Spectrophotometric Determination of Iron

Similar documents
Experiment 7: Adsorption Spectroscopy I, Determination of Iron with 1,10 Phenanthroline

EXPERIMENT #3 A Beer's Law Study

2014 NJIT RET Program. MODULE TOPIC: Two Methods of Determining the Concentration of Soluble Compounds or Analytes..

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Rate law Determination of the Crystal Violet Reaction Using the Isolation Method

Experiment 1 (Part A): Plotting the Absorption Spectrum of Iron (II) Complex with 1,10- Phenanthroline

Expt 4: Determination of Iron by Absorption Spectrophotometry Calibration solutions: Iron(II)ammonium sulfate = ferrous ammonium sulfate

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Lab #12: Determination of a Chemical Equilibrium Constant

MOLEBIO LAB #4: Using a Spectrophotometer

Experiment 11 Beer s Law

Chem 321 Lecture 18 - Spectrophotometry 10/31/13

Experiment 11 Beer s Law

Chemistry 1215 Experiment #11 Spectrophotometric Analysis of an Unknown Brass Sample

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

Experiment 13H THE REACTION OF RED FOOD COLOR WITH BLEACH 1

Exercise 6: Determination of Hardness of Water

DETERMINATION OF K c FOR AN EQUILIBRIUM SYSTEM

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Colorimetric analysis of aspirin content in a commercial tablet

Experiment 13I THE REACTION OF RED FOOD COLOR WITH BLEACH 1

MORE LIGHTS, COLOR, ABSORPTION!

Determination of the Solubility-Product Constant for a Sparingly Soluble Salt

Chemistry 213. A KINETIC STUDY: REACTION OF CRYSTAL VIOLET WITH NaOH LEARNING OBJECTIVES

Preparation of Standard Curves. Principle

Spectrophotometric Determination of an Equilibrium Constant

Exp 03 - Reaction Rate

REVIEW OF LAB TECHNIQUES

The Fluorometric Determination of Acetylsalicylic Acid in an Aspirin Tablet

Aspirin Lab By Maya Parks Partner: Ben Seufert 6/5/15, 6/8/15

Experiment 7A ANALYSIS OF BRASS

EXPERIMENT 5: PHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Experiment 6: Determination of the Equilibrium Constant for Iron Thiocyanate Complex

LEAD (Colorimetric) 2. Muffle Furnace: Equipped with pyrometer and capable of operating at controlled temperatures up to 500 C

Determining the Concentration of a Solution: Beer s Law

CHM 152 updated May 2011 Lab 6: Experimentally Determining an Equilibrium Constant using Spectrophotometry

Experimental Procedure Lab 402

CHEM Lab 7: Determination of an Equilibrium Constant using Spectroscopy

Introduction to Spectroscopy: Analysis of Copper Ore

Experimental Procedure Overview

Chemistry 141 Laboratory Spectrometric Determination of Iron Concentration Lab Lecture Notes 8/29/2011 Dr. Abrash

REVIEW OF LAB TECHNIQUES

The Synthesis and Analysis of Aspirin

Introduction to Spectroscopy: Analysis of Copper Ore

Spectrophotometry Materials

Introduction. The amount of radiation absorbed may be measured in a number of ways: Transmittance, T = P / P 0 % Transmittance, %T = 100 T

METHOD 9200 NITRATE. 1.2 The applicable range of concentration is 0.1 to 2 mg NO -N per liter. 3 of sample.

Determination of Orthophosphate Ion

Determination of an Equilibrium Constant

Experiment 12H, Parts A and B

'.. _~ - BAUSCH & LOMB {i) ANALYTICAL SYSTEMS DIVISION

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Lab Investigation 4 - How could you make more of this dye?

ASCORBIC ACID METHOD FOR PHOSPHORUS DETERMINATION

Determination of Orthophosphate Ion

CHEM 334 Quantitative Analysis Laboratory

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Introduction to Spectroscopy: Analysis of Copper Ore

VISIBLE SPECTROSCOPY

The ROXI Colorimeter & Fluorimeter. Laboratory Application I. Colorimetric measurements via Beer s Law.

Standard Methods for the Examination of Water and Wastewater

For this lab, you will determine the purity of the aspirin by titration and by spectrophotometric analysis.

Concepts, Techniques. Concepts, Techniques 9/11/2012. & Beer s Law. For a simple, transparent, COLORED material, e.g. ROYGBV

Chem 321 Name Answer Key D. Miller

Determination of the Equilibrium Constant. HIn - + H 2 O = H 3 O + + In 2-, [H 3 O + ][In 2- ] K ai = [HIn - ]

Chemistry 112 SPECTROPHOTOMETRIC DETERMINATION OF AN EQUILIBRIUM CONSTANT

Answers to spectroscopy questions. 1. Consider the spectrum below. Questions a f refer to this spectrum.

1. Preliminary qualitative analysis of unknown substances (liquid or solid).

Introduction to Spectroscopy: Analysis of Copper Ore

Atomic Theory: Spectroscopy and Flame Tests

NITROGEN, TOTAL, HR, Test N Tube

Chemical Equilibrium: Finding a Constant, Kc

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

A Spectrophotometric Analysis of Calcium in Cereal

Spectrophotometry. Introduction

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

AP Chemistry Laboratory #15: Reaction Rate of Crystal Violet and Sodium Hydroxide. Lab days: Thursday and Friday, February 1-2, 2018

Determining the Concentration of a Solution: Beer s Law. Evaluation copy. Figure 1

Atomic Theory: Spectroscopy and Flame Tests

2 SPECTROSCOPIC ANALYSIS

Experiment #7. Determination of an Equilibrium Constant

Reflection = EM strikes a boundary between two media differing in η and bounces back

Determining the Concentration of a Solution: Beer s Law

Lab #16: Determination of the Equilibrium Name: Constant of FeSCN 2+ Lab Exercise. 10 points USE BLUE/BLACK INK!!!! Date: Hour:

1901 Application of Spectrophotometry

Spectroscopy. Page 1 of 8 L.Pillay (2012)

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

Introduction to Spectroscopy: Analysis of Copper Ore

Atomic Absorption Spectroscopy and Atomic Emission Spectroscopy

NITROGEN, TOTAL, Test N Tube * Method 10071

Chemical Equilibrium: Finding a Constant, Kc

Investigating Food Dyes in Sports Beverages. Sample

Atomic Theory: Spectroscopy and Flame Tests

AP Biology Lab 4 PLANT PIGMENTS AND PHOTOSYNTHESIS

Available online Research Article

A Study of Beer s Law Prelab

For simplicity, we ll represent BTB s ionization in a solution by the equilibrium: HBTB = H + + BTB -

LAB 05 Enzyme Action

Transcription:

Spectrophotometric Determination of Iron INTRODUCTION Many investigations of chemical species involve the interaction between light and matter. One class of these investigations, called absorbance spectrophotometry, involves the transfer of energy from a photon of light to an analyte to produce an excited state species. By accounting for the quantity of light absorbed, the amount of analyte can be determined. Common analytical applications of spectrophotometry take advantage of the Beer-Lambert Law (often simply called Beer s Law) that relates the concentration of an analyte to its absorbance of monochromatic radiation as shown below: A = -log T = bc Where A is absorbance, T is transmittance, is a constant called the molar absorptivity, b is the path length through the sample, c is analyte concentration. In practice, instruments monitor the transmittance of light by comparing the amount of light that leaves the light source, passes through the absorbing solution and strikes a detector (P) to the amount of light that strikes the detector when a non-absorbing solution is in the beam path (P 0 ). The ratio of these responses is the transmittance (or percent transmittance). P T or %T 100% x P P 0 P 0 This ratio assumes that the response at the detector in the absence of any light is zero, which if often not the case. Most detectors have a nonzero response, even if no light is striking the detector, called the dark current (P dark ), which must be compensated for in order for the transmittance (and absorbance) value to be meaningful. T P P dark P 0 P dark or %T 100% x P P dark P 0 P dark Therefore, to make a useful measurement, the instrument must know three things (1) the response when no light strikes the detector (P dark ), (2) the response when a blank is in the beam path (P 0 ), and (3) the response when the sample is in the beam path (P). Ideally, P dark and P 0 are constant for a particular instrument under a particular set of conditions. Therefore, by measuring P at a single wavelength for a series of analyte solutions of known concentration and using P 0 and P dark to calculate the absorbance (or transmittance) it is possible prepare a calibration curve for the analyte response. The Spectronic 20 is an example of a direct-reading, single-beam spectrophotometer. By virtue of being direct-reading, spectrophotometers feature fast operation, and because of their simple design they are relatively inexpensive and require a minimum of maintenance. They are particularly well suited for spectrophotometric determinations of a single component at a single wavelength where only moderate accuracy ( 1 to T) is required. The Spec-20 has only three controls (1) an On/Off/0%T knob used to set zero percent transmittance when the sample chamber is empty, (2) a 100% T knob used to set 100 percent transmittance when a blank is in the sample chamber, and (3) a wavelength control used to set the desired wavelength for the

measurement. Output on the Spec -20 is through an analog display that has both percent transmittance and absorbance scales. When recording data, it is wise to record both transmittance and absorbance values. The limitations of the single-beam instrument become apparent when an absorption spectrum over a wavelength range is required. The response of the phototube, the emissivity of the light source, and the intensity of the light diffracted by the grating are all a function of wavelength. Accordingly, in order to obtain the absorption spectrum of a compound, the instrument must be recalibrated each time the wavelength setting is changed. The reaction between ferrous ion and 1,10-phenanthroline to form a red complex serves as a sensitive method for determining iron. Fe 2+ 2+ + 3 phen Fe(phen) 3 N N 1,10 phenanthrolene (phen) The molar absorptivity of the complex, [(C 12 H 8 N 2 ) 3 Fe] 2+, is 11,100 at 508 nm. The intensity of the color is independent of ph in the range 2 to 9. The complex is very stable and the color intensity does not change appreciably over long periods of time. Beer's law is obeyed. The iron must be in the ferrous state, and hence a reducing agent is added before the color is developed. Hydroxylamine, as its hydrochloride, can be used to reduce any ferric ion that is present: 2 Fe 3+ + 2 NH 2 OH + 2 OH - 2 Fe 2+ + N 2 + 4 H 2 O The ph is adjusted to a value between 6 and 9 by addition of ammonia or sodium acetate. An excellent discussion of interferences and of applications of this method is given by Sandell.1

HANDLING OF CUVETTES The handling of the cuvettes is extremely important. Often two cuvettes are used simultaneously, one for the "blank" solution and one for the samples to be measured. Yet any variation in the cuvette (such as a change in the cuvette width or curvature of the glass, stains, smudges, or scratches) will cause varying results. Thus, it is essential to follow several rules in dealing with cuvettes: 1. Do not handle the lower portion of a cuvette through which the light beam will pass. 2. Always rinse the cuvette with several portions of the solution before taking a measurement. 3. Wipe off any liquid drops or smudges on the lower half of the cuvette with a clean Kimwipe (or other lens paper) before placing the the cuvette in the instrument. Never wipe the cuvette with paper towels or handkerchiefs. Inspect the cuvette to ensure that nor bubbles are clinging to the inside walls. 4. When inserting a cuvette into the sample holder: a. To avoid any possible scratching of the cuvette in the optical path, insert the cuvette with the index line facing toward the front of the instrument. b. After the cuvette is seated, line up the index lines exactly. 5. When using two cuvettes simultaneously, use one of the cuvettes always for the blank solution and the other for the various samples being measured. Mark the cuvette ( on the upper portion) accordingly and do not interchange the cuvettes during the remainder of the experiment. 6. Do not use a test tube brush to clean the cuvettes after the experiment is complete. REAGENTS 1,10-phenanthroline (0.1 g of 1,10-phenanthroline monohydrate in 100 ml of distilled water, warming to effect solution if necessary). Hydroxylamine hydrochloride (10 g in 100 ml of distilled water). Sodium acetate (10 g in 100 ml of distilled water). Ferrous ammonium sulfate hexahydrate. SAMPLE PREPARATION 1. Weigh accurately about 0.07 g of pure ferrous ammonium sulfate hexahydrate, dissolve it in water, and transfer the solution to a 1-liter volumetric flask. Add 2.5 ml of concentrated sulfuric acid and dilute the solution to the mark. Calculate the concentration of the solution in mg of iron per ml. (Remember, your solution was prepared using Fe(NH 4 ) 2 (SO 4 ) 2 6H 2 O). 2. Prepare the unknown sample as follows. Add about 0.12 g of the solid unknown and approximately 0.25 ml concentrated sulfuric acid into a 100 ml volumetric flask and dilute to the mark. Now transfer a 1 ml aliquot of this solution to another 100 ml volumetric flask

- do not dilute yet. This will be referred to as the "prepared unknown". 3. Into five 100 ml volumetric flasks, pipet (volumetrically) 1, 5, 10, 25, and 50 ml portions of the standard iron solution. Put 50 ml of distilled water into another flask to serve as the blank. To each flask, including the "prepared unknown", add 1 ml of the hydroxylamine solution, 10 ml of the 1,10 phenanthroline solution and 8 ml of the sodium acetate solution. Then dilute all the solutions to the 100 ml marks and allow them to stand for 10 minutes with occasional shaking of the flasks. PART A -- Determination of Spectral Response of a Single-Beam Spectrophotometer The purpose of this section is to familiarize you with the Spectronic 20 operation and to illustrate the inherent impact of the performance of the light source and detector on a measurement. The light source in the Spectronic 20 is a type S-4 (cesium-antimony) phototube. The relative response of the phototube to a beam of monochromatic light of constant intensity is greatest near 400 nm and decreases fairly rapidly above 475 nm. Its response at 625 nm is only 5% of that at its maximum near 400 nm. PROCEDURE The Spectronic 20 should be allowed to warm-up at least 10 minutes before use. 1. First, adjust the instrument, photocell dark (i.e. no cuvette in the sample holder), with the zero control. 2. Insert a cuvette filled with water into the sample holder. Rotate the 100% control until the meter needle reads near mid-scale. Now rotate the wavelength control in the range between 350 and 625 nm, adjusting the 100% control to allow the meter to stay on scale throughout the entire wavelength range. Note the wavelength where the transmittance is maximum (needle deflection is greatest). 3. Carefully balance the instrument at 100% transmittance at the wavelength of maximum transmittance. You should now be able to rotate the wavelength control between 350 and 625 nm without the meter going off scale. 4. Without changing either the dark current control or the 100%T control, vary the wavelength from 350 to 625 nm in 25 nm intervals and record the instrument response. 5. Plot the instrument response vs. wavelength. PART B -- Determination of Iron with 1,10-phenanthroline In this portion of the experiment, you will record the absorbance spectrum of the ironphenanthroline complex by measuring the absorbance as a function of wavelength, and use this data to select a wavelength at which to measure the rest of your standards and unknowns. Since you want your absorbance values to be meaningful, you will need to set zero transmittance to account for dark current and 100% transmittance with a blank each time you change wavelength. PROCEDURE 1. Using the blank as a reference and any one of the iron solutions prepared above, measure the absorbance at 20-nm intervals in the range from 400 to 600 nm. (Note that it is necessary to

re-adjust the 0% T and 100%T settings whenever the wavelength is changed). 2. Once you have reached 600 nm, return to the region where absorbance was maximum and collect absorbance readings at 5 nm intervals in this region. 3. Plot the absorbance vs. wavelength and connect the points to from a smooth curve. Select the proper wavelength to use for the determination of iron with 1,10-phenanthroline and calculate the molar absorption coefficient,, at the wavelength of maximum absorption ( max ) on the absorption curve (assume b = 1 cm). 4. Measure the absorbance of each of the standard solutions and the unknown at the selected wavelength. Plot the absorbance vs. the concentration of the standards. Note whether Beer's law is obeyed. Using the absorbance of the unknown solution calculate the % (w/w) iron in your original solid sample and its 95% confidence interval, remember to correct for dilutions. (1) E. B. Sandell, Colorimetric Determination of Traces of Metals, 3rd Ed. Interscience Publishers, Inc., New York, 1959.