CHEM*3440. Photon Energy Units. Spectrum of Electromagnetic Radiation. Chemical Instrumentation. Spectroscopic Experimental Concept.

Similar documents
R O Y G B V. Spin States. Outer Shell Electrons. Molecular Rotations. Inner Shell Electrons. Molecular Vibrations. Nuclear Transitions

Reference literature. (See: CHEM 2470 notes, Module 8 Textbook 6th ed., Chapters )

Reflection = EM strikes a boundary between two media differing in η and bounces back

Spectroscopy: Introduction. Required reading Chapter 18 (pages ) Chapter 20 (pages )

Because light behaves like a wave, we can describe it in one of two ways by its wavelength or by its frequency.

Classification of spectroscopic methods

9/28/10. Visible and Ultraviolet Molecular Spectroscopy - (S-H-C Chapters 13-14) Valence Electronic Structure. n σ* transitions

Skoog Chapter 6 Introduction to Spectrometric Methods

2001 Spectrometers. Instrument Machinery. Movies from this presentation can be access at

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Ultraviolet-Visible and Infrared Spectrophotometry

Chemistry 524--Final Exam--Keiderling May 4, :30 -?? pm SES

Instrumental Analysis: Spectrophotometric Methods

The Fundamentals of Spectroscopy: Theory BUILDING BETTER SCIENCE AGILENT AND YOU

Spectrophotometry. Introduction

Chapter 17: Fundamentals of Spectrophotometry

Chapter 17: Fundamentals of Spectrophotometry

PC Laboratory Raman Spectroscopy

Analytical Chemistry II

Ch 313 FINAL EXAM OUTLINE Spring 2010

Ultraviolet-Visible and Infrared Spectrophotometry

Design and Development of a Smartphone Based Visible Spectrophotometer for Analytical Applications

Model Answer (Paper code: AR-7112) M. Sc. (Physics) IV Semester Paper I: Laser Physics and Spectroscopy

10/2/2008. hc λ. νλ =c. proportional to frequency. Energy is inversely proportional to wavelength And is directly proportional to wavenumber

Lecture 0. NC State University

What is spectroscopy?

CHEM6416 Theory of Molecular Spectroscopy 2013Jan Spectroscopy frequency dependence of the interaction of light with matter

Chapter 13 An Introduction to Ultraviolet/Visible Molecular Absorption Spectrometry

MOLECULAR AND ATOMIC SPECTROSCOPY

Complete the following. Clearly mark your answers. YOU MUST SHOW YOUR WORK TO RECEIVE CREDIT.

n ( λ ) is observed. Further, the bandgap of the ZnTe semiconductor is

Vibrational Spectroscopies. C-874 University of Delaware

Molecular spectroscopy

IR Spectrography - Absorption. Raman Spectrography - Scattering. n 0 n M - Raman n 0 - Rayleigh

24 Introduction to Spectrochemical Methods

Electronic Excitation by UV/Vis Spectroscopy :

Chemistry Instrumental Analysis Lecture 3. Chem 4631

Lecture 3: Light absorbance

Course Details. Analytical Techniques Based on Optical Spectroscopy. Course Details. Textbook. SCCH 211: Analytical Chemistry I

NPTEL/IITM. Molecular Spectroscopy Lectures 1 & 2. Prof.K. Mangala Sunder Page 1 of 15. Topics. Part I : Introductory concepts Topics

two slits and 5 slits

An Introduction to Ultraviolet-Visible Molecular Spectrometry (Chapter 13)

Reference. What is spectroscopy? What is Light? / EMR 11/15/2015. Principles of Spectroscopy. Processes in Spectroscopy

Chapter 4 Ultraviolet and visible spectroscopy Molecular Spectrophotometry

Analytical Spectroscopy Review

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

Characterisation of vibrational modes of adsorbed species

Spectroscopy Problem Set February 22, 2018

Questions on Instrumental Methods of Analysis

An Introduction to Diffraction and Scattering. School of Chemistry The University of Sydney

CHAPTER 2. Preliminaries A Review of Optical Physics

Chemistry 524--Final Exam--Keiderling Dec. 12, pm SES

Homework Due by 5PM September 20 (next class) Does everyone have a topic that has been approved by the faculty?

Advanced Spectroscopy Laboratory

Wavelength λ Velocity v. Electric Field Strength Amplitude A. Time t or Distance x time for 1 λ to pass fixed point. # of λ passing per s ν= 1 p

CHEM Atomic and Molecular Spectroscopy

Ultraviolet-Visible Spectroscopy

Chem 155 Midterm Exam Page 1 of 10 Spring 2010 Terrill

Atomic Emission Spectroscopy

高等食品分析 (Advanced Food Analysis) I. SPECTROSCOPIC METHODS *Instrumental methods: 1. Spectroscopic methods (spectroscopy): a) Electromagnetic radiation

Chemistry 213 Practical Spectroscopy

( ) x10 8 m. The energy in a mole of 400 nm photons is calculated by: ' & sec( ) ( & % ) 6.022x10 23 photons' E = h! = hc & 6.

What happens when light falls on a material? Transmission Reflection Absorption Luminescence. Elastic Scattering Inelastic Scattering

25 Instruments for Optical Spectrometry

Comments to Atkins: Physical chemistry, 7th edition.

1901 Application of Spectrophotometry

Absorption spectrometry summary

Chemistry 304B, Spring 1999 Lecture 5 1. UV Spectroscopy:

Compact Knowledge: Absorbance Spectrophotometry. Flexible. Reliable. Personal.

Chapter 24 Photonics Question 1 Question 2 Question 3 Question 4 Question 5

Chem 310 rd. 3 Homework Set Answers

Analytical Technologies in Biotechnology Prof. Dr. Ashwani K Sharma Department of Biotechnology Indian Institute of Technology, Roorkee

General Considerations 1

Course: M.Sc (Chemistry) Analytical Chemistry Unit: III

DOWNLOAD OR READ : INFRARED AND RAMAN SPECTROSCOPY CONCEPTS AND APPLICATIONS PDF EBOOK EPUB MOBI

Outline of Recombinant DNA technology. Application of UV spectroscopy in recombinant DNA technology

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

Raman and stimulated Raman spectroscopy of chlorinated hydrocarbons

JABLONSKI DIAGRAM INTERACTIONS BETWEEN LIGHT AND MATTER LIGHT AS A WAVE LIGHT AS A PARTICLE 2/1/16. Photoelectric effect Absorp<on Emission ScaDering

24/ Rayleigh and Raman scattering. Stokes and anti-stokes lines. Rotational Raman spectroscopy. Polarizability ellipsoid. Selection rules.

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

Spectral Resolution. Spectral resolution is a measure of the ability to separate nearby features in wavelength space.

Teaching philosophy. learn it, know it! Learn it 5-times and you know it Read (& simple question) Lecture Problem set

Photochemical principles

Overview of Spectroscopy

University of Cyprus. Reflectance and Diffuse Spectroscopy

Characterisation & Use of Array Spectrometers

Singlet. Fluorescence Spectroscopy * LUMO

Instrumentelle Analytik in den Geowissenschaften (PI)

Stray Light Rejection in Array Spectrometers

Spectrophotometry. Dr. Shareef SHAIK ASST. PROFESSOR Pharmacology

EMISSION SPECTROSCOPY

Diffraction Gratings, Atomic Spectra. Prof. Shawhan (substituting for Prof. Hall) November 14, 2016

DEPARTMENT OF PHYSICS UNIVERSITY OF PUNE PUNE SYLLABUS for the M.Phil. (Physics ) Course

Module 4 : Third order nonlinear optical processes. Lecture 28 : Inelastic Scattering Processes. Objectives

Introduction to Spectroscopic methods

Mie vs Rayleigh. Sun

levels. The signal is either absorbance vibrational and rotational energy levels or percent transmittance of the analyte

Chem Homework Set Answers

Transcription:

Spectrum of Electromagnetic Radiation Electromagnetic radiation is light. Different energy light interacts with different motions in molecules. CHEM*344 Chemical Instrumentation Topic 7 Spectrometry Radiofrequency MHz Nuclear Spin NMR Microwave GHz Electron Spin Molecular Rotation ESR (EPR) Microwave Spec. Infrared THz Molecular Vibrations I.R. Spectroscopy UV-Visible Far UV - X-ray PHz PHz Valence Electron Transitions Core Electron Transitions Gamma Ray EHz Nuclear Transitions UV/Vis Spectroscopy XUV - X-ray Spectroscopy Mössbauer Spectroscopy Photon Energy Units Different units for light quanta energy are used for different experiments. E = h! = h c " = hc! ev = 865.6 cm - ev =.62 x -9 J ev = 24.8 THz ev = 239 nm MHz NMR GHz ESR cm - ev, nm IR UV-Visible Spectroscopic Experimental Concept Spectroscopic experiments involve probing a system with some particles so as to disturb its equilibrium and then monitor its response by observing the emission of other particles. Fluorescence Electrons Photons Sample Photons Photons or Scattering Experiment Absorption Experiment Correspondence between different units used in spectroscopy. ev, Å Far UV/X-ray Units used in different spectroscopic experiments or regions. Heat Chemical Reaction

Spectral Bandwidth Radiation emission is distributed over a certain wavelength range in a closeto-gaussian manner. Two defining parameters: Peak position Full Width at Half-Maximum (FWHM) The Spectrometer Spectroscopy infers the analysis of the distribution of radiant power by wavelength. Spectrometers separate the various spectral components into the spectrum. Generally we can categorize different spectrometers based upon the way they deal with the spectrum. Peak at 559 nm for both curves..75.5 9 nm.25 3 nm 54 55 56 57 58 Broadband radiation from which one selects a monochromatic band Dispersive spectrometers (diffraction gratings, prisms) Non-Dispersive spectrometers (filters, Fourier transform) Narrow emission lines Coherent sources (lasers, synchrotrons) Incoherent sources (atomic lamps) Dispersive Spectrometers Non-dispersive Spectrometers Array s Good to detect a single band of wavelengths. Interferometer. Light source Measure one wavelength at a time. A serial detection scheme. Grating Slits Light source Slits Lens Grating Measure a large range of wavelengths simultaneously. A parallel detection scheme. Array Light source Lens Filter Filter-selecting spectrometer Light source Lens Encoded beam Beam splitter Fixed mirror Moving mirror Lens Slits Monochromators. Encodes spectral data into time domain. Many advantages.

Spectral Resolution When are two different colours really different? The resolution if any spectral feature can be quantified by the peak separation and the valley depth. Two peaks are commonly said to have been resolved if the valley between them is 2%..75.5.25!" 9% valley 54 56 58 6 62 Resolving power is an instrument-specific measure: report as "/!". Measuring a Spectrum.25 Slit width just right.75.5.25 54 56 58 6 62.25.25.75.5.25 Slit width too narrow 54 56 58 6 62 Slit width too wide.75.5.25 54 56 58 6 62 Spectrometric Experiments Three general classes of spectrometric experiments: Light Scattering photon momentum not conserved measure particles size identify molecules Emission photons generated by sample quantify through atomic Absorption photons absorbed by sample identify and quantity molecules Scattering Experiments There are two types of light scattering experiments: Elastic, in which the energy of the photon does not change but its direction or momentum does, and Inelastic, in which the photon energy changes. Back Elastic Refractive index differences Polarizability Rayleigh, Mie Scattering Dynamic Light Scattering Right Angle Sample Inelastic Changing polarizabilities Brillouin (acoustic density fluctuations in solids) Raman (bond polarizability changes in molecular vibrations) Forward

Emission Experiments In these experiments, a sample is excited by injecting energy (heat, light, electrons, etc.). The sample relaxes by the emission of light and this light is analyzed. Arc Discharge Inductively Coupled Plasma (ICP) Fluorescence and Phosphorescence Energy-Dispersive (EDX) and Wavelength Dispersive (WDX) X-Ray Spectroscopy Absorption Experiments In this case, an incident beam impinges upon a sample and the light passes through it is monitored for its decrease in power due to absorption by the sample. Widely used for sample analysis. Incident Beam Transmitted Beam Excitation Energy Sample Emitted Light Monochromator Narrow band of emitted light Reflection Losses Scattering Losses The emerging beam has less power than the incident beam because of the absorption process of interest and also due to various loss mechanisms, such as scattering and reflection. or Transmittance The absorption process is measured in one of two ways: Transmittance: T = P/P : A = -log T = log [P/P] In order to account for the numerous losses other than absorption, an experimental approximation to the above definitions is to measure P with a cell containing everything (solvent) except the analyte of interest. The expectation is that the losses will be the same in both sample and reference cells, thereby canceling the effect in both. This approximation is acceptable and constitutes our working definition for A and T. Transmittance: T = Psample/Preference : A = log [Preference/Psample] Note that the sample and reference switch places in the definitions. THis makes T a number between and (the transmitted beam is always weaker than the incident beam) and A a positive number. Beer s Law Model the absorption process by assuming each absorbing molecule has a a cross-section within which a photon will absorbed if it reaches it. The proportion of the power absorbed from the beam is equal to the proportion of the absorbing cross-section to the total cross-section of the beam. When this relation is solved, it leads to the well-known Beer!s Law or more completely the Beer-Lambert Law. A = log[p/p] = #bc A spectroscopic experiment because # depends upon wavelength and so the absorbance is wavelength dependent too. adds in a multicomponent system (assuming no interaction between the components). Transmittance is the property measured in an experiment but absorbance is the property directly related to concentration. concentration path length molar absorptivity (molar extinction coefficient)

Limitation s to Beer s Law Is the absorbance really linear with respect to the variables? Path length (b): Essentially this is always found to be linear. Concentration (c): Nonlinearity can arise from Intermolecular interactions Shifting chemical equilibria Molar absorptivity (#): Non-linearity can arise from the solution!s index of refraction. The instrument itself can skew the behaviour away from linearity in a number of ways. Intermolecular Interactions Beer!s Law is strictly a limiting law for dilute solutions. At high concentrations ( $>. M) the average distance between analyte molecules is small enough that the charge distributions around one affects that around another. Sometimes high concentrations of inert electrolytes can themselves alter the absorptivity of a species present even at low concentrations. Some organic molecules can show deviations even at -6 M concentrations Need to be aware of concentration linearity and confirm its validity in a given experiment. Chemical Equilibrium When a substance is involved in a chemical reaction, the extent of that reaction is concentration dependent. If the alternate form of the molecule has a different absorption spectrum, there will be non-linear distortion away from Beer!s Law..8.6.4.2 of Acid-Base Indicator 4 8 2 6 Concentration (M x 5 ) Index of Refraction The molar absorptivity depends upon the index of refraction of the solution. In some cases, the index of refraction can change with concentration. Use an expression for molar absorptivity # which is dependent upon the index of refraction. Often this is approximated by the expression #(n) = #[n/(n 2 + 2) 2 ] When a concentration change causes a significant change in the refractive index, then this can cause a deviation in Beer!s Law away from linearity. In practice, this correction is never very large and is rarely significant at concentrations below. M. Example: An indicator dye with Ka =.42 x -5. is measured at two wavelengths (43 and 57 nm), Note the curvature in the plots.

Polychromatic Radiation Beer!s Law is strictly applicable only for monochromatic radiation. All excitation sources have a non-zero bandwidth. Analyte will have a different absorptivity at each wavelength. If variation is large, then the non-linearity can be observed. Photometric Accuracy Experimental noise leads to an uncertainty in determining absorbance. Partial differentiation leads to a relationship between error in T and error in A.!A = "( log e) T!T Remedy: Choose a spectral range where the absorptivity changes slowly with wavelength. Select an excitation radiation bandwidth that is <. of the analyte!s spectral FWHM. This error in absorbance naturally is connected to an error in concentration. Partial differentiation of Beer!s Law leads to the relationship!c = " b!a = # log ( e) " bt!t Error in A Assume a fixed relative T error of.5% (.5). measurements between A =.5 and A =.55 keeps the A error <5%. Relative Error Relative Error in..9.8.7.6.5.4.3.2..5.5 2 T Error Depends Upon T A careful analysis of the problem shows that different noise sources will contribute differently to T error. Three general cases have been identified: () T error is constant Limited readout resolution. Thermal detector (Johnson noise). Dark current and amplifier noise. (2) T error varies as (T 2 + T) /2 Shot noise (3) T error varies as T Cell positioning uncertainty Source /f noise

Constant Error in T Arises in inexpensive spectrometers which suffer from limited readout resolution. IR spectrometers will be subject to Johnson noise. Experiments where source intensity is low or detector sensitivity is low will be limited by dark current and amplifier noise. Relative Concentration Error..8.6.4.2 Relative Error in Concentration.5.5 2 Error % (T 2 + T) /2 High quality UV/Vis spectrometers are susceptible to this case. Relative Concentration Error..8.6.4.2 Relative Error in Concentration.5.5 2 Note how this process allows us to make measurements out to much higher absorbances. This can be done with high quality instruments. Need to be careful, however, that another error source doesn!t start to dominate. Error % T High quality UV/Vis and IR spectrometers will be subject to cell positioning errors. Inexpensive IR spectrometers will be subject to flicker noise. Relative Concentration Error Relative Error in Concentration..8.6.4.2.5.5 2 Choose Range Carefully The take-home message here is that just because a machine gives you a number, doesn!t mean you have to believe it. When making spectrometric measurements, you need to adjust the concentration of the sample so that the absorbance range covered falls in the region which will minimize the instrumental error. range between A =. and A = should give reliable results with almost all instruments. Good spectrometers can make absorbance measurements down to A = 2. Only the very best can make measurements beyond A = 3.

Slit Width Affects Measurements If a significant variation in absorptivity occurs over the spectral bandwidth admitted by the slot, a non-linear variation (non-beer!s Law) with concentration will be observed. This arises because the spectrometer measures the average transmissivity over the spectral bandwidth, but transmissivity and concentration are not linearly related. Other Problems Stray Light: it is a problem when working at the limits of a spectrometer's range. Cells and Solvents: everything besides the analyte should be as transparent as possible. Keep slit width large to increase S/N ratio but must keep it small enough to maintain a linear relationship with concentration changes. This effect is minimized if the absorptivity changes slowly with wavelength. Select a wavelength near a peak maximum. Use a slit width to provide a bandwidth that is about / of the spectral feature!s width. Sample Preparation: if two samples are prepared so that one carries along a greater concentration of insoluble particulates, then additional scattering will lead to an apparent greater absorption.