Trefoil Columns [ CARE AND USE MANUAL ] CONTENTS I. INTRODUCTION I. INTRODUCTION GETTING STARTED III. COLUMN USE IV.

Similar documents
ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

ACQUITY UPLC HSS Columns

[ CARE AND USE MANUAL ]

[ CARE AND USE MANUAL ] XSELECT HSS XP 2.5 µm COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. GETTING STARTED III. COLUMN USE

Basic Principles for Purification Using Supercritical Fluid Chromatography

for Acclaim Mixed-Mode HILIC-1 Column

[ Care and Use Manual ]

[ CARE AND USE MANUAL ]

FAQ's. 1 - Which column would be the most appropriate for my application?

[ Care and Use Manual ]

XSEL EC T C HA RGED SU RFAC E HYBRID (C SH) COLUMNS

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

ikey Separation Device

C18 Column. Care & Use Sheet

Optical Isomer Separation Columns and Packing Materials

[ Care and Use Manual ]

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Barry E. Boyes, Ph.D. Consumables and Accessories Business Unit May 10, 2000

Routine MS Detection for USP Chromatographic Methods

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

[ a ppl ic at ion no t e ]


TCI Chiral HPLC Column

ChromegaChiral TM CSP Media and Columns

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

ProPac WCX-10 Columns

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

Determination of Polymer Modifier in Asphalt

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

Chiral separations efficient, fast and productive

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

UPLC Intact MASS Analysis Application Kit

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Dilution(*) Chromatography

Acclaim Mixed-Mode WCX-1

penta-hilic UHPLC COLUMNS

Solid Phase Extraction Method Development Tips and Tricks

ChromegaChiral TM CSP Media and Columns

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Chiral Separation Using SFC and HPLC

Method Development in Solid Phase Extraction using Non-Polar ISOLUTE SPE Columns for the Extraction of Aqueous Samples

High Purity Chromasolv Solvents

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

Headspace Technology for GC and GC/MS: Features, benefits & applications

Basic principles of HPLC

Peptide Isolation Using the Prep 150 LC System

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Acclaim Mixed-Mode WAX-1 Columns

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Pelagia Research Library

IonPac Trace Cation Concentrator

LIQUID CHROMATOGRAPHY

Successfully Scaling and Transferring HPLC and UPLC Methods

MEDROXYPROGESTERONE INJECTION

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Acclaim HILIC-10 Columns

Definitive EtG/EtS LC-MS/MS Analysis:

HPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

This method describes the identification of the following prohibited colorants in cosmetic products:

Flash chromatography purification of high pka organic compounds with Teledyne Isco s specialty RediSep columns

Scott Anderson, Kathy Lawrence Dennis McCreary, Melissa Wilcox. Grace Davison Discovery Sciences

EVOLUTE WCX Columns for Solid Phase Extraction of Quaternary Amines and Strongly Basic Compounds from Aqueous Samples

for Acclaim Mixed-Mode WCX-1

[ Care and Use Manual ]

High Performance Liquid Chromatography

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

LC and LC/MS Column Selection Flow Chart

LC Technical Information

CHEM 429 / 529 Chemical Separation Techniques

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Solvent Flexibility for Size-Based Polymer Analysis Using the Advanced Polymer Chromatography (APC) System

Biochemistry. Biochemical Techniques HPLC

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

Experiment 1: Thin Layer Chromatography

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Nexera UC Unified Chromatography

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

penta-hilic UHPLC COLUMNS

SEPARATION OF FLURBIPROFEN AND IBUPROFEN ENANTIOMERS ON A CHIRAL STATIONARY PHASE USING SUPERCRITICAL FLUIDS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

UPC 2 Method Development for Achiral Impurity Analysis

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

Enantiomeric and Diastereomeric Separations of Pyrethroids Using UPC 2

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Dionex IonPac AS28-Fast-4µm column

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

contents of the currently official monograph. Please refer to the current edition of the USP NF for official text.

Investigation of Petasis and Ugi Reactions in Series in an Automated Microreactor System

Transcription:

CNTENTS I. INTRDUCTIN II. GETTING STARTED a. Safety Considerations b. Column Connectors and Installation c. ecord Installation d. Column Equilibration e. Initial Column Efficiency Determination III. CLUMN USE a. Sample Preparation b. Solvents c. Additives d. Pressure e. Temperature f. Band Spreading Minimization I. INTRDUCTIN Thank you for choosing a Trefoil Column. are specifically designed to use the complete range of capabilities of the ACQUITY UPC 2 System to achieve fast, robust chiral separations to enable both selectivity and speed and to reduce method development time. consist of modified polysaccharide coated stationary phases for broad-spectrum chiral selectivity. The higher order structure of the polysaccharides have specific spatial orientations that impart chiral selectivity through well-chosen modifications or selectors. Trefoil AMY1, CEL1, and CEL2 Column chemistries are complementary to each other and independently offer different retention characteristics for separating chiral compounds. Selectivity can be further enhanced by using co-solvents and additive blends that most favorably modulate chiral recognition. These columns are designed to separate enantiomers, stereoisomers, metabolites, degradants, and impurities with greater resolution and speed. IV. TRUBLESHTING V. CLUMN CLEANING, ADDITIVES, AND STRAGE a. Cleaning b. Effects of Additives c. Storage VI. INTRDUCING ECRD INTELLIGENT CHIP TECHNLGY a. Introduction b. Installation c. Column Use Information VII. WATERS SFC CLUMN FAMILY

for Chiral Separations Trefoil AMY1 2.5 µm Columns Amylose tris-(3,5-dimethylphenylcarbamate) min 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 Flavanone Trefoil CEL1 2.5 µm Columns Cellulose tris-(3,5-dimethylphenylcarbamate) H H N min 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 xprenolol Trefoil CEL2 2.5 µm Columns Cellulose tris-(3-chloro-4-methylphenylcarbamate) min 0.0 0.2 0.4 0.6 0.8 1.0 1.2 1.4 1.6 1.8 2.0 N N Tetramisole S Chiral separations were all run using the 2-minute screening method. Trefoil Column: 2.1 x 50 mm Co-solvent: 1:1 MeH:IPA w/20 mm NH4H Gradient: 3 60% in 1.5 min, 60% for 0.5 min Flow rate: 1.2 ml/min Temperature: 40 C ABPR: 3,200 psi UV Detection: 220 nm The Trefoil packing materials are designed for use with the ACQUITY UPC 2 Systems and are manufactured in an IS-certified plant using ultra-pure reagents. Each batch of Trefoil material is tested and the results are held to narrow specification ranges to assure excellent, reproducible performance. Every column is individually tested and a Performance Chromatogram and Certificate of Batch Analysis is provided on the ecord Intelligent Chip. are designed and tested specifically for use on ACQUITY UPC 2 Systems. will exhibit maximum chromatographic performance and benefit NLY when used on holistically-designed ACQUITY UPC 2 Systems since these systems and columns were created and designed to operate together. For these reasons, Waters cannot support the use of on any system other than an ACQUITY UPC 2 System. II. GETTING STARTED Each Trefoil Column comes with a Certificate of Analysis and a Performance Test Chromatogram. The Certificate of Analysis is specific to each batch of packing material and includes the gel batch number, physical characterization, analysis of unbonded particle, analysis of bonded particles, and an SFC chromatographic batch test. The Performance Test Chromatogram is specific to each individual column and contains the following information: gel batch number, column serial number, USP plate count, USP tailing factor, capacity factor, and chromatographic conditions under normal-phase LC conditions. These data should be stored for future reference. a. Safety Considerations An SFC column, while in use, is under pressure with C 2 and possible modifiers as a supercritical fluid. A major safety concern is frostbite caused by adiabatic cooling when C 2 decompresses from a fluid to a gas at atmospheric pressure. Pay attention to any frosting on the column or system connections. This indicates a leak, usually with temperatures far below 0 C. 2

Any small leak could produce a situation where the LEL (lower exposure limit) is reached. Laboratories should be equipped with C 2 and/or 2 sensors when carbon dioxide (C 2 ) is in use. b. Column Connectors and Installation ACQUITY UPC 2 Systems utilize tubing and connectors which have been designed to meet stringent tolerance levels and to minimize extra-column volume. For information on system tubing and connectors, please refer to the ACQUITY UPC 2 System perator s Guide (Literature Number 720004226EN). Note: Scale the flow rate up or down accordingly based upon the column I.D., length, particle size, and backpressure of the Trefoil Column being installed. 1. Make sure your co-solvent pump is primed and has an adequate solvent/modifier supply before performing injections. It is recommended to use C 2 with a purity level of 99.97% (food grade) and use high-quality chromatography grade solvents (see section IIIb). 2. Connect both the inlet and outlet of the column to the ACQUITY UPC 2 System. 3. If the column is still filled with a solvent, use a low flow rate and backpressure setting (100 bar) to start pumping C 2 and modifier through the column. 4. If you see frosting on the column at the inlet or outlet, tighten the finger-tight fitting or compression screw on that side. If you continue to see frosting, turn off the C 2 and vent the system. Allow the column to depressurize fully before disconnecting the inlet or outlet to troubleshoot the leaking issue. c. ecord Installation The ecord button should be attached to the side of the column heater module. The ecord button is magnetized and does not require specific orientation. e. Initial Column Efficiency Determination 1. Before use, perform an efficiency test on the column, using suitable analytes dissolved in a weaker injection solvent to avoid peak distortion from strong injection solvent effects (see section IIIb). It is recommended to use Waters Quality Control Reference Material (p/n 186007950) which contains a mixture of standards specifically chosen for the. For more information, please review the SFC Columns Brochure (Literature Number 720005706EN). ACQUITY UPC 2 Quality Control Reference Materials Intended use Provides convergence chromatographic performance information for both chiral and achiral modes Contents 1. 0.50 mg/ml (+/-) trans-stilbene oxide 2. 0.50 mg/ml Thymine 3. 0.50 mg/ml Sulfamethoxazole 4. 0.50 mg/ml Sulfamethizole In a 1 ml solution of 75:25 ACN:MeH 2. Determine the number of theoretical plates (N) and use this value as a benchmark for periodic comparisons. 3. Repeat the test at predetermined intervals to track column performance over time. (Slight variations may occur if performance tests are performed on two different ACQUITY UPC 2 Systems due to the quality of the connections, operating environment, system electronics, reagent quality and column condition.) Empty column volumes (ml) Column internal diameter (mm) Column length 2.1 mm 3.0 mm 50 mm 0.2 0.4 150 mm 0.5 1.0 Table 1. Empty column volumes in ml (multiply by 10 for flush solvent volumes). d. Column Equilibration are shipped dry. Equilibrate the column with a minimum of 10-column volumes of the mobile phase prior to use. (Refer to Table 2 for a listing of empty column volumes.) Thermal and chemical equilibration are equally important for chromatography. Ensure that the mobile phase is preheated before entering the column to prevent a thermal missmatch. This can be done by enabling active preheating on an ACQUITY UPC 2 System. 3

III. CLUMN USE To ensure the continued high performance of, follow these guidelines: a. Sample Preparation 1. Sample impurities often contribute to column contamination. Waters offers both solid-phase extraction (SPE) and supercritical fluid extraction (SFE) options. For SPE, use asis Solid-Phase Extraction Cartridges/Columns or Sep-Pak Cartridges of the appropriate chemistry to cleanup the sample before analysis. For more information, visit www.waters.com/sampleprep. Alternatively, Waters offers the MV-10 ASFE Supercritical Fluid Extraction System for high throughput extractions from a wide variety of sample matrices. 2. Consider preparing the sample in a weak solvent (such as heptane) for the best peak shape and sensitivity. Using weak sample diluents may avoid peak distortion due to strong solvent effects. In particular, stronger solvents can impact peak shapes of low retaining analytes. 3. If the sample is not dissolved in the mobile-phase modifier, ensure that the sample, solvent, and mobile phases are miscible in order to avoid sample precipitation. Filter sample with 0.2 μm membranes to remove particulates. If the sample is dissolved in a solvent, ensure that the membrane material does not dissolve in the solvent. Contact the membrane manufacturer with solvent compatibility questions. Acrodisc filters are recommended (refer to the Waters Quality Parts, Chromatography Columns and Supplies Catalog for additional information). Please consider that some analytes can be retained on certain membrane materials and result in lower recovery (or lower detector signal) than expected. Alternatively, centrifugation for 20 minutes at 8000 rpm, followed by the transfer of the supernatant liquid to an appropriate vial could be considered. b. Solvents To maintain maximum column performance, use high quality chromatography grade solvents. Solvents containing suspended particulate materials will generally clog the outside surface of the inlet distribution frit of the column. This will result in higher operating pressure, reduced column lifetime, and compromised performance. are designed for supercritical (compressed C 2 ) applications and may be used with compressed C 2 and the following co-solvents (individually and in miscible combinations): alcohols (typically methanol, ethanol, isopropanol), acetonitrile, and alkanes (typically heptane and hexane). Under SFC conditions, the co-solvent elution strength is in the following order: methanol > ethanol > isopropanol > acetonitrile > alkanes. When dissolving samples for analysis, use an injection solvent that is weaker to avoid peak distortion. Employing blends of methanol, isopropanol, acetonitrile, and/or heptane commonly maintains analyte solubility while avoiding strong injection solvent peak distortion. Water tends to have an irreversible effect on selectivity. While not damaging the polysaccharide, the selectivity will be permanently effected upon exposure. contain a modified polysaccharide coating which can be damaged when exposed to stronger solvents. Solvents that must be avoided include: Ketones such as acetone, methyl ethyl ketone (MEK) Halogenated solvents such as chloroform, dichloromethane, methylene chloride Ethyl acetate Tetrahydrofuran (THF) Methyl tert-butyl ether (MTBE) Toluene Dimethylformamide Dimethyl sulfoxide N-Methylformamide c. Additives can safely be used with commonly used acidic and basic additives used in supercritical fluid chromatography (SFC) such as trifluoroacetic acid (TFA), formic acid, ammonium acetate, ammonium formate, ammonium hydroxide, organic amines (such as diethylamine and triethylamine) and ammoniated methanol. Typical concentrations of 20 mm or 0.2% can be used safely. Consider the volatility, solubility, and detector compatibility when choosing an appropriate additive. Additives tend to improve peak shape and control the retention characteristics of analytes, but can also impart different selectivities. It is recommended to flush and remove additives and salts from the column before placing the column into storage. 4

d. Pressure Trefoil Column hardware and packing materials are packed and tested to pressures of up to 6,000 psi (414 bar or 41 Mpa). e. Temperature The recommended maximum operating temperature for is 40 C. f. Band Spreading Minimization The ACQUITY UPC 2 System has been designed to minimize band spreading. Deviation from Waters specified tubing could result in deterioration of chromatographic performance due to band spreading induced by inappropriate tubing I.D. Figure 1 shows the influence of tubing internal diameter on system band spreading and peak shape. As can be seen, the larger tubing diameter causes excessive peak broadening and lower sensitivity. 0.005 inches 0.020 inches 0.040 inches IV. TRUBLESHTING 1. ne of the most common problems with regards to columns is incorrect or insufficient priming of the co-solvent/modifier pump. If no peaks are observed after an injection or unusually long retention times are observed, check the priming of the co-solvent/modifier pump first. 2. If you see frosting on the column at the inlet or outlet, tighten the compression screw on that side. If tightening doesn t work, depressurize the system and the column, and then replace the fitting that is not sealing correctly. 3. If you continue to see frosting on the column, turn off the C 2 and vent the system. Allow the column to depressurize fully before disconnecting the inlet or outlet. Please contact your Waters representative for additional support. V. CLUMN CLEANING, ADDITIVES, AND STRAGE Changes in peak shape, peak splitting, shoulders on the peak, shifts in retention, change in resolution or increasing backpressure may indicate contamination of the column. Flushing with high concentrations of organic solvent, taking care not to precipitate buffers, is usually sufficient to remove the contaminant. If the flushing procedure does not solve the problem, purge the column using the following cleaning and regeneration procedures. Figure 1: Effect of connecting tubing on system. Diluted/distorted sample band a. Cleaning Use the cleaning routine that matches the properties of the samples and/or what you believe is contaminating the column (see Table 4). Placing the column on an ACQUITY UPLC System (or HPLC system) will enable the use of a wide range of solvents to improve cleaning.flush columns with 10 column volumes of solvent. Increasing column temperature increases cleaning efficiency. If the column performance is poor after cleaning and regenerating, call your local Waters office for additional support. Polar samples Non-polar samples 1. Methanol 1. Isopropanol 2. Isopropanol 2. Heptane 3. Methanol 3. Isopropanol 4. Mobile phase 4. Mobile phase Table 2. Column cleaning sequence. 5

Retest the column after using the cleaning procedure to determine if the specific problem has been fixed. If so, continue using the column, avoiding samples and solvents that may clog the column inlet. b. Effects of Additives Use of co-solvent additives (eg, ammonia, organic amines, formic acid, trifluoroacetic acid) can be used successfully with all Trefoil Columns. Consider the volatility and detector compatibility when choosing an appropriate additive. Additives tend to improve peak shape and control the retention characteristics of analytes, but can also impart different selectivities. c. Storage If additives were used, flush the column using a minimum of 5 column volumes 50% MeH/C 2. Store the columns in 100% C 2. At the time of manufacture, tracking and quality control information will be downloaded to the ecord. Storing this information on the chip will eliminate the need for a paper Certificate of Analysis. nce the user installs the column, the software will automatically download key parameters into a column history file stored on the chip. In this manual, we explain how the ecord will provide a solution for easily tracking the history of the columns, reduce the frustration of paperwork trails, and give users the reassurance that a well-performing column is installed onto their instruments. b. Installation Install the column into the ACQUITY Column Manager. Plug the ecord into the side of the column heater. nce the ecord is inserted into the column heater the identification and overall column usage information will be available allowing the user to access column information on their desktop. VI. INTRDUCING ECRD INTELLIGENT CHIP TECHNLGY a. Introduction The ecord Intelligent Chip will provide the history of a column s performance throughout its lifetime. The ecord is permanently attached to the column to assure that the column s performance history is maintained in the event that the column is moved from one instrument to another. Figure 3. Installing the ecord Intelligent Chip. ecord fob Figure 2. ecord Intelligent Chip. 6

c. Column Use Information The ecord Chip provides the user with column use data, column dimensions, and serial number. The overall column usage information includes the total number of samples, total number of injections, total sample sets, date of first injection, date of last injection, maximum pressure, and temperature. The information also details the column history by sample set including date started, sample set name, user name, system name, number of injections in the sample set, number of samples in the sample set, maximum pressure, and temperature in the sample set and if the column met basic system suitability requirements. Up to 50 sample sets can be stored on the ecord Chip. In addition, the ecord provides two-way communications between the ecord Chip and Empower Software. VII. ACQUITY UPC 2 CLUMN FAMILY The Care and Use Manual is one of three manuals for the Waters SFC Column family. Torus Columns Care and Use Manual 720005203EN Viridis Columns Care and Use Manual 720004349EN Waters, The Science of What s Possible, ACQUITY UPC, 2 Sep-Pak, asis, Waters Quality Parts, ACQUITY, Trefoil, Viridis, and Empower are registered trademarks of Waters Corporation. ecord is a trademark of Waters Corporation. All other trademarks are the property of their respective owners. 2016 Waters Corporation. Produced in the U.S.A. November 2016 720004828EN IH-PDF Waters Corporation 34 Maple Street Milford, MA 01757 U.S.A. T: 1 508 478 2000 F: 1 508 872 1990 www.waters.com