Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Similar documents
Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Modernizing the USP Monograph for Acetaminophen

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application Note. Abstract. Authors. Environmental

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Determination of EU-Banned Disperse Dyes by LC/MSD TOF Application

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Determination of Pharmaceuticals in Water by SPE and LC/MS/MS in Both Positive and Negative Ion Modes Application

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

The Analysis of Organophosphate Pesticides by LC/MS Application

Measuring Drug-to-Antibody Ratio (DAR) for Antibody-Drug Conjugates (ADCs) with UHPLC/Q-TOF

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Application Note. Edgar Naegele. Abstract

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Agilent ESI and APCI sources: for polar to non-polar compounds

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Clinical Toxicology. Biomass Component Extraction: The uneaten cooked plant specimen was prepared for

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Accurate Mass Analysis of Hydraulic Fracturing Waters

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

Ultrasensitive EPA Method 1694 with the Agilent 6460 LC/MS/MS with Jet Stream Technology for Pharmaceuticals and Personal Care Products in Water

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Highly sensitive quantitative estimation of genotoxic impurities from API and drug formulation using LC/MS/MS

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Rapid quantification of Chinese medicine Zuo Jin Pill using rapid resolution liquid chromatography

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Transcription:

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS Application Note Traditional Chinese Medicine Author Rongjie Fu Agilent Technologies (Shanghai) Co. Ltd. Abstract Hydrophilic interaction chromatography (HILIC) was used for the analysis of the polar compound stachydrine from a traditional Chinese medicine. Two isocratic methods were developed using an Agilent ZORBAX Rapid Resolution High Definition (RRHD) HILIC Plus LC column with Agilent evaporative light scattering detection and Agilent 6410 Triple Quadrupole LC/MS. Both methods used acetonitrile/ammonium acetate as the mobile phase, and an appropriate organic composition was optimized individually for the different detectors. Both methods were suitable for the quantitative analysis of stachydrine, but the LC/MS/MS method provided higher sensitivity with low level stachydrine detection in complex samples.

Introduction Stachydrine is a major constituent of the traditional Chinese medicine from Leonurus japonicas Hoult, which is used to promote blood circulation and dispel blood stasis. The structure of stachydrine is shown in Figure 1. It is a strong polar compound with a quaternary-n, which makes it very difficult to retain on a reversed phase column. Recently, the HPLC separation technique of hydrophilic interaction chromatography (HILIC) has become popular for polar compound analysis. The retention mechanism of the HILIC mode is unique, because the aqueous phase is used as the strong elution solvent. Due to the weak UV absorption of the compound, both evaporative light scattering and MS detectors were employed. Experimental The sample was treated as follows to extract stachydrine for analysis [1]. Weigh 1 g of the dried Leonurus japonicus powder, add to 20 ml of 70% ethanol, and extract using sonication for 1 hour. Filter the extract with a 0.45 µm regenerated cellulose membrane filter (p/n 5064-8221) before injection for HPLC. Dilute the sample with acetonitrile to an appropriate concentration before analysis with LC/MS/MS. The LC/ELSD method was performed with an Agilent 1290 Infinity LC System, including an Agilent 1290 Infinity Binary Pump (G4220A), Agilent 1290 Infinity Autosampler (G4226A), Agilent 1290 Infinity Thermostatted Column Compartment (TCC) (G1316C), and Agilent 1290 Evaporative Light Scattering Detector (ELSD) (G4218A). + Figure 1. Structure of stachydrine. LC/ELSD conditions Column: Agilent ZORBAX RRHD HILIC Plus, 2.1 100 mm, 1.8 µm (p/n: 959758-901) Mobile phase: 80% Acetonitrile/20% 10 mm ammonium acetate Flow rate: 0.5 ml/min Temperature: 30 C Sample: 2 µl injection of 0.5 mg/ml standard solution in acetonitrile and prepared extract from Leonurus japonicus ELSD: gain = 7, filter = 3 s, evaporation temperature = 50 C The method for LC/MS/MS used the same Agilent 1290 Infinity LC but coupled to an Agilent 6460 Triple Quadrupole LC/MS. LC/MS/MS conditions Column: Agilent ZORBAX RRHD HILIC Plus, 2.1 100 mm, 1.8 µm (p/n: 959758-901) Mobile phase: 70% Acetonitrile/30% 10 mm ammonium acetate Flow rate: 0.5 ml/min Temperature: 30 C Sample: 1-µL injection of 10 µg/l standard solution in acetonitrile and prepared extract from Leonurus japonicus diluted with acetonitrile by 1/50,000 MS source parameters Gas temperatue: 300 C Gas flow: 5 L/min Vaporizer: 350 C Nebulizer: 45 psi Positive capillary: 3,500 V Sheath gas temperature: 250 C Sheath gas flow: 11 L/min Nozzle voltage: 500 V ESI acquisition parameters and transitions Compound Precursor ion Product ion Fragmentor Collision energy Polarity Stachydrine 144 84.1 125 25 Positive 144 58.1 125 25 Positive 2

Results and Discussion The polar compound stachydrine was well retained on the HILIC column with a value of K over 5 (Figure 2). Due to the complex components in the sample, the target compounds are usually difficult to separate. However, in this example, an interfering compound was easily coeluted with stachydrine. The mobile phase composition was optimized to achieve the ideal separation (Figure 2). The overlay of standard and sample is shown in Figure 3. The peak of stachydrine in the sample was confirmed using retention time consistency with the standard. mv 80 Stachydrine Standard 70 60 50 40 30 20 10 0 1 2 3 4 5 6 7 8 9 min mv 80 70 60 Sample 50 40 Stachydrine 30 20 10 0 0 1 2 3 4 5 6 7 8 9 min Figure 2. Stachydrine standard and extract from Leonurus japonicus analyzed on an Agilent ZORBAX RRHD HILIC Plus, 2.1 100 mm, 1.8 µm column with evaporative light scattering detection. mv 80 70 Stachydrine 60 50 40 30 20 10 0 0 1 2 3 4 5 6 7 8 9 min Figure 3. Overlay of stachydrine standard and extract from Leonurus japonicus, analyzed on an Agilent ZORBAX RRHD HILIC Plus, 2.1 100 mm, 1.8 µm column with evaporative light scattering detection. 3

The mobile phase for HILIC separations usually involves high organic and volatile additives, such as ammonium acetate and formic acid. High organic phases have higher volatility than traditional reversed phase liquid chromatography (RPLC) mobile phases, making HILIC well suited for applications with mass spectrometers [2]. Previous work demonstrated that the HILIC mode allowed more sensitive MS detection, due to more efficient spraying and desolvation in the ESI-MS source, as a result of the volatile mobile phase used. Using LC/MS/MS, a product ion from stachydrine was monitored during the separation. The product ion selected is unique to stachydrine, which makes it easy to be chromatographed. Figure 4 shows multiple reaction monitor (MRM ) chromatograms of standard and sample; only one peak was found. LC/MS/MS is, therefore, suitable for quantitative analysis of stachydrine in complex samples. The sub-2 µm Agilent ZORBAX RRHD column used with LC/MS/MS provided a narrow peak, which helped to enhance sensitivity. Counts Counts 10 3 3.4 3.0 2.6 2.2 1.8 1.4 1.0 0.6 0.2 10 4 1 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 + TIC MRM (**->**) 70 acn-10 ppb standard 1.d + TIC MRM (**->**) 70 acn-1-50,000 standard 1.d Standard Sample 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 2.2 2.4 2.6 2.8 3.0 3.2 3.4 3.6 3.8 Acquisition time (min) Figure 4. MRM chromatograms of stachydrine standard and extract from Leonurus japonicus analyzed on an Agilent ZORBAX RRHD HILIC Plus, 2.1 100 mm, 1.8 µm column with LC/MS/MS. 4

Conclusion The polar compound stachydrine was well retained on the Agilent ZORBAX RRHD HILIC Plus column and was easily separated in the sample for quantitative analysis. An LC/ELSD method was suitable for the routine analysis of stachydrine, but an LC/MS/MS method provided higher sensitivity and permitted the detection of low levels of stachydrine in complex samples. References 1. Leonuri Herba. China Pharmacopoeia, Part 1, 272 (2010). 2. Anne E. Mack. Comparing HILIC and RPLC of Morphine Using Agilent ZORBAX RRHD columns with UHPLC/MS, Application Note, Agilent Technologies, Inc., Publication No. 5991-0245EN (2012). For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2012 Printed in the USA August 29, 2012 5991-1047EN