The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Similar documents
Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Agilent ICP-MS Journal September 2009 Issue 39

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Samples compliant with the WATER

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Analysis of domestic sludge using the Agilent 4200 MP-AES

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Fergus Keenan Shona McSheehy Julian Wills

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Trace Analyses in Metal Matrices Using the ELAN DRC II

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

ICP-OES Application Note Number 35

ICP/MS Multi-Element Standards

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Validating performance of an Agilent 7800 ICP MS for USP <232>/<233> & ICH Q3D/Q2(R1): Elemental Impurity Analysis in Pharmaceuticals

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

Agilent 8900 Triple Quadrupole ICP-MS

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3165 Lot Number: EDTA NIST SRM 928 Lot Number: 928

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Agilent Technologies at TIAFT 2013

Agilent 8800 Triple Quadrupole ICP-MS: Understanding oxygen reaction mode in ICP-MS/MS

ATOMIC SPECTROSCOPY APPLICATIONS IN THE CONTRACT ENVIRONMENTAL LABORATORY. Primer July 2015 Edition

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Rapid Sample Assessment and Simplified Method Development with IntelliQuant

The New Agilent 7700 Series ICP-MS

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Single Analyte Custom Grade Solution g/ml (measured at 20 ± 1 ºC) ICP Assay NIST SRM 3163 Lot Number:

SeCtiOn 5 [STOCK AND CUSTOM] Assurance. Multi-Element CLP Standards

Single Analyte Mass Spec Solution ± µg/ml g/ml (measured at 20 ± 1 ºC) METHOD NIST SRM# SRM LOT#

Single Analyte Custom Grade Solution. CGWH2O1 M2-W H2O µg/ml ea: Tungsten Ammonium Metatungstate

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Elemental Scientific. Spike Recovery. Elemental Scientific

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

Overview of X-Ray Fluorescence Analysis

Single Analyte Custom Grade Solution ± 4 µg/ml g/ml (measured at 20 ± 1 ºC) Acidimetric NIST SRM 84L Lot Number: 84L

CERTIFICATE OF. Single Analyte Mass Spec Solution ± 0.1 µg/ml g/ml (measured at 20 ± 4 ºC)

Analysis Repeatability of Trace and Major Elements in a Water Sample

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Transcription:

The Agilent 77x ICP-MS Advantage for Drinking Water Analysis Application Note Environmental Authors Steve Wilbur Agilent Technologies, Inc. 338 146th Place SE, Suite 3, Bellevue Washington, 987 USA Introduction The Agilent 77 Series ICP-MS brings 3rd generation Octopole Reaction System (ORS 3 ) technology to routine environmental analysis of trace metals in drinking waters. Building on the cell technology used in the successful Agilent 75 Series ICP-MS, the ORS 3 has been completely redesigned to significantly improve the performance of helium (He) collision mode for the removal of polyatomic interferences. He mode is simple and universal, enabling the 77x ICP-MS to remove all polyatomic interferences under a single set of conditions, without requiring any previous knowledge about the sample, and without the need for complex mathematical corrections. The result is easier set-up, lower detection limits, faster analysis, and improved interference removal, especially for the difficult plasma based interferences affecting iron and selenium. Selenium detection limits less than 2 parts per trillion (ppt) are now routinely achievable in He mode.

Better Sensitivity, Better Accuracy, Better Dynamic Range In addition to a more efficient Octopole Reaction System (the ORS 3 ), the 77x ICP-MS also includes a new ion lens and detector, which deliver higher sensitivity and lower background compared to the 75cx ICP-MS it replaces. No special tuning or optimization is required for the routine analysis of water samples ranging from clean municipal tap waters to high total dissolved solids (TDS) mineral waters. Complete instrument optimization is fully automated, including the new One-click Plasma Setting function for fast, automated plasma setup, and a faster, more reproducible auto tune function (Expert AutoTune). Figure 1 shows typical calibrations obtained with the 77x ICP-MS. All calibration standards contained 1% HNO 3 and.5% HCl in order to stabilize Ag, Sb and Hg. While the use of HCl has traditionally been avoided in ICP-MS due to the formation of Cl based interferences on As, Se, Cr and V, the ORS 3 is able to remove these interferences in He mode, eliminating the need for hydrogen or other reactive gases. The 3.7 ppt background equivalent concentration (BEC) for vanadium demonstrates the efficient removal of the ClO + interference (mass 51) in He mode on the 77x. This interference is difficult to remove by other means, typically requiring highly reactive cell gases such as NH 3 or NH 3 /He, which are not suitable for multi-element analysis. 1 _ 1 9 Be [1] ISTD: 6 Li [1] y =.54 * + 6.1467E-5 R = 1. DL =.7827 ppb BEC =.1149 ppb 1 75 As [2] ISTD: 72 Ge [2] y =.87 * +.E+ R = 1. DL = ppb BEC = ppb 5.5. 5. 1. 15.. 5. 1. 15. 1 _ 1 1.5 51 V [2] ISTD: 45 Sc [2] y =.115 * + 4.2141E-4 R =.9999 DL =.194 ppb BEC =.3664 ppb 3 238 U [1] ISTD: 29 Bi [1] y =.214 * + 3.219E-5 R = 1. DL =.24 ppb BEC =.141 ppb 1 2.5 1. 5. 1. 15.. 5. 1. 15. Figure 1. Representative calibration curves for drinking water method. Be and U were analyzed in no gas mode, while As and V were analyzed in He mode. All standards contain 1% HNO 3 and.5% HCl. BEC and DL for As calculated based on cps +/- as acquired. Actual BEC and DL would have required longer counting times than those used in this method. 2

Typical method detection limits (MDLs, 3 sigma of 1 replicates of the low calibration standard) in ppt are shown in Table 1. These MDLs were not generated under highly optimized conditions, but using the default robust plasma conditions obtained by the One-click Plasma Setting routine (CeO + /Ce + <1%). The high sensitivity of the 77x ICP-MS means that accurate analysis of trace level contaminants in waters can be accomplished with ease, while the built-in High Matrix Introduction (HMI) system assures superior robustness and long term stability when running long sequences of high matrix samples. A sequence consisting of drinking waters, two high TDS mineral waters - samples "A" (TDS of 19 mg/l) and "B" (TDS of 39 mg/l), and NIST 1643e reference water (diluted 1/1) was analyzed continuously for 11 hours. The sequence included USEPA mandated continuous calibration validation and continuous calibration blank samples (CCVs and CCBs), giving a total of 19 analyses, performed with an average run time of less than 3.5 minutes per sample. CCV recoveries for all method analytes are plotted in Figure 2 for the entire sequence. No element in any CCV exceeded the USEPA limits of +/ 1% (shown in red). Table 2 lists the mean recoveries and relative standard deviations (%RSD) for the 28 repeat analyses of NIST 1643e (diluted 1/1 to test sensitivity and precision at low concentrations) which were included in the sequence. Table 1. The 3 Sigma Method Detection Limits in Parts Per Trillion (ppt) for Trace Elements in Drinking Waters. Note Fe and Se Detection Limits are Less Than 2 ppt in He Mode. Note Excellent DL For Be and B at 5 ppt. Mass Element MDL (ppt) Cell mode 9 Be 5.2 No gas 11 B 5. No gas 23 Na 58.5 No gas 24 Mg 2.8 No gas 27 Al 7.9 No gas 39 K 76.9 He 42 Ca 57.8 He 51 V 14.3 He 52 Cr 4.3 He 55 Mn 8.5 He 56 Fe 14.8 He 59 Co 4.4 He 6 Ni 14.7 He 63 Cu 2.7 He Mass Element MDL (ppt) Cell mode 66 Zn 14. He 75 As 11.9 He 78 Se 17.6 He 88 Sr 2.1 He 95 Mo 6.9 He 17 Ag 2.3 He 111 Cd 2.9 He 121 Sb 6.1 He 137 Ba 5.7 He 22 Hg 1.2 He 25 Tl 2.4 He 28 Pb 1.3 He 232 Th 1.8 He 238 U 1.7 He 3

Figure 2. CCV recoveries (2 ppb) for all method analytes for the entire 19 sample sequence. No CCV failures occurred. Mode [1] = no gas, Mode [2] = He mode Table 2. Mean Values, Relative Standard Deviation and % Recoveries of Certified Values for NIST 1643e (1/1 Dilution) Standard Reference Water. Element/mode Measured (ppb) RSD (%) Recovery (%) 9 Be [no gas] 14.4 2. 13.3 11 B [no gas] 156.3 1.6 99. 23 Na [no gas] 19581 2.7 94.4 24 Mg [no gas] 7376.6 2.9 91.8 27 Al [no gas] 137.5 2.8 97. 39 K [He] 243.5.9 1.5 44 Ca [He] 34251 2.6 16. 51 V [He] 36.6.7 96.9 52 Cr [He] 19.8 1. 97.2 53 Cr [He] 19.7.7 96.6 55 Mn [He] 38.5.6 98.9 56 Fe [He] 1.9 1.2 12.9 57 Fe [He] 1.2 1.4 12.2 59 Co [He] 25.8.6 95.5 Element/mode Measured (ppb) RSD (%) Recovery (%) 6 Ni [He] 58.2.7 93.4 63 Cu [He] 21.1.8 92.8 66 Zn [He] 73.2.5 93.3 75 As [He] 59.9 2.2 99.2 78 Se [He] 1.9 2.8 91.2 88 Sr [He] 31.5 4.5 93.3 95 Mo [He] 114.9.6 94.7 17 Ag [He].94 1.4 88.1 111 Cd [He] 6.32 1. 96.2 121 Sb [He] 56.2.5 96.3 137 Ba [He] 617.1 1. 113.4 25 Tl [He] 6..7 99.2 28 Pb [He] 18.9.2 96.1 4

Conclusions The new Octopole Reaction System (ORS 3 ), ion optics and detector of the Agilent 77x ICP-MS have produced an instrument that is fully capable of measuring all required elements at ppt levels in drinking waters without the need for either reactive cell gases or complex interference equations. The result is absolute confidence in data quality with significantly improved productivity. The 77x has the largest operating dynamic range of any quadrupole ICP-MS, and the integrated HMI system provides ultimate tolerance to high matrix samples. Furthermore, the simple optimization conditions used for drinking water analysis are identical to those used for more complex environmental samples such as soils and sludges, eliminating the need for the sample specific optimizations which are required on instruments that can only use reactive cell gases. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem/icpms. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 29 Printed in the USA October 23, 29 599-4315EN