Using the Coriolis Acceleration to Design a Digital Mass Flow Meter

Similar documents
By-Pass. This voltage is proportional to the liquid level (threewire potentiometer circuit). The resistance reading can

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

Applied Fluid Mechanics

Applied Fluid Mechanics

CORIOLIS MASS FLOW METERS FOR GAS AND LIQUID MEASUREMENT. Michael Keilty (Presented by Hubbard Donald) Endress + Hauser Flowtec AG

An Essential Requirement in CV Based Industrial Appliances.

Today s menu. Last lecture. Measurement of volume flow rate. Measurement of volume flow rate (cont d...) Differential pressure flow meters

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class #

DEVELOPMENT OF DROP WEIGHT IMPACT TEST MACHINE

Time of Flight Sensor with a Flow Parallel Wire

ELECTRONIC FLOWMETERS FOR THERMAL ENERGY MEASUREMENT. By Dr. Crainic Monica Sabina

ME411 Engineering Measurement & Instrumentation. Winter 2017 Lecture 11

Flow Measurement: Physical principle employed in various types of flow meters. PART 4 PREPARED BY: ESO OLUWATOBILOBA

1. a) A flag waving in the breeze flaps once each s. What is the period and frequency of the flapping flag?

FLOW MEASUREMENT INC 331 Industrial process measurement 2018

Transducers. ME 3251 Thermal Fluid Systems

NAME: PHYSICS 6B SPRING 2011 FINAL EXAM ( VERSION A )

Process Control and Instrumentation Prof. D. Sarkar Department of Chemical Engineering Indian Institute of Technology, Kharagpur

DESIGN AND PERFORMANCE OF THE CONVERGING-DIVERGING VORTEX FLOWMETER

Unit 57: Mechatronic System

Overview. Sensors? Commonly Detectable Phenomenon Physical Principles How Sensors Work? Need for Sensors Choosing a Sensor Examples

10 Measurement of Acceleration, Vibration and Shock Transducers

Instrumentation & Data Acquisition Systems

Water Circuit Lab. The pressure drop along a straight pipe segment can be calculated using the following set of equations:

Coriolis Mass Flowmeter. (MCM Series) 100% Customer Satisfaction MAXIFLO TM. Coriolis Mass Flowmeter Series MCM

Control Engineering BDA30703

Flow Monitoring Technologies in Oil and Gas. Don Ford, Technical Specialist Spartan Controls Ltd.

Fundamentals of Flow Metering

Laboratory work No 2: Calibration of Orifice Flow Meter

VALLIAMMAI ENGINEERING COLLEGE

spring magnet Fig. 7.1 One end of the magnet hangs inside a coil of wire. The coil is connected in series with a resistor R.

UNIVERSITY OF CAMBRIDGE INTERNATIONAL EXAMINATIONS General Certificate of Education Advanced Level

CHAPTER (13) FLOW MEASUREMENTS

Given Find Water Properties

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Transducers. EEE355 Industrial Electronics

In which vector triangle does the vector Z show the magnitude and direction of vector X Y?

M13/4/PHYSI/HPM/ENG/TZ1/XX. Physics Higher level Paper 1. Monday 6 May 2013 (morning) 1 hour INSTRUCTIONS TO CANDIDATES

INTERNATIONAL STANDARD

Cambridge International Examinations Cambridge International Advanced Level

Calibrating a Pressure Transducer, Accelerometer, and Spring System

4. What is the speed (in cm s - 1 ) of the tip of the minute hand?

MX Series COMPACT AND USER FRIENDLY. Metal Tube Type Variable Area Flowmeter OUTLINE STANDARD SPECIFICATION MODEL CODE MX-400, MX-71 MX-52E

Circular Motion and Centripetal Force

A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS20

Sensing and Sensors: Fundamental Concepts

SIMPLE PENDULUM AND PROPERTIES OF SIMPLE HARMONIC MOTION

Development of Laser Thickness Gauge in Steel Plate Shearing Line

Applicable Fluid. Accuracy. Flow Rate Range. Fluid Temperature. Fluid Viscosity. Connection Size. Flange Rating. Housing. Material. Blade.

A longitudinal wave travels through a medium from left to right.

Theory and Design for Mechanical Measurements

A beam of coherent monochromatic light from a distant galaxy is used in an optics experiment on Earth.

PHYA4/2. General Certificate of Education Advanced Level Examination June Unit 4 Fields and Further Mechanics Section B

Magnetic #uids in aerodynamic measuring devices

MAHATMA GANDHI MISSION S JAWAHARLAL NEHRU ENGINEERING COLLEGE, FLUID MECHANICS LABORATORY MANUAL

US06CPHY06 Instrumentation and Sensors UNIT 2 Part 2 Pressure Measurements

Answer three questions from Section A and five questions from Section B.

1. For which of the following motions of an object must the acceleration always be zero?

Given Find Water Properties

Teacher s notes 19b An investigation into the energy changes occurring in a pendulum swing

2. (a) Differentiate between rare metal thermocouples and base metal thermocouples.

Linearity of Coriolis Flowmeters at Full Scale Value. Dr. Jesús Aguilera - Endress + Hauser Paul Ceglia - Endress + Hauser

Physics 101. Hour Exam 3 Spring Last Name: First Name Network-ID Discussion Section: Discussion TA Name:

INC 331 Industrial Process Measurement. Instrument Characteristics

Physical Quantities and Units

Year 7 Recall Booklet. Name: Class:

UNIT 1 MODULE 2: OSCILLATIONS AND WAVES GENERAL OBJECTIVES EXPLANATORY NOTES SPECIFIC OBJECTIVES. On completion of this Module, students should:

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

Design and construction of an automatic coefficient of friction measuring device

Amir Hossein Parvardi 1

Physics 6b Winter 2015 Final Campagnari Section Test Form D

Physics 6b Winter 2015 Final Campagnari Section Test Form A

Application of Density Meters to Liquid Hydrocarbon Measurement

Analysis and Experiments of the Linear Electrical Generator in Wave Energy Farm utilizing Resonance Power Buoy System

Standing waves [49 marks]

Available online at ScienceDirect. Procedia Engineering 150 (2016 )

arxiv: v2 [physics.acc-ph] 21 Jun 2017

Centripetal and centrifugal force

N09/4/PHYSI/SPM/ENG/TZ0/XX+ Physics Standard level Paper 1. Monday 16 November 2009 (afternoon) 45 minutes INSTRUCTIONS TO CANDIDATES

Siddharth Institute of Engineering & Technology

AP Physics B Summer Assignment

11 SEPTEMBER This document consists of printed pages.

Module I Module I: traditional test instrumentation and acquisition systems. Prof. Ramat, Stefano

SILVER OAK COLLEGE OF ENGINEERING & TECHNOLOGY

EF 152 Final Exam - Fall, 2016 Page 1 Copy 169

Ultrasonic Gas Flow Meter

Cooling Water Flow Regulator

Last Name: First Name Network-ID Discussion Section: Discussion TA Name: Write your seat number on the answer sheet

Auxiliary Fluid Flowmeter

FINAL DRAFT: October 25, 2010 API TECHNICAL REPORT FUEL GAS MEASUREMENT. FINAL DRAFT: October 25, 2010 Page 0

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz

Loss analysis of a 1 MW class HTS synchronous motor

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

PIZZI FRANCO via Ripoli, 207/E Firenze - Italy. Instruments and Systems for Dam Monitoring and Structural Controls

INTEGRATED THROUGHFLOW MECHANICAL MICROFLUIDIC SENSORS

ABB Flowmeters Review of industrial processing flowmeters

Chapter 6. Fiber Optic Thermometer. Ho Suk Ryou

Installation effects on an ultrasonic flow meter with implications for self diagnostics

Transcription:

Global Journal of Scientific Researches Available online at gjsr.blue-ap.org 2016 GJSR Journal. Vol. 4(2), pp. 9-15, 30 April, 2016 E-ISSN: 2311-732X Using the Coriolis Acceleration to Design a Digital Mass Flow Meter Hossein Safaeifar 1* and Fazel Nouroozian 2 1- Engineering Department, Bozorgmehr University of Qaenat, Qaen, Iran 2- Mechanical Engineer. * P.O.B. 311, Qaen, Iran Corresponding Author: Hossein Safaeifar Received: 25 March, 2016 Accepted: 13 April, 2016 Published: 30 April, 2016 A B S T R A C T There are several ways to measure fluid flow can be generally divided into three groups: analyzed, volumetric and the mass. In this project, the design and construction of a mass flow meter using the Coriolis acceleration is analyzed. In this design, measurement of the mass of the fluid is done without the involvement of other physical properties by an appropriate theory and design. So this flow meter is able to measure the mass flow with less error than other flow meters. Also one can measure the various fluids flow rates by using various valves. It will be cost affordable. The springs are used on sensor to change the measuring range can be changed. Using a spring with larger k, the measuring range increases. Keywords: Coriolis acceleration, mass flow meter. 2016 GJSR Journal All rights reserved. INTRODUCTION When the mass flow meter is based on Coriolis force, it is named Coriolis flow meter. Many industrial processes require accurate measurement of mass flow. Commonly used flow meters often measure volume of flow. Then, using the volume and density, mass of fluid can be calculated by equation 1. Mass=Volume*Density Eq. 1 When density is constant, volume of fluid can be converted to mass easily. But changes in temperature, pressure and viscosity can be effect on the density and would conclude lower accuracy at the mass measurement. For example, in areas where temperatures are high in summer, amount of mass of fuel that is showed by car screen, is more than what it imports fuel tank. So the difference between the amount of fuel indicated by the fuel station and the amount of fuel in the vehicle fuel tank is the reduction in the density of the fuel by the temperature decrement. We need measure mass directly to fix this problem. By using Coriolis mass flow meter, mass of fluid is calculated by pipe harmonic movement. Changes in the direction of flow by the tube vibration are caused the Coriolis force. Coriolis force that the tube is inserted in the y direction is calculated by designed sensor, then the mass is measured using this force. Advantages of this flow meter can be as follows: Providing advanced diagnostic tools to measure fluid conditions The ability to measure mass, volume, temperature and velocity of the fluid as well as the different conditions in a single device Accurate and repeatable data delivery in a wide range of flow rates in different situations

2- History Review of Coriolis flow meters that has been made clear that the building of these flow meters consist of two tubes with vibrating properties and the torsion of tube is measured by the detector system. Twin-tube design that fluid flows in two parallel pipes to remove the effects of vibration effectively. Tube of flow meter is the only part of Coriolis flow meter that is in contact with the liquid and typically is made of stainless steel. Vibration caused by pipe is recorded using chart of recorder devices. This process was done manually previously. With the advancement of technology, this process is done using the software is no longer necessary to use paper. This deices have an important memory and all changes resulting from inputs are kept in this memory. These devices have a screen that can be customized for each operator to observe the quantity [1]. 3- Methods of Measurement Methods for measuring fluid flow can be divided into three groups: A) Inferential flow meter method B) Volumetric flow meter method C) Mass flow meter method (A) Inductive flow meter This method is used in most fluid flow measurements. In this method, flow rate of the fluid can be calculated by measuring the different properties that have are in the fluid. Analytical methods for measuring flow con be used as follows: 1) The liquid column (inferred by measuring the pressure difference created by the barrier fluid path) 2) Target flow meter (inferred by measuring the force) 3) Variable area flow meter (inferred by measuring the position of the float) 4) Magnetic flow meter (inferred by measuring the resulting voltage of the working fluid passes through a magnetic field) 5) Turbine flow meter (inferred by a factor proportional to the velocity) 6) Vortex flow meter (perception of vibration, pressure or temperature) 7) Ultrasonic flow meter (perception of sound intensity) (B) Volumetric flow meter If the device is able to measure the volume of fluid filled and emptied regularly provided that the remaining fluid volume is constant and the number of full and empty container can be measured, the flow rate can be determined. One of the most common methods is positive displacement flow meter that is used to measure the products they are buying or selling. (C) Mass flow meter The Coriolis devices are used in many factories. Although this device has a high price, but because it can measure the mass of the fluid directly, used in some operations. Also in some units, this type of flow meter can be used to measure the fluid to turn with integrated valves. The operation of this device is measuring the mass of fluid without the involvement of other physical properties [1]. 4- Building of Flow Meter In order to design accurate flow meter building and prevent the creation of problems and also saving time, at first a simple model of flow meter was made. With the help of this model, the exact location and dimensions of the flow meter sensors was estimated. This model is made of a bracket and a mechanism for generating oscillating motion, as shown in Figure 1. In this simple model oscillating motion have done manually. But at the original design, an engine with adjustable angular velocity is used to provide constant oscillating motion in order to increase measurement accuracy. 10 P a g e

Figure 1. Simple model After several tests and check the results, the size and exact location of the flow meter gauge force sensor and the maximum measurable flow rate was estimated as follows: Maximum of flow rate: 2.260 kg/s Dimensions: 320*185*200 mm 3 Then, final modified plan was drawn using Solid Works software. The overall building of this Coriolis flow meter is shown in figure 2. As can be seen in the figure 2, construction of the flow meter is as follows: 1) A vibrating tube that can be said is a major component of the flow meter 2) Small engine leading to swing tube by adjustable angular velocity 3) Mechanism for transmission oscillating motion of the motor to the tube 4) Sensor of measuring the Coriolis force 5) Screen 6) The main body that the other components are installed on it Figure 2. Building of final Corilis flow meter In Figure 3, the location of the vibrating tube is shown by the blue line which is swing using designed mechanism. Figure 3. The Location of Vibrating Tube 11 P a g e

Oscillation motion does by an external force which is funded by a small engine. This engine is installed on the outside of building of flow meter. Axes of mechanisms is connected to engine by a belt, thus oscillating motion of the motor is transmitted to axes. 5- The creation of the Coriolis force Coriolis effect is a force overnight Coriolis force which distorts moving objects out of the straight line, from the perspective of an observer in a rotating device. As shown in Figure 4, it is visible that the flow meter has a vibrating pipe which has the adjustable swing speed. If fluid be at rest, two points A and B shown in the figure 4 oscillate in a swing phase and its period is proportion to the angular velocity of engine. If fluid is at moving state, difference phase occurs between the sensor A and B. The cause of this difference phase is the Coriolis force. This force is shown at figures 4 and 5 for counter clockwise and clockwise rotations respectively. Figure 4. Corilis force at counter clockwise rotation Figure 5. Corilis force at clockwise rotation 6- The effect of the Coriolis force on the mass of flow The Coriolis force leading to the creation of a phase shift in two part of tube as shown at figure 6. A fixed amount of mass is passes between signals A and B signal. If a bubble is firmed in path of fluid, amount of mass reduces and therefore Coriolis force reduces and distracting of pipe is reduced. 12 P a g e

Figure 6. The phase difference between points A and B The mass of flow is proportional to the phase difference between two points A, B. The phase difference can be obtained over the time by equation 2 where Δθ id the phase difference between two points and w is the angular velocity. t = θ w Eq. 2 The amount of fluid mass flow is proportional to the distance between the sensors 1 and 2 as equation 3 where the calibration factor (k) can be calculated according to boundary conditions. m = k t Eq. 3 Considering the tube as a volume control as shown at figure 7, mass of flow can be calculated using Coriolis force by equation 4. Figure 7. Considering Tube as a Volume Control a r = r w 2 a t = 2 w v Eq. 4 dm = ρv = ρax F coriolis = ρax 2wv mass flow = F coriolis 2 w x The components of the measurement system are as follows: (A) Atoniks Sensor (B) AVR (C) Screen In this method, proximity sensor of Atoniks is used for measuring the flow rate. Number of oscillation sent to the AVR by this sensor. The heart of a measurement system is AVR that all calculations are done in it. An example of AVR (ATMEGA16) is shown in figure 8. The output mass can be shown in screen using programming. 13 P a g e

Figure 8. ATMEGA16 The sensor that is used in this flow meter is composed of a variable resistor and a spring as shown in figure 9. This sensor measure the Coriolis force and so mass of flow can be determined using equation 4. Figure 9. Sensor for measuring the force 7- Discussion and conclusion Finally, after assembly the components of flow meter as shown in Figure 10, the designed flow meter is tested. Results of test are shown in table 1. Figure 10. Final Designed Flow Meter 14 P a g e

Row 1 2 3 Table 1. Results of tests The measured Mass The Actual Mass 3.7 kg 5 kg 9.2 kg 8 kg 16 kg 10 kg Percentage of error 26% 15% 60% The measured mass at row 3 has high difference from actual mass. This difference is because of decrease in the measurement accuracy of rheostat that is because the rheostat became hot. Using rheostat by high quality can be solve this problem. REFERENCES 1. Wouter Sparreboom, Jan van de Geest, Marcel Katerberg, Ferry Postma, Jeroen Haneveld, Jarno Groenesteijn, Theo Lammerink, Remco Wiegerink, Joost Lötters, Compact Mass Flow Meter based on a Micro Coriolis Flow Sensor:, Micromachines 2013, 4, 22-33; doi:10.3390/mi4010022. 2. Whitesides G.M., The origins and future of microfluidics, Nature 2006, 442, pp. 368 373. 3. Nguyen N.T., Micromachined flow sensors-a review, Flow Meas. Instrum. 1997, 8, pp. 7 16. 4. Haneveld, J.; Lammerink, T.S.J.; Dijkstra, M.A.; Drogendijk, H.; de Boer, M.J.; Wiegerink, R.J., Highly sensitive micro coriolis mass flow sensor, Proceedings of the 21st IEEE International Conference on Micro Electro Mechanical Systems, Tucson, AZ, USA, November 2008, pp. 920 923. 5. Haneveld, J.; Lammerink, T.S.J.; de Boer, M.J.; Sanders, R.G.P.; Mehendale, A.; Lötters J.C., Dijkstra M., Wiegerink R.J., Modeling, design, fabrication and characterization of a micro Coriolis mass flow sensor, J. Micromech. Microeng. 2010, 20, doi:10.1088/0960-1317/20/12/125001. 15 P a g e