SEPARATION OF FLURBIPROFEN AND IBUPROFEN ENANTIOMERS ON A CHIRAL STATIONARY PHASE USING SUPERCRITICAL FLUIDS

Similar documents
Application Note. Authors. Abstract. Pharmaceuticals

SUPERCRITICAL FLUID CHROMATOGRAPHY AS SUCCESSFUL SEPARATION TOOL IN CHEMICAL AND PHARMACEUTICAL INDUSTRY

Optical Isomer Separation Columns and Packing Materials

Basic Principles for Purification Using Supercritical Fluid Chromatography

High Performance Liquid Chromatography

SUPERCRITICAL FLUID CHROMATOGRAPHY PROCESS OPTIMISATION OF THE SEPARATION OF TOCOPHEROL HOMOLOGUES

Chiral Flash Columns

CHIRAL STATIONARY PHASES

Chapter content. Reference

Instrumental Chemical Analysis

CHIRAL SEPARATION USING THIN LAYER CHROMATOGRAPHY

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Determination of Ibuprofen Isotherm Using Supercritical Fluid Chromatography

Chiral Separation Using SFC and HPLC

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Supercritical Fluid Chromatography

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

ChromegaChiral TM CSP Media and Columns

High Pressure/Performance Liquid Chromatography (HPLC)

Liquid Chromatography

M. Sajewicz, M. Gontarska, D. Kronenbach, Ł. Wojtal, G. Grygierczyk, and T. Kowalska

ON PROBLEMS WITH LIQUID CHROMATOGRAPHIC QUANTIFICATION OF CHIRAL 2-ARYLPROPIONIC ACIDS BY USE OF UV-ABSORPTION-BASED DETECTION

CHEM 429 / 529 Chemical Separation Techniques

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Chapter 29 Supercritical Fluid Chromatography and Extraction 29A Properties of supercritical fluids

Expediting Achiral SFC Method Development Using a Multi-Channel SFC System with MS Detection

Separation of Chiral Pesticides by SFC and HPLC

Enantiomeric and Diastereomeric Separations of Pyrethroids Using UPC 2

LIQUID CHROMATOGRAPHY

Use of the SFC-MS Splitter Kit G

Lab.2. Thin layer chromatography

penta-hilic UHPLC COLUMNS


ChromegaChiral TM CSP Media and Columns

What is Chromatography?

Chiral Columns for enantiomer separation by HPLC

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

Chromatography and other Separation Methods

Chapter 1. Chromatography. Abdul Muttaleb Jaber

Abstract: An minimalist overview of chromatography for the person who would conduct chromatographic experiments, but not design experiments.

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

FAQ's. 1 - Which column would be the most appropriate for my application?

LC Technical Information

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

LIQUID CHROMATOGRAPHY

Nexera UC Unified Chromatography

Thermo Scientific Accucore XL HPLC Columns. Technical Manual

CHROMTECH - CHIRAL HPLC COLUMNS

Introduction to Pharmaceutical Chemical Analysis

Chromatography. Gas Chromatography


Principles of Gas- Chromatography (GC)

Investigation of Petasis and Ugi Reactions in Series in an Automated Microreactor System

Simulated Moving Bed Chromatography (SMB) F DIN EN ISO 9001 E D Q U A L T N E T Y M A N E M A G SMB. The next step in chromatography

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

A C18 Silica Column With Exceptional Temperature and ph Stability

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

Optimized Flash Chromatography Purification: From TLC to large scale in three steps

Study of a Simulated Moving Bed (SMB) Preparative Liquid Chromatographic Appliance and Technology

New 3-Micron Polysaccharide-based Chiral Columns for Fast HPLC and SFC Geoffrey B. Cox, Norbert M. Maier Chiral Technologies, Inc

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

The enrichment of loxoprofen enantiomer by 6-columns simulated moving bed chromatography

CHROMATOGRAPHIC SEPARATION TECHNIQUES SUPERCRITICAL FLUID CHROMATOGRAPHY

Ester Synthesis And Analysis: Aspirin and Oil of Wintergreen. Vanessa Jones November 19, 2015 Thursday 8:30 Lab Section Lab Partner: Melissa Blanco

Automating Gradient Method Development in Flash Chromatography For Greater Productivity and Minimizing Solvent Use

ZirChrom Columns and the Metalox 200-C High Temperature Liquid Chromatography Column Heater - The Future of HPLC has Arrived.

Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Chemistry Instrumental Analysis Lecture 31. Chem 4631

Introduction to Chromatographic Separations

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Experiment UPHPLC: Separation and Quantification of Components in Diet Soft Drinks

IDENTIFICATION AND DETERMINATION OF HYDROQUINONE IN COSMETIC PRODUCTS 2 14/11/17 ACM 003 BY TLC AND HPLC

Chiral separations efficient, fast and productive

LEARNING OBJECTIVES CHEM 212: SEPARATION SCIENCE CHROMATOGRAPHY UNIT. Thomas Wenzel, Bates College. In-class Problem Set Extraction.

Open Column Chromatography, GC, TLC, and HPLC

2501 High Performance Liquid Chromatography

EVALUATION OF EXPERIMENTAL PARAMETER INFLUENCE ON HPLC SEPARATION OF SOME AMINES AND

Separation Methods Based on Distributions in Discrete Stages (02/04/15)

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

penta-hilic UHPLC COLUMNS

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Application. Gas Chromatography February Introduction

viridis Columns Bringing a New Dimension to SFC Combining state-of-the-art media manufacturing with industry leading column technology, Viridis

Trefoil Columns [ CARE AND USE MANUAL ] CONTENTS I. INTRODUCTION I. INTRODUCTION GETTING STARTED III. COLUMN USE IV.

LUMEFANTRINUM LUMEFANTRINE

Chromatographic Methods: Basics, Advanced HPLC Methods

Chiral Separation Techniques: A Practical Approach

A biphasic oxidation of alcohols to aldehydes and ketones using a simplified packed-bed microreactor

If you like us, please share us on social media. The latest UCD Hyperlibrary newsletter is now complete, check it out.

High Performance Liquid Chromatography

High Performance Liquid Chromatography

Determination of Pesticides in Aqueous Samples by On-Line Coupling Solid-Phase Extraction to Gas Chromatography with At-Column Concentrating Interface

CYCLOSERINE Final text for addition to The International Pharmacopoeia. (November 2008) CYCLOSERINUM CYCLOSERINE

ChromegaChiral TM Columns. The right choice for chiral purification

Method Development for Chiral LC/MS/MS Analysis of Acidic Stereoisomeric Pharmaceutical Compounds with Polysaccharide-based Stationary Phases

Transcription:

SEPARATION OF FLURBIPROFEN AND IBUPROFEN ENANTIOMERS ON A CHIRAL STATIONARY PHASE USING SUPERCRITICAL FLUIDS John Whelan, Mickey Martin, Raquel Carvallo, Sermin G. Sunol, Aydin K. Sunol* Chemical Engineering Department, University of South Florida, Tampa FL 33620 USA e-mail: sunol@eng.usf.edu fax: (813) 974 3651 A Supercritical Chromatography System is developed for separation of chiral isomers. The systems studied include ibuprofen as well as of flurbiprofen, a non-steroidal Anti-Inflammatory Drug (NSAID). S-isomer is used for various pain treatments such as Rheumatoid Arthritis and Osteoarthritis while R-isomer is used in cancer research. The column utilized for this method was a Pirkle brush-type Whelk-O 1 analytical column packed with 10µm particles. Carbon dioxide is used as a mobile phase with several different organics employed as a modifier. The effect of pressure, temperature, flow rate and modifier content on the separation factor, peak resolution and retention time is analyzed. Elution chromatography is used to abstract isotherm information for scale up and synthesis of operating policies for fractionation through simulated moving bed. INTRODUCTION Supercritical fluid chromatography (SFC) is most often used with carbon dioxide as a mobile phase and an organic modifier such as some type of organic alcohol (1). It has some significant advantages over standard HPLC methods such as less pressure drop across the columns, faster column equilibration, faster method development, higher efficiency separations and significantly less generation of hazardous waste. Some of the main advantages for preparatory chromatography include solvent waste reduction, facilitated product recovery, lower solvent cost and the possibility for solvent recycling (2). Two chiral drug systems were studied in this body of work. The first is Ibuprofen, a nonsterodial anti-inflammatory drug which as of late is being studied for its chiralrelated properties. There are several reasons to separate the two isomers (R and S). The first is that the S-isomer works more than twice as fast as the racemic mixture. This allows for therapeutic applications that require more speed than would be impossible with the mixture. The other reason is that the effective dosage of the drug is decreased. The second system is flurbiprofen, a non-steroidal Anti-Inflammatory Drug (NSAID). S- isomer is used for various pain treatments such as Rheumatoid Arthritis and Osteoarthritis while R-isomer is used in cancer research. Chromatographic techniques and technologies are currently in wide spread use for analytical and preparative purposes in industry (3). In many cases, the only method to obtain pure components is chromatography; this is especially true for isomers and

enantiomers. Chiral separations are relatively new capabilities that are based on the elution chromatography model, but special stationary phases have to be utilized to achieve separation. The most often produced stationary phases are made of polysaccharides or proteins, but these are much more delicate than the standard C 18 and C 8 columns and tend to have a much shorter operational life expectancy (4). The reason for this delicate nature is that chiral stationary phases are not covalently bonded to the silica support like the C 18 and C 8 columns. The effects of pressure on separation factors and peak resolutions were found at 100bar, 150bar, 200bar, and sometimes at 250bar. The effects of temperature on separation factors and peak resolutions were found at 35 C, 40 C and 50 C. Some experiments were done to find the effect of flow rate on peak resolutions and separation factors. Several different modifiers were used in the experiments to find what organic or combination of organics produced the best results. The modifiers that were used in the experiments were around 10 volume percent in carbon dioxide. The specific organics used as modifiers were isopropanol, ethanol, methanol, ethyl acetate, acetone, and combinations of equal amounts by volume of isopropanol and ethanol, isopropanol and methanol, ethanol and methanol, ethyl acetate and isopropanol, as well as a 9.4% and 0.6% combination of hexane and isopropanol. The objective of the research is to find optimal conditions for scale-up to preparative supercritical fluid chromatography applications. EXPERIMENTAL SFC SET-UP The SFC apparatus consisted of a 100mL Isco syringe pump used for the delivery of carbon dioxide. A Waters 600E HPLC pump was used for delivery of the various modifiers. A Hewlett Packard 1050 UV/vis detector, fitted with a 400bar flow cell, was used for detection. Sample delivery was accomplished via a Perkin-Elmer ISS-900 auto sampler equipped with a Rheodyne valve with a pressure rating up to 350bar. A circulating bath column heater was designed and built. The outlet line of the syringe pump was run down and through the back part of the column heater to assure that the carbon dioxide was also at the desired temperature before sample introduction. The Regis (R,R) Whelk-O 1 chiral chromatographic column used in this work was packed with 10µm particles. The 10µm particles were used because the larger packing is more durable and better encapsulated than the 5µm particles. The Whelk-O 1 column is pressure rated to 7000 psi, and is stable up to 60 C. An automated backpressure regulator made by Thar Designs, BPR-A-200B, was used to keep the pressure at desired levels. A recirculating chiller, Lauda E100 econoline RE120, supplied the coolant to the Isco syringe pump cooling jacket. A BNC-1120 connector block along with a 6240E data acquisition card was purchased from National Instruments for use with the Labview 6.1 data acquisition software. The data was analyzed by employing the Microcal Origin 6.0 graphical software.

RESULTS AND DISCUSION Effect of temperature and pressure using isopropanol as a modifier Higher flow rates and higher temperatures produced lower separation factors, a trend that was expected and displayed in Figure 1. Higher pressures produced slightly higher separation factors and slightly decreasing peak resolution numbers. Peak Resolution 1.4 1.2 1 0.8 0.6 0.4 0.2 0 0 50 100 150 200 250 300 Pressure (bar) Figure 1. Effect of Pressure on Peak Resolution at 35 o C Effect of various solvents as modifier Several different modifiers were used in the experiments to find what organic or combination of organics produced the best results. The modifiers that were used in the experiments were around 10 volume percent in carbon dioxide. The specific organics used as modifiers were isopropanol, ethanol, methanol, ethyl acetate, acetone, and combinations of equal amounts by volume of isopropanol and ethanol, isopropanol and methanol, ethanol and methanol, ethyl acetate and isopropanol, as well as a 9.4% and 0.6% combination of hexane and isopropanol. The ones that worked best were isopropanol and ethanol. Combining the two in a 50:50 mixture and 10% v/v gave the best results. Good results were also achieved using a combination of methanol and ethanol 50:50, and isopropanol and methanol 50:50. Figure 2 depicts the various data points obtained from the before mentioned results. Ethyl Acetate was used by itself at 10% v/v and no separation was observed. Methanol worked and the retention times were noticeably reduced, but the separation and the resolution were drastically worse. A 50:50 combination of methanol and isopropanol was used which combined the lower retention times of methanol and the better separations of isopropanol. In comparison, the separation and resolution was more desirable than just isopropanol. Since a mobile phase of 94% hexane and 6% isopropanol had worked so well for separating ibuprofen in HPLC, that mixture was combined with SFC. Carbon dioxide was combined with a 9.4% hexane and 0.6% isopropanol (total v/v 10%) mixture and used as a modifier. The hexane modifier did not work well at all at supercritical conditions so no comparison of results in various publications and in this work was possible with isopropanol and

combinations of isopropanol and ethanol. This figure includes results from sub-critical experiments. The effect of pressure and modifier is summarized in Table 1. 1.6 1.4 1.2 Peak Resolution 1 0.8 0.6 0.4 Isopropanol Modifier Ethanol/Methanol Modifier Isopropanol/Ethanol 0.2 0 0 50 100 150 200 250 300 Pressure(bar) Figure 2. Effect of Pressure on Peak Resolution, Various Modifiers Effect of Modifier Content v/v % on Peak Resolution and Separation Factor in SFC The results mirrored reported research with other columns: as the v/v% of modifier is lowered the greater the separation [5]. The peak resolution was also better as the modifier percentage was decreased. However, as the modifier percent was decreased the retention time of the sample increased greatly. In this work, more time was spent on shorter analysis times because rather long retention times are undesirable if the conditions are to be transferred to a preparatory chromatography process. Some of the key results such as theoretical plate numbers with the 10% Isopropanol as the modifier is given on Table 2. Other Chiral Systems and Path to Scale Up The second system studied is flurbiprofen, a non-steroidal Anti-Inflammatory Drug (NSAID). S-isomer is used for various pain treatments such as Rheumatoid Arthritis and Osteoarthritis while R-isomer is used in cancer research. The separation factors and Peak resolutions were satisfactory but not high as ibuprofen case. Furthermore, Elution chromatography is used to abstract isotherm information for scale up and synthesis of operating policies for fractionation through simulated moving bed.

Table 1. Peak Resolutions and Separation Factors at 35 o C Separation Factor Modifier Comparison @ 35 C vs. Pressure (bar) P (bar) IPA IPA/Ethanol Ethanol/Methanol 100 1.0953 1.1919 1.1012 150 1.0878 1.2232 1.2130 200 1.0803 1.2437 1.2441 250 1.0769 1.2947 1.2011 Peak Resolution Modifier Comparison @ 35 C vs. Pressure (bar) P (bar) IPA Ethanol/methanol IPA/ ethanol 100 1.2435 1.3126 1.3929 150 1.1196 0.8798 1.1087 200 0.9637 0.8078 1.0034 250 0.8986 0.7362 1.0186 Table 2. Theoretical Plate Numbers for Isopropanol Experiments 10% IPA Theoretical Plate Numbers P 35 C 50 C (bar) N H n H 100 2988.295 0.008366 2720.357 0.00919 150 3046.473 0.008206 2670.208 0.009363 200 2708.176 0.009231 2311.156 0.010817 250 2806.794 0.008907 3087.301 0.008098 CONCLUSIONS A method was developed to separate the enantiomers of ibuprofen and flurbiprofen on a Whelk-O 1 stationary phase and using various modifiers in Supercritical Fluid Chromatography. Several modifiers were used for analysis. One combination of 5% isopropanol and 5% ethanol modifier (total v/v 10%) achieved the best peak resolution of 1.39 at 35 C. Isopropanol came close at 1.24 at 35 C and 100bar. 5% ethanol and 5% methanol produced a peak resolution of 1.31 at 35 C and 100bar. When the 9.4% hexane and 0.6% isopropanol modifier were used in the supercritical region it produced results that were not very good and not as good as just using isopropanol. The results were found to be similar to previous experiments done in 1994 by Blum, Lynam and Nicolas, except that the columns that are produced today have advanced in several areas (2). The improvements of the column led to improved results for the separation.

REFERENCES 1. Poole, C.F., Poole, S. K., Chromatography today. 5th ed. 1991, Amsterdam: Elsevier Science. 2. Blum, A.M., Lynam, K. G.,Nicolas, N. C., Use of a new Pirkle-type Chiral stationary phase in Analytical and Preparative Subcritical Fluid Chromatography of Pharmaceutical Compounds. Chirality 6, 1994. 6(4): p. 302-313. 3. Kniep, H., Seidel-Morgenstern, A., SMB Guide 1.1. 1997. 4. Pirkle, W.H., Welch, C. J., Burke, J. A., Lamm, B., Design, syntesis and evalutation of an improved enantioselective Naproxen selector. Journal of Chromatography, 1992. 57: p. 3854-3860. 5. Johannsen, M., Separation of enantiomers of ibuprofen on chiral stationary phases by packed column supercritical fluid chromatography. Journal of Chromatography A, 2001. 937: p. 135-138.