High Performance Liquid Chromatography (HPLC) of Three Analgesics

Similar documents
Determination of Chloride using Potentiometry

UV-Visible Absorption Analysis of a Two Component System

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

Safety in the Chemistry Laboratory

Chemical Reactions: The Copper Cycle

DATES: LAB: Liquid Chromatography Separation of Grape Kool-Aid

Lab Safety Rules GENERAL GUIDELINES. 1. Conduct yourself in a responsible manner at all times in the laboratory.

Chemistry Gas Chromatography: Separation of Volatile Organics

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Determination of Caffeine by HPLC

Figure 1. Structures for Vitamin B2 and Vitamin B1.

TECHNICAL BRODIFACOUM

Objective: Science Classroom Laboratory Safety

6. Common Laboratory Techniques

High Pressure/Performance Liquid Chromatography (HPLC)

Soil Cation Analysis Using High-Performance Capillary Zone Electrophoresis Last Modified: October 20, 2006

General Rules. Be alert and responsible at all times in the laboratory.

Experiment 20-Acid-Base Titration: Standardization of KOH and Determination of the Molarity and/or Percent Composition of an Acid Solution

3590/ Laboratory

Laboratory Safety SAFETY IN THE LABORATORY

Liquid storage: Holds the solvent which is going to act as the mobile phase. Pump: Pushes the solvent through to the column at high pressure.

Volumetric Analysis. Quantitative analysis answers the second question

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

Working with Solutions. (and why that s not always ideal)

Experiment 2: Analysis of Commercial Bleach Solutions

#26 Preparation of Solutions

Chemistry 143 Acid Base Titration Dr. Caddell. Titrating Acid

RATE LAW DETERMINATION OF CRYSTAL VIOLET HYDROXYLATION

HPLC Praktikum Skript

Quantitative Analysis of Caffeine in Energy Drinks by High Performance Liquid Chromatography

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

Ascorbic Acid Titration of Vitamin C Tablets

Titration with an Acid and a Base

Plop Plop, Fizz Fizz, Oh What A Relief It Is (Which Pain Reliever Works Fastest)

Diquat 1,1 -ethylene-2,2 -bipyridium dibromide salt Paraquat 1,1 -dimethyl-4,4 -bipyridium dichloride salt Initial Preparation

Chemistry 143 Experiment #11 Acid Base Titration Dr. Caddell. Titrating Acid

To understand concept of limiting reagents. To learn how to do a vacuum filtration. To understand the concept of recrystallization.

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Physical Separations and Chromatography

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Chromatography. Gas Chromatography

2. Synthesis of Aspirin

EXPERIMENT: LIMITING REAGENT. NOTE: Students should have moles of reactants in DATASHEET converted into masses in grams prior to the lab period.

Dr. Jonathan Gutow Fall Looking for PCBs in Water or Can PCBs Wash out of Landfills and Contaminate Ground Water?

Newton s Second Law of Motion

8 Enthalpy of Reaction

Exercise 4-4. Titration of a Buffer Solution EXERCISE OBJECTIVE DISCUSSION OUTLINE DISCUSSION. Buffer solutions

EXPERIMENT 9 ENTHALPY OF REACTION HESS S LAW

EXPERIMENT 20. Solutions INTRODUCTION

(b) Write the chemical equation for the dissolution of NaOH in water.

Exp 1 Column Chromatography for the Isolation of Excedrin Components. Reading Assignment: Column Chromatography, TLC (Chapter 18)

Exercise 4-3. Titration of Weak Acids EXERCISE OBJECTIVE DISCUSSION OUTLINE. The 5% rule DISCUSSION

C. Eugene Bennett Department of Chemistry

Plop Plop, Fizz Fizz, Oh What A Relief It Is (Which Pain Reliever Works Fastest)

Experiment 13. Dilutions and Data Handling in a Spreadsheet rev 1/2013

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

Prelab Reading Assignment: Laboratory Techniques in Organic Chemistry, 4 th Ed. Chapter 19

EOSMS Guidelines Date: 01/16/2014 Page 1 of 5

#30 Thermochemistry: Heat of Solution

Ascorbic Acid Titration of Vitamin C Tablets

Thermodynamics and the Solubility of Sodium Tetraborate Decahydrate

Determination of trace anions in concentrated hydrofluoric acid

METHANOLYSIS OF ACETAL

Experiment #10: Analysis of Antacids

CHAPTER CHROMATOGRAPHIC METHODS OF SEPARATIONS

Experiment 3: Analysis of Unknown Hydrocarbons by GC, NMR, and Chemical Tests for Unsaturation and Aromaticity

Introduction to Spectroscopy: Analysis of Copper Ore

SCHOOLS ANALYST COMPETITION 2012 MIDLANDS REGIONAL HEAT INSTRUCTION BOOKLET ROYAL SOCIETY OF CHEMISTRY ANALYTICAL DIVISION

MEASUREMENT: PART II

SCHOOLS ANALYST COMPETITION 2013 MIDLANDS REGIONAL HEAT INSTRUCTION BOOKLET

Experiment 7: ACID-BASE TITRATION: STANDARDIZATION OF A SOLUTION

PYRIPROXYFEN TECHNICAL

Chem 2115 Experiment #7. Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution, analysis of vinegar & antacid tablets

Photochemistry of Nitro-Polycyclic Aromatic Hydrocarbons. August 9, 2011

Chemical Reactions: Titrations

Standardizing a Solution of Sodium Hydroxide. Evaluation copy

7-A. Inquiry INVESTIGATION. 322 MHR Unit 3 Quantities in Chemical Reactions. Skill Check. Safety Precautions

Skill Building Activity 2 Determining the Concentration of a Species using a Vernier Spectrometer

CHM Salicylic Acid Properties (r16) 1/11

Experiment 2: THE DENSITY OF A SOLID UNKNOWN AND CALIBRATION WITH DATASTUDIO SOFTWARE

Buffers for Biological Systems Laboratory Instructor s Manual

aa + bb cc + dd Equation 1

Introductory Remarks:

3. Separation of a Mixture into Pure Substances

Density of Aqueous Sodium Chloride Solutions

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

Completion Match each each piece of equipment with its description. Please only put one number in the blank.

LIQUID CHROMATOGRAPHY

ANALYTICAL CHEMISTRY. Basic Skills in Chemical Laboratory

EXPERIMENT 7. Determination of Sodium by Flame Atomic-Emission Spectroscopy

experiment7 Explaining the difference between analyte and standard solutions. Know the definition of equivalence point.

Chemistry 1B Experiment 17 89

Column Liquid Chromatography Experiment Adapted for Use in Secondary Schools

ARTEMETHER AND LUMEFANTRINE ORAL SUSPENSION:Final text for addition to The International Pharmacopoeia (November 2008)

Experiment: Titration

Thin Layer Chromatography

INVESTIGATING GAS CHROMATOGRAPHY rev 8/12

Ascorbic Acid Titration of Vitamin C Tablets

Chem 2115 Experiment #7. Volumetric Analysis & Consumer Chemistry Standardization of an unknown solution and the analysis of antacid tablets

Experiment 7A ANALYSIS OF BRASS

Transcription:

High Performance Liquid Chromatography (HPLC) of Three Analgesics OVERVIEW/THEORY Real-life samples are often complex mixtures of many substances (e.g., sea water contains many salts and other impurities along with the water). When samples contain several components, the individual analytes often must be separated from each other prior to their detection so that their signals do not interfere with one another. High performance liquid chromatography (HPLC) is a powerful separation technique in which a liquid mobile phase (eluent) is passed through a solid stationary phase contained in a column. The analytes are separated (resolved) from each other as they are pumped through the column based upon their relative affinities for the stationary phase versus the mobile phase. After being separated on the column, the analytes pass directly to an in-line detector that generates a chromatogram showing the individual signal peaks versus time. The schematic diagram of a typical HPLC with UV-vis detector is shown below (Figure 1) Figure 1. Schematic of an HPLC. a, solvent reservoir; b, high-pressure liquid pump; c, highpressure injector valve; d, column; e, detector; f, chart recorder; g, waste reservoir. HPLC Page 1

Figure 2: HPLC system. a, solvent reservoir; b, high-pressure liquid pump; c, high-pressure injector valve; d, column; e, detector; f, chart recorder; g, waste reservoir (not shown). In this experiment, three analgesics are separated using reversed-phase HPLC in which a polar mobile phase is used with a nonpolar stationary phase. The column in this experiment contains a bonded C 18 (octadecylsilane) nonpolar organic stationary phase. The mobile phase is a polar methanol/water mixture also containing some acetic acid. The three analytes are eluted in order of decreasing polarity, with the most polar analyte, acetaminophen, coming off first followed by acetanilide and then phenacetin. Since the molecules all have some aromaticity, signal is provided by a UV-vis absorbance detector with the resulting chromatogram showing signal (absorbance units; AU) versus time. The concentrations of the three analytes will be determined by comparing their peak areas from a diluted sample mixture to the corresponding peak areas from a diluted standard mixture you prepare according to the directions below. METHOD Preparation of the standard mixture Using the analytical balance, accurately weigh between 0.1 to 0.2 grams acetaminophen on a small weighing boat and record the weight. Zero the balance, clean the spatula and weigh between 0.1 and 0.2 grams acetanilide into the same weighing boat and record the weight. Zero HPLC Page 2

the balance again, clean the spatula, and weigh between 0.1 and 0.2 grams phenacetin into the same weighing boat. Carefully transfer this mixture to a 250-mL volumetric flask with the HPLC grade methanol wash bottle from the chemical cabinet (do not use the methanol meant for rinsing). If you already have a few ml s of methanol in the flask before you transfer, this will aid in the dissolution of the solid. Add just enough methanol (about 20-25 ml) to dissolve the mixture. DO NOT CONTINUE TO THE NEXT STEP UNTIL THE SOLID HAS COMPLETELY DISSOLVED IN THE METHANOL. Add nanopure water while swirling the flask but stop before the mark. The methanol/water mix heats up slightly so let the flask cool to room temperature before bring the solution to the mark and mixing completely. The solution is too concentrated for analysis so dilute it by a factor of 20 and mark the diluted sample diluted working standard. Preparation of the unknown mixture Obtain the unknown from the instructor. This sample contains all three analytes that have been accurately weighed to four decimal places. Transfer the sample to a 250-mL volumetric flask with about 25-mL methanol and allow to fully dissolve before diluting with nanopure water. Fill the flask almost to the mark while swirling and then let it cool to room temperature before diluting to the mark and mixing. Dilute this sample also by a factor of 25 and label diluted unknown. Use of the HPLC * NOTE: the figures for the handheld controller for the HPLC is in the instruction only to give you an idea what to expect when you look at the controller screen. All the important information that you should be checking is in the TEXT of the instruction, not in the figures. Take the diluted standard and unknown to the instrument room. Pick up the handheld controller. Press ESC button until you see the screen that resembles below. If you have already done GC or GC/MS for this course, you should be familiar with the handheld controller. If you have not done those experiments, then, you should know one thing: the handheld controller does not use a touchscreen control. Instead, you should use the Function keys (F1, F2, F3, etc) near the screen. Figure 3: the Analysis screen. Note: this figure is shown as an example. You should see the screen that resembles this, but will not be identical. Make sure the Method listed on the screen says 466 (not ISOCRAT as shown above the Figure 3). If not, ask for assistance from the laboratory instructor. Press F7 to turn on the system (the pump and the detector). You should see the screen that resembles Figure 4. HPLC Page 3

Figure 4: System On and System Off Note: this figure is shown as an example. You should see the screen that resembles this, but will not be identical. Use the arrow keys to move from System Off to System On if necessary. Once you have System On selected, press Enter key. This should start the HPLC pump as well as turn on the lamp in the UV detector. You should hear the pump running. Wait about 5 minutes to allow the instrument to equilibrate. In your notebook, record the pump flow rate (ml/min), the approximate pump pressure (bar), and the wavelength (nm) of the UVvis detector. Each values should be approximately the following: Flow rate: 0.9 ml/min Pressure: 250-350 bar Wavelength: 254 nm At this point, you are ready to make your first injection of the diluted standard mixture. From here on, you must follow the instruction for each sample you run. First, make sure the injector is in the Load position (Figure 5). Figure 5: the injector. From the left, the injector without the syringe, the injector with the syringe in Load position, and the injector with syringe in Inject position. Rinse the 1-mL syringe several times with the diluted standard solution and then fill it completely. Insert the syringe at the center of the injector all the way in. Inject almost the entire contents of the syringe into the instrument, but be sure to leave a little left in the syringe in case an air bubble is there. What you have done is fill a small 20-µL loop many times, with the excess draining into the waste bottle. Leave the syringe in the injector once you have loaded the sample. You are now ready to begin the analysis. Quickly turn the injector to Inject position. The run should start at this point and the integrator should start moving. Once you see the first peak comes out, turn the injector back to Load position for the next injection. Do not inject until the run is complete. HPLC Page 4

After less than three minutes, the three analytes will be eluted as three separate peaks on the chromatogram. The run should stop in 3 minutes. The integrator should print out the result below the collected chromatogram. The chromatogram obtained should resemble Figure 6. Figure 6: sample chromatogram Your goal is to get three successive trials that the area of all of the three peaks agree with each other within about 1%. After running sufficient trials of the standard mixture, run the unknown. When you are finished, turn off the instrument by selecting the System Off as you turned on the system (see Figure 3). After turning off the system, the pump should stop and the pressure should go to zero. Make a table in your notebook showing the integrated peak areas for the three analytes for both the standard and unknown. You will also turn in all the integrator chart paper. Remember, the unknown and standard must both be run on the same day and lab period, ideally within 60 minutes of each other. If too much time passes between the injections of the standard and the unknowns, then results can be skewed because the response of the instrument changes over time (this is known as signal drift ). HPLC Page 5

CALCULATIONS To determine the amount of each analyte in your unknown, set up simple proportion between its average area in the unknown chromatogram versus its average signal in the standard chromatogram. Report the weight of each of component in the original sample to four decimal places. HPLC Page 6

Instrumental Analysis Laboratory Safety Rules A. Instructions: Carry out all manipulations in accordance with instructions and the safety rules and procedures given herein. B. Eye Protection: All students and staff working in the laboratory must wear safety glasses at all times. If a student needs to be reminded more than three times to wear goggles, she/he will be dismissed from lab for the remainder of the day, and will not be given an opportunity to make up the work. C. Apparel: The clothes you wear in lab are an important part of your safety equipment, and should offer protection from splashes/spills. Closed toed shoes (sneakers are fine), Full-length pants or a full-length skirt, and A shirt that completely covers your torso (i.e. at minimum, a t-shirt). In other words, you must NOT wear shorts to lab. You must NOT wear flip-flops, sandals, or crocs. You must NOT wear tank tops, halter tops, spaghetti-strap tops, or low cut jeans to lab. Exposed abdomens, hips, and backs are not safe in the lab. D. Gloves: Gloves are an important part of personal protection. Gloves will be available at all times in the laboratory. Your instructor will require their use when appropriate. E. Food: Food, drinks, and gum are not allowed in lab. None at all, not even water bottles. F. Sanitation Issues: Be sure to wash your hands before leaving lab, before you eat anything outside of lab, and before you answer your cell phone. G. Music: Individual headphones are not allowed. Your may choose to play music for the entire class. H. Cell Phones and Other Electronic Devices: Cellular phones and other electronic devices that you do not need to perform your laboratory work should be put away. I. Other: All students are explicitly prohibited from: 1. conducting any unauthorized experiments. 2. removing chemicals or apparatus from the laboratory for any reason. 3. working in the lab alone, or at other than regularly scheduled lab periods. 4. smoking in the laboratory or within 20 feet of any doorway. 5. impeding movement in aisles or through doorways with bags, skateboards, etc.