Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Similar documents
Fast methods for the determination of ibuprofen in drug products

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Tomorrow s quantitation with the TSQ Fortis mass spectrometer: quantitation of phenylephrine hydrochloride for QA/QC laboratories

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Multiple Fragmentation Methods for Small Molecule Characterization on a Dual Pressure Linear Ion Trap Orbitrap Hybrid Mass Spectrometer

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Determination of urea in ultrapure water by IC-MS/MS

HR/AM Targeted Peptide Quantification on a Q Exactive MS: A Unique Combination of High Selectivity, High Sensitivity, and High Throughput

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Automated and accurate component detection using reference mass spectra

Improved Screening for 250 Pesticides in Matrix using a LC-Triple Quadrupole Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

High-Throughput LC-MS/MS Quantification of Estrone (E1) and Estradiol (E2) in Human Blood Plasma/Serum for Clinical Research Purposes

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of trace anions in concentrated hydrofluoric acid

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Multi-residue analysis of pesticides by GC-HRMS

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Sensitive and rapid determination of polycyclic aromatic hydrocarbons in tap water

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Improved Throughput and Reproducibility for Targeted Protein Quantification Using a New High-Performance Triple Quadrupole Mass Spectrometer

A Strategy for an Unknown Screening Approach on Environmental Samples Using HRAM Mass Spectrometry

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Routine MS Detection for USP Chromatographic Methods

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Utility of H-SRM to Reduce Matrix Interference in Food Residue Analysis of Pesticides by LC-MS/MS Using the TSQ Quantum Discovery

Relative quantification using TMT11plex on a modified Q Exactive HF mass spectrometer

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

Metoprolol and Select Impurities Analysis Using a Hydrophilic Interaction Chromatography Method with Combined UV and Charged Aerosol Detection

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

A novel high resolution accurate mass Orbitrap-based GC-MS platform for routine analysis of Short Chained Chlorinated Paraffins

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Monitoring Protein PEGylation with Ion Exchange Chromatography

Automated, intelligent sample preparation: Integration of the ESI prepfast Auto-dilution System with the Thermo Scientific icap 7400 ICP-OES

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Dionex IonPac AS28-Fast-4µm column

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

LC Technical Information

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Mass Spectral Studies of Polypropylene Chromatographic Well Plates

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

Enhancement of Linearity and Response in Charged Aerosol Detection

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

Comparison of Solid Core HPLC Column Performance: Effect of Particle Diameter

APPLICATION NOTE Goal. Introduction. Keywords

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Modernizing the USP Monograph for Acetaminophen

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Peptide Isolation Using the Prep 150 LC System

Mixed-Mode, Weak Anion-Exchange, Solid-Phase Extraction Method for the Extraction of Niflumic Acid from Human Plasma

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Analysis of Silicone Oils by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Multi-Channel SFC System for Fast Chiral Method Development and Optimization

Low-level Determination of 4-Hydrazino Benzoic Acid in Drug Substance by High Performance Liquid Chromatography/Mass Spectrometry

Routine-grade performance of a new static headspace autosampler for the analysis of residual solvents according to USP <467> method

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Transcription:

APPLICATION NOTE 72385 Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer Authors Stephan Meding, Katherine Lovejoy, Martin Ruehl Thermo Fisher Scientific, Germering, Germany Keywords Single quadrupole mass spectrometer, Medicinal chemistry, Synthesis control, Quantitative LC-MS, High throughput Goal Demonstrate quantitative synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer and show its benefit for medicinal chemistry. Introduction Medicinal chemistry has a central place in drug discovery, lead optimization, and drug development. It enables pharmaceutical companies to research and develop new active pharmaceutical compounds or even compound classes. The success of organic synthesis of new molecules or of the isolation of naturally sourced compounds needs to be established. An important tool for this is LC-MS, which allows verification of the analyte by mass confirmation to determine the yield of the reaction and to check for impurities related to the proposed drug candidate in the same analysis. Another requirement is fast analysis to achieve high throughput. Often large numbers of candidates need to be screened, putting a strain on LC-MS resources. 1

Modern single quadrupole mass spectrometers, such as the Thermo Scientific ISQ EC single quadrupole mass spectrometer (ISQ EC MS), are reliable workhorses designed for routine applications. The ISQ EC MS can operate in Full Scan and Single Ion Monitoring (SIM) mode, to either scan a mass range for all detectable analytes or focus on a specific compound. It can run at scan rates suitable for fast UHPLC applications while delivering picogram detection limits. The new orthogonal source design provides high levels of instrument robustness, even with difficult matrices. Full integration into the Thermo Scientific Chromeleon 7.2 Chromatography Data System (CDS) and the Thermo Scientific AutoSpray smart method setup make LC-MS operation straightforward and intuitive. The synthesis of acetaminophen, also known as paracetamol, was selected as an example to prove that the ISQ EC MS can be used for fast, quantitative synthesis control. is produced in a onestep synthesis by acetylation of 4-aminophenol (Figure 1). The ratios of pre-mixed reactant to product sample mixes were determined and compared with the theoretical values to determine the accuracy and precision of the presented workflow. + + Acetic Anhydride Acetic Acid Figure 1. Reaction equation of acetaminophen (paracetamol). Experimental Sample reagents were purchased from Sigma-Aldrich. Table 1 lists the analytes used and Table 2 lists the solvents and additives. Table 1. Analytes. Analyte Purity CAS Chemical Formula Monoisotopic Mass [M] [M+H] + 98% 123-30-8 C 6 H 7 NO 109.05 110.06 98.0 102.0% 103-90-2 C 8 H 9 NO 2 151.06 152.07 Table 2. Solvents and additives. Reagent Grade Supplier Part Number Acetonitrile Optima LC-MS Fisher Chemical A955-212 Formic acid Optima LC-MS Fisher Chemical A117-50 Methanol Optima LC-MS Fisher Chemical A456-212 Water Ultra-Pure, 18.2 MΩ at 25 C Thermo Scientific Barnstead GenPure xcad Plus Ultrapure Water Purification System 2

Chromatographic separation was performed on a Thermo Scientific Vanquish Flex Binary UHPLC system consisting of a binary pump, an autosampler, a column compartment, and a variable wavelength detector (Table 3). A 75 cm long MP35N capillary with 100 µm inner diameter (P/N 6042.2390) was used for connecting to the ISQ EC MS. The HPLC conditions are listed in Table 4, and the MS conditions are listed in Table 5. Table 3. Vanquish Flex Binary UHPLC system modules. Module Vanquish System Base F Vanquish Binary Pump F (with 35 µl mixer) Vanquish Split Sampler FT Vanquish Column Compartment H Vanquish Variable Wavelength Detector F (2.5 µl SST flow cell) Part number VF-S01-A VF-P10-A (6044.3870) VF-A10-A VH-C10-A VF-D40-A (6074.0360) Table 4. HPLC conditions. Parameter Value Column Thermo Scientific Hypersil GOLD, 1.9 µm, 2.1 x 50 mm (P/N 25002-052130) Mobile phase Gradient Flow rate A: Water with 0.1% formic acid B: Acetonitrile with 0.1% formic acid 0 2 min: 5 50% B 2 3 min: 50% B 3 5 min: 5% B 0.6 ml/min Column temperature Forced air, 40 C Active preheater, 40 C Injection volume 1 µl UV detection 280 nm, 50 Hz, easy mode Table 5. MS conditions. Parameter Value Vaporizer temperature 550 C Ion transfer tube temperature 350 C Source voltage SIM scan Compound Time Mass Source CID voltage Compound Time Mass Source CID voltage Full Scan Time Mass range Source CID voltage +1000 V 0 0.5 min 110.1 m/z 20 V 0.5 1.18 min 152.1 m/z 20 V 0 5 min 105 250 m/z 20 V 3

The ISQ EC MS is fully integrated into the Chromeleon 7.2 CDS, which was used for system operation and subsequent data analysis. Calibration standards (10 ppb 10 ppm) were prepared by serially diluting stock solutions (1 µg/µl 4-aminophenol and acetaminophen in 50:50 methanol:water) in 5% methanol in water. Samples for measuring the reactant to product ratios were prepared diluting the stock solutions in 5% methanol in water. Prepared samples are listed in Table 6. Table 6. Sample analyte ratios and used sample concentrations (1 ppm = 1 ng/µl). Theoretical Ratio (Reactant:Product) (ppm) (ppm) 100:1 10 0.1 50:1 5 0.1 20:1 2 0.1 10:1 1 0.1 5:1 0.5 0.1 2:1 0.2 0.1 1:1 0.1 0.1 1:2 0.1 0.2 1:5 0.1 0.5 1:10 0.1 1 1:20 0.1 2 1:50 0.1 5 1:100 0.1 10 1:200 0.05 10 1:500 0.02 10 1:1000 0.01 10 4

Results and discussion Synthesis control requires fast, quantitative analysis of the reactants and products while they are present in different ratios. In this work, ratios between 100:1 and 1:1000 (reactant to product) were premixed and quantified in order to mimic the time course of acetaminophen synthesis from the very start to near completion. To assure the best quantitation results, a high-throughput method was developed, which provided baseline separation of 4-aminophenol and acetaminophen (Figure 2). Total Ion Chromatogram from Full Scan 50% B SIM scans 5% B 110.1 m/z 152.1 m/z Timed SIM scans Full Scan m/z 105-250 Figure 2. Total ion chromatogram of 4-aminophenol and acetaminophen (50 pg) in full scan acquisition. The gradient part of the high throughput method is depicted. The gradient starts with 5% B and increases to 50% B. The acquisition windows for the SIM scans are depicted in orange and blue. 5

The analytes were detected with SIM mode to improve detection limits. The SIM window (0.4 amu) and dwell time (0.2 s) parameters were selected to increase signal intensity and to assure at least 15 scans over the peak for good quantitation results. The SIM scans for 4-aminophenol (110.1 m/z) and acetaminophen (152.1 m/z) were timed using retention time windows fully encompassing the retention time of the expected chromatographic peaks (Figure 2). For initial method verification, full scan acquisition together with SIM mode acquisition was performed. For quantification, only SIM scans were performed to assure optimal analyte detection. Mass spectrometric detectors usually outperform UV detectors in terms of detection limits. Thus, the detection limits of the ISQ EC MS and the Vanquish Flex variable wavelength UV detector (VWD) were compared. Looking at the signal responses of an ISQ EC MS and a VWD revealed differences in detetction limits of up to three orders of magnitude (Figure 3). With the VWD, 5 ng acetaminophen on column were detectable with a signal-to-noise ratio (peak to peak) of 18. Therefore, a detection limit of ~1 ng can be expected (S/N 3). Using the ISQ EC MS extracted ion chromatograms from the Full Scan afforded detection down to 50 pg of analyte; in SIM mode signal-to-noise was 55 at 50 pg and 4 at 5 pg. Thus, 5 pg is the lower limit of detection. on column 50 ng 5 ng 500 pg 50 pg 5 pg 1 pg SIM Scan Intensity [counts] EIC / Full Scan Intensity [counts] UV Signal Intensity [mau] Figure 3. Comparison of signal response between UV and ISQ EC detector for acetaminophen (EIC: Extracted Ion Chromatogram). 6

Calibration curves for 4-aminophenol and acetaminophen spanning the relevant sample concentrations were generated (Figure 4). All injections were done in quintuplicate. Afterwards reinjections of calibrants were done in triplicate to verify the accuracy of the calibration. and acetaminophen showed good recovery rates. The standard deviation between the reinjection replicates was below 5% indicating high precision (data not shown). 30000 25000 Area [counts*min] 20000 15000 10000 5000 0 70000 62500 0 2500 5000 7500 10000 11000 Concentration (ppb) 50000 Area [counts*min] 37500 25000 12500 0 0 2500 5000 7500 10000 11000 Concentration (ppb) Figure 4. Calibration curves for 4-aminophenol and acetaminophen. 1 ppb equals 1 pg on column since 1 µl was injected. Quadratic fit with 1/x weighting was applied. : 3.7220x - 0.0001x 2, R 2 = 0.9962; : 663.3313 + 10.2319x 0.0004x 2, R 2 = 0.9966. 7

For synthesis control, 4-aminophenol / acetaminophen mixtures at different ratios were analyzed starting with 100:1 and finishing with 1:1000 (Table 6 and Figure 5). All samples were analyzed in quintuplicate. The change in the composition of the samples could be accurately detected. With changing ratios, the relative amount of acetaminophen increased in the expected manner. Even relative abundances of 4-aminophenol below 1 percent could be accurately determined (zoom-in, Figure 5). These results show that the ISQ EC MS can accurately quantify analytes. This highlights that the presented single quadrupole mass spectrometer is a suitable tool for high-throughput monitoring of chemical syntheses in medicinal chemistry. 100% Relative amount of analyte 80% 60% 40% 20% 0% 5.0% 0.5% 0.1% Pre-mixed Ratio (Reactant:Product) Pre-mixed Ratio (Reactant:Product) Figure 5. Synthesis control. Observed 4-aminophenol:acetaminophen ratios are plotted. The pre-mixed ratios are stated on the x-axis. Zoom-in shows 1:20 to 1:1000 ratios for 4-aminophenol on a logarithmic scale. Conclusion High-throughput, quantitative synthesis control can be done with the ISQ EC single quadrupole mass spectrometer. SIM mode greatly increases sensitivity over UV detection and can be used for targeted quantitative analyses. Full Scan mode results in general detection of present analytes, e.g. monitoring other potential products, and can be used for generic methods. Find out more at thermofisher.com/isqec 8 2017 Thermo Fisher Scientific Inc. All rights reserved. All trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representatives for details. AN72385-EN 0617S