Calibration of Volumetric Glassware. Prepared by Allan Fraser May 2016 APPLICATION Note 1

Similar documents
Practice Lab. Balances and calibration of volumetric tools

CHEM 334 Quantitative Analysis Laboratory

Chemistry Calibration of a Pipet and Acid Titration

Experiment 1. Determination of the Density of Water and an Unknown Solid Sample

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

2. MASS AND VOLUME MEASUREMENTS Mass measurement Analytical and standard laboratory balances Pre-lab Exercises

Spectrophotometric Determination of Ferrocyanide in Effluents

MEASUREMENTS AND ERRORS

Basic Equipments and Instruments used in Chemistry laboratory: Balance: It is an instrument for measuring mass.

STANDARD PROCEDURE: NWSP R2 (15) Polyacrylate Superabsorbent Powders Determination of the Particle Size Distribution by Sieve Fractionation

Density of an Unknown

Tex-620-J, Determining Chloride and Sulfate Contents in Soil

Chemistry 119: Experiment 6. Sampling and Analysis of a Solid Drain Cleaner

Ascorbic Acid Titration of Vitamin C Tablets

Density of Aqueous Sodium Chloride Solutions

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Chemistry 143 Experiment #11 Acid Base Titration Dr. Caddell. Titrating Acid

Functional Genomics Research Stream. Lecture: February 17, 2009 Masses, Volumes, Solutions & Dilutions

Chemistry 143 Acid Base Titration Dr. Caddell. Titrating Acid

LAB 15. Traceability: Volumetric Apparatus. Edition 3 February 2019 Draft for Consultation. LAB 15 / Issue: 3 Page 1 of 12

Lab 1: Safety Lab; Introduction to Volumetric and Weighing Techniques

Determination of Orthophosphate Ion

Chemistry Lab: Introduction to Measurement

Accuracy and Precision of Laboratory Glassware: Determining the Density of Water

CHEM Practice to be done before the lab. Experiment 9 Introduction to Volumetric Techniques II. Objectives

Name: Date: AP Chemistry. Titrations - Volumetric Analysis. Steps for Solving Titration Problems

Introduction to Spectroscopy: Analysis of Copper Ore

INDIANA DEPARTMENT OF TRANSPORTATION OFFICE OF MATERIALS MANAGEMENT. VERIFYING THERMOMETERS ITM No T

Technical Procedure for General Laboratory Equipment

Chemistry 101 Experiment 12B-ANALYSIS OF COMMERCIAL VINEGARS

Thermodynamics and the Solubility of Sodium Tetraborate Decahydrate

Experiment 8 Introduction to Volumetric Techniques I. Objectives

IODINE CLOCK REACTION KINETICS

In this lab you are asked to make a series of volume and temperature measurements and record the number of significant figures in each measurement.

QAM-I-116 Preparation of Labware

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Chapter 9. Volumetric Analysis

LAB 8: DETERMINATION OF ACETIC ACID CONTENT IN VINEGAR

Introduction to Spectroscopy: Analysis of Copper Ore

Determination of Orthophosphate Ion

Basic Equipment. 2D.1 Equipment for Measuring Mass

ANALYTICAL WEIGHING, LIQUID TRANSFER, AND PENNY DENSITY REPORT

Density of Aqueous Sodium Chloride Solutions

EXPERIMENT 1 Chemistry 110 LABORATORY SAFETY

# 12 ph-titration of Strong Acids with Strong Bases

Experimental Procedure. Lab 406

Identification Of The Common Laboratory Glassware, Pipettes And Equipment. BCH 312 [Practical]

Montgomery County Community College Document Number: MET DeKalb Pike Revision Number: 1

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

Chemistry 119: Experiment 4. Titrimetric Determination of Acetic Acid Samples And Preparation of Youden Plot

ISO INTERNATIONAL STANDARD. Paints and varnishes Determination of density Part 3: Oscillation method

Method 8326 (0.006 to mg/l Al 3+ ) Powder Pillows

Making Measurements. Units of Length

To become acquainted with simple laboratory measurements and calculations using common laboratory equipment.

SOIL ORGANIC CONTENT USING UV-VIS METHOD

Sodium Chloride - Analytical Standard

6. Common Laboratory Techniques

Chemistry 143 Dr. Caddell Laboratory Experiment 1

The Molecular Weight of Carbon Dioxide

To use calorimetry results to calculate the specific heat of an unknown metal. To determine heat of reaction ( H) from calorimetry measurements.

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

CHM101 Lab Measurements and Conversions Grading Rubric

The Thermodynamics of the Solubility of Borax

Determination of Total Bromine and Iodine Emission Spectrometric Method (ICP-OES) EuSalt/AS

Acid-Base Titration. M M V a

Assessment of Accuracy and Precision

ASTM Designation: D Standard Test Method for Determination of Iodine Number of Activated Carbon

Titration of Acetic Acid in Vinegar Minneapolis Community and Technical College v.9.17

PART I: MEASURING MASS

Core practical 11: Find the amount of iron in an iron tablet using redox titration

CHEM 30A EXPERIMENT 8 & 9: ACID- BASE TITRATION. Learning Outcomes. Introduction. Upon completion of this lab, the student will be able to:

(b) Write the chemical equation for the dissolution of NaOH in water.

Introduction to Spectroscopy: Analysis of Copper Ore

Experimental Procedure

Partner: Judy 29 March Analysis of a Commercial Bleach

Materials: Micropipettes (2-20 µl range pipette, µl range, µl range), tips, test tubes with color dye, well plates

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Measuring Length Equipment: Laser pointer, English ruler, millimetric scale screen, beaker, water, hexane, acetone, and methyl alcohol.

Safety Note: Safety glasses and laboratory coats are required when performing this experiment

Analyzing Hydrogen Peroxide Solution Chemistry 30

Reference method for the determination of real alcoholic strength by volume of spirit drinks of viti-vinicultural origin: measurement by pycnometry

EXPERIMENT 30A1: MEASUREMENTS. Learning Outcomes. Introduction. Experimental Value - True Value. 100 True Value

XHCO3(s) + HCl(aq) XCl(aq) + H2O(l) + CO2(g)

Test Method: CPSC-CH-E

To explore solubilities and reactivities of different metal ions. To identify ions present in unknown solutions using separation methods.

The Uncertainty of a Waterdraw Calibration vs. Gravimetric Calibration on Small Volume Provers

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

TITRATION STAKES 2018 INSTRUCTIONS TO TEAM MEMBERS

Product Safety Reference Manual. Book 5 - Laboratory Policies and Procedures C

Determination of Isocyanate Equivalent Weight (in toluene solution)

CALIBRATION OF LABORATORY GLASSWARE: BUTYROMETER, PIPETTES, LACTOMETERS, THERMOMETER AND BURETTES

Chem(Bio) Week 6 Expt F: Amino Acid pka from potentiometry. Identification of an Amino Acid through Potentiometric Titration

The Nitrofurantoin Capsules Revision Bulletin supersedes the currently official monograph.

Measurement and Lab Equipment Basics Guided Notes

Experiment 7A ANALYSIS OF BRASS

Experiment 2: Analysis of Commercial Bleach Solutions

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Advanced Unit 7: Chemistry Practical Examination (SET A) Candidates must have: Scientific calculator Ruler

ANALYTICAL TASK EXPERIMENTAL PROCEDURE

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 1900 DR 5000 DR The fill line is to the right.

Transcription:

Calibration of Volumetric Glassware Prepared by Allan Fraser May 2016 APPLICATION Note 1

TABLE OF CONTENTS TABLE OF CONTENTS... i LIST OF TABLES... ii 1. 2. 3. 4. 5. 6. 7. 8. Purpose... 1 Scope... 1 Principle... 1 Abbreviations and Definitions... 1 Material Types... 2 Range... 2 Apparatus and Equipment... 2 Volumetric Flask Calibration... 3 REFERENCES... 10 APPENDIX... 11 i

LIST OF TABLES Table 1. Density of pure water for various temperatures. Corrections for buoyancy for stainless steel weights and the change in the volume of the container have been applied as per Skoog & West (1982), page 727.... 6 Table 2. Required tolerances of A- and B class volumetric flasks (source: ASTM E-288-94).... 7 Table 3. Required tolerances of A-class pipettes (source: ASTM E-288-94).... 7 Table 4. Required tolerances of A-class burettes (source: ASTM E-288-94).... 7 Table 5. Example calculation for the calibration of a 100ml A-class volumetric flask of tolerance 0.08ml. 98.0336 ml is outside of the ±2s tolerance and is therefore excluded in the data evaluation.... 8 ii

1. Purpose The purpose of this procedure is to provide a method for the calibration of volumetric glassware. 2. Scope The scope includes volumetric glassware (i.e. designed to contain or deliver an accurate volume of solution), including pipettes, volumetric flasks and burettes. 3. Principle The weight of deionized water contained in the flask or pipette is used to determine the actual volume contained in the glassware. Water weight is corrected for density differences due to temperature. The method involves weighing a specific volume of water either dispensed or to contain. Corrections for buoyancy are very small and are therefore can be neglected. The temperature of the water is measured at time of the calibration after the water has been allowed to equilibrate to a constant temperature. Multiplication of the obtained mass of water at some temperature is converted to the volume it would occupy at 20 o C. The difference between the apparent volume (i.e. that at some temperature) and the true volume is the absolute error, which should be less than the glassware tolerance. If so, the glassware is for all practical purposes, acceptable for quantitative work. If the absolute error > the tolerance, a correction factor needs to be introduced to correct to the true volume. 4. Abbreviations and Definitions g mg m ml Vol. gram milligram mass millilitres Volume 1

5. Material Types All volumetric glassware for quantitative work requires calibration. 6. Range All volumetric glassware used for accurate delivery or to contain specific volumes needs to be calibrated. 7. Apparatus and Equipment 7.1 Analytical balance accurate to 0.1mg (0.001g) Analytical balance accurate to 0.1g. (For volumes of 100 ml or less, the balance used must be capable of measuring in increments of 0.001 g. For volumes greater than 100 ml, the balance used must be capable of measuring in increments of 0.1 g). 7.1.1 Thermometer (0-100 o C) High Purity Water Pipette Bulb Beaker or weighing boat 7.2 Cleaning of Glassware All glassware to be calibrated must be scrupulously clean before calibration. 1. The test for cleanliness of glass apparatus is that on being filled with deionised water, and the water discarded an unbroken film of water remains. 2. If the vessel is very dirty and greasy, it can be filled with a warm soap solution for 15 minutes, rising several times with deionised water, followed by 50% v/v hydrochloric acid and then rinsed several times with deionised water. 2

8. Volumetric Flask Calibration 1. 2. 3. 4. 5. 6. 7. 8. 9. Allow a few liters of deionised water in a beaker to equilibrate in terms of temperature by monitoring the temperature of a thermometer placed in the water. The water should be in the same environment in which the calibration of the volumetric apparatus is to be done. Record the equilibrated water temperature (experimental temperature). Record the serial number of the flask. If it does not already exist, add a serial number to the flask in permanent ink or by attaching a label. Wash, dry, weigh, and record the mass of the volumetric flask to be calibrated. Fill the volumetric flask to a few millimeters below the mark with deionized water. Fill to the mark carefully using a medicine dropper or pipette. Wipe dry the outside of the flask and then weigh. Record the weight of the flask with the water in it. Using Table 1, look up and record the density of the water corresponding to the temperature recorded in step 2. 10. Calculate the expected mass of the flask at the experimental temperature (W Tx ) as: V flask = W flask+water W flask ρ water [1] Where: V flask is the nominal volume of the flask (in ml); W flask + water is the weight of the water recorded in step 8 (in g); W flask is the weight of the flask recorded in step 4 (in g); and ρ water is the density of water recorded in step 9 (in gml -1 ) 3

11. Repeat steps 4 through 10 for at least six replicates. 12. Calculate and record the mean, standard deviation (s), and twice the standard deviation (2s) of the flask volume based on the replicates. These data are then entered into the MS Excel sheet Calibration Example 1. 13. Identify any aliquots for which the volume is greater than 2s from the average value. If any aliquots are greater than 2s from the average value remove this datum and treat as an outlier. 14. Calculate and record the flask absolute error (i.e. ignore the sign of the computation) based on the following formula: Absolute Error = Nominal flask volume Mean apparent volume [2] 15. Ensure that the absolute error as in [2] is within the limits of a Class A (or B, if that is being calibrated) volumetric glassware as listed in Table 2. 16. If the absolute error, calculated with the aid of Table 2 falls within the tolerance of the class of glassware, the glassware may be taken as accurate for all quantitative purposes. If however, the error is larger than the certified tolerance of the glassware, the simplest procedure is to affix a small label to the flask bearing the true volume and the date of calibration. It should be noted that the true volume should then be used in the calculation of concentrations made up in this flask. 17. Sign and date the data sheet as shown in Appendix 1. (Skoog & West, 1982) For pipettes the same procedure as described above is to be followed. With the exception of graduated cylinders, the tolerances for Class B glassware is typically twice that of a Class A glassware. (ASTM E694). 4

8.1.1 Example Calibration A 100ml A-class volumetric flask was found to give a mean value (of six replicates) of 99.430g of water at 24 o C. Refer to Table 5 for the data in this example. The data in Table 1 is used to calculate the volume occupied at this temperature and is found to be: 99.430g x 1 0.99732g/ml = 99.697ml The standard deviation of the six replicates is found to be: 0.82ml. Calculating the ±2s for the mean based on 2 x 0.82ml: (98.360 101.640) Since one of the values, 98.0336 is outside of the range, it is rejected as an outlier and not included in the mean and standard deviation calculations. A new mean excluding this datum is calculated along with a new standard deviation. The ±2s is now: (99.947 100.05) The absolute error is calculated and found to be within the tolerance of 0.08 for the 100ml volumetric flask: Absolute error = 100.00 100.041 = 0.041 Since the criterion of the absolute error < tolerance has been satisfied (i.e. 0.041<0.08), the 100ml volumetric flask is taken as accurate for all quantitative purposes. Should the absolute error > tolerance of the flask (e.g. 0.10>0.08), then a correction is made for the difference in volume. 5

Table 1. Density of pure water for various temperatures. Corrections for buoyancy for stainless steel weights and the change in the volume of the container have been applied as per Skoog & West (1982), page 727. Temperature ( C) Density (g ml -1 ) Temperature ( C) Density (g ml -1 ) 16.0 0.99897 23.5 0.99745 16.5 0.99889 24.0 0.99732 17.0 0.99880 24.5 0.99720 17.5 0.99871 25.0 0.99707 18.0 0.99862 25.5 0.99694 18.5 0.99853 26.0 0.99681 19.0 0.99843 26.5 0.99668 19.5 0.99833 27.0 0.99654 20.0 0.99823 27.5 0.99640 20.5 0.99812 28.0 0.99626 21.0 0.99802 28.5 0.99612 21.5 0.99791 29.0 0.99597 22.0 0.99780 29.5 0.99582 22.5 0.99768 30.0 0.99567 23.0 0.99757 6

Table 2. Required tolerances of A- and B class volumetric flasks (source: ASTM E-288-94). Capacity Tolerances ( ml) (ml) Class A Class B 5 0.02 0.04 10 0.02 0.04 25 0.03 0.06 50 0.05 0.10 100 0.08 0.16 200 0.10 0.20 250 0.12 0.24 500 0.20 0.40 1000 0.30 0.60 2500 0.50 1.00 Table 3. Required tolerances of A-class pipettes (source: ASTM E-288-94). Capacity (ml) Tolerance (ml) 0.5 0.006 1.0 0.006 2.0 0.006 5.0 0.01 10.0 0.02 20.0 0.03 25.0 0.03 50.0 0.05 100.0 0.08 Table 4. Required tolerances of A-class burettes (source: ASTM E-288-94). Capacity (ml) Tolerance (ml) 5.0 0.01 10.0 0.02 25.0 0.03 50.0 0.05 100.0 0.2 7

Table 5. Example calculation for the calibration of a 100ml A-class volumetric flask of tolerance 0.08ml. 98.0336 ml is outside of the ±2s tolerance and is therefore excluded in the data evaluation. Run No. Before calibration (ml) W(empty) g T,water o C Density(g ml -1 ) W, water + flask W(empty+water)- (W,empty) ml V. flask (ml) Range (±2s) 1 100 56.534 24 0.99723 156.26 99.723 100.000 99.947 100.053 2 100 56.886 24 0.99723 156.66 99.777 100.054 99.947 100.053 3 100 56.165 24 0.99723 155.94 99.778 100.055 99.947 100.053 4 100 56.321 24 0.99723 156.11 99.789 100.066 99.947 100.053 5 100 56.815 24 0.99723 154.58 97.762 98.0336 6 100 55.557 24 0.99723 155.31 99.753 100.030 99.947 100.053 Mean Volume= 100.041 s= 0.026 Criteria Tolerance (ml) = 0.08 Error (ml)= 0.041 TRUE Calibration factor= 0.9996 Corrected Vol.= 100.00 RSD ( % )= 0.026 ASTM Classification A 8

8.2 Uncertainties in Measurement Weight of water Density of water Error in sighting the level of the water Error in temperature reading 8.3 How to cite this application note: To cite this applications note, use the following: Fraser A.W. (2016). Volumetric Glassware Calibration. Applications note 1. www.allanfraserandassociates.com and the date that the article was accessed. 9

REFERENCES ASTM E288-94. Standard speciation for laboratory glass volumetric flasks. Skoog, D., West, D. (1982). Fundamentals of analytical chemistry. 4 th Edition. CBS College Publishing. Holt Saunders International Editions:725-728 10

APPENDIX Appendix 1. Document for recording of relevant information for volumetric flask calibration. The same format can be used for pipettes. Volumetric Flask Calibration Date: Calibration Due Date: Flask Identification Manufacturer: Nominal Volume: ml Serial Number: Step 3. Tolerance: ± ml Calibration of Flask Step 4 - Mass of Empty Flask: g Run Step 7 (g) Step 2 ( C) Step 9 (g ml -1 ) Step 12 (ml) Average Volume: Absolute Error: ml ml 1 2 3 4 5 6 Analyst: Chemist: Signature Reviewer Date Date 11