CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS

Similar documents
OPTICAL PARTICLE SIZER MASS CALIBRATION METHOD

NANOSCAN SMPS SPECTROMETER COMPARED TO THE TSI SMPS SPECTROMETER

TSI AEROTRAK REMOTE 4-20 ma SENSORS LIFE SCIENCE

SIZE-SELECTED AEROSOL FILTER ANALYSIS USING ELECTROSTATIC CLASSIFICATION AND LIBS

CARBON NANOTUBE ANALYSIS USING THE TSI LIBS DESKTOP ANALYZER

Fundamentals of Particle Counting

MACROIMS MACROION MOBILITY SPECTROMETER MOLECULAR WEIGHT AND PARTICLE SIZE DETERMINATION OF POLYVINYLPYRROLIDONE (PVP)

Determination of particle size distribution Single particle light interaction methods. Part 4:

ELECTROSTATIC CLASSIFIER MODEL 3082

MEASURING NANOPARTICLE EXPOSURE

ISO INTERNATIONAL STANDARD

Particle number concentration standards at NMIJ, including a new calibration service with an inkjet aerosol generator.

SPECIFICATIONS PARTICLE SENSOR KS-28B Higashimotomachi, Kokubunji, Tokyo , Japan

ISO INTERNATIONAL STANDARD

Technical Paper 3 Verifying Particle Counter Calibration: Size Verification

Maintianing Validation and Calibration. Version 01. Rev 01.

Estimation of Mass with the Model 3321 APS Spectrometer

VIBRATING ORIFICE AEROSOL GENERATOR MODEL 3450

Micro-Channel Water Cooled Single Bar Diode Laser (QCW)

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

Is NIST SRM2806b Responsible for the Sudden Increase in Particle Counts you have Seen on your Oil Analysis Reports?

The Fundamentals of Moisture Calibration

PARTICLE MEASUREMENT IN CLEAN ROOM TECHNOLOGY

Latest Technology and Standardization Trends for Liquid-borne Particle Counters

of mass spectrometry

Heat Flow Calculations Made Using the Models 8386 and 8386A VELOCICALC Plus Air Velocity Meters Application Note ITI-124

Application of Micro-Flow Imaging (MFI TM ) to The Analysis of Particles in Parenteral Fluids. October 2006 Ottawa, Canada

ITS. One of the highest resolutions for particle sizing and counting. Multisizer 3 COULTER COUNTER SCIENCE

SOP SMPS BaltimorePM supersite Version /29/01 Page 1 of 13 SMPS

Model 3450 Vibrating Orifice Aerosol Generator

Mobility-Based Particle Size Classification and Detection down to 2nm Theory to Practice

Absolute activity measurement of 85 Kr with proportional counters

Standardization of Optical Particle Counters

Introduction to Fourier Transform Infrared Spectroscopy

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

ENVR 116 Aerosol Technology Laboratory Session Fall 2005

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

Controlled Environment Testing Report

Calibration checks of particle counter using primary and other techniques at the laboratory level

The new 11-R (one part of the Mini WRAS) and GRIMM MINI-WRAS Andreas Jaksch. Symposium Stockholm Mai

The solution for all of your

Differential Mobility Particle Sizer (Aerosol measurements)

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

The Elzone II Particle Count and Size Analyzer

AN1755 APPLICATION NOTE

Intercomparison of Mobility Particle Size Spectrometers

Introduction to Fourier Transform Infrared Spectroscopy

The Coulter Principle for Outlier Detection in Highly Concentrated Solutions

ISO/DTR 22681:2018 ISO/TC 131/SC 6/WG 1. 4 October Secretariat: BSI DTR 22681:2018

Spa & Wellness for Pipettes

Brian Sargent, Assistant Director, Designee 3/21/18. See Thread for Approval

Holographic Characterization of Protein Aggregates

Intercomparison of Mobility Particle Size Spectrometers

KNOWLEDGE

Performance and Control of the Agilent Nano Indenter DCM

Supplementary information

22.S902 IAP 2015 (DIY Geiger Counters), Lab 1

Extrel is widely respected for the quality of mass spectrometer systems that are

The Plus for your lab!

Figure 2. Size distribution and concentration of a cocktail of PSL particles measured by three different sizing instruments.

FSA TM Multi-bit Digital Processors

ISO STANDARD INTERNATIONAL. Cleanrooms and associated controlled environments Part 1: Classification of air cleanliness

TRACEABILITY STRATEGIES FOR THE CALIBRATION OF GEAR AND SPLINE ARTEFACTS

Rapid Measurements of Aerosol Size Distributions Using a Fast Integrated Mobility Spectrometer (FIMS)

HCF4089B BINARY RATE MULTIPLIER

Measurments with Michelson interferometers

Know thy Enemy : Understand and Control Particle Contamination. March 10th, 2008 Kirk Buecher

Alan Mortimer PhD. Ideas of Modern Physics

Determining the Optimum High Voltage for PMT s

FRITSCH "analysette 22" NanoTec The Result of 30 Years of Experience in Particle Size Measurement. 1 The Inventor of "Reverse Fourier Optics"

ISO defines the Rockwell superficial hardness value, HRN, as:

Traceability research activities in the field of airborne particle number/ mass concentration measurement Liu Junjie, Zhang Wenge, Song Xiaoping

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005. ALLOMETRICS, INC Bayport Blvd. Seabrook, TX Terry Baldwin Phone: CALIBRATION

ICP-MS. plasma 17OO. High Resolution Multi-Collector ICP-MS.

Certificate of Accreditation

Intercomparison of Mobility Particle Size Spectrometers

Technical Note A Quantification and NIST Traceability of Micromatter Thin Film Standards for XRF Analysis. Version March 16, 2017

Applications SU 9 V- R

HOW ADVANCED PYROMETERS INCREASE THERMAL PROCESS REPEATABILITY AND PRODUCT QUALITY

PhysioCare Concept. Make your Lab a better place

CSCB63 Winter Week 11 Bloom Filters. Anna Bretscher. March 30, / 13

PROCESS CONTROL BASIS FOR A COST-EFFECTIVE SELECTIVE SOLDERING PROCESS

For the rapid, sensitive and accurate measurement of Caspase 9 activity in cell and tissue lysates

Evaluating the functionality of Eppendorf BioPhotometer and Eppendorf BioSpectrometer using a Secondary UV-VIS Filter Set


To: Mitch Cottrell, Instructor From: Brian Garber, Crawford Binns, Jeff Hollis (Group 3, Sec. J) Subject: Project Deliverable #3 Date: 4 May 2010

Bring a printed copy of this lab to your lab section. We will answer the lettered questions in class.

UNIVERSITY OF MINNESOTA. This is to certify that I have examined this copy of a master s thesis by XIAOLIANG WANG

For the rapid, sensitive and accurate measurement of Caspase 3 Activity in cell and tissue lysates

Fast, Effective XPS Point Analysis of Metal Components

BRIDGE CIRCUITS EXPERIMENT 5: DC AND AC BRIDGE CIRCUITS 10/2/13

ab Cervix cancer tissue array of 102 cases (1.5mm)

IES LM MEASUREMENT AND TEST REPORT For

Caspase Substrate Kit (Colorimetric)

SCOPE OF ACCREDITATION TO ISO/IEC 17025:2005 & ANSI/NCSL Z & ANSI/NCSL Z

Applications SS 11 VL- R

Model 2300XP PSL & Process-Particle Wafer Deposition System

What can laser light do for (or to) me?

International Atomic Energy Agency. Department of Nuclear Sciences and Applications. IAEA Environment Laboratories

Comparing the Microlab 600 to Volumetric Glassware and Air Displacement Pipettes. January 2013

Transcription:

CALIBRATION CERTIFICATE FAILURE EXPLANATION FOR AEROTRAK PARTICLE COUNTERS APPLICATION NOTE CC-119 (A4) This application note works through the As Found Calibration Certificate of the AeroTrak Counters and explains common failures. An explanation of the failure mode, possible reasons, and impact are discussed for each failure a customer may see. Also included, is an explanation of Counting Efficiency and why a 0.3 µm instrument is not needed to measure 100% of the 0.5 µm particles, as commonly believed. This application note assumes that the reader already has a basic understanding of how particle counting technology works. For a more in-depth look at each elements of a particle counter calibration, please refer to Application Note CC-117. Calibration and Verification of Setting Nominal Gain Stage Previous Digital Cutpoint Calibration and Verification of Setting As Found Digital Cutpoint Measured Error Expanded Uncertainty 0.3 µm A 30 24 0.29 µm -4.8% ±10% Pass 3.9% 0.5 µm A 325 325 0.50 µm 0.0% ±10% Pass 3.9% 0.7 µm B 3 3 0.70 µm 0.0% ±10% Pass 3.9% 1 µm B 8 8 1.00 µm 0.0% ±10% Pass 3.9% 2 µm B 20 20 2.00 µm 0.0% ±10% Pass 3.8% 5 µm B 179 179 5.00 µm 0.0% ±10% Pass 3.8% This section of the Certificate shows the current sizing capabilities of the particle counter as compared to the last time it was calibrated. The Digital Cutpoint is a value associated with a given size of a particle. When polystyrene latex (PSL) particles of a known size are measured during calibration, the intensity of the signal interpreted by the optics is then associated with that size of particle (see Figure 1). During the As Found Calibration, PSL particles of a known size are again measured to see if the value previously found still matches. The difference between the Digital Cutpoints is the Error. There are two possible failure modes for this section.

Figure 1 If the Error fails on the positive side, it means that the Digital Cutpoint has risen since the last calibration (see Figure 2). One possible reason for this failure may be that the airflow through the system is too slow. s passing through spend more time in the laser beam, causing more light to scatter and a higher intensity signal to be interpreted. This failure s impact is that some particles measured will be interpreted as larger than they actually are. Figure 2 Page 2 of 6

If the Error fails on the negative side, it is because the Digital Cutpoint has fallen since the last calibration (See Figure 3). Common reasons for this failure are contaminated optics (some scattered light is lost), laser misalignment or laser power drop (not enough light hitting the particles), or airflow is too fast (particles do not spend enough time in the laser beam). This type of failure may mean that the lower end of the channel was undercounted, and the higher end of the channel was over counted. Figure 3 Counting Efficiency Counting Efficiency Actual 0.3 µm 32% 50% ± 20 Pass 0.5 µm 92% 100% ± 10 Pass Counting Efficiency is tested to ensure that the particle counter is accurately counting the number of particles that pass through it. The particle counter under test is compared to a reference instrument (see Figure 4). Counting Efficiency is tested at two particle sizes. The first size must be close to the minimum detectable particle size, and the second size is to be 1.5 to 2 times larger than the minimum detectable particle size. ISO 21501-4 requires that the counting efficiency near the minimum detectable size be 50 ± 20%, and the larger particle size be 100 ± 10% (the 50% requirement will be explained later in this application note). Figure 4 While very uncommon, a particle counter may have a high failure for the Counting Efficiency (the instrument under test (IUT) counted too many particles compared to the reference instrument). If this failure occurs, the most likely explanation is because the flow to the IUT was higher than the reference instrument, or there was a leak in the system allowing extra particles to get into the IUT. This would be a conservative error, meaning the instrument was possibly counting more particles than it should have been. Page 3 of 6

More commonly, if the IUT fails this aspect of the calibration, it is a low failure, meaning the IUT counted too few particles compared to the reference instrument. Possible reasons for this failure are low flow to the IUT, contaminated or misaligned optics, or low laser power. The impact of the IUT counting too few particles is that the environment may have been dirtier than the instrument interpreted. Resolution Resolution is the ability of the instrument to differentiate between similarly sized particles. It is calculated in relation to the standard deviation of the distribution measured with the pulse height analyzer. As the standard deviation of the distribution increases, the instrument s ability to differentiate sizes is reduced. Resolution Measured 0.5 µm 5.3% 15% Pass The only failure possible is a high failure, and is most likely caused by contaminated and/or misaligned optics. When this happens, the instrument is not as able to differentiate similarly sized particles. False Count Rate Sample Time (min) Sampled (L) Measured Counts (#) False Count Rate Concentration (#/m 3 ) 95% UCL (#/m 3 ) (#/m 3 ) 20 57 0 0.00 53.0 70.7 ±10% The purpose for the False Count Rate test is to ensure that the particle counter is not counting particles when it shouldn t be. A HEPA filter is placed over the inlet, and the instrument is set to the minimum detectable size. Ideally, no particles would be counted in this configuration. However, if too many particles are counted, and the test is failed, it may be because of a dirty instrument, leaks in the system (particles getting introduced into the flow somewhere other than the inlet), or particles stuck to the side of the flow system became dislodged and counted during the False Count Rate test. This is a conservative failure, and indicates that the particle counter potentially counted too many particles. Sampling Flow Rate The Sampling Flow Rate test is done to ensure that there is the correct flow of air through the instrument. If the particle counter fails the test because the flow is too high, it may be because the critical orifice is Sampling Flow Rate (L/min) Nominal Actual Error 2.83 2.83 ±5% Pass damaged (remotes only), or because the pump and/or pressure transducer is operating incorrectly. If a high failure has occurred, the particles may have been sized too small (in the laser beam for too little time). Additionally, since the flow was higher, there may have been more particles flowing through the system, which means the counting efficiency may have been too high. More commonly, if there is a Sampling Flow Rate failure, it is a low failure. This could be caused by a leak in the system, contamination, pump or pump electronics issues. This failure may cause particles to be measured as larger than they really are (in the laser beam for too long). Also, since there are fewer particles flowing through the system, the counting efficiency may be too low. Page 4 of 6

Measuring All of the 0.5 µm s A Simple Explanation Previous to ISO 21501-4 calibration standards, many particles counters had lower quality and questionable performance when measuring smaller sized particles. Now, ISO 21501-4 ensures that particle counters perform adequately, even at the smallest cutpoint. In the past, there was a general misunderstanding that if a 0.5 µm instrument counted only 50% of the 0.5 µm particles, then a 0.3 µm instrument would be needed to count 100% of the 0.5 µm particles. Misunderstanding of this concept, and of ISO 21501-4 calibration standards, led many competitors to upsell customers to more expensive particle counters. When a particle counter is calibrated, it uses Polystyrene Latex (PSL) particles of a known size with a known standard deviation from that size. Even though these particles are created in a lab to NIST traceable standards, nothing is perfect. If 0.5 µm PSL particles are used during calibration, some of them will invariably be smaller than 0.5 µm and some will be larger than 0.5 µm. When these PSL particles are introduced to the particle counter, they are sized and counted. The larger the particle, the higher the voltage of the signal interpreted by the instrument. Since the standard deviation of the PSL particle size is known, the median of the histogram is taken to be the threshold for the size of particle being measured (Figure 5), in accordance with ISO 21501-4. The 50% of counts above that threshold are larger than 0.5 µm and the 50% of counts below the threshold are smaller than 0.5 µm. Figure 5 This is why a 0.5 µm channel counting 50% of the 0.5 µm calibration PSL particles is nominal. The other 50% of the PSL calibration particles are smaller than 0.5 µm and should not be counted as 0.5 µm particles. When using an instrument with a 0.3 µm channel and a 0.5 µm channel, 50% of the calibration PSL particles will still be counted in the 0.5 µm channel, the remainder will be counted in the 0.3 µm channel because they are actually smaller than 0.5 µm and therefore not large enough to be counted in the 0.5 µm channel. Page 5 of 6

In summary, the 50% is referring to calibration PSL particles of a known size and standard deviation. In real world applications, a 0.5 µm instrument will still count all of the 0.5 µm particles in the sampled air. TSI, TSI logo, and AeroTrak are registered trademarks of TSI Incorporated. TSI Incorporated Visit our website www.tsi.com for more information. USA Tel: +1 800 874 2811 UK Tel: +44 149 4 459200 France Tel: +33 1 41 19 21 99 Germany Tel: +49 241 523030 India Tel: +91 80 67877200 China Tel: +86 10 8219 7688 Singapore Tel: +65 6595 6388 CC-119-A (3/12/2015) 2015 TSI Incorporated Printed in U.S.A.