphthylamine. The deconvoluted spectrum is very different from the raw spectrum in Figure 1B, as the software was able to separate out all of the interfering ions. 115 115 126 143 143 70 80 90 100 110 120 130 140 150 Mass-to-charge (m/z) spectrum of the identified compound. The analysis method also looks for alternative peaks in the retention time window that produce high library match scores. This additional peak search aids in the review of the data, especially when analyzing complex matrices. The software then compares and matches each SureTarget deconvoluted mass spectrum to mass spectra in the NIST library and ranks each of the matches in the NIST hit list; a hit list is generated for each SureTarget identified compound. Figure 3 displays a graphical representation of the SureTarget workflow steps. Deconvolution Removes ions from background and coeluting interferences Library search Matches against a SureTarget reference library Alternative peaks in RT window Looks for alternate peaks that are a better match NIST search Looks for alternate hits that are a better match Results and Discussion Each prepared AccuStandard mixture included 86 compounds for identification, which are also found in the water screener library. These samples were run at 1 ppm, 200 ppb, and 100 ppb in RGO (complex matrix). After data acquisition, the SureTarget deconvolution workflow was completed on 10 samples per concentration level, and the results (number of identified compounds and LMSs) were averaged. The unfiltered tap water sample (Lehigh County, Pennsylvania), primary effluent, and final effluent of wastewater treatment (Wilmington, Delaware) were each averaged over five replicate runs to provide more analytical data for the deconvolution comparison. The number of identified compounds and their LMS values at each concentration were observed and documented for all AccuStandard mixtures and real-world water samples. Once the SureTarget workflow on the extracted water samples was complete, the data analysis was repeated with the DRS, which uses AMDIS (NIST software). Both sets of data analysis results were then compared (Table 2 for AccuStandard results, and Table 3 for tap water results). The data files were analyzed with DRS (AMDIS) and SureTarget to evaluate the ability of SureTarget. At each concentration level for AccuStandard samples (Table 2), both DRS and SureTarget reported a similar number of compounds, within a margin of error (fewer than four compounds difference on average), and similar LMS values (within ~1 point per concentration level). These results indicate that SureTarget performs very well and similarly to DRS (AMDIS), with respect to identified compounds and corresponding LMSs. Table 2. Approximate AccuStandard concentration Number of Identified Compounds and Average Library Match Scores (LMSs) of the AccuStandard 8270 Semivolatiles Mixture with a 0.1 % Reference Gas Oil (RGO) Spike per Software Package Deconvolution tool (software package) No. of compounds found (of 86) Average LMS 1 ppm AMDIS (DRS) 82 90.6 SureTarget (in Quant B.08) 84 89.4 200 ppb AMDIS (DRS) 69 80.2 SureTarget (in Quant B.08) 72 79.3 100 ppb AMDIS (DRS) 59 76.0 SureTarget (in Quant B.08) 62 75.6 Figure 3. The four major Agilent SureTarget workflow steps include mass spectral deconvolution in the RT window, library search of deconvoluted mass spectrum, identification of alternative peaks (in the RT window), and search of the NIST library. 3

Table 3. RT Extract of Lehigh County, PA, USA Unfiltered Tap Water Compound AMDIS (DRS) Avg. LMS 2.328 Trichloroethylene 82.2 85.3 2.388 Bromodichloromethane 90.2 93.5 3.419 Chlorodibromomethane 92.4 92.3 3.630 Tetrachloroethylene 87.3 79.9 4.817 Bromoform 83.0 84.2 21.986 Bisphenol A 59.3 64.4 SureTarget (Quant B.08) Avg. LMS In the unfiltered tap water sample (Lehigh County, Pennsylvania), six compounds were identified by both analysis programs with similar match scores (within five LMS points). Trihalomethanes (sanitation by-products), trichloroethylene, and bisphenol A were identified with high match scores in the tap water by both deconvolution analyses. With the confidence in the SureTarget deconvolution workflow, the wastewater samples (Wilmington, Delaware; wastewater plant) were examined with the SureTarget workflow in MassHunter Quantitative Analysis. Each wastewater sample was averaged over five replicate runs to achieve more confident results. Several compounds were identified in the primary and final wastewater effluent samples, including a mixture of concerning compounds. These compounds include 1,4-dioxane, tetrachloroethylene, codeine, and 4-tert-octylphenol. Other compounds identified in the primary effluent included caffeine, phentermine (weight loss drug), DEET (insect repellant), and cholesterol. Table 4 displays the compounds identified in the primary and final effluent. Table 4 also provides a picture as to what compounds were destroyed, or not destroyed, in the wastewater treatment process. DEET, triacetin, caffeine, TAED, and cholesterol were destroyed in wastewater treatment, while bromodichloromethane was introduced upon sanitation of the wastewater in the final stage. In contrast, several compounds were retained through the wastewater treatment process, including codeine, 1,4-dioxane, 4-tert octylphenol, phentermine, and diisobutyl phthalate. Table 4. RT Compounds Identified in Primary Wastewater Effluent (After Solid Separation Phase) and Final Effluent (After Sanitation of Wastewater) with Agilent SureTarget Deconvolution Compound Average LMS Primary effluent Final effluent 2.345 Bromodichloromethane 58.7 2.366 1,4-Dioxane 68.5 80.4 3.606 Tetrachloroethylene 79.7 54 9.619 a,a-dimethylphenethylamine (Phentermine) 69.1 65.2 10.031 Tributylamine 94.6 92.6 12.383 Triacetin 60.2 13.307 2,4,7,9-Tetramethyl-5-decyne-4,7-diol 75.3 55.6 15.500 N,N-Diethyl-m-toluamide (DEET) 83.8 15.776 4-tert-Octylphenol 84.8 60.1 16.223 N,N,N,N -tetraacetylethylenediamine (TAED) 59.1 18.610 Caffeine 91.1 18.804 Diisobutyl phthalate 84.3 67.8 24.278 Codeine 97.2 90.1 29.724 Cholesterol 79.1 Obtaining the results may be the first part of any water screening analysis, but data reports are needed for communicating the results. The SureTarget GC/MSD Water Pollutants Screener also supplies PDF reports to preview, summarize, and graphically display the data (Figure 4). The data preview report is designed to show the details (compound name, CAS number, RT, and LMS) associated with a primary peak of an identified compound, and any alternative peak in the RT window with a high LMS (greater than minimum) for that compound. The preview report indicates which compounds have alternative peaks (with the alternative RT) for further user review in the Quantitative Analysis batch Table (Figure 4). The summary report is designed to summarize the identified compounds, all compound related details (compound name and CAS number, RT, LMS, approximate concentration, and difference from reference RT in minutes), as well as the rank and score of the compound in the NIST hit list in a tabular view. If the SureTarget identified compound is not found in the NIST list (user-defined list, sized between 1 and 100, and set in method parameters), then the top hit will be recorded on a second line (Figure 4). The detailed graphics report dedicates one page to each identified compound in the sample (Figure 4). Each page will display the overlaid target and qualifier EICs, raw mass spectrum, deconvoluted spectrum, reference library spectrum, and the NIST spectrum (if the NIST hit is different from the identified compound). 4

Figure 4. Examples of PDF reports showing the preview report (left), summary report (middle) and a detailed graphics report page for one compound (right). Conclusions to support the complete optimal solution. These features include alternative peak identification in the RT window, NIST search, and customized report templates (preview, summary, and detailed graphics). The use of automated deconvolution data analysis considerably reduces the manual review time, while increasing the number of reviewable compounds to greater than 1,000. The library match score, generated from the deconvolution analysis, helps to communicate confidence in compound identification along with the ability to visually review a deconvoluted mass spectrum next to the reference library spectrum. The Agilent SureTarget GC/MSD Water Pollutants Screener with the SureTarget deconvolution workflow offers streamlined data analysis and reporting by enabling the automated separation and identification of compounds in complex matrices, such as tap and wastewater samples. The water screener includes the SureTarget deconvolution workflow, which is activated in Agilent MassHunter Quantitative Analysis (B.08.00), with additional features 5

For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2017 Printed in the USA February 16, 2017 5991-7834EN