Enhancement of Linearity and Response in Charged Aerosol Detection

Similar documents
Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Advances in Sample Preparation for Accelerated Solvent Extraction

Discover what you re missing

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Monitoring Protein PEGylation with Ion Exchange Chromatography

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

HILIC Method Development in a Few Simple Steps

Automated and accurate component detection using reference mass spectra

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Thermo Scientific ConFlo IV Universal Interface. Continuous Flow Interface. Isotope Ratio MS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Introduction to Fourier Transform Infrared Spectroscopy

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Paul A. Ullucci, Bruce Bailey, Ian N. Acworth, Christopher Crafts, Marc Plante Thermo Fisher Scientific, Chelmsford, MA, USA

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

Fast methods for the determination of ibuprofen in drug products

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

of mass spectrometry

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

Introduction to Fourier Transform Infrared Spectroscopy

Expanded capabilities in ammonium and amine detection

IN QUALITATIVE ANALYSIS,

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Corona Veo the Next Generation Charged Aerosol Detector

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

APPLICATION NOTE Goal. Introduction. Keywords

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

New Multi-Collector Mass Spectrometry Data for Noble Gases Analysis

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

[ Care and Use Manual ]

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Determination of Carbonyl Compounds in Workplace Air

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

[ OSTRO PASS-THROUGH SAMPLE PREPARATION PRODUCT ] The Simpler Way to Cleaner Samples

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Utility of the Charge Detector in Ion Chromatography Applications

Key Words Nanoparticles, spicp-ms, ICP-MS, Qtegra ISDS Software, AF4-ICP-MS

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Charged Aerosol Detection 101

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

Thermo Fisher Scientific, San Jose, CA; 2 Kelleher Lab, Northwestern University, Evanston, IL; 3

Chem 230, Fall, 2014 Homework Set # 4 LONG ANSWER SOLUTIONS

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

Charged Aerosol Detection: Factors for consideration in its use as a generic quantitative detector

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

Transcription:

Enhancement of Linearity and Response in Charged Aerosol Detection Christopher Crafts, Marc Plante, Bruce Bailey, Ian Acworth, Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: This work evaluates the application of a power transformation algorithm during data acquisition with a charged aerosol detection. Its effect on increased linear range, peak shape, response, resolution, signal-to-noise and limits of detection will be discussed. Methods: The effect of changing the power function variable was examined using standards ranging from nanogram to mid-microgram amounts on column. It was also evaluated using a range of analytes showing different physicochemical properties and chromatographic chemistries. Results: Linearity was improved over the mass range analyzed with correlation coefficients generally >0.999 or greater. Significant improvements in signal-to-noise and resolution along with decreases in peak asymmetry and peak width were also observed but were dependent upon the algorithm value used. Introduction Raw data collected by analytical detectors rarely correlates directly to the data points that are analyzed by the chromatographic software. Rather the data has already been manipulated in some way within the detector, typically post data collection. The most common examples are filtering and smoothing algorithms. The power transformation is simply another internal data manipulation aimed at improving the outputted data quality without changing the true separation. The use of power transformation can often improve the appearance of non-linear detectors. While these settings are typically not fully disclosed or described to the end user, they may be implemented to improve the appearance or linearity of the results. The Thermo Scientific Dionex Corona ultra RS Charged Aerosol Detector (CAD ) is a highly sensitive universal detector that can deliver near uniform response for nonvolatile analytes. 1 Charged aerosol detection works by nebulizing the HPLC eluent into fine droplets that are then dried to form particles. The size of these particles directly correlates to the mass injected. After drying particles then collide with a stream of charged nitrogen and the charge is transferred to the surface of the particle (Figure 1). The charge on the particles continue is collected and converted to current using a highly sensitive electrometer. Charged aerosol detection is non-linear as the amount of analyte being measured depends on the charge that can be placed on the surface of the particle. This in turn is dependent upon the volume of the particle being formed. The relationship between particle surface area and volume is not linear. This is observed experimentally as the response curves for the Corona ultra RS Charged Aerosol Detector are typically expressed as the 2 nd order polynomial function over a range of up to 4 orders of magnitude. 2 The application of power function can now correct for this physical characteristic and deliver a more linear response over a larger dynamic range. FIGURE 1: Mixing Chamber of the Corona ultra RS Charged Aerosol Detector. Analyte Particles with Charges Charges remain on surface of analyte particles. Particles remain intact and do not ionize. The more surface area. The more charge is carried by the particle Dry Analyte Particles Charged Nitrogen Gas A power function is a simple mathematic transformation where the individual data points are raised to a preset value. The possible values are between 1.0 and 2.0 for the Corona ultra RS detector. This power function value (PFV) can vary depending on the chemistry in use and the mass range being analyzed. While its primary function is to increase linear range, several other benefits may result from a power transformation 2,3,4. Data are presented to demonstrate the impact of the power function on linearity, peak width, resolution, and signal-to-noise (S/N). 2 Enhancement of Linearity and Response in Charged Aerosol Detection

Methods Sample Preparation Standards of 14 compounds were dissolved in appropriate solvents at concentration of ~ 2 mg/ml. These compounds were then diluted using serial dilution. The mass range tested was from ~50 ng to as high as 14 µg on-column (7 to 10 point response curves) depending on the analysis in question. Further detail on the analysis range can be found in the figures below. Liquid Chromatography Thermo Scientific Dionex UltiMate 3000 RSLC system along with a Corona ultra RS Charged Aerosol Detector was used for all data generation. Exact analytical conditions varied for the different compounds analyzed and are not relevant to the data presented here. Method conditions are available upon request. Data Analysis Thermo Scientific Chromeleon Chromatography Data System 6.8 SR.11 (CDS) was used for all data collection and processing. The power function value (PFV) was set in the program method conditions and stored in the audit trail. The response curves were generated at a minimum of 6 different PFVs, with most being run at 7. This resulted in the generation and processing of ~100 response curves. Results Linearity All 14 data sets showed an increase in linear dynamic range. The PFV that returned the greatest correlation coefficient was then listed as the optimal value. Figure 2 shows the response curves for bovine serum albumin (BSA) sample analyzed using a monolith column chemistry on the CAD without a power transformation (PFV=1.0) and with a PFV of 1.7. The data without the PFV=1.0 was processed with a 2 nd order polynomial fit (quadratic in CDS software), while the PFV=1.7 was processed with a standard linear regression. Neither used the origin as a data point. The results showed an improvement in correlation when using the power transformation over the 2.5 orders of magnitude analyzed. FIGURE 2: Nine point response curves for bovine serum albumin. The optimal PFV value and the resulting correlation coefficient for all samples analyzed over the full range are presented in Table 1A. The average PFV for these samples was 1.55. The values in green represented data points that had acceptable back calculations from the linear equation of the response curves. While the R 2 values for all 14 curves were greater then 0.995, many of the lower level back calculations were not within the 5% range and are colored according to the percent deviation. The highest PFV used was 1.7, which was not high enough for the pyridoxine sample resulting in poor back calculations. Overall, the samples showed a similar trend: good correlation for high mass concentrations but declining when decreasing by two or three orders of magnitude. In Table 1B, the same data are analyzed but excluding the highest two levels. Over this reduced mass range the average of the optimal PFVs was 1.38. The majority of back calculations now fell within the acceptable range and the correlation coefficients improved. 3 Enhancement of Linearity and Response in Charged Aerosol Detection

The linearity and back calculation results demonstrate some of the advantages of the power transformation, as well as some guidelines for use. The use of a large PFV can give the appearance of linearity over a wide dynamic range but it is always important to understand how each point is weighted over the entire range. Note, with any curve over two orders of magnitude, the points at the top of the curve will have a greater impact on the correlation coefficient than those at lower levels. This can result in high R 2 values but poor back calculations. If this function is intended to be used to measure low levels of analytes based on area percent, the quality of the curve must be evaluated. The major improvement noticed in this work was the ability to obtain linear response over a larger targeted range than previously available. If a curve still does not correlate after a power transformation has been used (i.e., the PFV value was either too high or too low), a second order polynomial fit can be used (4 data points or more required). In most cases, this will provide an improved fit over the use of the 2 nd order polynomial alone (data not shown), with the added benefit of additional chromatographic improvements (see below). Table 1: Results observed at the optimal power function value over (A) a large range of 2 orders of magnitude or greater. (B) smaller ~ 1.5 orders of magnitude range. The color key to the correlation coefficient (R 2 ) and back-calculated percent recovery is shown below. 1A Primidone Hydrocortisone Ketoprofen Warfarin Progesterone Pyridoxine Procainamide C18 Diclofenac BSA Diclofenac C18 Erucic Acid Stearic Acid Palmitic Acid Myristic Acid Power Function 1.6 1.5 1.5 1.5 1.4 1.7 1.7 1.6 1.7 1.7 1.4 1.6 1.6 1.3 R2 Value 0.9992 0.9996 0.9990 0.9997 0.9996 0.9952 0.9999 0.9991 0.9993 0.9999 0.9996 0.9996 0.9992 0.9993 Std 1 99.9% 100.6% 100.9% 100.5% 99.4% 97.3% 100.0% 99.4% 98.9% 99.8% 100.1% 99.9% 99.5% 100.2% Std 2 101.7% 98.1% 97.8% 98.9% 102.5% 107.8% 100.5% 103.9% 103.9% 99.8% 100.8% 101.5% 103.3% 98.7% Std 3 97.2% 98.3% 95.0% 97.9% 101.2% 113.5% 99.7% 97.2% 102.5% 101.9% 96.4% 96.5% 98.8% 99.1% Std 4 87.4% 95.5% 93.0% 93.8% 94.3% #N/A 96.7% 91.8% 98.1% 99.3% 93.4% 95.5% 90.7% 108.7% Std 5 91.2% 99.2% 98.0% 98.1% 88.2% 90.3% 93.8% 85.8% 94.3% 93.8% 93.7% 90.7% 84.4% 122.1% Std 6 119.2% 113.3% 127.6% 114.5% 98.1% 65.2% 94.4% 96.9% 84.5% 88.1% 101.4% 95.7% 93.1% 112.1% Std 7 192.0% 161.9% 200.1% 165.5% 134.5% 4.2% 115.2% 131.4% 77.4% 93.6% 137.3% 131.5% 133.0% 43.0% Std 8 #N/A #N/A #N/A #N/A #N/A -79.6% 174.3% 214.5% 88.5% 119.6% 223.0% 225.3% 234.6% -59.3% Std 9 #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A 191.5% #N/A #N/A #N/A #N/A 1B Primidone Hydrocortisone Ketoprofen Warfarin Progesterone Pyridoxine Procainamide Diclofenac BSA Diclofenac C18 Erucic Acid Stearic Acid Palmitic Acid Myristic Acid Power Function 1.2 1.3 1.2 1.2 1.2 1.7 1.6 1.3 1.7 1.6 1.2 1.4 1.3 1.5 R2 Value 0.9999 0.9999 0.9999 0.9997 0.9997 1.0000 0.9999 0.9999 0.9998 0.9999 0.9997 0.9994 0.9998 0.9988 Std 1 #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A Std 2 #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A Std 3 99.8% 100.0% 100.1% 99.5% 100.1% 100.0% 100.2% 99.8% 100.4% 100.1% 99.5% 99.4% 99.8% 99.7% Std 4 100.8% 100.6% 100.3% 102.3% 100.3% #N/A 99.6% 101.3% 98.5% 99.5% 102.3% 103.0% 101.3% 99.7% Std 5 101.4% 97.5% 97.2% 100.9% 96.5% 98.9% 98.6% 98.9% 99.3% 100.4% 100.1% 98.5% 97.7% 106.8% Std 6 97.0% 98.6% 99.7% 96.5% 97.8% 102.1% 97.4% 98.1% 99.0% 97.4% 95.3% 94.8% 97.0% 104.8% Std 7 97.6% 107.8% 108.2% 93.9% 113.9% 99.6% 110.9% 101.8% 110.5% 105.6% 98.7% 101.6% 107.3% 70.5% Std 8 #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A Std 9 #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A #N/A < 80% 80% - 90% 90% - 95% 95% - 105% 105% - 110% 110% - 120% >120% Signal-to-Noise and Limits Testing The impact of applying a power transformation to a number greater than 1 is to increase its value (e.g., with a PFV=1.3; 2 becomes 2.5, 20 becomes 49.1, and 200 becomes 280.2). However, the impact of applying that same power function to a value less than 1 is to decrease its value (e.g., 0.2 becomes 0.123). This mathematical principle along with the need to apply a scaling factor in order to keep all the data within the range of the detector, results in a dramatic decrease in the observed noise. While this has a greater impact on larger peaks, smaller peaks can also benefit. The degree to which the S/N was affected for the entire analytical range was analyzed. The signal-to-noise and peak width at both half height and base were measured and evaluated for every sample point at all the PFVs used above. This resulted in more than 750 S/N data points. For all 14 samples and across all mass ranges, the S/N was found to increase as the PFV increased. In general it was observed that the S/N was 1-2 times greater at the low end and 50-1000 times greater for high mass values (with a PFV of 1.5). However, the effect on S/N can vary depending on the baseline noise and gradient effects. For example, Figure 4 shows data for stearic acid analyzed as part of a mix of fatty acid standards using a water/acetonitrile to acetonitrile/ipa gradient. The S/N for the 250 ng on-column injection improved by a factor of 31 as shown. 4 Enhancement of Linearity and Response in Charged Aerosol Detection

FIGURE 4: Analysis of stearic acid with PFVs of 1.0 and 1.6 with peak attributes shown in table. PFV = 1.0 PFV = 1.6 Peak Width @ 50% Signal/Noise Ratio (~250 ng O.C.) Peak Asymmetry (ep) Resolution (ep) PFV = 1.0 1.62 sec 662 / 1 1.14 9.2 PFV = 1.6 1.23 sec 21000 / 1 1.08 12.0 It is important to ensure that no value is lost by application of the power function. Peaks that are near or below the limit of detection (LOD) can be harder to discern after a power transformation. For this reason, trace analysis of unknown samples should initially be analyzed with a PFV=1.0 in order to ensure accuracy of the data. This however does not mean that the power transformation has no value at low levels. Figure 5 shows the analysis of procainamide at a PFV of 1.0 and 1.7. The S/N for the none-power transformed data of 13ng on column procainamide is just above the limit of quantification (LOQ) at 18.8/1 but with the PFV=1.7, the S/N was 126.1/1. This 13 ng on-column amount represents a 0.1% w/w impurity level for a 13µg API injection. Setting system suitability requirements for S/N at 20/1 (2x LOQ) for a point below the release criteria is needed for a robust method. In this example the 0.05 w/w% level would now meet the system suitability requirements and result in a more robust method. FIGURE 5: Analysis of procainamide with power function values of 1.0 and 1.7 illustrating signal-to-noise improvements at low levels. PFV =1.0 PFV =1.7 Peak Width and Resolution The effect of the power transformation on other chromatographic results was also observed during the analysis of the linearity samples. The effect on peak shape also correlated as expected: greater impact on higher levels with the peaks becoming taller and narrower. This can be seen for the stearic acid peak (Figure 4), which showed a decreased peak width at half height of 24%. This narrowing has two effects on data quality. First, is an improved peak symmetry and second the narrowing of peaks throughout the chromatogram results in an improvement in the calculated resolution. The mathematical impact of power transformations on peak resolution is discussed in detail in work from Dasgupta. 2 This is experimentally shown again with the data from the fatty acid mix (Figure 4) but examined in a more complex-amino acid sample as well. 5 Enhancement of Linearity and Response in Charged Aerosol Detection

Figure 6 illustrates the analysis of 21 underivatized amino acids using the Corona ultra RS detector at PFVs of 1.0, 1.5 and 2.0. The two areas highlighted in red accounted for the regions with poorest peak resolution. Peak resolution was calculated by the software using the EP method with calculations based on next integrated peak. The resolution increased as the selected PFV increased with an average increase of 20% at PFV=1.5 and 39% at PFV=2.0. This improvement in the observed data for linearity, signal-to-noise, and resolution can be implemented without negatively impacting data quality or accuracy for these samples. FIGURE 6: Analysis of 21 underivatized amino acids using reversed phase chemistry from 0.4% heptafluorobutyric acid to 50% acetonitrile at multiple PFVs. The peak resolution was calculated for all peaks highlighted in red and the results are shown in table below. Conclusion Implementation of a power transformation to raw data with charged aerosol detection results in improvements in observed linear range, signal-to-noise, and peak resolution, along with reduction in peak width and asymmetry. The power function value should be chosen based on the linear range needed and experimental results as one preset value is not suited for all analyses. The improvement in signal-to-noise at levels above the limit of quantification provides a powerful tool for improving reproducibility and quality when doing limits testing with the Corona ultra RS charged aerosol detector. The data presented here illustrated that use of power transformation in the Corona ultra RS is auditable and can improve data quality. References Resolution of Selected Peaks 2.85 min 3.06 min 5.30 min 5.42 min 5.57 min 5.81 min PFV = 1.0 2.99 0.88 1.62 1.61 2.77 0.78 PFV = 1.5 3.64 1.07 1.81 2 3.39 0.96 PFV = 2.0 4.15 1.28 2.17 2.28 3.79 1.11 1. Górecki, T.; Lynen, F.; Szucs, R.; Sandra, P. Universal Response in Liquid Chromatography Using Charged Aerosol Detection. Anal. Chem. 2006, 78, 3186 3192. 2. Dasgupta PK; Chen Y; Serrano CA; Guiochon G; Liu H; Fairchild JN; Shalliker RA. Black Box Linearization for Greater Linear Dynamic Range: The Effect of Power Transforms on the Representation of Data. Anal Chem. 2010 Dec 15;82(24):10143-50. 3. Shalliker RA; Stevenson PG; Shock D; Mnatsakanyan M; Dasgupta PK; Guiochon G. Application of Power Functions to Chromatographic Data for the Enhancement of Signal to Noise Ratios and Separation Resolution. J. Chrom A, 2010, 1217, 5693-5699. 6 Enhancement of Linearity and Response in Charged Aerosol Detection

www.thermofisher.com/dionex 2016 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. U.S./Canada (847) 295 7500 Brazil (55) 11 3731 5140 Austria (43) 1 616 51 25 Benelux (31) 20 683 9768 (32) 3 353 42 94 Denmark (45) 36 36 90 90 France (33) 1 39 30 01 10 Germany (49) 6126 991 0 Ireland (353) 1 644 0064 Italy (39) 02 51 62 1267 Sweden (46) 8 473 3380 Switzerland (41) 62 205 9966 United Kingdom (44) 1276 691722 Australia (61) 2 9420 5233 China (852) 2428 3282 India (91) 22 2764 2735 Japan (81) 6 6885 1213 Korea (82) 2 2653 2580 Singapore (65) 6289 1190 Taiwan (886) 2 8751 6655 PN70003_E 07/16S