Instruction Manual HI 96738C Chlorine dioxide ISM

Similar documents
Instruction Manual HI 96728C Nitrate ISM

Instruction Manual HI 96759C Maple Syrup Transmittance Analyzer

Instruction Manual HI Potassium LR and MR ISM

Instruction Manual HI 96720C Calcium Hardness ISM

HI Hardness ISM

Instruction Manual HI 96710C ph & Chlorine ISM

PHOTOMETER FOR BOAR SEMEN ANALYSIS

HI REDUCING SUGARS ISM for wine analysis

Instruction Manual HI Refractometer for Sodium Chloride Measurements

HI Refractometer for Sodium Chloride Measurements INSTRUCTION MANUAL

Instruction Manual HI Refractometer for Sucrose in Wine and Grape Products Measurements

Instruction Manual HI Refractometer for Invert Sugar Measurements

HI Free & Total Chlorine ISM

HI Propylene Glycol Refractometer INSTRUCTION MANUAL

HI Propylene Glycol Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA881 Refractometer for Invert Sugar Measurements

HI Seawater Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA886 Refractometer for Sodium Chloride Measurements

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

INSTRUCTION MANUAL Milwaukee Refractometer MA888 Refractometer for Ethylene Glycol Measurements

INSTRUCTION MANUAL Milwaukee Refractometer MA872 Refractometer for Fructose Measurements

Test Equipment Depot Washington Street Melrose, MA TestEquipmentDepot.com INSTRUCTION MANUAL. Milwaukee Refractometer

Instruction Manual C401/301/201/102/101. Portable Colorimeter. 99 Washington Street Melrose, MA Phone Toll Free

Instruction Manual. Waterproof Portable Colorimeter C401 / 301 / 201 / 101 / 102 / 103 / 104 / 105. Part of Thermo Fisher Scientific

HI 96811, HI HI 96813, HI 96814

HI HI 8633 HI HI 8734 HI

Instruction Manual. HI Refractometer for Sucrose in Wine and Grape Products Measurements

MA882, MA883, MA884, MA885

PC100 Handheld ph/cond./tds/sal. Meter. Instruction Manual

Tetraphenylborate Method Method to 7.0 mg/l K Powder Pillows

DR/4000 PROCEDURE NITRATE. Using Powder Pillows

Cadmium Reduction Method Method to 0.50 mg/l NO 3 N (LR) Powder Pillows

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

Cadmium Reduction Method Method to 30.0 mg/l NO 3 N (HR) Powder Pillows or AccuVac Ampuls

1,5-Diphenylcarbohydrazide Method* Powder Pillows or AccuVac Ampuls (0 to mg/l Cr 6+ )

Bante210 Benchtop ph Meter Instruction Manual

Mercaptoacetic Acid Method*

Cadmium Reduction Method (Using Powder Pillows or AccuVac Ampuls)

SC400 Colorimeter. Silica. Photometer-System. Instruction Manual. SiO 2. Bedienungsanleitung Seite Instruction Manual Page 19 33

Operating Manual. Turb 355 IR/T. Portable turbidimeter. ba75454e03 09/

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

DR/4000 PROCEDURE. IRON, Total

TPH (Total Petroleum Hydrocarbons)

Chlorine, Free and Total, High Range

FerroZine Method 1 Method to 100 µg/l Fe (10-cm cell) Reagent Solution. Instrument Adapter Sample cell DR 6000 LZV

Chloramine (Mono) and Nitrogen, Free Ammonia

Cadmium Reduction Method Method to 10.0 mg/l NO 3 N (MR, spectrophotometers) 0.2 to 5.0 mg/l NO 3 N (MR, colorimeters)

Chromium, Total. Alkaline Hypobromite Oxidation Method 1 Method to 0.70 mg/l Cr (spectrophotometers) 0.01 to 0.60 mg/l Cr (colorimeters)

NITROGEN, TOTAL, HR, Test N Tube

IR/Thermistor Pen. Instruction Manual

Cadmium Reduction Method Method 8171 MR (0.1 to 10.0 mg/l NO 3

DR/4000 PROCEDURE NITRATE. 2. The display will show:

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

ph electrode Instruction Manual For use with the RAH-210 & RPH-250

Chromotropic Acid Method Method to 30.0 mg/l NO 3 N (HR) Test N Tube Vials

Today s Weather 24 Hour Forecaster

Operation Manual. SPECTRO-NANO4 Nucleic Acid Analyzer PLEASE READ THIS MANUAL CAREFULLY BEFORE OPERATION

Spectrum 22ED Spectrophotometer #FOT Absorbency 0A to 1.99A Photometric accuracy 2.0% T Power requirement 115/230 VAC; Hz automatic Size

Powder Pillows 0.01 to 0.80 mg/l Al 3+ (colorimeters)

Instrument Sample cell orientation Sample cell DR 6000 DR 3800 DR 2800 DR 2700 DR 5000 DR The fill line is to the right.

HL-800K Infrared Laser Thermometer. -50 C to +800 C (13:1 ratio) User Manual

Chematest 25. Operator s Manual A Swan Analytical Instruments AG CH-8340 Hinwil/Switzerland ANALYTICAL INSTRUMENTS

DIAGNOSTICS NO TOUCH FOREHEAD THERMOMETER

Nitrogen, Ammonia. Test Preparation. Powder Pillows Method Method 8155 Salicylate Method 1 Powder Pillows

Chlorine, Free and Total, High Range

National Optical & Scientific Instrument Inc Tri-County Parkway Schertz, Texas Phone (210) Fax (210)

Determining the Concentration of a Solution: Beer s Law

Method to 0.50 mg/l NH 3 N Powder Pillows

Persulfate Digestion Method Method to 40 mg/l N (HR) TNTplus 827

OPERATING INSTRUCTIONS

Dimethylglyoxime Method Method to 6.0 mg/l Ni TNTplus 856

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

Powder Pillows 0.01 to 1.60 mg/l SiO 2 (LR, colorimeters)

Instruction Manual HI Grow Master Basic for Nutrient Analyses.

Phenolphthalein and Total Alkalinity Method to 4000 mg/l as CaCO 3 Digital Titrator

Method to 0.50 mg/l NH 3 N Powder Pillows

Package Insert for FASafe/AciSafe

ph 5 Meter Manual The LaMotte 5 Series

Sulfide DOC Methylene Blue Method 1 (0 0.70, and mg/l) Method Test procedure

SKELESCOPE REFLECTOR TELESCOPE WITH TABLE TOP TRIPOD SK-100 INSTRUCTION MANUAL

SkyScan EWS-PRO - Manual -

Experiment 18 - Absorption Spectroscopy and Beer s Law: Analysis of Cu 2+

Rapid RH Portable ph Meter. User's Manual

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

Persulfate Digestion Method Method to 25.0 mg/l N (LR) Test N Tube Vials

Persulfate Digestion Method Method to 150 mg/l N (HR) Test N Tube Vials

Introduction. Concepts Kinetics Order of reaction Reaction rate Colorimetry. Background

INSTRUCTION MANUAL. Laser Diode Sarcomere Length. Aurora Scientific Inc. 360 Industrial Parkway South, Unit 4 Aurora, Ontario, Canada L4G 3V7

INSTRUCTION MANUAL ANALOG COLORIMETER

Dimethylphenol Method Method to mg/l NO 3 N or 1.00 to mg/l NO

User Manual. Stick-Style Water-Resistant Infrared Thermometer with 8:1 Distance-to-Sight Ratio. Model with NIST-Traceable Calibration

Salicylate Method Method HR (2 to 47 mg/l NH 3 N) TNTplus 832

Phosphorus, Total. USEPA 1 PhosVer 3 with Acid Persulfate Digestion Method Method to 3.50 mg/l PO. Test preparation

Richter Optica. Instructions for Model: S850 Stereo Zoom Microscope

Mercuric Thiocyanate Method Method to 25.0 mg/l Cl Reagent Solution

Vet Photometer DP 700 Operating Manual. Version V5.9

C 214 Multiparameter Bench Photometer

CHROMA Colorimeter Model 252

Transcription:

Instruction Manual HI 96738C Chlorine dioxide ISM www.hannainst.com 1

Dear Customer, Thank you for choosing a Hanna product. This manual will provide you with the necessary information for the correct use of the instrument. Please read it carefully before using the meter. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. TABLE OF CONTENTS PRELIMINARY EXAMINATION... 3 GENERAL DESCRIPTION... 4 ABBREVIATIONS... 4 SPECIFICATIONS... 5 PRECISION AND ACCURACY... 5 PRINCIPLE OF OPERATION... 6 FUNCTIONAL DESCRIPTION... 7 ERRORS AND WARNINGS... 9 GENERAL TIPS FOR AN ACCURATE MEASUREMENT... 11 STARTUP... 13 MEASUREMENT PROCEDURE... 13 VALIDATION PROCEDURE... 15 CALIBRATION PROCEDURE... 17 GLP... 20 BATTERY MANAGEMENT... 21 BATTERY REPLACEMENT... 22 ACCESSORIES... 22 WARRANTY... 23 PRELIMINARY EXAMINATION Please examine this product carefully. Make sure that the instrument is not damaged. If any damage occurred during shipment, please notify your Dealer. Each HI 96738 Ion Selective Meter is supplied complete with: Two Sample Cuvettes and Caps Two CAL CHECK TM standard cuvettes HI 96738KIT 9V Battery Scissors Cloth for wiping cuvettes Instrument quality certificate Instruction Manual Rigid carrying case All rights are reserved. Reproduction in whole or in part is prohibited without the written consent of the copyright owner, Hanna Instruments Inc., Woonsocket, Rhode Island, 02895, USA. Note: Save all packing material until you are sure that the instrument works correctly. Any defective item must be returned in its original packing. 2 3

GENERAL DESCRIPTION The HI 96738 is an auto diagnostic portable microprocessor meter that benefits from Hanna s years of experience as a manufacturer of analytical instruments. It has the advanced optical system based on a special tungsten lamp and a narrow band interference filter that allows most accurate and repeatable readings. All instruments are factory calibrated and the electronic and optical design minimizes the need of frequent calibration. With the powerful CAL CHECK TM validation function, you are able to validate good performance of your instrument at any time. The validation procedure is extremely user friendly. Just use the exclusive HANNA ready-made, NIST traceable standards to verify the performance of the instrument and recalibrate if necessary. The instrument is splash waterproof and the lamp and filter units are protected from dust or dirt by a transparent cup. This makes the instruments fulfill field applications. Display messages aid the user in routine operation. The meter has an auto-shut off feature that will turn off the instrument after 10 minutes of non use in measurement mode or after 1 hour if left in calibration mode. The meter uses an exclusive positive-locking system to ensure that the cuvette is in the same position every time it is placed into the measurement cell. It is designed to fit a cuvette with a larger neck making it easier to add both sample and reagents. The cuvette is made from special optical glass to obtain best results. The HI 96738 meter measures the chlorine dioxide content in water samples in the 0.00 to 2.00 mg/l (ppm) range. The method is an adaptation of the Chlorophenol Red method. The reagents are liquid and powder form and are supplied in bottles. The amount of reagent is precisely dosed to ensure the maximum repeatability. C: F: mg/l: ml: LCD: ABBREVIATIONS degree Celsius degree Fahrenheit milligrams per liter. mg/l is equivalent to ppm (parts per million) milliliter Liquid Crystal Display SPECIFICATIONS Range 0.00 to 2.00 mg/l Resolution 0.01 mg/l Accuracy ±0.10 mg/l ±5% of reading @ 25 C Typical EMC Deviation ±0.01 mg/l Light Source Tungsten Lamp Light Detector Silicon Photocell with narrow band interference filter @ 575 nm Method Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and reagents causes a colorless to purple tint in the sample. Environment 0 to 50 C (32 to 122 F); max 95% RH non-condensing Battery Type 1 x 9 volt Auto-Shut off After 10 of non-use in measurement mode; after 1 hour of non-use in calibration mode; with last reading reminder. Dimensions 192 x 104 x 69 mm (7.6 x 4.1 x 2.7 ) Weight 360 g (12.7 oz.). REQUIRED REAGENTS Code Description Quantity/test HI 93738A-0 Reagent A 0.5 ml HI 93738B-0 Reagent B 1 packet HI 93738C-0 Reagent C 0.5 ml HI 93738D-0 Reagent D 0.5 ml Precision is how closely repeated measurements agree with each other. Precision is usually expressed as standard deviation (SD). Accuracy is defined as the nearness of a test result to the true value. Although good precision suggests good accuracy, precise results can be inaccurate. The figure explains these definitions. In a laboratory using a standard solution of 0.80 mg/l chlorine dioxide and a representative lot of reagents, an operator obtained with a single instrument a standard deviation of 0.02 mg/l. PRECISION AND ACCURACY 4 5

PRINCIPLE OF OPERATION Absorption of Light is a typical phenomenon of interaction between electromagnetic radiation and matter. When a light beam crosses a substance, some of the radiation may be absorbed by atoms, molecules or crystal lattices. If pure absorption occurs, the fraction of light absorbed depends both on the optical path length through the matter and on the physical-chemical characteristics of the substance according to the Lambert-Beer Law: -log I/I o = ε λ c d or A = ε λ c d Where: -log I/I o = Absorbance (A) I o = intensity of incident light beam I = intensity of light beam after absorption ε λ = molar extinction coefficient at wavelength λ c = molar concentration of the substance d = optical path through the substance Therefore, the concentration "c" can be calculated from the absorbance of the substance as the other factors are known. Photometric chemical analysis is based on the possibility to develop an absorbing compound from a specific chemical reaction between sample and reagents. Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements. The optical system of Hanna's HI 96738 colorimeter is based on a tungsten lamp combined with a narrow-band interference filter to guarantee both high performance and reliable results. A microprocessor controlled special tungsten lamp emits radiation which is first optically conditioned and beamed to the sample contained in the cuvette. The optical path is fixed by the diameter of the cuvette. Then the light is spectrally filtered to a narrow spectral bandwidth, to obtain a light beam of intensity I o or I. The photoelectric cell collects the radiation I that is not absorbed by the sample and converts it into an electric current, producing a potential in the mv range. The microprocessor uses this potential to convert the incoming value into the desired measuring unit and to display it on the LCD. The measurement process is carried out in two phases: first the meter is zeroed and then the actual measurement is performed. The cuvette has a very important role because it is an optical element and thus requires particular attention. It is important that both, the measurement and the calibration (zeroing) cuvettes, are optically identical to provide the same measurement conditions. Whenever possible use the same cuvette for both. It is necessary that the surface of the cuvette is clean and not scratched. This to avoid measurement interference due to unwanted reflection and absorption of light. It is recommended not to touch the cuvette walls with hands. Furthermore, in order to maintain the same conditions during the zeroing and the measuring phases, it is necessary to close the cuvette to prevent any contamination. INSTRUMENT DESCRIPTION FUNCTIONAL DESCRIPTION 1) GLP/ key 2) CAL CHECK key 3) ZERO/CFM key 4) READ/ /TIMER key 5) ON/OFF key 6) Liquid Crystal Display (LCD) 7) Cuvette alignment indicator 8) Cuvette holder HI 96738 block diagram (optical layout) 6 7

KEYPAD DESCRIPTION ON/OFF: to turn the meter on and off. ZERO/CFM: this is a bi-functional key. Just press to zero the meter prior to measurement, or to confirm edited values. In or to confirm factory calibration restore. READ/ /TIMER: this is a multi-functional key. In measurement mode, press to make a measurement, or press and hold for three seconds to start a pre-programmed countdown prior to measurement. In GLP mode press to view the next screen. CAL CHECK: this is a bi-functional key. Just press to perform the validation of the meter, or press and hold for three seconds to enter calibration mode. GLP/ : this is a bi-functional key. Just press to enter GLP mode. In calibration mode press to edit the date and time. ERRORS AND WARNINGS The instrument shows clear messages when erroneous condition appears. Messages are also displayed when the obtained values are outside expected range. The beeper is playing a beep on errors. a) on zero reading Light High: There is too much light to perform a measurement. Please check the preparation of the zero cuvette. OPERATING MODES Measurement mode: default operation mode, enables both validation and measurement. Calibration mode: may be entered by keeping CAL CHECK pressed for three seconds (the CAL tag appears), it enables calibration of the instrument. GLP mode may be entered by pressing GLP/ ( GLP appears), it enables consulting of user calibration date or restore factory calibration. DISPLAY ELEMENTS DESCRIPTION 5 6 7 Light Low: There is not enough light to perform a measurement. Please check the preparation of the zero cuvette. No Light:The instrument cannot adjust the light level. Please check that the sample does not contain any debris. 4 3 2 1 1) The measuring scheme (lamp, cuvette, detector), appears during different phases of zero or reading measurement 2) Error messages and warnings 3) The battery icon shows the charging level of the battery 4) The hourglass appears when an internal checking is in progress 5) Status messages 6) The chronometer appears when the reaction timer is running 7) The month, day and date icons appear when a date is displayed 8) Four digit main display 9) Measuring units 10) Four digit secondary display 8 9 10 b) on sample reading Inverted cuvettes: The sample and the zero cuvette are inverted. Zero: A zero reading was not taken. Follow the instructions of the measurement procedure for zeroing the meter. 8 9

Under range: A blinking 0.00 indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure you use the same cuvette for reference (zero) and measurement. Battery low: The battery must be replaced soon. Over Range: A flashing value of the maximum concentration indicates an over range condition. The concentration of the sample is beyond the programmed range: dilute the sample and re-run the test. Dead battery: This indicates that the battery is dead and must be replaced. Once this indication is displayed, the meter will lock up. Change the battery and restart the meter. c) during calibration procedure Standard Low: The standard reading is less than expected. Standard High: The standard reading is higher than expected. GENERAL TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure best accuracy. Color or suspended matter in large amounts may cause interference, therefore these should be removed by treatment with active carbon and by prior filtration. d) other errors and warnings Cap error: Appears when external light enters in the analysis cell. Assure that the cuvette cap is present. For a correct filling of the cuvette: the liquid in the cuvette forms a concavity on the top; the bottom of this concavity must be at the same level of the 10 ml mark. Proper use of the powder reagent packet: (a) use scissors to open the powder packet; (b) push the edges of the packet to form a spout; (c) pour out the content of the packet. Cooling lamp: The instrument waits for the lamp to cool down. 10 11

It is important that the sample does not contain any debris. This would corrupt the reading. In order to avoid reagent leaking and to obtain more accurate measurements, it is recommended to close the cuvette first with the supplied HPDE plastic stopper and then with the black cap. Each time the cuvette is used, the cap must be tightened to the same degree. Whenever the cuvette is placed into the measurement cell, it must be dry outside, and completely free of fingerprints, oil or dirt. Wipe it thoroughly with HI 731318 or a lint-free cloth prior to insertion. Shaking the cuvette can generate bubbles in the sample, causing higher readings. To obtain accurate measurements, remove such bubbles by swirling or by gently tapping the cuvette. Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost. It is possible to take multiple readings in a row, but it is recommended to take a new zero reading for each sample and to use the same cuvette for zeroing and measurement. After the reading it is important to discard immediately the sample, otherwise the glass might become permanently stained. All the reaction times reported in this manual are referred to 20 C (68 F). As a general rule of thumb, they should be doubled at 10 C (50 F) and halved at 30 C (86 F). In order to maximize accuracy, prior to a measurement follow the validation procedure to be sure that the instrument is properly calibrated. If necessary, calibrate the instrument. STARTUP Prepare the instrument for measurement as follows: Unpack the instrument by removing the dust protection sleeve from the instrument cuvette holder. Place the battery in the instrument as described in the BATTERY REPLACEMENT chapter. Place the instrument on a flat table. Do not place the instrument under direct sun light. MEASUREMENT PROCEDURE To compensate the meter for the sample turbidity or color, the measurement takes place in two phases. First, the meter is zeroed using the unreacted sample. After the reagents are added the reacted sample is measured. Turn the meter on by pressing ON/OFF. The display briefly shows all tags on. When the beeper sounds briefly and the LCD displays dashes, the meter is ready. The blinking ZERO indicates that the instrument needs to be zeroed first. Fill two graduated mixing cylinders (#1 & #2) up to the 25 ml mark with the sample. Add 0.5 ml of HI 93738A-0 Reagent A to each cylinder (#1 & #2), close them and invert several times to mix. Add the content of one packet of HI 93738B-0 Reagent B to only one of the two cylinders (#1), close it and invert several times to mix. This is the blank. 12 13

Add precisely 0.5 ml of HI 93738C-0 Reagent C to each cylinder (#1 & #2), close them and invert several times to mix. Add 0.5 ml of HI 93738D-0 Reagent D to each cylinder (#1 & #2), close them and invert several times to mix. Cylinder 2 is the sample. Fill a cuvette up to the 10 ml mark with blank (#1) and replace the cap. Place the blank (#1) into the holder and ensure that the notch on the cap is positioned securely into the groove. Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. After a few seconds, the display will show -0.0-. The meter is now zeroed and ready for measurement. Press READ/ and the display will show the lamp, cuvette and detector icons, depending on the measurement phase. At the end of measurement, the instrument directly displays concentration in mg/l of chlorine dioxide on the LCD. SAMPLING PROCEDURE: It is recommended to analyze chlorine dioxide samples immediately upon collection. Chlorine dioxide samples must be stored in dark, glass stoppered bottles, with minimal head space. Excessive heat (greater than 25 C/78 F), agitation and exeposure to light must be avoided. INTERFERENCES: Interference may be caused by strong oxidants. VALIDATION PROCEDURE Use the validation procedure to ensure that the instrument is properly calibrated. Warning: Do not validate the instrument with any standard solutions other than the HANNA CAL CHECK TM Standards, otherwise erroneous results will be obtained. For accurate validation, please perform test at room temperature: 18 to 25 C (64.5 to 77.0 F) Fill a cuvette up to the 10 ml mark with the reacted sample (#2) and replace the cap. Insert the sample into the instrument and ensure that the notch on the cap is positioned securely into the groove. Turn the meter on by pressing ON/OFF. When the beeper sounds briefly and the LCD displays dashes, the meter is ready. 14 15

Place the CAL CHECK TM Standard HI 96738-11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. CALIBRATION PROCEDURE Note: It is possible to interrupt calibration procedure at any time by pressing CAL CHECK or ON/OFF keys. Warning: Do not calibrate the instrument with standard solutions other than the HANNA CAL CHECK TM Standards, otherwise erroneous results will be obtained. For accurate calibration, please perform test at room temperature: 18 to 25 C (64.5 to 77.0 F) Turn the meter on by pressing ON/OFF. After a few seconds, the display will show -0.0-. The meter is now zeroed and ready for validation. When the beeper sounds briefly and the LCD displays dashes, the meter is ready. Remove the cuvette. Place the CAL CHECK TM Standard HI 96738-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. Press CAL CHECK TM and the lamp, cuvette and detector icons together with CAL CHECK will appear on the display, depending on the measurement phase. At the end of the measurement the display will show the validation standard value. The reading should be within specifications as reported in the CAL CHECK TM Standard Certificate. If the value is found out of the specifications, please check that the cuvettes are free of fingerprints, oil or dirt and repeat validation. If results are still found out of specifications, then recalibrate the instrument. Press and hold CAL CHECK for three seconds to enter calibration mode. The display will show CAL during calibration procedure. The blinking ZERO asks for instrument zeroing. Place the CAL CHECK TM Standard HI 96738-11 Cuvette A into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. Press ZERO/CFM and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. 16 17

After a few seconds the display will show -0.0-. The meter is now zeroed and ready for calibration. The blinking READ asks for reading calibration standard. Remove the cuvette. Place the CAL CHECK TM Standard HI 96738-11 Cuvette B into the cuvette holder and ensure that the notch on the cap is positioned securely into the groove. DATE INPUT Press GLP/ to edit the desired year (2000-2099). If the key is kept pressed, the year number is automatically increased. When the correct year has been set, press ZERO/CFM or READ/ /TIMER to confirm. Now the display will show the month blinking. Press GLP/ to edit the desired month (01-12). If the key is kept pressed, the month number is automatically increased. or Press READ/ /TIMER and the lamp, cuvette and detector icons will appear on the display, depending on the measurement phase. After measurement the instrument will show for three seconds the Cal Check Standard value. Note: If the display shows STD HIGH, the standard value was too high. If the display shows STD LOW, the standard value was too low. Verify that both CAL CHECK TM Standards HI 96738-11 Cuvettes, A and B are free from fingerprints or dirt and that they are inserted correctly. Then the date of the last calibration (e.g.: 01.08.2005 ) appears on the display, or 01.01.2005 if the factory calibration was selected before. In both cases the year number is blinking, ready for date input. When the correct month has been set, press ZERO/CFM or READ/ /TIMER to confirm. Now the display will show the day blinking. Press GLP/ to edit the desired day (01-31). If the key is kept pressed, the day number is automatically increased. Note: It is possible to change the editing from day to year and to month by pressing READ/ /TIMER. Press ZERO/CFM to save the calibration date. The instrument displays Stor for one second and the calibration is saved. or 18 19

The instrument will return automatically to the measurement mode by displaying dashes on the LCD. Press READ/ /TIMER to enter in the factory calibration restore screen. The instrument asks for confirmation of user calibration delete. In the GLP mode, the last user calibration date can be consulted and the factory calibration can be restored. LAST CALIBRATION DATE To display the calibration date: Press GLP/ to enter GLP mode. The calibration month and day will appear on the main display and the year on the secondary display. If no calibration was performed, the factory calibration message, F.CAL will appear on the main display and the instrument returns to measurement mode after three seconds. FACTORY CALIBRATION RESTORE It is possible to delete the calibration and restore factory calibration. Press GLP/ to enter GLP mode. GLP Press ZERO/CFM to restore the factory calibration or press GLP/ again to abort factory calibration restore. The instrument briefly notifies done when restores factory calibration and returns to measurement mode. BATTERY MANAGEMENT To save battery, the instrument shuts down after 10 minutes of non-use in measurement mode and after 1 hour of non-use in calibration mode. If a valid measurement was displayed before auto shut off, the value is displayed when the instrument is switched on. The blinking ZERO means that a new zero has to be performed. One fresh battery lasts for around 750 measurements, depending on the light level. The remaining battery capacity is evaluated at the instrument startup and after each measurement. The instrument displays a battery indicator with three levels as follows: 3 lines for 100 % capacity 2 lines for 66 % capacity 1 line for 33 % capacity Empty battery icon blinking if the capacity is under 10 %. If the battery is empty and accurate measurements can t be taken anymore, the instrument shows dead batt and turns off. To restart the instrument, the battery must be replaced with a fresh one. 20 21

BATTERY REPLACEMENT WARRANTY To replace the instrument s battery, follow the steps: Turn the instrument off by pressing ON/OFF. Turn the instrument upside down and remove the battery cover by turning it counterclockwise. Extract the battery from its location and replace it with a fresh one. Insert back the battery cover and turn it clockwise to close. HI 96738 is warranted for two years against defects in workmanship and materials when used for its intended purpose and maintained according to the instructions. This warranty is limited to repair or replacement free of charge. Damages due to accident, misuse, tampering or lack of prescribed maintenance are not covered. If service is required, contact your dealer. If under warranty, report the model number, date of purchase, serial number and the nature of the failure. If the repair is not covered by the warranty, you will be notified of the charges incurred. If the instrument is to be returned to Hanna Instruments, first obtain a Returned Goods Authorization Number from the Customer Service Department and then send it with shipment costs prepaid. When shipping any instrument, make sure it is properly packaged for complete protection. To validate your warranty, fill out and return the enclosed warranty card within 14 days from the date of purchase. ACCESORIES REAGENT SET HI 93738-01 HI 93738-03 Reagents for 100 tests Reagents for 300 tests OTHER ACCESORIES HI 96738-11 CAL CHECK TM Standard Cuvettes (1 set) HI 721310 9V battery (10 pcs.) HI 731318 Cloth for wiping cuvettes (4 pcs.) HI 731331 Glass cuvettes (4 pcs.) HI 731335 Caps for cuvettes HI 93703-50 Cuvette cleaning solution (230 ml) HI 741218 Carrying case Recommendations for Users Before using these products, make sure that they are entirely suitable for your specific application and for the environment in which they are used. Operation of these instruments may cause unacceptable interferences to other electronic equipments, this requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance. To avoid damages or burns, do not put the instrument in microwave oven. For yours and the instrument safety do not use or store the instrument in hazardous environments. Hanna Instruments reserves the right to modify the design, construction and appearance of its products without advance notice. 22 23

Hanna Instruments Inc. Highland Industrial Park 584 Park East Drive Woonsocket, RI 02895 USA Technical Support for Customers Tel. (800) 426 6287 Fax (401) 765 7575 E-mail tech@hannainst.com www.hannainst.com Local Sales and Customer Service Office Printed in EUROPE (ROMANIA) MAN96738 12/10 24