PRODUCT DATA. Sound Intensity Calibrator Type 3541-A. Uses and Features

Similar documents
Product Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541

Product Data. Brüel & Kjær B. Sound Power Source Type Introduction

WEATHER STATION FIELD GUIDE

CERTIFICATE OF CALIBRATION

CERTIFICATE OF CALIBRATION

Report on Key Comparison COOMET.AUV.A-K5: Pressure calibration of laboratory standard microphones in the frequency range 2 Hz to 10 khz

Consultancy Report Ref: 9383-R01

Uncertainties associated with the use of a sound level meter

Provläsningsexemplar / Preview TECHNICAL REPORT INTERNATIONAL SPECIAL COMMITTEE ON RADIO INTERFERENCE

Static pressure and temperature coefficients of working standard microphones

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

THE ACOUSTIC IMPEDANCE MEASUREMNET SYSTEM USING TWO MICROPHONES

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Microphone HALT Report

Latest generation testo air velocity & IAQ probes with cable handle. Instruction manual

ECMA Measurement of airborne noise emitted and structure-borne vibration induced by small air-moving devices Part 1: Airborne noise measurement

Certificate of Calibration Sound Level Calibrator

Accurate Determination of Loudspeaker Parameters using Audio Analyzer Type 2012 and Laser Velocity Transducer Type 3544

Testing Procedure: AS ISO

MEASUREMENTS OF SOUND ABSORPTION TIMBERCRETE

Tel: Fax: PRODUCT INFORMATION

A comparison of two different sound intensity measurement principles a)

CERTIFICATE OF ACCREDITATION

Environmental Effects and Control of Noise Lecture 2

Determination of sound absorption coefficient in impedance tubes according to ISO (2 appendices)

EMC Standards. Samuel Burri 2013, EPFL

DS 300. Electronic Pressure Switch. with IO-Link interface. Stainless Steel Sensor. accuracy according to IEC 60770: 0.35 % FSO.

792A AC/DC Transfer Standard

792A AC/DC Transfer Standard

5730A High Performance Multifunction Calibrator. Extended specifications

FINAL REPORT ON KEY COMPARISON APMP.AUV.A-K3

MAC Aura XB Acoustic Test Report

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HC5FW 500-S

Measuring HRTFs of Brüel & Kjær Type 4128-C, G.R.A.S. KEMAR Type 45BM, and Head Acoustics HMS II.3 Head and Torso Simulators

Method of Measuring Machinery Sound Within an Equipment Space

The Ordering Code for various standard model Recorders with an AC supply and without any additional options are as follows:

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HC5FW 200-S/SP1

RIVERBANK ACOUSTICAL LABORATORIES

Sitting height... 6 Back tilt and back height... 6 Armrest... 7 Brakes... 7 Charging VELA Tango 300E... 7

Insertion Loss Analysis of the Acoustic Panels with Composite Construction

Uncertainties in Acoustics

ROTANTA 460/460 R Benchtop, Underbench and Floorstanding Centrifuges classic/cooled

A L I O N Technical Center. Test Report. Port Coquitlam, British Columbia, Canada

Review of Anemometer Calibration Standards

Page 1 of 10 pages Appendix, 5 pages Report no.: P Glostrup, October 12th 2012 Project: Claimant:

LABORATORY MEASUREMENT OF SOUND ABSORPTION OF SONOGLASS SPRAY-ON TREATMENT IN THREE CONFIGURATIONS

Measurement method for the proficiency testing program

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Analog - Digital multimeters

Acoustic Quantities. LMS Test.Lab. Rev 12A

ISO INTERNATIONAL STANDARD. Acoustics Acoustic insulation for pipes, valves and flanges

Simulation of Horn Driver Response by Direct Combination of Compression Driver Frequency Response and Horn FEA

A practical implementation of microphone free-field comparison calibration according to the standard IEC

T20WN. Data Sheet. Torque transducers. Special features. Installation example with bellows couplings. B en

The information included in the following report presents the results of sound pressure and sound power testing.

LINEAR SOURCES WITH NON-LINEAR DISTRIBUTION OF SOUND ENERGY. Miroslav Kučera, Richard Nový, Jiří Bašta

Acoustic test report AU01+N01

A L I O N Technical Center. Test Report

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

ISO 6395 INTERNATIONAL STANDARD. Earth-moving machinery Determination of sound power level Dynamic test conditions

Number Title Year Organization Page

DETERMINATION OF AIRBORNE SOUND POWER LEVELS EMITTED BY GEAR UNITS

ODEON APPLICATION NOTE Calibration of Impulse Response Measurements

Evaluation of standards for transmission loss tests

Measurement of Acoustic Properties of light weight concrete SL-Deck

Active Minimization of Acoustic Energy Density to Attenuate Radiated Noise from Enclosures

TEST CERTIFICATES OR REPORTS ISSUED BY VEIKI-VNL LTD.

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

IS INTERNATIONAL STANDARD. Acoustics - Determination of sound power levels of noise sources using sound intensity - Part 2:

Vernier Probe Inventory List

AUTOMOTIVE CURRENT TRANSDUCER OPEN LOOP TECHNOLOGY HC6H 400-S/SP1

CCQ-800 AutoFocimeter. CCQ-800 AutoFocimeter. Operating Instruction. R/No.: CCQ8.0ESM REV: 3.45

ISO 3741 INTERNATIONAL STANDARD

On determination of microphone response and other parameters by a hybrid experimental and numerical method

ED 701 General Industry Pressure Transmitter

VK2 volume flow controller Overview

REPORT ON THE DETERMINATION OF SOUND ABSORPTION COEFFICIENTS OF WOVEN IMAGE ECHO PANEL 7MM TESTED WITH A 20MM AIR GAP IN A REVERBERATION ROOM.

OIML R 50-3 RECOMMENDATION. Edition 2014 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

Standard ECMA-108 3rd Edition - December Measurement of High-frequency Noise emitted by Information Technology and Telecommunications Equipment

Technical Report. Project. The Laboratory Determination of The Random Incidence sound Absorption Coefficient of VertiQ Panels.

CHAPTER 5 SOUND LEVEL MEASUREMENT ON POWER TRANSFORMERS

OIML D 28 DOCUMENT. Edition 2004 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION. Conventional value of the result of weighing in air

Addendum to ARI Standard , Air Terminals September 2002

Aeroacoustic simulation of automotive ventilation outlets

ALUMINIUM CENTRIFUGAL FANS. 4 x M6 x OUTLET x M6 x 20

ALUMINIUM CENTRIFUGAL FANS x M6 x x M6 x 16

A Transfer Matrix Method for Estimating the Dispersion and Attenuation of Plane Waves in a Standing Wave Tube

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

Nonlinear Losses in Electro-acoustical Transducers Wolfgang Klippel, Daniel Knobloch

Sound power level measurement in diffuse field for not movable sources or emitting prominent discrete tones

Signal types. Signal characteristics: RMS, power, db Probability Density Function (PDF). Analogue-to-Digital Conversion (ADC).

ALUMINUM ELECTROLYTIC CAPACITORS Capacitor Life

Turbines and turbine sets Measurement of emitted airborne noise Engineering/survey method

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

Safety Precautions WARNING If critical situations that could lead to user s death or serious injury is assumed by mishandling of the product.

Masterpact. Installation recommendations

OIML R 141 RECOMMENDATION. Edition 2008 (E) ORGANISATION INTERNATIONALE INTERNATIONAL ORGANIZATION

1 Wind Turbine Acoustics. Wind turbines generate sound by both mechanical and aerodynamic

Operating limits and tolerances of platinum resistance thermometers per EN 60751: 2008

Part 2: Methods for special reverberation test rooms

Transcription:

PRODUCT DATA Sound Intensity Calibrator Type 3541-A Type 3541-A enables calibration of sound intensity measuring instruments by using a coupler designed especially for sound intensity calibrations. Intensity-probe microphones are positioned in the coupler, which in conjunction with a pistonphone simulates a plane sound wave propagating along the axis of the probe. The instrument s sensitivity to both sound intensity and particle velocity can then be calibrated. The pistonphone can also be used for sound pressure calibrations. In addition to calibration, Type 3541-A can be used to measure the pressure-residual intensity index of instruments used for measuring sound intensity at 250 Hz. Type 3541-A components in their case are shown on the right. The main components are the Intensity Coupler and Pistonphone. Other components are the Correction Barometer, Dummy Microphone (1/4 ), Microphone Adaptors (two), Microphone Adaptors for Pistonphone (1/2, 1/4 and 1/8 ). Calibration Charts are stored in the lid. Uses and Features Uses Sound intensity and particle velocity calibrations Sound pressure calibration Measurement of pressure-residual intensity index at 250 Hz Features Intensity coupler for simulating a plane sound wave in a free field (US. pat. No. 4715219) Pistonphone for sound intensity, particle velocity and sound pressure calibrations

Introduction Sound Intensity Calibrator Type 3541-A enables instruments that measure sound intensity to be calibrated against simulated sound intensity and particle velocity levels. Instruments and microphones cannot be considered fully calibrated if only the sound pressure sensitivities of the individual microphone channels are calibrated. Calibration with the simulated sound intensity and particle velocity levels of Type 3541-A ensure that these parameters can be measured correctly. Type 3541-A is intended for use with Brüel & Kjær Sound Intensity Probes with a Sound Intensity Microphone Pair. Other microphone pairs have much higher vent sensitivities and this restricts their use (see Microphones and Vent Sensitivity ). Fig. 1 Simplified crosssection of intensity coupler Dummy microphone (removable) Microphone B Pistonphone Microphone A Upper chamber Coupling element Lower chamber A simplified cross-section of the intensity coupler is shown in Fig. 1. The coupler consists of two chambers connected by a coupling element. When the pistonphone is attached to the coupler there is a phase difference between the sound pressures in the upper and lower chambers. The amplitude of the sound pressure is the same in both chambers, so a plane sound wave propagating in a free field is simulated. If one microphone is positioned in the upper chamber and the other in the lower chamber, the simulated sound wave can be used for calibrating the sensitivity of the measuring instrument to sound intensity and particle velocity. The coupler and pistonphone can also be used 871887/2 for sound pressure sensitivity calibration. For this, the microphones are both positioned in the upper chamber. Then they are exposed to exactly the same sound pressure (amplitude and phase). The supplied calibration chart states the levels that should be detected during calibration. The chart also gives information about corrections to the calibration levels for use when conditions are different from the stated reference conditions. A correction barometer determines correction terms to the sound pressure and particle velocity calibration levels due to changes in atmospheric pressure. The sound intensity calibration level is independent of any change in atmospheric pressure. Calibration Procedure Fig. 2 Arrangement for sound pressure calibration. It can also be used to measure residual intensity and pressureresidual intensity index at 250 Hz Full calibration of an intensity measuring instrument and its microphones includes: Sound pressure calibration of the individual microphone channels Sound intensity and particle velocity calibration Measurement of the pressure-residual intensity index spectrum of the system Sound Pressure Calibration Fig. 2 shows Pistonphone Type 4228 fitted to the coupler, with both microphones positioned in the upper chamber of the coupler. With this arrangement, the pistonphone produces the same sound pressure level at each microphone. The microphone channels are calibrated against this known sound pressure level. 2

Sound Intensity and Particle Velocity Calibrations Fig. 3 Simplified block diagram of an intensity measuring instrument. The signals from two pressure microphones, p A and p B, are used to determine the pressure at the midpoint on the probe axis, p, and the particle velocity along the probe axis, u r. Multiplying p and u r gives the intensity reading I r. r is the microphone spacing and the density of air p A Microphones p B + + + - Σ Σ p + p A B p = 2 ur = p A - p B ρδr dt X lr = p ur 871673/3 Fig. 3 is a simplified block diagram showing how sound intensity is measured. The particle velocity signal is obtained by integrating, with respect to time, the instantaneous difference in sound pressure between the two microphones. This signal is zero during a sound pressure calibration, so the correct functioning of the instrument is not confirmed. Fig. 4 Arrangement for sound intensity and particle velocity calibrations Fig. 4 shows the pistonphone fitted to the coupler, and the microphones positioned in different chambers of the coupler. With this arrangement, the coupler causes a phase change between the sound pressures at the microphones, corresponding to a nominal spacing of 50 mm with no reflections. The phase change between the sound pressures simulates the sound intensity and particle velocity levels so that the pressure-difference signal for the integrator is not zero. Only now is the correct functioning of the instrument confirmed. Pressure-Residual Intensity Index Measurement Fig. 5 Arrangement for measuring residual intensity and pressureresidual intensity index over a wide range of frequencies using Sound Intensity Calibrator Type 4297 Fig. 5 shows an arrangement for measuring pressure-residual intensity index over a wide range of frequencies using Sound Intensity Calibrator Type 4297. The intensity probe is fitted in the calibrator, which delivers broad-band pink noise, so the sound pressure spectrum measured in the coupler is constant (in octave bands) over a wide frequency range. Both microphones are exposed to the same sound pressure, so any intensity detected is residual intensity. The residual energy explains how small differences in the phase responses of the microphones and input channels result in the detection of residual intensity. Residual intensity is a parameter that should be taken into account when interpreting measured intensity data. It is worth noting that the residual intensity spectrum is not a fixed one; it is tied to (rises and falls with) the measured sound pressure level. It can be shown that, for a given measurement system and frequency, the difference between measured sound pressure level and detected residual intensity level will be a constant. This constant difference is called the pressure-residual intensity index. The pressure-residual intensity index spectrum can be measured with the arrangement shown in Fig. 5 by subtracting the detected intensity spectrum from the sound pressure spectrum. An example of this is shown in Fig. 6. Type 3541-A, by means of the Pistonphone Type 4228 and Coupler UA-0914, can be used to measure the pressure-residual intensity index in the 250 Hz 1/3-octave band. For measurements of pressure-residual intensity index over a wide frequency range, it is recommended to use Sound Intensity Calibrator Type 4297. 3

Fig. 6 Intensity and sound pressure levels measured with the arrangement shown in Fig. 5. The pressureresidual intensity index spectrum is characteristic of the measurement system and is obtained by subtracting these two spectra db 70 60 50 40 30 SPL in coupler Intensity level in coupler 20 Pressure-residual intensity index 10 0 63 125 250 500 1k 2k 4k Hz 871670/4 Residual Intensity Level If a pressure-residual intensity index spectrum is to be used to assess the accuracy of sound intensity measurements, the mean sound pressure spectrum in the field must also be measured. The residual intensity level is then quickly established by subtracting the pressure-residual intensity index spectrum from the measured mean sound pressure spectrum. An example of this is shown in Fig. 7. The residual intensity level is then compared to the measured sound intensity level. It can be shown that, for a certain frequency, the residual intensity level must be at least 7 db lower to ensure a measurement error of less than 1 db. The residual intensity level shown in Fig. 7 is dependent on the sound pressure level measured in the field and should not be confused with the intensity level, which is measured with the arrangement shown in Fig. 5. Fig. 7 Sound intensity and mean sound pressure spectra measured in the field with an instrument calibrated using Type 4297. The residual intensity spectrum is obtained by subtracting the pressure-residual intensity index spectrum in Fig. 6 from the measured SPL spectrum db 60 50 40 30 20 Measured SPL Residual intensity level Measured intensity level 10 63 125 250 500 1k 2k 4k Hz 871671/2 4

Microphones and Vent Sensitivity Coupler UA-0914 has been designed to work with Microphone Pair Types 4181 and 4197, which have an extremely low sensitivity to sound pressure at their pressure-equalization vents. When microphones are inserted into the coupler, their diaphragms are exposed to the sound pressure in the coupler but their pressure-equalization vents are not. Coupler UA-0914 cannot be used to measure the pressure-residual intensity index with conventional microphone pairs as they have vent sensitivities several orders of magnitude higher than that of Type 4181 and 4197. The coupler can, however, be used with conventional microphone pairs for calibration of sound pressure, sound intensity, and particle velocity. Residual Intensity Fig. 8 A sound wave is incident on a probe axis at 90. There is no flow of acoustic energy along the probe axis. The signals from the microphones are in phase and no intensity is detected Fig. 9 If a sound wave is incident at an angle other than 90, then acoustic energy flows along the probe axis. The microphone signals are out of phase and intensity is detected Signals are in phase Ideal probe with no phase mismatch Intensity Signals are not in phase Ideal probe with no phase mismatch No intensity is detected 870897/2 Fig. 10 In practice, if a sound wave is incident at 90, small differences between the phase responses of the microphones cause a small phase difference between the microphone signals. There now appears to be a flow of acoustic energy along the probe axis Intensity is detected 871838/2 Fig. 11 It is this apparent flow of acoustic energy that is detected and called residual intensity Signals are not in phase Real probe with phase mismatch db SPL Residual intensity Intensity is detected 871837/3 Frequency 870899/3 Even under controlled laboratory conditions, it is very difficult to create a free-field situation where the angle between the propagation of the sound wave and the probe axis is exactly 90 degrees (as shown in Fig. 8 and Fig. 10). However, for practical applications this situation can easily be simulated using the setup shown in Fig. 5. 5

Compliance with Standards, CE-mark indicates compliance with: EMC Directive and Low Voltage Directive. C-Tick mark indicates compliance with the EMC requirements of Australia and New Zealand. Safety EMC Emission EMC Immunity EN/IEC 61010 1: Safety requirements for electrical equipment for measurement, control and laboratory use. ANSI/UL 61010 1: Safety requirements for electrical equipment for measurement, control and laboratory use. EN/IEC 61000 6 3: Generic emission standard for residential, commercial and light industrial environments. EN/IEC 61000 6 4: Generic emission standard for industrial environments. CISPR 22: Radio disturbance characteristics of information technology equipment. Class B Limits. FCC Rules, Part 15: Complies with the limits for a Class B digital device. This ISM device complies with Canadian ICES 001 (interference causing equipment standard) EN/IEC 61000 6 1: Generic standards Immunity for residential, commercial and light industrial environments. EN/IEC 61000 6 2: Generic standards Immunity for industrial environments. EN/IEC 61326: Electrical equipment for measurement, control and laboratory use EMC requirements. Note: The above is only guaranteed using accessories listed in this Product Data sheet. Specifications Sound Intensity Calibrator Type 3541-A Intensity Coupler UA-0914 First Chamber: Ports 1 and 2 Second Chamber: Port 3 Chamber Volume: 10 cm 3 each Equivalent Load for Each Port: 250 mm 3 Correction Barometer UZ-0004 Pressure Range: 650 hpa to 1080 hpa Accuracy: Better than ±2.0% at 20 C Signal Levels obtained in Intensity Coupler UA-0914 REFERENCE CONDITIONS: Pressure: 1013 hpa Temperature: 20 C Relative Humidity: 65% CALIBRATION USING TYPE 4228 PORTS 1 AND 2: Sound Pressure Level: 118.0 ±0.4 db re 20 Pa Calibration Tolerance: ±0.2 db Ambient Pressure Coefficient: 8.4 10 3 db/hpa Temperature Coefficient: <±0.002 db/ C Humidity Coefficient: Negligible PORTS 1 OR 2, AND 3: Simulated Sound Intensity Level: 117.85 ±0.5 db re 1 pw/m 2 Calibration Tolerance: ±0.25 db Nominal Microphone Spacing: 50 mm Ambient Pressure Coefficient: 1.25 10 4 db/hpa Temperature Coefficient: 0.024 db/ C Humidity coefficient: Negligible SIMULATED PARTICLE VELOCITY LEVEL: 117.7 ±0.6 db re 50 nm/s Calibration Tolerance: ±0.3 db Nominal Microphone Spacing: 50 mm Temperature Coefficient: 0.05 db/ C Ambient pressure coefficient: 8.3 10 3 db/hpa Humidity Coefficient: Negligible PRESSURE-RESIDUAL INTENSITY INDEX OF SOUND FIELD IN UA-0914 MEASURED WITH: 50 mm Nominal Microphone Spacing: >30 db 12 mm Nominal Microphone Spacing: >24 db Dimension and Weight (Case) Height: 280 mm (11 ) Width: 230 mm (9 ) Depth: 63 mm (2.5 ) Weight: 2.3 kg (5.1 lb) Note: All values are typical at 20 C (77 F), unless measurement uncertainty or tolerance field is specified. All uncertainty values are specified at 2 (i.e., expanded uncertainty using a coverage factor of 2) Ordering Information Type 3541-A Sound Intensity Calibrator Includes the following accessories: UA-0914: Intensity Coupler Type 4228: Pistonphone UZ-0004: Correction Barometer 2 UA-1314: Two 1/4 Microphone Adaptors DB-3111: Intensity Coupler Base-Plate 6 QB-0013: 1.5 V Alkaline Battery IEC Type LR 6 QB-0016: 9 V Alkaline Battery IEC Type 6 LF 22 UC-5298: Dummy Microphone Microphone Adaptors for Pistonphone: DP-0776: 1/2 Adaptor DP-0775: 1/4 Adaptor DP-0774: 1/8 Adaptor Calibration Charts OPTIONAL ACCESSORIES Type 4297 Sound Intensity Calibrator 6

7

Brüel & Kjær reserves the right to change specifications and accessories without notice. Brüel & Kjær. All rights reserved. ËBP-2325---/Î BP 2325 13 2012-01 HEADQUARTERS: Brüel & Kjær Sound & Vibration Measurement A/S DK-2850 Nærum Denmark Telephone: +45 7741 2000 Fax: +45 4580 1405 www.bksv.com info@bksv.com Local representatives and service organisations worldwide