Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Similar documents
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Highly Sensitive and Rugged GC/MS/MS Tool

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Author. Abstract. Introduction

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Agilent G2350A Atomic Emission Detector (AED)

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

U.S. EPA Method 8270 for multicomponent analyte determination

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Agilent 5977A Series GC/MSD System

Improved Forensic Toxicology Screening Using A GC/MS/NPD System with a 725-Compound DRS Database. Application. Author. Introduction.

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Application. Gas Chromatography February Introduction

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

USP<467> residual solvents

Application Note. Abstract. Authors. Introduction. Food Safety

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Agilent Intuvo 9000 Gas Chromatograph System

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Detection of Chemical Warfare Agents by Transportable GC/MS

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Application Note. Abstract. Authors. Introduction

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

GC Analysis of Polybrominated Flame Retardants Application

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Useful GC/MS Sensitivity for Improved Environmental Analyses. GC/MS Agilent Restricted

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Introduction to Capillary GC

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

COOKBOOK FOR THE PERKINELMER SWAFER TM TECHNOLOGY FOR GC APPLICATIONS

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

brochure brief 5973Network Mass Selective Detector General Description and Configurations

Understanding Gas Chromatography

The Tool Box Getting More Productivity from Your GC and GC/MSD

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

Agilent G3212 GC-APCI Source

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Comprehensive Pesticide Screening by GC/MSD using Deconvolution Reporting Software Application

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Multi-residue analysis of pesticides by GC-HRMS

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Activity in the FID Detection Port: A Big Problem if Underestimated

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Expectations for GC-MS Lab

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

The Analysis of Organophosphate Pesticides by LC/MS Application

The Third Dimension of Information Vol. IV Applications with the Agilent Technologies Atomic Emission Detector. O H Cl. Fe Se

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Determination of EU-Banned Disperse Dyes by LC/MSD TOF Application

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of Volatile Substances Proof of Food Adulteration

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

A New Web-Based Application for Modeling Gas Chromatographic Separations

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Gas chromatography. Advantages of GC. Disadvantages of GC

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Introduction to Capillary GC

Application Note 032

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

Transcription:

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application Food Safety, Environmental Authors Chin-Kai Meng and Bruce Quimby Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract In this application note, a gas chromatography/mass spectrometry (GC/MS) system capable of providing up to four signals from a single injection is described. When a three-way micro-fluidic splitter is added to the end of the column, two additional signals from GC detectors can be acquired together with the MS data from a single injection. This multi-signal configuration provides: full-scan data for library searching, selective ion monitoring (SIM) data for trace analysis, micro-electron capture detector and flame photometric detector data for excellent selectivity and sensitivity from complex matrices. A combination of element selective detectors, SIM/Scan, and deconvolution reporting software makes a very powerful pesticide analysis system. Examples for trace-level compound quantitation/confirmation or for screening are discussed. Introduction Many laboratories in the world are analyzing pesticide residue levels in both foods and the environment to protect human health. The process usually involves homogenizing the sample, extracting the pesticides, and analyzing the target compounds with a Gas Chromatograph (GC) or a Liquid Chromatograph (LC) depending on the nature of the compounds. For GC amenable compounds, the traditional detectors are NPD (Nitrogen Phosphorus Detector), µecd (micro-electron Capture Detector), and FPD (Flame Photometric Detector) for their excellent sensitivity and selectivity. However, even with dual-column confirmation analysis, these GC detectors cannot be used to verify the identity of the compounds with high confidence. Full scan mass spectral data and library searching are typically used for final compound verification. However, full-scan analysis has a worse (higher) detection limit (DL) compared to selective detectors on a GC. To improve the DL, the technique selective ion monitoring (SIM) is often used. With SIM, the MS monitors only a few characteristic ions for each target compound within the retention time (RT) range that the target elutes from the column. By monitoring only a few specific ions, the signal-to-noise ratio (S/N) improves significantly. The ions monitored are time programmed in groups corresponding to the RTs of the targets. SIM analyses with closely eluting targets require precise alignment of chromatographic RTs with the time programming of SIM groups. The Retention Time Locking (RTL) technique can be applied to eliminate the need to adjust SIM group time-windows after column maintenance or replacement. In this application note, a GC/MS system capable of providing up to four signals from a single injection is described. The benefits of the multi-signal detection include: Confirmatory information Full-scan data for library search capability

Maximum sensitivity SIM data enables trace analysis Excellent selectivity µecd and FPD detect trace-level hetero-compounds from complex matrices Experimental A recent technical note describes Synchronous SIM/Scan, which takes advantages of the Performance Electronics in the 5975 inert MSD to get both SIM and full-scan signals in a single run without sacrificing performance [1]. The SIM method can be easily developed automatically using the ChemStation s AutoSIM tool [2]. By simply selecting a checkbox in the method, the SIM and fullscan data can be acquired together. The trade-off is giving up some cycles per second but gaining an additional signal (full-scan data or SIM data) for the whole analysis. With properly chosen acquisition parameters, for example, increasing the scan speed, the decrease of cycles per second is usually not significant and does not affect peak quantitation or the quality of results (for example, S/N). 2 Injection Split vent EPC At the end of the column, effluent flow is split three ways according to the length and diameter of the capillary tubing (restrictor) used. Figure 1. Splitter MSD (SIM/Scan) µecd FPD (P) A schematic of the multi-signal configuration. Note: the EPC flow adds to the column flow into the splitter. Besides the SIM/Scan data, the ChemStation software can simultaneously acquire up to two additional GC detector signals, for example, FPD (in phosphorus- or sulfur- mode) and NPD (nitrogenphosphorus detector) signals or both P- and S- signals from a dual-wavelength FPD (DFPD). See Figure 1. Figure 1 is a schematic for multi-signal detection. At the end of the column, a three-way micro-fluidic splitter was used to split the column effluent to different detectors [3]. For this study, an FPD and a µecd were installed. Notice on the figure that an Auxiliary Electronic Pneumatics Control (Aux EPC) gas channel was connected to the splitter to maintain the pressure at the end of the column so that the split ratios/flows are kept constant throughout a run. Figure 2 shows a close-up view of the microfluidic splitter installed in the GC oven. Figure 2. A close-up view of the micro-fluidic three-way splitter in the 6890 GC oven. The size of the micro-fluidic plate is 1.25 inches (3.2 cm) wide and 2.5 inches (6.4 cm) tall. The device was designed to eliminate the common problems of large thermal mass, excess dead volume, and leaky connection due to oven temperature cycling etc. The splitter's flow paths and connection points are laid out and etched onto a thin, stainless steel plate using photolithography and chem-milling technologies. The plate is diffusion bonded, mounted with column connectors, and surface deactivated, resulting in an integrated and compact micro-fluidic splitter. Metal ferrules are used at the connectors that are leak-free after temperature cycling and will not absorb solvents or sample matrix, improving sensitivity for trace analysis applications. Deactivated capillary tubing between the splitter and each detector was used as a flow restrictor. Aux EPC pressure and the restrictor dimensions were determined using a spreadsheet-like calculator program to achieve the proper split ratio among all detectors. The three-way splitter can easily turn into a two-way splitter when a connector is capped. Other advantages of a splitter include backflushing [3] and quick-swapping. The Aux EPC flow can be run-time programmed to a higher pressure, while at the same time the inlet pressure is lowered to near ambient. This causes the column flow to reverse direction, back-flushing the less volatile materials out of the split vent of the inlet. The Aux EPC on the splitter also allows column changing and inlet maintenance without cooling and venting the MSD. The splitter s flow paths and connection points were designed in such a way

that when the column fitting is removed, the helium gas from the Aux EPC purges the fitting, preventing air from entering the splitter/msd. See Table 1 for hardware details and settings. Table 1. Gas Chromatograph, Mass Spectrometer, and Three-Way Splitter Operating Parameters GC Agilent Technologies 6890 Inlet EPC Split/Splitless Mode Splitless, 1.0 µl injected (7683 ALS) Inlet temp 280 C Pressure ~27 psi (chlorpyrifos methyl RT locked to 16.596 min) Purge flow 50.0 ml/min Purge time 0.75 min Total flow 55.3 ml/min Gas saver Off Gas type Helium Inlet liner Siltek Cyclosplitter, 4-mm id, Restek p/n 20706-214.1 Oven Oven ramp C/min Final ( C) Hold (min) Initial 70 2.00 Ramp 1 25 150 0.00 Ramp 2 3 200 0.00 Ramp 3 8 280 15 Total run time 46.87 min (last standard elutes around 35 min) Equilibration time 0.5 min Oven max temp 325 C Column Agilent Technologies HP 5-ms, p/n 19091S-433 Length 30.0 m Diameter 0.25 mm Film thickness 0.25 µm Mode Constant pressure Nominal initial flow 2.5 ml/min Outlet Unspecified Outlet pressure 3.8 psi (Aux EPC pressure to splitter) Front detector (FPD) Phosphorus mode Sulfur mode Hydrogen flow: 75.0 ml/min Hydrogen flow: 50.0 ml/min Oxidizer flow: 100.0 ml/min Oxidizer flow: 60.0 ml/min Temperature: 250 C Oxidizer gas type: Air Mode: Constant makeup flow Makeup flow: 60.0 ml/min Makeup gas type: Nitrogen Lit offset: 2.00 Data rate: 5 Hz 3

Table 1. Gas Chromatograph, Mass Spectrometer, and Three-Way Splitter Operating Parameters (Continued) Back detector (µecd) Temperature: 300 C Mode: Constant makeup flow Makeup flow: 60.0 ml/min Makeup gas type: Nitrogen Date rate: 5 Hz Thermal AUX 2 Use: MSD Transfer line heater Initial temp: 280 C Pressure AUX 5 Gas type: Helium Initial pressure: 3.80 psi Initial time: 0.00 min (this value will follow oven ramp) MSD Agilent Technologies 5975 inert MSD Tune file Atune.U Mode Scan Solvent delay 3.00 min EM voltage Atune voltage Low mass 45 amu High mass 555 amu Threshold 100 Sampling 2 A/D Samples 4 Scans/s 2.89 Quad temp 150 C Source temp 230 C Three-way splitter Agilent 6890N Option 890, when installed on the GC during factory assembly Split ratio 10:10:1 MSD:FPD:µECD MSD restrictor 1.444 m 0.18-mm id Deactivated fused silica tubing FPD restrictor 0.532 m 0.18-mm id Deactivated fused silica tubing µecd restrictor 0.507 m 0.10-mm id Deactivated fused silica tubing Flow to MSD (at 280 C) 1.53 ml/min Flow to FPD (at 280 C) 1.53 ml/min Flow to µecd (at 280 C) 0.153 ml/min Makeup flow (at 280 C) 1.38 ml/min Software Used in this Application Note GC/MSD ChemStation G1701DA Deconvolution Reporting Software (DRS) G1716AA NIST Library G1033A AMDIS (included for free with the NIST library CD) 4

Results and Discussion Figure 3 shows four signals that were simultaneously acquired from a single injection of a pesticide mixture. Due to the high sensitivity of the µecd, the split ratios for the three detectors was set to MSD:FPD:µECD = 10:10:1. This split ratio distributes the sample of a 1-µL splitless injection of a 1-ppm (1000 pg/µl) sample to the different detectors as labeled in Figure 3. Scan: ~ 480 pg SIM: ~ 480 pg µecd: ~ 48 pg FPD (P): ~ 480 pg Figure 3. Signals acquired simultaneously from a 1-µL splitless injection of 1-ppm standard. The split ratios were MSD:FPD:µECD = 10:10:1. 5

Figure 4 shows the signals when the pesticide standard was diluted 100-fold in a produce matrix. The total ion chromatogram (TIC) from full scan was not shown due to the lack of sensitivity. The FPD(P) and µecd were able to detect all the pesticides spiked in this extract. For trace-level target compound analysis, the SIM signal can be used for quantitation and the GC signals used for further confirmation. 1800 1700 1500 1300 1100 900 700 500 300 SIM Produce extract 2688, spiked 5.00 10.00 15.00 20.00 25.00 30.00 35.00 700000 600000 FPD(P) 500000 400000 300000 5.00 10.00 15.00 20.00 25.00 30.00 35.00 1240000 1200000 1160000 1120000 1080000 1040000 1000000 µecd 5.00 10.00 15.00 20.00 25.00 30.00 35.00 Figure 4. Data of a produce extract spiked at 10 ppb. FPD and µecd were able to detect the respective standards spiked into the extract. 6

Another application for this multi-signal system is for screening. In screening, no target list is available for the analysis; therefore, SIM acquisition or MS/MS is not possible. Figure 5 shows three signals (no SIM) from a produce extract. 800000 700000 600000 500000 400000 300000 200000 100000 50000 45000 40000 35000 30000 25000 20000 15000 10000 110000 100000 90000 80000 70000 60000 50000 40000 30000 Produce extract 13-10927 SCAN FPD (P) µecd Figure 5. Full-scan, FPD(P), and µecd data for extract 13-10927. 7

The Deconvolution Reporting Software (DRS) [3, 4] found several pesticides in the TIC as shown in Figure 6. Figure 6. Report for extract 13-10927 generated from DRS. The possible pesticides in the sample were benzophenone, chlorpyrifos methyl, and thiabendazole. Propoxur and metamitron were not confirmed by both AMDIS and NIST; therefore, they were most likely false positives. Due to the complexity of the sample matrix and other interferences, it is sometimes difficult to get a high library match factor from peaks in the TIC, even after background subtraction. Therefore, element selective detectors would be very useful in providing the supporting information for compound confirmation. The multi-signal system was retention time locked, therefore, from the RT and the aligned peaks from the FPD(P) and the µecd responses, chlorpyrifos methyl (C 7 H 7 Cl 3 NO 3 PS) was confirmed. 8

It usually takes less than 3 minutes to turn off the FPD photomultiplier, swap the P-filter with the S-filter, and turn the photomultiplier back on. After the swap, adjust the detector gas flows to optimize the response in either P- or S- mode. A new injection of the same extract was made in FPD(S) mode. The FPD(S) result is shown with previously acquired signals in Figure 7. Two major peaks were seen on the FPD(S) chromatogram. From the peak RTs, they supported the presence of chlorpyrifos methyl and thiabendazole (C 10 H 7 N 3 S) respectively. Note that the full-scan TIC barely showed a peak for either compound, which made it impossible for traditional data analysis to identify both compounds. The FPD(S) mode is very selective, but it is not as sensitive as the FPD(P) mode. Although the µecd is very sensitive, it is not as selective as the FPD. A combination of GC detectors, SIM/Scan, and DRS makes a very powerful pesticide analysis system. Conclusion The Synchronous SIM/Scan provides users with library searchable full-scan spectra as well as trace level SIM data in a single analysis. When a threeway micro-fluidic splitter is added to the end of the column, two additional signals from element selective detectors can be acquired together with the MS data from a single injection. This configuration makes it very attractive for the analysis of trace-level pesticide residues in foods or environmental samples. This multi-signal configuration provides: full-scan data for library searching, SIM data for trace analysis, µecd and FPD data for excellent selectivity and sensitivity from complex matrices. In this application note, examples of µecd signal and FPD signal (P- or S- mode) were acquired together with the SIM/Scan data from a single injection for trace-level compound quantitation/confirmation, or for screening. 800000 700000 600000 500000 400000 300000 200000 100000 Produce extract 13-10927 SCAN 23000 21000 19000 17000 15000 13000 50000 45000 40000 35000 30000 25000 20000 15000 10000 Chlorpyrifos methyl Thiabendazole FPD (S) (new injection) FPD (P) Figure 7. Full-scan, FPD(S), and FPD(P) data for extract 13-10927. 9

References 1. Chin-Kai Meng, Improving Productivity with Synchronous SIM/Scan, Agilent Technologies, publication 5989-3108, www.agilent.com/chem 2. Harry Prest and David W. Peterson, New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods, Agilent Technologies, publication 5988-4188, www.agilent.com/chem 3. Michael J. Szelewski and Bruce Quimby, New Tools for Rapid Pesticide Analysis in High Matrix Samples Agilent Technologies, publication 5989-1716, www.agilent.com/chem 4. Philip L. Wylie, Michael J. Szelewski, Chin-Kai Meng, and Christopher P. Sandy, Comprehensive Pesticide Screening by GC/MSD Using Deconvolution Reporting Software Agilent Technologies, publication 5989-1157, www.agilent.com/chem www.agilent.com/chem For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2005 Printed in the USA July 6, 2005 5989-3299EN