Understanding Gas Chromatography

Similar documents
Introduction to Capillary GC

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Introduction to Capillary GC

Chemistry Instrumental Analysis Lecture 27. Chem 4631

Capillary GC Column Selection and Method Development A Primer on Column Parameters and Instrument Conditions

Secrets of GC Column Dimensions

CHAPTER 6 GAS CHROMATOGRAPHY

Trajan SGE GC Columns

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Gas Chromatography. Introduction

CH 2252 Instrumental Methods of Analysis Unit V Gas Chromatography. M. Subramanian

Ultra-Inert chemistry for Trace Level Analysis

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Gas Chromatography (GC)

2401 Gas (liquid) Chromatography

Chemistry Gas Chromatography: Separation of Volatile Organics

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Selection of a Capillary

Introduction to Gas Chromatography

Water Injections in GC - Does Water Cause Bleed?

Activity in the FID Detection Port: A Big Problem if Underestimated

Volatile organic compounds (VOCs):

Theory and Instrumentation of GC. Sample Introduction

Chapter 27: Gas Chromatography

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Theory and Instrumentation of GC. Chromatographic Parameters

Selection of a Capillary GC Column

Disadvantage: Destructive Technique once analyzed by GC, the sample is lost

Introduction and Principles of Gas Chromatography

New ZB-5HT Inferno The World s Highest Temperature Non-Metal GC Column

An Advanced Base Deactivated Capillary Column for analysis of Volatile amines Ammonia and Alcohols.

Chromatography. Gas Chromatography

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Chapter 31 Gas Chromatography. Carrier Gas System

Harris: Quantitative Chemical Analysis, Eight Edition CHAPTER 23: GAS CHROMATOGRAPHY

Gas Chromatography. Presented By Mr. Venkateswarlu Mpharm KTPC

Instrumental Chemical Analysis

Fall 2012 Due In Class Friday, Oct. 19. Complete the following on separate paper. Show your work and clearly identify your answers.

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Chem 230, Fall, 2014 Homework Set # 3 Short Answer SOLUTIONS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Ch24. Gas Chromatography (GC)

Gas Chromatography. Rosa Yu, David Reckhow CEE772 Instrumental Methods in Environmental Analysis CEE 772 #16 2

GC Resolution Do You See What I See?

How To Select the Correct GC Column. Simon Jones Application Engineer

Chromatography- Separation of mixtures CHEM 212. What is solvent extraction and what is it commonly used for?

Gas chromatography. Advantages of GC. Disadvantages of GC

Agilent Intuvo 9000 Gas Chromatograph System

Understanding the Capillary GC Column: How to Choose the Correct Type and Dimension

COOKBOOK FOR THE PERKINELMER SWAFER TM TECHNOLOGY FOR GC APPLICATIONS

2] The plate height in chromatography is best described as 2

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Gas Chromatography (Chapter 2 and 3 in The essence of chromatography)

Headspace Technology for GC and GC/MS: Features, benefits & applications

-xt. -xt SYSTEM. Specifications for PAL-xt Systems. Valid for PAL-xt System models only. Prep and Load Platform

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

So Many Columns! How Do I Choose? Daron Decker Chromatography Technical Specialist

Paints and varnishes Determination of volatile organic compound (VOC) content. Part 2: Gas-chromatographic method

GAS CHROMATOGRAPHY. Mobile phase is a gas! Stationary phase could be anything but a gas

GAS CHROMATOGRAPHY (GC)

Fundamentals of GC: Introduction: Sample introduction:

Chromatographic Methods of Analysis Section: 5 Gas Chromatography (GC) Prof. Tarek A. Fayed

GC and GCMS Fundamentals of Instrument Troubleshooting and Maintenance. Kirk Lokits Ph.D. Applications Chemist

High Performance Liquid Chromatography

Chapter 11 Conventional Gas Chromatography

PRINCIPLES AND APPLICATION OF CHROMATOGRAPHY. Dr. P. Jayachandra Reddy Mpharm PhD Principal & professor KTPC

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Partitioning. Separation is based on the analyte s relative solubility between two liquid phases or a liquid and solid.

Course goals: Course goals: Lecture 1 A brief introduction to chromatography. AM Quality parameters and optimization in Chromatography

Mass Spectrometry. Fundamental GC-MS. GC Considerations

Luminescence transitions. Fluorescence spectroscopy

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Open Column Chromatography, GC, TLC, and HPLC

Practical Fast Gas Chromatography for Contract Laboratory Program Pesticide Analyses *

Title Experiment 7: Gas Chromatography and Mass Spectrometry: Fuel Analysis

Determination of Volatile Substances Proof of Food Adulteration

Detection of Chemical Warfare Agents by Transportable GC/MS

Chapter 27: Gas Chromatography. Principles Instrumentation Detectors Columns and Stationary Phases Applications

Analytical Chemistry

Principles of Gas- Chromatography (GC)

TOTALLY INNOVATIVE MULTIMODE AUTOSAMPLER NEW KONIK ROBOKROM Laboratory Gas Generators An Overview +8 OPERATIONAL MODES MAIN FEATURES

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

The end of. mass-speculation. MS Certified Vials Pre-cleaned and certified vials for mass spectrometry

Gas Chromatography. 1. Experiment Category: 2. Experiment Name: 3. Date and Issue number: 4. Instructor Name: 5. Institution: Ain Shams University

Experiment 6 Simple and Fractional Distillation

Dynamic headspace (DHS) technique: set-up and parameter control for GC/MS analysis of odorant formulations

CEE 772: Instrumental Methods in Environmental Analysis

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Instrumentation. Components of a gas chromatograph

environmental Rtx -CLPesticides and Rtx -CLPesticides2 Columns: The Ideal Confirmational Pair for Analyzing Polychlorinated Biphenyls (PCBs)

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

U.S. EPA Method 8270 for multicomponent analyte determination

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

Transcription:

Understanding Gas Chromatography What is Really Going on Inside the Box? Simon Jones GC Applications Engineer Page 1 Group/Presentation Title Month ##, 200X

?? K? Page 2

Typical GC System Gas supply Injector Detector Data handling GAS Column Oven Page 3

CARRIER GAS Carries the solutes down the column Selection and velocity influences efficiency and retention time Page 4

VAN DEEMTER CURVES 1.00 Diffusion N 2 Poor mass transfer H 0.75 0.50 He Small 0.25 Large H 2 10 20 30 40 50 60 u (cm/sec) Page 5

CARRIER GAS Type Velocity Range (u opt OPGV) Nitrogen 8-16 Helium 20-40 Hydrogen 30-55 Page 6

Sample Introduction Purpose: To introduce a representative portion of sample onto the column in a reproducible manner, while minimizing sample bandwidth Syringe Injection Autosampler injection Valve Injection Gas sampling valve Liquid sampling valves Objective: The sample must not be chemically altered, unless desired (e.g., derivatization). Success is not contamination, degradation, or discrimination. Group/Presentation Title Month ##, 200X

Injection Port Types Purged Packed Split/Splitless Cool-On-Column PTV MMI Volatiles Inlet Page 8

SPLIT/SPLITLESS INJECTOR Carrier gas source Septum purge (~2ml/min) Ferrule Column Split vent Flow through injector = Column flow + Split Vent Flow Page 9

Splitless Injector Purge Off At Injection Carrier gas source Septum purge Split vent Flow through injector = Column flow only

Splitless Injector Purge On After Injection Carrier gas source Septum purge Split vent Flow through injector = Column flow + Split Vent Flow

Influence of Injection Efficiency Short Concentrated Solute Bands Long Diffuse Same column, same chromatographic conditions

Split Injector Major Variables Split ratio - determines amount of sample onto column and efficiency of injection (sensitivity vs peak shape) Liner - influences efficiency of vaporization/discrimination Temperature - hot enough to vaporize sample without degradation or causing backflash Injection volume - typically 1-3uL, increasing it does not have as much of an effect as one might think

Split Liner C 10 Packed with Glass Wool Peak Area Ratio n-c 40 /n-c 10 = 0.64 C 40 C 10 Without Glass Wool Packing Peak Area Ratio n-c 40 /n-c 10 = 0.37 C 40 Page 14

Split Injector - 200:1 vs 5:1 200:1 5:1 1 2 3 4 5 6 DB-1, 15 m x 0.25 mm I.D., 0.25 µm 60 C for 1 min, 60-180 C at 20 /min; Helium at 30 cm/sec 1. n-heptane 2. toluene 3. n-decane 4. n-butylbenzene 5. n-tridecane

Split Liners What s What? Straight tube Straight tube with glass wool Fixed glass wool Inverted cup Baffle

GLASS WOOL Placement in Liner Near top of liner: Wipes syringe needle of sample More thermal mass Aids in sample volatilization Can improve injector precision Helps to prevent backflash Near bottom of liner: Helps in volatilization of high MW components Increases mixing Page 17

Split Injector Injection Volume 2 1 µl 3 µl 4 5 2 4 5 1 3 1 3 1 2 3 4 5 6 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 60 C for 1 min, 60-180 C at 20 /min; Helium at 30 cm/sec 1. n-heptane 2. toluene 3. n-decane 4. n-butylbenzene 5. n-tridecane 1 2 3 4 5 6 Time (min.)

Splitless Injector Overview Most of the sample is introduced into the column Used for low concentration samples Wider peaks are obtained than for split injections

Splitless Injector Major Variables Purge activation time - determines amount of sample onto column and efficiency of injection (sensitivity vs peak shape) Liner - preventing backflash more critical than vaporization properties (double tapered type recommended) Injection volume - typically 1uL or less (backflash) Temperature long residence times allow for lower temps

Splitless Injector Purge Activation Time 0.5 min 0.75 min 2 3 4 2 3 4 1 1 2 4 6 8 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 60 C for 1 min, 60-180 C at 20 /min; Helium at 30 cm/sec 1. n-decane 2. n-dodecane 3. n-tetradecane 4. n-hexadecane 2 4 6 8 Time (min.)

Splitless Injection Liners Liner Part No. Comments Side hole 5181-3316 5062-3587 5181-3315 G1544-80730 G1544-80700 Single taper, deactivated, 900 L volume. Taper isolates sample from metal seal, reducing breakdown of compounds that are active with metals. For trace samples, general application. Single taper, deactivated, with glass wool, 900 L volume. Glass wool aides volatilization and protects column. For trace (dirty) samples. Double taper, deactivated, 800 L volume. Taper on inlet reduces chance for backflash into carrier gas lines. High efficiency liner for trace, active samples. Direct connect liners, single and dual taper, deactivated. Capillary column press fits into liner end, eliminating sample exposure to inlet. Ultimate protection for trace, active samples. Side hole permits use with EPC. Page 22

Splitless Injector Injection Volume 1 µl 3 µl 2 3 4 1 2 3 4 1 2 4 6 8 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 60 C for 1 min, 60-180 C at 20 /min; Helium at 30 cm/sec 1. n-decane 2. n-dodecane 3. n-tetradecane 4. n-hexadecane 2 4 6 8 Time (min.)

Splitless Injector Injector Temperature 200 C 250 C 1 2 3 4 5 6 1 2 3 4 5 6 0 2 4 6 8 10 12 14 Time (min.) 0 2 4 6 8 10 12 14 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C for 0.5 min, 50-325 C at 20 /min; Helium at 30 cm/sec Phthalates: 1. dimethyl 2. diethyl 3. dibutyl 4. benzylbutyl 5.bis(2-ethylhexyl) 6. dioctyl

Splitless Injector Sample Re-focusing Sample re-focusing improves efficiency Use low column temperature to refocus solvent - called the solvent effect Use cold trapping

Splitless Injector Solvent Effect Initial column temperature at least 10 C below sample solvent boiling point Required to obtain good peak shapes unless cold trapping occurs Rule of thumb, if solute BP >150 C above initial column temperature, the solute will cold trap Cold trapping has greater efficiency than solvent effect 1. 2. 3. 4. Solvent and solutes Solvent film

Splitless Injector Initial Column Temperature Hexane Solvent (BP = 68-69 C) 70 C 3 4 50 C Solvent Effect Cold Trapping 2 3 4 2 1 1 2 4 6 8 Time (min.) 2 4 6 8 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C or 70 C for 0.5 min, to 210 C at 20 /min; Helium at 30 cm/sec 1. n-decane 2. n-dodecane 3. n-tetradecane 4. n-hexadecane

Splitless Injector Reverse Solvent Effect Hexane 5 6 Methanol 4 1 4 2 5 6 2 3 1 3 0 1 2 3 4 5 6 Time (min.) 1 2 3 4 5 6 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C for 1 min, 50-210 C at 20 /min; Helium at 30 cm/sec 1. 1,3-DCP 2. 3-hexanol 3. butyl acetate 4. 1-heptanol 5. 3-octanone 6. 1,2-dichlorobenzene

Retention Gap Also Called A Guard Column Injector Detector Deactivated Fused Silica Tubing Union Column Usually 2-10 meters long and same diameter as the column (or larger if needed)

Splitless Injector 3 m x 0.25 mm I.D. Retention Gap Hexane 1 Methanol 4 5 6 2 3 4 1 2 3 5 6 0 1 2 3 4 5 6 Time (min.) 1 2 3 4 5 6 Time (min.) DB-1, 15 m x 0.25 mm I.D., 0.25 µm 50 C for 1 min, 50-210 C at 20 /min; Helium at 30 cm/sec 1. 1,3-DCP 2. 3-hexanol 3. butyl acetate 4. 1-heptanol 5. 3-octanone 6. 1,2-dichlorobenzene

COMPOUND REQUIREMENTS FOR GC Only 10-20% of all compounds are suitable for GC analysis The compounds must have: Sufficient volatility Thermal stability NO Inorganic Acids and Bases Be mindful of salts! Page 31 Month ##, 200X Group/Presentation Title

Typical Capillary Column Polyimide Coating Fused Silica Stationary Phase Expanded view of capillary tubing Page 32

SEPARATION PROCESS 1 3 2 4 Page 33

TWO PHASES Mobile Phase Stationary Phase Solute molecules distribute into the two phases Page 34

DISTRIBUTION CONSTANT (K C ) Mobile Phase Stationary Phase K C = conc. of solute in stationary phase conc. of solute in mobile phase K C formerly written as K D Page 35

SOLUTE LOCATION In stationary phase = Not moving down the column In mobile phase = Moving down the column Page 36

SEPARATION PROCESS Movement Down the Column Mobile phase A A Stationary phase B 1 B 2 B B A 3 A 4 Page 37

KC AND RETENTION Fused Silica Tubing Stationary Phase Gas Flow Stationary Phase K c => Large retention K c => Small retention Page 38

KC AND PEAK WIDTH Time of Elution Fused Silica Tubing Gas Flow Stationary Phase Stationary Phase K c => Large K c => Small Page 39

THREE PARAMETERS THAT AFFECT K C Solute: different solubilities in a stationary phase Stationary phase: different solubilities of a solute Temperature: K C decreases as temperature increases Page 40

RETENTION TIME t r 1.25 4.41 Time for a solute to travel through the column Page 41

ADJUSTED RETENTION TIME t r ' Actual time the solute spends in the stationary phase t r ' = t r - t m t r = retention time t m = retention time of a non-retained solute Page 42

ADJUSTED RETENTION TIME tr 1.25 4.41 t m t r t r = tr - tm t r = 4.41-1.25 t r = 3.16 min = time spent in stationary phase Page 43

TIME IN THE MOBILE PHASE All solutes spend the same amount of time in the mobile phase. Page 44

RETENTION FACTOR (k) Ratio of the time the solute spends in the stationary and mobile phases k = t r - t m t m t r = retention time t m = retention time of non-retained compound Formerly called partition ratio; k' Page 45

RETENTION FACTOR (k) Relative retention Linear Factors out carrier gas influence Page 46

PHASE RATIO ( ) = r 2d f r = radius (µm) d f = film thickness (µm) Page 47

DISTRIBUTION CONSTANT (Kc) K c = kβ k = t r t m = r 2d f Page 48

RANGE OF RETENTION Average Fastest Slowest Time Page 49

PEAK SYMMETRY Symmetry = A B A B 10% height Tailing : Symmetry <1 Fronting : Symmetry >1 Page 50

PEAK WIDTH Peak width at half height Half height Peak width at base Page 51

PEAK WIDTH Page 52

EFFICIENCY Theoretical Plates (N) Large number implies a better column Often a measure of column quality Relationship between retention time and width Page 53

THEORETICAL PLATES (N) N = 5.545 t r W h 2 t r = retention time W h = peak width at half height (time) Page 54

EFFICIENCY MEASUREMENT Cautions Actually, measurement of the GC system Condition dependent Use a peak with k>5 Page 55

ISOTHERMAL VS. TEMPERATURE PROGRAMMING Efficiency 9.31 N = 104,000 100 C isothermal 0 2 4 6 8 10 9.50 N = 433,200 75-135 C at 5 /min 0 2 4 6 8 10 DB-1, 30 m x 0.25 mm ID, 0.25 um He at 37 cm/sec C10, C11, C12 Page 56

SEPARATION VS. RESOLUTION Separation: time between peaks Resolution: time between the peaks while considering peak widths Page 57

SEPARATION FACTOR ( = k 2 k 1 co-elution: = 1 k 2 = retention factor of 2nd peak k 1 = retention factor of 1st peak Page 58

RESOLUTION (Rs) R = 1.18 s t r2 W h1 - t r1 + W h2 t r = retention time W h = peak width at half height (time) Page 59

RESOLUTION Baseline Resolution: Rs = 1.5 10.59 10.77 10.59 10.77 10.59 10.83 W = 0.105 h R = 0.84 % = 50 W h = 0.059 R = 1.50 % = 100 W h = 0.059 R = 2.40 % = 100 Page 60

Resolution R s = N k 4 k 1 1 N k = Theoretical plates = Retention factor = Separation factor Page 61

INFLUENCING RESOLUTION Variables: N: column dimensions, carrier gas a: stationary phase, temperature k: stationary phase, temperature, column dimensions Page 62

DETECTORS Purpose: Responds to some property of the solutes Converts the interaction into a signal Immediate Predictable Page 63

Detectors Detector Dynamic Range MDL TCD 10 5 Universal 400 pg Tridecane FID 10 7 Responds to C-H bonds 1.8 pg Tridecane ECD 5x10 5 Responds to free electrons 6 fg/ml Lindane NPD 10 5 Specific to N or P 0.4 pgn/s 0.06 pg P /s FPD 10 3 S, 10 4 P Specific to S or P 60 fg P/s 3.6 pg S/s SCD 10 4 Specific & Selective to S 0.5 pg S/s NCD 10 4 Specific & Selective to N 3 pg N/s MSD Universal S/N 400:1 1 pg/ul OFN

DATA HANDLING Converts the detector signal into a chromatogram Integrator Software Program Page 65

Conclusions The GC is comprised of an inlet, column and detector that all work together to produce good chromatography Separation (via K c ) is based on 3 things: Solute: different solubilities/interaction in a given stationary phase Stationary phase: different solubilities/interaction of a solute (correct column selection is critical!) Temperature: K C decreases as temperature increases When in doubt, contact Agilent Technical Support! Page 66

Agilent J&W Scientific Technical Support 800-227-9770 (phone: US & Canada) * * Select option 3..3..1 866-422-5571 (fax) GC-Column-Support@agilent.com www.chem.agilent.com Page 67