Optimizing Peak Capacity in Nanoscale Trap-Elute Peptide Separations with Differential Column Heating

Similar documents
UPLC Intact MASS Analysis Application Kit

Routine MS Detection for USP Chromatographic Methods

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

The Use of the ACQUITY QDa Detector for a Selective, Sensitive, and Robust Quantitative Method for a Potential Genotoxic Impurity

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Analysis of Serum 17-Hydroxyprogesterone, Androstenedione, and Cortisol by UPLC-MS/MS for Clinical Research

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

Electron Transfer Dissociation of N-linked Glycopeptides from a Recombinant mab Using SYNAPT G2-S HDMS

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

[application note] ACQUITY UPLC/SQD ANALYSIS OF POLYMER ADDITIVES. Peter J. Lee, and Alice J. Di Gioia, Waters Corporation, Milford, MA, U.S.A.

Analysis of Labeled and Non-Labeled Proteomic Data Using Progenesis QI for Proteomics

Conformational Characterization of Calmodulin by Hydrogen Deuterium Exchange Mass Spectrometry

UPC 2 Strategy for Scaling from Analytical to Preparative SFC Separations

Jonathan P. Danaceau, Erin E. Chambers, and Kenneth J. Fountain Waters Corporation, Milford, MA USA APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Effects of Aqueous Sample Content and Aqueous Co-Solvent Composition on UPC 2 Separation Performance

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

LC Technical Information

Online Reaction Monitoring of In-Process Manufacturing Samples by UPLC

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

Traditional Herbal Medicine Structural Elucidation using SYNAPT HDMS

Basic Principles for Purification Using Supercritical Fluid Chromatography

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Successfully Scaling and Transferring HPLC and UPLC Methods

C HA R AC T E RIZ AT IO N O F INK J E T P RINT E R C A RT RIDG E INK S USING A CHEMOMETRIC APPROACH

APPLICATION BENEFITS INTRODUCTION WATERS SOLUTIONS KEY WORDS. Simultaneous high-throughput screening of melamine (MEL) and cyanuric acid (CYA).

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

POST-COLUMN ADDITION AS A TOOL TO ENHANCE PERFORMANCE IN MICROFLOW LC-MS

Improved Extraction of THC and its Metabolites from Oral Fluid Using Oasis PRiME HLB Solid Phase Extraction (SPE) and a UPLC CORTECS C 18

DEMONSTRATING SUPERIOR LINEARITY: THE ACQUITY UPLC PHOTODIODE ARRAY DETECTOR

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

ANALYSIS OF BASIC COMPOUNDS

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

penta-hilic UHPLC COLUMNS

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Peptide Isolation Using the Prep 150 LC System

Performing Peptide Bioanalysis Using High Resolution Mass Spectrometry with Target Enhancement MRM Acquisition

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

A Case of Pesticide Poisoning: The Use of a Broad-Scope Tof Screening Approach in Wildlife Protection

Biochemistry. Biochemical Techniques HPLC

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Chemistry Instrumental Analysis Lecture 37. Chem 4631

PosterREPRINT AN INTERACTIVE PHYSICOCHEMICAL PROPERTY PROFILING SOFTWARE FOR EARLY CANDIDATE ANALYSIS IN DRUG DISCOVERY INTRODUCTION

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

penta-hilic UHPLC COLUMNS

ImprovIng ADmE ScrEEnIng productivity In Drug DIScovEry

Improved 6- Plex TMT Quantification Throughput Using a Linear Ion Trap HCD MS 3 Scan Jane M. Liu, 1,2 * Michael J. Sweredoski, 2 Sonja Hess 2 *

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Thermo Finnigan LTQ. Specifications

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

Analysis of Organic Light Emitting Diode Materials by UltraPerformance Convergence Chromatography Coupled with Mass Spectrometry (UPC 2 /MS)

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Multi-stage Mass Spectrometry up to MS 4 on a QTof System

Impact of Instrument Dispersion on Performance of HPLC Capillary Columns

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

Analysis of Pharmaceuticals and Personal Care Products in River Water Samples by UHPLC-TOF

Sensitive and Repeatable Analysis of Pesticides in QuEChERS Extracts with APGC-MS/MS

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

HPLC Praktikum Skript

Live Webinar : How to be more Successful with your ACQUITY QDa Detector?

Toxicity, Teratogenic and Estrogenic Effects of Bisphenol A and its Alternative. Replacements Bisphenol S, Bisphenol F and Bisphenol AF in Zebrafish.

Extraction of Methylmalonic Acid from Serum Using ISOLUTE. SAX Prior to LC-MS/MS Analysis

Identification of Human Hemoglobin Protein Variants Using Electrospray Ionization-Electron Transfer Dissociation Mass Spectrometry

Bruker Daltonics. EASY-nLC. Tailored HPLC for nano-lc-ms Proteomics. Nano-HPLC. think forward

Reports of the Death of Fully Porous Particles are Greatly Exaggerated

2007 Waters Corporation. Application of UPLC to Peptide Mapping

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Optimization of the sample preparation and extraction methodology

ultimate efficiency unleashed [ Cortecs 1.6 µm columns ]

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Hints and Tips for APGC Coupled with Xevo TQ-S for Food and Environmental Analysis

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

How proteins separate on reverse-phase HPLC

BEST PRACTICES IN THE ANALYSIS OF RAPIFLUOR-MS LABELED GLYCANS USING THE ACQUITY QDa DETECTOR (PERFORMANCE MODEL)

A S ENSIT IV E M E T HO D FO R T H E D E T E RM INAT IO N O F ENDO C RIN E- D IS RU P T ING COM P OUNDS IN RIV E R WAT E R BY L C / MS/MS

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

[ a ppl ic at ion no t e ]

Fast Protein and Peptide Separations Using Monolithic Nanocolumns and Capillary Columns

High Pressure/Performance Liquid Chromatography (HPLC)

Bringing uhplc Performance to the Separation of Peptides

Q-TOF PREMIER DYNAMIC RANGE ENHANCEMENT AND ACCURATE MASS MEASUREMENT PERFORMANCE

Breaking the speed limit: Fast, high-resolution peptide and tryptic digest separations using fused-core particles

Impact of Pore Exclusion on Reversed-Phase HPLC Column Performance

INTRODUCTION OVERVIEW

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Direct Analysis of Opioids and Metabolites from Whole Blood Using Ostro Sample Preparation Plates Combined with UPLC-MS/MS for Forensic Toxicology

LC and LC/MS Column Selection Flow Chart

APPLICATION COMPENDIUM

AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Transcription:

Optimizing Peak Capacity in Nanoscale Trap-Elute Peptide Separations with Differential Column Heating Matthew A. Lauber, Stephan M. Koza, and Kenneth J. Fountain Waters Corporation, Maple Street, Milford, MA, USA APPLICATION BENEFITS Optimized analyte focusing through differential trap and analytical column heating Up to 80% improvement in the peak capacity of trap-elute nanoscale peptide separations through use of differential heating and reverse elute trapping Exemplary loadability with reverse-elute trapping and a CSH C 18 analytical column provides an analyst the capability to more readily detect low abundance species WATERS SOLUTIONS ACQUITY UPLC M-Class System ACQUITY UPLC M-Class CSH C 18 Column MassPREP Enolase Digest with Phosphopeptides Peptide Retention Standard SYNAPT G2-S Xevo G2 QTof MassLynx Software KEY WORDS ACQUITY UPLC M-Class System, CSH C 18, Symmetry C 18, peptides, nanolc, trap-elute, phosphopeptides INTRODUCTION Nanoscale LC enables the analysis of minimal sample quantities with comparatively higher mass spectrometric sensitivity versus higher flow rate chromatographic separations, such as those achieved with narrow bore (2.1 mm I.D.) columns. 1-2 Not surprisingly, studies aiming to characterize the proteolytic digests of complex biological systems have been made possible by modern nanoscale LC techniques. Notable examples include the use of nanolc-ms to elucidate cellular mechanisms and to discover biomarkers. - With nanolc, however, it is generally not reasonable to analyze large volume samples using just a single column. Instead, it is preferable to perform online trapping of analytes at microscale flow rates and to subsequently elute and separate those analytes across an analytical column, wherein a significantly lower nanoscale flow rate is employed. Often, tens of microliters of sample must be analyzed, and in these samples, analytes may be present at concentrations differing by more than several orders of magnitude. Performing trapping on a trap column amenable to high flow rates allows for greater throughput efficiency as well as desalting in a step that diverts unwanted solutes away from a downstream mass spectrometer without adding additional connections and tubing after the analytical column (Figure 1). A trap column can also function as a guard column to preserve the performance of the analytical column. Beyond its role as a guard column, an effective trapping column is packed with larger diameter particles, typically µm, to enable higher velocity loading and retention of large, possibly dilute sample volumes. To be effective, however, a trap column should have lower retentivity when compared to the analytical column. This match between trapping and analytical columns during gradient elution ensures refocusing of analytes on to the analytical column and that high peak capacity separations are delivered to the MS detector. In this application note, we discuss the concepts to be considered to achieve high peak capacity nanoscale trap-elute peptide separations. A 1K psi capable, ACQUITY UPLC M-Class System is used to demonstrate high peak capacity trap-elute peptide separations with a charge surface hybrid (CSH) C 18 analytical column. 1

EXPERIMENTAL Method conditions (unless otherwise noted): For the study of peptide retention: LC conditions System: Detection: Wavelength: Scan rate: ACQUITY UPLC H-Class System Bio with a 20-cm Column Heater ACQUITY UPLC TUV Detector 21 nm Hz Columns: XSelect Peptide CSH C 18, Å,. µm, 2.1 x 0 mm (p/n 180090) XBridge Peptide BEH C 18, Å,. µm, 2.1 x 0 mm (p/n 1800) Symmetry C 18 Column, 0Å,. µm, 2.1 x 0 mm (p/n WAT2000) XSelect HSS T Column, 0Å,. µm, 2.1 x 0 mm (p/n 1800) Column temp.: 0 80 C Sample temp.: C Injection volume: µl Flow rate: Mobile phase A: Mobile phase B: Vials: 0.2 ml/min 0.1% formic acid (v/v), water 0.1% formic acid (v/v), acetonitrile (ACN) LCGC Certified Clear Glass 12 x 2 mm Screw Neck Qsert Vial (p/n 1800112C) Gradient: Time Flow rate (min) %A %B (ml/min) 0 0 0 0.2 0 0 0 0.2 2 90 0.2 90 0.2 0 0 0.2 0 0 0.2 MS conditions MS system: Xevo G2 QToF Ionization mode: ESI+ Analyzer mode: Resolution Capillary voltage:.0 kv Cone voltage: 2 V Source temp.: 0 C Desolvation temp.: 0 C Cone gas flow: 0.0 L/Hr Desolvation gas flow: 800 L/Hr Calibration: NaI, 2 µg/µl from 0 2000 m/z Acquisition: 0 1990 m/z, Hz scan rate Data management: MassLynx Software (v.1) For the development of nanoscale trap-elute separations: LC conditions System: ACQUITY UPLC M-Class System with a Trap Valve Manager (and heater) and a CH-A 20-cm Column Heater (p/n 1802) Analytical Column: ACQUITY UPLC M-Class CSH C 18, Å, 1.7 µm, 7 µm x mm (p/n 180077) Sample description For the study of peptide retention on C 18 stationary phases, MassPREP Peptide Mixture (p/n 180027) was mixed with Peptide Retention Standard (p/n 1800) to make a solution in 0.1% formic acid, water that was comprised of ~20 pmol/µl MassPREP peptide mixture and 170 pmol/µl Peptide Retention Standard. For the development of nanoscale trap-elute separations, MassPREP Enolase Digest with Phosphopeptides Mix (p/n 180028) was reconstituted in 200 µl of 0.1% (v/v) TFA in water to make a solution of pmol/µl. 2

Trapping column: ACQUITY UPLC M-Class Symmetry C 18, V/V, 0Å, µm, 180 µm x 20 mm (p/n 1800700) Analytical column temp.: 0-0 C Trapping column temp.: Ambient - 0 C Sample temp.: C Injection volume: Flowrate: Mobile phase A: Mobile phase B: Weak needle wash: Vials: Trapping: 0.2 µl (intermediate mass load), 2 µl (high mass load) 1 µl/min (trapping), 0. µl/min (analytical) 0.1% formic acid (v/v), water 0.1% formic acid (v/v), acetonitrile (ACN) 0.1% trifluoroacetic acid (TFA) (v/v), water LCGC Certified Clear Glass 12 x 2 mm Screw Neck Qsert Vial (p/n 1800112C) Trapping column effluent diverted to waste Time Flow Rate (min) %A %B (µl/min) 0 99. 0. 1 1 99. 0. 1 Analytical gradient: Trapping column in-line with analytical column Time Flow Rate (min) %A %B (µl/min) 0 99. 0. 0. 0 0.0 0.0 0. 2 1.0 8.0 0. 1.0 8.0 0. 8 99. 0. 0. 90 99. 0. 0. MS conditions MS system: SYNAPT G2-S (with NanoLockSpray ion source) Ionization mode: ESI+ Analyzer mode: Resolution Capillary voltage:.0 kv Cone voltage: 0 V Source offset: 0 V Source temp.: 0 C Nanoflow gas pressure: 1. Bar Calibration: NaI, 2 µg/µl from 0 2000 m/z Acquisition: 0 1900 m/z, 2 Hz scan rate Data management: MassLynx Software (v.1)

RESULTS AND DISCUSSION Developing Trap-Elute Peptide Separations for CSH C 18 When used for low ph, MS-friendly reversed-phase peptide separations CSH C 18 can deliver ~2-fold greater peak capacities when compared to conventional C 18. -8 It has additionally been shown that the performance of CSH C 18 is largely related to its improved loadability.,7 Interestingly, the performance advantages of CSH C 18 can be attributed to its novel composition. CSH C 18 is bonded with both a trifunctional C 18 ligand and a controlled, low level coverage of basic residues. CSH C 18 consequently exhibits a positive surface potential at acidic ph. This surface charge, estimated to be about +20 mv, 9 repels similarly charged analytes, such as peptides at low ph, while minimizing undesired secondary interactions. But while the coloumbic repulsion between CSH C 18 and peptides improves peak capacity and loadability, this improvement comes with a small compromise to retentivity. Compared to uncharged silica and hybrid-silica C 18 columns, peptides tend to elute from CSH C 18 columns in 2-% less acetonitrile. -8 Employing CSH C 18 in a LC configuration involving just a single column simply requires adjustment of the initial gradient conditions to compensate for the reduced retentivity of the phase. In nanolc, the use of a single column is referred to as direct injection. Wherever possible, direct injection should be favored in nanolc as it is a simple configuration and the one most likely to produce the highest peak capacities. As mentioned before, it may nonetheless be preferable to perform nanolc with a trap-elute configuration, such as one of those portrayed in Figure 1. In trap-elute separations, phase retentivity becomes very important. It is necessary that a trapping stationary phase exhibit a lower retentivity than the analytical column phase so that eluting peptides can be optimally refocused on the analytical column. In that way, a large particle (i.e. µm) trapping phase can be used, keeping back pressures during trapping low, without detriment to the resolving power of the separation on the analytical column. A Waste Pin Plug 2 1 Analytical Column to nanospray emiiter ACQUITY UPLC M-Class System Dual Pump Reverse Flush Trapping Dual Temperature Control Syringe Injection Needle Valve 2 1 Sample Loop Trap Pin Plug Pin Plug Trapping Valve Single Pump Trapping Single Pump Trapping Dual Temperature Control TVM Heater TVM CH-A Heater BSM TVM CH-A Heater TVM Heater BSM SM SM B Injection Needle Valve Dual Pump Reverse Flush Trapping 2 1 BSM BSM/ASM Syringe 2 1 Sample Loop BSM 2 Waste Trapping Valve BSM 1 Analytical Column to nanospray emiiter *A second BSM or an ASM pump can be used for loading and trapping C Injection Needle Valve 2 1 Sample Loop Direct Injection Analytical Column to nanospray emiiter Syringe BSM Figure 1. Trap-elute fluidic configurations on an ACQUITY UPLC M-Class System. (A) Single pump trapping. (B) Dual pump reverse flush trapping. (C) Direct injection. (TVM = Trap Valve Manager, µsm = Micro Sample Manager, µbsm = Micro Binary Solvent Manager, ASM = Auxilliary Solvent Manager).

Recently, the combination of a 180 µm x 20 mm trap packed with µm Symmetry C 18 and a 7 µm I.D. analytical column packed with a C 18 stationary phase, known as high strength silica (HSS) T, was shown to be useful for nanoscale UPLC. -1 Symmetry C 18 is known to exhibit weak peptide retentivity, whereas HSS T exhibits a characteristically higher peptide retentivity. Trap-elute nanolc where Symmetry C 18 is combined with HSS T can thus be highly effective; that is, as long as low mass loads are injected. Looking to develop a trap-elute nanolc separation capable of high mass loading, we consequently explored the considerations that must be made in order to optimize a trap-elute peptide separation in which the less retentive CSH C 18 is used as the analytical phase instead of HSS T. Understanding the Peptide Retentivity of Common C 18 Stationary Phases The ability of various stationary phases to retain peptides was investigated using narrow bore (2.1 mm I.D.) columns packed with. µm diameter particles, a test mixture composed of 1 different peptides, an ACQUITY UPLC H-Class Bio System, and a Xevo G2 QTof mass spectrometer. Figure 2 presents three chromatograms obtained with three common analytical column phases, CSH C 18, BEH C 18, and HSS T. Many observations made previously during comparison of these phases are again observed in this work, namely differences in selectivity and loadability. -8 The objective of this analysis was to catalog the retentivity of these phases. To this end, the retentivity of the mentioned analytical phases is reported in Figure A in terms of the acetonitrile composition needed to elute each individual peptide. To obtain additional information about the retentivity of these phases, their dependence on temperature was studied, from 0 C up to 0 or 80 C depending on the maximum suggested operating temperatures of the various phases. These same experiments were performed to study the retentivity of the Symmetry C 18 phase, which is used in ACQUITY UPLC M-Class trapping columns (Figure A). Upon interpretation of these data, it becomes apparent that the retentivity of the phases can be ranked from low to high follows: Symmetry C 18 < CSH C 18 < BEH C 18 < HSS T. Moreover, it is clear that peptide retentivity decreases as column temperature increases. Figure B summarizes these observations as an average of the individual peptide retention measurements. A 21 A 21 A 21 1. 1.0 1 0. 7 12 8 11 1 2 0.0 0 1 20 2 0 1. 1.0 0. 0.0 0 1 20 2 0 1. CSH C 18 Å. m C BEH C 18 Å. m C 1 HSS T 0Å. m C 2 7 8 1.0 9, 0. 1 2 7 8 11 12 1 0.0 min 0 1 20 2 0 9 9 11 12 1 High Mass Load ~1700 pmoles ID Sequence 1 DGYGK DTVGYGVK DFVGYGVK 9 DFVGYGVKDFVGVGVK Intermediate Mass Load ~0-00 pmoles ID Sequence 2 RGDSPASSKP DRVYIHP RPPGFSPFR 7 DRVYIHPF 8 DRVYIHPFHL DRVYIHPFHLLVYS 11 WLTGPQLADLYHSLMK 12 YPIVSIEDPFAEDDWEAWSHFFK 1 GIGAVLKVLTTGLPALISWIKRKRQQ Figure 2. C 18 stationary phases and peptide retention. Separations of a peptide mixture (MassPREP Peptide Mixture and Peptide Retention Standard) were performed with an ACQUITY UPLC H-Class Bio System, 2.1 x 0 mm columns packed with various C 18,. µm stationary phases, and mobile phases modified with 0.1% formic acid.

These results corroborate the effectiveness of pairing Symmetry C 18 trapping with HSS T analytical separations. The two phases exhibit such different retentivities ( % ACN >%) that the analytical column can be operated over a wide range of temperatures independently of the trapping column. For instance, no matter the temperature at which the Symmetry C 18 trap column is maintained, peptides will elute from it at a substantially lower elution strength than that required to elute the peptides from HSS T. Peptides will consequently refocus on the head of the analytical column and elute as sharp detectable chromatographic peaks under a very broad range of experimental conditions. In contrast, the results shown in Figure suggest that methods involving the pairing of Symmetry C 18 trapping and CSH C 18 analytical separations need to be more carefully controlled, due to the highly similar retentivity characteristics of the two phases, in order to ensure optimal analytical column focusing. A GIGAVLKVLTTGLPALISWIKRKRQQ B 0C C YPIVSIEDPFAEDDWEAWSHFFK C C WLTGPQLADLYHSLMK 0C C DRVYIHPFHLLVYS DFVGYGVKDFVGVGVK DRVYIHPFHL DRVYIHPF RPPGFSPFR DFVGYGVK DRVYIHP DTVGYGVK RGDSPASSKP Temperature 80 0 0 C Light Dark HSS T BEH C18 CSH C18 Symmetry C18 %ACN to Elute Peptides (Average Peptides -12) 20 1 80C C DGYGK 0 20 0 0 %ACN to Elute Peptides Figure. Retentivity as a function of temperature. (A) Acetonitrile composition needed to elute different peptides from Symmetry C 18, CSH C 18, BEH C 18, and HSS T at temperatures ranging from 0 to 80 C. (B) Average acetonitrile composition needed to elute peptides 12 from Symmetry C 18, CSH C 18,BEH C 18, and HSS T at temperatures ranging from 0 to 80 C. All values are based on separations achieved with. µm diameter particle columns.

Use of differential heating to optimize analytical column focusing One way to optimize trap and analytical column retentivity is by independently controlling the temperature of the two devices. By adjusting phase retentivity through separately controlled column heaters, peptides can be induced to elute from a trapping column and re-focus on the analytical column. In standard configurations the ACQUITY UPLC M-Class System allows for a single heating zone. However, the system can add dual temperature control through addition of an ACQUITY Column Heater (CH-A, p/n 1802), as shown in Figure 1. With such a setup, the trap column can be housed in the CH-A compartment at a temperature that is different from the analytical column residing in the standard ACQUITY UPLC M-Class heater. Chromatographic separations of MassPREP enolase tryptic digest obtained with a Symmetry C 18 trapping column and a CSH C 18 analytical column in a typical trap-elute configuration (single-pump, forward elute, Figure 1A) are shown in Figure. In the example trap-elute separations, the analytical column was maintained at 0 C, while the trapping column was heated at increasing temperatures of 0 C, C, and 0 C. Based on the peak widths of enolase tryptic peptides, peak capacities were calculated for each scenario. A direct injection separation of this sample was found to yield a peak capacity (P c,0% height) of. Meanwhile, these trap-elute separations yielded peak capacities of 22, 02, and 9, respectively. This observation indicates that phases with similar retentivity can indeed be optimized for use in both trapping and analytical columns. Maintaining the trapping column at a higher temperature than the analytical column reduces its retentivity, resulting in peptides eluting from the trap at a lower concentration of acetonitrile. Since the eluting peptides are delivered at lower percentage acetonitrile, the peptides are readily retained and focused on the analytical column. In turn, peak capacities more closely resembling direct injection can be achieved, while maintaining the advantages of trap-elute, such as being able to load and desalt large volumes of sample quickly. T T1 Trapping: Symmetry C 18, m, 0 C Analytical: CSH C 18, 1.7 m, 0 C P c,0% = 22 (0% of Direct Injection) T2 T T27 T Trapping: Symmetry C 18, m, C Analytical: CSH C 18, 1.7 m, 0 C P c,0% = 02 (% of Direct Injection) T1 T2 T T27 T Trapping: Symmetry C 18, m, 0 C Analytical: CSH C 18, 1.7 m, 0 C P c,0% = 9 (77% of Direct Injection) T1 T2 T T27 min 1 20 2 0 0 0 0 Figure. Base peak intensity chromatograms from Nanoscale LC-MS of MassPREP Enolase Digest with Phosphopeptides. Single pump trap-elute fluidics along with a fixed analytical column temperature (0 C) and trapping column temperatures ranging from 0 to 0 C were employed. Separations were performed with an ACQUITY UPLC M-Class System, an ACQUITY UPLC M-Class Symmetry C 18 Trap Column, and a 7 µm x mm ACQUITY UPLC M-Class CSH C 18 Column. Peak capacities (P c,0% ) were calculated from the peak widths of T2, T, T1, T, and T27. A direct injection separation of the sample was found to produce a P c,0% of. 7

Fine tuning analytical column temperatures and employing reverse elute The ability to heat a Symmetry C18 trapping column up to 0 C is clearly a useful asset to the development of high peak capacity peptide separations. Results displayed in Figure demonstrate that the recoveries of weakly retained hydrophilic peptides are only marginally affected. Comparing the operation of the trapping column at 0 versus 0 C, only 1 or 2 peptides were lost due to reduced trap retentivity. Moreover, stability testing of Symmetry C18 indicates that there is only a minor retentivity shift, corresponding to -0.0% ACN/day, due to incubation at 0 C (Figure ). Every th Separation across 2 Days 0 C DFVGYGVKDFVGVGVK 0 to 0% ACN in 0 min Intensity A DFVGYGVK Avg Retention Shift -0.001 min/run -0.0% ACN/day DTVGYGVK DGYGK min Retention Time (min) B 0 1 Time (min) Time (min) 20 1 20 2 y = -0.001x + 1.98 R = 0.08 y = -0.001x + 11.7 R = 0.987 y = -0.0012x + 7.827 R = 0.2888 min 0 20 0 Run Run 0 Figure. Temperature stability of Symmetry C18. Separations of the Peptide Retention Standard performed at 0 C with an ACQUITY UPLC H-Class Bio System, 2.1 x 0 mm column packed with Symmetry C18,. µm and mobile phases modified with 0.1% formic acid. (A) Every th separation is displayed in the form of summed extracted ion chromatograms for the peptides of the mixture. (B) A plot of retention time versus run and corresponding linear fits. An estimation of retention loss per day is provided. Having established the value and feasibility of operating a Symmetry C18 trapping column at 0 C, it was important to optimize the temperature for best resolution of peptides from the CSH C18 analytical column. As before, peak capacities were measured based on the separation of an enolase digest using a single-pump, forward elute configuration. Unlike before, the trapping column was held at 0 C, while the CSH C18 analytical column was subjected to increasing temperatures, from 0 to 0 C. Peak capacities measured for these varying temperature settings are provided in Figure (blue bars). The trend in the data indicates that a 0 C Symmetry C18 trapping column pairs effectively with a CSH C18 analytical column operating between 0 and 0 C. Despite this optimization though, the peak capacities obtainable with a single-pump, forward elute trap-elute configuration remain less than those possible by direct injection. 8

An alternative to the single-pump, forward elute configuration, was thus investigated, specifically the configuration shown in Figure 1B. While this configuration is more complicated, since it relies on a two pump system, it has notable advantages. In particular, this configuration makes it possible to load the trapping column in one direction and to elute its contents in the reverse direction. This so-called reverse flush/elute configuration can be exploited to obtain peak capacities more representative of the full capabilities of an analytical column as observed in direct inject mode. In this mode, analytes will not pass through the entirety of the trapping column and will not be subjected to the band broadening imparted by the µm diameter particles and larger inner diameter hardware typical of a trapping column. Although outside of the scope of this work, it should also be noted that a two pump system such as this allows for the use of an ion pairing agent (e.g. TFA) to be used in the mobile phase of the trapping/loading pump. This has the effect to increase peptide retention/ recovery without hindering ionization of peptides since they will be subsequently eluted in MS-friendly formic acid mobile phases delivered by the second pump system. In this work with only formic acid mobile phases, the reverse-elute configuration proved to yield significantly higher peak capacities (red bars, Figure ). In fact, reverse-elute separations with a Symmetry C 18 trap heated to 0 C and a CSH C 18 analytical column maintained between 0 and 0 C were found to produce peak capacities within % difference of those obtained by direct injection. 700 0 Dual Pump Reverse-Flush (Fig 1B) Single Pump Trapping (Fig 1A) 00 0 00 Peak Capacity (0% Height) 0 00 0 00 20 200 0 0 0 0 C 0 C 0 C 0 C CSH C 18 Analytical Column Temperature (Symmetry C18 Trapping Temperature = 0 C) Figure. Peak capacities of trap-elute nanoscale separations using either a single pump trapping (blue) or dual pump reverse flush trapping configuration (red). Trapping was performed with an ACQUITY UPLC M-Class Symmetry C 18 Trap Column at 0 C and the analytical separation was performed across a ACQUITY UPLC M-Class CSH C 18, 7 µm x mm Column at temperatures ranging from 0 to 0 C. 9

Loadability of nanoscale reverse-elute peptide separations with CSH C 18 A primary motivation for developing methods suitable for CSH C 18 is to take advantage of its unique loadability. To highlight the utility of an optimized trap-elute CSH C 18 separation, we investigated loading more sample in a nanolc-ms analysis. This can be important when needing to extend the dynamic range of an analysis. Researchers may want to study a low abundance protein or peptide in the midst of a larger population of proteins and peptides: for example, phosphopeptides in a cellular or plasma matrix. In order to detect and quantify the very minimally expressed phosphopeptides, a large amount of the starting material must be loaded. It has been found that CSH C 18 can be used to obtain a comprehensive nanolc analysis of such samples, because it is capable of producing high peak capacities even at relatively high mass loads. A base peak intensity (BPI) chromatogram obtained for a picomole (pmole) load of an enolase tryptic digest, containing phosphopeptides illustrates the use of CSH C 18 in high mass load applications (Figure 7). It is apparent that this pmole load of the enolase digest yields a comprehensive representation of the different peptides in the mixture, including the diphosphorylated species, VNQIGTLpSEpSIK. The inset extracted ion chromatogram for VNQIGTLpSEpSIK on Figure 7 shows that signal-to-noise for this peptide is remarkably high. Moreover, the distribution of signals seen in Figure 7 correlate well to data obtained for this sample when using narrow bore (2.1 mm I.D.) column chromatography. 1 And since CSH C 18 was employed in this nanolc analysis, peak capacity decreased by only % when the mass load was increased from 1 pmole (Figure ) to pmoles, as shown in Figure 7. 9.0E+ 7.0E+ VNQIGTLpSEpSIK High Mass Load pmoles Tryptic Digest P c,0% = 0 (only a % decrease versus 1 pmole load) 0.0E+0 0 0.0E+0 min 1 20 2 0 0 0 0 Figure 7. High mass load of a dual pump reverse flush trap-elute nanoscale separation with an ACQUITY UPLC M-Class Symmetry C 18 Trap Column (0 C), and a 7 µm x mm ACQUITY UPLC M-Class CSH C 18 Column (0 C). Base peak intensity chromatogram of MassPREP Enolase Digest with Phosphopeptides from analysis of a high mass load ( pmoles). The inset shows an extracted ion chromatogram for VNQIGTLpSEpSIK to demonstrate that the diphosphorylated peptide species in this sample can be readily detected. Peak capacities were calculated from the peak widths of T2, T1, T1, T, and T27.

CONCLUSIONS The presented work defines the considerations that must be made in order to optimize trap-elute peptide separations, including the retentivity of stationary phases and the configuration with which a trap is paired with an analytical column. By employing the dual heating capabilities possible with the combined use of an ACQUITY UPLC M-Class System and a CH-A heater, we have developed high peak capacity trap-elute nanolc peptide separations using a novel CSH C 18 analytical column. This configuration has been optimized for use with a Symmetry C 18 trapping column by adjusting column temperatures to affect phase retentivity as needed to maximize re-focusing of peptide analytes on the analytical column. Along with a two pump reverse-flush, trap-elute configuration, very high peak capacities can be reliably obtained. More importantly, this method development ensures that the unique attributes of the CSH C 18 stationary phase, such as its loadability, can be readily exploited. Optimized, high sample load nanoscale trap-elute peptide separations with CSH C 18 can be used to extend the dynamic range of an analysis and to facilitate detection of low abundance species. References 1. Vissers, J. P., Recent developments in microcolumn liquid chromatography. J Chromatogr A 1999, 8 (1-2), 117-. 2. Meiring, H. D.; van der Heeft, E.; ten Hove, G. J.; de Jong, A. P. J. M., Nanoscale LC-MS(n): technical design and applications to peptide and protein analysis. J. Sep. Sci. 2002, (2), 7-8.. Paweletz, C. P.; Andersen, J. N.; Pollock, R.; Nagashima, K.; Hayashi, M. L.; Yu, S. U.; Guo, H.; Bobkova, E. V.; Xu, Z.; Northrup, A.; Blume-Jensen, P.; Hendrickson, R. C.; Chi, A., Identification of direct target engagement biomarkers for kinase-targeted therapeutics. PLoS One 2011, (), e29.. Andersen, J. N.; Sathyanarayanan, S.; Di Bacco, A.; Chi, A.; Zhang, T.; Chen, A. H.; Dolinski, B.; Kraus, M.; Roberts, B.; Arthur, W.; Klinghoffer, R. A.; Gargano, D.; Li, L.; Feldman, I.; Lynch, B.; Rush, J.; Hendrickson, R. C.; Blume-Jensen, P.; Paweletz, C. P., Pathway-based identification of biomarkers for targeted therapeutics: personalized oncology with PIK pathway inhibitors. Sci Transl Med 20, 2 (), ra.. Lauber, M. A.; Koza, S. M.; McCall, S. A.; Alden, B. A.; Iraneta, P. C.; Fountain, K. J., High-Resolution Peptide Mapping Separations with MS-Friendly Mobile Phases and Charge-Surface-Modified C 18. Anal Chem 201, 8 (1), 9-.. Lauber, M. A.; Koza, S. M.; Fountain, K. J., Increasing Peak Capacity in Reversed Phase Separations of Peptides with Charged Surface Hybrid (CSH) C 18 Columns. Waters Application Note 720008EN 201. 7. Lauber, M. A.; Koza, S. M.; Fountain, K. J., Charged Surface Hybrid C 18 for High Peak Capacity nanolc Peptide Separations. Waters Application Note 72000917EN 201. 8. Lauber, M. A.; Koza, S. M.; Fountain, K. J., Peptide Mapping and Small Protein Separations with Charged Surface Hybrid (CSH) C 18 and TFA-Free Mobile Phases. Waters Application Note 7200071EN 201. 9. Gritti, F.; Guiochon, G., Adsorption behaviors of neutral and ionizable compounds on hybrid stationary phases in the absence (BEH-C 18 ) and the presence (CSH-C 18 ) of immobile surface charges. J Chromatogr A 201, 1282, 8-71.. Brownridge, P. J.; Harman, V.; Cubbon, S.; Vissers, J. P. C.; Lawless, C.; Hubbard, C. J.; Beynon, R. J., Absolute Quantification of Yeast Kinases by LC/MS/MS using QconCAT and MRM Technologies. Waters Application Note 7200092EN 201. 11. Schenauer, M. R.; Flynn, G. C.; Goetze, A. M., Identification and quantification of host cell protein impurities in biotherapeutics using mass spectrometry. Anal Biochem 2012, 28 (2), -7. 12. Schenauer, M. R.; Flynn, G. C.; Goetze, A. M., Profiling the effects of process changes on residual host cell proteins in biotherapeutics by mass spectrometry. Biotechnol Prog 201, 29 (), 91-97. 1. Gao, X.; Zhang, Q.; Meng, D.; Isaac, G.; Zhao, R.; Fillmore, T. L.; Chu, R. K.; Zhou, J.; Tang, K.; Hu, Z.; Moore, R. J.; Smith, R. D.; Katze, M. G.; Metz, T. O., A reversed-phase capillary ultra-performance liquid chromatography-mass spectrometry (UPLC-MS) method for comprehensive top-down/bottom-up lipid profiling. Anal Bioanal Chem 2012, 02 (9), 292-. 1. Pizzatti, L.; Panis, C.; Lemos, G.; Rocha, M.; Cecchini, R.; Souza, G. H.; Abdelhay, E., Label-free MS E proteomic analysis of chronic myeloid leukemia bone marrow plasma: disclosing new insights from therapy resistance. Proteomics 2012, 12 (17), 218-1. 1. MassPREP Enolase Digest with Phosphopeptides Mix. Waters Care and Use Manual 70171 2008. Waters, The Science of What s Possible, ACQUITY UPLC, SYNAPT, Xevo, XSelect, XBridge, Symmetry, MassLynx, and UPLC are registered trademarks of Waters Corporation. CSH, MassPREP, and NanoLockSpray are trademarks of Waters Corporation. All other trademarks are the property of their respective owners. 201 Waters Corporation. Produced in the U.S.A. May 201 7200007EN AG-PDF Waters Corporation Maple Street Milford, MA 0177 U.S.A. T: 1 08 78 2000 F: 1 08 872 1990 www.waters.com