Mechanism of Microwave Sterilization in the Dry State

Similar documents
Inactivation of Bacillus Spores in Dry Systems at Low and Wgh Temperatures

Effect of Microwaves on Escherichia coli and Bacillus subtilis

Kinetics of Escherchia coli Destruction by Microwave Irradiation

Thermal Injury and Recovery of Salmonella typhimurium and Its Effect on

Thermal Resistance of Bacillus subtilis var. niger in a

EFFECT OF MICROWAVES ON MOULDS ISOLATED FROM SURFACES

Kinetics of Microbial Inactivation for Alternative Food Processing Technologies Oscillating Magnetic Fields (Table of Contents)

Solutions With Formaldehyde-Water Solutions

Dry Heat Inactivation of Bacillus subtilis var.

Relationship Between Atmospheric Temperature

DNA And The Microwave Effect Post From MailBag 4/8/2002: DNA and the Microwave Effect Dear Technical staff,

ANTIMICROBIAL TESTING. E-Coli K-12 - E-Coli 0157:H7. Salmonella Enterica Servoar Typhimurium LT2 Enterococcus Faecalis

Rule 2. Rule 1. Rule 4. Rule 3. Rule 5. Rule 6. Rule 7. Rule 8

was prepared by the method of Beeby and Whitehouse and sodium hypochlorite were tested periodically; no changes were detected over the experimental

ANTIBACTERIAL ACTIVITY OF PLANT EXTRACTS IN FOOD PRODUCTS

INTRODUCTION Radiation differs from conduction and convection in that it does not require the presence of a material medium to take place.

Module 2. Measurement Systems. Version 2 EE IIT, Kharagpur 1

Supporting Information

IGNITABILITY ANALYSIS USING THE CONE CALORIMETER AND LIFT APPARATUS

Ultraviolet Bactericidal Irradiation of Ice

Concentrated Milk. Wisconsin, Madison, Wisconsin data generated by this study would enable the

Salmonella typhimurium in Glucose-Mineral Salts Medium

of Freeze-dried Salmonella typhimurium1

A Selective Medium for Bacillus anthracis

Thermochemistry: Calorimetry and Hess s Law

The Effect of Static Magnetic Field on E. coli, S. aureus and B. subtilis Viability

of the work reported here was to define the point in the developmental process at which the curing salts act to prevent outgrowth.

BIT TM Detailed Technical Discussion

Tineke Jones Agriculture and Agri-Food Canada Lacombe Research Centre Lacombe, Alberta

Simple method to study the mechanism of thermal and non thermal bactericidal action of microwave radiations on different bacterial species

SCITECH Volume 4, Issue 1 RESEARCH ORGANISATION November 09, 2017

Differential Damage in Bacterial Cells by Microwave Radiation on the Basis of Cell Wall Structure

Evaluation of the efficiency of Mxxxx as a barrier against microrganisms crossing

RELATIONSHIP OF CELL WALL STAINING TO GRAM DIFFERENTIATION'

Microwave Dielectric Properties of Four Types of Rhizomes from Zingiberaceace Family

USING SCANNING ELECTRON MICROSCOPY TO VISUALIZE PHOTOCATALYTIC MINERALIZATION OF AIRBORNE MICROORGANISMS

Chapter 6. Heat is transferred from one place to another by three different processes

A METHOD FOR CALCULATING THE EFFECTIVE PERMITTIVITY OF A MIXTURE SOLUTION DURING A CHEMICAL REACTION BY EXPERIMENTAL RESULTS

Interpretation Guide. Aerobic Count Plate

Mechanisms of Actions of Hypochlorous Acid as Cleaning and Disinfecting Agents in Relation to Injury to Bacteria

Collaborators. Page 1 of 7

CEINT/NIST PROTOCOL REPORTING GUIDELINES FOR THE PREPARATION OF AQUEOUS NANOPARTICLE DISPERSIONS FROM DRY MATERIALS. Ver. 2.0

Biomedical Instrumentation

Infrared Temperature Calibration 101 Using the right tool means better work and more productivity

SPH3U1 Lesson 03 Energy

LAB. FACTORS INFLUENCING ENZYME ACTIVITY

gskin Application Note: Calorimetry

ACTIVITY OF PERA SAFE AGAINST BACILLUS ANTHRACIS SPORES

Detection limit: grain, feed 500 ppb; milk 50 ppb; cream, cheese 5 ppb

Inactivation of spores using Ultraviolet (UV) in combination with heat

Experimental Procedure

Biological Process Engineering An Analogical Approach to Fluid Flow, Heat Transfer, and Mass Transfer Applied to Biological Systems

Comparing the actual value and the experimental value on heat. By conservation of energy

Antimicrobial Activity of Cinnamic Acid, Citric Acid, Cinnamaldehyde, and Levulinic Acid Against Foodborne Pathogens

Resistance of Escherichia coli and Salmonella typhimurium to Carbenicillin

Inactivation of Bacillus cereus Spores by High Hydrostatic Pressure at Different Temperatures

Topic 5 Practice Test

QUANTUM DOTS, ADVANTAGES AND DRAWBACKS FOR LIGHTING APPLICATIONS. David Schmidmayr, Johann Zehetner

STERILISATION BY ECR PLASMA

Thermo-Mechanical Analysis of a Multi-Layer MEMS Membrane

Solar Matters III Teacher Page

Which type of electromagnetic wave has a wavelength longer than that of yellow light? A. Infrared radiation C. X-rays B. Gamma Rays D.

Matter mass space atoms solid, a liquid, a gas, or plasm elements compounds mixtures atoms Compounds chemically combined Mixtures not chemically

Growth from Spores of Clostridium perfringens

Growth of Salmonella typhimurium in Skim

A thermodynamic system is taken from an initial state X along the path XYZX as shown in the PV-diagram.

Solid State Science and Technology, Vol. 19, No 1 (2011) ISSN MICROWAVE DIELECTRIC CHARACTERIZATION OF TURMERIC

Output intensity measurement on a diagnostic ultrasound machine using a calibrated thermoacoustic sensor

* Defining Temperature * Temperature is proportional to the kinetic energy of atoms and molecules. * Temperature * Internal energy

LASERS. Amplifiers: Broad-band communications (avoid down-conversion)

Experiment 1: Thin Layer Chromatography

It s Time to Take the Guesswork Out of Your Sterilization Process Apply Accurate Biological and Physical Information

Working with Hazardous Chemicals

The Effect of Cooling Systems on HTS Microstrip Antennas

Mathematical Models of Cobalt and Iron Ions Catalyzed Microwave Bacterial Deactivation

This watermark does not appear in the registered version - Laser- Tissue Interaction

Free Radical Chlorination

SECTION 3 & 4 LIGHT WAVES & INFORMATION TRANSFER

Heat Transfer. Conduction, Convection, and Radiation. Review: Temperature

A Comparison of Clear-Sky Emission Models with Data Taken During the 1999 Millimeter-Wave Radiometric Arctic Winter Water Vapor Experiment

OPTIMIZATION OF TIME AND ROTATIONAL SPEED OF DISC FOR TINNED MILK USING NUMERICAL SIMULATIONS FOR UNIFORM HEAT DISTRIBUTION

Originally published as:

Tentative Identification of Methanogenic Bacteria by Fluorescence Microscopy

Computation of Electromagnetic Energy Absorption in the Human Body Tissues by High Frequency Structure Simulator

ph / Temperature Meter for Cheese with Application Specific Probe

Effect of Several Environmental Conditions on the "Thermal Death Rate" of Endospores of Aerobic, Thermophilic Bacteria

3D Thermal-Diffusion Analysis on a Moisture Loaded Epoxy Sample

High temperature He is hot

Experiment 11: Dehydration of Cyclohexanol

Laboratory Report. Thus both gram-positive and gram-negative bacteria, yeast, dermatophytes and mould are detected.

Growth of Clostridium perfringens in Cooked Chili during Cooling

Comparative Bacteriology Analysis: Source, cultivation, and preparation of bacterial samples:

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

A DIELECTRIC STUDY ON HUMAN BLOOD AND PLASMA

There are four phases of matter: Phases of Matter

Response Characteristics of Dew Point Sensor with Aluminum Oxide by means of Correlation between Purging Rate and Tube Length

Investigation of the Biocidal Effect of Electrochemically Activated Aqueous Sodium Chloride Solution on Gram-negative Pathogenic Bacteria

Supporting information

Measurement of wettability for polymer materials using non-contact surface resistivity

1P22/1P92 Exam Review Problems 2013 Friday, January 14, :03 AM. Chapter 20

Transcription:

APPLIED AND ENVIRONMENTAL MICROBIOLOGY, Sept. 1987, p. 2133-2137 99-224/87/92133-5$2./O Copyright C 1987, American Society for Microbiology Vol. 53, No. 9 Mechanism of Microwave Sterilization in the Dry State DAVID K. H. JENG,* KURT A. KACZMAREK,t ARCHIE G. WOODWORTH, AND GLENN BALASKY Advanced Sterilization Technology, Baxter Travenol Laboratories, Inc., Round Lake, Illinois 673 Received 17 March 1987/Accepted 25 June 1987 With an automated computerized temperature control and a specialized temperature measurement system, dry spores of Bacillus subtilis subsp. niger were treated with heat simultaneously in a convection dry-heat oven and a microwave oven. The temperature of the microwave oven was monitored such that the temperature profiles of the spore samples in both heat sources were nearly identical. Under these experimental conditions, we unequivocally demonstrated that the mechanism of sporicidal action of the microwaves was caused solely by thermal effects. Nonthermal effects were not significant in a dry microwave sterilization process. Both heating systems showed that a dwelling time of more than 45 min was required to sterilize 15 inoculated spores in dry glass vials at 137 C. The D values of both heating systems were 88, 14, and 7 min at 117, 13, and 137 C, respectively. The Z value was estimated to be 18 C. Electromagnetic energy in the microwave region (225 MHz to 1 GHz, typically 2,45 MHz) is extensively studied as one of the alternative energy sources for sterilization. The efficiency of microwave sterilization is essentially a function of both the electromagnetic field strength and the exposure time. The electromagnetic energy is expressed largely in two forms: (i) the factors that depend on the dielectric properties of the dipole molecules of the irradiated materials in the form of heat (thermal effect) and (ii) the factors that do not depend on the dipole molecules in the form of a direct effect of the radiofrequency (nonthermal effect). Although studies to differentiate the thermal and nonthermal effects of microwaves on microbiological systems have been attempted, the mechanism of sporicidal action by microwaves has not been resolved. A few recent studies have conflicting conclusions. Olsen et al. believe that the nonthermal effect of microwaves plays a role in inactivation of microorganisms in suspension via formation of hydrogen peroxide and other chemical transformation of small molecules such as chemical bond cleavage (15). However, details of the evidence put forward to support these conclusions have not been given. Culkin and Fung observed that heat generated during the microwave exposure alone is inadequate to fully account for the nature of the lethal effects of microwaves for microorganisms in soup (5). Goldblith and Wang (8) and Lechowich et al. (1), however, found that in liquid systems the bactericidal and sporicidal activities caused by microwaves were caused solely by thermal effects. Recently, Furia et al. found essentially similar results with yeast cells (7). All biological systems are electrochemical in nature. It would, therefore, not be surprising if electromagnetic fields influenced the physiology of microorganisms. However, a liquid suspension containing microorganisms is not an efficient system for the purpose of studies because of its reduction of the local electric field strength (21). In addition, a microbiological liquid sample placed in an electromagnetic field rises in temperature as dielectric absorption converts * Corresponding author. t Present address: Department of Electrical and Computer Engineering, University of Wisconsin, Madison, WI 5376. 2133 electrical energy into heat, which could mask the nonthermal functions. Therefore, in any experimental attempt to differentiate the thermal and nonthermal effects of microwaves on microbial physiology, the microbiological samples should be in a dry state. It is critically important that the microwaveheated and non-microwave-heated samples are simultaneously studied under the same temperature profiles. Vela and Wu (22) studied the mechanism of microwave bactericidal action under dry conditions at a subbactericidal temperature and concluded that the dry or lyophilized microorganisms are not capable of absorbing microwave energy and are not damageable by the microwaves. Wayland et al., however, used microwaves to heat the dry spores in comparison with convection thermal heating and found that the thermal and electromagnetic nonthermal functions of microwaves are interdependent (23). Several investigators have proposed the use of microwaves for sterilizing materials in the dry state for medical use (12, 18, 19) (U.S. patent 3,753,651; Japanese patent 4786). Particularly intriguing is the possibility of sterilizing heat-sensitive products at relatively low temperatures for a short time. Lohman and Manique recently reported that a microwave treatment for as short as 2.7 min in a constantly rising temperature up to 14 C is effective to sterilize 16 Bacillus subtilis subsp. niger spores in well-insulated and properly loaded dry glass vials (12). Similar results are claimed in U.S. and Japanese patents. Generally, a typical dry-heat sterilization process by convection heat would require a few hours of dwelling time at 16 C. The effect of heat sterilization is a function of temperature and exposure time. If microwaves can sterilize dry products more efficiently in a shorter time and at lower temperatures than convection heat under the same humidity conditions, the effect of the sporicidal activities of microwaves must be caused by the nonthermal effects, rather than or in addition to the thermal effects. The quantal energy of microwaves per se (about 1-5 ev) is not sufficient to cause chemical bond cleavage (11, 2). Effects of microwaves on biological systems other than chemical bond cleavage have been reported (2-4, 25) and reviewed (L. Furia, Ph.D. dissertation, University of Utah, Salt Lake City, 1986). If the nonthermal effects of microwaves can cause bactericidal and sporicidal

2134 JENG ET AL. APPL. ENVIRON. MICROBIOL. activities, they must be a result of nonthermal effects other than chemical bond cleavage. In conducting mechanism studies of microwaves, perfect temperature control is of prime importance since even a temporary temperature variation of only a few degrees may completely obscure the results. By using a computerized temperature control and a specialized temperature measurement system, experimental conditions were created in the current study so that dry spores were simultaneously exposed to microwaves and to convection dry heat with nearly identical temperature profiles. Under these conditions, the possibility of developing a short sterilization cycle at a lower temperature by microwaves to inactivate the dry spores was evaluated and the mechanism of the sporicidal activities of microwaves was studied. MATERIALS AND METHODS Bacterial spores. The spores of B. subtilis subsp. niger (ATCC 9372) were used because of their extreme resistance to dry heat (9). The dry-heat resistance is related to water activity in spores. The optimal water activity for the maximal heat resistance is about 2 to 4% relative humidity (1). A 17/ml spore suspension (1,ul) in distilled water was inoculated on the inside side wall of borosilicate glass vials (65 by 15 mm diameter) approximately 1 cm from the bottom and air dried. The vials were then situated in a closed chamber containing a beaker of saturated calcium chloride, which provided 33% relative humidity at 22 C (24), for 7 days before being used. The relative humidity was verified by a calibrated humidity sensor (model HM 31; Vaisala, Inc.). Spore recovery. After heat exposures, 5 ml of sterile distilled water was added to the sample vials, which were then sonicated for 15 min in an ultrasonic bath. Duplicates of 1-ml samples were assayed by the standard pour plate methods in Trypticase soy agar (BBL Microbiology Systems). All plates were incubated at 3 to 35 C for 48 h. Convection oven. An electric dry-heat convection oven (GCA/Precision model 19; Fisher Scientific Co.) was used. It was preheated to approximately the desired maximum test temperature before each experiment. A calibrated mercuryglass thermometer was used to measure the oven temperature. The sample vial was situated in a 1-ml beaker, which supported and slowed the heating rate of the sample vial in order to accommodate the microwave oven temperature tracking. Microwave oven. The microwave oven (model BPH-6; Cober Electronics) used in this study is an industrial unit capable of up to 6 kw of microwave output power at 2,45 MHz. The maximum experimental power output was 4 kw. The oven was modified so that an external computer system could control the microwave power automatically. Inside the microwave oven, a styrofoam block (14 by 15 by 15 cm) contained one sample vial. The styrofoam provided the necessary thermal insulation to prevent the vial from losing too much heat to the environment (12). Temperature probes. The model 2A Fluoroptic thermometer (Luxtron, Inc.) measured the temperature of the glass vial during each experiment. This specialized instrument uses nonmetallic and microwave-transparent fiber optic probes. Transparent cellophane tape secured two probe tips, approximately 2 mm apart, to the outside surface of each glass vial opposite the spore inoculum. Temperatures collected from these two temperature probes were averaged for datum analysis, except for the data shown in Fig. 2, in FIG. 1. System acquisition. Serial interface RS-232 Plotter Houston Instruments DMP-29 set-up for temperature control and datum which only one probe was used for temperature measurement. Since the thermal mass of the.7-mm-diameter probe is much lower than that of the glass vial, the probe and vial were in thermal equilibrium, a necessary condition for accurate temperature measurement. An external temperature reference (Kaye Instruments model HTR-3), a 1-fl platinum resistance thermometer (Yellow Springs Instrument Co.) and a platinum resistance thermometer monitor (Kaye Instruments model 373) were used to calibrate each Fluoroptic probe before all experiments. The temperature measurements were accurate to +.5 C over the temperature range of interest (1 to 15 C). Automation. A computer system (MacSym 12; Analog Devices) acquired data from the Luxtron thermometer for convenient datum printout. The computer also controlled the microwave generators. As the vial in the thermal oven heated via convection, the computer adjusted the microwave power in real time so that the vial in the microwave oven heated at the same rate as the vial in the thermal oven. The temperatures of samples in the convection and microwave vials tracked within 2 C. The entire experimental setup is shown in Fig. 1. Exposure. To accurately control and measure the exposure temperature, only one pair of sample vials was exposed for each experiment, one vial in the microwave oven and the other in the convection oven. There were, therefore, slight temperature variations (less than 2 C from target temperatures) between tests. The exposure time is defined as the total time the dry-heat vial was situated in the convection dry-heat oven, regardless of temperature. Upon completion of exposure, both the dry-heat vial and the microwave vial were quickly removed from their respective ovens and cooled with a stream of filtered compressed air. Figure 2 shows a typical temperature-time profile of vials exposed to

VOL. 53, 1987 both thermal and microwave treatments at various temperatures. Postcycle cooling is also shown. The microwave power output varied from to 4 kw during experiments. Since the objective of this study was only to determine the possible existence of nonthermal effects of microwaves on microorganisms but not the magnitude of such effects, no attempt was made to measure the microwave field strength or power in the vicinity of the sample. RESULTS Temperature profiles. The actual temperatures achieved in the experiments were 17, 117, 13, and 137 C. Figure 2 shows the representative comparative temperature profiles of both heat sources at 17, 117, 13, and 137 C. Except for a few time points at which the temperature of the microwave oven overshot during tracking, the computer temperature monitoring and measuring systems were able to minimize the temperature difference between the two heat sources to less than 2 C. The narrow range of the temperature variations between the two heat sources assured the validity of the results. Sporicidal activities. Figure 3 shows the representative sporicidal kinetics of convection dry heat versus microwave heat at various temperatures. From these results, it is clear that temperatures below 117 C are not effective for dryproduct sterilization because of a low lethal rate. At 13 and 137 C, the sample come-up time (the time required to reach maximum experimental temperatures) was approximately 2 min. Once the maximum experimental temperatures were reached, the spores were exponentially inactivated. The total exposure times to reduce 15 spores were 75 and 48 min at 13 and 137 C, respectively. There was no significant difference observed between convection and microwave heat in spore inactivation in all temperature ranges studied. The D values (the time needed to reduce the population by 9% at one temperature) at 117, 13, and 137 C were estimated to be 88, 14, and 7 min, respectively. The Z value (increase in temperature required for 9% reduction in D value) was estimated to be 18 C. U) co L- L. E U) 1 2 3 4 5 6 7 Time (minutes) FIG. 2. Representative temperature-time profiles of convection oven- and microwave (MW) oven-heated samples. The microwave heating was designed to track the convection dry heat with minimum temperature variations. MICROWAVE STERILIZATION IN THE DRY STATE 2135 *z D s- o CO) -J 5 > 4-3 > 2 JL_. 5 A AMBw a MW CH och 4 A M 3/. 2 1 II 3 9 15 3 6 9 12 C ^ AMW CH.. 2 4 6 8 1 2 3 4 5 6 Exposure Time (min) Exposure Tiime (mii) 15 13 6 11 o 1D 9 R 7 E 15 13, 11 9 = -1 7 E FIG. 3. Sporicidal kinetics of microwave (MW) versus convection dry heat (CH) at 17 C (A), 117 C (B), 13 C (C), and 137 C (D). The solid line is only for the purpose of visualization. The dotted line is an average temperature profile obtained from two temperature probes of the dry-heat sample of the longest exposure. DISCUSSION D A MW CH The development of a short cycle for microwave sterilization at relatively low temperatures was attempted under the assumption that the nonthermal effects of microwave energy may have an impact on sporicidal activities. In addition, the mechanism of the microwave sporicidal action was studied by comparing the spore inactivation kinetics from convection dry heat with those from microwaves. Under the experimental conditions, microwave heating at temperatures below 117 C was not effective for sterilization of dry medical devices because of a low lethal rate. Furthermore, within the temperature range studied, there was no significant difference in sporicidal activities between the microwave and convection thermal treatments. Both convection and microwave heating showed similar inactivation kinetics, provided that the microwave oven tracked the convection oven at nearly identical temperatures. These results strongly suggest that the sporicidal activities of microwave energy are merely a function of heat. No other factors seem significant enough to influence the sterilization process. Nonthermal effects are not significant in a dry microwave sterilization process. Our experience in preliminary tests revealed difficulties in accurately tracking the microwave oven temperature to the convection oven (data not shown). The difficulties were especially true in experiments dealing with higher temperatures because the lethal rate increases exponentially as a function of temperature. At temperatures higher than 135 C, a slight temperature variation would cause a significant change in sporicidal activities. However, once the accurate temperature tracking was achieved, similar spore inactiva- A\ r,..i r.

2136 JENG ET AL. tion kinetics were repeatedly obtained from both heat sources. The temperature control, therefore, is the key to the experimental evaluation. Various observations found in the earlier literature (5, 12, 15, 23) (U.S. patent 3,753,651; Japanese patent 4786) were believed to be caused by variations in either temperature control or temperature measurement or both. With the aid of the computerized control system and Fluoroptic temperature probes, we have now unequivocally demonstrated that spore inactivation by microwaves in a dry state is caused solely by thermal effects. The current study showed that a total dwelling time of at least 45 min was required to inactivate 15 inoculated dry spores at 137 C by both convection and microwave heating. The Z value of 18 C shows a slight variation from the published Z values of 17.5 (6), 21 (9), or 22 C (13) for still-air convection dry heat on B. subtilis subsp. niger spores and from the Z values of 21 (14) to 2 C (16) for dry-heat sterilization process validation. The sporicidal activities in the present studies are close to those found in the work of Wayland et al. (23). However, the thermal-electromagnetic interdependence found in their studies was not observed. In the study by Wayland et al., it is not clear why the sum of the interaction of microwaves and thermal heat was synergistic below 135 C but antagonistic at temperatures above 135 C. The experimental temperature variations in their study (±4 to ±9 C) could seriously affect the lethal rate (17) in the temperature range studied and thus the validity of their conclusion. Extrapolation from the Z value in the current study shows that inactivation of 16 B. subtilis subsp. niger dry spores would require at least 26.4 min with a D value of 4.4 min at 14 C. These observations varied from the experimental results obtained by Lohman and Manique (12) in which 16 dry spores of the same strain were inactivated by microwaves in less than 3 min cycle time with a constantly rising temperature up to 14 C. In their study, multiple vials were used for microwave treatment. A gradient of the spore inactivation rate among vials situated in different locations was observed, presumably because of heterogeneous heat loss. Within a temperature range of 13 to 14 C, the D value from their work is approximately.1 min. These results drastically disagree with the observed D value of 8 min at 137 C and the calculated D value of 4.4 min at 14 C from the current studies. The heat distribution in vials during microwave treatment was found to be surprisingly uneven. On the same insulated vial, a temperature variation of 4 to 5 C was detected between a distance of only 1 to 2 cm (data not shown). Temperature probes, therefore, must be located at the exact location where the spores are inoculated in order to obtain accurate data. The method of temperature measurement used by Lohman and Manique (12) required an opening of the microwave oven door and removal of the insulated materials. Samples in a microwave oven undergo a rapid decrease in temperature when the oven door is opened. The temperature recorded by the infrared method used by Lohman and Manique after these procedures may therefore significantly underestimate the true temperature of the samples. In dry-heat sterilization cycle development, the mass of the glassware being sterilized essentially determines the length of the come-up time. In general practice, to sterilize an industrial-scale load of glassware by convection dry heat in an overkill approach requires a dwelling time of 2 to 3 h to establish a safety assurance level of 1-6 at 16 C. The exposure cycle would be considerably longer if the come-up APPL. ENVIRON. MICROBIOL. time and heat distribution issues were considered. In our microwave heating system with a constant power output of 4 kw and a single-vial load, the maximum rate of temperature increase from 22 to 14 C is 5 min (data not shown), compared with a minimum of 2 min required for the convection dry-heat oven with the same load. A larger load would further increase the difference in the come-up time between these two methods. It seems, therefore, that the only advantage to microwave sterilization of dry products lies in a shorter come-up time. Once the sterilization maximum temperature is obtained, there are no advantages in sporicidal activities to a microwave sterilization process. ACKNOWLEDGMENTS We are indebted to James M. Fuller and Larry V. Hervey for their excellent technical support in microwave sterilization and microbiological assays. LITERATURE CITED 1. Angellotti, R., J. H. Maryanski, T. F. Butler, J. T. Peeler, and J. E. Campbell. 1968. Influence of spore moisture content on the dry-heat resistance of Bacillus subtilis var. niger. Appl. Microbiol. 16:735-745. 2. Barnes, F. S., and C.-L. J. Hu. 1977. Model of some nonthermal effects of radio and microwave fields on biological membranes. IEEE Trans. Microwave Theory Tech. 25:742-746. 3. Cleary, S. F. 1977. Biological effects of microwave and radiofrequency radiation, p. 121-166. In Critical review in environmental control. CRC Press, Inc., Boca Raton, Fla. 4. Cope, F. W. 1976. Superconductivity-a possible mechanism for non-thermal biological effects of microwaves. J. Microwave Power 11:267-27. 5. Culkin, K. A., and Y. C. Fung. 1975. Destruction of Escherichia coli and Salmonella typhimurium in microwave-cooked soups. J. Milk Food Technol. 38:8-15. 6. Fox, K., and I. J. Pflug. 1968. Effect of temperature and gas velocity on the dry-heat destruction rate of bacterial spores. Appl. Microbiol. 16:343-348. 7. Furia, L., D. W. Hill, and. P. Gandhi. 1986. Effect of millimeter-wave irradiation on growth of Saccharomyces cerevisiae. IEEE Trans. Biomed. Eng. 33:993-999. 8. Goldblith, S. A., and D. I. C. Wang. 1967. Effect of microwaves on Escherichia coli and Bacillus subtilis. Appl. Microbiol. 15:1371-1375. 9. Gurney, T. R., and L. B. Quesnel. 198. Thermal activation of dry-heat inactivation of spores of Bacillus subtilis MD 2 and Bacillus subtilis var. niger. J. Appl. Bacteriol. 48:231-274. 1. Lechowich, R. V., L. R. Beuchat, K. I. Fox, and F. H. Webster. 1969. Procedure for evaluating the effects of 2,45-megahertz microwaves upon Streptococcus faecalis and Saccharomyces cerevisiae. Appl. Microbiol. 17:16-11. 11. Lin, J. C. 1979. Health effects of radio and microwave radiation. J. Environ. Pathol. Toxicol. 2:1413-1432. 12. Lohman, S., and F. Manique. 1986. Microwave sterilization of vials. J. Parenter. Sci. Technol. 4:25-3. 13. Molin, G., and K. Ostlund. 1976. Dry-heat inactivation of Bacillus subtilis var. niger spores with special reference to spore density. Can. J. Microbiol. 22:359-363. 14. National Aeronautics and Space Administration. 1978. Advances in sterilization and decontamination (a survey), chapter 2, p. 37. Publication 515. National Aeronautics and Space Administration, Washington, D.C. 15. Olsen, C. M., C. L. Drake, and S. L. Bunch. 1966. Some biological effects of microwave energy. J. Microwave Power 1:45-56. 16. Parenteral Drug Association, Inc. 1981. Validation of dry heat processes used for sterilization and depyrogenation, p. 35. Technical report no. 3. Parenteral Drug Association, Inc., Philadelphia. 17. Pflug, I. J. 1982. Textbook for an introductory course in the

VOL. 53, 1987 MICROWAVE STERILIZATION IN THE DRY STATE 2137 microbiology and engineering of sterilization processes, 5th ed., p. 11.8. Environmental Sterilization Laboratory, Minneapolis. 18. Rohrer, M. D., and R. A. Bulard. 1985. Microwave sterilization. J. Am. Dent. Assoc. 11:194-198. 19. Rohrer, M. D., M. A. Terry, R. A. Bulard, D. C. Graves, and E. M. Taylor. 1986. Microwave sterilization of hydrophilic contact lenses. Am. J. Ophthalmol. 11:49-57. 2. Rosen, C.-G. 1972. Effects of microwaves on food and related materials. Food Technol. 26:36-4, 55. 21. Stuckly, S. S. 197. Dielectric properties of some granular solids containing water. J. Microwave Power 5:62-68. 22. Vela, G. R., and J. F. Wu. 1979. Mechanism of lethal action of 2,45-MHz radiation on microorganisms. Appl. Microbiol. 37:554-553. 23. Wayland, J. R., J. P. Brannen, and M. E. Morris. 1977. On the interdependence of thermal and electromagnetic effects in the response of Bacillus subtilis spores to microwave exposure. Radiat. Res. 71:251-258. 24. Weast, R. C. (ed.). 1982. Handbook of chemistry and physics, 63rd ed., p. E-46. CRC Press, Inc., Boca Raton, Fla. 25. Webb, N. J., and D. D. Dopps. 1968. Inhibition of bacterial cell growth by 136 gc microwaves. Nature (London) 218:51.