MSE 321 Structural Characterization

Similar documents
The illumination source: the electron beam

Part II: Thin Film Characterization

AP5301/ Name the major parts of an optical microscope and state their functions.

= 6 (1/ nm) So what is probability of finding electron tunneled into a barrier 3 ev high?

Why microscopy?

Basic structure of SEM

Scanning Electron Microscopy

A system of two lenses is achromatic when the separation between them is

Part I Basics and Methods

*Specifications subject to change without notice.

Chapter 2 Instrumentation for Analytical Electron Microscopy Lecture 5. Chapter 2 CHEM 793, 2011 Fall 1

SEM Doctoral Course MS-636. April 11-13, 2016

tip conducting surface

High-Resolution. Transmission. Electron Microscopy

M2 TP. Low-Energy Electron Diffraction (LEED)

object objective lens eyepiece lens

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Electron Microprobe Analysis 1 Nilanjan Chatterjee, Ph.D. Principal Research Scientist

Electron microscopy in molecular cell biology I

Electron beam scanning

Massachusetts Institute of Technology. Dr. Nilanjan Chatterjee

Ecole Franco-Roumaine : Magnétisme des systèmes nanoscopiques et structures hybrides - Brasov, Modern Analytical Microscopic Tools

Nano-Microscopy. Lecture 2. Scanning and Transmission Electron Microscopies: Principles. Pavel Zinin HIGP, University of Hawaii, Honolulu, USA

Chapter 2 Structure and Imaging of a Transmission Electron Microscope (TEM)

Transmission Electron Microscopy

Gaetano L Episcopo. Scanning Electron Microscopy Focus Ion Beam and. Pulsed Plasma Deposition

1986 s Nobel Prize in Physics

h p λ = mν Back to de Broglie and the electron as a wave you will learn more about this Equation in CHEM* 2060

Physical Principles of Electron Microscopy. 2. Electron Optics

Ultrafast nanophotonics - optical control of coherent electron -

Invited Lecture. "Different Aspects of Electron Microscopy. Sardar Vallabhbhai National Institute of Technology, Surat. Deepak Rajput & S.K.

Modern Optical Spectroscopy

For more sample papers visit :

Analytical Methods for Materials

Everhart-Thornley detector

Scanning Electron Microscopy & Ancillary Techniques

Chapter 9. Electron mean free path Microscopy principles of SEM, TEM, LEEM

LC circuit: Energy stored. This lecture reviews some but not all of the material that will be on the final exam that covers in Chapters

Name Final Exam May 1, 2017

Nova 600 NanoLab Dual beam Focused Ion Beam IITKanpur

Sparks in Gases: Line Spectra

Experimental methods in Physics. Electron Microscopy. Basic Techniques (MEP-I) SEM, TEM

Dual Nature of Radiation and Matter-I

Biochemistry 9001 Protein structure determination by 3D Electron Microscopy

Scanning Electron Microscopy

Practical course in scanning electron microscopy

April 10th-12th, 2017

School. Team Number. Optics

CHEM 681 Seminar Mingqi Zhao April 20, 1998 Room 2104, 4:00 p.m. High Resolution Transmission Electron Microscopy: theories and applications

Phys 100 Astronomy (Dr. Ilias Fernini) Review Questions for Chapter 5

A small object is placed a distance 2.0 cm from a thin convex lens. The focal length of the lens is 5.0 cm.

Introduction to Electron Microscopy Andres Kaech. Instrumentation

Lecture 7: Electron Emission

Recent Status of Polarized Electron Sources at Nagoya University

MSE 321 Structural Characterization

HOW TO APPROACH SCANNING ELECTRON MICROSCOPY AND ENERGY DISPERSIVE SPECTROSCOPY ANALYSIS. SCSAM Short Course Amir Avishai

Structure analysis: Electron diffraction LEED TEM RHEED

Magnifying Glass. Angular magnification (m): 25 cm/f < m < 25cm/f + 1. image at 25 cm (= normal near point) relaxed eye, image at (normal) far point

Astronomy 203 practice final examination

Weak-Beam Dark-Field Technique

LECTURE 11 ELECTROMAGNETIC WAVES & POLARIZATION. Instructor: Kazumi Tolich

General Physics II Summer Session 2013 Review Ch - 16, 17, 18

Lecture 20 Optical Characterization 2

Introduction to Electron Beam Lithography

MSE 321 Structural Characterization

CHARACTERIZATION of NANOMATERIALS KHP

SOLID STATE PHYSICS PHY F341. Dr. Manjuladevi.V Associate Professor Department of Physics BITS Pilani

Lecture 2: Basic Astronomical Optics. Prisms, Lenses, and Mirrors

A Question. Simple Magnifier. Magnification by a Lens 11/29/2011. The last lecture

Model : JEOL JSM 7610f

Auger Electron Spectroscopy

PHYS General Physics II Lab The Balmer Series for Hydrogen Source. c = speed of light = 3 x 10 8 m/s

MEMS Metrology. Prof. Tianhong Cui ME 8254

PhysicsAndMathsTutor.com 1

Inside this Issue. Articles. Atmospheric pressure plasma system for Textiles

QUANTUM PHYSICS. Limitation: This law holds well only for the short wavelength and not for the longer wavelength. Raleigh Jean s Law:

PHYS 102 Exams. PHYS 102 Exam 3 PRINT (A)

SEM stands for Scanning Electron Microscopy. The earliest known work describing

PhysicsAndMathsTutor.com 1

Needle cathodes for high-brightness beams. Chase Boulware Jonathan Jarvis Heather Andrews Charlie Brau

The Electromagnetic Properties of Materials

Spatial Frequency and Transfer Function. columns of atoms, where the electrostatic potential is higher than in vacuum

20. Aberration Theory

Techniques EDX, EELS et HAADF en TEM: possibilités d analyse et applications

Electromagnetic Waves

Select the response that best answers the given statement. Be sure to write all final multiple choice answers on your Scantron answer sheet.

Solution set for EXAM IN TFY4265/FY8906 Biophysical microtechniques

Supporting Information s for

A) n L < 1.0 B) n L > 1.1 C) n L > 1.3 D) n L < 1.1 E) n L < 1.3

Optical/IR Observational Astronomy Telescopes I: Optical Principles. David Buckley, SAAO. 24 Feb 2012 NASSP OT1: Telescopes I-1

Electron Diffraction

The Liquid Metal Ion Source A Hot Ion Source. Jon Orloff Professor Emeritus, University of Maryland and FEI Company Hillsboro, Oregon

Electron Microscopy. SEM = Scanning Electron Microscopy TEM = Transmission Electron Microscopy. E. coli, William E. Bentley, Maryland, USA

Physics 23 Fall 1998 Lab 4 - The Hydrogen Spectrum

ABC s of Electrochemistry series Materials Characterization techniques: SEM and EDS Ana María Valenzuela-Muñiz November 3, 2011

PHYSICS PAPER 1 (THEORY)

AS 101: Day Lab #2 Summer Spectroscopy

Test 4 Preparation Questions

LECTURE 11 ELECTROMAGNETIC WAVES & POLARIZATION. Instructor: Kazumi Tolich

Technology for Micro- and Nanostructures Micro- and Nanotechnology

Transcription:

Optical Microscope Plan Lenses In an "ideal" single-element lens system all planar wave fronts are focused to a point at distance f from the lens; therefore: Image near the optical axis will be in perfect focus on a flat image sensor/screen. Non-axial rays will come into focus before the image sensor. This problem is reduced when the imaging surface is curved, as in the eye. φ Most current photographic lenses are designed to minimize field curvature (plan lenses), and so effectively have a focal length that increases with ray angle. f = f 0 /cosφ

Achromatic Lenses f (focal length) = f(µ) µ blue > µ red (dispersion) Chromatic aberration means that a lens will not focus different colors in exactly the same place because the focal length depends on refraction and the index of refraction for blue light (short λ) is larger than that of red light (long λ). The amount of chromatic aberration depends on the dispersion of the glass. Image Formation Three ways to form an image: Projection (shadow) Optical (with lenses) } all parts of image simultaneously Scanning Three requirements for image formation: Brightness Contrast Resolution

Visible Spectrum Electrons vs Light Electron Optics 4000 nm λ 7000 Å Wavelength of a photon is inversely proportional to its energy, λ = ch/e hν e- 100 Å λ 2000 Å X-Rays Wavelength of an electron is dependent on its energy (accelerating voltage) JEOL 2100-HR TEM: Hitachi 4500 SEM: λ = 0.0251 Å (200 kv) 5 orders of magnitude } smaller than for visible light! λ = 0.0859 Å (20 kv) Electrons Electrons are much more strongly scattered by matter (EM operates in vacuum) Lenses are magnetic fields (n = 1) Electrons carry a charge r d = 0.61λ α Electrons vs Light Electron Optics hν e- Compared to optical microscopy, electron microscopy offers the following advantges: Higher resolution Higher magnification Greater depth of field Crystallographic information Chemical analysis

Electron Generation β - The illumination system in an EM consists of a source of electrons (electron gun) and a lens system to focus and control emitted electrons. Electron guns overcome the work function of the electrons by either resistive heating or a strong electric field (or both). Thermionic Emission: A heated wire or crystal is given enough thermal energy to overcome work function (Edison Effect) plus an electric potential (0.5 300 kv) to accelerate the electrons. Tungsten (hairpin) W has very high T m (3422 C) so can have more thermal energy But work function still high (φ = 4.5 ev) even at operating T 2500 C Cheap ($30) Hitachi S-3400N SEM LaB 6 /CeB 6 Lower work function (φ = 2.5 ev) than W at operating T 1500 C, so it is easier to pull electrons off and can have beams 30 times brighter than with W hairpin. Require high vacuum, suffer from thermal shock, and are expensive ($1000). Higher T causes increased volatilisation and premature failure. T m (LaB 6 ) = 2530 C, T m (CeB 6 ) = 2540 C. JEOL 2100 HR TEM, CAMECA SXFive EPMA Electron Generation Field Effect Emission: The field effect gun (FEG) uses a W single crystal and high field (> 10 9 V/m), which lowers the height of the potential barrier via the Schottky effect. In addition, if the field is sufficiently high, the width of the potential barrier becomes small enough to allow electrons to escape via tunnelling (field emission). Field reduces the work function by φ = 3.8x10-5 F ½ ev Cold FEG operates at room temperature, but requires flashing Thermally-Assisted FEG (Schottky emitter) pointed W crystal (coated with ZrO 2 ) is welded to V-shaped W filament and heated to 1500 C ZrO 2 lowers work function (φ = 2.8 ev) improved stability and no flashing enhanced emission by giving some thermal energy, which reduces tunneling distance (really a field-assisted thermionic source) FEI Teneo F-SEM

Thermionic Emission Current density at filament surface: j = AT 2 e -φ/kt [A/m 2 ] A = 1.20173x10 6 [A/m 2 /K 2 ] (Richardson s constant) T = absolute temperature [K] increases rapidly as T increases & φ decreases (so use high-t m low-φ material W) Brightness: B = j/(πα 2 ) (j = beam current density, α = convergence semi-angle) Example: (at cross-over: Langmuir equation): B = (jv)/(πkt) [A/m 2 /sr] V = voltage [V] k = 8.617x10-5 ev/k For W (φ = 4.5 ev) operating at T = 2500 K and V = 100 kv: j = (1.20173x10 6 A/m 2 K 2 )(2500 K) 2 exp(-4.5 ev/[(8.617x10-5 ev/k)(2500 K)] = 6364 A/m 2 B = (6364 A/m 2 )(100000 ev)/[(π sr)(8.617x10-5 ev/k)(2500 K)] = 9.4035x10 8 A/m 2 /sr ~9x10 4 A/cm 2 /sr Field Effect Emission Current density = j = 6.2x10-10 (E f /φ) ½ F 2-6.8x10 exp 9 φ 3/2 Acm -2 E f + φ F Temperature-independent E f 5 ev for W at R.T. For a field F > 5x10 9 V/m, field emission exceeds thermionic emission. To apply such a high field, W has to be prepared as a sharp point of diameter 100 nm, which makes it very delicate; therefore, ultra-high vacuum systems (10-8 10-9 torr) are necessary adds cost (JEOL 2100 LaB 6 TEM operates at 10-7 torr). Low energy spread (δe 0.3 ev) for improved resolution and better interpretation of EELS data More coherence Highly sensitive to contamination - cold FEG must be flashed every few hours

Brightness increases linearly with voltage. Electron Generation A comparison of electron emission characteristics of W, LaB 6, CeB 6 and FEG guns W wire LaB 6 <100> CeB 6 <100> FEG (cold) FEG (Schottky) Operating temperature ( C) 2500 1500 1500 27 1500 Work function (ev) 4.5 2.7 ~2.5 4.5 2.8 Emission current density (A/cm 2 ) 3 30 30 17,000 5,300 Brightness (A cm -2 sr -1 ) 10 4 10 5 10 5 2x10 7 10 7 Crossover diameter (mm) 30 10 10 0.01 0.02 Short-term beam current stability (%RMS/hr) <1 <1 <1 3-5 <1 Energy spread at 100 kv (ev) 2 1.5 1.5 0.2-0.4 0.5 Typical service life (hr) 30-100 1000 1,500 10 3 10 4 10,000 Operating vacuum (torr) 10-5 10-7 10-7 10-10 10-9 Evaporation rate (g cm -2 s -1 ) NA 2.2 x 10-9 1.6 x 10-9 NA NA Resolution determined largely by diameter of beam and δe Thermionic guns have large areas of emission beam is demagnified x10-4 very stable and insensitive to contamination Cold FEG has small area of emission beam is reduced just 5-10x to 1 nm unstable and very sensitive to contamination, but nearly coherent End of Lecture Yes, in your face! I found Waldo! Goodfellow Cartoon Caption Contest, September 2012 First place: Daniel Glover, Ohio