Enhancing the productivity of food sample analysis with the Agilent 7700x ICP-MS

Similar documents
Trace elemental analysis of distilled alcoholic beverages using the Agilent 7700x ICP-MS with octopole collision/ reaction cell

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Rapid Analysis of High-Matrix Environmental Samples Using the Agilent 7500cx ICP-MS. Application. Author. Abstract. Introduction.

Direct Measurement of Metallic Impurities in 20% Ammonium Hydroxide by 7700s/7900 ICP-MS

The ultratrace determination of iodine 129 in aqueous samples using the 7700x ICP-MS with oxygen reaction mode

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Multi-Element Analysis of Petroleum Crude Oils using an Agilent 7900 ICP-MS

Ultra-fast determination of base metals in geochemical samples using the 5100 SVDV ICP-OES

Determination of Chromium in Gelatin Capsules using an Agilent 7700x ICP-MS

Simple, reliable analysis of high matrix samples according to US EPA Method 6020A using the Agilent 7700x/7800 ICP-MS

High Throughput Water Analysis using Agilent 7900 ICP-MS coupled with ESI prepfast

Multi-Element Analysis of Cannabis using the Agilent 7800 ICP-MS

Simple and Reliable Soil Analysis using the Agilent 7800 ICP-MS with ISIS 3

Semiquantitative Screening of Pharmaceutical Antiviral Drugs using the Agilent 7500ce ICP-MS in Helium Collision Mode

Direct Analysis of Trace Metal Impurities in High Purity Nitric Acid Using ICP-QQQ

Comparing Collision/Reaction Cell Modes for the Measurement of Interfered Analytes in Complex Matrices using the Agilent 7700 Series ICP-MS

Determination of challenging elements in ultrapure semiconductor grade sulfuric acid by Triple Quadrupole ICP-MS

Thermo Scientific icap RQ ICP-MS: Typical limits of detection

ICP-OES Application Note Number 35

Determination of Impurities in Silica Wafers with the NexION 300S/350S ICP-MS

Analysis of Trace Metal Impurities in High Purity Hydrochloric Acid Using ICP-QQQ

Ultra trace measurement of potassium and other elements in ultrapure water using the Agilent 8800 ICP-QQQ in cool plasma reaction cell mode

High-Throughput Semiquantitative Screening of Ambient Air Samples by ORS-ICP-MS and Integrated Sample Introduction System (ISIS) Application

Determination of major, minor and trace elements in rice fl our using the 4200 Microwave Plasma- Atomic Emission Spectrometer (MP-AES) Authors

Elemental analysis of river sediment using the Agilent 4200 MP-AES

Quantitative analysis of high purity metals using laser ablation coupled to an Agilent 7900 ICP-MS

Application of the Agilent 7900 ICP-MS with Method Automation function for the routine determination of trace metallic components in food CRMs

Direct Measurement of Trace Metals in Edible Oils by 7500cx ICP-MS with Octopole Reaction System

China Soil Pollution Survey: Elemental Analysis of Soil and Sediment Digests by ICP-MS

Thermo Scientific icap TQ ICP-MS: Typical limits of detection

Lead isotope analysis: Removal of 204 Hg isobaric interference from 204 Pb using ICP-QQQ in MS/MS mode

Analysis of domestic sludge using the Agilent 4200 MP-AES

Real World Analysis of Trace Metals in Drinking Water Using the Agilent 7500ce ICP-MS with Enhanced ORS Technology. Application. Authors.

Implementing EPA Method 6020 with the Agilent ICP-MS Portfolio

Fast Analysis of Water Samples Comparing Axially-and Radially- Viewed CCD Simultaneous ICP-OES

Total Elemental Analysis of Food Samples for Routine and Research Laboratories, using the Thermo Scientific icap RQ ICP-MS

Determination of Pb, Cd, Cr and Ni in Grains Based on Four Chinese National Methods via Zeeman GFAAS

Application note. Trace level analysis of sulfur, phosphorus, silicon and chlorine in NMP using the Agilent 8800 Triple Quadrupole ICP-MS

Trace Analyses in Metal Matrices Using the ELAN DRC II

Direct Analysis of Photoresist Using Inductively Coupled Plasma Mass Spectrometry (ICP-MS) Application

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Applications of ICP-MS for Trace Elemental Analysis in the Hydrocarbon Processing Industry

Single particle analysis using the Agilent 7700x ICP-MS

Samples compliant with the WATER

Application. Determination of Trace Metal Impurities in Semiconductor-Grade Hydrofluoric Acid. Authors. Introduction. Abstract.

Analysis of Arsenic, Selenium and Antimony in Seawater by Continuous-Flow Hydride ICP-MS with ISIS

Determination of ultratrace elements in photoresist solvents using the Thermo Scientific icap TQs ICP-MS

Agilent ICP-MS. Fundamentals of ICP-MS Analysis and Its Applications for Low Level Elemental Determination in Cannabis

Rapid, simple, interference-free analysis of environmental samples using the XSERIES 2 ICP-MS with 3 rd generation CCT ED

Application Note. Abstract. Authors

Fergus Keenan Shona McSheehy Julian Wills

Techniques for the Analysis of Organic Chemicals by Inductively Coupled Plasma Mass Spectrometry (ICP-MS)

Agilent ICP-MS Journal September 2009 Issue 39

Analysis of high matrix samples using argon gas dilution with the Thermo Scientific icap RQ ICP-MS

Performance of the Agilent 7900 ICP-MS with UHMI for high salt matrix analysis

Today s Agilent Solutions for Determining Heavy Metals in Food using Atomic Spectroscopy

The Investigation of Fertilizer Analyses Using Microwave Digestion and the Agilent 720-ES

Analysis Repeatability of Trace and Major Elements in a Water Sample

Key Analytical Issues: Sample Preparation, Interferences and Variability. Tim Shelbourn, Eli Lilly and Company

Total elemental analysis of food samples for routine and research laboratories using the Thermo Scientific icap RQ ICP-MS

Fundamentals of the ICP-MS Technique and How to Resolve Issues for Pharmaceutical Materials (In 20 Minutes) Tim Shelbourn, Eli Lilly and Company

Accurate Quantification of Cadmium, Chromium, Mercury and Lead in Plastics for RoHS Compliance using the Agilent 7500ce ICP-MS

Accurate Analysis of Trace Mercury in Cosmetics using the Agilent 8900 ICP-QQQ

U.S. EPA SW-846 Method 6010C using the Prodigy High Dispersion ICP

Following documents shall be used for reference on quantities, units, prefixes and other technical vocabulary in this document:

On-line isotope dilution analysis with the 7700 Series ICP-MS: Analysis of trace elements in high matrix samples

Hands on mass spectrometry: ICP-MS analysis of enriched 82 Se samples for the LUCIFER experiment

Determination of Phosphorus and Sulfur with Dynamic Reaction Cell ICP-MS

Agilent s New 8800 ICP-QQQ. Transforming ICP-MS Technology

Hydride Generation for the Determination of As, Sb, Se and Bi Using the Teledyne Leeman Lab s Prodigy 7 ICP-OES

The Determination of Toxic, Trace, and Essential Elements in Food Matrices using THGA Coupled with Longitudinal Zeeman Background Correction

ICP-3000 Inductively Coupled Plasma Optical Emission Spectrometer

February 20, Joe Cerniglia The International Group for Historic Aircraft Recovery (TIGHAR) Job Number: S0CHG688. Dear Joe:

Meeting the Challenges of Soil Analysis with the Avio 200 ICP-OES

OES - Optical Emission Spectrometer 2000

Elemental Scientific. Spike Recovery. Elemental Scientific

Florida State University Libraries

Analysis of noble metals at low levels in geological reference materials and ores

U.S. EPA SW-846 Method 6010C Using the Prodigy7 High- Dispersion ICP Introduction. Application Note - AN1305. Experimental

Determination of Elements at Sub-ppb Concentrations in Naphtha Mixtures Using the NexION 300 ICP-MS

Rapid determination of five arsenic species in polished rice using HPLC-ICP-MS

Agilent 7500a Inductively Coupled Plasma Mass Spectrometer (ICP-MS)

The Characterization of Nanoparticle Element Oxide Slurries Used in Chemical-Mechanical Planarization by Single Particle ICP-MS

Not Just for Paint Chips Anymore. High Throughput, Open Vessel Microwave Digestion of Matrices Requiring Moderate Conditions

Unmatched Removal of Spectral Interferences in ICP-MS Using the Agilent Octopole Reaction System with Helium Collision Mode Application

Automated Determination of PPQ Levels of Thorium in High Purity Copper

Agilent 8900 Triple Quadrupole ICP-MS

The New Agilent 7700 Series ICP-MS

Application note. Accurate determination of sulfur in biodiesel using Isotope Dilution-Triple Quadrupole ICP-MS (ID-ICP-QQQ) Petrochemical

Basic and Application Inductively coupled plasma mass spectrometry (ICPMS)

Analytical Approaches to Arsenic

Sample Analysis Design PART III

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Agilent Technologies at TIAFT 2013

icap TQ ICP-MS Applications Compendium

Simplified workflows for the Thermo Scientific icap TQ ICP-MS

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Elemental Scientific. Spike Recovery. Elemental Scientific

1.0 ACCREDITATION / REGISTRATION 2.0 PRODUCT DESCRIPTION 3.0 CERTIFIED VALUES AND UNCERTAINTIES. Assay Information:

Transcription:

Enhancing the productivity of food sample analysis with the Agilent 77x ICP-MS Application note Foods testing Authors Sebastien Sannac, Jean Pierre Lener and Jerome Darrouzes Agilent Technologies Paris, France Introduction To ensure food safety and safeguard human health, the characterization of the elemental composition of a wide range of food types is required. Since the concentration ranges for different elements vary significantly across different foods, various techniques have been used for sample characterization. In this work, we investigated the possibility of measuring all the required elements with a single configuration using the Agilent 77x ICP-MS. The determination of trace and major elements simultaneously is possible in part due to the 9 orders of magnitude provided by the detector. Interferences generated by the plasma and the sample matrix present another challenge in food analysis. In this study, we investigated the use of collision cell technology with a single cell gas, helium, to eliminate those interferences.

We also evaluated the use of a discrete sampling system to increase sample throughput. The Agilent ISIS-DS discrete sampling system was evaluated for its ability to reduce analysis times by minimizing uptake and washout time during the analytical run. Experimental Standard and sample preparation Four certified reference materials (CRMs) were analyzed as part of this work. These included NIST (US) SRM 1548a (Typical Diet), NIST SRM 2976 (Mussel Tissue), NRC (Canada) DORM3 (Fish Tissue) and NIST SRM 8415 (Whole Egg). 25 mg of each CRM sample was digested with 3 ml of nitric acid and 1 ml of hydrogen peroxide using a microwave device (program details are listed in Table 1). The digested samples were made up to 5 ml with DI water. The resulting samples contained up to 5 g/l of matrix. Standards for calibration were prepared from 1 to 5 µg/l (1 more concentrated for the major elements) and from.5 to 5 µg/l for mercury by diluting multi-element solutions with 6% HNO 3 /.5% HCl. No matrix matching of the standards to the samples was necessary. Table 1. Detailed program for the microwave digestion of the food CRM samples Step Time (min) Power (W) 1 1 28 2 5 3 1 55 4 5 5 6 72 6 7 7 1 28 Instrumentation An Agilent 77x ICP-MS featuring the 3 rd generation Octopole Reaction System (ORS 3 ) and fitted with a standard sample introduction system (MicroMist glass concentric nebulizer, a quartz Peltier-cooled spray chamber, and quartz torch with 2.5 mm internal diameter injector) was used for all measurements. The oxide ratio was set at.8% (CeO + /Ce + ) resulting in excellent sample decomposition in the plasma and minimum matrix effects. For interference suppression, the ORS 3 was operated in helium collision mode (He mode) only, which is effective at removing a wide range of plasma and matrix-based polyatomic species using kinetic energy discrimination (KED). Instrument operating conditions are summarized in Table 2. Table 2. Agilent 77x ICP-MS operating parameters Parameter Plasma power Plasma gas flow rate Auxiliary gas flow rate Carrier gas flow rate Dilution gas flow rate Sample depth Value 155 W 15. L/min 1. L/min.89 L/min.15 L/min 8. mm Spray chamber temperature 2 C KED Helium gas flow rate 3 V 4.5 ml/min Helium mode provides several critical advantages for food analysis when compared to reactive cell gases. He mode effectively removes all polyatomic interferences, not just reactive polyatomics. Since He is inert, no new interferences are produced, regardless of the matrix. Unlike a reactive cell gas, He does not react with any analytes, so consistent and predictable sensitivity is maintained. Elements that do not suffer from polyatomic interferences can be analyzed with He mode as well. But to achieve better limits of detection, they were analyzed with no gas in the cell (no gas mode). During analysis of a sample, the system automatically switches between no gas and He mode as needed, thereby permitting the measurement all elements under optimum conditions without the need for multiple analyses for a given sample. The transition time needed for switching modes is very rapid (~5 seconds), so high throughput is not significantly impacted. 2

ISIS-DS discrete sampling system Figure 1 shows the basic operation of the Agilent ISIS-DS system, which was employed to improve the throughput after the initial method was developed. The sample is rapidly drawn into the sample loop using the high speed ISIS pump (P1), while blank carrier with online internal standards is constantly pumped to the nebulizer (P2). Rotation of the 6-port valve then diverts the carrier through the loop, pushing the sample ahead of it to the nebulizer. At the same time, the autosampler probe goes to rinse position and begins rinsing before the next sample. ISIS-DS provides several benefits for this analysis: Sample uptake and rinse-out times are significantly reduced, resulting in very fast analysis times. The total exposure of the ICP-MS cones and lenses to the sample matrix is minimized, resulting in extended long-term stability. Carryover is reduced due to the elimination of peristaltic pump tubing from the sample path. Sample introduction system maintenance and cleaning are reduced. Results and discussion Calibration Table 3 shows the details obtained from the calibration curves, and Figure 2 shows the calibration curves obtained for a typical selection of elements. The major elements (Ca, K, Mg, Na) were calibrated up to 5 mg/l, while the trace elements were measured up to 5 µg/l. This figure underlines the sensitivity of the system, with detection limits at the ng/l level (ppt). Table 3. Details of the calibration curves. R represents the coefficient of linearity. DL is the detection limit calculated as 3 sigma of the blank. Mass Element Tune step R DL (ppb) 23 Na He 1..16 24 Mg He 1..31 27 Al He 1..23 39 K He 1. 1.8 44 Ca He 1. 5.7 47 Ti He 1..41 51 V He 1..13 52 Cr He 1..38 55 Mn He 1..18 56 Fe He 1..21 59 Co He 1..14 6 Ni He 1..39 63 Cu He 1..13 66 Zn He 1..17 75 As He 1..84 78 Se He 1..38 95 Mo No gas 1..22 17 Ag No gas 1..16 111 Cd No gas 1..7 118 Sn No gas 1..28 121 Sb No gas 1..5 137 Ba No gas 1..2 21 Hg No gas 1..3 28 Pb No gas 1..13 Figure 1. Details of the function of the Agilent ISIS-DS discrete sampling device 3

24 Mg [2] ISTD :45 Sc [2] x1 1 y =.7* x + 5.546E-4 R =.9999 DL =.3126 ppb BEC =.7846 ppb 4 47 Ti [2] ISTD :13 Rh [2] x1-2 y = 2.1254E-4* x +9.5153E-6 1.5 R =.9997 DL =.4114 ppb BEC =.4477 ppb 1 2.5 2. 4. 6. 2. 4. 6. 75 As [2] ISTD :13 Rh [2] x1-2 y = 7.658E-4* x +8.3975E-6 R =.9998 DL =.848 ppb BEC =.114 ppb 4 111 Cd [1] ISTD :13 Rh [1] x1-1 y =.22* x +3.1328E-6 1.5 R = 1. DL =.7243 ppb BEC =.1457 ppb 1 2.5 2. 4. 6. 2. 4. 6. 21 Hg [1] ISTD :193 Ir [1] x1-2 y =.19* x +2.4143E-5 R =.9996 DL =.2975 ppb BEC =.1283 ppb 1 28 Pb [1] ISTD :193 Ir [1] y =.362* x +4.321E-4 R = 1. DL =.1268 ppb BEC =.1187 ppb 2.5 1 2. 4. 6. Figure 2. Typical calibration curves obtained during the analysis of the food CRM samples 2. 4. 6. 4

Interference suppression Initial measurements were made in order to evaluate the effectiveness of using He mode alone to manage interferences. Multiple isotopes, where available, were monitored and compared for each element in the reference materials. The results are given in Table 4. Table 4. Comparison of measured concentrations between isotopes of the same element (µg/l) DORM3 SRM2976 SRM8415 SRM1548a 24 Mg 2781.99 2247.9 151.3 2717.79 26 Mg 2749.48 23311.89 1531.2 266.32 Difference (%) 1-4 -2 2 43 Ca 6249.1 35189.24 1744.7 7489.41 44 Ca 6511.34 34589.19 11192.78 7532.89 Difference (%) -4 2-4 -1 47 Ti 142.9 21.38 43.91 1.91 49 Ti 152.4 22.49 42.85 11.4 Difference (%) -7-5 2-5 52 Cr 6.72 2.65 2.16.57 53 Cr 6.74 2.58 1.99.42 Difference (%) 3 8 27 56 Fe 115.83 99.14 576.99 182.53 57 Fe 119.94 966.56 564.96 18.4 Difference (%) 4 2 2 1 6 Ni 4.46 4.39 1.11 5.48 62 Ni 4.45 4.45 1.1 5.44 Difference (%) -1 1 1 63 Cu 49.67 19.86 15.23 11.61 65 Cu 49.5 19.83 15.16 11.6 Difference (%) As can be seen from Table 4, excellent agreement between isotopes was demonstrated, confirming the effective suppression of multiple interferences in varying matrices using He mode. In contrast, the use of a reactive gas such as H 2 or NH 3 cannot be applied to multiple, unknown interferences on multiple isotopes simultaneously in a single quadrupole ICP-MS. By quantifying an element using multiple isotopes, the results can be compared; good agreement validates the data, indicating that the reported concentration was not affected by any interference. The 77x ICP-MS is unique in its ability to allow the efficient suppression of multiple interferences on multiple isotopes using only helium gas. Method validation Table 5 summarizes the final results, comparing the measured values for each CRM with the certified values. All elements in all of the CRMs showed excellent agreement with the certified values, including the major elements (Ca, K, Mg or Na), the interfered elements (As, Se, Fe ) and the non-interfered elements (Hg, Pb ). This clearly demonstrates that it is practical to measure a wide range of elements in a variety of food types over a very wide range of concentrations in a single analytical run using the Agilent 77x ICP-MS. Application of discrete sampling to food analysis Prior to the addition of ISIS-DS, the method described above required an analysis time of 5 minutes per sample. With the addition of discrete sampling using ISIS-DS, the total analysis time per sample was reduced to 1.2 minutes. While the ISIS-DS system significantly reduces the analysis time per sample, there was no compromise in the other performance attributes of the system, that is, the capacity to measure trace, majors or interfered elements simultaneously in a range of food matrices with the use of He mode to remove interferences. Table 6 summarizes the parameters used for the ISIS-DS method. 5

Table 5. Comparison of measured versus certified values for all analyte elements in the certified reference materials (units: mg/kg except * in %) SRM 1548a Typical Diet SRM 2976 Mussel Tissue DORM3 Fish Tissue SRM 8415 Whole Egg Result Certified Result Certified Result Certified Result Certified Na 8459 8132 ± 942 3.4 3.5 ±.1* - -.317.377* Ca 1869 1967 ± 113.73.76 ±.3* - -.235.248* Mg 63 58 ± 26.7.48.53 ±.5* - - 297 35 K 6684 697 ± 125.99.97 ±.5* - -.319.319* Al 73.5 72.4 ± 1.52 14 134 ± 34 - - 563 54 As.21.2 ±.1 14.9 13.3 ± 1.8 6.61 6.88 ±.3.15 (.1) Cd.35.35 ±. 15.79.82 ±.16.284.29 ±.2.1 (.5) Cu 2.57 2.32 ±.16 4.9 4.2 ±.33 15.9 15.5 ±.63 3 2.7 Cr - -.54.5 ±.16 2.15 1.89 ±.17.42.37 Fe 4.4 35.3 ± 3.77 24 171 ± 4.9 368 347 ± 2 114 112 Ni 1.21.369 ±.23.9.93 ±.12 1.42 1.28 ±.24 - - Pb.12.44 ±.9 1.14 1.19 ±.18.39.395 ±.5.59.61 Se.259.245 ±.28 1.76 1.8 ±.15 - - 1.45 1.39 Sn 14.3 17.2 ± 2.57.12.96 ±.39.1.66 ±.12 - - Zn 23.3 24.6 ± 1.79 144 137 ± 13 47.5 51.3 ± 3.1 65.8 67.5 Hg - -.14.61 ±.36.412.49 ±.27 - - Table 6. Parameters for the ISIS-DS method Conclusions Parameter Load time Load speed Probe rinse time Probe rinse speed Post rinse time Post rinse speed Value 12 s 1. rps 6 s.2 rps 1 s.5 s Food matrices present complex and challenging samples for elemental analysis, particularly using a single analytical technique, due to the varying matrix composition, very wide range of element concentrations, and potential for matrix based interferences. The Agilent 77x ICP-MS, operated in standard configuration and using only no gas mode and He mode was demonstrated to easily and accurately measure all certified elements in a range of food certified reference materials. Excellent sensitivity was demonstrated with detection limits in the ppt (ng/l) range while at the same time maintaining sufficient dynamic range to measure elements in the high ppm range. The use of discrete sampling via ISIS-DS did not compromise the performance of the method in any way, while reducing the run time from 5 minutes to 1.2 minutes per sample. 6

7

www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc. 212 Published March 23, 212 Publication number: 5991-17EN