British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

Similar documents
Method - Determination of aromatic amines in mainstream cigarette smoke

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

CORESTA Recommended Method No. 83

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Sulfotepp impurities in Chlorpyrifos EC formulations

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

DIQUAT DIBROMIDE. The Determination of Ethylene Dibromide in Diquat Dibromide and Diquat Dibromide / Paraquat Dichloride SL Formulations

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Chemistry Gas Chromatography: Separation of Volatile Organics

STANDARD OPERATING PROCEDURES

Technical Report. Determination of Nitrite and Nitrate in Smokeless Tobacco Products by Ion Chromatography and Continuous Flow Analysis

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Technical Procedure for Solid Phase Extraction of THC and THC-COOH for GC-MS Analysis

BUTYL ALCOHOL in urine by GC-FID Code GC05510

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

ENVIRONMENTAL analysis

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

Determination of Polymer Modifier in Asphalt

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

TEST METHOD. Ethoxyl Substitution in ETHOCEL TM Ethylcellulose Polymers. 1. Scope. 2. Principle. 3. Safety

METHYLISOBUTYLKETONE (4-methyl-2 pentanone) IN URINE BY GC/MS Headspace Code GC10510

NORBUPRENORPHINE (Buprenorphine s Metabolite ) BUPRENORPHINE in urine by GC/MS Code GC Method of Confirmation by GC-MS

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

METHANOLYSIS OF ACETAL

METHADONE and EDDP in urine by GC-MS Code GC Method of Confirmation by GC-MS

TETRACHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC15010

Certified Reference Material - Certificate of Analysis

H 3 CO H 3 CO S CH 3

637. Thiamethoxam. HPLC method

TRICHLOROETHYLENE IN URINE BY GC/MS Headspace Code GC16110

METHOD: 1403, Issue 3 EVALUATION: FULL Issue 1: 15 August 1990 Issue 3: 15 March 2003

Certified Reference Material - Certificate of Analysis Methylone, Primary Standard

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

FORMIC ACID in URINE by GC/MS-headspace - Code GC74010

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson

CORESTA Recommended Method No. 86

ACETONE IN URINE BY GC/MS Headspace Code GC10110

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Determination and quantification by GC-MS of melamine (1 mg/kg), ammeline and cyanuric acid (20 mg/kg)

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

METHYLETHYLKETONE (M.E.K.) IN URINE BY GC/MS in head space Code GC10010

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Uncontrolled Copy. SOP Ethylene Glycol Quantitation by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

The analysis of organic acid content of additives, premix, feed, and water.

Volatile Organic Compounds in Every Day Food

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

NCASI METHOD CI/SG/PULP-94.02

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF THE

Report AFK0242/18 TABLE OF CONTENTS

Validation of TNCO, TSNAs, and PAHs

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS

The Focus Robotic Sample Processor as a Tool for the Multiple Analysis of Samples using Complementary Techniques

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Acknowledgements. For more information about this method, contact:

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Spectrophotometric Determination of Ferrocyanide in Effluents

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Certified Reference Material - Certificate of Analysis

ANALYTICAL METHOD PROCEDURES

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Determination of Volatile Substances Proof of Food Adulteration

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

STANDARD OPERATING PROCEDURES

METHYLTIN CHLORIDES 5526

Application Note. Abstract. Authors. Environmental

JWH-018 and JWH-022 as Combustion Products of AM2201

BRIEFING. (EM2: K. Moore.) RTS C Add the following: Methylcellulose

HYDROCARBONS, AROMATIC 1501

Spring 2010 updated March 24 Determination of Aromatics in Gasoline by Gas Chromatography- Mass Spectrometry: Comparison of Grades and Brands

GC-MS Screen for the Presence of Melamine, Ammeline, Ammelide and Cyanuric Acid

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

PYRIPROXYFEN TECHNICAL

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

Analysis of BTEX in Natural Water with SPME

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Soils, sludges and treated bio-waste Organic constituents - LAS by HPLC with fluorescence detection (LC-FLD) and mass selective detection (LC- MSD)

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB/T Translated English of Chinese Standard: GB/T

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Method for the determination of dimethyl sulfate

Analysis of Bakken Oil Samples

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

NCFS Sample Preparation and Instrumental Parameters

Technical Procedure for General Laboratory Equipment

TERTIARY BUTYLHYDROQUINONE

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

Transcription:

British American Tobacco Group Research & Development Method - Determination of phenols in mainstream cigarette smoke 1 SCOPE OF APPLICATION The method is applicable to the quantitative determination of the yields of phenol, o-cresol, m- cresol, p-cresol, catechol, resorcinol and hydroquinone in mainstream cigarette smoke, using gas chromatography with mass selective detection. 2 NORMATIVE REFERENCES ISO 3308:2000 Cigarettes - Routine analytical cigarette smoking machine definitions and standard conditions ISO 3402:1999 ISO 4387:2000 ISO 8243:2006 Tobacco and tobacco products atmospheres for conditioning and testing Cigarettes - Determination of total and nicotine-free dry particulate matter using a routine analytical smoking machine Cigarettes - Sampling 3 PRINCIPLE Five conditioned cigarettes are smoked using a 20 port rotary Borgwaldt smoking machine. The mainstream smoke is collected on to a 44mm Cambridge filter pad (CFP). After smoking, the CFP is extracted with tert-butyl methyl ether (TBME) containing the internal standard, o-chlorophenol. The extract is derivatised with bis(trimethylsilyl)trifluoroacetamide (BSTFA) and analysed by GC/MS using Selective Ion Monitoring (SIM). 4 HEALTH & SAFETY Read and understand the Material Safety Data Sheets for the chemicals used in this method. Read and understand the method risk assessment. Ensure that you understand the hazards and follow control measures relevant to the operation of this method. All preparation of standards and extraction of samples must be performed in a fume cupboard. 5 REAGENTS AND MATERIALS All reagents are Analytical Grade or equivalent unless otherwise stated. N,O-bis(trimethylsilyl)trifluoroacetamide (BSTFA) Tert-butyl methyl ether (TBME) (SLR grade) o-chlorophenol Phenol o-cresol m-cresol p-cresol Catechol Resorcinol Hydroquinone Page 1 of 9

6 APPARATUS 20 port rotary Borgwaldt smoking machine Soap bubble manometer to measure puff volume Analytical balance capable of measuring to at least four decimal places 44mm Cambridge filter pads Wrist action shaker Heating block capable of holding 2mL GC vials and incubating to 76 C Certified thermometer Agilent GC/MS with autosampler 2mL amber GC crimp top vials with Teflon/Silicone caps 5L dispenser bottle 5L volumetric flask (class A) 100mL amber volumetric flask 0.5mL pipette (class A) 1mL pipette (class A) 2mL pipette (class A) 3mL pipette (class A) 50mL amber volumetric flasks (class A) with stoppers multidispenser pipette capable of dispensing 250µL and 500µL 500µL gas tight syringe J&W DB5 MS 60m x 0.25mm i.d x 0.25µm Column 5m of 0.25mm deactivated retention gap 7 PRELIMINARY SAMPLE PREPARATION Cigarettes should be conditioned according to normal procedures (ISO 3402:1999). Unless specifically requested, cigarettes are not subjected to any selection criteria other than the rejection of any obviously defective or damaged cigarettes. Butt marking is as specified in ISO 4387:2000 unless otherwise requested. 8 ANALYTICAL PROCEDURE SOLUTION PREPARATION 8.1 Extracting Solution Weigh 80mg (±1mg) of o-chlorophenol and transfer to a 5L volumetric flask (class A) and make to volume with TBME. Transfer to a 5L dispenser bottle. 8.2 Stock Solution To a 100mL amber volumetric flask (class A) add the following amounts of each compound and make to volume with extracting solution (8.1). Target Weight (mg) Phenol 40 Page 2 of 9

o-cresol 10 m-cresol 10 p-cresol 10 Catechol 100 Resorcinol 10 Hydroquinone 80 8.3 Calibration s Dilute the Stock Solution (8.2) as follows in a 50mL amber volumetric flask (class A) and make to volume with extracting solution. Calibration Volume (ml) of Stock Solution Final Volume (ml) 1 0.10 50 2 0.25 50 3 0.50 50 4 1.00 50 5 1.50 50 6 2.50 50 7 3.50 50 The calibration standards will contain the following concentrations of each compound: 1 2 3 Concentration (μg/ml) 4 5 6 7 Phenol 0.8 2.0 4.0 8.0 12.0 20.0 28.0 o-cresol 0.2 0.5 1.0 2.0 3.0 5.0 7.0 m-cresol 0.2 0.5 1.0 2.0 3.0 5.0 7.0 p-cresol 0.2 0.5 1.0 2.0 3.0 5.0 7.0 Catechol 2.0 5.0 10.0 20.0 30.0 50.0 70.0 Resorcinol 0.2 0.5 1.0 2.0 3.0 5.0 7.0 Hydroquinone 1.6 4.0 8.0 16.0 24.0 40.0 56.0 Page 3 of 9

8.4 QC Stock Solution To a 100mL amber volumetric flask (class A) add the following amounts of each compound and make to volume with extracting solution (8.1). Target Weight (mg) Phenol 40 o-cresol 10 m-cresol 10 p-cresol 10 Catechol 100 Resorcinol 10 Hydroquinone 80 8.5 QC Dilute 1.0mL of the QC Stock Solution (8.4) with extracting solution in a 50mL amber volumetric flask (class A). The QC standard will contain the following concentrations: QC concentration (μg/ml) Phenol 8 o-cresol 2 m-cresol 2 p-cresol 2 Catechol 20 Resorcinol 2 Hydroquinone 16 All standards are stored in a freezer until required and defrosted thoroughly prior to use. Expiry date: 3 months from date of preparation. 9 ANALYTICAL PROCEDURE SAMPLE PREPARATION 9.1 Sample Collection Cigarettes are smoked on a Borgwaldt Rotary 20 port smoking machine. Warm-up the smoking machine for 20 minutes before smoking. Check the linear airflow is 200 mm/s ±30 mm/s, the system has no leaks and puff volume is 35mL (± 0.3 ml) (for ISO smoking). Page 4 of 9

5 cigarettes are loaded and smoked. Record the number of lit puffs and the weight of Total Particulate Matter (TPM). 9.2 Sample Extraction Place the CFP into a 50mL conical flask and stopper immediately. The CFP holder is wiped with two clean CFP quarters and these are added to the flask. To the flask add 10mL of extracting solution using the preset dispenser, seal with parafilm and shake for 20 minutes (180 revs/minute). 9.3 Sample Clean Up 9.3.1 Dispense a 250μL aliquot of the sample extract or standard followed by 250μL of BSTFA and 500μL of TBME into a 2mL amber autosampler vial. 9.3.2 Cap the vial using Teflon/Silicone 11mm crimp caps and invert to mix. Place the vial in a preheated heating block for 30 minutes at 76 C (±5 C). The sample is ready for analysis by GC/MS. 10 ANALYTICAL PROCEDURE INSTRUMENTAL ANALYSIS 10.1 Instrument Set Up Parameters Analysis is performed on an Agilent 6890 Gas Chromatograph (GC) fitted with autosampler and 5973 Mass Selective Detector (MSD). Column Type Injection type and temperature Splitless / 250 C Column temperature programme Carrier gas J&W DB5 MS 60m x 0.25mm i.d x 0.25µm, preceded by 5m of 0.25mm deactivated retention gap Initial 60 C hold for 6 minutes Ramp 10 C/min to 180 C 15 C/min to 300 C;hold for 2 minutes Total run time is 28.0 minutes Helium Transfer line temperature 280 C Injection Volume 1µL Column Flow Solvent Delay 1mL/min (constant flow) 14 minutes MS Source temperature 230 C MS Quadrupole temperature 150 C MS Mode Ion Dwell time SIM 100ms The following ions are used as target and qualifier ions Page 5 of 9

Group m/z Approximate Retention time (minutes) Phenol 1 166*(100%) 151(350-380%), 152 (15-25%) o-cresol 2 180*(100%), 165(140-160%), 91(110-130%) m-cresol 2 165*(100%), 180(25-40%), 91(10 20%) p-cresol 2 180*(100%), 165(240-270%), 91(45 60%) o-chlorophenol 3 200*(100%), 185(300-340%), 149(240-265%), 93(260-310%) Catechol 4 254*(100%), 239(25-35%), 73(440 520%) Resorcinol 4 254*(100%), 239(135-155%), 73(75 110%) Hydroquinone 4 254*(100%), 239(120-140%), 73(55 75%) 15.4 16.8 16.9 17.1 18.0 19.6 20.4 20.6 * quantitation ion s and samples are quantified on the target ions stated above. The qualifier ion ratios are used to confirm that the peaks in the standards are correctly identified. In addition, the ion ratios of the sample peaks should be within ±20% of the standard ion ratios in that run. 10.2 System Suitability Criteria 10.2.1 MS Tuning Tune the MS prior to every run. Check the following criteria are met on the tune report: Air and water peaks <10% EM Volts 1000 3000 Ion ratios of m/z 69:219:502 are approximately 10:4:1 Peak width approximately 0.6+10% 10.2.2 Peak Resolution Check Open a chromatogram for the control cigarette smoke extract, and use the instrument software to integrate the m-cresol peak and the o-cresol peak. The valley should be >90% of the m-cresol peak height. If the valley is 90%, investigate the problem before further analysis takes place. Record the value in the maintenance log. Page 6 of 9

10.2.3 Peak Shape Check Open a chromatogram of calibration standard 1, and use the instrument software to assess the phenol peak shape. The value for tailing, as defined in the instrument software, should be <3.26. If the result is >3.26 the problem should be investigated before further analysis. 10.2.4 Ion Ratio Check Check the ion ratios are within the limits shown in section 10.1. 10.2.5 Calibration Linearity The R 2 value of the calibration graphs must be >0.99. 10.3 Run Order Start with two conditioning samples* Calibration standards in ascending order QC standard 10 samples (including the reference cigarette sample) QC standard 10 samples (including the reference cigarette sample) etc Calibration standards in ascending order *NB: These conditioning samples should be underivatised smoke extracts. 11 CALCULATIONS Using the instrument software, plot calibration graphs of calibration standards concentration against peak area ratio, without forcing the line through zero. Peak area ratio = compound peak area/o-chlorophenol peak area Check the plots, coefficient of determination (R 2 ) and intercept before accepting the calibrations. Calculate the concentration of the phenolic compounds in the sample solutions. The results obtained from the GC/MS are in μg/ml. To convert results to μg/cigarette, use the following equation: Individual compound concentration (μg/cigarette) = C x V N Where: C = Individual compound concentration in extract (μg/ml) V = Volume of extracting solution (usually 10mL) N = Number of cigarettes smoked (usually 5) Page 7 of 9

12 PRECISION AND REPORTING LIMITS Five replicate smokings and analyses are performed to determine the precision of the analysis. Longer-term precision is monitored through the maintenance of control charts. The method quantitation limits are defined as ten times the standard deviation of the lowest calibration standard analysed ten times. The practical lower reporting limits are defined by the concentration of the lowest calibration standard and are as follows: Quantitation Limit (μg/cigarette) Lower Reporting Limit (μg/cigarette) Phenol 0.11 1.6 o-cresol 0.03 0.4 m-cresol 0.03 0.4 p-cresol 0.07 0.4 Catechol 0.24 4.0 Resorcinol 0.06 0.4 Hydroquinone 0.19 3.2 13 QUALITY ASSURANCE AND CONTROL Control charts of the QC standard and the reference cigarette are maintained to allow inspection of the method performance. 14 SPECIAL CASES Under more intense smoking regimes, the number of cigarettes per smoking run may need to be reduced in order to avoid smoke breakthrough on the Cambridge filter pad. Date of Issue: 31 March 2008 Page 8 of 9

APPENDIX A SAMPLE CHROMATOGRAMS Calibration - Total ion chromatogram Smoke Extract - Total ion chromatogram Page 9 of 9