Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Similar documents
Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Separation of Large and Small Peptides by Supercritical Fluid Chromatography and Detection by Mass Spectrometry

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Detection of Trace Level Pharmaceuticals in Drinking Water by Online SPE Enrichment with the Agilent 1200 Infinity Series Online-SPE Solution

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

SAMHSA-Compliant LC/MS/MS Analysis of Phencyclidine in Urine with Agilent Bond Elut Plexa PCX and Agilent Poroshell 120

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Application Note. Abstract. Authors. Environmental

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Quantitation of Cystine and Identification of Related Metabolites in White Blood Cells Using a High Resolution Accurate Mass LC/MS Approach

Determination of Pharmaceuticals in Environmental Samples

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

Computer-assisted analysis of complex natural product extracts

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Agilent ESI and APCI sources: for polar to non-polar compounds

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Application Note. Edgar Naegele. Abstract

Application Note. Edgar Naegele. Abstract

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Test method for the determination of NDMA and NDEA by LC-MS/MS in Sartan containing film coated tablets

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Agilent All Ions MS/MS

Routine Multiresidue Pesticide Analysis using the Agilent 6470 Triple Quadrupole Mass Spectrometer

materials analysis Solutions for Your Analytical Business Markets and Applications Programs

Fast Screening Methods for Beta Blockers by HPLC with Agilent Poroshell 120 Columns

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Accurate Mass Analysis of Hydraulic Fracturing Waters

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

Screening and Quantitation of 240 Pesticides in Difficult Food Matrixes Using the Agilent 6545 Q-TOF LC/MS System

LC/Q-TOF Workflows for Comprehensive Micropollutant Analysis

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Use of the SFC-MS Splitter Kit G

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Application Note. Author. Abstract. Food Safety. Syed Salman Lateef Agilent Technologies, Inc. Bangalore, India

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Supporting Information

Measuring Drug-to-Antibody Ratio (DAR) for Antibody-Drug Conjugates (ADCs) with UHPLC/Q-TOF

The Agilent 1290 Infinity II Multi Method Solution

Application Note. Authors. Abstract. Food Testing & Agriculture. Stephen E. Reichenbach University of Nebraska Lincoln, NE, USA

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Fast methods for the determination of ibuprofen in drug products

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Assay Robustness Improvement for Drug Urinalysis Using FAIMS and H-SRM on a Triple- Quadrupole Mass Spectrometer

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Routine MS Detection for USP Chromatographic Methods

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Transcription:

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry Application Note Doping Control Authors Prof. Maria Kristina Parr Freie Universität Berlin Institute of Pharmacy Berlin, Germany Edgar Naegele and Bernhard Wuest Agilent Technologies, Inc. Waldbronn, Germany Abstract This Application Note demonstrates the use of the Agilent 126 Infinity Analytical SFC System for the separation of a library of polar pharmaceutical compounds. It is focused on the larger number of beta-blockers included in the used library. For all beta-blockers, performance data such as linearity, limit of detection (LOD) and limit of quantification (LOQ), and retention time and area RSDs are discussed. Finally, a urine sample spiked with a beta blocker was measured.

Introduction A subsection of a larger library of polar compounds, listed as prohibited substances by the World Anti-Doping Agency (WADA) 1, has been analyzed under HILIC conditions 2. In this study, this compound library was investigated by supercritical fluid chromatography (SFC) to evaluate the separation capabilities for such polar compounds and their detection by triplequadrupole mass spectrometry. Due to the high sample load in doping control analysis, a fast method for the analysis of known doping compounds is required. Due to its fast separation capabilities, SFC can play a significant role to cope with the number of samples in this application area. For the evaluation, the 13 beta-blockers inherent in the used library were used to determine typical limit of detection (LOD), limit of quantification (LOQ), retention time and area RSDs, as well as precision and accuracy data from the quantification of a sample. Experimental Instrumentation Agilent 126 Infinity Analytical SFC System (G439A): Agilent 126 Infinity SFC Control Module Agilent 126 Infinity SFC Binary Pump Agilent 126 Infinity High Performance Degasser Agilent 126 Infinity SFC Standard Autosampler Agilent 126 Infinity Thermostatted Column Compartment Agilent 126 Infinity Diode Array Detector with high-pressure SFC flow cell Agilent 646 Triple Quadrupole LC/MS System (G646C) with Agilent Jet Stream Agilent 126 Infinity Isocratic Pump (G131B) Agilent Splitter kit G439-68715 Instrumental setup Figure 1 shows the recommended configuration of the Agilent 126 Infinity Analytical SFC System with the Agilent 646 Triple Quadrupole LC/MS System. Column Agilent ZORBAX NH2, 4.6 15 mm, 5 µm (p/n 883952-78) Software Agilent MassHunter Data Acquisition Software for triple quadrupole mass spectrometer, Version 7.1 Agilent MassHunter Qualitative Software, Version 7. Agilent MassHunter Quantitative Software, Version 7. Connection of the SFC to the MS by splitting and make-up flow Make up composition Methanol/Water (95/5),.5 mm Ammonium Formate, +.2 % formic acid Make-up flow.5 ml/min Isocratic pump BPR Column Waste Splitter 1 Splitter 2 Figure 1. Configuration of the Agilent 126 Infinity Analytical SFC System with the Agilent 646 Triple Quadrupole LC/MS System. The column is directly connected to splitter 1 in the splitter assembly (BPR = backpressure regulator, UV detector not used, splitter kit p/n G439-68715). 2

Standards The compound library comprised polar pharmaceutical compounds including 13 beta-blockers at a concentration of 1 mg/ml each in acetonitrile. To prepare the stock solution for the described work, a 1:1 dilution in methanol was made. Chemicals All chemicals were purchased from Sigma-Aldrich, Taufkirchen, Germany. All solvents were LC/MS grade. Methanol was purchased from J.T. Baker, Germany. Fresh ultrapure water was obtained from a Milli-Q Integral system equipped with LC-Pak Polisher and a.22 μm membrane point-of-use cartridge (Millipak). Sample preparation A urine sample was spiked with penbutolol (25 ng/ml), diluted 1:5 with methanol, vortexed, and centrifuged at 14, g for 5 minutes. The supernatant was filtered, and the filtrate was used directly for injection. SFC method Parameter SFC flow SFC gradient Stop time Post time Modifier Value 3 ml/min minutes, 2 %B; 1 minutes, 5 %B 1 minutes 2 minutes BPR temperature 6 C BPR pressure Methanol +.1 % formic acid (FA) 15 bar Column temperature 4 C Injection volume MS method Parameter Ionization mode Capillary voltage Nozzle voltage Gas flow 5 µl, three-times loop overfill Value positive 2,5 V 2, V 8 L/min Gas temperature 22 C Sheath gas flow 12 L/min Sheath gas temperature 38 C Nebulizer pressure MRM conditions 25 psi See Table 1, showing detailed retention time, retention time window, and fragmentor and collision energies Table 1. MRM conditions for the studied beta-blockers, listing precursor ion mass, fragment ion mass, and fragmentor and collision energies. Precursor ion (m/z) Fragmentor (V) Quantifier CE (V) Qualifier CE (V) Penbutolol 292 113 236 12 74 2 Alprenolol 25 11 116 16 56 28 Oxprenolol 266 113 116 12 72 16 Bisoprolol 326 141 116 16 74 28 Esmolol 296 121 145 24 56 32 Propranolol 26 95 116 16 56 28 Celiprolol 38 136 251 2 74 32 Acebutolol 337 145 116 2 56 28 Nebivolol 46 151 151 28 44 44 Pindolol 249 11 116 16 56 28 Sotalol 273 83 255 8 133 28 Atenolol 267 11 145 28 56 28 Nadolol 31 88 254 12 56 36 3

Results and Discussion The separation of the 13 beta-blockers inherent in the 44-compound drug mix was performed on an amino column with a gradient of methanol (+.1 %FA) as modifier, starting at 2 % and going up to 5 % in 1 minutes. With this gradient, all 44 compounds could be separated. In this example, only the 13 beta blockers were monitored, and are discussed in detail (Figure 2). The beta blockers eluted between 4.68 and 7.38 minutes with a minimum of coelution. The first compound of the whole drug mix eluted at a retention time of 2.27 minutes, with the last one eluting at 8.19 minutes, respectively. For all beta-blockers, calibration curves were generated from 1, ng/ml down to 1 ng/ml (in modifier) to measure linearity, LOQ (S/N > 1), and LOD (S/N > 3) (Table 2). Typical LODs are below 1.5 ng/ml, and typical LOQs are below 5 ng/ml. Linear calibration was obtained for all compounds, with an R 2 better than.999. The 1 ng/ml calibration concentration was injected 1 times for a statistical evaluation (Table 2). The measured retention time RSDs are typically below.25 %, and the area RSDs are below 5 %. 1 6 1. Counts.8.6.4.2 1. 2. 3. 4. 5. 6. 7. 8. 9. Acquisition time (min) Figure 2. Separation of 13 beta-blockers from a library of 44 compounds. The applied gradient to separate all compounds went from 2 to 5 % methanol in 1 minutes. The earliest eluting compound has a retention time of 2.27 minutes, and the latest eluting compound has a retention time of 8.198 minutes. The beta-blockers eluted between 4.68 and 7.38 minutes. Table 2. Retention time, RSD s of retention time and areas as well as LODs, LOQs, and linearity of the 13 beta-blockers. The LODs and LOQs are calculated from calibration curves created between 1 and 1, ng/ml. The RSDs are calculated from 1 injections of the 1 ng/ml calibration point. Compound RT (min) RT RSD (%) Area RSD (%) LOD (ng/ml) LOQ (ng/ml) R 2 Penbutolol 4.68.18 3.54.2 1.6.9994 Alprenolol 4.784.16 4.42 1.5 5..999 Oxprenolol 4.912.16 2.76.1.25.9992 Bisoprolol 5.45.15 3..19.6.9991 Esmolol 5.111.15 6.83 1.41 4.71.9992 Propranolol 5.611.18 8.5 1.71 5.7.9997 Celiprolol 6.158.26 3.29.37 1.25.9998 Acebutolol 6.397.26 1.68.66 2.19.9997 Nebivolol 6.463.24 5.31 2.85 9.52.9996 Pindolol 6.854.28 3.67 3.64 1.9.9994 Sotalol 7.88.24 6.29 5.51 18.38.9995 Atenolol 7.322.28 3.73 1.98 6.66.9995 Nadolol 7.383.29 2.53.41 1.38.9997 4

As an example of a real-life sample, penbutolol was spiked into urine at a concentration of 25 ng/ml, and prepared as described in the experimental section. Because pure aqueous samples should not be injected into the SFC, the sample was diluted 1:5 in methanol. For the quantitative determination of the penbutolol, a calibration curve from 1, ng/ml down to 2 ng/ml was created in solvent (Figure 3A). The measured average concentration of the diluted sample was 45.59 ng/ml, and the determined sample concentration 228.45 ng/ml. The compound eluted at 4.53 minutes with good signal intensity (Figure 3B) and the quantifier/quantifier ratio was in the expected range (Figure 3C). For a statistical evaluation, the sample was injected 1 times, and the retention time and area RSD were calculated to be.14 and 2.29 %, respectively. The concentration precision and the accuracy was calculated to be 2.28 and 91.38 %, respectively. Counts Responses 1 6 3. 2. 1. 1 4 3. 2. A Penbutolol y = 3268.68287*x 436.716917 R 2 =.9994 1 2 3 4 5 6 7 8 9 1, Concentration (ng/ml) Penbutolol RT RSD (%).14 Area RSD (%) 2.29 Average concentration (ng/ml) 45.69 Concentration precision (%) 2.28 Concentration accuracy (%) 91.38 + MRM (292. & 236.) Penbutolol Ratio = 44. (13.6 %) + MRM (4.336 4.592 min)(292. &**) 1 4 1 5 236. B 4.531 min C D 292. & 236. 292. & 74. 1.5 3. Counts 2. O H OH H N CH 3 CH 3 CH 3 Counts 1. 74. 1. 1..5 4. 4.4 4.8 5.2 Acquisition time (min) 4. 4.4 4.8 5.2 Acquisition time (min) 292. 1 15 2 25 3 Mass-to-charge (m/z) Figure 3. Determination of penbutolol in urine. A) Calibration curve of penbutolol between 2 and 1, ng/ml. The concentration measured in the diluted sample is indicated by the arrow. B) Quantifier transition of penbutolol from the diluted sample. C) Quantifier/qualifier ratio of the measured sample. D) MRM transitions measured for penbutolol in the sample. The table inset shows retention time RSD and area RSD as well as average concentration, concentration precision and concentration accuracy calculated from 1 injections of the sample. 5

Conclusion This Application Note demonstrates the possibility to separate a large number of highly polar pharmaceutical compounds within a short run time by SFC. Their detection by connecting the SFC to a triple-quadruple mass spectrometer with limits of detection typically below 1.5 ng/ml has been shown. The retention time RSDs were below.25 %, and area RSDs were below 5 %. It was demonstrated that pharmaceutical compounds could be measured in an aqueous real-life sample by dilution with organic solvents with sufficient sensitivity and concentration precision and accuracy. References 1. The World Anti-Doping Code, The 215 Prohibited List International Standard, World Anti-Doping Agency, Montreal, Canada, 2 September 214, Available at www.wada-ama.org (accessed November 1, 215) 2. Mazzarino, M.; et al. Screening and confirmation analysis of stimulants, narcotics and beta-adrenic agents in human urine by hydrophobic interaction liquid chromatography coupled to mass spectrometry. Journal of Chromatography A 211, 1218, 8156 8167 6

7

www.agilent.com/chem This information is subject to change without notice. Agilent Technologies, Inc., 215-217 Published in the USA, October 15, 217 5991-6437EN