Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Similar documents
Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Direct Analysis of Surfactants using HPLC with Charged Aerosol Detection

Sensitive HPLC Method for Triterpenoid Analysis Using Charged Aerosol Detection with Improved Resolution

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

Determination of the Composition of Natural Products by HPLC with Charged Aerosol Detection. Introduction. Black Cohosh. Corona Detector Parameters

Quantitation and Characterization of Copper and Nickel Plating Bath Additives by Liquid Chromatography

Enhancement of Linearity and Response in Charged Aerosol Detection

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Charged Aerosol Detection and Evaporative Light Scattering Detection Fundamental Differences Affecting Analytical Performance

Monitoring Protein PEGylation with Ion Exchange Chromatography

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

Analytical Methods to Qualify and Quantify PEG and PEGylated Biopharmaceuticals

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

Advances in Sample Preparation for Accelerated Solvent Extraction

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

Determination of Carbonyl Compounds in Workplace Air

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Use of Charged Aerosol Detection as an Orthogonal Quantification Technique for Drug Metabolites in Safety Testing (MIST)

Simultaneous Determination of Pharmaceutical Peptides and Acetate Counterions by HPLC Using a Mixed-Mode Weak Anion-Exchange Column

Determination of trace anions in concentrated hydrofluoric acid

Comprehensive Neurochemical Profiling of Brain Tissue Samples

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Semi-Targeted Screening of Pharmaceutically- Related Contaminants in the Thames Tideway using LC-HRMS

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

HILIC Method Development in a Few Simple Steps

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

of mass spectrometry

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Fast methods for the determination of ibuprofen in drug products

The Low-Temperature Evaporative Light-Scattering Detector (LT-ELSD)

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Method Development for a Simple and Reliable Determination of PCBs in Mineral Insulating Oil by SPME-GC-ECD

New Developments in Surfactant Analysis Using HPLC

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Expanded capabilities in ammonium and amine detection

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Fitting of ETD Rate Constants for Doubly and Triply Charged Ions in a Linear Ion Trap

Varian 380-LC. ELSD for Routine HPLC and GPC Analysis. Key Benefits

Discover what you re missing

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Varian 385-LC. ELSD for Low Molecular Weight Compounds

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

A Total Solution for Explosives Analysis by Reversed-Phase HPLC with a Parallel HPLC System

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

[ Care and Use Manual ]

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of Methylamine in Drug Products

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

Supported Liquid Extraction (SLE) Guide and FAQ s. User Guide

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Direct Determination of Small Organic Acids in Sea Water by IC-MS

APPLICATION NOTE Goal. Introduction. Keywords

Introduction to Fourier Transform Infrared Spectroscopy

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Determination of urea in ultrapure water by IC-MS/MS

No Chromophore? No Problem! Universal HPLC Detection. The world leader in serving science

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Transcription:

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection Marc Plante, Bruce Bailey, and Ian N. Acworth Thermo Fisher Scientific, Chelmsford, MA, USA

Overview Purpose: To develop HPLC methods for the quantitation of surfactants in a variety of products and matrices. Methods: Methods using reversed phase and the specialty Thermo Scientific Acclaim Surfactant Plus HPLC column, along with the Thermo Scientific Dionex Corona ultra RS charged aerosol detector and UltiMate 3000 LC systems are outlined. Results: Surfactants in machine oils were detered qualitatively, while poloxamers in consumer products were detered quantitatively. Other consumer products were analyzed to show method specificity. Introduction High performance liquid chromatography (HPLC) is the ideal tool for the analysis of surfactants in samples. However, since surfactants do not typically possess a chromophore, sensitive detection by ultraviolet or fluorescence is difficult. Universal detectors, such as refractive index (RI) and evaporative light scattering (ELS) exist, but each has significant limitations with sensitivity and capability. Mass spectrometry (MS) is a useful tool for these analytes but is more expensive to own and operate in a quality control environment than a simpler detector such as the Corona charged aerosol detector. Sample preparation is simple and does not require derivatization since the detector responds to all non-volatile analytes. Samples can simply be diluted and analyzed. Different methods are presented that allow either characterization or quantification of surfactants. Examples include both aqueous and eral oil-based matrices. The charged aerosol detector is a sensitive, mass-based detector, especially wellsuited for the deteration of any nonvolatile and most semi-volatile analytes. The charged aerosol detector uses nebulization to create aerosol droplets (Figure 1). The mobile phase evaporates in the drying tube, leaving analyte particles, which become charged in the mixing chamber. The charge is then measured by a highly sensitive electrometer, providing reproducible, nanogram-level sensitivity. This technology has greater sensitivity and precision than ELSD and RI, is gradient compatible and is simpler to operate than a mass spectrometer. Compounds do not have to possess a chromophore (unlike UV detection) or be ionized (as with MS). Analyte response for the charged aerosol detectors is largely independent of chemical structure). With little chemical dependence, clear relationships among different analytes in a sample can be inferred. These attributes when combined with the specificity of the new Acclaim Surfactant Plus column for reversed phase analysis, provide a unique analytical solution for sensitive, reproducible, and routine analysis of surfactant-containing samples. FIGURE 1. Schematic and functioning of charged aerosol detection 1 3 10. 2 4 5 8 6 7 9 1. Liquid eluent enters from HPLC system 2. Pneumatic nebulization occurs 3. Small droplets enter drying tube 4. Large droplets exit to drain 5. Dried particles enter mixing chamber 6. Gas stream passes over corona needle 7. Charged gas collides with particles and charge is transferred 8. High mobility species are removed 9. Charge is measured by a highly sensitive electrometer 10. Signal transferred to chromatographic software 2 Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Methods Sample Preparations Samples and standards were dissolved in methanol/tetrahydrofuran (1:1), isopropyl alcohol/water (1:1), or water. Centrifugation (10,000 g for 3 ) was used if insoluble materials were present. Liquid Chromatography Metal Working Oils HPLC System: Thermo Scientific Dionex UltiMate 3000 LPG-3400SD pump (normal phase), WPS-3000RS autosampler, and TCC-3000RS column oven HPLC Column: Thermo Scientific Acclaim C8, 4.6 150 mm, 5 µm Column Temperature: 40 C Mobile Phase A: 50 mm ammonium acetate, ph 5 Mobile Phase B: 500 mm ammonium acetate, ph 5/methanol (:900) Mobile Phase C: tetrahydrofuran (unstabilized) Injection Volume: 10 50 µl Sample: Mineral oils in methanol/tetrahydrofuran (1:1) Detector: Corona ultra RS Nebulizer Temperature: 15 C Filter: 5 Data rate: 10 Hz PowerFunction: 1.00 Flow Gradient: Time () Flow Rate %A %B %C (ml/) 0 0.5 67 33 0 3 0.5 67 33 0 10 0.9 0 96 2 12 0.9 0 96 2 15 0.9 0 15 85 20 0.9 0 10 90 20 0.9 67 33 0 25 0.5 67 33 0 Liquid Chromatography Poloxamer Quantitation, Surfactants in Products HPLC System: Thermo Scientific Dionex UltiMate 3000 DGP-3600RS pump, WPS-3000RS autosampler, and TCC-3000RS column oven HPLC Columns: Acclaim Surfactant Plus, 3 µm, 3.0 150 mm and Acclaim Phenyl-1, 3 µm, 2.1 150 mm (2 nd column for poloxamer in product) Column Temperature: 50 C Mobile Phase A1: mm ammonium formate, ph 4 Mobile Phase A2 50 mm ammonium acetate, ph 5 in acetonitrile/water (1:1) Mobile Phase B1: n-propanol/acetone (1:1) Mobile Phase B2: 50 mm ammonium acetate, ph 5 in acetonitrile/methanol/water (5:4:1) Flow Rate: 0.4 ml/ (poloxamer), 0.6 ml/ (surfactants) Injection Volume: 20 µl (poloxamer), 5 µl (surfactant products) Samples: Fluoride mouth wash, containing Polaxamer 407, undiluted Other surfactant products were made into solutions of 50 µl or 50 mg in 1 ml of isopropyl alcohol/water (1:1) Detector: Corona ultra RS Nebulizer Temperature: ambient Filter: 5 Data Rate: 10 Hz PowerFunction: 1.75 (Poloxamer),1.00 (Surfactants) Flow Gradient: Poloxamers Surfactants Time () %A1 %B1 Curve Time () %A2 %B2-5 95 5 5-5 95 5 0 95 5 5 0 95 5 1 95 5 5 1 95 5 7 50 50 2 7 50 50 10 10 90 5 10 10 90 10 10 90 5 10 10 90 19 10 90 5 14 10 90 19 95 5 5 Data Analysis All HPLC chromatograms were obtained and compiled using Thermo Scientific Dionex Chromeleon Chromatography Data Station, 7.1 SR 1 software. Thermo Scientifi c Poster Note PN70539_E 03/13S 3

Results Analysis of Machine Oils No standards were available for these surfactants, so calibration curves were generated based on varying injection volumes. This provided a means of detering calibration fit, accuracy, and precision, with each injection volume analyzed in triplicate. A chromatogram of a metal working oil generated on a C8 column is shown in Figure 2, as well as a peak for unretained ions, the surfactants, and the eral oil. FIGURE 2. HPLC-Corona detector chromatogram of a eral oil (200 nl oil o.c.) containing surfactants 220 pa Mineral Oil 150 50 Surfactant Peaks 2 3 1 4 5 20-0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 17.0 18.0 19.0 20.0 A calibration curve, using analyte peak area (surfactant peak #4 vs. volume of eral oil injected, as nanoliters of oil per injection) was generated with good precision results, and this fit is shown in Figure 3. The correlation coefficient, r 2, of a second-order polynomial fit on inverted axed was 0.998, and the replicate injections has peak area percent residual standard deviations (%RSD) of 0.2 to 5% over the range of 40 to 200 nl of oil (as prepared). The method was specific, with a number of oils exhibiting different peaks in the surfactant region of the chromatogram. FIGURE 3. Calibration curve of peak 4 (in Figure 2) in eral oil from 50 to 200 nl (as prepared) of oil on column (n=3) 250 225 nl Oil o.c. 200 175 150 125 75 50 25 Area [pa*] 0 0.00 0.25 0.50 0.75 1.00 1.25 1.50 1.75 2.00 2.25 2.50 2.75 3.00 Based on the calibration results above and with the appropriate standard references, the method can be used for the deteration of surfactant materials in oil-based products. Analysis of Poloxamers in an over-the-counter product Poloxamers are a special class of non-ionic surfactants, consisting of a triblock copolymer of one polypropylene oxide connected to two polyethylene blocks. They are used in many detergents, consumer, and pharmaceutical products. They typically lack a chromophore so cannot be measured using ultraviolet absorption detectors. Furthermore, their polymeric nature often produces broad peaks which make the use of evaporative light scattering detectors challenging (poor sensitivity and reproducibility). However, such issues are not problematic for the Corona ultra RS detector when used with the specificity of the Acclaim Surfactant Plus column. Other matrix components can contribute to interference problems, as was encountered with a fluoride mouth rinse containing the Poloxamer 407 surfactant. 4 Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

A small shoulder was found on the front of the Polxamer 407 peak in a chromatogram using the Acclaim Surfactant Plus column, and mobile phase changes were not sufficient to resolve the two analytes. The use of a second HPLC column was required to separate the Poloxamer 407 (Pluronic 407) surfactant from the matrix peak. Based upon the high number of double bonds in the poloxamer compounds, an Acclaim Phenyl-1 column was chosen and placed after the Acclaim Surfactant Plus column. Consequently, the polaxamer moved from the central peak to the last peak in the chromatogram. A linear calibration curve was then fit to data generated from triplicate injections from 156 to 20,000 ng o.c. The fit exhibited high correlation (r 2 = 0.9993) and precision (calibration %RSD= 2.30). Replicate precision of peak area varied from 0.28 to 2.09 % RSD, and the limit of quantitation was detered at 156 ng o.c., based on a signal-to-noise ratio of 10.0. A fluoride mouth rinse was then analyzed without dilution. Three samples were prepared for spike recovery deterations: a sample comprised of 980 µl of fluoride mouth wash with 20 µl each of water, a 1 mg/ml solution (+200 ng o.c., 10 ppm) of Pluronic F127, and a 0.5 mg/ml (+400 ng o.c., 20 ppm) solution of Pluronic F127. The overlaid chromatograms are shown in Figure 4. The results of the analysis of the fluoride mouth wash are presented in Table 1. Spike recovery values were calculated by dividing the found-spiked amount by the theoretical-spiked amount, and these values were within 15% of theoretical values. This demonstrates good sensitivity and accuracy, considering that the spike levels were less than 7% of the productʼs poloxamer concentration. FIGURE 4. HPLC chromatogram overlay of over-the-counter fluoride mouth rinse product, spiked and unspiked, and the Poloxamer 407 standard at 10,000 ng o.c. 30.0 pa 25.0 Cetylpyridinium 20.0 15.0 10.0 Unknown Poloxamer 407 10,000 ng o.c Product + 20 ppm Product 5.0 0.0-2.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 16.8 TABLE 1. Spike recovery results for Poloxamer 407 in fluoride mouth rinse at two spike values. Sample Theoretical Amount (ng o.c., ppm) Experimental Amount Found (ng o.c., ppm) Spike Recovery (%) Fluoride mouth rinse product -- 5894, 294.7 -- Spiked + 10 ppm 6094, 304.7 6123, 306.2 114.5 Spiked + 20 ppm 6294, 314.7 6342, 317.1 112.0 Other examples of surfactants in consumer products Additional examples of consumer products that contain a variety of surfactants were prepared and analyzed using the single column Acclaim Surfactant Plus method. Samples were dissolved in either water or a sample solvent that was compatible with reversed phase mobile phases. Concentrations of 50 mg/ml or 50 µl/ml were prepared and centrifuged. Figure 5 shows an HPLC chromatogram of a hair conditioning product containing five surfactants. Good resolution was achieved and the five surfactants were well resolved. Figure 6 shows the separation of two types of surfactants (alcohol ethoxylates and alkylsulfonates) in a sample of a laundry detergent. As can be seen, this approach clearly separates the different classes of surfactants as well as the different surfactants within each surfactant class and congeners. Thermo Scientifi c Poster Note PN70539_E 03/13S 5

FIGURE 5. HPLC chromatogram of a hair conditioner containing five different surfactants. 50 mg/ml in tetrahydrofuran/methanol/water/chloroform (12:40:5:40) 250 pa 200 150 Surfactants peaks: 1. Cetrimonium chloride 2. Stearalkonium chloride 3. Benhentrimonium methosulfate 4. Palmityl alcohol 5. Stearyl alcohol 5 3 50 1 2 4-5 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 FIGURE 6. HPLC chromatogram of sustainable laundry detergent containing two different classes of surfactants. Sample is 50 mg/ml in water. 160 pa 140 Alkylsulfonates 120 Ions, glycerol, alcohols 80 Alcohol ethoxylates 60 40 20-2 0.0 1.0 2.0 3.0 4.0 5.0 6.0 7.0 8.0 9.0 10.0 11.0 12.0 13.0 14.0 15.0 16.0 Conclusions Three different methods were used to characterize surfactants in a wide variety of different products. Analytes were detected using a charged aerosol detector which offers the best sensitivity and the most homogenous response factors of universal HPLC detectors. Surfactants in machine oil were separated on a C8 column and surfactants in a hair conditioner and a laundry detergent were resolved using the Acclaim Surfactant Plus column. A quantitative analysis was performed on a poloxomer contained in an over-the-counter fluoride mouth wash product: spiked recovery values were 112.0 114.5 % at two levels. Methods using the Acclaim Surfactant Plus column require use of a buffered mobile phase and a volatile, reversed-phase organic eluent. The design of the solid phase enables high specificity for the different surfactant classes, with high resolution between all analytes. Acknowledgement We thank BASF Corporation for the gift of the Pluronic polymers used in this evaluation. 6 Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

www.thermoscientific.com/dionex 2013 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. Pluronic is a registered trademark of BASF Corporation. All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific Inc. products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9001:2008 Certified. Australia +61 3 9757 4486 Austria +43 1 333 50 34 0 Belgium +32 53 73 42 41 Brazil +55 11 3731 5140 China +852 2428 3282 Denmark +45 70 23 62 60 France +33 1 60 92 48 00 Germany +49 6126 991 0 India +91 22 2764 2735 Italy +39 02 51 62 1267 Japan +81 6 6885 1213 Korea +82 2 3420 8600 Netherlands +31 76 579 55 55 Singapore +65 6289 1190 Sweden +46 8 473 3380 Switzerland +41 62 205 9966 Taiwan +886 2 8751 6655 UK/Ireland +44 1442 233555 USA and Canada +847 295 7500 PN70539_E 03/13S