LIQUID VISCOSITY DETERMINATION BY CORIOLIS FLOW METER

Similar documents
ME 305 Fluid Mechanics I. Part 8 Viscous Flow in Pipes and Ducts. Flow in Pipes and Ducts. Flow in Pipes and Ducts (cont d)

Determination of Pressure Losses in Hydraulic Pipeline Systems by Considering Temperature and Pressure

Water Circuit Lab. The pressure drop along a straight pipe segment can be calculated using the following set of equations:

ME 305 Fluid Mechanics I. Chapter 8 Viscous Flow in Pipes and Ducts

Non Newtonian Fluid Dynamics

Major and Minor Losses

Lesson 37 Transmission Of Air In Air Conditioning Ducts

CHARACTERISTICS OF SILICA SLURRY FLOW IN A SPIRAL PIPE. Keywords: Drag reduction; Particle concentration; Pitch ratio; Silica slurry flow; Spiral pipe

Hydraulics and hydrology

FORMULA SHEET. General formulas:

EXPERIMENT No.1 FLOW MEASUREMENT BY ORIFICEMETER

Countercurrent heat exchanger

Fig.8-1 Scheme of the fluidization column

Reynolds, an engineering professor in early 1880 demonstrated two different types of flow through an experiment:

CHME 302 CHEMICAL ENGINEERING LABOATORY-I EXPERIMENT 302-V FREE AND FORCED CONVECTION

Q1 Give answers to all of the following questions (5 marks each):

The Mechatronics Design for Measuring Fluid Friction Losses in Pipe Flows Rıza Gurbuz

FLUID MECHANICS PROF. DR. METİN GÜNER COMPILER

8.1 Technically Feasible Design of a Heat Exchanger

CHAPTER 3. CONVENTIONAL RHEOMETRY: STATE-OF-THE-ART. briefly introduces conventional rheometers. In sections 3.2 and 3.

Study fluid dynamics. Understanding Bernoulli s Equation.

ME332 FLUID MECHANICS LABORATORY (PART II)

Chapter 6. Losses due to Fluid Friction

Piping Systems and Flow Analysis (Chapter 3)

Experimental Study on Port to Channel Flow Distribution of Plate Heat Exchangers

LEAKLESS COOLING SYSTEM V.2 PRESSURE DROP CALCULATIONS AND ASSUMPTIONS

Nicholas J. Giordano. Chapter 10 Fluids

Applied Fluid Mechanics

ME332 FLUID MECHANICS LABORATORY (PART I)

CVE 372 HYDROMECHANICS EXERCISE PROBLEMS

Laboratory work No 2: Calibration of Orifice Flow Meter

1-Reynold s Experiment

Experiment (4): Flow measurement

B.E/B.Tech/M.E/M.Tech : Chemical Engineering Regulation: 2016 PG Specialisation : NA Sub. Code / Sub. Name : CH16304 FLUID MECHANICS Unit : I

CHAPTER THREE FLUID MECHANICS

5. MODELING OF NON-STRATIFIED MIXTURE FLOWS (Pseudo-homogeneous flows)

Lesson 6 Review of fundamentals: Fluid flow

Instruction Manual. Equipment for Engineering Education

THE INFLUENCE OF THERMODYNAMIC STATE OF MINERAL HYDRAULIC OIL ON FLOW RATE THROUGH RADIAL CLEARANCE AT ZERO OVERLAP INSIDE THE HYDRAULIC COMPONENTS

Steven Burian Civil & Environmental Engineering September 25, 2013

SCHOOL OF CHEMICAL ENGINEERING FACULTY OF ENGINEERING AND TECHNOLOGY SRM UNIVERSITY COURSE PLAN

Annubar Primary Element Flow Calculations

V/ t = 0 p/ t = 0 ρ/ t = 0. V/ s = 0 p/ s = 0 ρ/ s = 0

What s important: viscosity Poiseuille's law Stokes' law Demo: dissipation in flow through a tube

UNIT II Real fluids. FMM / KRG / MECH / NPRCET Page 78. Laminar and turbulent flow

S.E. (Mech.) (First Sem.) EXAMINATION, (Common to Mech/Sandwich) FLUID MECHANICS (2008 PATTERN) Time : Three Hours Maximum Marks : 100

FACULTY OF CHEMICAL & ENERGY ENGINEERING FLUID MECHANICS LABORATORY TITLE OF EXPERIMENT: MINOR LOSSES IN PIPE (E4)

MECHANICAL PROPERTIES OF FLUIDS:

Centrifugal pumps - characteristics

COURSE CODE : 3072 COURSE CATEGORY : B PERIODS/ WEEK : 5 PERIODS/ SEMESTER : 75 CREDIT : 5 TIME SCHEDULE

Evaluating New Generation Vibrating Tube Sensor for Density Measurement under Process Conditions

Investigation of Thermal-Hydraulic Characteristics of Pillow Plate Heat Exchangers Using CFD

Fluid Flow Analysis Penn State Chemical Engineering

Performance Assessment of PV/T Air Collector by Using CFD

Fluids. Fluids in Motion or Fluid Dynamics

Chapter 6. Losses due to Fluid Friction

Chapter (6) Energy Equation and Its Applications

Universität Duisburg-Essen Fakultät für Ingenieurwissenschaften WS 2012 Maschinenbau, IVG, Thermodynamik Dr. M. A. Siddiqi

Evaluation of pump characteristic from measurement of fast deceleration

LDE/LME/LMI pressure sensors in bypass configuration for gas flowmeters

Pipe Flow/Friction Factor Calculations using Excel Spreadsheets

Applied Fluid Mechanics

AGITATION/GAS-LIQUID DISPERSION. CHEM-E Fluid Flow in Process Units

Calculation of Power and Flow Capacity of Rotor / Stator Devices in VisiMix RSD Program.

FE Exam Fluids Review October 23, Important Concepts

Calculation of Pipe Friction Loss

Heat Exchangers for Condensation and Evaporation Applications Operating in a Low Pressure Atmosphere

Applied Fluid Mechanics

Figure 3: Problem 7. (a) 0.9 m (b) 1.8 m (c) 2.7 m (d) 3.6 m

FLOW MEASUREMENT. INC 102 Fundamental of Instrumentation and Process Control 2/2560

REE 307 Fluid Mechanics II. Lecture 1. Sep 27, Dr./ Ahmed Mohamed Nagib Elmekawy. Zewail City for Science and Technology

Hydraulic Fundamentals Hydraulics Definition Advent of Oil Hydraulics Pascal s Law Pressure

Modelling of dispersed, multicomponent, multiphase flows in resource industries Section 4: Non-Newtonian fluids and rheometry (PART 1)

EFFECT OF PIPE ROUTING ON TURBINE FLOW METER READINGS AND SUBSEQUENT DISCREPANCIES IN CONDENSATE LOADING AT BANGORA GAS FIELD

Viscosity. appearance.

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

A Comparative Second Law Analysis of Microchannel Evaporator with R-134A & R-22 Refrigerants

WTS Table of contents. Layout

Experimental Investigation of Single-Phase Friction Factor and Heat Transfer inside the Horizontal Internally Micro-Fin Tubes.

Pressure Losses for Fluid Flow Through Abrupt Area. Contraction in Compact Heat Exchangers

Lecture 13 Flow Measurement in Pipes. I. Introduction

TOTAL HEAD, N.P.S.H. AND OTHER CALCULATION EXAMPLES Jacques Chaurette p. eng., June 2003

APPLICATION OF DENSITOMETERS TO LIQUID MEASUREMENT Class #

Heat transfer coefficient of near boiling single phase flow with propane in horizontal circular micro channel

FE Fluids Review March 23, 2012 Steve Burian (Civil & Environmental Engineering)

GRAVITY EFFECT ON THE DISTRIBUTION OF REFRIGERANT FLOW IN A MULTI-CIRCUITED CONDENSER

INTERNATIONAL STANDARD

Hydraulic Design Of Polyethylene Pipes

Modelling of dispersed, multicomponent, multiphase flows in resource industries. Section 3: Examples of analyses conducted for Newtonian fluids

Experiment- To determine the coefficient of impact for vanes. Experiment To determine the coefficient of discharge of an orifice meter.

1/54 Circulation pump, safety valve, expansion vessel

What we know about Fluid Mechanics. What we know about Fluid Mechanics

The effect of geometric parameters on the head loss factor in headers

Heat Transfer Coefficient Solver for a Triple Concentric-tube Heat Exchanger in Transition Regime

BERNOULLI EQUATION. The motion of a fluid is usually extremely complex.

Microfluidics 1 Basics, Laminar flow, shear and flow profiles

arxiv: v1 [physics.app-ph] 25 Mar 2018

R09. d water surface. Prove that the depth of pressure is equal to p +.

Polymers Reactions and Polymers Production (3 rd cycle)

LECTURE 1 THE CONTENTS OF THIS LECTURE ARE AS FOLLOWS:

Transcription:

Biblid: 181-4487 (010) 14; 4; p.178-18 UDK: 53.13:303. Original Scientific Paper Originalni naučni rad LIQUID VISCOSITY DETERMINATION BY CORIOLIS FLOW METER ODREĐIVANJE VISKOZNOSTI TEČNOSTI KORIOLISOVIM MERILOM PROTOKA Siniša BIKIĆ*, Maša BUKUROV*, Mirko BABIĆ**, Slobodan TAŠIN*, Ivan PAVKOV**, Milivoj RADOJČIN** *Faculty of Technical Sciences, 1000 Novi Sad, Trg Dositeja Obradovića 6, Serbia **Faculty of Agriculture, 1000 Novi Sad, Trg Dositeja Obradovića 8, Serbia email: bika@uns.ac.rs ABSTRACT In this paper continuous measuring of water viscosity by Coriolis flow meter is analyzed. The idea of technical solution for determining fluid viscosity by Coriolis flow meter and Hagen-Poiseuille equation was elaborated. Pressure drop was measured at the Coriolis tube instead of at the straight pipeline section. The aim of the research is to consider the possibility of applying the proposed technical solution for reliable determination of water viscosity. It was proved that by taking into account the temperature of water t and hydraulic parameters of measurement system Δp/Q, the proposed technical solution can be a reliable method for continuous monitoring of water viscosity, even if Hagen-Poiseuille flow was not achieved. The proposed solution decreases the number of necessary sensors for viscosity measuring, and makes the system and facility more compact. Compared to the conventional parallel mounted capillary viscometer with constant flow pump, the proposed technical solution is much cheaper. The presented technical solution of the measurement system requires a long process of calibration in real terms. Key words: Coriolis flow meter, Hagen-Poiseuille equation, viscosity measuring. REZIME Predmet istraživanja u radu je određivanje viskoznosti vode pomoću Koriolisovog merila protoka. U radu je razrađena ideja tehničkog rešenja da se za određivanje viskoznosti fluida koristi Koriolisovo merilo protoka i Hagen Poazejeva jednačina, gde bi se umesto na pravoj deonici pad pritiska merio na samoj Koriolisovoj cevi. Cilj istraživanja je da se sagleda mogućnost primene predloženog rešenja za pouzdano merenje dinamičke viskoznosti vode. Dokazano je da uzimanjem u obzir temperature vode t i hidrauličkog parametra Δp/Q, predloženi koncept mernog sistema može pouzdano da se koristi za kontinualno merenje dinamičke viskoznosti vode, pa čak i pri narušavanju uslova Hagen Poazejevog strujanja. Predloženo rešenje mernog sistema smanjuje broj senzora za merenje viskoznosti, pa se dobija na kompaktnosti uređaja u koji se ugrađuje merni sistem i na pojednostavljenju samog postrojenja. U poređenju sa klasičnim kapilarnim viskozimetrom paralelno postavljenim na cevovod i pumpom za ostvarivanje konstantnog protoka, predloženo tehničko rešenje je jeftinije. Međutim, predloženo rešenje mernog sistema zahteva dugotrajan proces kalibrisanja u realnim uslovima. Ključne reči: Koriolisovo merilo protoka, Hagen Poazejeva jednačina, merenje viskoznosti. INTRODUCTION At the Faculty of Agriculture in Novi Sad, at the Laboratory for Engineering of Biosystems, a device for evaporation of sucrose solution was developed (Babic, et al., 008). One of the goals of evaporator development was continual monitoring of evaporation process. In the process of osmotic dehydration and evaporation of the sucrose solution, density, viscosity and volumetric flow rate depend on temperature and concentration of sucrose. For the purpose of continuous flow measurement of solution in the frame of development of the evaporator of the sucrose solution, the possibility of using angle valves and Coriolis flow meter was examined. It was found that angle valve can be successfully used to measure the flow of sucrose solution. But the calibration process is time consuming and expensive, because the flow solution must be compensated for the value of changes in concentration and temperature of the solution (Bikic, S, et al, 010). Coriolis mass flow meter measures fluid density and temperature of fluids, and consequently the volume of fluid flow, with only one sensor. It was found that it can reliably measure the flow of sucrose solution without the need for compensation of flow due to changes in concentration and temperature of the sucrose solution. It was also observed that there is a need for deeper understanding of the error made by Coriolis flow meter due to two-phase flow of the sucrose solution (Bukurov, et al, 009). Continual measuring of viscosity is one of the most difficult goals to achieve in the frame of monitoring of evaporation process parameters. Fluid viscosity is an important parameter, which is, in most cases, determined by sampling and analysis. Laboratory studies are expensive and slow, and in some cases and under certain conditions unreliable. The general objective is to perform continuous measurement of the production process parameters, with a trend to measure several parameters with one sensor. In line with this trend, an idea of the Kalotay (Kalotay, 1999) was worked out. He proposed determination of fluid viscosity by Hagen-Poiseuille equation, where pressure drop was measured at the Coriolis tube instead of at the straight pipeline section. Kalotay (Kalotay, 1999) only gave an idea of technical solution without proving it. If reliability of the proposed technical solution is proved, more compact fluid viscosity measurement system could be obtained. The scope of the research is continuous measurement of water viscosity. Experimental research was carried out with water at two different temperatures t and different values of hydraulic parameters of measurement system Δp/Q. The aim of the research is to consider the possibility of applying the proposed technical solution and Hagen-Poiseuille equation to determine the viscosity of water. The hypothesis which was tested was: when taking into account the temperature of water t and the hydraulic parameters of the measurement system Δp/Q, the proposed technical solution can be a reliable method for con- 178 Journal on Processing and Energy in Agriculture 14 (010) 4

tinuous monitoring of water viscosity, even if Hagen-Poiseuille flow was not achieved. Nomenclature A [rad /s kg] spring constant A, B constants d [m] inner diameter of the tube, width I u [Ns ] inertia of the tube K [m 3 ] conversion factor K u [Nm/rad] temperature dependent stiffness of the tube L [m] the length of the straight tube between the pressure connections m tube [kg] mass of Coriolis tube m [kg/s] mass flow rate Re [-] Reynolds s number t [ o C] temperature T [K] temperature t [s] period of frequency v [m/s] velocity V [m 3 ] volume of fluid in Coriolis tube Q [m 3 /s] volumetric flow rate Greek Symbols ρ [kg/m 3 ] density υ [m /s] viscosity λ [-] friction factor η [Paxs] absolute viscosity ε [%] relative difference ω [rad/s] the vibration frequency τ[s] time lag Δp [Pa] pressure drop Subscripts l literary m model tube Coriolis tube MATERIAL AND METHOD Darcy-Weisbach equation for the straight pipeline section of the pipeline is L v Δ p = ρλ (1) d If the previous equations is restricted to the laminar flow regime vd Re = < 300 () ν 64 λ = (3) Re and including the equation of continuity 4Q v = (4) d π Hagen-Poiseuille equation is obtained 4 d π Δp η = (5) 18 LQ Previous equation has the following limitations (White, 1998): - fluid flow regime must be laminar; - measuring section must be straight; - diameter of the pipe must be constant and - liquid has to be Newtonian. Kalotay (Kalotay, 1999) proposed determination of fluid viscosity by Hagen-Poiseuille equation, where pressure drop was measured at the Coriolis tube instead of at the straight pipeline section. The Coriolis tube is curved in plane and in many cases it is split in two tubes, Figure 1. In this way two of four above mentioned limitations are not respected and the Hagen-Poiseuille flow was not achieved. Fig. 1. Illustration of curved Coriolis tube As long as Hagen-Poiseuille flow is not achieved through the Coriolis flow meter, conversion term K must be noticed in previous equation and determined in real terms 4 d π K =. (6) 18 L Previous equation becomes Δp η = K, (7) Q where a unit of conversion factor is m 3. Coriolis flow meter measures flow rate and density of the fluid: K m I w = u u t Ad (8) At m ρ = tube, 4π V V (9) Q (10) = m ρ If the value of conversion factor K is known, it is possible to determine the absolute viscosity of fluid using equation (7), by measuring the pressure drop Δp at the Coriolis tube and volume flow rate Q through the Coriolis tube. In the proposed concept of the measurement system, any two reasonably separated temperature points are enough to determine the fluid viscosity. The Arrhenius equation is used to compute viscosity at any given temperature, as shown below: B T A= η1e (11) T1 T η B = ln (1) T η 1 1 B η = Ae (13) Industrial viscometers using Hagen-Poisseuille equation operate as capillary viscometers, where the capillary is parallel connected to the main flow stream, and constant flow rate through capillary is performed by dosing pumps (Cvijanovic, 1998). In this way capillary viscometer has constant value of hydraulic parameter Δp/Q=const., where absolute viscosity η depends only on temperature t. Application of parallel capillary pipe with the dosing pump is an expensive and complicated solution. The proposed measurement system is serial connected to Journal on Processing and Energy in Agriculture 14 (010) 4 179

the main flow stream, where in many cases flow rate through facilities is not constant. The proposed measurement system could be calibrated and used to determine the dependence of the conversion factor K, temperature t and hydraulic parameter of the system Δp/Q: Δp K = K T, (14) Q The following procedure is suggested. Absolute viscosity is measured at two different temperatures by laboratory capillary viscometer. Laboratory obtained data are modeled by equations (11), (1) and (13). Coriolis flow meter with connections for measuring of pressure drop is serial connected to the pipe of facility. At the formed system, temperature T and hydraulic parameters of system Δp/Q are measured. If equations (7) and (13) are equalized: B Δp Ae K T, = (15) Q Δp ( T ) Q the conversion factor of measurement system could be determined. Simple experiment with water as liquid was carried out at the Laboratory for Fluid Mechanics, Faculty of Technical Sciences in Novi Sad. Figure shows scheme of installation for factor conversion K determination. The water is heated by the boiler, and constant flow rate is set up by the regulator. Flow rate Q, density ρ and water temperature t are measured by Coriolis flow meter, while pressure drop Δp is measured by differential pressure gauge. Coriolis flow meter is checked by volume method. Differential manometer for pressure drop measuring at Coriolis tube - Yokogava manufacturer, model EJA510A, measuring range 0-500 mbar (Yokogava, User s Manual EJA110A Differential Pressure and Pressure Transmitters). Temperature of water was measured using a PTC sensor embedded in the Coriolis flow meter. PTC probe is measuring range - 40 to 150 C and 1:01 C resolution. RESULTS AND DISCUSSION The constants for water A = 1,3991 x 10-6 and B = 1930.5 in equation (13) are determined by equations (11) and (1). For this purpose the well known temperature t and absolute viscosity η function is used (Bukurov, M., 009), Figure 4. At the installation for viscosity measurement system calibration, Figure, function of flow rate Q and pressure drop Δp for two water temperatures t = 15 o C and t = 60 o C are measured. The results of measurement are shown in Table 1 and. Dependence of the conversion factor K and hydraulic parameter of measurement system Δp/Q is determined for two measured water temperatures by equation (15). Fig.. Schema of installation for viscosity measurement system calibration Fig. 3. Elements of viscosity measurement system Figure 3 shows elements of the proposed viscosity measurement system. The following equipment was used: Coriolis flow meter manufacturer Micro Motion Coriolis, model R05, measuring range 1:60, accuracy (0.1-0.) % and repeatability (0.05-0.1) %. The accuracy of density measurement is (0.5 - ) % of the current density measured values. Inner diameter of the Coriolis tube is 14 mm (Instruction Manual, P / N 3101009-eu). Fig. 4. Absolute viscosity as a function of water temperature (Bukurov, 009) Table 1. The results of measuring for flow rate, density and pressure drop at the temperature of water t=15 o C t = 15 o C ρ = 999.05 kg/m 3 No. Q [L/s] Δp [mbar] 1 0.0087.1 0.017 6 3 0.046 10.6 4 0.031 16.3 5 0.0353 19.1 6 0.046 6.3 Table. The results of measuring for flow rate, density and pressure drop at the temperature of water t=60 o C t = 60 o C ρ = 988 kg/m 3 No. Q [L/s] Δp [mbar] 1 0.0137 1 0.09.6 3 0.0305 6.8 4 0.0417 16.5 Figure 5 graphically presents the dependence (14) of the conversion factor K and hydraulic parameter of measurement system Δp/Q for two measured water temperatures t = 15 o C and t = 60 o C. Circles and squares represent the values obtained by 180 Journal on Processing and Energy in Agriculture 14 (010) 4

measurement, while the solid lines represent the values obtained from the measured values modeled by the "spline" function. "Spline" function proved to be the best model for the dependence shown in Figure 5. Fig. 5. Conversion factor as a function of hydraulic flow conditions at different temperatures of water To confirm the mathematical model, the comparison of the data obtained by the mathematical model and literary data (Bajic at all, 1980) was done: ηl ηm ε = 100%. (16) η l Confirmation of the mathematical model is shown in Figure 6, where one can see the relative difference between the literary data and the data obtained by the mathematical model. It can be observed that the function "spline" quite well represents experimental data in the whole operation regime. The maximal relative differences between the data are order of magnitude 10-4 and 10-5, which is negligible. It was interesting to check whether the achieved flow regime was laminar (equation ). The velocity of flow was determined using equation (4) and Reynolds's number using equation (). Figure 7 shows that through almost the whole operation mode of measurement system flow regime is turbulent. Despite the fact that flow regime was turbulent, the model was verified because conversion factor K was determined in real terms. In this way the presented model is valid only under these conditions. Fig. 6. The relative difference between the literal data and the data obtained by mathematical model at differential temperature of water Flow regime could be controlled by proper choice of Coriolis flow meter. These types of meters are produced with different flow measuring ranges. For example, company Küppers Elektromechanik offers meter KCM 300 with flow measuring range from 4.5 to 300 kg/h and inner diameter of 4 mm (www.eeberhardt.com), while company Kobold offers meter TM UMC3 with flow measuring range from 0.8 to 8 kg/h and inner diameter of 6 mm (www.koboldusa.com). In situation like in Figure 7, water temperature t=15 o C, flow regime could be also controlled by setting the hydraulic parameters of system Δp/Q. Fig. 7. Flow regime of water through the measurement system Journal on Processing and Energy in Agriculture 14 (010) 4 181

CONCLUSION In the paper the idea of determining fluid viscosity by Hagen-Poiseuille equation was elaborated, while pressure drop was measured at the Coriolis tube instead of at the straight pipeline section. This technical solution makes measurement system and facility at the same time more compact and simpler. Compared to conventional parallel mounted capillary viscometer with constant flow pump, the proposed technical solution is much cheaper. However, the presented technical solution of the measurement system requires a long process of calibration in real terms. The proposed measurement system is serial connected to the main flow stream, where in many cases flow rate through facilities is not constant. The proposed measurement system must be calibrated in real terms. In this way, the dependence of conversion factor K, temperature t and hydraulic parameter of the system Δp/Q is valid only under the calibrated conditions. It was proved that by taking into account the temperature of water t and the hydraulic parameters of measurement system Δp/Q, the proposed technical solution can be a reliable method for continuous monitoring of water viscosity, even if Hagen- Poiseuille flow was not achieved. ACKNOWLEDGEMENT: The project No. 0065 The Quality of Dried Fruit Production is a grant by the Serbian Ministry of Science and Technology. REFERENCES Babić, M, Babić, Ljiljana, Pavkov, I, Radojčin, M. (008). Rationalization of sucrose solution using during the fruit osmotic dehydration. Third international Congress of Technologists for Post-harvest Technology. Stubičke Toplice, Croatia, Proceeding, 104-105. Bajić, D, i dr.: Priručnik za industriju šećera II, Savez hemičara i tehnologa Jugoslavije, Beograd, 1980. Bikić, S., Bukurov, Maša, Babić, M., Tašin, S., Pavkov, I., Radojčin, M. (009). Application of angle valve for continual measuring of sugar water solution flow rate. Journal on processing and energy in agriculture,14(), 11 115. Bukurov, Maša. (009). Uređaji za mehaničko prečišćavanje vazduha. FTN Izdavaštvo, Novi Sad. Bukurov, M., Bikić, S., Babić, M., Tašin, S., Pavkov, I., Radojčin, M. (009). Coriolis flow meter application for continual monitoring of sucrose solution evaporation process. Journal on processing and energy in agriculture, 13(4), 98 30. Cvijanović, P. (1998). Merenje fluidnih veličina, Stylos, Novi Sad. Kalotay, P. (1999). Density and viscosity monitoring systems using Coriolis flow meters. ISA Transactions, 38(4), 303-310. Küppers Elektromechanik GmbH (010). C Flow KCM and KCE 5000 Manual. www.eeberhardt.com. Kobold (010). Coriolis mass flow meter operation manual model TM UMC3. www.koboldusa.com Micro Motion R-Series Sensor. Instruction Manual, P/N 3101009-eu, Rev. C 06/01, and Data Sheets PS0000- PS00016. www.documentation.emersonprocess.com Yokogava. Yokogava, User s Manual Yokogava, User s Manual EJA110A Differential Pressure and Pressure Transmitters. www.yokogawa.com Received:0.1.010. Accepted:5.1.010. 18 Journal on Processing and Energy in Agriculture 14 (010) 4