Evaluating Leaks in a Soil Gas Sample Train

Similar documents
The Application of Method TO-15 to Naphthalene Measurements in Indoor Air

Vapor Intrusion Sampling Options: Performance Data for Canisters, Badges, and Sorbent Tubes for VOCs

Chapter 3: Source Measurement Techniques

TO-15 Checklist Determination of VOCs in Air by GC-MS

Introduction to IH Analytical Chemistry

An Innovative Field-Based Analytical Method For Low-Level Detection of Chlorinated Solvents in Groundwater and Soil Samples

Real Time On-Site Odor and VOC Emission Measurements Using a znose

Measurement of Acrylonitrile, Acetonitrile and Nitromethane in Ambient Air and Vehicle Exhaust by Gas Chromatography with Thermionic Specific Detector

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

TPH (Total Petroleum Hydrocarbons)

Detection and Identification of Stachybotrys chartarum

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Analysis of TO-15/TO-17 air toxics in urban air using TD GC/TOF MS and automated compound identification software

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

Atmospheric Analysis Gases. Sampling and analysis of gaseous compounds

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

AUTOMATED ONLINE IDENTIFICATION AND MONITORING OF IMPURITIES IN GASES

METHOD 3520B CONTINUOUS LIQUID-LIQUID EXTRACTION

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

SAB Soil Vapour Forum July 8, Overview of Soil Vapour Assessment From A Laboratory Industry Perspective

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

ASTM D5116 Emissions Test Report with CDPH SM V Loading Scenarios per Cradle to Cradle Standard v3.0 Sect. 5.8

Experiment C-15 Distillation - part 1

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Solvothermal Technique Guidelines

Safe Method of Use 2 Workshop and Arts Studios

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Semi-continuous Benzene Monitoring in Burlington, Vermont July 2007-June 2008

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Fast Determination of Impurities in Propane- Propylene Streams Using a Pulsed Flame Photometric Detector (PFPD) and a New Capillary.

GAW Expert Workshop VOC, C. Plass-Dülmer et al., FEHp-B 1

A- Determination Of Boiling point B- Distillation

METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION

Raw Material Finished product By product Exposure standards Length of shift Physical environment

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Characterization Survey Techniques and Some Practical Feedback

Key points. Discuss why contact lenses are not allowed in the chemical laboratory. Short pants Skirt Scandals Open-toed shoes

ICE MANTLE MAKER MODEL 452 User Maintenance Manual/Handbook

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Adding these quantities together gives the total kilogram moles.

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

SYNOPSIS OF CHANGES TO BC LABORATORY MANUAL, MARCH 2017

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

METHOD 3510B SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

Chlorine, Total. USEPA DPD Method 1 Method to mg/l as Cl 2 Chemkey Reagents. Test preparation. Before starting.

10 May Rod Walls - Director Facilities New Construction, Sustainability Hays CISD IH 35 Kyle, Texas 78640

Maine DEP DRO. MY05TP220A(0-0.5) MY05TP222A(0-0.5) FD MY05TP223A(0-0.5) FD MY05TP220A (EB) EB-Equipment Rinsate Blank FD- Field Duplicate Samples


Electrolysis SCIENCE TOPICS PROCESS SKILLS VOCABULARY

Chlorphosphonazo Colorimetric Method Method to 1000 µg/l Ca and Mg as CaCO 3 (ULR) Solution Pillow. Instrument Adapter Sample cell

METHOD 7B - DETERMINATION OF NITROGEN OXIDE EMISSIONS FROM STATIONARY SOURCES (ULTRAVIOLET SPECTROPHOTOMETRIC METHOD)

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

January 19, 2012, Workshop on Chemical Data Reporting (CDR) Rule Case Studies for Byproduct/Recycling Reporting

A Micro-GC Based Chemical Analysis System

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

HPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

A STUDY OF VOLATILE ORGANIC CARBON POLLUTION ON A UNIVERSITY CAMPUS DUE TO TRAFFIC

Nitrogen, Total Inorganic

Facilities Management

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Lab 1: Safety Lab; Introduction to Volumetric and Weighing Techniques

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Title of experiment and short description of the purpose of the experiment.

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

STANDARD OPERATING PROCEDURES

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Determination of Carbonyl Compounds In Water by Dinitrophenylhydrazine Derivatization and HPLC/UV*

GUIDANCE DOCUMENT. ESTCP Project ER

Accelerator Facility Accident Report

Examination Heat Transfer

FVBSS fixed volume bottle

Soil Sampling Results Former Truck Maintenance Garage

VSMOW Triple Point of Water Cells: Borosilicate versus Fused- Quartz

Chemical Hygiene Plan for Laboratories

3. Chemical industry. Because of their modular design, the instruments in the TOC-L series can be equipped for any possible measurement

Chlorine and Hydrogen Chloride Monitoring Utilizing Ion Mobility Spectroscopy (IMS)

Working with Hazardous Chemicals

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

Gases. and all other variables are directly proportional. a. From what laws is this equation derived?

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Sampling. Information is helpful in implementing control measures for reducing pollutant concentration to acceptable levels

of uranium hexafluoride

GUIDELINES FOR THE SAFE USE OF PYROPHORIC LIQUID REAGENTS

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

R021. BioTector TOC Analyzer Reagents_Mixer Reactor Systems

Transcription:

Evaluating Leaks in a Soil Gas Sample Train Paper #45 Extended Abstract Diane J. Benton Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 Nathan S. Shafer Air Toxics, Ltd., 180 Blue Ravine Rd. Ste. B, Folsom, CA 95630 ABSTRACT The current advisory for active soil gas investigations in California 1 requires leak tests to be conducted at every soil gas probe while taking a sample. A leak may allow ambient air into the sample train thus diluting the soil gas and underestimating the results. Also, the leak check compound itself may have impurities thereby contaminating the sample. The advisory suggests leak check compounds and lists potential leak points in the sample train that should be investigated. If a leak has been detected but unable to be resolved, the soil gas probe is decommissioned, and a new soil gas probe is installed, or consultation with agency staff required. Since the leak check results may draw soil gas data into question, the presence of any leak check compound in the sample is a concern. This paper investigates the response of several leak check compounds at determined leak rates into the soil gas sampling train. The diffusion of leak check compounds through the walls of commonly used tubing for soil gas investigation is also evaluated. Finally, considerations for performing leak checks and interpreting leak check data are presented. INTRODUCTION In order to insure the highest quality data for soil gas measurements, leak checks are often performed by exposing the potential leak sites to a known compound during sample collection. The compound is then analyzed for in the soil gas sample to determine if the sample was compromised by ambient air intrusion. A wide array of leak check compounds are used in active soil gas investigations. The presence of a leak check compound in soil gas data often raises more questions for the data user than answers on how to interpret the results. Understanding the relationship between a leak and concentration of a leak check compound, the role of diffusion of leak check compounds through sample train tubing, and potential for interference by the check compound with VOC measurements is important in answering these data usability questions. APPROACH Definition of a Leak

Leaks which may occur at fittings and connections of a sampling train exist on a continuum from negligible to severe. To evaluate the presence and/or extent of the leak, check compounds are used to test the potential leak points during sample collection. When the sample is analyzed by the laboratory, the check compound is measured and reported with the soil gas results. The question for the data user is how to translate a leak check concentration to a determination of data usability. Different interpretations regarding the definition of leak are suggested by various agencies. For example, the state of California Advisory-Active Soil Gas Investigations recommends a detection limit of 10 µg /L or less for a leak check compound. 1 A detection limit of 0.1 µg /L for the leak check compound has also been suggested in subsurface vapor sampling guidelines. 2 To model the impact of differing leak rates on concentrations of the leak check compound in the sample, experimental calculations were performed using the following two equations. Examples are presented using varied leak rates and several common leak check compounds. For these calculations, the sample flow rate was fixed at 100 ml/min through the entire sampling period. Equation 1. Concentration (ppmv) = (Leak Volume (ml)) * (Purity of check standard) * 1,000,000 (Sample Volume (ml)) Where: Leak Volume (ml) = Leak Rate (ml/min) * Sample Time (min) Sample Volume (ml) = Sample Rate (ml/min) * Sample Time (min) Equation 2. Concentration (µg/l) = (Conc. ppmv) * (MW) 24.45 Where: 24.45 is the molar volume of any gas at normalized pressure and temperature, derived from the ideal gas law (PV=nRT) and MW = Molecular Weight of compound.

Table 1. Leak rate relationship. Check Compound Leak Rate (ml/min) % Leak Volume* Introduced Calc. Conc. (µg/l) Calc. Conc.* (ppmv) Isobutane or Butane** 0.5 0.5 11,000 4500 0.05 0.05 1100 450 0.0005 0.0005 11 4.5 2-Propanol 0.5 0.5 12000 5000 0.05 0.05 1200 500 0.0005 0.0005 12 5.0 1,1,1,2-Tetrafluoroethane** 0.5 0.5 19,000 4500 (Freon134a) 0.05 0.05 1900 450 (Office Depot Duster) 0.0005 0.0005 19 4.5 Sulfur Hexafluoride 0.5 0.5 30000 5000 0.05 0.05 3000 500 0.0005 0.0005 30 5.0 *Sample flow rate assumed to be 100 ml/min. ** Purity assumed to be 90% for these compounds. Diffusion of Leak Check Compounds When a leak check compound is detected in a soil gas sample, it is often assumed that a leak in the train must be the cause. However, compounds can also diffuse through the sample collection tubing. Diffusion of some common leak check compounds through the walls of different types of tubing used in soil gas sampling was investigated. Experimental Conditions A five-foot section of tubing was attached to an evacuated 1 L Summa canister. The other end was attached to a cylinder of humidified zero air with a flow meter set at 90ml/min. Using a vacuum check, the system was demonstrated to be leak free prior to the test. A two-foot mid-section of the tubing was sealed into a 5.0 L Tedlar bag. The bag was filled with leak check compound and allowed to stand for 10 min. Next, zero air was passed through the tubing into the evacuated 1.0L canister at a flow rate of 90 ml/min. for 8.0 minutes to yield a volume of 720ml. The canisters were pressurized with nitrogen to 15 psig and analyzed by GC/MS or GC/ECD. Three types of tubing were evaluated: Teflon, PEEK, and Nylaflow. The presence of leak check compounds at varying concentrations was observed in most of the tests. The presence of these compounds may be due to diffusion. During different periods of these tests, cross contamination could not be ruled out from one test to the next. This was due in part to the highly concentrated nature of the sources used to obtain the leak check compounds. This same difficulty may also cause contamination during

field sampling and extreme care must be used in the handling of neat materials during an air sampling event. Analytical Considerations The choice of a leak check compound should be made carefully to insure that both sampling and analytical data quality objectives are met. Some initial questions should be asked before choosing a leak check compound. 1) Is there a potential that the leak check compound is present at the contaminated site? It may be necessary to collect a test sample to characterize the soil gas in order to select a non-interfering leak check compound. 2) What reporting limits are needed? The samples may be diluted due to compounds present in the sample, but may also need to be diluted because of the leak check compound. All compounds in Table 3, with the exception of sulfur hexafluoride, exhibit similar volatility to many of the compounds of concern for soil gas analysis. As such, presence of these compounds at high levels may necessitate dilution to limit any negative effect on the instrumentation. This may compromise reporting limit objectives for other VOCs of concern. For several of the check compounds, even small leaks <10 µg/l can be problematic when subppbv reporting limits are required. 3) Does the source of leak check compound have the required purity? Some of the over the counter leak checks contain other VOCs that may be compounds of concern for soil gas analysis. A summary of some common leak check compounds and their attributes are presented in Table 3. Table 3. Leak Check Compound Comparison Leak Check Compound Reporting Limit Detector Availability Purity Isobutane 0.5ppbv GC/MS Readily 88% (Shaving Cream) 2-Propanol 2.0ppbv GC/MS Readily 70-99% (Rubbing Alcohol) 1,1,1,2-Tetrafluoroethane 0.5ppbv GC/MS Readily Variable (Freon134a) (Office Depot Duster) Butane 0.5ppbv GC/MS Readily Variable (Butane Lighter)

Sulfur Hexafluoride 0.5ppbv GC/ECD Special Order 99.8% CONCLUSION The presence of a leak in a soil gas sampling train can compromise the sample either through dilution or contamination. Because of this, many regulatory agencies are requiring leak tests. Careful thought should be used when choosing a leak check compound, whether it is site specific, detection limit or cost driven. REFERENCES 1. Advisory Active Soil Gas Investigations, Department of Toxic Substances Control: Glendale, CA and California Regional Water Quality Board Los Angeles Region: Los Angeles, CA, January 2003. 2. Subsurface Vapor Sampling Using a Geoprobe and Summa Canisters (GPP Guidelines) DRAFT, San Mateo County Environmental Health Services Agency Ground Water Protection Program, June 2004.