Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Similar documents
Standard Operating Procedure for the Analysis of Dissolved Inorganic Carbon CCAL 21A.1

Standard Operating Procedure for the Analysis of Silica in Fresh Waters CCAL 32A.2

Standard Operating Procedure for the Analysis of Nitrate/Nitrite and Nitrite in Fresh Waters CCAL 31B.1

Standard Operating Procedure for the Analysis of Fresh Water Samples for Orthophosphorus CCAL 34B.2

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

USGS Troy WSC Laboratory Ion Chromatography SOP EPA 300.0, Rev Jordan Road SOP. No. 1 Rev. No. 1.7 Troy, NY Date 01/25/18 Page 1 of 7

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

ALLOWAY METHOD OUTLINE

Revision: 11 (MBAS) ALLOWAY METHOD OUTLINE. Standard Laboratory Method:

California Environmental Protection Agency Air Resources Board

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Chloride by Flow Injection Analysis (FIA)

METHOD DETERMINATION OF SULFATE BY AUTOMATED COLORIMETRY. Edited by James W. O'Dell Inorganic Chemistry Branch Chemistry Research Division

Determination of trace anions in concentrated hydrofluoric acid

Standard Operating Procedures. Determination of Anions and/or Cations Extracted from Nylon Filters by Ion Chromatography (IC)

Standard Operating Procedure for the Digestion and Analysis of Fresh Water Samples for Total Phosphorus and Total Dissolved Phosphorus CCAL 35B.

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Hexavalent Chromium by Flow Injection Analysis (FIA)

Total Dissolved Residue (Total Dissolved Solids, TDS) in Aqueous Matrices

Hach Company TNTplus TM Phosphorus Spectrophotometric Measurement of Phosphorus in Water and Wastewater

Method for Nitrite determination on Low-Range Samples

Timberline Ammonia-001. Determination of Inorganic Ammonia by Continuous Flow Gas Diffusion and Conductivity Cell Analysis

Nitrate plus Nitrite and Nitrite in Seawater by Segmented Flow Analysis (SFA)

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

columns IonPac AS17-C Anion Exchange Column

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

METHOD DETERMINATION OF PHOSPHORUS BY SEMI-AUTOMATED COLORIMETRY

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

Standard Operating Procedure VUSP M

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

O Sodium Bicarbonate, NaHCO 3 Ammonium Molybdate Tetrahydrate, (NH 4. O 4 Sodium Chloride, NaCl Ascorbic Acid, C 6 H 8 O 6

LOW-RANGE TOTAL CHLORINE

Ohio EPA Total (Extracellular and Intracellular) Microcystins - ADDA by ELISA Analytical Methodology Ohio EPA DES Version 2.

Standard Operating Procedure for: Conductivity Using Cole-Parmer Traceable Portable Conductivity Meter. Missouri State University.

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography 1

METHOD DETERMINATION OF AMMONIA NITROGEN BY SEMI-AUTOMATED COLORIMETRY

British American Tobacco Group Research & Development. Method - Determination of ammonia in mainstream smoke

columns IonPac AS22 Anion-Exchange Column

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

SECTION D.2 AMMONIA NITROGEN

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

anion-exchange column

Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

Inline sample preparation for the determination of anions in sodium hydroxide

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

Alkalinity, USEPA by Flow Injection Analysis (FIA)

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Determination of Anions and Cations on Multi (2) - Ion Chromatography System

Laboratory ID. Laboratory Name. Analyst(s) Auditor. Date(s) of Audit. Type of Audit Initial Biennial Special ELCP TNI/NELAP.

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of urea in ultrapure water by IC-MS/MS

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Dionex IonPac AS28-Fast-4µm column

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Method to 0.50 mg/l NH 3 N Powder Pillows

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Method Abstract. Flow Solution Fluoride, USGS by ISE and FIA

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Method to 0.50 mg/l NH 3 N Powder Pillows

STANDARD OPERATING PROCEDURES

United States EPA Method 415.3

Standard Operating Procedure VUSP N

Bromide by Flow Injection Analysis (FIA)

METHOD DETERMINATION OF TOTAL CYANIDE BY SEMI-AUTOMATED COLORIMETRY

Determination of Adsorbable Organic Halogen in Wastewater

Determination of Inorganic Anions in Wastewater by Ion Chromatography

METHOD DETERMINATION OF INORGANIC ANIONS IN DRINKING WATER BY ION CHROMATOGRAPHY. Revision 1.0

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

The determination of trace anions in concentrated phosphoric acid

METHOD #: Recommended for Approval for NPDES (August, 1991)

columns IonPac AS18 Anion-Exchange Columns

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

Analysis of pulping liquors by suppressed ion chromatography

DETERMINATION OF TOTAL HALOGEN CONTENT IN HALOGEN-FREE FLUXES BY INDUCTIVELY COUPLED PLASMA AND SOME LIMITATIONS OF ION CHROMATOGRAPHY

Determination of Common Inorganic Anions in Environmental Waters Using a Hydroxide Selective Column

DR/4000 PROCEDURE NITRATE. 2. The display will show:

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of silicate in high-purity water using ion chromatography and online sample preparation

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Determination of trace anions in high-nitrate matrices by ion chromatography

Transcription:

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Cooperative Chemical Analytical Laboratory College of Forestry Oregon State University 321 Richardson Hall Corvallis, Oregon Prepared by Kathryn Motter And Cameron Jones Revised

Page 2 of 11 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 Table of Contents 1.0 Scope and Application... 3 2.0 Summary of Method... 3 3.0 Definitions... 3 4.0 Interferences... 4 5.0 Safety... 4 6.0 Equipment and Supplies... 5 7.0 Reagents and Standards... 5 7.1 Preparation of Reagents... 5 7.2 Preparation of Standards... 6 8.0 Sample Handling and Storage... 6 9.0 Quality Control... 7 10.0 Calibration and Standardization... 7 11.0 Procedure... 8 11.1 Dionex ICS-1500 Instrument Operating Parameters... 8 11.2 Procedure... 8 11.3 System Notes... 8 12.0 Data Analysis and Calculations... 8 13.0 Method Performance... 9 14.0 Pollution Prevention... 9 15.0 Waste Management... 9 16.0 References... 10 17.0 Tables, Diagrams, Flowcharts, and Validation Data... 10 18.0 Document Revision History... 10

Page 3 of 11 Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2 1.0 Scope and Application 1.1 This method is applicable to the sequential measurement of anions in surface, ground, drinking and waste waters by ion chromatography. Measurement is direct on a single instrument after sample filtration. Currently, CCAL quantifies chloride, bromide, sulfate and fluoride exclusively; nitrate and phosphate may also be determined simultaneously. Detection limits of 0.01 mg Cl/L, 0.01 mg Br/L, 0.01 mg SO4/L and 0.01 mg F/L have been determined for this method using a 300 ul sample loop and SRS current of 43 ma. 2.0 Summary of Method 2.1 Samples are injected into a stream of carbonate/bicarbonate eluent, passed through a series of ion exchange columns and into a conductivity detector. The first column, a guard column, protects the analytical column by removing particulate and organic matter. The analytical column separates anions by their relative affinities for a low-capacity, strongly basic anion exchanger. The suppressor (between the analytical column and the conductivity detector) provides continuous suppression of background conductivity of the eluent and enhances response of the target analytes by acidification. The separated anions in their acidified form are measured by conductivity. The anions are identified based on retention times and quantified by conductivity relative to standard response. 3.0 Definitions 3.1 DI water: Water that has been through a deionization system to produce water similar to ASTM Type I reagent with 16.7 Mohms resistivity (ASTM) (Reference 16.3). 3.2 Method Detection Limit (MDL): The minimum concentration of an analyte that can be measured and reported with 99% confidence, based on a one-sided 99% confidence interval (t-value at a significance level of 0.01

Page 4 of 11 and n-1 degrees of freedom) from at least seven repeated measurements of a low concentration standard measured within an analysis run. MDL = ts Where, t = Student s t value at a significance level of 0.01 and n-1 degrees of freedom s = standard deviation of at least seven repeated measurements of a low level standard 4.0 Interferences 4.1 Any substance with a retention time similar to that of an anion of interest, that produces a detector response, will interfere. A high concentration of a component with similar elution times may also interfere with resolution, and sometimes retention, of other analytes. Peak separation may be achieved by sample dilution, reduced eluent concentration or change in instrument parameters (eg sample size, flow rate or suppressor current). 4.2 Introduction of contaminants may lead to spurious peaks or elevated baseline. 4.3 Samples should be pre-filtered to prevent damage to columns and flow system. 5.0 Safety 5.1 The toxicity or carcinogenicity of each reagent has not been precisely determined; however, each chemical should be regarded as a potential health hazard. Exposure to these chemicals should be reduced to the lowest possible level. Cautions are included for known extremely hazardous materials. 5.2 The following chemicals have the potential to be highly toxic or hazardous. For detailed explanations, consult the MSDS. 5.2.1 Sulfuric acid (used in regeneration of the suppressor) 5.2.2 Oxalic acid (used in regeneration of the suppressor) 5.2.3 Methanol (cleaning agent)

Page 5 of 11 6.0 Equipment and Supplies Note: Brand names, suppliers and part numbers are for illustrative purposes only. No endorsement is implied. Equivalent performance may be achieved using apparatus and materials other than those specified here, but demonstration of equivalent performance that meets the requirements of this method is the responsibility of the laboratory. 6.1 Dionex ICS-1500 Chromatography System 6.1.1 Dionex AS-40 Autosampler 6.1.2 Instrument Controller 6.1.3 Data Collection Software (Chromeleon) 6.1.4 Anion Guard Column (AG14A-7um, 4 x 50 mm) 6.1.5 Anion Analytical Column (AS14A-7um, 4 x 250 mm) 6.1.6 Self-Regenerating Suppressor (AERS 500, 4-mm) 6.1.7 Eluent reservoirs 6.2 Nitrogen gas and regulator 6.3 Dionex 5 ml sample vials and filter caps 6.4 Laboratory glassware and pipettes 6.5 Balance 6.6 Safety glasses 6.7 Nitrile gloves 6.8 Lab coat or apron 6.9 Laboratory exhaust fume hood 6.10 High Density Polyethylene (HDPE) bottles 7.0 Reagents and Standards 7.1 Preparation of Reagents 7.1.1 Sodium Carbonate/Bicarbonate Eluent Concentrate (0.8 M Na 2 CO 3 /0.1 M NaHCO 3 ) Dissolve 84.79 g sodium carbonate (Na 2 CO 3, FW 105.99) and 8.40 g sodium bicarbonate (NaHCO 3, FW 84.01) in 700 ml DI water in a 1 L volumetric flask. When dissolution is complete, bring up to

Page 6 of 11 1000 ml with DI water. Final concentration; 0.8 M Na 2 CO 3 /0.1 M NaHCO 3. 7.1.2 Working Carbonate Eluent Solution (8 mm Na 2 CO 3 /1 mm NaHCO 3 ) Pipette 10.0 ml of eluent concentrate into a 1 L volumetric flask and bring to volume with DI water. Exclude particulates from the system with Dionex 0.2 um end-line filter. 7.1.3 Supressor Regenerant Dilute solutions of sulfuric or oxalic acid may be required to clean away build-up over time. See the suppressor manual for concentration specific to application. 7.2 Preparation of Standards 7.2.1 Calibration Standards: Standards are prepared by dilution of a mixed standards purchased from a vendor that provide traceability to NIST standards. A mixed standard of concentration 100 mg/l of each chloride, bromide, fluoride and sulfate is used to prepare working standards in the table below. mg Cl/L mg Br/L mg SO 4 /L mg F/L 1 0.20 0.20 0.20 0.20 2 0.50 0.50 0.50 0.50 3 1.00 1.00 1.00 1.00 4 2.00 2.00 2.00 2.00 5 5.00 5.00 5.00 5.00 7.2.2 Second Source Check Standard: Standard prepared by dilution of mixed element standard purchased from a vendor that provide traceability to NIST standards, this from a source other than that of the calibration standards. Check standard concentration should be mid-calibration range; currently 1.00 mg/l for each chloride, bromide, fluoride and sulfate. 8.0 Sample Handling and Storage 8.1 If required, unfiltered samples are filtered upon receipt through glass fiber filters into clean HDPE bottles and stored at 4 C in the dark. If samples must be held longer than 28 days prior to analysis, they are stored frozen at -18ºC unless otherwise requested.

9.0 Quality Control IC/CCAL 50B.2 Page 7 of 11 9.1 Preparation of stock standards is recorded on worksheets and documented by weight of standard added to a given flask before dilution to volume with DI water. All records of certification are kept on file at CCAL Laboratory. 9.2 Blank: DI water run before and after the calibration, and after every check standard. 9.3 Second Source Check Standard: Run every ten samples to monitor drift and column efficiency. 9.4 Method Detection Limit (MDL): Established for each analyte. Based on a one-sided 99% confidence interval (t-value) from at least seven repeated measurements of a low concentration standard. The t-distribution value is multiplied by the standard deviation of the population (n-1) to obtain the MDL. 9.5 Analytical Duplicate: Separate analysis from the same sample aliquot. One per ten samples. 9.6 A bulk surface water Quality Control Check Standard is used to generate control charts and is tracked over time to monitor overall performance. 9.7 Samples with concentration greater than the highest calibration standard are diluted and reanalyzed. 10.0 Calibration and Standardization 10.1 Balances: calibrated yearly by external vendor. 10.2 Pipette delivery checked by weight to within 1% of theoretical weight of aliquot volume. 10.3 The IC is calibrated every analysis run. The Chromeleon software is setup to use quadratic curve fit of six working standards (see section 7.2). The r 2 value of each calibration curve should be greater than 0.995. 10.4 Check standard recovery must be within 10% of theoretical value.

11.0 Procedure IC/CCAL 50B.2 Page 8 of 11 11.1 Dionex ICS-1500 Instrument Operating Parameters 11.2 Procedure Nitrogen Gas Pressure: tank regulator 40 psi eluent regulator 6 psi Flow Rate 1.0 ml/min Sample Loop Size 300 ul Aquisition Time 15.00 Pressure Limit 200-3000 psi Injection Duration 60 sec Cell T 30ºC Column T 30ºC Data Collection Rate 5 Hz Supressor current 43 ma 11.2.1 Prime the pumps 11.2.2 Allow instrument to equilibrate at normal operating parameters until baseline is stable. Normal baseline conductivity is approximately 25 us/cm. Normal operating pressure approximately 1700 psi. Values vary with column life. 11.3 System Notes 11.3.1 Yearly preventative maintenance as suggested by the manufacturer is required for optimum performance. 11.3.2 Column and suppressor must be maintained or replaced as needed to retain peak separation. 11.3.3 Peak separation for individual sample matrices may be achieved by sample dilution, reduced eluent concentration or change in instrument parameters (eg sample size, flow rate or suppressor current). 12.0 Data Analysis and Calculations 12.1 The Chromeleon software prepares a standard curve by plotting standard response against known concentration. Sample response is compared to the standard curve to determine concentration. 12.2 Samples with concentration greater than the highest calibration standard are diluted and reanalyzed.

Page 9 of 11 13.0 Method Performance 13.1 This method was validated through inter-laboratory studies. The CCAL Water Analysis Laboratory participates in the USGS Standard Reference Water QA program and the National Water Research Institute s (NWRI) Environment Canada Proficiency Testing (PT) Program. 14.0 Pollution Prevention 14.1 The chemicals used in this method pose little threat to the environment when properly managed. 14.2 All standards and reagents should be prepared in volumes consistent with laboratory use to minimize the volume of disposable waste. 14.3 For further information on pollution prevention consult Less is better: Laboratory Chemical Management for Waste Reduction, available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington D.C. 20036, (202) 872-4477. 15.0 Waste Management 15.1 It is the laboratory s responsibility to comply with all federal, state and local regulations governing waste management, and to protect the environment by minimizing and controlling all releases from fume hoods and bench operations. Compliance with all sewage discharge permits and regulations is required. 15.2 For further information on waste management, consult The Waste Management Manual for Laboratory Personnel, and Less is Better: Laboratory Chemical Management for Waste Reduction, both available from the American Chemical Society s Department of Government Relations and Science Policy, 1155 16 th Street NW, Washington DC, 20036.

16.0 References IC/CCAL 50B.2 Page 10 of 11 16.1 Standard Methods For The Examination of Water and Wastewater, Method 4110 B; Ion Chromatography with Chemical Suppression of Eluent Conductivity. American Public Health Association. 21 st Edition, 2005. 16.2 Code of Federal Regulations. Protection of Environment. Section 40, Appendix B to Part 136. Definition and procedure for the determination of the method detection limit. Revision 1.11. Revised July 1, 1990. Office of the Federal Register, National Archives and Records. 16.3 ASTM. American Society for Testing and Materials. Standard Specifications for Reagent Water. D1193-77 (Reapproved 1983). Annual Book of ASTM Standards, Vol. 11.01. ASTM: Philadelphia, PA, 1991. 16.4 Dionex Instrument, Software and Accessory manuals. Dionex Corporation; Sunnyvale, California. 16.5 Pfaff, J.D., Hautman, D.P. and Munch, D.J., 1997. EPA Method 300.1, Revision 1.0. Determination of Inorganic Anions in Drinking Water by Ion Chromatography, US EPA NERL, ORD, Cincinnati, Ohio. 16.6 US EPA, 1994. Determination of Inorganic Anions by Ion Chromatography, Method 9056, SW-846, US Environmental Protection Agency, Office of Solid Wate and Emergency Response, Washington, DC. 17.0 Tables, Diagrams, Flowcharts, and Validation Data 17.1 NA 18.0 Document Revision History Original Document: April 2006 Version: 50A.0 Title: Standard Operating Procedure for the Analysis of Chloride and Sulfate in Fresh Waters by Ion Chromatography Edit Date: May 2008 New Version: 50B.0 Title and documentation changed to include bromide Section 1.1: change from 250 to 300 ul sample loop

Page 11 of 11 Section 7.2.1 and 7.2.2: change concentrations of calibration and check standards. Change from serial dilution to dilution. Change from individual standards to mixed standards. Edit Date: February 2010 New Version: 50B.1 Address update Section 1.1: change sulfate detection limit to 0.01 mg SO4/L Section 11.1: change sample loop size to 300 ul Section 11.2.2: change baseline conductivity to 25 us/cm Section 13.1: add Environment Canada Proficiency Testing Program participation. Edit Date: New Version: 50B.2 Section 1.1 & 7.2: add fluoride to routine analyses Section 5.2: add methanol Section 6.1: update suppressor information Section 8.1: add hold time Section 9.6: edit performance standard General edits throughout