Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9000 Gas Chromatograph

Similar documents
Semivolatile Organics Analysis Using an Agilent J&W HP-5ms Ultra Inert Capillary GC Column

Faster Semivolatiles Analysis with a Scaled- Down Method and GC Accelerator Kit

The Role of the Carbon Cartridge in the One-Pass System

Meeting Challenging Laboratory Requirements for USEPA Method 8270 Using a Highly Sensitive, Robust, and Easy-to-Use GC/MS

Developing Large Volume Injection (LVI) in Split / Splitless Inlets. Philip J. Koerner Phenomenex Inc. NEMC 2014

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

POTENTIAL VAPOR INTRUSION ENVIRONMENTAL

Amec Foster Wheeler Environment & Infrastructure, Inc BIG SHANTY ROAD, NW, SUITE 100 KENNESAW, GEORGIA (770)

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Appendix 2 Data Distributions for Contaminants in Water and Sediment Samples in Response to the Deepwater Horizon Oil Spill, 2010

STANDARD OPERATING PROCEDURES SOP: 1821 PAGE: 1 of 33 REV: 0.0 DATE: 11/25/94

2. September 2015 Sub-Concrete Slab Soil Investigation

STANDARD OPERATING PROCEDURES

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Multi-residue Analysis for PAHs, PCBs and OCPs on Agilent J&W FactorFour VF-Xms

October 26, Dear Mr. White:

Non-Potable Water A2LA Certificate

Semivolatile Organic Compounds By Gas Chromatography / Mass Spectrometry (GC/MS)

Semivolatile Organic Reference Materials

USEPA Methods 8270 and 8260 on a Single GCMS Without Changing Columns

Certificate of Accreditation

317 Elm Street Milford, NH (603) Fax (603)

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Certificate of Accreditation

Certificate of Accreditation

Appendix C. Validation Standard Operating Procedures

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Table 1 Soil and Groundwater Sampling Summary Table

Certificate of Accreditation

Methods and Analytes Requiring Laboratory Certification Attachment 3

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Ultra-Inert chemistry for Trace Level Analysis

Table of Contents ~ 1 ~

DUAL CARTRIDGE SYSTEM FOR THE EXTRACTION OF ACIDS, BASES, AND NEUTRALS IN WATER

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Author. Abstract. Introduction

MULTIRESIDUE ANALYSIS OF PRIORITY POLLUTANTS IN SURFACE WATERS USING EXACT MASS GC-TOF MS

STANDARD OPERATING PROCEDURES

ENVIRONMENTAL analysis

Pittcon P

Rely on Rxi -PAH Columns

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Certificate of Accreditation

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Method 8270C PAH. FD- Field Duplicate Samples

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Selection of a Capillary

Presentation Title. Isotope Dilution: Where it doesn t work Where we should expand its use. August 10, David I. Thal Principal Chemist

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Facility: Mansfield Revision 3. Title: PAHs by SPME Page 1 of 11

Selection of a Capillary GC Column

Analysis of Organic Pollutants by Micro Scale Liquid-Liquid Extraction. and On-column Large Volume Injection Gas Chromatography. Mark S.

The Suite for Environmental GC Analysis

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

The Design of High Temperature Capillary Gas Chromatography Columns Based on Polydimethylsiloxane

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Katherine K. Stenerson, Michael Ye, Michael Halpenny, Olga Shimelis, and Leonard M. Sidisky. Supelco, Div. of Sigma-Aldrich Bellefonte, PA USA

METHOD 8410 GAS CHROMATOGRAPHY/FOURIER TRANSFORM INFRARED (GC/FT-IR) SPECTROMETRY FOR SEMIVOLATILE ORGANICS: CAPILLARY COLUMN

SUBSURFACE (PHASE II) INVESTIGATION REPORT

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Application Note. Abstract. Authors. Environmental

Determination of Polycyclic Aromatic Hydrocarbons (PAHs) in Seafood Using GC/MS

Determination of 24 PAHs in Drinking Water

Polycyclic Aromatic Hydrocarbons in Water by GC/MS PBM

Residual solvents analysis using an Agilent Intuvo 9000 GC system

STANDARD OPERATING PROCEDURES

Page 1 of 7 ADEQ Water Div NPDES 9/18/2017 8:52 AM

LABORATORY DATA CONSULTANTS, INC Loker Ave. West, Suite 220, Carlsbad, CA Bus: Fax:

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

STANDARD OPERATING PROCEDURES SOP: 1810 PAGE: 1 of 31 REV: 0.0 DATE:03/14/05 ANALYSIS OF POLYNUCLEAR AROMATIC HYDROCARBONS (PAHs) IN DUST BY GC/MS-SIM

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Uncontrolled Copy. SOP-060 Chromatography, Tune Evaluation and Troubleshooting. Table of Contents. 1. Principle... 3

TETRA TECH, INC. December 8, Lynn Nakashima Berkeley Regional Office 700 Heinz Avenue, Suite 200C Berkeley, California 94710

Understanding Gas Chromatography

Analysis. Priority LC GC. Zebron PAH Accurately quantitate EU and EPA PAHs in less than 28 minutes

This project has been funded with support from the European Commission. This publication reflects the views only of the authors, and the Commission

March 6, Mr. John Meyer DRI Coordinator Tampa Bay Regional Planning Council 4000 Gateway Center Blvd. Pinellas Park, FL 33782

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to Environmental Aqueous Samples

Results of a Sediment Survey in the Near Offshore Waters of the Proposed Quarry Site in the Vicinity of Whites Cove, Digby Neck, Nova Scotia

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

Methanol Extraction of high level soil samples by USEPA Method 8260C

A Single-Method Approach for the Analysis of Volatile and Semivolatile Organic Compounds in Air Using Thermal Desorption Coupled with GC MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Sample Pre-Treatment : Introduction to Commonly Used Techniques for Gas

APPENDIX D Laboratory Analytical Reports

ALLOWAY STANDARD OPERATING PROCEDURES FOR METHOD 525.2

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

NEW Agilent J&W Ultra Inert Capillary GC Columns. Proven performance that raises the bar on. consistent. column inertness.

What Does the EPA Method Round Robin Data Really Tell Us?

~WJuliJ LABORATORY DATA CONSULTANTS, INC.

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

DETERMINATION OF PHENOLS IN DRINKING WATER BY SOLID PHASE EXTRACTION AND CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY (GC/MS) Revision 1.

U.S. EPA Method 8270 for multicomponent analyte determination

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Transcription:

Analysis of Semivolatile Organic Compounds Using the Agilent Intuvo 9 Gas Chromatograph Application Note Author Matthew Giardina, Ph.D., Applications Chemist, Gas Phase Separations Division, Agilent Technologies, Inc. Abstract Gas chromatography/mass spectrometry is a workhorse for the analysis of semivolatile organic compounds in environmental samples. The Agilent Intuvo 9 GC represents a significant advancement in gas chromatography, and provides a number of advantages for semivolatile analysis. This Application Note demonstrates the performance of the Intuvo 9 GC for the analysis of semivolatiles based upon the specifications of USEPA method 827D. Included in this study is an evaluation of the Intuvo during repetitive injections of a composite soil extract to gauge instrument durability under conditions modeling a high throughput environmental laboratory.

Introduction Gas chromatography/mass spectrometry (GC/MS) is widely regarded as the analytical technique of choice for the analysis of semivolatile organic compounds (SVOCs) 1. A number of compounds within the class of SVOCs are considered environmental pollutants. Government agencies with regulatory authority have established methods and set performance criteria for the measurement SVOCs in a number of environmental and industrial matrices. For example, United States Environmental Protection Agency (USEPA) method 827D contains a list of 243 compounds that are suitable for analysis by GC/MS in solid waste, soil, air and water extracts 2. Method 827D contains detailed specifications and requirements for the quantitative analysis of SVOCs. The durability and data quality from the Agilent 789 GC has set the standard for the analysis of SVOCs and these instruments are used in environmental testing laboratories throughout the world. With the development of new technology, it is of critical importance to demonstrate that the next generation of GC instrumentation can meet the rigors of SVOC analysis for environmental testing. The Agilent Intuvo 9 GC is a step forward in the advancement of gas chromatography, and includes a number of design innovations making it ideally suited for SVOC analysis 3. Intuvo Flow Technology (IFT) incorporates the use of an easy to install and replace Intuvo Column and Guard Chip which acts as a precolumn to prevent particulate and nonvolatile contamination of the column and flow path 4. Direct heating technology reduces power requirements, and facilitates more rapid column cooling for faster cycle time. This Application Note demonstrates that the Intuvo 9 GC can easily achieve the rigorous performance specifications established in USEPA method 827D, while maintaining equivalent retention time, resolution, and analyte response of the benchmark 789 GC. In addition, the system durability is demonstrated by the repetitive injection of a composite soil extract. 2

Experimental Table 1. Target compounds, surrogates and internal standards. Standards and samples A stock standard containing 77 target compounds and surrogates was purchased from AccuStandard (New Haven, CT). The standard was selected to provide a representative mixture of acids, bases, and neutrals. The stock standard was diluted in dichloromethane containing six internal standards purchased from Supelco (Bellefonte, PA). The nominal concentration of targets and surrogates in the calibration standards were.1,.2,.8, 1, 1.6, 2, 5, 1, 15, 2, 35, 5, 75, and 1 µg/ml. The concentration of the internal standards in each calibration standard was 4 µg/ml. Table 1 lists the compounds used in the study. The compound numbers in Table 1 were assigned based upon retention order of the targets and surrogates with the internal standards listed at the end of the table out of retention order. Compound numbers were assigned to reduce complexity of the graphs. The tuning standard, containing a mixture of benzidine, pentachlorophenol, 4,4 -dichlorodiphenyltrichloroethane (4,4 DDT), and decafluorotriphenylphosphine (DFTPP), was purchased from AccuStandard. The tuning standard was diluted in dichloromethane to a final concentration 25 µg/ml. A composite mixture of soils extracted with dichloromethane prepared for method 827 was donated from ESC Lab Sciences (Mt. Juliet, TN). The extracts selected for the composite mixture contained the heaviest matrix residue typically encountered in their laboratory. No. Compound 1 N-Nitrosodimethylamine 2 Pyridine 3 2-Fluorophenol (surrogate) 4 Phenol-d 5 (surrogate) 5 Phenol 6 Aniline 7 Bis(2-chloroethyl) ether 8 2-Chlorophenol 9 1,3-Dichlorobenzene 1 1,4-Dichlorobenzene 11 Benzyl alcohol 12 1,2-Dichlorobenzene 13 2-Methylphenol 14 Bis(2-chloroisopropyl) ether 15 4-Methylphenol 16 N-Nitrosodi-n-propylamine 17 Hexachloroethane 18 Nitrobenzene-d 5 (surrogate) 19 Nitrobenzene 2 Isophorone 21 2-Nitrophenol 22 2,4-Dimethylphenol 23 Benzoic acid 24 Bis(2-chloroethoxy)methane 25 2,4-Dichlorophenol 26 1,2,4-Trichlorobenzene 27 Naphthalene 28 4-Chloroaniline 29 Hexachlorobutadiene 3 4-Chloro-3-methylphenol 31 2-Methylnaphthalene 32 Hexachlorocyclopentadiene 33 2,4,6-Trichlorophenol 34 2,4,5-Trichlorophenol 35 2-Fluorobiphenyl (surrogate) 36 2-Chloronaphthalene 37 2-Nitroaniline 38 Dimethyl phthalate 39 2,6-Dinitrotoluene 4 Acenaphthylene 41 3-Nitroaniline 42 Acenaphthene 43 2,4-Dinitrophenol No. Compound 44 4-Nitrophenol 45 2,4-Dinitrotoluene 46 Dibenzofuran 47 Diethyl phthalate 48 4-Chlorophenyl-phenyl ether 49 Fluorene 5 4-Nitroaniline 51 4,6-Dinitro-2-methylphenol 52 N-Nitrosodiphenylamine 53 Azobenzene 54 2,4,6-Tribromophenol (surrogate) 55 4-Bromophenyl phenyl ether 56 Hexachlorobenzene 57 Pentachlorophenol 58 Phenanthrene 59 Anthracene 6 Carbazole 61 Di-n-butylphthalate 62 Fluoranthene 63 Benzidine 64 Pyrene 65 p-terphenyl-d 14 66 Butylbenzylphthalate 67 3,3 -Dichlorobenzidine 68 Benzo[a]anthracene 69 Bis(2-ethylhexyl)phthalate 7 Chrysene 71 Di-n-octyl phthalate 72 Benzo[b]fluoranthene 73 Benzo[k]fluoranthene 74 Benzo[a]pyrene 75 Indeno[1,2,3-cd]pyrene 76 Dibenzo[a,h]anthracene 77 Benzo[g,h,i]perylene 78 1,4-Dichlorobenzene-d 4 (internal standard) 79 Naphthalene-d 8 (internal standard) 8 Acenaphthalene-d 1 (internal standard) 81 Phenanthrene-d 1 (internal standard) 82 Chrysene-d 12 (internal standard) 83 Perylene-d 12 (internal standard) 3

Instrument methods The Agilent Intuvo 9 GC was configured with a single MS flow path for interfacing to an Agilent 5977B MSD with an inert EI ion source and a 3 m Intuvo DB-5ms Ultra Inert column. For comparative data, an Agilent 789 GC was also interfaced to a 5977B MSD with an inert EI ion source and a 3 m Agilent J&W DB-5ms Ultra Inert column. The same method parameters were used on both the Intuvo 9 GC and the 789 GC, with the exception of the Intuvo Guard Chip, which was not present on the 789 GC. Table 2 lists the instrument conditions used during the study. Results and Discussion Equivalency to an Agilent 798 GC In addition to the many innovations in flow path and column heating technology that have been incorporated into the Intuvo 9 GC, it has been designed to provide equivalent performance to the 789 GC, with respect to chromatographic retention and response. This is particularly important for the analysis of SVOCs in environmental testing where the 789 GC has set the performance benchmark. As an initial validation of equivalency, a standard prepared with a target and surrogate concentration of 2 µg/ml, and an internal standard concentration of 4 µg/ml was injected on a 789 GC coupled to a 5977B MSD and a 3 m Agilent J&W DB-5ms Ultra Inert column, and an Intuvo 9 GC coupled to a 5977B MSD and a 3 m Intuvo DB-5ms Ultra Inert column. The same column temperature program and detector conditions were used for analysis (Table 2). Figure 1 shows a normalized total ion chromatogram obtained on both systems. The chromatograms are virtually indistinguishable with only slight differences in regions of high peak density (12.5 and 16.5 minutes). Table 2. Common GC/MS Parameters. Normalized response Normalized response Parameter Value Injection volume 1 µl Inlet Split/Splitless 3 C Pulsed splitless 6 psi to.5 minutes Purge 5 ml/min at.5 minutes Septum purge switched flow mode 3 ml/min Liner Intuvo Guard Chip Agilent Ultra Inert splitless single taper liner with glass wool (p/n 519-2293) 6 C for 2 minutes, 2 C/min to 26 C, 6 C/min to 33 C, Hold 1.333 minutes (standards) or 1.333 minutes (soil extract) Column Agilent J&W DB-5ms and Intuvo DB-5ms UI 3 m.25 mm,.5 µm (p/n 122-5536UI for 789 and 122-5536UI-INT for Intuvo) Flow Column temperature Transfer line temperature 33 C Drawout plate Ion source temperature 33 C Quadrupole temperature 2 C Scan Gain factor 1 Threshold 5 A/D samples 2 1 A 2 ml/min constant flow 4 C for 2 minutes, 2 C/min to 26 C, 6 C/min to 33 C, Hold 1.333 minutes (standards) or 1.333 minutes (soil extract) 6 mm (optional) 35 to 55 m/z Agilent Intuvo 9 GC 3.5 5. 6.5 8. 9.5 11. 12.5 14. 15.5 17. 18.5 2. 21.5 23. 24.5 Time (min) 1 B Agilent 789 GC 3.5 5. 6.5 8. 9.5 11. 12.5 14. 15.5 17. 18.5 2. 21.5 23. 24.5 Time (min) Figure 1. Comparison of SVOC chromatograms generated with an Agilent Intuvo 9 GC (A) and an Agilent 789 GC (B). 4

To provide a more quantitative comparison of retention, Figure 2 shows a plot of the relative retention times (that is, ratio of target and surrogate retention times to the internal standards) for the Intuvo 9 GC and 789 GC. The agreement between the two instruments was quite good. The average difference in relative retention time between the two instruments was.6. Figure 2 shows that the first two eluting compounds, N-nitrosodimethylamine and pyridine, show a greater degree of difference in relative retention time. This is a result of a small amount of precolumn volume introduced by the Intuvo Guard Chip. Figure 3 shows, total ion chromatograms for N-nitrosodimethylamine and pyridine on an expanded axis for the 789 GC and Intuvo 9 GC. As expected, the additional volume causes a slight increase in retention time compared to 789 GC. However, the effect on peak shape was negligible. Method 827D stipulates that for reporting close eluting structural isomers as separate analytes, the valley between the two peaks cannot be greater than 5 % of the average maximum height of the isomers. Benzo[b]fluoranthene and benzo[k]fluoranthene are generally selected as a measure of the system s ability to resolve isomers. Figure 4 shows the resolution achieved on the 789 GC and Intuvo GC. In both cases, the valley between the isomers was well below 5 %. As with many GC methods, quantitation with method 827D is based upon the relative response with respect to the internal standards. Figure 5 shows a comparison of response factors for the 789 GC and Intuvo 9 GC for an injection of the 2 µg/ml standard. Again, agreement was very good. The average difference in response factors across all the target compounds was 4.6 %. Relative retention time 1.2 1.1 1..9.8.7.6 Agilent 789 GC Agilent Intuvo 9 GC.5 1 2 3 4 5 6 7 8 Compound number Figure 2. Comparison of relative retention times for SVOCs on an Agilent Intuvo 9 GC and an Agilent 789 GC. A N-Nitrosodimethylamine Pyridine 3.6 3.7 3.8 3.9 4. 4.1 4.2 4.3 4.4 4.5 Time (min) B N-Nitrosodimethylamine Pyridine Agilent 789 GC Agilent Intuvo 9 GC 3.6 3.7 3.8 3.9 4. 4.1 4.2 4.3 4.4 4.5 Time (min) Figure 3. Comparison of first eluting SVOCs on an Agilent 789 GC (A) and an Agilent Intuvo 9 GC (B). A B Benzo[k]fluoroanthene Benzo[b]fluoranthene 19. 19.1 19.2 19.3 19.4 19.5 19.6 19.7 19.8 Time (min) Benzo[b]fluoranthene Benzo[k]fluoroanthene Agilent 789 GC 19. 19.1 19.2 19.3 19.4 19.5 19.6 19.7 19.8 Time (min) Figure 4. Resolution of isomers on an Agilent 789 GC (A) and an Agilent Intuvo 9 GC (B). Agilent Intuvo 9 GC 5

In Figure 5, note that the relative responses of the last eluting polyaromatic hydrocarbons (PAHs) indeno[1,2,3-cd] pyrene, dibenzo[a,h]anthracene, and benzo[g,h,i]perylene was slightly greater on the Intuvo 9 GC compared to the 789 GC. This demonstrates that the thermal profile across the Intuvo 9 GC flow path is consistent, allowing the higher boiling point PAHs to pass through the flow path, and maintain recovery and peak shape (Figure 1). Method 827D suitability For the analysis of SVOCs by GC/MS, it is of paramount importance to verify the system is suitable for quantitative analysis before the collection of data, particularly when data are used for regulatory reporting. Method 827D specifies the use of a tuning or control standard for verifying performance of the GC/MS. The standard consists of a mixture of DFTTP, 4,4 -DDT, pentachlorophenol, and benzidine to test MS tuning and inertness. The DFTPP is used to verify suitable ionization and detection of the mass spectrometer. The 4,4 -DDT is used to gauge system inertness by determination of the breakdown products 4,4 -DDD and 4,4 -DDE. Benzidine is used as a probe for base activity, and pentachlorophenol is used as a probe for acid activity. If the minimum performance criteria as established by the method cannot be achieved, the system is deemed unsuitable for analysis. Figure 6 shows a chromatogram of the tuning standard at a concentration of 25 µg/ml. Method 827D suggests a concentration of 5 µg/ml with a caveat that lower concentrations may be used to accommodate instruments with greater sensitivity. In this case, 25 µg/ml was selected as not to overload the column, and bias peak symmetry measurements. Relative response 2.5 2. 1.5 1..5 Agilent 789 GC Agilent Intuvo 9 GC 1 2 3 4 5 6 7 8 Compound number Figure 5. Comparison of relative retention times for SVOCs on an Agilent Intuvo 9 GC and an Agilent 789 GC. Normailzed response 1 9 8 7 6 5 4 3 2 1 Pentachlorophenol TF = 1. DFTPP Ion ratios (Table 3) Benzidine TF =.8 Breakdown = 1.4 % 11.5 12. 12.5 13. 13.5 14. 14.5 15. 15.5 Time (min) Figure 6. Chromatogram of a method 827D tuning mix on the Agilent Intuvo 9 GC. Tailing factor (TF) was used as the determinant of acid/base activity of pentachlorophenol and benzidine. Based upon method 827D requirements, the TF measured at 1 % peak height for the extracted quantitation ion should be no greater than 2. For pentachlorophenol and benzidine, the measured tailing factors were 1. and.8, respectively. DDT A percent breakdown of 4,4 -DDT was used to determine system inertness. According to method 827D, the combined area sum of the extracted ions for 4,4 -DDD and 4,4 -DDE should not exceed 2 %. The percent breakdown was measured at 1.4 % on the Intuvo 9 GC. 6

Table 3 lists the measured DFTPP ion ratios along with the specified ratios and ranges for method 827D. All measured ratios are well within the required limits. The Intuvo 9 GC easily passed the system suitability metrics as specified in method 827D. Calibration requirements Calibration is, perhaps, the most challenging requirement to achieve and maintain with method 827D. The target list comprises a range of acidic, basic, and neutral molecules. The type of calibration and calibration range for a selected analyte is largely based upon the sensitivity of the instrument and the nature of the compound. Some compounds are more sensitive to surface activity, thermal conditions, or detection efficiency. Because of this, several methods of calibration are acceptable for quantitation. The simplest and most widely applied calibration is based upon average response factor. According to the method, a minimum of five standard levels must be used, and the relative Table 3. DFTPP tuning check. Target mass Relative to mass Lower limit % Upper limit % Rel. abn % Pass/Fail 51 442 1 8 31.1 Pass 68 69 2 Pass 7 69 2.4 Pass 127 442 1 8 39.4 Pass 197 442 2 Pass 198 442 5 1 84 Pass 199 198 5 9 6.1 Pass 275 442 1 6 22.4 Pass 365 198 1 1 4.2 Pass 441 442 24 15.4 Pass 442 442 1 1 1 Pass 443 442 15 24 18.7 Pass standard deviation (RSD) in response factors should be within ± 2 %. Figure 7 shows the percent RSD achieved for 57 of the 77 compounds over a range of.1 to 1 µg/ml using 14 calibration levels (with the exception of number 23, benzoic acid, with a calibration from 4 to 1 µg/ml over eight levels). A total of 14 calibration levels were used To precisely determine the linear range of each analyte. The average RSD for the 57 compounds was 4.98 %. 3 25 Relative response factor %RSD 2 15 1 Response factor RSD limit 5 1 2 3 4 5 6 8 9 1 11 12 13 14 15 17 18 19 2 21 2223 24 25 26 27 28 29 3 31 3233343536 37 38394 41 42 4546 47 48495 52 535556 66566 67 68 71 74 Compound number Figure 7. Percent RSD in average response factor for calibration up to 1 µg/ml. 7

Certain compounds with higher sensitivity such as PAHs tend to overload the stationary phase and saturate the detector at high concentration. In this case, the upper calibration limit was lowered to 5 µg/ml to achieve the calibration criteria using average response factors. Figure 8 shows the percent RSD for 14 of the 77 compounds over a range of.1 to 5 µg/ml using 12 calibration levels (with the exception of number 69, benz[a]anthracene, with a calibration from.8 to 5 µg/ml over 1 levels). The average RSD for the 14 compounds was 4.1 %. Some compounds that are active or labile tend to have response factors that vary as a function of concentration. For these analytes, method 827D allows curve fitting for calibration. The method stipulates that the correlation coefficient (R) must be greater than.99, and the calculated concentration of the lowest standard must be within ±3 % of the actual concentration. Table 4 lists the calibration results for the remaining six of the 77 compounds using a weighted linear least squares regression with a 1/x weighting factor. In all the cases, the specified calibration criteria were achieved. Note that the calibration range was selected to achieve the widest dynamic range while satisfying the calibration criteria using a linear model. The percent deviation of the lowest level standard would be closer to zero if the dynamic range were narrowed, or a higher order calibration model were used. Relative response factor RSD (%) 3 25 2 15 1 5 Response factor RSD limit 7 16 58 59 61 62 64 69 7 72 73 75 76 77 Figure 8. Percent RSD in average response factor for calibration up to 5 µg/ml. Table 4. Calibration results using weighted least squares regression. Compound no. Compound R 2 range (µg/ml) Calibration 43 2,4-Dinitrophenol.9984 1.6 1 23.9 44 4-Nitrophenol.9994.8 1 22.9 51 4,6-Dinitro-2-methylphenol.9991.8 1-1.3 54 2,4,6-Tribromophenol.9997.8 1 12.4 57 Pentachlorophenol.9992.8 1 23.4 63 Benzidine.9966 4 1 16.5 However, in this study, a test strategy was used where matrix samples were injected until suitability or calibration failure, then returned to performance through corrective maintenance. Figure 9 Initial QC and calibration Inject matrix set (N = 2) Yes Percent difference of lowest level standard (± 3 % Required) shows a flowchart illustrating the testing scheme. The test study was gated by performance checks in between Matrix study To gauge the durability of the Intuvo 9 GC, an iterative cycle of matrix injections and performance checks was carried out. Typically, environmental testing laboratories perform preventative maintenance (for example, liner replacement, column clipping) at regular intervals. This is done to maintain system suitability and calibration integrity over an extended period of time by preempting column and source contamination. Check QC/CCV/ ISTD Check QC/CCV/ ISTD Yes Pass check? No MSD Maintenance No Pass check? Corrective maintenance attempt 1? Yes Change septum and liner No Check QC/CCV/ ISTD Corrective maintenance attempt 2? Yes Replace Guard Chip New column Figure 9. Matrix study testing scheme. 8

2 matrix injections. The check consisted of three measurements related to the specifications listed in method 827D including: QC Correct DFTPP tuning ratios, tailing factor for benzidine and pentchlorphenol less than 2, and percent breakdown for 4,4 -DDT of less than 2 % CCV Midpoint calibration drift is within ± 2 % for greater than 1 % of the target compounds ISTD Verify the area of internal standard peak area drift is within ± 5 % To begin the study, the system was calibrated using the compounds listed in Table 1, and the method 827D parameters listed in Table 2. The matrix sample was supplied by ESC Lab Sciences (Mt. Juliet, TN), and was composed of multiple soil samples extracted with dichloromethane and combined. The composite extract was representative of the heaviest matrix sample typically encountered in their laboratory. Figure 1 shows that the extract was opaque, and contained significant matrix residue. Study results QC results A total of 68 matrix injections were performed over the course of the study. After each interval of 2 sample injections, the matrix load was sufficient to almost cause a complete breakdown of 4,4 -DDT. Following the testing protocol, the liner and septum were replaced, and the system was retested. After the liner replacement, the percent breakdown dropped below 2 % (Figure 11). Figure 12 shows a liner after 2 injections of the soil extract. Clearly visible is a layer of soil residue deposited on the glass wool. This residue was likely the cause of the 4,4 -DDT breakdown since replacement of the liner restored breakdown to below 2 %. The glass wool packing was sufficient to protect the system from particulate contamination. Figure 1. Dichloromethane soil extract 1 After liner replacement After matrix injections 9 8 7 4,4 -DDT % Breakdown 6 5 4 3 2 Breakdown limit 1 2 4 6 8 1 12 14 16 18 2 22 24 24 26 28 3 32 34 36 38 4 42 44 46 48 5 52 54 54 56 58 6 62 64 66 68 Number of matrix injections Figure 11. Breakdown and recovery of DDT after liner change (blue) and after matrix injection (orange). 9

In addition to 4,4 -DDT, the QC sample contains pentachlorophenol and benzidine. Figure 13 shows the tailing factor for benzidine and pentachlorophenol measured after liner replacement. From 18 to 24 matrix injections, the benzidine tailing factor increased from 1. to 1.8 close to the limit of 2. Liner replacement was not sufficient to reduce the tailing factor. Following the study protocol, the Intuvo Guard Chip was replaced, and the system was retested, then the tailing factor of benzidine dropped to 1.2. After 52 injections, the benzidine tailing factor increased to 1.7. In this case, liner replacement was sufficient to reduce the tailing factor to 1. CCV Results According to method 827D, every 12 hours the calibration needs to be verified by injecting a standard at the midpoint of the calibration range. The calculated concentration must be within ± 2 % of the actual concentration for the curve to be validated. If more than 2 % of the compounds fail the ± 2 % calibration check, the system is deemed unsuitable, and corrective action must be taken. In this study, a corrective action limit was set to be more stringent at a failure rate of 1 % for the 77 targets and surrogates (that is, 7 compounds failing CCV). Figure 14 shows the CCV result. After every 2 matrix injections, the number of compounds failing calibration exceeds the 1 % limit. After a liner change, the number of compounds failing the calibration dropped to below 1 %. Figure 12. Glass wool packed liner after 2 matrix injections. Tailing factor at 1 % peak height 3. 2.5 2. 1.5 1..5 Tailing factor limit Intuvo Guard Chip replacement 2 4 6 8 1 12 14 16 18 2 22 24 26 28 3 32 34 36 38 4 42 44 46 48 5 52 54 56 58 6 62 64 66 68 7 Number of matrix injections Pentachlorophenol Benzidine Figure 13. Tailing factor measurements after liner replacement for pentachlorophenol and benzidine. 1

Figure 14 also shows replacement of the Intuvo Guard Chip after 24 matrix injections triggered by the benzidine tailing factor (Figure 13). As indicated in Figure 14, one compound, hexachlorocyclopentadiene, failed the CCV check even after Intuvo Guard Chip replacement. Although the percent error dropped from 31.3 % to 26.7 %, the Intuvo Guard Chip replacement was not sufficient to reduce the percent error below 2 %. The sensitivity of this compound to matrix can be used as an indicator for determining Intuvo Guard Chip replacement frequency. Figure 15 shows a plot of the CCV checks for hexachlorocyclopentadiene as a function of the number of matrix injections. Based on this data, a preventative maintenance cycle can be estimated of one Intuvo Guard Chip replacement after every 6 matrix injections. Considering that the extract contained significant matrix residue, it may be surmised that this Intuvo Guard Chip replacement frequency would be sufficient for most soil matrices. Number of CCVs >2 % RSD 17 16 15 14 13 12 11 1 9 8 7 6 5 4 3 2 1 2 After liner replacement After matrix injections 4 Figure 14. Number of CCV failures after liner change (blue), and after matrix injections (orange). Guard Chip Replacement 6 8 1 12 14 16 18 2 22 24 24 26 28 3 32 34 36 38 4 42 44 46 48 5 52 54 56 58 6 62 64 66 68 68 Number of matrix injections Guard Chip Replacement Column Replacment Method 827D limit (2 %) Study limit (1 %) Number of matrix injections 5 1 15 2 25 3-5 CCV Percent error -1-15 -2-25 -3-35 Figure 15. CCV check percent error for hexachlorocyclopentadiene. 11

To gauge the level of system contamination as a result of a heavy matrix load, matrix injections were continued. Although the 1 % threshold was not reached within the 68 injections, the study was concluded. Figure 14 shows that liner replacement after the 68 th injection resulted in 1 CCV failures dropping to four. Intuvo Guard Chip replacement resulted in a further reduction from four to three CCV failures. Column replacement resulted in complete system restoration. The only failing compound after column replacement was benzidine which gave a response greater than the initial calibration (Figure 16). This indicates that the cause of these CCV failures was isolated to the column, and not the flow path or ion source. ISTD Results Method 827D requires that the variation of the area of internal standards should not exceed a factor of 2. If it does, the system is not suitable for analysis, and corrective action must be taken. Typically, loss in response in the internal standard is a result of ion source contamination. Figure 17 shows the normalized area for the internal standards over 68 injections after liner replacement. Throughout the study, the internal area was within the specified range. Comparison to an Agilent 789A GC As a point of reference, three similar matrix studies were carried out on the 789A GC coupled to a 5977 MSD using the protocol shown in Figure 9. Table 5 lists the results. Interestingly, the behavior of the 789A GC combined with an 5977 MSD appeared significantly different from the Intuvo 9 GC. For column numbers 1 and 3, the mechanism of the failure was a combination of ISTD responses falling below 5 % (which were restored with ion source cleaning), and greater than 1 % failure of the CCV checks (which were not restored by inlet maintenance or column trimming). For column 2, the mechanism of failure was a result of greater than 1 % failure of the CCV checks. Calibration check % error Normalized internal standard area 4 3 2 1-2 -3-4 Figure 16. Calibration checks after initial calibration, after 68 matrix injections, and after column replacement. 1.6 1.4 1.2 1..8.6.4 1,4-Dichlorobenzene-d4 Naphthalene-d8.2 Acenaphthene-d1 Phenanthrene-d1 Chrysene-d12 Perylene-d12 1 2 3 4 5 6 7 Number of matrix injections Figure 17. Normalized internal standard peak area across 68 matrix injections. Table 5. Results of the Agilent 789A GC combined with an Agilent 5977 MSD study. Column number After calibration After 68 matrix injections After column replacement Number of liners and septa replaced 2,4,6-Trichlorophenol Number of inlet seals replaced 4-Nitrophenol Number of column trims (3.5 cm each) Pentachlorophenol 1 12 2 6 24 2 4 2 5 8 3 6 3 4 12 Benzidine 1 2 3 4 5 6 7 8-1 Compound number Total number of matrix injections 12

Another difference in performing maintenance on the 789 GC compared to Intuvo 9 GC is that the Intuvo Guard Chip can be replaced more quickly and more assuredly on the Intuvo 9 GC compared to column trimming on the 789 GC. In addition, column trimming on the 789 GC required readjustment of the retention time windows after a sufficient portion of the column was removed. This did not occur with the replacement of the Intuvo Guard Chip. Figure 18 shows the change in retention time after replacement of an Intuvo Guard Chip on the Intuvo 9 GC compared to trimming 3.5 cm of the column on a 789 GC. The shift in retention times are clearly visible after trimming on the 789 GC compared to the Intuvo Guard Chip replacement on the Intuvo 9 GC where the overlays are virtually identical. Conclusion This study demonstrates the suitability of the Agilent Intuvo 9 GC for the analysis of SVOCs. The Intuvo 9 GC can easily meet the performance requirements as specified by USEPA method 827D. Compared to the Agilent 789 GC, the Intuvo 9 GC provided equivalent results in terms of relative retention time and relative response. In addition, repetitive injections of a soil extract illustrated the resilience of the Intuvo 9 GC to a substantial matrix challenge, and it was easier to maintain compared to the 789 GC. Intuvo Guard Chip replacement was more expedient than column trimming in terms of maintenance time, and did not require retention time adjustment. Acknowledgement The author would like to thank John Romesburg for his significant contribution to this study, ESC Lab Sciences for donation of matrix extracts, and Michael Szelewski for his expertise in environmental analysis. 13

References 1. Padilla-Sánchez, J. A.; Plaza Bolaños, P.; Frenich, A. G. Applications and Strategies Based on Gas Chromatography Low Resolution Mass Spectrometry (GC-LRMS) for the Determination of Residues and Organic Contaminants in Environmental Samples. In Advanced Techniques in Gas Chromatography Mass Spectrometry (GC-MS-MS and GC-TOF-MS) for Environmental Chemistry; Ferrer, I.; Thurman, M. Eds.; Elsevier Oxford, 213; Volume 61, pp 181-199. 2. Semivolatile Organic Compounds by Gas Chromatography/Mass Spectrometry (GC/MS); Method 827D; United States Environmental Protection Agency, Revision 4, February 27. 3. Analysis of Semivolatile Organic Compounds, Agilent Technologies, publication number 5991-718EN, 216. 4. Analysis of Semivolatiles Intuvo Guard Chip Protection, Agilent Technologies, publication number 5991-7182EN, 216. www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., 216 Published in the USA, September 1, 216 5991-7256EN