Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

Similar documents
Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Methanol Extraction of high level soil samples by USEPA Method 8260C

Evaluation of a New Analytical Trap for Gasoline Range Organics Analysis

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Analytical Trap Comparison for USEPA Method 8260C

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Volatile Organic Compounds in Every Day Food

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

ANNE JUREK. Abstract: soils and are found. in can also with GC/MS. poster. in series. option for. There are problems. analytes. in methanol.

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Solid Phase Micro Extraction of Flavor Compounds in Beer

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

Amy Nutter, Applications Chemist; Teledyne Tekmar P a g e 1

Determination of Volatile Organic Compounds in Air

Optimizing. Abstract: is standardd. procedures. altered to

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Maximizing Sample Throughput In Purge And Trap Analysis

U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

Solid Phase Micro Extraction of Flavor Compounds in Beer

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Study of Residual Solvents in Various Matrices by Static Headspace

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Headspace Technology for GC and GC/MS: Features, benefits & applications

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

Optimizing Standard Preparation

GC/MS: A Valid Tool for Soil Gas Hydrocarbons Speciation

Analysis of BTEX in Natural Water with SPME

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Author. Abstract. Introduction

Maximizing Production While Minimizing Costs

USP <467> Headspace Residual Solvent Assay with a HT3 Headspace Instrument

Ed George and Anaïs Viven Varian, Inc.

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Determination of VOC in Artificial Runway by Multiple Headspace Extraction-GC/MS. Gas Chromatography/ Mass Spectrometry APPLICATION.

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

TDTS 16 Round-the-clock, on-line and cryogen-free monitoring of hydrocarbons from acetylene to trimethylbenzene in ambient air

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Applying the Technology of the TurboMatrix 650 ATD to the Analysis of Liquid Accelerants in Arson Investigation

EPA TO-17 Volatile Organic Compounds

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

U.S. EPA Method 8270 for multicomponent analyte determination

STANDARD OPERATING PROCEDURES

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Volatile Hydrocarbons (VH) in Waters by GC/FID

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

2017 EPA Method Update Rule and EPA Method 624.1

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Application Note. Abstract. Introduction. Determination of Polycyclic Aromatic Hydrocarbons in Seafood by an Automated QuEChERS Solution

Atlantic RBCA. Guidelines for Laboratories. Tier I and Tier II Petroleum Hydrocarbon Methods

Thermal Desorption Technical Support

THE NEW QUANTITATIVE ANALYTICAL METHOD FOR ULTRATRACE SULFUR COMPOUNDS IN NATURAL GAS

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

STANDARD OPERATING PROCEDURES

Automated Characterization of Compounds in Fire Debris Samples

TO-17 Extending the Hydrocarbon Range above Naphthalene for Soil Vapor and Air Samples Using Automated

Low-Level Sulfur Compounds in Beer by Purge and Trap with a Pulsed Flame Photometric Detector (PFPD)

United States EPA Method 415.3

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

USP<467> residual solvents

TPH Methods and Measurements and Petroleum Vapor Intrusion (PVI) Risks

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Application Note # Performance of Method 8270 Using Hydrogen Carrier Gas on SCION Bruker SCION GC-MS

Transcription:

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System Application Note Abstract The Massachusetts Department of Environmental Protection (MassDEP) developed the Method for the Determination of Volatile Petroleum Hydrocarbons (VPH) by Gas Chromatography/ Mass Spectrometry to measure the collective concentrations of volatile aliphatic and aromatic petroleum hydrocarbons in aqueous and soil/sediment matrices 1. The previous method for VPH determination required both Flame Ionization (FID) and Photoionization (PID) detectors, while this method utilizes Mass Spectrometry. This method uses Purge and Trap technology, which allows for the determination of volatile organic compounds (VOCs) at low concentration levels with precision and accuracy. The Teledyne Tekmar Atomx automated VOC sample prep system will be used to validate this method for both water and soil matrices. Calibration curves and detection limits will be established for the aliphatic and aromatic petroleum hydrocarbons listed in the MassDEP VPH method. Introduction Regulatory agencies, such as the Massachusetts DEP, develop analytical testing procedures to ensure clean air, water, and soil. Due to their potential impact on water and soil, petroleum hydrocarbon contamination from leaking underground storage tanks is a major concern. Many of these compounds, such as benzene and xylenes, have EPA-established Maximum Contaminant Levels (MCLs) due to their potential health effects 2. Studies have shown that acute exposure to high levels of many of these compounds can affect the central nervous system with symptoms such as lethargy, confusion, dizziness, and nausea. Chronic effects can also include effects on the liver, kidney, blood, and nervous system. With the potential risks of exposure to volatile petroleum hydrocarbons, it is important to have methods to analyze and monitor these compounds. The Method for the Determination of Volatile Petroleum Hydrocarbons by Gas Chromatography was developed by the Massachusetts DEP (MassDEP) to measure the collective concentrations of volatile aliphatic and aromatic hydrocarbons 1. Originally this analysis was performed using flame ionization (FID) as the preferred detection method, but interest has grown in utilizing a mass spectrometry. Table 1 lists and categorizes the compounds to be monitored using a GC/MS. This study employs the GC conditions outlined specifically in the draft method. Further evaluation may be required, pending any changes, once the method is finalized.

Analyte Analyte Type Retention Time Quantitation Ion Confirming Ions Pentafluorobenzene Internal Standard 5.202 96 70 Methyl tertiary butyl ether Target Analyte 3.255 73 45 n-pentane C 5 -C 8 Aliphatic 2.204 43 57, 72 n-hexane C 5 -C 8 Aliphatic 3.472 57 41, 43, 56 2,3-Dimethylpentane C 5 -C 8 Aliphatic 5.355 56 43, 57, 41 Cyclohexane C 5 -C 8 Aliphatic 5.743 56 84, 41 Benzene Target Analyte 6.436 78 52, 51 n-heptane C 5 -C 8 Aliphatic 9.101 43 71, 57, 100 n-octane C 5 -C 8 Aliphatic 9.103 43 85, 57, 71 Chlorobenzene-d5 Internal Standard 11.529 117 119, 82 Toluene Target Analyte 9.403 91 92 Toluene-D8 Surrogate Standard 9.279 98 100, 42 Ethylbenzene Target Analyte 11.672 91 106 m,p-xylene Target Analyte 11.784 91 106, 105 o-xylene Target Analyte 12.409 91 106, 105 1,4-Dichlorobenzene-d4 Internal Standard 14.586 152 115, 150 2,3-Dimethylheptane C 9 -C 12 Aliphatic 10.506 43 84, 85 n-nonane C 9 -C 12 Aliphatic 11.422 43 57, 85 Isopropylbenzene C 9 -C 10 Aromatic 12.906 105 120 n-decane C 9 -C 12 Aliphatic 13.289 57 43, 71, 85 1-Methyl-3-ethylbenzene C 9 -C 10 Aromatic 13.535 105 120 1,3,5-Trimethylbenzene C 9 -C 10 Aromatic 13.605 105 120 Butylcyclohexane C 9 -C 12 Aliphatic 14.168 83 55, 82 p-isopropyltoluene C 9 -C 10 Aromatic 14.407 119 105, 134 1,2,3-Trimethylbenzene C 9 -C 10 Aromatic 14.581 105 120 n-undecane C 9 -C 12 Aliphatic 14.646 57 43, 71, 85 n-dodecane C 9 -C 12 Aliphatic 15.727 57 43, 71, 85 Naphthalene Target Analyte 16.827 128 Table 1: MassDEP Volatile Petroleum Hydrocarbon Analytes This method has been designed to complement the toxicological approach to evaluate human health effects that may result from exposure to petroleum hydrocarbons. Table 2 lists the current Massachusetts regulatory limit on petroleum hydrocarbons.

Current Massachusetts Regulatory Limit C 5 -C 8 Aliphatic 0.3 mg/l C 9 -C 12 Aliphatic 0.7 mg/l C 9 -C 10 Aromatic 0.2 mg/l Table 2: Current Massachusetts Regulatory Limits for Petroleum Hydrocarbons These regulatory limits require instrumentation capable of analysis at low concentrations. Purge and Trap technology allows for the determination VOCs at these low levels with precision and accuracy. This study will employ a Teledyne Tekmar Atomx automated VOC sample prep system to validate the MassDEP GC-MS VPH method for both water and soil matrices. Soil samples are analyzed utilizing automated methanol extraction, exclusive to the Atomx, and quantified using a water-based calibration. Experimental-Instrument Conditions An Agilent 7890A GC and 5975C inert XL MSD with Triple Axis Detector (TAD) was used in this study. Tables 3 and 4 list the GC/MS conditions. An Atomx automated VOC sample prep system was employed using the purge and trap parameters outlined in Table 5 and a #9 analytical trap. GC Parameters MSD Parameters GC: Agilent 7890A MSD: 5975C inert XL with TAD Column Oven Program: Restek RTX-502.2 30m x 0.25 mmid x 1.4 µm 50 C for 5 min; 11 C/min to 130 C for 0 min; 20 C /min to 250 C for 1 min Source: 250 C Quad: 200 C Inlet: 240 C Solvent Delay: 0.5 min Column Flow 1.0 ml/min Scan Range: m/z 35-250 Gas: Helium Scans: 5.19 scans/sec Split: 50:1 Threshold: 150 Pressure: 12.401 psi MS Transfer Line Temp: 250 C Inlet: Split Tables 3 and 4: GC and MSD Parameters

Atomx Water Parameters Variable Value Variable Value Valve oven Temp 140 C Dry Purge Flow 100 ml/min Transfer Line Temp 140 C Dry Purge Temp 20 C Sample Mount Temp 90 C Methanol Needle Rinse Off Water Heater Temp 90 C Methanol Needle Rinse Volume 3.0 ml Sample Vial Temp 20 C Water Needle Rinse Volume 7.0 ml Sample Equilibrate Time 0.00 min Sweep Needle Time 0.25 min Soil Valve Temp 40 C Desorb Preheat Time 245 C Standby Flow 10 ml/min GC Start Signal Start of Desorb Purge Ready Temp 40 C Desorb Time 2.00 min Condensate Ready Temp 45 C Drain Flow 300 ml/min Presweep Time 0.25 min Desorb Temp 250 C Prime Sample Fill Volume 3.0 ml Methanol Glass rinse On Sample Volume 5.0 ml Number of Methanol Glass Rinses 1 Sweep Sample Time 0.25 min Methanol Glass Rinse Volume 7.0 ml Sweep Sample Flow 100 ml/min Number of Bake Rinses 1 Sparger Vessel Heater Off Water Bake Rinse Volume 7.0 ml Sparge Vessel Temp 20 C Bake Rinse Sweep Time 0.25 min Prepurge Time 0.00 min Bake Rinse Sweep Flow 100 ml/min Prepurge Flow 0 ml/min Bake Rinse Drain Time 0.40 min Purge Time 11.00 min Bake Time 7.00 min Purge Flow 40 ml/min Bake Flow 200 ml/min Purge Temp 20 C Bake Temp 280 C Condensate Purge Temp 20 C Condensate Bake Temp 200 C Dry Purge Time 0.00 min Table 5: Atomx Parameters (parameters highlighted in yellow were not used.) Calibration Data and Method Detection Limits A stock solution in was prepared at 100 ppm in methanol. From this stock, an eight-point calibration from 1-200 ppb was prepared. According to this MassDEP VPH method, evaluation using Relative Response Factors (RRFs) is preferred, but linear regression is permissible. For this study, linear regressions were used to evaluate the data and determine the Method Detection Limit (MDL). Agilent ChemStation software was used to process the calibration data. Figure 1 is a total ion chromatogram (TIC) of a 10 ppb VPH standard. Figure 2 shows the calibration curve for ethylbenzene, a target analyte from the MassDEP VPH method, as an example. According to the method, a passing calibration has a linear regression factor of 0.995 or greater. Calibration data and MDLs for all compounds can be found in Table 6.

Figure 1: Chromatogram of a 10 ppb standard for MassDEP VPH by GC/MS method. r 2 =0.999762 Figure 2: Calibration curve for Ethylbenzene, a target analyte for the VPH by GC/MS method.

Analyte Group Linear Regression Factor (r 2 ) MDL* Total C5-C8 Aliphatic 0.9965 0.23 Total C9-C12 Aliphatic 0.9982 0.20 Total C9-C10 Aromatic 0.9965 0.20 Target Analytes Methyl Tert-butyl Ether 0.9997 0.18 Benzene 0.9994 0.11 Toluene 0.9993 0.15 Ethylbenzene 0.9998 0.12 m,p-xylenes 0.9993 0.11 o-xylene 0.9992 0.12 Naphthalene 0.9987 0.07 Table 6: Calibration and MDL Results for MassDEP VPH Analytes (*Analysis at 2 ppb, n=7) The Minimum Detection Limits (MDL) for this study were determined by analyzing seven samples at 2 ppb and using the equation below to obtain a value. MDL= (Standard Deviation)*(Student s t value for 99% confidence level) T value for 7 replicates = 3.14 Conclusions Studies have shown that exposure to high levels of petroleum hydrocarbons can have negative effects on human health. With these implications in mind, it is important to have a method to detect these compounds at low levels. The Massachusetts Department of Environmental Protection developed a method to measure volatile petroleum hydrocarbons using FID. Using Teledyne Tekmar s Atomx automated VOC sample prep system, the MassDEP VPH method has been successfully validated using GC/MS. Due to the draft status of this method, further evaluation may be required once it has been finalized. Calibration criteria of this method were met, with linear regression values greater than 0.995 for all compounds. Using the Atomx s automated methanol extraction for soil samples can also lead to considerable time savings. References 1) Method for the Determination of Volatile Petroleum Hydrocarbons (VPH) by Gas Chromatography/Mass Spectrometry, Massachusetts Department of Environmental Protection, February 2012. 2) Drinking Water Contaminants, United States Environmental Protection Agency, http://water.epa.gov/drink/contaminants/index.cfm