The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Similar documents
The determination of residual solvents in pharmaceuticals using the Agilent G1888 headspace/6890n GC/5975 inert MSD system Application Note

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Study of Residual Solvents in Various Matrices by Static Headspace

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

USP<467> residual solvents

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Residual solvents analysis using an Agilent Intuvo 9000 GC system

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

A Generic Method for the Analysis of Residual Solvents in Pharmaceuticals Using Static Headspace-GC-FID/MS

Introducing New Functionalities in Liquid Stationary Phases in GC Columns for Confirming Organic Volatile Impurity Testing in Pharmaceutical Products.

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Headspace Technology for GC and GC/MS: Features, benefits & applications

The Analysis of Residual Solvents in Pharmaceutical Products Using GC-VUV and Static Headspace

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Author. Abstract. Introduction

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Solid Phase Micro Extraction of Flavor Compounds in Beer

Methanol Extraction of high level soil samples by USEPA Method 8260C

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Maximizing Sample Throughput In Purge And Trap Analysis

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Application. Gas Chromatography February Introduction

Application Note. Abstract. Authors. Introduction

Optimizing. Abstract: is standardd. procedures. altered to

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Fast, Quantitative Analysis of Residual Solvents in Cannabis Concentrates

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

A Capillary Gas Chromatographic Procedure for the Analysis of Nine Common Residual Solvents in Water-Insoluble Bulk Pharmaceuticals

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Thermal Desorption Technical Support

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

GC METHOD FOR QUANTITATIVE DETERMINATION OF RESIDUAL 2-(2-CHLOROETHOXY)ETHANOL (CEE) AND N-METHYL-2-PYRROLIDINONE (NMP) IN QUETIAPINE

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

C146-E209. Headspace Samplers. HS-20 Series

Optimizing Standard Preparation

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Analytical Trap Comparison for USEPA Method 8260C

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

STANDARD OPERATING PROCEDURES

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Volatile Organic Compounds in Every Day Food

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

A Fast, Simple FET Headspace GC-FID Technique for Determining Residual Solvents in Cannabis Concentrates

Determination of Volatile Organic Compounds in Air

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Application. Gas Chromatography March 1998

Analysis of BTEX in Natural Water with SPME

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

The Analysis of Organophosphate Pesticides by LC/MS Application

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Thermal Desorption Technical Support

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Transcription:

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application Pharmaceuticals Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville Road Wilmington, DE 19808-1610 USA Abstract The G1888 Network Headspace Sampler interfaced to 6890N gas chromatographs configured with either an FID or 597 inert MSD were used for the determination of regulated residual solvents. Standard mixtures in water were used at various concentrations including levels below published acceptance guidelines to demonstrate system performance. Included were Class 1 and Class 2 solvents, according to the International Conference on Harmonization, and those listed in USP 467. Repeatability, inertness, and carryover reduction are improved compared to previous generation samplers, using the automated 70-sample G1888 with Siltek flow path. Integrated control and sequencing of the sampler is incorporated into the Agilent GC ChemStation through an add-on software module. Introduction Organic volatile impurities (OVIs) can result from the manufacture of active pharmaceuticals or other drug products. Many are used to enhance yields, improve crystallization, or increase solubility [1]. Other factors such as packaging, transportation, and storage can also impact the level of residual solvents. Gas chromatography (GC) coupled with static headspace sampling, acknowledged as an easy-to-use high-throughput analytical tool for the determination of low-level solvent impurities in drugs, can be found in nearly all Quality Control (QC) laboratories in pharmaceutical manufacturing facilities. Sample prep is relatively simple and the methodology is easily validated as per specific monographs. General guidelines established by the International Conference on Harmonization (ICH) divide solvents into three classes [2]. The Class 1 solvents should not be used in pharmaceutical manufacture because of toxicity or environmental impact, while use of Class 2 solvents should be limited due to potential toxicity. Class solvents are regarded as posing a lower risk to human health. Solvents listed in USP 467 include a subset of specific Class 1 and Class 2 solvents. This application note will demonstrate the analysis and quantitation of Class 1 and Class 2 solvents. See Table 1 for a listing of residual solvents.

Table 1A. Class 1 Solvents in Pharmaceutical Products - To Be Avoided [2] Concentration Solvent limit (ppm) Concern Benzene 2 Carcinogen Carbon tetrachloride 4 Toxic and environmental hazard 1,2-Dichloroethane 5 Toxic 1,1-Dichloroethene 8 Toxic 1,1,1-Trichloroethane 1500 Environmental hazard Table 1B. Class 2 Solvents in Pharmaceutical Products [2] Permissible daily exposure (ppm) Concentration Solvent (mg/day) limit (ppm) Acetonitrile 4.1 410 Chlorobenzene.6 60 Chloroform 0.6 60 Cyclohexane 8.8 880 1,2-Dichloroethene 18.7 1870 Dichloromethane 6.0 600 1,2-Dimethoxyethane 1.0 100 N,N-Dimethylacetamide 10.9 1090 N,N-Dimethylformamide 8.8 880 1,4-Dioxane.8 80 2-Ethoxyethanol 1.6 160 Ethyleneglycol 6.2 620 Formamide 2.2 220 Hexane 2.9 290 Methanol 0.0 000 2-Methoxyethanol 0.5 50 Methylbutyl ketone 0.5 50 Methylcyclohexane 11.8 1180 N-Methylpyrrolidone 48.4 4840 Nitromethane 0.5 50 Pyridine 2.0 200 Sulfolane 1.6 160 Tetralin 1.0 100 Toluene 8.9 890 1,1,2-Trichloroethene 0.8 80 Xylene* 21.7 2170 * Usually 60% m-xylene, 14% p-xylene, 9% o-xylene with 17% ethyl benzene. Table 1C. Solvents in Pharmaceutical Products, According to USP 467 Concentration Solvent limit (ppm) Methylene chloride 600 Chloroform 60 Benzene 2 Trichloroethylene 80 1,4-dioxane 80 2

Residual solvent and other contaminant levels, designated as safe, have trended downward in recent years as information about potential harmful effects of long-term and/or low-level exposures accumulate and as the detection sensitivity of analytical instrumentation improves. For example, based on new toxicity data, a 200 change in the regulations for residual solvents stipulate a 10-fold reduction of the 1997 PDE (permitted daily exposure) and residual concentration limits for the solvent N-methylpyrrolidone []. It also reclassifies tetrahydrofuran from a Class to a Class 2 category solvent with PDE and concentration limitations more restrictive than toluene []. Table 1B also lists PDE and concentration limits for Class 1 and Class 2 residual solvents in pharmaceutical products [4]. Experimental Two systems are described in this work. The first, based on flame ionization detection is considered the system of choice for routine QC work, while the second system with mass selective detection provides unknown determination, possible quantitation of near-coeluting peaks, and solvent confirmation. Ten-milliliter headspace vials were used for all experiments containing 5 ml water as the matrix, with 1-g sodium sulfate added to assist with analyte extraction. The headspace sampler was equipped with a 1-mL sample loop. Since a sufficient flow must be maintained through the system to avoid excessive peak broadening, a split injection is used. A 2:1 split ratio is the lowest recommended with typical 0.5-mm id column flows. Table 2. Instrument Conditions FID system 6890N GC Injection port Volatiles interface Temperature 160 C Split ratio 2:1 to 5:1 Carrier gas Helium Carrier flow 9 ml/min Detector FID, 250 C GC oven program Initial temperature 5 C Initial time 20 min Rate 25 C/min Final temperature 250 C Final time 15 min Columns 0 m 0.5 mm µm DB-624 0 m 0.45 mm 2.55 µm DB-624 G1888A Headspace Sampler Loop size 1 ml Vial pressure 14.0 psig Headspace oven 85 C Loop temp 100 C Transfer line temp 120 C Equilibration time 0 min, low shake GC Cycle time 50 min Pressurization 0.15 min Vent (loop fill) 0.15 min Inject 0.5 min 597 inert system 6890N GC Injection port Volatiles interface Temperature 160 C Split ratio 2:1 to 5:1 Carrier gas Helium Inlet pressure 2.7 psi Column flow 1.7 ml/min GC oven program Initial temperature 5 C Initial time 20 min Rate 20 C/min Final temperature 250 C Final time 15 min Column 0 m 0.2 mm 1.8 µm DB-624 G1888A Headspace Sampler Same settings as FID system 597 inert Scan 0 to 200 amu, samples 2 SIM 100 ms dwell Source temperature 20 C Quad temperature 150 C Tune BFB.u Standards USP 467 Restek #6228 AccuStandard NF-467-R ICH Class 1 and 2 Restek #6228 (Class 1) #6229 (Class 2A) #620 (Class 2B)

Discussion GC System Most quality control labs in pharmaceutical manufacturing employ gas chromatography (GC) for the determination of residual solvents that are included in either USP 467 or in the more extensive list covered in ICH guidelines. Capillary GC based on the 624 phase (USP G4) is widely used for solvent separation. A different stationary phase such as DB-1701, DB-5, or DB-WAX (USP G16) can be used in specific methods when coelution is identified. Headspace sampling has many advantages over direct liquid injection including the avoidance of large water injections that can result in column degradation. Table lists concentrations and identifications of Class 1 and Class 2 solvents used to produce the chromatogram shown in Figure 1. 1 4 5 6 7,8 10, 11, 12 17 20 22, 2, 24, 25, 26,27, 28 29 21 16 2 9 1,14 15 18 19 5 10 15 20 25 0 Figure 1. Class 1 and Class 2 residual solvents. G1888, 6890N with FID and volatiles interface. 4

These concentrations equal the guideline limits based on a 100-mg sample of the pharmaceutical dissolved in 5 ml. USP 467 solvents are shown in Figure 2 at concentrations below the required levels (10 µl Restek std. #6228). Excellent signalto-noise ratio is still achieved. Concentrations used throughout this work are defined as the analyte concentration present in the headspace vial prior to sampling. 1 4 2 5 5 10 15 Figure 2. Identifications and concentrations: 1. Methylene chloride 1.2 µg/ml, 2. Chloroform 0.12 µg/ml,. Benzene 0.004 µg/ml, 4. Trichloroethylene 0.16 µg/ml, 5. 1,4-dioxane 0.76 µg/ml. Coelutions that occur on the DB-624 column under the chromatographic conditions and concentrations employed are listed in Table 4. Using the 0 m 0.45 mm 2.55 µm DB-624 column, benzene and 1,2-dichloro-ethane can be resolved at a 5 C oven temperature, as seen in Figure. 1 2 9 9.25 9.5 9.75 10 10.25 10.5 10.75 11 Figure. Resolution of benzene and 1,2 dichloroethane. Peak identifications: 1. benzene, 2. 1,2-dichloroethane,. 1,2-dimethoxyethane. 5

Calibration curves for selected solvents included in USP 467 are shown in Figure 4. Linear results are seen over a concentration range that extends well below recommended maximum concentrations. The concentration range is well within the linear dynamic range of the thick film 0.5 mm and 0.45-mm id columns. 5000 Methylene chloride 400 Chloroform 4000 00 Area 000 2000 1000 y = 252.82x + 19.987 R 2 = 0.9988 Area 200 100 y = 192.41x + 0.1851 R 2 = 0.9991 0 0 5 10 15 20 µg/ml 0 0 0.5 1.0 1.5 2.0 µg/ml Area 140 Benzene 120 100 80 60 40 y = 2106.2x - 0.0015 R = 0.9995 20 0 0 0.01 0.02 0.0 0.04 0.05 0.06 0.07 µg/ml Area 1400 Trichloroethylene 1200 1000 800 600 400 y = 55.9x +.7229 R 2 = 0.9991 200 0 0 0.5 1.0 1.5 2.0 2.5.0 µg/ml 200 1,4 dioxane 160 Area 120 80 40 y = 15.268x + 0.29 R 2 = 0.9996 0 0 2 4 6 8 10 12 µg/ml Figure 4. Calibration plots for selected solvents. 6

Headspace equilibration time is normally set at 60 min; however, in most cases 0 min is sufficient. Figure 5 illustrates an overlay of a 0- and 60-min headspace equilibration for a selected portion of the chromatogram (Class 1 and 2 solvents). Little overall difference is seen in the peak areas; although for some compounds 0-min equilibration produces marginally larger areas. 0 min 60 min 2 1 4 5 2 24 25 Figure 5. Overlay of selected compounds after 0- and 60-min headspace equilibration times. 1. Chlorobenzene, 2. Ethylbenzene and DMF,. m-xylene, p-xylene, 4. o-xylene, 5. N,N-dimethylacetamide. Table. Class 1 and Class 2 Residual Solvent Concentrations*. Table ID Column Corresponds to Chromatogram Numbering Solvent ID Conc. (µg/ml) Solvent ID Conc. (µg/ml) Methanol 1 60 Trichloroethylene 16 1.6 1,1-Dichloroethane 2 0.16 Methyl cyclohexane 17 26 Acetonitrile 8.2 1,4-Dioxane 18 7.6 Methylene chloride 4 12 Pyridine 19 4 Hexane 6 5.1 Toluene 20 17.8 cis-1,2-dichloroethane 7 7.4 2-Hexanone 21 1 Nitrobenzene 8 1 Chlorobenzene 22 7.6 Chloroform 9 1.2 Ethylbenzene 2 7.8 Carbon tetrachloride 10 0.08 N,N-dimethylformamide 24 17.6 Cyclohexane 11 77.6 m-xylene 25 26.04 1,1,1-Trichloroethane 12 0 p-xylene 26 6.08 Benzene 1 0.04 o-xylene 27.9 1,2-Dichloroethane 14 0.1 N,N-dimethylacetamide 28 21.8 1,2-Dimethoxyethane 15 2 Tetraline 29 2 *Concentrations shown are headspace vial solution concentrations prior to sampling. Peak 5 (trans 1,2 dichloroethane) is not listed in the table as a Class 1 or Class 2 solvent. 7

Table 4. Coelutions on 0.5-mm id DB-624 Coelution group Solvents 1 (partial) Benzene, 1,2 dichloroethane 2 Nitrobenzene, cis-1,2 dichloroethene Carbon tetrachloride, Cyclohexane, 1,1,1- trichloroethane* + 4 Ethylbenzene, DMF 5 m-xylene, p-xylene * Trichloroethane will separate from CCl 4 in absence of cyclohexane + Figure 6 shows separation on a 0 m x 0.45 mm x 2.55 µm DB-624 Agilent part no.124-14. USP 467 and Class 1 Residual Solvents 6, 7 1 2 1. 1,1 Dichloroethylene 2. Dichloromethane. Chloroform 4. 1,1,1 Trichloroethane 5. Carbon tetrachloride 6. Benzene 7. 1,2 Dichloroethane 8. Trichloroethylene 9. 1,4 Dioxane 4 8 5 9 Figure 6. 2 4 6 8 10 min USP 467 and Class 1 solvents at 1 ppm each on the 0 m x 0.45 mm x 2.55 µm DB-624 column. Starting GC oven temperature was 40 C. 8

MSD System A TIC of Class 1 and Class 2 solvents produced with a G1888/6890N/597 inert system is shown in Figure 7. 4 5 6 7,8 9 10,11,12 16 17 22,2,24,25,26,27,28 20 21 29 1,14 2 15 19 1 18 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Time Figure 7. TIC of Class 1 and Class 2 solvents. See Table for compound identifications. Analyte solution concentrations and peak identifications are as indicated in Table. Gas chromatography/mass spectrometry (GC/MS) offers an alternative methodology to flame ionization detection (FID) that can be useful to solve coelution problems or identify unknowns. Also, excellent sensitivity and selectivity can be achieved in Selected Ion Monitoring (SIM) mode, which may be useful for drug manufacturing process development to identify and quantify trace impurities. Carryover In practice, nonaqueous solvents are commonly used in testing since extraction of many common solvents used in pharmaceutical manufacturing are not water soluble. Some common solvents include DMA (N.N-dimethylacetamide), DMSO, pyridine, and DMI (1,-Dimethyl-2-imidazolidinone). Because many of the solvents used are high boiling, the possibility of headspace carryover exists. Improvements in the thermal zone temperature uniformity in the G1888 reduce the condensation of high-boiling materials in various flow path lines and vent valve. 9

The G1888 incorporates a new feature that allows users to program the vent purge time, labeled Sequence Vent Purge in the advanced functions menu. This represents the time interval when the vent line is purged as a post injection event. The default time of 0 seconds can be increased up to an approximate maximum of the cycle time. For the carryover experiments in this work, a vent purge time of 20 min was used to further reduce the possibility of solvent carryover. One hundred microliters of pure solvent was introduced into 10-mL vials. Larger amounts of solvent will not result in an increase in the amount injected. A 10-vial sequence was set up with alternating solvent and water blank vials using the chromatographic program shown in the experimental section. This test was performed for pyridine, DMSO, and DMA. For all three solvents, carryover ( [amount solvent area from blank vial/solvent area from solvent vial] 100) was under 0.006%. In addition, the solvent areas for all blanks had ±% RSDs, indicating an absence of trending. As an additional carryover check, 10 consecutive vials of DMA (100 µl per 10-mL vial) were run at a Headspace oven temperature of 100 C. These were followed by two water blanks. The first blank showed carryover of 0.004% and the second 0.001%. One of the most effective solvent systems used today by pharmaceutical companies is DMI with a boiling point of 225 C. Table 5 lists the system set points used in a carryover test with this solvent. Alternating vials of DMI and water blank were run in a headspace sequence. Results are shown in Figure 8. The large concentration of DMI overloads the column and leads to some inconsistency in areas, however, it is not a concern given the purpose of this test. Measured carryover is under 0.00%. Table 5. Setting Used for DMI Carryover Test Headspace oven 220 C Loop 250 C Transfer line 250 C Vial eq. time 0 min Sequence vent purge 20 min Sample 100-µL DMI in 10-mL vial Blanks 5-µL water in 10-mL vial Volatiles interface 250 C Split ratio 10:1 Oven program 5 C (0 min) to 260 C (15 min) at 25 C/min To check for carryover of the actual analytes, a test scheme similar to that used for the pure solvents was chosen. One hundred microliters of the USP 467 standard (Restek # 6007) was placed in 5-mL water/1 g Na 2 SO 4. These vials were alternated with pure water blanks in a 10-vial sequence at 85 C equilibration temperature. No measurable area for the analytes could be integrated in any of the runs. DMI Carryover Test 14.0 60 12.0 50 10.0 40 Area 100,000 8.0 6.0 4.0 0 20 Area DMI (Left scale) Water blank (Right scale) 2.0 10 0.0 2 4 6 8 10 12 14 16 18 Vial number 0 Figure 8. DMI carryover results. 10

Conclusions Manufacturers of pharmaceuticals must ensure that residual solvents, OVIs, and related contaminants are not present in their products, or are present below levels stipulated as safe by regulation. One of the impediments to accurate determination of impurities at very low levels is the tendency for analyte interaction and/or reaction with the internal surfaces of the instrument sample path. To eliminate this problem, a new inert headspace sampler, the G1888 system was developed. This instrument possesses a nonreactive, nonadsorptive sample flow path from the point of injection through detection. This significantly reduces carryover, a common problem with older instrumentation. High temperature heated zones extend the choice of solvent systems that can be used. When coupled to the 597 inert MSD, which uses a solid inert source, superior results are obtained when the need for unknown identification or confirmation is required. Analytical results obtained for broad classes of solvents, used in medicinal products by the G1888 Headspace Sampler systems, described in this application show reduced carryover, excellent detection sensitivity, and good linear response over the ppm to ppb range. References 1. Anil M. Dwivedi, Residual Solvent Analysis in Pharmaceuticals, (Nov., 2002) Pharmaceutical Technology, 42-46. 2. Guidance for Industry, QC Impurities: Residual Solvents, U.S. Department of Heath and Human Services, FDA, Center for Drug Evaluation and Research (CDER), Center for Biologics Evaluation and Research (CBER) Dec. 1997, ICH.. Revised PDEs for NMP and THF, Federal Register, 68, (219), Nov. 200, Notices, 645. 4. Limits of Residual Solvents, Federal Register, 62, (247), Dec. 1997, Notices 6780-6781. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. The methods and procedures outlined in this work illustrate potential approaches to the analysis of residual solvents. Laboratories should perform system suitability studies and validate their methods according to ICH or USP guidelines. 11

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2004 Printed in the USA June 21, 2004 5989-126EN