Appendix A. Glossary of Definitions, Acronyms, Abbreviations and Symbols. March 1997

Similar documents
Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

Do Now NOTHING! Someone 6 foot tall and 182 centimeters? Someone who is 182 lbs, 13 stone, or 82.5 kg? 1.5 minutes and 90 seconds?

2 Standards for Measurement. Careful and accurate measurements of ingredients are important both when cooking and in the chemistry laboratory!

LAMBTON SCIENTIFIC (A Division of Technical Chemical Services Inc.)

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

The Metric System and Measurement

Copyright ENCO Laboratories, Inc. II. Quality Control. A. Introduction

oz ounce (mass) = L = cm 3

Scientific Units of Measurement & Conversion

1 gram = kg = 1.0 x 10 3 kg. 1 milligram (1 mg) = 1.0 x 10 3 g. 1 microgram (1µg) = 1.0 x 10 6 g

3.2 Units of Measurement > Chapter 3 Scientific Measurement. 3.2 Units of Measurement. 3.1 Using and Expressing Measurements

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

ALLOWAY METHOD OUTLINE

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

3.2 The International System of Units HR/Name. temperature: Celsius scale: Kelvin scale: Vocabulary. absolute zero:

METHOD 8000A GAS CHROMATOGRAPHY

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

Calculations Used in Analytical Chemistry

STANDARD OPERATING PROCEDURES

Measurements in Chemistry Chapter 2

Sulfotepp impurities in Chlorpyrifos EC formulations


Chapter 4: Calculations Used in Analytical Chemistry

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Unit 1 - INTRODUCTION MEDICAL MATH Listening guide

The Metric System and Measurement

Measurements in Chemistry Chapter 2

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Co Curricular Data Analysis Review

Chemistry Basic Science Concepts. Observations: are recorded using the senses. Examples: the paper is white; the air is cold; the drink is sweet.

The number of stars in a galaxy is an example of an estimate that should be expressed in scientific notation.

5) If you count 7 pennies, you can only report one significant figure in that measurement. Answer: FALSE Diff: 1 Page Ref: 2.3

PREFIXES AND SYMBOLS SI Prefixes you need to know by heart

Required math skills:

Data Glossary. Acceptance criteria - specific limits placed on characteristics of an item, process, or service defined in requirements documents.

Measurements UNITS FOR MEASUREMENTS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Adding these quantities together gives the total kilogram moles.

INTRODUCTORY CHEMISTRY Concepts and Critical Thinking Seventh Edition by Charles H. Corwin

Bell Work: Pre AP 22-Aug Why is it important to be able to a document to somebody, give two reasons that are not school focused?

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

1 centimeter (cm) 5 10 millimeters (mm) 1 meter (m) centimeters. 1 kilometer (km) 5 1,000 meters. Set up equivalent ratios and cross multiply.

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

MEASUREMENTS. Significant Figures

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

APPENDIX B ABBREVIATIONS, SYMBOLS AND CONVERSION FACTORS Abbreviations

Name Date Class MEASUREMENTS AND THEIR UNCERTAINTY

Chapter 1 Introduction: Matter and Measurement Honors Chemistry Lecture Notes. 1.1 The Study of Chemistry: The study of and the it undergoes.

A. GENERAL NOTICES. Ninth Edition, which may be abbreviated as JSFA-IX.

Fundamental Concepts C020

APPENDIX H CONVERSION FACTORS

Math Skills Needed For Chemistry

Name Date Class MEASUREMENTS AND THEIR UNCERTAINTY

Lecture Presentation. Chapter 1. Chemistry in Our Lives. Karen C. Timberlake

ACCURACY is the nearness of a result or the mean of a set of results to the true or accepted value.

In recording measurements, it is necessary to understand 1. SIGNIFICANCE of numbers 2. importance of UNITS.

Name Period Date. Measurements. Fill-in the blanks during the PowerPoint presentation in class.

General Chemistry I Introductory Concepts. Units, dimensions, and mathematics for problem solving

Number vs. Quantity. Quantity - number + unit UNITS MATTER!! for a measurement to be useful, must include both a number and unit

Chloride by Flow Injection Analysis (FIA)

Results of the EPA Method 1631 Validation Study

LECTURE 4 - Units Used in Measurements

Clinical Chemistry Lecture Guide

Unit Conversions, Important Constants and Relationships

Chemistry 104 Chapter Two PowerPoint Notes

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Quality Assurance is what we do to get the right answer for our purpose FITNESS FOR PURPOSE

Measurement and Calculations

Solution Concentration

Chapter 1 (Part 2) Measurements in Chemistry 1.6 Physical Quantities

International System of Units 3.2. Slide 1of 33

Scientific Measurement

Chemistry Section Review 2.2

Laboratory 101: A Guide to Understanding your Testing Laboratory

2 Standards of Measurement

Ohio EPA Formula/Conversion Table for Water Supply and Distribution Exams

In addition to the information at the end of the exam, you will be given a periodic table.

Hexavalent Chromium by Flow Injection Analysis (FIA)

Skill Building Topic 1 UNDERSTANDING THE USES OF NUMBERS

Chemistry and Measurement

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

cular-view-of-our-world-5th-edition-by-tro/

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

PREFIXES AND SYMBOLS SI Prefixes you need to know by heart

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

Lecture 2 METRIC SYSTEM

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Everyday Conversion: Money

2.1 Units of Measurement. Copyright 2011 Pearson Education, Inc.

METHOD 3510B SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

METHOD 3520B CONTINUOUS LIQUID-LIQUID EXTRACTION

Glossary of Common Laboratory Terms

Handout Unit Conversions (Dimensional Analysis)

Functional Genomics Research Stream. Lecture: February 17, 2009 Masses, Volumes, Solutions & Dilutions

Welcome to Chemistry 121

Matter & Measurement. Chapter 1 Chemistry 2A

Transcription:

Appendix A Glossary of Definitions, Acronyms, Abbreviations and Symbols March 1997

Glossary Of Definitions, Acronyms, Abbreviations And Symbols 1.0 Standard Acronyms, Abbreviations, And Symbols 1.1 Units of weight and measure and their abbreviations ampere amp centimeter cm cubic centimeter cm 3 day do not abbreviate degree Celsius o C diameter dia foot ft gallon gal gram g grams per Liter g/l greater than > hour h hydrogen ion concentration, ph negative logarithm of inch in. inside diameter ID kilogram kg less than < linear dynamic range LDR liter L logarithm (common) log logarithm (natural) log e or ln maximum max meter m microgram µg microgram per liter µg/l microliter µl milliequivalent meq milligram mg milligram per gram mg/g milligram per liter mg/l milligram per milliliter mg/ml milliliter ml millimeter mm millimeter of mercury (pressure) mm Hg millisecond ms millivolt mv minimum min minute min A-1 Draft, March 1997

molal do not abbreviate molar M mole mol month do not abbreviate most probable number MPN nanogram ng nanogram per liter ng/l nanometer nm normal N number No. (only when followed by a numeral) outside diameter OD part per billion ppb part per million ppm part per quadrillion ppq part per trillion ppt per / (when used in expressions with unit symbols) percent % pico (prefix) p picogram pg pound lb pounds-per-square inch gauge psig quart qt relative centrifugal force RCF revolutions per minute rpm second s specific gravity sp gr micrometer um volt V volume (of a publication) Vol. (only when followed by a numeral) volume per unit volume v/v watt W weight per unit volume w/v year do not abbreviate 1.2 Alphabetical Abbreviations average avg calibration standard CAL equation Eq figure Fig. (only when followed by a numeral) Flame Ionization Detector FID gas chromatograph/chromatography GC gas chromatography/ GC/MS mass spectroscopy gel permeation chromatograph/ GPC chromatography A-2 Draft, March 1997

high performance liquid HPLC chromatograph/chromatography high resolution GC HRGC high resolution MS HRMS Kuderna-Danish concentrator K-D linear dynamic range LDR mass spectrometer/spectrometry MS mass to charge ratio m/z meta m molecular ion M most probable number MPN ortho o page p pages pp. para p quality assurance QA quality control QC quart qt reference Ref relative centrifugal force RCF selected ion current profile SICP solid phase extraction SPE Soxhlet/Dean-Stark extractor SDS 2.0 Definitions 2.1 Many of the following definitions were compiled from SW-846 Test Methods for Evaluating Solid Wastes (Section 3.1), the Environmental Methods Monitoring Council (EMMC) Guidelines and Format for Methods to Be Proposed at 40 CFR Part 136 or Part 141 (Section 3.2) and Standard Methods for the Examination of Water and Wastewater (Section 3.3). Where applicable the definition is followed by the appropriate reference, SW-846, EMMC or SM, respectively. 2.2 Accuracy the nearness of a result or the mean of a set of results to the true value. Accuracy is assessed by means of reference samples and percent recoveries (SW-846). 2.3 Analytical Batch the basic unit for analytical quality control is the analytical batch. The analytical batch is defined as samples which are analyzed together with the same method sequence and the same lots of reagents and with the manipulations common to each sample within the same time period or in continuous sequential time periods. Samples in each batch should be of similar composition. (SW-846). 2.4 Blank an artificial sample designed to monitor the introduction of artifacts into the process. For aqueous samples, reagent water is used as a blank matrix; however, a universal blank matrix does not exist for solid samples, and therefore, no matrix is used. A-3 Draft, March 1997

The blank is taken through the appropriate steps of the process. See also Reagent Blank, and Field Blank (SW-846). 2.5 Calibration Check verification of the ratio of instrument response to analyte amount, a calibration check is done by analyzing for analyte standards in an appropriate solvent. Calibration check solutions are made from a stock solution which is different from the stock used to prepare standards (SW-846). 2.6 Control Charts three types of control charts commonly are used in laboratories. Means charts for standards, a means chart for background or reagent blank results and a range chart for replicate analyses. These charts are essential instruments for quality control. See warning and control limits (SM). 2.7 Control or Action Limit plus or minus three times the standard deviation of the mean of a series of standards. These values are derived from stated values for standard reference materials, or replicate analyses of laboratory control standards or calibration check standards. If one measurement exceeds a control limit, repeat the analysis immediately. If the repeat is within the control limit, continue analyses; if it exceeds the control limit, discontinue analyses and correct the problem. See warning limits (SM). 2.8 Environmental or Field Sample environmental or field samples are representative samples of any material (aqueous, non-aqueous, or multimedia) collected from any source for which determination of composition or contamination is requested or required (SW-846). 2.9 Field Blank an aliquot of analyte-free water or solvents brought to the field in sealed containers and transported back to the laboratory with the sample containers. Trip blanks and equipment blanks are two specific types of field blanks. Trip blanks are not opened in the field. They are a check on sample contamination originating from sample transport, shipping and site conditions. Equipment blanks are opened in the field and the contents are poured appropriately over or through the sample collection device, collected in a sample container, and returned to the laboratory as a sample. Equipment blanks are a check on sampling device cleanliness (SW-846). 2.10 Initial precision and recovery (IPR) four aliquots of the diluted PAR analyzed to establish the ability to generate acceptable precision and accuracy (EMMC). 2.11 Laboratory control sample (LCS) see ongoing precision and recovery standard (OPAR). 2.12 Matrix spike (MS) and matrix spike duplicate (MSD) aliquots of an environmental sample to which known quantities of the analytes are added in the laboratory. The MS and MSD are prepared and/or analyzed exactly like a field sample. Their purpose is to quantify any additional bias and imprecision caused by the sample matrix. The background concentrations of the analytes in the sample matrix must be determined in a separate aliquot and the measured values in the MS and MSD corrected for background concentrations. Percent recoveries are calculated after background subtraction for each of A-4 Draft, March 1997

the analytes detected. The relative percent difference between the samples is calculated and used to assess analytical precision (EMMC). 2.13 May this action, activity or procedural step is neither required or prohibited (EMMC). 2.14 Method detection limit (MDL) the lowest level at which an analyte can be detected with 99% confidence that the analyte concentration is greater than zero (EMMC). 2.15 Method quantification limit (MQL) is the minimum concentration of a substance that can be measured and reported (SW-846). 2.16 Minimum level (ML) the lowest level at which the entire analytical system gives a recognizable signal and acceptable calibration point for the analyte. It is equivalent to the concentration of the lowest calibration standard, assuming that all method-specified sample weights, volumes and clean-up procedures have been employed (EMMC). 2.17 Must this action, activity or procedural step is required (EMMC). 2.18 Must Not this action, activity or procedural step is prohibited (EMMC). 2.19 Ongoing precision and recovery (OPR) also called a laboratory control sample or check sample. A laboratory blank spiked with known quantities of analytes. The OPR is analyzed exactly like a sample. Its purpose is to assure that the results produced by the laboratory remain within the limits specified in the method for precision and accuracy (EMMC). 2.20 Precision precision is the measurement of agreement of a set of replicate results among themselves without assumption of any prior information as to the true result. Precision is assessed by means of duplicate/replicate sample analysis (SW-846). 2.21 Precision and recovery standard (PAR) a secondary standard that is diluted and spiked to form the IPR and OPR (EMMC). 2.22 Practical quantitation limit (PQL) is the lowest level that can be reliably achieved within specified limits of precision and accuracy during routine laboratory operating conditions (SW-846). 2.23 Quality control sample (QCS) a sample containing analytes of interest at known concentrations. The QCS is obtained from a source external to the laboratory or is prepared from standards obtained from a different source than the calibration standards. The purpose is to check laboratory performance using test materials that have been prepared independently from the normal preparation process (EMMC). 2.24 Reagent Blank the reagent blank is an aliquot of analyte-free water or solvent analyzed with the analytical batch. Reagent, analyte-free, or laboratory pure water, means distilled A-5 Draft, March 1997

or deionized water or Type II reagent water which is free of contaminants that may interfere with the analytical test in question (SW-846). 2.25 Reagent Grade analytical reagent (AR) grade, ACS reagent grade and reagent grade are synonymous terms for reagents which conform to the current specifications of the Committee on Analytical Reagents of the American Chemical Society (SW-846). 2.26 Relative response factor (RRF) the relative response (peak area or height) of an internal standard of known concentration to the response of the analyte of interest. 2.27 Replicate Sample replicate samples are samples prepared by dividing a sample into two or more separate aliquots. Duplicate samples are considered to be two replicates. Field duplicates are two environmental samples that are collected from the same sample location, at the same time using the same sampling equipment etc. and are a measure of the precision of field sampling equipment and procedures. Laboratory duplicates are samples prepared by taking two separate aliquots of the same sample to measure the precision of the analytical testing procedure and analysis (SW-846). 2.28 Response factor (RF) the ratio of the instrument response, peak height or area responses, to the concentration of analyte injected. 2.29 Standard Curve a curve which plots concentrations of known analyte standard versus the instrument response to the analyte (SW-846). 2.30 Surrogate Spike Recovery Standard surrogates are organic compounds which are similar to analytes of interest in chemical composition, extraction, and chromatography, but which are not normally found in environmental samples. These compounds are spiked into all blanks, standards, samples and spiked samples prior to analysis. Percent recoveries are calculated for each surrogate (SW-846). 2.31 Warning Limits plus or minus two times the standard deviation of the mean of a series of standards. These values are derived from stated values for standard reference materials, or replicate analyses of laboratory control standards or calibration check standards. If two out of three successive points exceeds a warning limit, analyze another sample. If the next point is less than the warning limit, continue analyses; otherwise, discontinue analyses and correct the problem. See control limits (SM). 3.0 References 3.1 Test Methods for Evaluating Solid Wastes, EPA Office of Solid Waste and Emergency Response, Washington, DC., November 1986, Third edition. 3.2 Guidelines and Format for Methods to Be Proposed at 40 CFR Part 136 or Part 141 EPA, Office of Water Engineering and Analysis Division, EPA-821-B-96-003, Washington, DC, July 1996. A-6 Draft, March 1997

3.3 Standard Methods for the Examination of Water and Wastewater, 19 th edition, American Public Health Association, Washington, DC. 1995 (Standard Method). A-7 Draft, March 1997