The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Similar documents
Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Application. Gas Chromatography February Introduction

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Application Note. Abstract. Authors. Introduction

Author. Abstract. Introduction

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Introduction to Capillary GC

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

Introduction to Capillary GC

Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

GAS CHROMATOGRAPHY (GC)

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Understanding Gas Chromatography

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Chapter 31 Gas Chromatography. Carrier Gas System

Agilent G2350A Atomic Emission Detector (AED)

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Activity in the FID Detection Port: A Big Problem if Underestimated

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Agilent Intuvo 9000 Gas Chromatograph System

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Sulfotepp impurities in Chlorpyrifos EC formulations

1.0 Reactive Gas Chromatography Design and Setup Procedure for Catalyst Pelletization & Inlet Packing Software and Method.

Application. Gas Chromatography March 1998

STANDARD OPERATING PROCEDURES

Optimizing GC Parameters for Faster Separations with Conventional Instrumentation

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Fast and Accurate EPA 8270 Quantitation Using Deconvolution

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Application Note. Abstract. Authors. Introduction. Food Safety

Application Note. Abstract. Authors. Environmental

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

CHAPTER 6 GAS CHROMATOGRAPHY

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

A New Web-Based Application for Modeling Gas Chromatographic Separations

GC Analysis of Polybrominated Flame Retardants Application

Highly Sensitive and Rugged GC/MS/MS Tool

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Application Note 032

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

VOC Analysis of Water-Based Coatings by Headspace-Gas Chromatography

Gas Chromatography (GC)! Environmental Organic Chemistry CEE-PUBH Analysis Topic 5

U.S. EPA Method 8270 for multicomponent analyte determination

A Comparative Analysis of Fuel Oxygenates in Soil by Dynamic and Static Headspace Utilizing the HT3 TM Automatic Headspace Analyzer

Uncontrolled Copy. SOP-060 Chromatography, Tune Evaluation and Troubleshooting. Table of Contents. 1. Principle... 3

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

Determination of Volatile Substances Proof of Food Adulteration

A New Web-Based Application for Modeling Gas Chromatographic Separations. Dr. Hansjoerg Majer Restek Corporation Market Development Manager Europe

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Gas Chromatography. Introduction

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

Analysis of BTEX in Natural Water with SPME

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Trace analysis of mesityl oxide and diacetone alcohol in pharmaceuticals by capillary gas chromatography with flame ionization detection

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

Fast and Accurate Analysis of PBDEs in a Single Run, Including

Agilent G3212 GC-APCI Source

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

USP<467> residual solvents

Transcription:

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note Abstract Area and retention time are the two primary measurements in gas chromatography. The precision with which they are measured ultimately determines the validity of the generated quantitative data. Key parameters affecting their precision are discussed. Introduction Two measurements govern all chromatographic data: retention time (t R ) and peak area. Precise t R measurement is the mainstay of all modern gas chromatographs. Increasing t R precision leads to increased confidence in qualitative reporting as well as an increased use of automation in quantitative analysis. The importance of t R precision takes on greater significance as chromatographic efficiency increases (the use of small diameter capillary columns) and/or analysis time decreases (fast chromatography). Both of these conditions result in a potential for a large number of eluents per unit time. A lack of precision in t R can easily lead to misidentifications, resulting in gross errors by matching the calibration data with the wrong analyte. In gas chromatography (GC), t R precision is controlled by both carrier flow and oven temperature precision. Secondary factors include auto-sampler speed and the precision of the start run signal. The Agilent 6890N is capable of t R precision better than 0.0008 minutes (0.048 seconds). With measurements this precise, small changes in ambient operating conditions (laboratory temperature and pressure) can dramatically affect retention precision. For this reason, the Agilent 6890N uses ambient pressure and temperature compensation. Area precision ultimately determines the degree of accuracy of a quantitative measurement. While not directly related to t R, area precision can be affected by many of the same variables affecting t R. Precise flow control through a detector is critical for the maintenance of precise responses. The temperature of the detector also plays a part in determining some detector responses; the electron capture detector (ECD) or thermal conductivity detector (TCD) are more sensitive to temperature than others. Other factors influencing area repeatability are auto-sampler draw precision, speed of injection, and syringe performance. For active compounds, the inertness of the entire flow path can dramatically affect area precision, especially at low levels. In short, precision of both t R and area in GC are influenced by many parameters. Some of the parameters are more obvious than others. Retention time and area precision, compatable with highefficiency and/or fast chromatography, require that all parameters (temperature, pressure, and flow) are controlled to exacting tolerances.

6890N Area and t R Precision Specification t R Precision The t R precision of the Agilent 6890N GC is specified at better than 0.0008 minute or better than 0.008% RSD (relative standard deviation). The or between the two specifications is a true Boolean, meaning that one of the two conditions will always hold, and in most cases both will hold. These specifications were determined using the conditions listed in Table 1. In order to meet these specifications the instrument must be powered up, with flows, inlets, and detectors at temperature for at least 2 hours. The column is conditioned and is cleaned of any residual contaminants by performing at least one blank run (with no sample injection) immediately prior to test. No further equilibration of the instrument is necessary to meet these specifications. The resulting chromatogram is shown in Figure 1. Tetradecane (2 ng on-column was used for both the t R and area precision specifications. Table 1. Gas Chromatograph Conditions Agilent 6890N GC with 7683N ALS 5-µL syringe: p/n 5181-1273 Crimp-cap vials: p/n 5181-1210 and crimp caps (p/n 5181-3375) Inlet EPC: Split/Splitless Mode splitless: 1.0 µl injection Inlet temp: 250 C Purge flow: 40.0 ml/min Purge time: 0.5 min Constant flow: 6.5 ml/min Gas saver: Off Carrier gas: Helium Detector FID Temperature: 300 ºC H 2 flow: 30 ml/min Air flow: 400 ml/min Makeup flow: 25 ml/min Flame on offset: <15 pa Makeup gas: N 2 Inlet Liner Agilent multi-purpose inlet liner p/n 5183-4647 Septum: p/n 5183-4757 Oven 120 V or 240 V Oven ramp C/min Next C Hold min Initial 75 0.5 Ramp 1 20 190 0.00 Total run time 6.25 min Equilibration time 1.0 min Column Agilent HP-5.625, p/n 19091J-413, 30 m 0.32 mm, 0.25 µm Agilent ChemStation (32-bit) Signal 1 set to FID at a data rate of 5 Hz Integration Events Table Initial slope sensitivity: 5 Initial peak width: 0.02 Initial area reject: 5 Initial height reject: 5 Initial shoulders: Drop Sequence At least one blank run Run sample six times 2

Figure 1. Test chromatogram. The instrument conditions chosen are generic, meaning they are representative of many different analysis methods. Great care was also taken not to use conditions or procedures that would not be encountered in the normal day-to-day operation of a GC. Chromatography, like many measurement techniques, is highly dependent upon the conditions chosen by the operator. The conditions and results of the specifications are representative of what can be done by an average operator performing a typical method. How Precise Do GC Measurements Need To Be? As we move towards faster chromatography, peaks get narrower and the need for better t R control increases. We know from theory that tr = tm(1 + k) where: t r is the t R t m is the t R of an un-retained peak k is the partition ratio t m is a function of the average linear velocity µ L is the column length L t m = ì Now µ is a function of the pressure drop across the column and the viscosity of the gas, which has a temperature dependence. So for a given column, µ will change with changes in both temperature and pressure. k is a thermodynamic function with dependence on the film thickness (capillary columns) and the temperature of the column. So to achieve good t R repeatability, we must control column temperature and linear velocity. Changes in film thickness will also effect this, but these changes are usually slow, unidirectional (the film gets thinner), and are not an instrument parameter subject to our control other than an initial choice of film thickness in the column. The need of accurate t R precision is a function of peak width and resolution. The primary use of t R is to identify peaks. If the t R of two peaks move so much that they can t be correctly identified (fall outside a set t R window) then there is too much variation. Consider for instance, the two peaks illustrated in Figure 2, with a resolution of 1 and equal peak widths: From the formula for resolution, R, we know that t R = 1 = In this expression, w is the peak width at the base: w = 4σ t w r2 where σ has the usual relationship for a Gaussian peak: Ae y = ó 2ð It should be noted that the peak width at half height, w 1/2, which is usually measured, is given by: r1 w 1/2 = 0.589w =2.355σ 2 (t tr ) /2ó 2 3

Now for our two peaks, t r2 t r1 = w. To be able to confidently say that a peak is either 1 or 2, we must be confident that neither peak can move more than half the distance between them. Another way of saying this is that our t R window for identification is less than ±w/2. Note that this is an absolute, not a relative window. It is not uncommon for peaks to exhibit w 1/2 of 60 ms with the use of a 100- or 150-µm column or larger diameter columns operated well above optimal flow in fast mode. Therefore, t R repeatability must be better than ±60 ms (0.001 min). Faster or more efficient peaks will require even better reproducibility. The above argument assumes a resolution of 1. Less resolved peaks will require more stringent t R precision for accurate peak identification. t r2 - t r1 costly to the operator (in terms of time) depending upon the number of analytes being measured. Changing the t R window for six or less components will take less than 10 minutes. For those performing analysis with more components (such as VOA or semi-volatiles) the time requirement becomes much more substantial. Along with the cost of time the manual process adds many sources for error. By achieving extremely good t R precision, it is possible to update t R windows automatically as part of the method every time the method is run. The requirement for this is identical to the previous argument or that absolute precision be better than one half the peak width so that any deviation between two subsequent runs does not result in the misidentification of the peak. Automatic updating of t R windows is a user-settable option with the 6890N when using any Agilent data station. Retention Time Locking (RTL) w (1/2) w Figure 2. Example of two peaks with resolution of one and equal peak widths. Retention time precision can also be influenced by the algorithm used by the integrator to determine the peak maximum and assign a t R. Wider peaks (with a relatively large "flat" maximum) are the most difficult for the computer algorithms to handle accurately. Fortunately, peaks with wider peaks have larger t R differences (for a resolution of 1), negating the need for extreme t R precision. Other Benefits of Extreme t R Precision Automated t R Window Updating In normal operation, a gas chromatograph performing a method will see small changes in t R, or shifts, occur as a part of the normal aging process of the column. The difference in t R in two sequential runs is insignificant. The difference after many runs or weeks of use will lead to the need for the t R windows to be adjusted. This can be extremely The t R precision is most commonly thought of as repeatability. That is the precision obtained on a single instrument, performing a particular method. For a single instrument to achieve this performance both temperature and column pressure (column flow) settings must be precise but not necessarily accurate. To achieve the same level of precision between two different instruments performing the same method requires that the same settings be accurate. That is the same temperature or pressure setting on different instruments be the same (within small tolerances) as measured by another monitoring device. The Agilent 6890N shows this type absolute accuracy. Because of this, it is possible to perform RTL on separate instruments performing the same method to produce identical (t R ) results. RTL is dependent upon the fact that the temperatures and pressures of one gas chromatograph are identical to that of another. Small differences in column length between units, which would ordinarily cause slight t R differences, can be normalized by small changes in column pressure (flow). Benefits to the chromatographer include ease of set-up after column maintenance or replacement (no need to change t R windows), consistent results over many years and different instruments, and the ability to directly compare results or transfer methods precisely between different laboratories. Area Precision Area precision of an analyte peak is the basis for all quantitative measurement in chromatography. Area precision as it relates to a GC is directly related to many aspects of the measurement. The 4

most important being, the ability of data processing equipment to precisely integrate signal from the gas chromatographic detector, the ability of the detector to provide the same signal for the same amount of material introduced to it and the ability to load a precise amount of sample and standard onto the chromatographic column. More than any other chromatographic measurement, area is a function of the entire chromatographic system and chromatographic method. Area precision of the Agilent 6890N is specified at better than 1% RSD. This is specified using a splitless injection so that 2 ng of tetradecane is loaded on column and the same generic chromatographic conditions listed in Table 1 employed. The response of the signal is analyzed without the use of an internal standard (ISTD). Conclusion Area and t R are the two primary measurements in GC. The precision with which they are measured ultimately determines the validity of quantitative data generated. For More Information For more information on our products and services, visit our Web site at www.agilent.com/chem. 5

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 2005 Printed in the USA September 16, 2005 5989-3425EN