Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Similar documents
Improved GC/MS Analysis of Tomato Pesticides with Agilent Deactivated Silica Tubing

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

High Efficiency, Broad Scope Screening of Pesticides Using Gas Chromatography High Resolution Orbitrap Mass Spectrometry

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Screening for Hundreds of Pesticide Residues Using a GC/Q-TOF with an Exact Mass Pesticide Database in Food

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Author. Abstract. Introduction

An Optimal Method for the Analysis of Pesticides in a Variety of Matrices

Analysis of Pesticides and Contaminants using HS-SPME-GC/MS and DRS Detection. Blair Surridge, Tyler Dyer, Rene Bilideau, Paloma Bertolin Presoti

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

GC/Q-TOF MS Surveillance of Pesticides in Food

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Highly Sensitive and Rugged GC/MS/MS Tool

LC/MS/MS Analysis of Pesticide Residues in Apples Using Agilent Chem Elut Cartridges

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Challenging Pesticide Analysis Using an Agilent J&W DB-35ms Ultra Inert GC Column

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Sensitive Detection of Pyrethroids in Surface Water and Sediment

Application Note. Abstract. Authors. Introduction. Food Safety

Rugged GC/MS/MS Pesticide Residue Analysis Fulfilling the USDA Pesticide Data Program (PDP) Requirements

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Analysis of Pesticides in Citrus Oil Using a Shimadzu GCMS-TQ8030. No. GCMS No. SSI-GCMS-1404

The Analysis of Organophosphate Pesticides by LC/MS Application

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

ENVIRONMENTAL analysis

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Multi-residue GC-MS analysis. Richard Fussell CSL York, UK

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Identification and Quantitation of Pesticides in Chamomile and Ginger Extracts Using an Agilent 6460 Triple Quadrupole LC/MS system with Triggered MRM

Simultaneous analysis of pesticides by GC/MS

Trim Time Not Columns Multiresidue Pesticides Analysis with an Intuvo GC/TQ

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Simultaneous determination of 74 kinds of pesticides in Chinese herbal medicine listed in Chinese pharmacopoeia (2015) by GPC-GC-MS/MS

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

AppNote 4/2004. KEYWORDS Multi-residue method, pesticides, vegetables, fruits, green tea, Stir bar sorptive extraction (SBSE), GC-MS.

Application Note. Authors. Abstract. Xin Sun Shandong Agriculture University, Taian, , China

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

QuEChERS Pesticide Analysis for Fresh Produce*

Ultra-fast screening of pesticides in foods and agricultural products with desorption corona beam ionization (DCBI) tandem mass spectrometry

Application Note. Abstract. Authors. Introduction

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Three-Fold Increase in Productivity for Pesticide Residue Analysis in Baby Food Using Fast Triple Quadrupole GC-MS/MS

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Multi-residue analysis of pesticides by GC-HRMS

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

U.S. EPA Method 8270 for multicomponent analyte determination

Institute of Plant Protection Belarusian State University

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

LC/MS/MS of Fungicides and Metabolites in Orange Juice with Agilent Bond Elut Plexa and Poroshell 120

GC/MS Application Note

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Strategies for Analysis of Pyrethroid Insecticides in Complex Matrices

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Analysis of pesticide residues using various GC/MS systems. Radim Štěpán Czech Agriculture and Food Inspection Authority Prague, Czech Republic

LC/MS/MS of Fungicides and Metabolites in Apple Juice with Agilent Bond Elut Plexa and Poroshell 120

Determination of pesticides in baby food by UHPLC/MS/MS using the Agilent 1290 Infinity LC system and the Agilent 6460 triple quadrupole LC/MS

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Organophosphorus Pesticides in Apple Matrix by GC/MS/FPD using an Agilent J&W DB-35ms Ultra Inert GC Column

Application. Gas Chromatography February Introduction

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

The Pulsed Flame Photometric Detector for analysis of phosphorus pesticides in wastewater and food commodities

IMPROVING PESTICIDE RESIDUE DETECTION PROTOCOL FOR FRUIT

MultiVap8: Automated concentration system

PYRETHROID ANALYSIS. MARKET FOCUSED - Environmental TECHNICAL ARTICLE

Analysis of BTEX in Natural Water with SPME

STANDARD OPERATING PROCEDURES

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Sulfotepp impurities in Chlorpyrifos EC formulations

Application of supercritical fluid extraction in multi-residue pesticide analysis of plant matrices

Validation of Commercially Available ELISA Assay Kits

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Transcription:

Pesticides Analysis Using the Agilent 5977A Series GC/MSD Application Note Food Testing and Agriculture Authors Jia-jia Wu and Wen-wen Wang Agilent Technologies Co. Ltd (China) Beijing, People s Republic of China Abstract A method has been developed using the Agilent 5977A Series GC/MSD that can deliver excellent linearity (R 2 0.991), reproducibility ( 11% in solvent and apple matrix), and sensitivity, most minimum detection limits (MDLs) ~ 6.7 ppb of quantitation for 42 pesticides commonly analyzed in China. Introduction With the increasing globalization of the food industry, there is greater scrutiny on food safety, resulting in major changes in the number of pesticides that are being regulated and monitored, as well as the allowable levels of those pesticides in food. There are more than 1,000 registered pesticides worldwide, and many countries employ strict regulations on pesticide residues in food and animal feed. Currently, the lowest maximum residue level (MRL) for most pesticides in China is 0.01 mg/kg, parts per million (ppm), and can range as high as 5 mg/kg, depending on the food matrix being tested. This application note demonstrates the ability of the 5977A Series GC/MSD to provide sensitive, accurate, and reproducible analysis for 42 commonly tested pesticides in China, with MDLs well within the required MRLs.

Experimental Standards and Reagents Standards for 42 pesticides were commercially obtained (Table 1). The working calibration and matrix spiking standards were prepared in hexane using the Agilent 7696A Sample Prep WorkBench. Instruments The study was performed on an Agilent 7890B gas chromatograph equipped with a Split/Splitless Inlet and coupled to an Agilent 5977A Series GC/MSD, using SIM and Electron Ionization (EI) acquisition modes. Table 2 lists the instrument conditions. Table 1. 42 Pesticide Standards No. Classification CAS Name 1 Conazole 43121-43-3 Triadimefon 2 Dicarboximide 32809-16-8 Procymidone 3 Dicarboximide 50471-44-8 Vinclozolin 4 Dicarboximide 36734-19-7 Iprodione 5 Organochlorine 319-84-6 BHC-alpha 6 Organochlorine 319-85-7 BHC-beta 7 Organochlorine 58-89-9 BHC-gamma 8 Organochlorine 319-86-8 BHC-delta 9 Organochlorine 72-55-9 DDE-p,p' 10 Organochlorine 72-54-8 DDD-p,p' 11 Organochlorine 789-02-6 DDT-o,p' 12 Organochlorine 50-29-3 DDT-p,p' 13 Organochlorine 82-68-8 Pentachloronitrobenzene 14 Organophosphorus 62-73-7 Dichlorvos 15 Organophosphorus 298-02-2 Phorate 16 Organophosphorus 122-14-5 Fenitrothion 17 Organophosphorus 55-38-9 Fenthion 18 Organophosphorus 2921-88-2 Chlorpyrifos 19 Organophosphorus 56-38-2 Parathion 20 Organophosphorus 24353-61-5 Isocarbophos 21 Organophosphorus 732-11-6 Phosmet No. Classification CAS Name 22 Organophosphorus 2310-17-0 Phosalone 23 Organophosphorus 60-51-5 Dimethoate 24 Organophosphorus 333-41-5 Diazinon 25 Organophosphorus 298-00-0 Parathion-methyl 26 Organophosphorus 121-75-5 Malathion 27 Organophosphorus 41198-08-7 Profenofos 28 Organophosphorus 24017-47-8 Triazophos 29 Organophosphorus 99675-03-3 Isofenphos-methyl 30 Pyrazole 120068-37-3 Fipronil 31 Pyrethroid 39515-41-8 Fenpropathrin 32 Pyrethroid 91465-08-6 Cyhalothrin 33 Pyrethroid 68359-37-5 Cyfluthrin 34 Pyrethroid 70124-77-5 Flucythrinate 35 Pyrethroid 102851-06-9 Fluvalinate-tau 36 Pyrethroid 82657-04-3 Bifenthrin 37 Pyrethroid 51877-74-8 Permethrin 38 Pyrethroid 52315-07-8 Cypermethrin 39 Pyrethroid 51630-58-1 Fenvalerate 40 Pyrethroid 52918-63-5 Deltamethrin 41 Pyrimidine 53112-28-0 Pyrimethanil 42 Substituted Benzene 1897-45-6 Chlorothalonil Table 2. Agilent 7890B GC and Agilent 5977A Series GC/MSD Conditions GC run conditions Analytical column HP-5ms Ultra Inert 30 m 250 µm, 0.25 µm (p/n 19091S-433UI) Injection volume 1 µl Injection mode Splitless Inlet temperature 280 C Liner UI, splitless, single taper, glass wool (p/n 5190-2293) Plated seal kit Gold Seal, Ultra Inert, with washer (p/n 5190-6144) Carrier gas Helium, constant flow, 1 ml/min Oven program 60 C for 1 minute, then 40 C/min to 170 C, then 10 C/min to 310 C, then hold for 2 minutes Transfer line temperature 280 C MS conditions Solvent delay 3.5 minutes Acquisition mode SIM Tune Etune.u Gain factor 5.00 Source temperature 250 C Quadrupole temperature 150 C TID On 2

Sample Preparation A 20 g apple sample was homogenized with 40 ml of acetonitrile and mixed vigorously for 1 minute in a vortex mixer. A 5-g amount of NaCl was added, and the sample was mixed on the vortex mixer for another minute. The sample was then centrifuged at 4,200 rpm for 5 minutes. A 20-mL aliquot of the supernatant, equivalent to 10 g of sample, was removed and concentrated to approximately 1 ml. The extract was then cleaned up using a Bond Elut carbon/nh 2 column (500 mg/500 mg, 6 ml, p/n 12252202). The effluent was collected and evaporated to near dryness with a gentle stream of nitrogen. The residue was dissolved in hexane to a volume of 10 ml. One milliliter of the resulting blank extract, corresponding to 1 g of sample, was used to prepare a 50 ng/ml parts per billion (ppb) fortified matrix-matching standard (pesticide standards added to blank extracts). Acquisition Parameters Table 3 shows the ions used for acquisition. Table 3. Retention Times and Acquisition Parameters Target compound RT Q 0 Q 1 Q 2 Q 3 Dichlorvos 4.57 109 185 79 187 Phorate 7.33 75 121 260 97 BHC-alpha 7.47 181 219 183 217 Dimethoate 7.61 87 93 125 143 BHC-beta 7.84 217 181 183 219 BHC-gamma 7.96 217 183 219 111 Pentachloronitrobenzene 8.04 237 249 295 214 Pyrimethanil 8.10 198 199 200 77 Diazinon 8.11 179 137 152 199 BHC-delta 8.31 181 219 183 217 Chlorothalonil 8.39 266 264 268 270 Vinclozolin 8.91 212 285 198 187 Parathion-methyl 8.94 263 109 125 79 Fenitrothion 9.38 277 125 109 260 Malathion 9.53 173 127 125 93 Fenthion 9.71 278 125 109 169 Chlorpyrifos 9.75 197 199 314 97 Parathion 9.76 291 109 97 139 Triadimefon 9.79 57 208 85 210 Isocarbophos 9.87 136 121 120 110 Isofenphos-methyl 10.19 199 58 121 231 Target compound RT Q 0 Q 1 Q 2 Q 3 Fipronil 10.43 367 369 213 351 Procymidone 10.62 96 283 285 67 Profenofos 11.31 339 139 206 208 DDE-p,p' 11.39 246 318 316 248 DDD-p,p' 12.13 235 237 165 236 DDT-o,p' 12.20 235 237 165 236 Triazophos 12.41 161 162 172 77 DDT-p,p' 12.79 235 237 165 236 Iprodione 13.45 314 187 189 244 Phosmet 13.65 160 161 77 93 Bifenthrin 13.69 181 165 166 182 Fenpropathrin 13.81 181 97 125 265 Phosalone 14.33 182 121 184 367 Cyhalothrin 14.63 181 197 208 209 Permethrin 15.40 183 163 165 184 Cyfluthrin 16.12 163 206 165 227 Cypermethrin 16.24 181 163 165 77 Flucythrinate 16.44 199 157 44 207 Fenvalerate 17.14 167 125 181 152 Fluvalinate-tau 17.32 250 252 209 181 Deltamethrin 17.85 181 253 251 255 RT retention time in minutes Q 0 quantifier ion Q 1, Q 2, Q 3 qualifier ions 3

Results and Discussion Linearity Calibration curves were constructed from 20 to 200 ng/ml (ppb) in hexane for most target compounds. All 42 compounds had calibration coefficient values 0.991 (Table 4). Table 4. Calibration Coefficients (R 2 ) for the 42 Pesticides Target compound R 2 Target compound R 2 Dichlorvos 0.998 Fipronil 0.997 Phorate 0.998 Procymidone 0.998 BHC-alpha 0.998 Profenofos 0.997 Dimethoate 0.995 DDE-p,p' 0.998 BHC-beta 0.995 DDD-P,P' 0.998 BHC-gamma 0.998 DDT-o,p' 0.998 Pentachloronitrobenzene 0.997 Triazophos 0.995 Pyrimethanil 0.998 DDT-p, p' 0.996 Diazinon 0.998 Iprodione 0.995 BHC-delta 0.998 Phosmet 0.995 Chlorothalonil 0.999 Bifenthrin 0.996 Vinclozolin 0.999 Fenpropathrin 0.996 Parathion-methyl 0.993 Phosalone 0.995 Fenitrothion 0.991 Cyhalothrin 0.995 Malathion 0.998 Permethrin 0.992 Fenthion 0.998 Cyfluthrin 0.995 Chlorpyrifos 0.998 Cypermethrin 0.995 Parathion 0.989 Flucythrinate 0.994 Triadimefon 0.994 Fenvalerate 0.996 Isocarbophos 0.991 Fluvalinate-tau 0.995 Isofenphos-methyl 0.998 Deltamethrin 0.992 Reproducibility and Minimum Detection Limits in Solvent Table 5 illustrates the good reproducibility obtainable with the 5977A Series GC/MSD in solvent. The majority of the relative standard deviation values (s) across all 42 pesticide compounds were ~ 5%, with none exceeding 11%, across eight injections. The corresponding MDLs, calculated as the product of the multiplied by the Student t test value, ranged from 3 to 14.3 ppb, with the majority being ~ 6.7 ppb. Table 5. Compound Reproducibility s and Calculated MDLs for a 50 ppb Standards Mix Sample in Solvent* MDL (ppb) Target compound MDL (ppb) Dichlorvos 5 6.4 Fipronil 6 8.3 Phorate 5 6.3 Procymidone 4 4.7 BHC-alpha 3 4.3 Profenofos 10 12.0 Dimethoate 7 8.3 DDE-p,p' 3 4.5 BHC-beta 3 3.6 DDD-p,p' 5 6.7 BHC-gamma 2 3.0 DDT-o,p' 3 3.6 Pentachloronitrobenzene 3 3.8 Triazophos 8 9.9 Pyrimethanil 4 5.0 DDT-p,p' 3 4.0 Diazinon 4 4.7 Iprodione 10 12.2 BHC-delta 3 3.5 Phosmet 8 9.7 Chlorothalonil 3 4.1 Bifenthrin 6 8.2 Vinclozolin 3 4.2 Fenpropathrin 6 8.6 Parathion-methyl 5 5.9 Phosalone 7 9.1 Fenitrothion 5 6.6 Cyhalothrin 8 10.6 Malathion 5 6.2 Permethrin 8 9.6 Fenthion 5 5.9 Cyfluthrin 9 11.3 Chlorpyrifos 4 5.7 Cypermethrin 9 11.0 Parathion 5 5.6 Flucythrinate 11 13.1 Triadimefon 4 5.5 Fenvalerate 9 12.0 Isocarbophos 8 9.1 Fluvalinate-tau 11 14.3 Isofenphos-methyl 5 6.7 Deltamethrin 9 11.2 *Eight consecutive injections were used to calculate the s. Reproducibility and Minimum Detection Limits in Apple Matrix Reproducibility was quite good in apple matrix, with many of the s being lower than those calculated for analysis in solvent (Table 6 versus Table 5). All but two of the MDLs were ~ 7.5 ppb, none was higher than 10.1 ppb, and all were well below the Chinese MRL requirements (Table 6). 4

Table 6. Reproducibility s and Calculated MDLs for a 50 ppb Standards Mix Sample Spiked into Apple* Target compound Calculated MDL (ppb) Dichlorvos 2 2.6 200 Phorate 3 4.2 10 BHC-alpha 3 4.1 50** Dimethoate 3 3.7 1,000 BHC-beta 4 5.9 50** BHC-gamma 3 5.2 50** Required MRL (ppb) Pentachloronitrobenzene 3 4.0 20 (watermelon) Pyrimethanil 3 3.9 1,000 (pear) Diazinon 3 4.1 100 (cereal) BHC-delta 3 4.3 50** Chlorothalonil 2 2.9 1,000 Vinclozolin 3 4.0 1,000 (cucumber) Parathion-methyl 3 4.1 10 Fenitrothion 2 3.2 500 Malathion 3 3.7 2,000 Fenthion 3 5.0 50 Chlorpyrifos 3 4.5 1,000 Parathion 3 4.5 10 Triadimefon 3 3.7 1,000 Isocarbophos 3 6.0 10 Isofenphos-methyl 3 3.7 10 Target compound Calculated MDL (ppb) Required MRL (ppb) Fipronil 2 3.4 20 (cereal) Procymidone 3 4.5 5,000 (grape) Profenofos 4 5.8 50 DDE-p,p' 3 3.5 50*** DDD-p,p' 1 1.8 50*** DDT-o,p' 2 2.4 50*** Triazophos 2 3.7 200 DDT-p,p' 2 2.9 50*** Iprodione 3 5.3 5,000 Phosmet 3 4.1 5,000 (orange) Bifenthrin 3 5.0 500 Fenpropathrin 3 3.8 5,000 Phosalone 2 2.6 1,000 (spinach) Cyhalothrin 4 5.1 200 Permethrin 4 5.5 2000 Cyfluthrin 7 10.1 500 Cypermethrin 5 7.3 2,000 Flucythrinate 5 7.5 500 Fenvalerate 4 5.7 1,000 Fluvalinate-tau 3 4.1 500 (spinach) Deltamethrin 2 3.1 100 * Eight consecutive injections were used to calculate the s. Chinese regulation GB 2763-2012, MRL: Maximum Residue Level ** Four isomers in total *** Four DDT, DDE, DDD in total Conclusion The Agilent 5977A Series GC/MSD can deliver sensitive, accurate and reproducible results for the analysis for 42 commonly tested pesticides in China, including MDLs in some food matrices that can be well within the required limits of Chinese regulation GB 2763-2012. The use of automation such as the Agilent 7696A Sample Prep WorkBench to prepare standards also results in more linear calibration curves. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2013 Printed in the USA April 12, 2013 5991-2212EN