STUDY OF DETECTION LIMITS AND QUANTITATION ACCURACY USING 300 MHZ NMR

Similar documents
^i^l^z CHEMICAL BIOLOGICAL CENTER \ U.S. ARMY RESEARCH, DEVELOPMENT AND ENGINEERING COMMAND

Report Documentation Page

Report Documentation Page

REGENERATION OF SPENT ADSORBENTS USING ADVANCED OXIDATION (PREPRINT)

P. Kestener and A. Arneodo. Laboratoire de Physique Ecole Normale Supérieure de Lyon 46, allée d Italie Lyon cedex 07, FRANCE

Use of Wijsman's Theorem for the Ratio of Maximal Invariant Densities in Signal Detection Applications

Analysis Comparison between CFD and FEA of an Idealized Concept V- Hull Floor Configuration in Two Dimensions. Dr. Bijan Khatib-Shahidi & Rob E.

Attribution Concepts for Sub-meter Resolution Ground Physics Models

Quantitation and Ratio Determination of Uranium Isotopes in Water and Soil Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

System Reliability Simulation and Optimization by Component Reliability Allocation

Thermo-Kinetic Model of Burning for Polymeric Materials

REPORT DOCUMENTATION PAGE

Diagonal Representation of Certain Matrices

Catalytic Oxidation of CW Agents Using H O in 2 2 Ionic Liquids

Broadband matched-field source localization in the East China Sea*

Crowd Behavior Modeling in COMBAT XXI

CRS Report for Congress

High-Fidelity Computational Simulation of Nonlinear Fluid- Structure Interaction Problems

Molecular Characterization and Proposed Taxonomic Placement of the Biosimulant 'BG'

Predictive Model for Archaeological Resources. Marine Corps Base Quantico, Virginia John Haynes Jesse Bellavance

FATE OF NERVE AGENT SIMULANTS ON CONCRETE

Optimizing Robotic Team Performance with Probabilistic Model Checking

Report Documentation Page

Real-Time Environmental Information Network and Analysis System (REINAS)

7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) 8. PERFORMING ORGANIZATION REPORT NUMBER

Dynamics of Droplet-Droplet and Droplet-Film Collision. C. K. Law Princeton University

DIRECTIONAL WAVE SPECTRA USING NORMAL SPREADING FUNCTION

Volume 6 Water Surface Profiles

Extension of the BLT Equation to Incorporate Electromagnetic Field Propagation

Wavelets and Affine Distributions A Time-Frequency Perspective

Closed-form and Numerical Reverberation and Propagation: Inclusion of Convergence Effects

A report (dated September 20, 2011) on. scientific research carried out under Grant: FA

Coastal Engineering Technical Note

Erik L. Swanberg 1 and Steven G. Hoffert 2. Veridian Systems 1, Autometric 2. Sponsored by Defense Threat Reduction Agency

SW06 Shallow Water Acoustics Experiment Data Analysis

INFRARED SPECTRAL MEASUREMENTS OF SHUTTLE ENGINE FIRINGS

IMPROVED SYNTHESIS AND REACTION CHEMISTRY OF FN 3

Super-Parameterization of Boundary Layer Roll Vortices in Tropical Cyclone Models

Metrology Experiment for Engineering Students: Platinum Resistance Temperature Detector

A GENERAL PROCEDURE TO SET UP THE DYADIC GREEN S FUNCTION OF MULTILAYER CONFORMAL STRUCTURES AND ITS APPLICATION TO MICROSTRIP ANTENNAS

Scattering of Internal Gravity Waves at Finite Topography

Estimation of Vertical Distributions of Water Vapor and Aerosols from Spaceborne Observations of Scattered Sunlight

Effects of Constrictions on Blast-Hazard Areas

Comparative Analysis of Flood Routing Methods

USMC Enlisted Endstrength Model

Development and Application of Acoustic Metamaterials with Locally Resonant Microstructures

PIPS 3.0. Pamela G. Posey NRL Code 7322 Stennis Space Center, MS Phone: Fax:

On Applying Point-Interval Logic to Criminal Forensics

HIGH-POWER SOLID-STATE LASER: LETHALITY TESTING AND MODELING

Award # N LONG TERM GOALS

LAGRANGIAN MEASUREMENTS OF EDDY CHARACTERISTICS IN THE CALIFORNIA CURRENT

Sensitivity of West Florida Shelf Simulations to Initial and Boundary Conditions Provided by HYCOM Data-Assimilative Ocean Hindcasts

High Resolution Surface Characterization from Marine Radar Measurements

REPORT DOCUMENTATION PAGE. Theoretical Study on Nano-Catalyst Burn Rate. Yoshiyuki Kawazoe (Tohoku Univ) N/A AOARD UNIT APO AP

Fleet Maintenance Simulation With Insufficient Data

Improvements in Modeling Radiant Emission from the Interaction Between Spacecraft Emanations and the Residual Atmosphere in LEO

FRACTAL CONCEPTS AND THE ANALYSIS OF ATMOSPHERIC PROCESSES

Super-Parameterization of Boundary Layer Roll Vortices in Tropical Cyclone Models

Design for Lifecycle Cost using Time-Dependent Reliability

Direct Numerical Simulation of Aeolian Tones

Z-scan Measurement of Upconversion in Er:YAG

Nonlinear Internal Waves: Test of the Inverted Echo Sounder

USER S GUIDE. ESTCP Project ER

MTIE AND TDEV ANALYSIS OF UNEVENLY SPACED TIME SERIES DATA AND ITS APPLICATION TO TELECOMMUNICATIONS SYNCHRONIZATION MEASUREMENTS

VLBA IMAGING OF SOURCES AT 24 AND 43 GHZ

2008 Monitoring Research Review: Ground-Based Nuclear Explosion Monitoring Technologies SPECTRAL ANALYSIS OF RADIOXENON

Experimental and Theoretical Studies of Ice-Albedo Feedback Processes in the Arctic Basin

Characterization of Caribbean Meso-Scale Eddies

Understanding Near-Surface and In-cloud Turbulent Fluxes in the Coastal Stratocumulus-topped Boundary Layers

Ocean Acoustics Turbulence Study

Mine Burial Studies with a Large Oscillating Water-Sediment Tunnel (LOWST)

Aviation Weather Routing Tool: A Decision Aid for Manned/Unmanned Aircraft Routing

Playing Abstract games with Hidden States (Spatial and Non-Spatial).

Nano and Biological Technology Panel: Quantum Information Science

Regional Stratification and Shear of the Various Streams Feeding the Philippine Straits ESR Component

Improved Parameterizations Of Nonlinear Four Wave Interactions For Application In Operational Wave Prediction Models

REPORT DOCUMENTATION PAGE

PREDICTION OF HIGH TEMPERATURE GREASE LIFE USING A DECOMPOSITION KINETIC MODEL

Predicting Tropical Cyclone Formation and Structure Change

EFFECTS OF LOCAL METEOROLOGICAL VARIABILITY ON SURFACE AND SUBSURFACE SEISMIC-ACOUSTIC SIGNALS

Marginal Sea - Open Ocean Exchange

Flocculation, Optics and Turbulence in the Community Sediment Transport Model System: Application of OASIS Results

The Mechanics of Bubble Growth and Rise in Sediments

HYCOM Caspian Sea Modeling. Part I: An Overview of the Model and Coastal Upwelling. Naval Research Laboratory, Stennis Space Center, USA

Curating and Maintenance of the Sediment Library and Dredge Collection

Experimental and Theoretical Studies of Ice-Albedo Feedback Processes in the Arctic Basin

MODELING SOLAR UV/EUV IRRADIANCE AND IONOSPHERE VARIATIONS WITH MT. WILSON MAGNETIC INDICIES

Coastal Mixing and Optics

uniform distribution theory

Determining the Stratification of Exchange Flows in Sea Straits

Wavelet Spectral Finite Elements for Wave Propagation in Composite Plates

HIGH PERFORMANCE CONTROLLERS BASED ON REAL PARAMETERS TO ACCOUNT FOR PARAMETER VARIATIONS DUE TO IRON SATURATION

Guide to the Development of a Deterioration Rate Curve Using Condition State Inspection Data

Expressions for the Total Yaw Angle

SMA Bending. Cellular Shape Memory Structures: Experiments & Modeling N. Triantafyllidis (UM), J. Shaw (UM), D. Grummon (MSU)

Investigation of the Air-Wave-Sea Interaction Modes Using an Airborne Doppler Wind Lidar: Analyses of the HRDL data taken during DYNAMO

Predicting Deformation and Strain Behavior in Circuit Board Bend Testing

Discovery of Planetary Systems With SIM

Study of Electromagnetic Scattering From Material Object Doped Randomely WithThin Metallic Wires Using Finite Element Method

THE EULER FUNCTION OF FIBONACCI AND LUCAS NUMBERS AND FACTORIALS

Transcription:

STUDY OF DETECTION LIMITS AND QUANTITATION ACCURACY USING 300 MHZ NMR William R. Creasy EAI Corporation, 1308 Continental Drive, Suite J, Abingdon MD 21009 David J. McGarvey, Jeffrey S. Rice, Richard O'Connor, and H. Dupont Durst R&T Directorate, Edgewood Chemical and Biological Center, Aberdeen Proving Ground, MD ABSTRACT The analytical precision and accuracy of 300 MHz nuclear magnetic resonance (NMR) spectrometers was determined. The instruments that were used include a Bruker AC-300, a Bruker Avance 300, and a Varian 300. The minimum detection limits for proton ( 1 H) and phosphorus ( 31 P) were determined and compared for different instruments. The Minimum Detection Limits (MDL) for proton measurements were in the range of 3-10 µg/g. The MDL for phosphorus measurements were in the range of 50-70 µg/g. The quantitation accuracy for purity determinations was measured to be in the range of 1% relative to an internal standard. Specific MDL values should be determined for each particular matrix, internal standard, and instrument conditions. INTRODUCTION The analytical accuracy and precision of 300 MHz Nuclear Magnetic Resonance (NMR) spectrometers were studied under typical sample analysis conditions. Instruments used were Bruker Avance, Bruker AC-300, and Varian 300 MHz. Low concentration trace detection and high concentration purity determinations of chemical weapons simulants were done. Minimum detection limits (MDL) based on the 99% confidence limits were determined using the EPA criteria 1 for 1 H and 31 P NMR in standard solutions and aqueous samples. EXPERIMENTAL METHODS The sample or standard solutions were transferred to a 5 mm NMR tube to run on the 300 MHz NMR instruments. Unless otherwise stated, the same sample was analyzed on two days with 8 repetitions each day to determine the reproducibility of the NMR signals. Data processing was done with Acorn NUTS software, Bruker XWIN-NMR, or Varian software. Standard deviations of the data were used to determine 99% confidence limits, which were used for the MDL determinations. 1

Report Documentation Page Form Approved OMB No. 0704-0188 Public reporting burden for the collection of information is estimated to average 1 hour per response, including the time for reviewing instructions, searching existing data sources, gathering and maintaining the data needed, and completing and reviewing the collection of information. Send comments regarding this burden estimate or any other aspect of this collection of information, including suggestions for reducing this burden, to Washington Headquarters Services, Directorate for Information Operations and Reports, 1215 Jefferson Davis Highway, Suite 1204, Arlington VA 22202-4302. Respondents should be aware that notwithstanding any other provision of law, no person shall be subject to a penalty for failing to comply with a collection of information if it does not display a currently valid OMB control number. 1. REPORT DATE 01 JUL 2003 2. REPORT TYPE N/A 3. DATES COVERED - 4. TITLE AND SUBTITLE Study Of Detection Limits And Quantitation Accuracy Using 300 Mhz Nmr 5a. CONTRACT NUMBER 5b. GRANT NUMBER 5c. PROGRAM ELEMENT NUMBER 6. AUTHOR(S) 5d. PROJECT NUMBER 5e. TASK NUMBER 5f. WORK UNIT NUMBER 7. PERFORMING ORGANIZATION NAME(S) AND ADDRESS(ES) EAI Corporation, 1308 Continental Drive, Suite J, Abingdon, MD 21009 8. PERFORMING ORGANIZATION REPORT NUMBER 9. SPONSORING/MONITORING AGENCY NAME(S) AND ADDRESS(ES) 10. SPONSOR/MONITOR S ACRONYM(S) 12. DISTRIBUTION/AVAILABILITY STATEMENT Approved for public release, distribution unlimited 13. SUPPLEMENTARY NOTES See also ADM001523., The original document contains color images. 14. ABSTRACT 15. SUBJECT TERMS 11. SPONSOR/MONITOR S REPORT NUMBER(S) 16. SECURITY CLASSIFICATION OF: 17. LIMITATION OF ABSTRACT UU a. REPORT unclassified b. ABSTRACT unclassified c. THIS PAGE unclassified 18. NUMBER OF PAGES 5 19a. NAME OF RESPONSIBLE PERSON Standard Form 298 (Rev. 8-98) Prescribed by ANSI Std Z39-18

SAMPLE PREPARATION An internal standard (IS) of hexamethylphosphoramide (HMPA, 99% purity) from Sigma Aldrich was added as a quantitation reference internal standard (IS). IS and samples were weighed. Dilute standards were diluted by weight to weight solvent. Typically, the IS was added at 10 wt.% of the sample or standard. This was used to obtain an accurate IS and solution weight. By comparison, this IS weight is much higher than the dilute analytes at 50-200 µg/g (ppm), giving an indication of instrumental dynamic range. MDL FOR 1 H IN A STANDARD (AC-300) RESULTS The MDL was determined for proton NMR on the Bruker AC-300 with a standard solution. A standard of 45 ppm thiodiglycol (TDG) in D 2 O was prepared. Water solvent suppression was used. The number of scans that were signal averaged was 128. As shown in Table 1, the measured MDL was 10 ppm. Recovery was 97%, indicating good quantitation accuracy. Table 1. MDL Results for 1 H standard (AC-300) Prepared Concentration (ppm) 45.81 Average Found Concentration (ppm) relative 44.45 to internal standard Standard deviation 3.95 Calculated MDL (ppm) 10.27 Recovery % 97% MDL FOR 1 H IN A STANDARD (VARIAN 300) For comparison, the MDL was also determined for proton NMR on the Varian 300. A standard solution of 45 ppm thiodiglycol (TDG) in D 2 O was prepared. Water solvent suppression was used. As shown in Table 2, the MDL was 3 ppm, lower than the AC-300. Recovery was 96%. Table 2. MDL Results for 1 H in Varian 300 Prepared concentration (ppm) 44.32 Average Found Concentration (ppm) 42.68 Standard deviation (ppm) 1.03 Calculated MDL (ppm) 3.09 Recovery % 96% MDL FOR 1 H: EFFECT OF WEIGHING ERROR (VARIAN 300) The MDL was determined for multiple standard preps using proton NMR on the Varian 300 to determine the error associated with weighing. Seven standard solutions of 45 ppm TDG in D 2 O were prepared and analyzed for seven reps each. The average standard deviation for the seven NMR reps was 0.99 ppm in the measurement of the found concentration. The standard deviation for the seven different standards was 0.28 ppm, which indicates the error from weighing is less than the error from the NMR data repetitions. 2

Table 3. MDL for 1 H: Results for effect of weighing error (Varian 300) Std. Dev. Average 0.990 Std. Dev. Deviation 0.276 Recovery Average 94.46% Recovery Std. Dev. 1.58% MDL Average 2.961 MDL Std. Dev. 0.824 MDL FOR 1 H IN AN AQUEOUS SAMPLE An aqueous sample was studied without water suppression on the AC-300. The solution was mostly H 2 O. The sample was spiked to 200 ppm with TDG. The same methodology was used. As shown in Table 4, the MDL was higher, at 116 ppm, due to interference from the large peak tailing of the water peak. The tailing made accurate integration of the peak more difficult, so recovery was only 76%. This is a worst case, but these results indicate the MDL can vary significantly depending on sample and instrument conditions. Of course, the MDL can also depend on the instrument parameters such as number of scans, excite pulse power, and excite pulse length. Table 4. MDL Data for 1 H in an Aqueous Sample Prepared concentration 202.30 Average Found Concentration (ppm) 153.97 Standard deviation 44.62 MDL (ppm) 116.11 Recovery % 76% MDL FOR 31 P IN AN AQUEOUS SAMPLE The MDL for 31 P NMR was determined for both AC-300 and Avance instruments, shown in Table 5 and 6. A standard solution of 100 ppm of diisopropyl methylphosphonate (DIMP) in H 2 O was prepared. The major source of error was from the phasing of the spectra. The IS peak was much larger than the DIMP peak, which made it difficult for both peaks to have the optimal phasing. Small changes in the large peak made much larger effects on the integration of the nearby small peak. As a result, it is likely that the MDL for 31 P detection can be made significantly lower, similar to those for 1 H, by using less internal standard or by choosing an internal standard that is further away from the analyte peak. Table 5. MDL Results for 31 P in a Standard for the AC-300 Prepared concentration (ppm) 99.48 Average Found Concentration (ppm) 98.06 Standard deviation 26.20 MDL (ppm) 68.18 Recovery % 98% 3

Table 6.MDL Results for 31 P in a Standard for the Avance Prepared concentration (ppm) 99.48 Average Found Concentration (ppm) 100.44 Standard deviation 20.54 MDL (ppm) 53.45 Recovery % 101% MDL is slightly higher for the AC-300 that for the Avance, which is a newer instrument, but there is not a large difference for the analysis of the same sample. QUANTITATIVE PURITY DETERMINATIONS NMR is used for purity determinations of agent standards, as well as trace analysis. A detailed comparison of quantitation accuracy and precision was made between the two Bruker 300 MHz instruments for each of the nucleii 1 H, 31 P, and 13 C. A sample was prepared by weighing about approximately equal amounts of HMPA (99% purity) and triethylphosphate (TEP, 99+% purity) and transferring it to an insert in a 5 mm NMR tube. Results that are shown in Figure 1 were corrected for measured weights and molecular weights so the expected relative purity is 1.00. The measured accuracy for both instruments is within 1% of the expected purity. Although the AC-300 typically had somewhat larger standard deviations, there was generally good agreement between the AC-300 and the Avance instruments. The one exception was case B, in which the excitation frequency was not centered between the two large phosphorus peaks. The AC-300 gave a much lower result due to the roll-off of the excite waveform. NMR Relative Purity of HMPA and TEP 1.1 Purity 1.05 1 A C D E 0.95 B 0.9 Experiment A = P-31, excite centered B = P-31, excite not centered C = P-31, no H decoupling D = C-13 with H decoupling E = Proton AC-300 Avance 300 Figure 1. Quantitative Purity Determination Results for the comparison of the AC-300 to the Avance spectrometers. Error bars indicate one standard deviation from seven reps except for case B, which had three reps. 4

CONCLUSIONS Although the older AC-300 instrument has slightly poorer performance, all three 300 MHz instruments give signal precision and accuracy that is reasonably similar. MDL results for proton measurements were in the range of 3-10 ppm in concentration, although the MDL depends on instrument conditions. MDL for phosphorus determinations were in the range of 50-70 ppm. Purity determinations gave accuracy in the range of 1% or less. Additional information is given in technical reports. 2-5 ACKNOWLEDGEMENTS This work was performed under contract number DAAM01-97-0005 at SBCCOM, U. S. Army. The work was done at the Edgewood Chemical and Biological Center (ECBC, formerly ERDEC), Aberdeen Proving Ground-Edgewood Area, MD. REFERENCES 1. Code of Federal Regulations, 40 CFR 136 Appendix B. 2. R. J. O Connor, M. D. Brickhouse, D. J. McGarvey, H. D. Durst, W. R. Creasy, J. L. Ruth, NMR Method for the Quantitative Purity Analysis of Neat Feed Stock Sample, ECBC Technical Report #ECBC-TR-253, August 2002. 3. R. J. O Connor, M. D. Brickhouse, D. J. McGarvey, W. R. Creasy, J. L. Ruth, H. D. Durst, NMR Method for the Quantitative Analysis of Neat Materials, ECBC Technical Report Draft, October 2002. 4. D. J. McGarvey, R. J. O Connor, M. D. Brickhouse, W. R. Creasy, J. S. Rice, J. Pence, J. L. Montgomery, H. D. Durst, NMR Method for the Quantitative Analysis of Trace-level Analytes in Liquid Samples, ECBC Technical Report Draft, October 2002. 5. D. J. McGarvey, R. J. O Connor, M. D. Brickhouse, W. R. Creasy, J. S. Rice, H. D. Durst, NMR Method for the Quantitative Analysis of VX Hydrolysates, ECBC Technical Report Draft, October 2002. 5