Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber

Similar documents
AppNote 6/2004. Thermal Desorption GC Analysis of High Boiling, High Molecular Weight Hydrocarbons SUMMARY

AppNote 8/2008. A New Purge Tool for Use with Automated Headspace Analysis KEYWORDS ABSTRACT

AppNote 6/2014. Automated Online Desorption and Analysis of DNPH Derivatives of Airborne Aldehydes and Ketones KEYWORDS ABSTRACT

AppNote 14/2012. Thermal Gravimetric Analysis/Mass Spectrometry Simulation using the GERSTEL Automated Pyrolyzer KEYWORDS ABSTRACT

AppNote 5/2008. Automation of Sample Preparation Steps using a Robotic X-Y-Z Coordinate Autosampler with Software Control KEYWORDS ABSTRACT

AppNote 4/2008. Automated Dynamic Headspace Sampling of Aqueous Samples Using Replaceable Adsorbent Traps KEYWORDS ABSTRACT

TechNote 5/2002 KEYWORDS ABSTRACT. SBSE, Twister, Thermal Desorption

AppNote 3/2007 KEYWORDS INTRODUCTION. Beverage, Detector, Headspace, Quality Control

AppNote 3/2014. Analysis of Aroma Compounds in Edible Oils by Direct Thermal Desorption GC/MS using Slitted Microvials KEYWORDS ABSTRACT

AppNote 2/1996. Andreas Hoffmann Gerstel GmbH & Co.KG, Eberhard-Gerstel-Platz 1, D Mülheim an der Ruhr, Germany

Off-gassing of Rubber Particles Used for Athletic Fields using Automated Dynamic Headspace Sampling

AppNote 1/2007. Automated Dynamic Headspace Sampling using Replaceable Sorbent Traps KEYWORDS ABSTRACT

AppNote 2/2000. Stir Bar Sorptive Extraction (SBSE) applied to En vi ron men tal Aqueous Samples INTRODUCTION

AppNote 5/2002. Characterization of a Mass Spectrometer based Electronic Nose for Routine Quality Control Measurements of Flavors KEYWORDS ABSTRACT

AppNote 7/2003. Coupling Retention Time Locked Methods and Libraries to Automated SPME or SBSE for Analysis of Flavors and Fragrances KEYWORDS

AppNote 4/1998. New Developments in Multi-Dimensional Capillary Gas Chromatography KEYWORDS ABSTRACT

AromaOffi ce: Application of a Novel Linear Retention Indices Database to a Complex Hop Essential Oil

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

AppNote 8/2000. Volatile Organic Compounds from Adhesives and their Contribution to Indoor Air Problems KEY WORDS ABSTRACT

VOC EMISSION TEST REPORT BREEAM NOR

AppNote 2/2002. Discrimination of Soft Drinks using a Chemical Sensor and Principal Component Analysis KEYWORDS ABSTRACT

Thermal Desorption Unit TDU

AppNote 1/2004. Use of a Mass Spectral Based Chemical Sensor to Discriminate Food and Beverage Samples: Olive Oils and Wine as Examples KEYWORDS

AppNote 6/2010. part a. Alternative Procedure for Extraction and Analysis of PAHs in Seafood by QuEChERS-SBSE-GC-MS KEYWORDS ABSTRACT

VOC TEST REPORT AgBB

AppNote 8/2002 KEYWORDS ABSTRACT. Electronic Nose, Fast GC, Headspace Analysis, Chemometrics,

VOC TEST REPORT M1. 23 October Regulation or protocol Conclusion Version of regulation or protocol

VOC EMISSION TEST REPORT ISO 16000

VOC TEST REPORT EMICODE

AppNote 8/2012. Rapid Automated Extraction and Confi rmation of Buprenorphine and Norbuprenorphine in Urine by DPX-LC/MS/MS KEYWORDS ABSTRACT

VOC TEST REPORT CDPH

AppNote 15/2012. Large Volume Injection with Solvent Venting - Application to Trace Detection of Analytes in Water INTRODUCTION

VOC TEST REPORT EMICODE

VOC TEST REPORT M1. 6 April Regulation or protocol Conclusion Version of regulation or protocol

Analysis of Food Samples using Thin Film Solid Phase Microextraction (TF-SPME) and Thermal Desorption GC/MS

VOC TEST REPORT CDPH

AppNote 8/2009. Fragrance Profi ling of Consumer Products using a Fully Automated Dynamic Headspace System KEYWORDS ABSTRACT

AppNote 7/2014. Automating Liquid-Liquid Extractions using a Bench-top Workstation KEYWORDS ABSTRACT. Sample Preparation, Lab Automation, LC/MS/MS

AppNote 6/2012. Flavor and Fragrance Analysis of Consumer Products - Dynamic Headspace Compared to Some Traditional Analysis Approaches KEYWORDS

VOC TEST REPORT M1. 16 August Regulation or protocol Conclusion Version of regulation or protocol

Test Report. Tremco-Illbruck GmbH & Co KG. Product Emissions Test (AgBB/DIBt Test Protocol) Butyl und Bitumenprimer. June 2011

AppNote 3/2012. Rapid Automated Screening of Extractable Compounds in Materials for Food Packaging, Medical or Technical Purposes KEYWORDS ABSTRACT

AppNote 2/2012. Determining Phenolic Compounds in Whisky using Direct Large Volume Injection and Stir Bar Sorptive Extraction KEYWORDS ABSTRACT

AppNote 10/2012. Large Volume Injection with GC/MS Multi Column Switching for Direct Determination of Ethyl Carbamate in Alcoholic Beverages KEYWORDS

VOC TEST REPORT CDPH

VOC TEST REPORT Indoor Air Comfort GOLD

AppNote 1/2011 KEYWORDS ABSTRACT

AppNote 2/2001. Stir Bar Sorptive Extraction: Enhancing Selectivity of the PDMS Phase KEYWORDS ABSTRACT

AppNote 2/2007. Automated Solid Phase Microextraction using the GERSTEL MPS Prepstation and MAESTRO Software KEYWORDS ABSTRACT

SURVEY OF CHEMICALS EMISSION FACTORS FROM BUILDING MATERIALS FOR INTERIORS

Emission measurements according to, M1 (3 appendices)

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No. G12871B_02

AppNote 1/2013. Determination of Barbiturates and 11-Nor-9-carboxy- 9 -THC in Urine using Automated Disposable Pipette Extraction (DPX) and LC/MS/MS

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

EMICODE Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No A_04

Test Report. Muovihaka Oy. Product Emissions in accordance with the Swedish Building Material Assessment Ultima graphite.

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

VOC TEST REPORT Tarkett Indoor Air Quality

AppNote 2/2004. Analysis of Packaging Materials using a Mass Spectral Based Chemical Sensor KEYWORDS ABSTRACT

Test Report- VOC emission regulations in Europe

VOC TEST REPORT Indoor Air Comfort GOLD

Test Report ROMA USA LLC. Product Emissions of Furniture in accordance with. Cradle to Cradle section 5.8 ECODOMUS SATIN BASE TR.

Test Report. Glava AS. Emission test of Ecophon Venus according to M1 classification. October / November Smedeskovvej 38, DK-8464 Galten

AppNote 5/2001. The Analysis of the Bitter and Other Flavour Compounds in Beer and Wort by Stir Bar Sorptive Extraction (SBSE) Followed by HPLC

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

VOC TEST REPORT Indoor Air Comfort

M1 Test Report. 1 Sample Information. 2 Evaluation of the Results. Report No A_04

VOC TEST REPORT Indoor Air Comfort

VOC TEST REPORT Indoor Air Comfort GOLD

Test Report in accordance CDPH-IAQ

VOC EMISSION TEST REPORT Indoor Air Comfort GOLD

Test Report- VOC emission regulations in Europe

AppNote 12/2012. Comparison of EG-Silicone-SBSE and Derivatization-PDMS-SBSE for the Analysis of Phenolic Compounds and Off-fl avors in Water KEYWORDS

Test Report- VOC emission regulations in Europe

KEYWORDS ABSTRACT AppNote 6/2002

Test Report- VOC emission regulations in Europe

VOC TEST REPORT Indoor Air Comfort GOLD

Test report No. MAIC Testing and evaluating of a digital printed paper material according to AgBB/DIBt-scheme.

AppNote 7/2002. Classifi cation of Food and Flavor Samples using a Chemical Sensor KEYWORDS ABSTRACT

VOC EMISSION TEST REPORT Indoor Air Comfort GOLD

ASTM D5116 Emissions Test Report with CDPH SM V Loading Scenarios per Cradle to Cradle Standard v3.0 Sect. 5.8

VOC EMISSION TEST REPORT Indoor Air Comfort

VOC TEST REPORT Indoor Air Comfort GOLD

Test Report in accordance CDPH-IAQ

AppNote 8/2011 KEYWORDS ABSTRACT INTRODUCTION

THEORETICAL DETERMINATION OF THE SAMPLING RATES OF DIFFUSION SAMPLERS FOR VOCS AND ALDEHYDES

VOC EMISSION TEST REPORT Indoor Air Comfort GOLD

VOC TEST REPORT Indoor Air Comfort GOLD

GERSTEL MultiPurpose Sampler MPS. Versatile autosampler and sample preparation robot

AppNote 4/2005. A Selectable Single or Multidimensional. Fraction Collection and Dual Detection for Trace Analysis of Complex Samples KEYWORDS

Intertek Environmental VOC Emissions

VOC TEST REPORT Indoor Air Comfort GOLD

AppNote 9/2012. Solid Phase Micro-Extraction (SPME) Coupled with Selectable 1 D/ 2 D GC-MS for the Determination of Food Product Flavors KEYWORDS

AppNote 2/1998. Direct Injection of Distilled Spirits with PTV Matrix Removal: The Perfect Splitless Injection? KEYWORDS INTRODUCTION

VOC EMISSION TEST REPORT Indoor Air Comfort GOLD

Influence Of The Substrates On Water-based Paints Emissions Fabio Abbà,1*, Mikaela Decio 2, Tiziano Cerulli 2 and Roberto Leoni 2

AppNote 12/2008 KEYWORDS ABSTRACT

COMPLIANCE EMISSIONS TEST California Dept. of Public Health Standard Method Version 1.1 and FloorScore Flooring Evaluation

Determination of Carbonyl Compounds in Workplace Air

AppNote 7/2005. Detection of Packaging Emissions using a Flexible Headspace Sampler Combined with a Multi Sensor System and a Separation Unit KEYWORDS

Transcription:

Determination of Formaldehyde and VOCs in Wood-based Products using an Automated Micro-Scale Chamber Yunyun Nie, Oliver Lerch, Eike Kleine-Benne GERSTEL GmbH & Co. KG, Eberhard-Gerstel-Platz 1, 45473 Mülheim an der Ruhr, Germany ABSTRACT An automated micro-scale chamber method based on 2,4-dinitrophenylhydrazine (DNPH) derivatization and Tenax TA adsorption, for sampling formaldehyde, acetaldehyde, acetone, acrolein, as well as volatile organic compounds (VOCs) was developed. Following the sampling step, the Tenax TA tubes were immediately transferred to a thermal desorption (TD)-GC/MS system for analysis. The DNPH tubes were transferred to an autosampler tray for temporary storage and later eluted and analyzed by high performance liquid chromatography (HPLC) using a diode array detector (DAD). Different wood-based products were analyzed for their emissions of the target compounds. Different emission values of formaldehyde, acetaldehyde and acetone were obtained, no acrolein was detected in any of the samples. The hexanal emission values obtained using the DNPH-HPLC method and Tenax-GC method were compared. The results showed that using the DNPH-HPLC method for determination of hexanal concentrations in chamber air resulted in analyte discrimination with lower emission values being reported. Finally, the long term emission behavior of a medium-density fiber board (MDF) was determined based on both DNPH-HPLC-DAD and Tenax TA-TD GC/MS analysis. INTRODUCTION Wood-based products, such as particle board, orientedstrand board (OSB), high-density fiber board (HDF), medium-density fiber board (MDF) and plywood are widely used indoors in buildings and they frequently contain formaldehyde-based adhesives. For this reason, wood-based products are generally a major source of formaldehyde in indoor air along with materials such as flooring, insulation and coating [1]. Exposure to formaldehyde in air can cause irritation of the mucous membranes. Formaldehyde is classified as a probable human carcinogen (EPA 1991) and a suspected carcinogen (European Union 2004). The World Health Organization (WHO) has recommended that formaldehyde concentrations be kept below 100 μg/m³ even for a half hour short-time exposure in order to prevent nose and throat irritation (1987). This value (100 μg/m³) is used as reference in countries such as Germany, the U.K. (0.5 h), Norway (0.5 h) and China (1 h) [1]. Trapping of formaldehyde and other volatile aldehydes and ketones on DNPH impregnated sorbents with subsequent HPLC-DAD determination is widely used in international standards [2, 3]. In addition to the very volatile organic compounds formaldehyde, acetaldehyde and acrolein, VOCs ranging from n-hexane to n-hexadecane are emitted from wood-based products and have an impact on the indoor air quality as well as on the health of building occupants. According to the German AgBB scheme, the total VOC (TVOC) values after 3 days in an

environmental test chamber should meet the criteria 10 mg/m³ and after 38 days 1.0 mg/m³ [4]. For each individual VOC, the lowest concentration of interest (LCI) is specified [4]. For determination of VOCs in emission test chamber, active sampling using Tenax TA Tube followed by thermal desorption - GC-MS/ FID is widely used [5]. Environmental test chambers are widely used to determine emission values of building materials and other products that come into contact with indoor air. These chambers are normally operated off-line. The specimen is put into a closed chamber under specified temperature and relative humidity conditions. Into the chamber, pure air or inert gas is supplied at a required air exchange rate. After specified time periods, air samples are collected by manually placing a Tenax TA sorbent tube and/or DNPH cartridge into the outlet flow and sampling a specified volume at a specified flow using a pump. Afterwards, the loaded Tenax TA sorbent tube and/or DNPH-cartridge are removed and placed in individually sealed storage containers for later analysis. In this study, an automated micro-scale chamber method based on the GERSTEL DHS Large autosampler (Figure 1) was developed, which offers a simple-touse efficient screening method for material emission testing. DHS Large enables analysis of heterogenous or bulk samples with a volume up to 1 L without the need for cutting, grinding or other means of homogenizing normally used in order to take a smaller representative sample. The DHS L sample containers are made of stainless steel with an inert coating. The vessels are air tight and exhibit very low VOC/SVOC background values. Different types of inlet gases can be selected, such as synthetic air, nitrogen or helium to name the most widely used. A controlled purge flow from 5-100 ml/min results in an air change rate from 0.3-6/h when a one liter container is used. The DHS Large autosampler provides 11 sample positions and the PrepAhead function enables efficient parallel processing: During the ongoing GC/MS analysis the next sample is extracted in the DHS L autosampler for efficient sample throughput and GC/MS instrument utilization. Figure 1. DHS-Large autosampler. EXPERIMENTAL Materials. For HPLC-DAD calibration, a standard mixture of aldehydes/ketones-dnph TO11/IP-6A was obtained from Sigma Aldrich. ORBO TM -DNPH tubes (LpDNPH on silica gel, 120 mg) were purchased from SUPELCO. These are 90 mm length glass tubes with a frangible seal at one end. The tubes contain a 120 mg bed of DNPH-coated silica gel and both ends are sealed with PTFE caps to avoid air ingress and contamination. Before use, the PTFE caps are removed and the frangible end is broken. For micro-scale chamber-dnph-hplc-dad system calibration, an aqueous mix of formaldehyde and acrolein was prepared, the formaldehyde solution (37 wt. % H 2 O, contains 10-15 % methanol) and acrolein ( 99.0 %) were purchased from Sigma Aldrich. This additional complete system calibration was performed due to the well-known formation of multiple derivatization products of acrolein [3]. For Tenax-TD-GC/MS calibration, a mixture of pentanal, toluene, hexanal, 2-hexanoic acid, octanal, and dodecane was made, the individual standards were purchased from Sigma Aldrich. Tenax TA tubes were obtained from GERSTEL GmbH & Co. KG. Different wood-based samples were used: oriental-strand board (OSB); medium-density fiber board (MDF); and plywood. All samples were cut to 6 cm x 9 cm size with front and back surfaces exposed to the chamber and all edges covered with aluminum foil. In addition, several wood toys were weighed and tested including puzzle pieces and wood blocks. All samples were placed vertically in the chamber. - 2 -

Micro-scale chamber method. DHS Large method parameters used for the micro-scale chamber analysis are shown in Table 1. A surface specific air flow (q) of 0.2778 m³/m²h was chosen following the AgBB specification, which requires an air exchange rate of 0.5/h and a loading factor of 1.8 m²/m³ for wall, flooring and ceiling area [4]. The loading factor and air exchange rate can be varied as long as the surface specific air flow for the product type is met [6]. Since a one liter container is used, an air exchange rate of 0.5/h results only in a flow rate of 8.3 ml/min, which is too small for the method, so the air exchange rate was varied to 3/h for the micro-scale chamber. To achieve the same surface specific air flow, a specimen size of 108 cm² was analyzed. For dual side exposure, each side must be 54 cm², therefore a sample size of 6 cm x 9 cm was chosen. Several different surface specific air flows are recommended in different standards methods. The AgBB recommendation is 0.2778; in ASTM D6007-14, different surface specific air flows are recommend, for example, 1.905 for MDF, 1.172 for particleboard flooring panels, industrial particleboard panels, industrial hardwood plywood panels, and 0.526 for hardwood plywood wall paneling [6]. In this study, we followed the AgBB recommendation, the goal was to use the same chamber parameters for quantitative comparison of different wood-based samples, not to achieve the same emission values obtained from standard-sized, large emission chambers. In contrast to standard emission chambers no humidification of the purge gas was used. For formaldehyde, acetaldehyde, acetone and acrolein, a three liter air volume was sampled on ORBO TM -DNPH tubes followed by elution with 2 ml acetonitrile and HPLC-DAD analysis. For other VOCs 500 ml air volumes were sampled on Tenax TA-filled tubes followed by thermal desorption-gc-ms analysis. Between each sampling, an empty tube was inserted to simulate emission chamber conditions. The tube exchange resulted in a short interruption of the gas flow. A schematic diagram of the automated micro-scale chamber sampling process in the DHS Large system is shown in Figure 2. Please note: In this study, elution of the DNPH derivatives into an open vial was performed manually under nitrogen atmosphere, no syringe filter and needle was used. A comparison of results obtained with and without syringe filter indicated that the filter is not needed, a pre-column provided sufficient protection of the analytical column. Table 1. DHS Large parameters for the micro-scale chamber method. Micro-scale chamber parameter Setting Loading factor L [m²/m 3 ] 10.8 Air exchange rate n [h -1 ] 3 Surface specifi c air fl ow q [m³/m 2 h] 0.2778 Chamber volume V [m³] 0.001 Sample surface A [m²] 0.0108 Temperature [ C] 23 ± 1 Relative humidity [%] 0 Figure 2. Schematic diagram of DNPH sampling (top) and Tenax sampling (bottom) using DHS Large as a micro-scale chamber. - 3 -

Analysis conditions DNPH-HPLC-DAD. DNPH elution: 2 ml Acetonitrile HPLC column: Nucleoshell RP 18plus 50 mm x 2 mm x 2.7 μm (Macherey Nagel) Pump: gradient, flowrate = 0.8 ml/min Mobile phase: A - water, B - acetonitrile Gradient: Initial 20 % B 15 min 60 % B Injection volume: 5.0 μl (full loop injection) Detector: 1200 SL HPLC-DAD (Agilent Technologies) 360 nm, bandwidth 30 nm Analysis conditions Tenax-GC-MSD. TDU: 20 C; 720 C/min; 275 C (5 min) Column: 30 m Rxi-5ms (Restek) d i = 0.25 mm d f = 0.5 μm Pneumatics: He, constant flow = 1.5 ml/min split 1:5 Oven: 40 C (4 min); 5 C/min; 90 C; 8 C/min 200 C; 20 C/min; 280 C (4 min) Detector: 5973 MSD (Agilent Technologies) Scan, 33-300 amu RESULTS AND DISCUSSION Linearity, LOD, LOQ. Using HPLC-DAD, good linearity was obtained for aldehyde/ketone-dnph derivatives in the range from 0.01 to 15 μg/ml with R² larger than 0.9991. Formaldehyde, acetaldehyde and acetone were found to have unavoidable blank values caused by the DNPH sorbent tube (background per tube: formaldehyde < 25 ng, acetaldehyde < 35 ng, acetone < 120 ng) as well as the collection chamber and the elution process due to contact with laboratory air. Therefore, the blank value method according to DIN 32645 was used to determine the limits of detection (LOD) and limits of quantification (LOQ).Three liter samples of air were collected from an empty DHS Large container through DNPH tubes. Five blank samples were run using five different DHS Large containers and five ORBO TM -DNPH tubes (Table 2). Table 2. Linearity of liquid injection of an aldehyde/ketone-dnph standard and LOD and LOQ of the microscale chamber-dnph-hplc-dad method (3 L sampling, 2 ml ACN) according to DIN 32645. Aldehyde/Ketone-DNPH Linear range [µg/ml] Linearity (R²) LOD [µg/ml], (n=5) LOQ [µg/ml], (n=5) Formaldehyde-DNPH 0.01-15 1.0000 0.010 0.032 Acetaldehyde-DNPH 0.01-15 0.9995 0.041 0.123 Acetone-DNPH 0.01-15 0.9998 0.504 1.511 Acrolein-DNPH 0.01-15 0.9995 n.d. n.d. Others-DNPH 0.01-15 > 0.9991 n.d n.d. Since the calibration standards contain the DNPH derivatives of aldehydes and ketones, the LODs and LOQs need to be recalculated to the underivatized compounds taking into account the molecular weight of the respective molecules (Figure 3). Furthermore, liquid concentrations of aldehydes/ketones need to be converted into air concentrations for the micro-scale chamber-dnph-hplc-dad method based on a 3 liter sampling volume and an ACN elution volume of 2 ml (equation 1 and table 3). The LOQs were as follows: formaldehyde: 3.05 μg/m³; acetaldehyde: 14.8 μg/m³; and acetone: 246 μg/m³. The high LOQ for acetone was caused by the high background concentration in the laboratory air. Air concentration [µg/m 3 ] = Liquid concentration [µg/ml] * Volume of acetonitrile [ml] Sampling volume [m³] Equation 1-4 -

Figure 3. Reaction process of aldehyde or ketone with 2,4-dinitrophenylhydrazine (DNPH). Table 3. LODs and LOQs of aldehydes/ketones with the micro-scale chamber-dnph-hplc-dad method (3 L sampling volume, 2 ml ACN). Liquid concentration Air concentration Aldehyde/Ketone LOD [µg/ml] (n=5) LOQ [µg/ml] (n=5) LOD [µg/m³] (n=5) LOQ [µg/m³] (n=5) Formaldehyde 0.0014 0.0046 0.95 3.05 Acetaldehyde 0.0081 0.0222 5.37 14.81 Acetone 0.1229 0.3684 81.93 245.62 Acrolein n.d. n.d. n.d. n.d. HPLC-MS. HPLC-MS detection in SIM mode was also used at the beginning of the study, but the method was only linear up to concentrations of around 0.06 μg/ml. HPLC-MS was found to be more sensitive than HPLC- DAD, but in reality LODs and LOQs of formaldehyde, acetaldehyde and acetone were dictated by the relative high blank values rather than by the sensitivity of the detector. Consequently, a DAD detector was deemed the more the suitable choice for the DNPH method. Repeatability & Recovery. For validation of the micro-scale chamber method, a 3 μl sample of water containing 0.6 μg each of formaldehyde and acrolein was spiked into a 1 L DHS Large container and 3 L of air subsequently sampled onto a DNPH tube. This recovery experiment was repeated 5 times using 5 different containers. Average recovery for formaldehyde was found to be 102 % (RSD 6.0 %) when calculated against an HPLC-DAD calibration curve based on direct injection of liquid standard DNPH derivatives into the HPLC system, and 102 % (RSD 6.1 %) when calculated against a micro-scale chamber-dnph-hplc-dad calibration curve based on direct injection of a liquid standard into the micro-scale chamber. The recovery of acrolein was calculated to be only 43 % (RSD 10.7 %) based on a direct injection HPLC-DAD calibration curve, but 82 % (RSD 11.7 %) based on a micro-scale chamber-dnph-hplc-dad calibration curve (Table 4). Sampling of acrolein on DNPH cartridges leads to the formation of more than one derivative, whereas the aldehyde/ketone-dnph standard mix contains only the main acrolein-dnph derivative [1]. Therefore an underestimation occurs when using the HPLC-DAD calibration. - 5 -

Table 4. Recoveries and RSDs (%, n=5) for formaldehyde and acrolein using the micro-scale chamber- DNPH-HPLC-DAD method (3 L sampling, 2 ml ACN elution). Compound Amount spiked (µg) Recovery [%], (n=5) HPLC-DAD calibration micro-scale chamber-dnph- HPLC-DAD calibration HPLC-DAD calibration RSD [%], (n=5) micro-scale chamber-dnph- HPLC-DAD calibration Formaldehyde 0.6 102 102 6.0 6.1 Acrolein 0.6 43 82 10.7 11.7 Analysis of wood samples. Three different wooden samples (OSD, MDF and plywood) were chosen as test samples and analyzed using the micro-scale chamber-dnph-hplc-dad method. All the board/plate samples were prepared in 6 cm x 9 cm size with front and back surfaces exposed to the chamber. All edges were covered with aluminum foil and the sample was placed vertically in the DHS Large container (Figure 4). Figure 4. Medium density fi ber board (MDF) prepared in 6 cm x 9 cm size and all edges covered with aluminum foil. The sample was placed vertically in the DHS Large container. Detected aldehydes/ketones with their calculated air concentrations and surface specific emission rates (SER, in brackets) are listed in Table 5. All samples show emissions of formaldehyde, acetaldehyde and acetone, but no acrolein was found. The OSB sample emitted mainly higher aldehydes, like propionaldehyde, butyraldehyde and (iso)valeraldehyde. The highest formaldehyde and acetone emissions were found for the plywood sample. The average concentration of formaldehyde was 147 μg/m³ with a SER of 41 μg/m²h and an RSD of 8.8 % (n=4). This sample was part of a wooden base plate of a box containing wooden building blocks for children of 1 year and older. Base plates of the same product boxes were tested by the German Stiftung Warentest also showing high amounts of formaldehyde emissions [7]. Although the methods used in these two cases were not the same, the results are comparable. - 6 -

Table 5. Air concentrations [μg/m 3 ] and surface specifi c emission rate (SER [μg/m²h]) obtained from wood plates using micro-scale chamber-dnph-hplc-dad method. Sample Plywood* OSB 12 mm MDF 12 mm Compound (RSD %, n=4) Formaldehyde 3 (1) 33 (9) 147 (41, 8.8 %) Acetaldehyde 24 (7) 26 (7) 13 (4, 16.0 %) Acetone < LOQ < LOQ 312 (87, 8.6 % ) Acrolein < LDD < LDD < LDD Propionaldehyde 6 (2) < LDD < LDD Butyraldehyde 6 (2) < LDD < LDD (Iso)valeraldehyde 53 (15) < LDD < LDD Hexanal 191 (53) 100 (28) < LDD * Plywood is from a base plate of a box containing building blocks for children. The surface specific emission rate (SER) was calculated using the following equation: A B SER = c * N L = c * N * V A Equation 2 c: analyte concentration in air sample [μg/m³] N: air exchange rate [3/h] V: Sampling volume [0.003 m³] A: Sample surface area [0.0108 m²] q: Surface specific air flow rate [0.2778 m³/m²h] L: Loading factor [m²/m 3 ] C D Several wooden toys (Figure 5) were also tested using the micro-scale chamber method under the same conditions except that the sizes of the wooden toys varied. The surface area of the samples were measured, surface specific emission rates are shown in Table 6. Figure 5. Puzzles (A: sun & B: balloon ) and building blocks (C: nature & D: pink ) Table 6. Air concentrations [μg/m 3 ] and surface specifi c emission rates (SER [μg/m²h]) obtained from wooden toys using the micro-scale chamber-dnph-hplc-dad method. Compound Puzzle sun plywood (103 cm²) Puzzle balloon plywood (146 cm²) Building block nature (56 cm²) Building block pink (78 cm²) Formaldehyde 21 (6) 32 (4) < LOQ 10 (7) Acetaldehyde 26 (8) 23 (7) 33 (18) 18 (5) Acetone < LOQ < LOQ < LOQ < LOQ Acrolein 8 (2) < LOD < LOD < LOD Butyraldehyde 6 (2) < LOD < LOD < LOD (Iso)valeraldehyde 17 (5) < LOD < LOD < LOD Hexanal 77(22) 25 (6) 17 (9) 16 (5) - 7 -

As can be seen in the table, the building block nature emitted the lowest amount of formaldehyde (< 3 μg/m 3 ) whereas to the puzzle piece balloon made of plywood emitted the highest amount of formaldehyde (32 μg/m 3 ). By contrast, the building block nature emitted the highest amount of acetaldehyde (33 μg/m 3 ). Acrolein was detected only in pieces from the puzzle sun in a concentration of 8 μg/m 3 along with butyraldehyde and (iso)valeraldehyde. The sun puzzle piece tested also emitted the highest amount of hexanal (77 μg/m 3 ). The same wooden toys were tested by the German Stiftung Warentest, which found high levels of formaldehyde emissions in the puzzle piece balloon, but no formaldehyde in the wood building block nature or in the puzzle piece Sun [7]. Emission behavior. In order to demonstrate the possibility for automated monitoring of VOC emissions over time from wooden samples, a monitoring program over three days was set up for the MDF sample, alternating between the DNPH-HPLC-DAD and the Tenax-GC-MS methods. Between each DNPH tube sampling (interval at least 5 hours), several Tenax TA samplings were done automatically every hour. As an example, emission profiles for formaldehyde and hexanal obtained using the DNPH-HPLC-DAD method are shown in Figure 5. The emission value of formaldehyde decreases slowly, from 9 μg/m²h after 1 hour under micro-scale chamber sampling conditions (3/h, 23 ± 1 C, RH = 0 %) to 4 μg/m²h after 72.87 hours under micro-scale chamber sampling conditions (3/h, 23 ± 1 C, RH = 0 %). Hexanal shows a strong decrease from 28 μg/m²h at 1 hour to 2 μg/m²h at 72.87 hours. Hexanal emission - DNPH vs. Tenax. The hexanal emission profile obtained with the Tenax-GC-MS method is shown in Figure 7. A pronounced decay was observed, from 104 μg/m²h at 2 hours to 7 μg/m²h at 53.33 hours. The data shows that the DNPH-HPLC method results in a significantly lower hexanal concentration (around one-quarter) compared to the Tenax-GC method. Results are in good accordance with the literature which also reports poor recovery for the analysis of hexanal using DNPH sorbent tubes due to slow derivatization kinetics [8, 9]. Figure 7. Hexanal emission profi le for MDF using the Tenax-GC-MS method and the DNPH-HPLC method. TVOCs in toluene equivalents. The total volatile organic compound (TVOC) emission values from n-hexane to n-hexadecane were calculated as toluene equivalents. A decay curve is shown in Figure 8, the SER value of TVOCs decreased from 399 μg/m²h at 2 hour to 209 μg/m²h at 53.33 hours. Using a power function trend line, the TVOC SER values after 3 days (72 hours) and after 28 days (672 hours) were extrapolated and calculated to be 182 μg/m²h and 96 μg/m²h respectively. An estimation of emission values after 3 days or even 28 days with only one or two days of continuous measurements would be a powerful way to study the material emission behavior and therefore shorten the overall test time needed. More tests of different materials would be needed in order to build solid emission model. Figure 6. Formaldehyde and hexanal emission profi les of MDF using the DNPH-HPLC-DAD method. - 8 -

Figure 8. TVOCs emission profi le and power trend line for MDF using the Tenax-GC-MS method. Humidity. It is well known that a higher relative humidity leads to higher formaldehyde emission values [10]. Under low relative humidity, formaldehyde emission values can be underestimated due to a suppressed hydrazine-to-hydrazone conversion [11]. The impact of relative humidity on the emission behavior of formaldehyde in building materials was also studied by Huang et al. [12], the model was built using an airtight chamber according to the c-history method [13] for relative humidity from 20 % to 85 %. Unfortunately, no modeling or relationship was available to estimate the emission value at another relative humidity using the emission value obtained at zero % relative humidity. Our goal was to quantitatively compare the woodbased products under the same micro-scale chamber conditions and not to achieve the same emission values as those obtained in emission tests in bigger environmental chambers. We have succeeded in developing a time, labor and cost saving screening method for product quality control and development. CONCLUSIONS This study demonstrates the possibility of combining DNPH derivatization and Tenax TA adsorption, for automated sampling of formaldehyde and other low molecular aldehydes and ketones as well as VOCs in parallel in one micro-scale chamber. VOC analysis via GC-MS was performed automatically after sampling on the same instrument using thermal desorption, while DNPH sorbents were first stored on the instrument and later eluted followed by HPLC-DAD analysis. Automation of the elution step and LC determination is planned. The micro-scale chamber method using DHS Large containers (1 L) and DNPH sorbent tubes showed high recovery with low RSDs for formaldehyde and acrolein as well as low LODs and LOQs in addition to good linearity. Method applicability was demonstrated by the analysis of several wooden samples. The DNPH sorbent was found to be suitable for C1- C4 aldehydes, especially for formaldehyde, but not for hexanal and higher aldehydes, a fact that has also been reported by other authors [9]. Acetone showed high and variable background values, probably caused by differing acetone air concentrations in the laboratory. Acrolein forms many derivatives on a DNPH tube, as a consequence calibration needs to be done for the whole sampling process. A time and labor saving automated sampling and analysis method was developed that is suitable as screening method for quality control in production processes or for product development. It is also ideal for studying the emission behavior of materials and for building emission models. For example, Xiong et al. have developed a c-history method in a closed chamber to build models in order to calculate the influence of environmental factors (temperature, humidity and air exchange rate) and to compensate for the influence of different chamber types. - 9 -

REFERENCES [1] Tunga Salthammer, Sibel Mentese, Rainer Marutzky, Formaldehyde in the Indoor Environment, Chemical Reviews 2010, Vol. 110, 2536-2572. [2] ISO 16000-3, Determination of formaldehyde and the carbonyl compounds in indoor air and test chamber air-active sampling method, ISO, 2011. [3] D5197-09, Standard Test Method for Determination of Formaldehyde and Other Carbonyl Compounds in Air (Active Sampler Methodology), ASTM International, 2010. [4] AgBB Scheme, Health-related Evaluation of Emissions of Volatile Organic Compounds (VVOC, VOC and SVOC) from Building Products, Umweltbundesamt, 2015. [5] ISO 16000-6, Determination of volatile organic compounds in indoor and test chamber air by active sampling on Tenax TA sorbent, thermal desorption and gas chromatography using MS/ FID, ISO, 2010. [6] D6007-14, Standard Test Method for Determining Formaldehyde Concentrations in Air from Wood Products Using a Small-Scale Chamber, ASTM, 2014. [7] Der Lack ist ab, Stiftung Warentest, 12/2013, pp. 70-75. [8] Schlomski, Stefan, Entwicklung von Methoden zur Analyse von Carbonylverbindungen in der Atmosphäre, http://tuprints.ulb.tu-darmstadt. de/72/, [Online] 2000. [9] Tunga Salthammer, Sibel Mentese, Comparison of analytical techniques for the determination of aldehydes in test chambers. Chemosphere 2008, Vol. 73, 351-1356. [10] ISO 16000-2, Indoor Air - Sampling Strategy for Formaldehyde, ISO, 2004. [11] A. Wisthaler, et al., Intercomparison of formaldehyde measurements at the atmosphere simulation chamber SAPHIR, Atmos. Chem. Phys. 2008, Vol. 8, 2189 2200. [12] Shaodan Huang, Jiangyin Xiong, Yinping Zhang, The Impact of Relative Humidity on the Emission Behavior of Formaldehyde in Building Materials, Procedia Engineering 2015, Vol. 121, 59-66. [13] Jianyin Xiong, Yuan Yao, Yinping Zhang, C-History Method: Rapid Measurement of the Initial Emittable Concentration, Diffusion and Partition Corfficients for Formaldheyde and VOCs in Building Materials, Environ. Sci. Technol. 2011, Vol. 45, 3584-3590. [14] ISO 16000-9, Determination of the emission of volatile organic compounds from building products and furnishing-emission test chamber method, ISO, 2006. [15] EN 717-1, Wood-based panels - Determination of formaldeyhde release- part 1: Formaldehyde emission by the chamber method, DIN, 2005. [16] ISO16000-25, Determination of the emission of semi-volatile organic compounds by building products- Micro-chamber method, ISO, 2011. [17] D7706-1, Standard Practice for Rapid Screening of VOC Emissions from Products Using Microscale Chamber, ASTM International, 2011. [18] HJ/T 400-2007, Determination of Volatile Organic Compounds and Carbonyl Compounds in Cabins of Vehicles, HJ, 2007. [19] Shaodan Huang, Jiangyin Xiong, Yinping Zhang, The Impact of Relative Humidity on the Emission Behaviour of Formaldehyde in Building Materials, Proceldia Engineering 2015, Vol. 121, 59-66. [20] Wenjian Wei, Cynthia Howard-Reed, Andrew Persily, Yinping Zhang, Standard Formaldehyde Source for Chamber Testing of Material Emissions: Model Development, Experimental Evaluation, and Impacts of Environmental Factors, Environ. Sci. Technol. 2013, Vol. 47, 7848-7854. - 10 -

- 11 - GERSTEL Application Note No. 193, 2017

GERSTEL GmbH & Co. KG Eberhard-Gerstel-Platz 1 45473 Mülheim an der Ruhr Germany +49 (0) 208-7 65 03-0 +49 (0) 208-7 65 03 33 gerstel@gerstel.com www.gerstel.com GERSTEL Worldwide GERSTEL, Inc. 701 Digital Drive, Suite J Linthicum, MD 21090 USA +1 (410) 247 5885 +1 (410) 247 5887 sales@gerstelus.com www.gerstelus.com GERSTEL AG Wassergrabe 27 CH-6210 Sursee Switzerland +41 (41) 9 21 97 23 gerstelag@ch.gerstel.com www.gerstel.ch GERSTEL K.K. 1-3-1 Nakane, Meguro-ku Tokyo 152-0031 SMBC Toritsudai Ekimae Bldg 4F Japan +81 3 5731 5321 +81 3 5731 5322 info@gerstel.co.jp www.gerstel.co.jp GERSTEL LLP Level 25, North Tower One Raffles Quay Singapore 048583 +65 6622 5486 +65 6622 5999 SEA@gerstel.com www.gerstel.com GERSTEL Brasil Av. Pascoal da Rocha Falcão, 367 04785-000 São Paulo - SP Brasil +55 (11)5665-8931 +55 (11)5666-9084 gerstel-brasil@gerstel.com www.gerstel.com.br Information, descriptions and specifications in this Publication are subject to change without notice. GERSTEL, GRAPHPACK and TWISTER are registered trademarks of GERSTEL GmbH & Co. KG. Copyright by GERSTEL GmbH & Co. KG Awarded for the active pursuit of environmental sustainability