Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400

Similar documents
Gentian Canine CRP Immunoassay Application Note for Siemens ADVIA * 1800/1650

Application Note for Gentian Cystatin C Immunoassay on. Cobas c501 1 (Optimised Volumes II), Roche Diagnostics 1 v06-september 2010

BÜHLMANN fcal turbo. Reagent Kit. Calprotectin turbidimetric assay for professional use B-KCAL-RSET-U. Intended use

Activity. This procedure is to be used in conjunction with the package insert for PLAC Test for Lp-PLA 2

CEDIA CYCLOSPORINE PLUS (LOW RANGE) APPLICATION BECKMAN COULTER AU480 /AU680 /AU5800

The NGAL Test Reagent Kit

Human anti-deoxyribonuclease B, anti-dnase B ELISA Kit

Canine Erythropoietin,EPO ELISA Kit

SIEMENS DIMENSION VISTA 1 The NGAL Test Reagent Kit

Human papillomavirus,hpv ELISA Kit

DRI Ecstasy Assay Application BECKMAN COULTER UniCel DxC and Synchron System(s)

Instructions for use. Chromogranin A ELISA TM E-9000

IgM (Canine) ELISA. For the quantitative determination of IgM in canine serum and plasma. For Research Use Only. Not For Use In Diagnostic Procedures.

IgG (Canine) ELISA. Please see Appendix A for Reference Serum information. For Research Use Only. Not For Use In Diagnostic Procedures.

DRI OXYCODONE APPLICATION BECKMAN COULTER DxC 700 AU

Instructions for use. Glycine ELISA BA E-2100

Histamine Research ELISA

Human rheumatoid factor (RF) antibody (IgM) ELISA Kit

ELISA QUALITY ASSURANCE Analytical Phase

IgD ELISA. For the quantitative determination of IgD in Human serum and plasma. Please See Appendix A for Reference Serum Information

Human Creatinine Serum Detection Kit

IgD ELISA. For the quantitative determination of IgD in Human Sera. Please See Appendix A for Reference Serum Information

Monkey Kidney injury molecule 1,Kim-1 ELISA Kit

CEDIA Cyclosporine PLUS High Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System, VITROS 5,1 FS and 4600 Chemistry Systems

Mouse Retinol Binding Protein 4 ELISA

Mouse IgE ELISA. For the quantitative determination of IgE in mouse serum or plasma. For Research use Only. Not For Use In Diagnostic Procedures.

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. Please see Appendix A for Reference Serum Information

Cystatin C ELISA. For the quantitative determination of cystatin C in human biological samples.

Human Aspartate aminotransferase, AST ELISA Kit

CEDIA Cyclosporine PLUS Low Range Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems

Mouse Tri-iodothyronine, T3 ELISA Kit

Rat Ferritin ELISA. For the determination of ferritin in serum and plasma of rats. For Research Use Only. Not For Use In Diagnostic Procedures.

Rat Beta-2 Microglobulin ELISA

Human hepatitis B virus surface antigen(hbsag) ELISA Kit

Human Mullerian Inhibiting Substance/Anti-Mullerian hormone (MIS/AMH)Elisa Kit

CANINE COMPLEMENT FACTOR 3 (C3) ELISA

IgE (Dog) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only.

Human anti-gliadin antibody (IgA)ELISA Kit

IgG (Rabbit) ELISA Kit

Canine brain natriuretic peptide,bnp ELISA Kit

CHGA (Human) ELISA Kit

CEDIA Tacrolimus Application Ortho Clinical Diagnostics VITROS 5600 Integrated System and VITROS 5,1 FS and 4600 Chemistry Systems

Beta-2 M Beta-2-Microglobulin

Intended Use: Diagnostic reagent for quantitative in vitro determination of Urea / Blood Urea Nitrogen in serum, plasma and urine by colorimetry.

C3 (Mouse) ELISA Kit. Catalog Number KA assays Version: 15. Intended for research use only.

HUMAN MYELOPEROXIDASE

Serotonin (Human) ELISA Kit

Instructions for use 5-HIAA ELISA BA E-1900

Mouse Glutathione S Transferase Alpha 1 (GSTa1)

Rabbit IgA ELISA Kit

CTX-II (Human) ELISA Kit

RayBio Human Thyroid Peroxidase ELISA Kit

Prealbumin (Mouse) ELISA KitI

Mouse cross linked N-telopeptide of typeⅠcollagen, NTX ELISA Kit

Human anti-ganglioside IgG antibody (GM1-IgG) ELISA Kit

Mouse Cholecystokinin (CCK) ELISA Kit

Mouse KIM-1 ELISA. For the quantitative determination of Kidney Injury Molecule in mouse serum, plasma, or urine.

Pig immunoglobulin E (IgE) ELISA Kit

IgE (Rat) ELISA Kit. Catalog Number KA assays Version: 05. Intended for research use only.

Human Papillomavirus Antibody (IgG) ELISA Kit

BIOO FOOD AND FEED SAFETY. Histamine Enzymatic Assay Kit Manual. Catalog #: Reference #:

Potassium Enzymatic Assay Kit (Liquid Stable) Catalog Number: BQ 010-EAEL

KIM-1 ELISA. For the quantitative determination of Kidney Injury Molecule in various biological samples.

IgA ELISA. For the quantitative determination of IgA in human serum and plasma. For Research Use Only. Not for Use in Diagnostic Procedures.

Canine Lipocalin-2 (NGAL) ELISA KIT

Mouse cyclooxygenase-2, COX-2 ELISA Kit

Mouse estrogen, E ELISA Kit

User Defined Assay Reference Guide

Sheep myeloperoxidase(mpo) ELISA kit

Aflatoxin M1 (AFM1) ELISA Kit

Human Epinephrine/Adrenaline, EPI ELISA Kit

Porcine Immunoglobulin E (IgE)ELISA Kit

Human cyclophilin A, CyPA ELISA Kit

Dog NGAL ELISA. For the quantitative determination of NGAL (Lipocalin-2) in dog biological samples. Cat. No. KT-546. For Research Use Only.

For In Vitro Diagnostic Use. Rx Only. Reviewed by Date Reviewed by Date

Pig Vascular Endothelial Cell Growth Factor A (VEGF-A) ELISA

Human amyloid beta peptide 1-42, Aβ1-42 ELISA Kit

The Human IgD ELISA KIT is a highly sensitive two-site enzyme linked immunoassay (ELISA) for measuring IgD in Human Biological Samples.

Revised 24 Nov rm (Vers. 15.1)

SHBG-ELISA

Hydrogen Peroxide Colorimetric Detection Kit

Human anti-myelin associated glycoprotein antibody (MAG) Ab ELISA Kit

Human placenta lactogen, HPL ELISA kit

Cat IgA ELISA. Cat. No. KT-755 K-ASSAY KAMIYA BIOMEDICAL COMPANY. For the quantitative determination of IgA in cat biological samples

Human prolactin, PRL ELISA Kit

TF (Bovine) ELISA Kit

Human cross linked N-telopeptide of typeⅠcollagen, NTX ELISA Kit

Porcine Factor-related Apoptosis(FAS) ELISA Kit

For in vitro diagnostic use only. For Rx use only. Reviewed by Date Reviewed by Date

IgA (Guinea Pig) ELISA Kit

Bradford Reagent, 5x

Mouse Myeloperoxidase ELISA

Date. Urea nitrogen measurements are used in the diagnosis and treatment of certain renal and metabolic diseases.

Human Anti-Ovary Antibody (IgG)ELISA Kit

CALC PRINCIPLE REF & REF A28937 & A ANNUAL REVIEW Reviewed by: Date. Date INTENDED USE

Human Lipocalin-2/NGAL ELISA

Human antidiuretic hormone/vasopressin/arginine vasopressin, ADH/VP/AVP ELISA Kit

Gyrolab Protein A Kit Quantification of residual Protein A ligands in the presence of excess amounts of IgG

Human Paraneoplastic pemphigus (PNP)antibody ELISA Kit. Catalog No. MBS (96 T)

PRINCIPLE. Kit Reorder # Kit Reorder # Kit Reorder # A17652 Kit Reorder # A ANNUAL REVIEW Reviewed by: Date.

Transcription:

v02-jan2014 Gentian Canine CRP Immunoassay Application Note for ABX Pentra * 400 Intended Use The canine CRP immunoassay on Horiba Medical's ABX Pentra 400 is an in vitro diagnostic test for quantitative determination of canine CRP in dog serum and plasma. The measurement of canine CRP is used in the diagnosis and treatment of inflammatory diseases in dogs [1,2,3,4,5]. Measuring range The recommended measuring range of the Gentian canine CRP immunoassay on ABX Pentra 400 is 15-300 mg/l, with a security zone up to 700 mg/l. Normal Values Healthy dogs have CRP concentrations <15 mg/l with the Gentian canine CRP method on ABX Pentra 400. An exact reference range cannot be determined as CRP concentrations in healthy dogs are below the LoQ of the Gentian canine CRP assay. Clinical Decision Limits The diagnostic specificity of canine CRP can be enhanced without seriously impairing diagnostic sensitivity by using a cutoff limit somewhere above the normal range [2]. Each laboratory should establish its own cut-off. Assay Reagents Materials Provided by Gentian Gentian Canine CRP Reagent Kit R1 Reaction buffer (45 ml) R2 Immunoparticles (10,5 ml) Gentian Canine CRP Calibrator Kit (6 levels, 0,5 ml per level) Gentian Canine CRP Control Kit Control low (0,5 ml) Control high (0,5 ml) All materials are ready for use. *Not available for individual sale. REF 1501 REF 1507* REF 1514* REF 1551 REF 1519 REF 1520* REF 1521* Calibrator Standardization Gentian canine CRP calibrator values are established on the basis of internal canine CRP reference material. No international standard is available for canine CRP. Calibration Stability The stability of the calibration curve has not been tested on ABX Pentra 400. Recalibration every 4 th week is generally recommended. Material Storage and Stability All materials provided for the Gentian canine CRP test must be stored at 2-8 C. The expiry date is printed on the labels. Using an Abbott Architect c4000, the on board stability of the Gentian canine CRP reagents was found to be at least eight weeks. Sample Material Recommended sample material is canine serum, canine heparinized plasma or canine EDTA plasma. Analyze the samples as fresh as possible, and mix them well in advance. Sample stability testing showed that canine CRP (in serum) was stable for 14 days at 4-22 C [6]. The samples can be shipped without any special cooling and must then be analyzed within 14 days after shipment. Samples have been tested and shown to withstand up to four freeze and thaw cycles [6]. Assay Procedure Assay Principle The canine serum or plasma sample is mixed with canine CRP immunoparticles. Canine CRP from the sample and the immunoparticles' anti-canine CRP aggregate. The complex particles created absorb light, and turbidimetric measurements of absorption are related to canine CRP concentration via interpolation on an established standard calibration curve. Results are automatically calculated by Pentra 400. Application Parameter Setup The application must be installed with the instrument settings provided for the Gentian canine CRP method. For instructions on how to install a new application, consult the instrument manual. Reagent Preparation The reagents provided are ready for use. Mix the reagents gently before placing them into the assigned reagents positions. The bottles provided can be used directly on ABX Pentra 400. Calibration Curve Establishment Use the Gentian canine CRP calibrator kit to establish a calibration curve as described in the instrument manual. A recalibration must be performed when a new calibrator lot and/or a new reagent kit lot is to be used. The assigned concentration values of the calibrators are lot dependent. The relevant values are stated in the Analytical Value Sheet provided with the calibrator kit. * Registered trademark of Horiba Medical. 1

Measured canine CRP concentration Linearity ABX Pentra 400 QC Controls The Gentian canine CRP control samples should be assayed every day the method is in use, to validate the calibration curve. Each control has an assigned concentration value range that must be met before measuring regular samples. The relevant concentration ranges are given in the Analytical Value Sheet provided with the control kit. If the measured concentrations are outside the valid range, repeat the control measurements. Recalibrate if necessary. If the calibration cannot be performed without error, or valid control values cannot be reproduced, contact the local distributor for support. Measuring Patient Samples When a satisfactory calibration curve has been established and the control values are within the valid concentration range, canine serum or plasma samples may be measured. Check that minimum sample volume is present in sample cups/tubes and assay the samples according to the instructions given in the instrument manual. Results The sample concentration of canine CRP is calculated automatically by the analyzer and presented in mg/l. Performance Characteristics ABX Pentra 400 Lower Quantification Limit The lower quantification limit (LoQ) of canine CRP is defined as the lowest sample concentration that can be measured with a total error (TE) < 29,6 % [7]. Total error is calculated from the pooled standard deviation of the sample's measured concentration and the bias between the sample's mean and theoretical concentration value. A sample of 7,7 mg/l had a total error of 18 % with the Gentian canine CRP method. The LoQ of the Gentian canine CRP immunoassay on ABX Pentra 400 is set to 8 mg/l. Linearity Dilution of a high canine CRP serum was performed, and the concentration of the resulting samples was measured. Recovery from expected concentrations was calculated. Based on interpolation, the assay is considered satisfactory linear in the range 10-300 mg/l. Dilution factor Expected concentration Measured concentration Recovery (%) 100 % 320,1 315,7 98,6 80 % 256,1 255,4 99,7 60 % 192,1 197,8 103,0 40 % 128,0 134,5 105,1 20 % 64,0 67,4 105,2 10 % 32,0 32,0 99,9 5 % 16,0 17,3 108,3 2,5 % 8,0 8,6 107,9 400 300 200 100 0 Imprecision Four samples (A-D) were assayed over eight days on an ABX Pentra 400 analyzer. Four other samples (E-H) were assayed on another Pentra 400 instrument. Each sample was measured in duplicate, twice a day. Sample ID Mean Within run CV (%) Between run CV (%) Between day CV (%) Total CV (%) A 25,4 7,43 0 1 3,24 8,10 B 63,8 2,08 1,65 2,97 3,98 C 121,1 1,34 1,56 2,61 3,32 D 257,8 0,76 0,41 2,00 2,18 E 7,2 12,13 7,84 6,14 15,7 F 58,4 1,85 2,14 4,57 5,38 G 103,9 0,85 0,88 1,79 2,16 H 272,1 0,68 1,83 1,05 2,22 1: According to statistical convention and EP5-A2, vol 24, No 25, the CV is set to zero, if a negative value is returned. Interference Canine CRP samples of about 30 mg/l were spiked with 10 g/l hemoglobin, 10 g/l Intralipid or 800 mg/l bilirubin. The samples were compared to corresponding control samples. None of the spiked samples demonstrated clinically significant interference defined as more than 10 % difference between test sample and control sample. Control sample Test sample Mean Linear fit y = 0,99x + 2,8302 R² = 0,9992 0 100 200 300 400 Expected canine CRP concentration Recovery (%) Hemoglobin (10 g/l) 29,0 29,1 100,5 Intralipid (10 g/l) 27,9 28,2 101,2 Bilirubin (800 mg/l) 28,7 30,4 105,8 Security Zone Samples with a canine CRP concentration up to 700 mg/l return a value above 300 mg/l (upper limit of linearity range) and can be sent to automatic diluted rerun. 2

Gentian - Mean Analytical Recovery The analytical recovery of the canine CRP assay was determined by adding fixed amounts of canine CRP to samples and calculating the recovery from theoretical concentrations. Sample ID Expected concentration Measured concentration Recovery (%) 1 64,0 63,8 99,7 2 124,8 129,0 103,4 3 187,8 186,6 99,4 Instrument Variation A total of 32 samples was assayed on both ABX Pentra 400, Abbott Architect c4000 and Beckman Coulter AU400 with the Gentian canine CRP immunoassay. A statistical analysis of the measured instrument variation was performed. Instrument Variation Pentra 400 vs. Architect c4000 Bland Altman bias 1,4 % Bland Altman 95 % limits of agreement -8,2 % to 5,3 % Passing Bablok slope 0,96 Passing Bablok intercept 2,36 Instrument Variation Pentra 400 vs. AU400 Bland Altman bias -3,6 % Bland Altman 95 % limits of agreement -4,1 % to 11,2 % Passing Bablok slope 1,01 Passing Bablok intercept 2,06 Method Comparison The Gentian canine CRP test was compared to the Randox human CRP immunoassay, used for canine CRP applications (with canine CRP calibrators). A set of 45 serum samples was assayed on both methods at the Swedish University of Agriculture Sciences Animal Hospital, using an Abbott Architect c4000. Results were subjected to statistical analysis. Method Comparison N = 45 Bland Altman bias -3,0 % Bland Altman limits of agreement -19,4 to 13,4 % Passing Bablok slope 0,91 Passing Bablok intercept 7,79 400 300 200 100 0 Shipping Damage Please notify your distributor if the product received is damaged. Symbols Key Lot number Temperature limitations Expiration date Consult instructions for use Manufacturer Manufacturer Gentian AS, Bjornasveien 5 N-1509 Moss, Norway TEL: +47 99 33 99 05 FAX: +47 69 24 09 62 http://www.gentian.no References Method Comparison: Gentian vs Randox Catalogue number Identity Passing & Bablok (I) fit (7.79 + 0.91x) 0 100 200 300 400 Randox - Mean Distributor scil animal care company GmbH Dina-Weissmann-Allee 6 68519 Viernheim, Deutschland Tel: +49 (0) 6204 7890 0 Fax: +49 (0) 6204 7890 200 E-Mail: info-de@scilvet.com [1] Ceron et al. Vet Clin Pathol. 2005; 34: 85-99 [2] Kjelgaard-Hansen; PhD Thesis. 2004 [3] Kjelgaard-Hansen et al. Vet Clin Pathol. 2003; 32: 81-87 [4] Yamamoto et al. Vet Res Comun. 1993; 17: 259-266 [5] Eckersall et al. Vet J. 2010; 185 (1): 23-27 [6] Hillström et al. Vet Clin Pathol. 2014; in press. [7] Kjelgaard-Hansen et al. Comp Clin Path. 2003; 12: 69-74 3

Instrument Settings Canine CRP ABX Pentra 400 General Parameters Characteristics: Sample type: Serum / Plasma Number of reagents: 2 Reagent: Reagent short name: C-CRP* Reagent number: 999* On board stability: --- Cassette Automatic Rerun: Post dilution: --- Post concentration: --- Results: Decimal positions: 1 Manual patient validation Pre-dilution Diluent Name: Factor: --- Incubation time (cycles): --- Linearity: Correlation: Low limit: High limit: Slope: Intercept: 15.0 300.0 1.00 0.00 Delta Check: Delta check validity: Absolute variation: Relative variation: --- --- --- *Defined by user. 4

Reference Check : Man / Default: Women: Child: Low: --- --- --- High: --- --- --- Rerun Range : Man / Default: Women: Child: Low: --- --- --- High: --- --- --- Calibration Parameters Pre-dilution: Type: Calibrator diluent: --- --- Factor 1 Factor 2 Factor 3 Factor 4 --- --- --- --- Factor 5 Factor 6 Factor 7 Factor 8 --- --- --- --- Checks: Reagent limit absorbance check Reagent range low: --- Reagent range high: --- Reagent blank limit absorbance check Blank range low limit: --- Blank range high limit: --- Control Required Control used 1: --- Control used 2: --- Control used 3: --- 5

Validity Backup: Backup time frame without calibration required Interval: 30 Time unit: Day Calibration: Calibration mode: Logit / Log5 Level: 6 Calibration factor: --- Run(s): 2 Dev_rep. (%): 10.0 Dev_c. (%): --- Calibrator used: C-CRP* Validity: On request Time validity Interval: --- Time unit: --- Factor Calibration: Low limit check: --- High limit check: --- Relative variation check: --- Analysis Parameters Cleaner: Cleaner solution: H 2O Before After *Defined by user. 6

Wavelength (nm): Primary wavelength: 600 Secondary wavelength: --- Blank: Reagent blank Diluent: H 2O Analysis Sequence: Cycle: Reagent needle: Volume (µl): Sample needle: Volume (µl): H 2O volume (µl): 1 R1 270.0 SAMPLE 3.0 10.0 19 R2 75.0 --- --- --- --- --- --- --- --- --- --- --- --- --- --- --- Mixing Speed: 80 Test Name: C-CRP* Channel: 999* Code: C-CRP* Local Code: C-CRP* Enable Modified on: N/A Calculation Parameters Correlation Factor: Slope: 1.000 Intercept: 0.000 Sample Limit Check: Sample limit (Δ OD): --- Sample limit cycle: --- 7

Reaction Direction: Reaction direction check Reaction direction: Increase Antigen Excess Check: Antigen excess limit (%): --- Antigen excess point: --- Definition Step A: Calculation type: End point First reading (cycle): 21 Last reading (cycle): 42 Reaction Limit Check: Reaction limit absorbance: --- Cycle: --- OD Deviation Check: Linear regression First point Last point r 2 threshold SD First point threshold SD Factor Definition Step B (N/A) Calculation type: End point First reading (cycle): --- Last reading (cycle): --- Reaction Limit Check Reaction limit absorbance: --- Cycle: --- 8

OD Deviation Check: Linear regression First point Last point r 2 threshold SD First point threshold SD Factor Definition Step C (N/A) Calculation type: End point First reading (cycle): --- Last reading (cycle): --- Reaction Limit Check Reaction limit absorbance: --- Cycle: --- OD Deviation Check Linear regression First point Last point r 2 threshold SD First point threshold SD Factor Definition Step D (N/A) Calculation type: End point First reading (cycle): --- Last reading (cycle): --- Reaction Limit Check Reaction limit absorbance: --- Cycle: --- 9

OD Deviation Check: Linear regression First point Last point r 2 threshold SD First point threshold SD Factor Formula Unit Parameters Unit: Conversion Factor: mg/l 1.0000 --- --- --- --- Unit: Conversion Factor: mg/l 1.0000 --- --- --- --- 10