HPLC Columns. HILICpak VT-50 2D MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

Similar documents
HPLC Columns. STANDARD Pullulan P-82 MANUAL. Shodex HPLC Columns Europe, Middle East, Africa, Russia

for Acclaim Mixed-Mode HILIC-1 Column

FAQ's. 1 - Which column would be the most appropriate for my application?

LC III: HPLC. Originally referred to as High-Pressure Liquid Chromatography. Now more commonly called High Performance Liquid Chromatography

High Performance Liquid Chromatography

High Performance Liquid Chromatography

High Performance Liquid Chromatography

ION CHROMATOGRAPHY SYSTEM S 150


High Performance Liquid Chromatography

Basic principles of HPLC

User Manual. IonPac NS2 Columns Revision 02 October For Research Use Only. Not for use in diagnostic procedures.

Shodex TM ODP2 HP series columns

HPLC. GRATE Chromatography Lab Course. Dr. Johannes Ranke. September 2003

SRT -10/10C (10 µm ) Fast Protein Purification For FPLC, HPLC and Prep LC. Sepax Technologies, Inc. Preparative Size Exclusion Chromatography

High Performance Liquid Chromatography (HPLC) of Three Analgesics

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Guide to Choosing and Using Polymer Reversed Phase Columns!

ProPac WCX-10 Columns

Packed Column for Ultra-Fast Reversed-Phase Liquid Chromatography, TSKgel Super-ODS. Table of Contents

COLUMN GUIDE - Usage and maintenance of HPLC columns

C18 Column. Care & Use Sheet

Acclaim Mixed-Mode WCX-1

HPLC Praktikum Skript

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

ACQUITY UPLC HSS Columns

Hydrophilic Interaction Liquid Chromatography: Some Aspects of Solvent and Column Selectivity

for Acclaim Mixed-Mode WCX-1

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Biochemistry. Biochemical Techniques HPLC

COSMOSIL Applications

High Pressure/Performance Liquid Chromatography (HPLC)

LC Technical Information

Acclaim HILIC-10 Columns

Saccharides and Organic Acids

VMA, HVA, 5-HIAA Urine LC-MS/MS Analysis Kit User Manual

Acclaim Mixed-Mode WAX-1 Columns

for Acclaim Trinity TM P1

Dionex IonPac AS28-Fast-4µm column

Thermo Scientific. Acclaim Trinity P2. Column Product Manual. P/N: October Part of Thermo Fisher Scientific

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Determination of trace anions in concentrated hydrofluoric acid

Chemistry 3200 High Performance Liquid Chromatography: Quantitative Determination of Headache Tablets

HPLC Winter Webinars Part 2: Sample Preparation for HPLC

columns IonPac AS17-C Anion Exchange Column

It s hot! ZIC -chilic

TECHNICAL BRODIFACOUM

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

PRODUCT SAFETY SUMMARY (GPS) SODIUM-HYDROXIDE

PYRIPROXYFEN TECHNICAL

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

IonPac Trace Cation Concentrator

GPC/SEC Practical Tips and Tricks. Thomas Dent Applications Scientist Agilent Technologies. October, 2011 Gulf Coast Conference

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Lowry Protein Assay Kit KB test (96 well plate)

[ CARE AND USE MANUAL ]

H 3 CO H 3 CO S CH 3

INSTRUCTION MANUAL. Packed Columns for Ion Chromatography

Cation Exchange HPLC Columns

columns IonPac SCS 1 Silica Cation Separator

ACQUITY UPLC Peptide BEH C 18, 130Å and 300Å Columns

Anion and Cation analysis with Professional IC - automatic dilution and sample preparation with SPM

Automating Method Development with an HPLC System Optimized for Scouting of Columns and Eluents

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

Chromatography. Gas Chromatography

Advantages of polymerbased. an alternative for ODS

Determination of Polymer Modifier in Asphalt

Dilution(*) Chromatography

Figure 1. Structures for Vitamin B2 and Vitamin B1.

DATES: LAB: Liquid Chromatography Separation of Grape Kool-Aid

Micro Volume QuEChERS kit

101Bio.com. Total Aflatoxin ELISA Kit. better than big brands and save. Cat. #: T21131 (1 x 96-well plate)

penta-hilic UHPLC COLUMNS

N) manual. Biomaster Operating manual

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

for IonPac AG11-HC IonPac AS11-HC

RESOLUTION OENO 33/2004 DETERMINATION OF SHIKIMIC ACID IN WINE BY HPLC AND UV-DETECTION

Technical Information

Quick guide to selecting columns and standards for Gel Permeation Chromatography and Size Exclusion Chromatography SELECTION GUIDE

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Oligo Click S ROTI kit for DNA labeling

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Preparative Separation of Active Components from Natural Products using Low-pressure Gradient Preparative HPLC

Oligo Click M Reload ROTI kit for DNA labeling

HPLC Column. Technical Guide

The determination of trace anions in concentrated phosphoric acid

[ Care and Use Manual ]

Chapter content. Reference

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

The analysis of organic acid content of additives, premix, feed, and water.

Ion Chromatography (IC)

[ CARE AND USE MANUAL ] XSELECT HSS XP 2.5 µm COLUMNS I. INTRODUCTION CONTENTS I. INTRODUCTION II. GETTING STARTED III. COLUMN USE

CATALASE ACTIVITY ASSAY KIT KB test (96 well plate)

Liquid Chromatography

LC Column Troubleshooting. Do you have an equivalent column for my LC column?

Avery Dennison LW Series Retroreflective Window Sticker Film Issued: July 2013

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

A novel approach to the analysis of multivitamin by online supercritical fluid extraction/supercritical fluid chromatography

Transcription:

HPLC Columns MANUAL HILICpak VT-50 2D Shodex HPLC Columns Europe, Middle East, Africa, Russia For technical support please use contact details shown below: SHOWA DENKO EUROPE GmbH Shodex Business Konrad-Zuse-Platz 3 81829 Munich, Germany E-mail: support@shodex.de Phone: +49 (0)89 93 99 62 37 www.shodex.de

Operation Shodex HILICpak VT-50 2D (Please read this manual carefully before using the column to ensure performance and life.) 1. Introduction Shodex HILICpak VT series is packed with a hydrophilic polymeric gel containing chemically bonded quaternary ammonium group. Especially about analysis of anionic substances like phosphorylated sugar and organic acid, the VT column provides high resolution and reproducibility. 2. Handling instructions <Important> Caution! * Please consult the SDS for the reagents and solvents used with the columns for health concerns caused by acute exposure due to leakage from the column or adjoining tubing. * Use the column within the regular range of flow rate, pressure and temperature. The column performance may deteriorate when it is handled beyond the permissible range even for a short time. See the Section 4. Usable Conditions for the permissible ranges. 3. Specifications Product code F7630400 VT-50 2D 2.0 x 150 4,500 5 F6711300 VT-50G 2A 2.0 x 10 (Guard column) Packing material: Polyvinyl alcohol. Functional group: Column material: Screw type: Product name Quaternary ammonium group. PEEK. Size (mm) I.D. x L Particle size (µm) Unified Thread Standard No.10-32 UNF. Plate Number (per column) Shipping solvent: Acetonitrile / 25mM ammonium formate = 85 / 15 4. Usable conditions 2

Product name Flow rate (ml/min) Max pressure Standard Max (MPa) VT-50 2D 0.2-0.3 0.5 10 VT-50G 2A 0.2-0.3 0.5 - ph range Temperature range ( C) 2-13 4-60 * High-temperature operation may result in the generation of an air bubble, necessitating degassing. Low-temperature operation may require reduced flow rates, because of increased eluent viscosity. 5. System Clean up Clean up the LC system including the injector and the sample loop by switching the valve, and flow the eluent before column installation. * The previous eluent used for analyses in the system may damage the column, if it is not compatible with the column. * When replacing nonpolar solvent to water, replace first with methanol and then replace with water. When replacing buffer solution to acetonitrile/water, replace first with 100% water and then replace with eluent. Substances adsorbed in the pump and tubing may not be compatible with the column. 6. Column installation Install and use the VT column with the flow through the column matching the flow direction arrow on the column tag. Set the flow rate at 0.1 ml/min, and connect the column. Flow at the low rate until the column temperature increases to the setting temperature, and then increase the flow rate to the analytical condition. * The column should always be installed in the manner above, for safe and effective operation. 7. Eluents HILIC mode Operational solvents Aqueous solutions of acetonitrile (or methanol *1 ) at any ratio, is used as eluent. *2 Any salts *3 soluble in the solution above are available. 3

*1 With aqueous methanol solutions, the flow rate should be lower than normal, due to relatively high viscosity. *2 Other organic solvents are not guaranteed as the eluent. *3 Maximum concentration of the buffer depends on the ratio of acetonitrile (or methanol). Eluent modes Please note any precipitation of salt when using buffer! Isocratic, gradient, or stepwise elution * Filter the eluent with a membrane filter (0.45 µm) to prevent chromatogram noise and column performance deterioration by small particles or undissolved materials. * When the shipping solvent (Acetonitrile / 25mM ammonium formate = 85 / 15) is replaced with a buffer solution, please flow 100% water to the column before flowing the buffer solution to avoid the precipitation of the buffer. When replacing buffer solution to acetonitrile solution, also flow 100% water to the column before flowing acetonitrile solution. * Column equilibration for using acetonitrile/buffer eluent. Anions in the buffer interact with the quaternary ammonium functional groups of VT packing gel. The reproducibility of the analysis will not be attained until the ionic equilibrium is attained between the anions and the amino groups. In such case, the equilibration time may be shortened by the operation below. This equilibrium operation is not needed when acetonitrile/water or buffer eluents are used. HILIC mode Eluent Acetonitrile / buffer Acetonitrile/water Solution The same buffer without acetonitrile Quantity 10 to 20 times the column volume Unnecessary Flow rate Low 4

* When the eluent conditions change from buffer solution to acetonitrile/water, please use the standard cleaning method. See Section 9. Column Cleaning. 8. Sample preparation HILIC mode Dissolve sample in water or the eluent, and Eluent: add acetonitrile to obtain 50% or higher Acetonitrile/water acetonitrile aqueous solution. Solid sample Sample Dissolve sample in the buffer or the eluent; Eluent: preparation and add acetonitrile to obtain 50% or higher Acetonitrile/buffer acetonitrile aqueous solution. Aqueous Add acetonitrile to obtain 50% or higher acetonitrile aqueous sample solution. Injection volume 20 µl or less for analytical columns Optimum separation efficiency is generally obtained with sample matrix similar to the eluent composition. Use acetonitrile in the sample matrix whenever possible. * Filter the sample with a membrane filter (0.45 µm) to prevent blockage. * In case of gradient condition, dilute the sample with the initial eluent. 9. Column Cleaning Elution characteristics of a column may change considerably after long, repeated usage, due to the accumulation of pollution components on the packing material, for example metal ions from the LC system or the sample. The cleaning procedures outlined below may be used. Clean the guard column and analytical column separately by flowing the cleaning solution in the opposite direction of the arrow on the column tag. The applied flow rate should be lower than 0.1 ml/min. Cleaning method 1 (Standard cleaning) Cleaning solution VT-50 2D VT-50G 2A 1 H 2 O 6 min 1 min 2 0.1M NaOH aq. 75 min 5 min 3 H 2 O 12 min 2 min 4 Eluent 40 min 3 min 5

Cleaning method 2 (Cleaning metal ions) Cleaning solution VT-50 2D VT-50G 2A 1 H 2 O 6 min 1 min 2 0.1M HNO 3 aq. 75 min 5 min 3 H 2 O 6 min 1 min 4 0.1M NaOH aq. 75 min 5 min 5 H 2 O 12 min 2 min 6 Eluent 40 min 3 min * After cleaning the column by alkaline, in some substances, there are cases that the theoretical plate number becomes higher than the initial plate number. In such cases, the plate number will get back to the initial level gradually by running the eluent or water. * Complete the washing procedure steps continuously. Do not store the column with either acid or alkaline solution as it will advance the deterioration of the column. * Flow the alkaline waste into a separate bottle without flowing to the detector. The strong alkaline solution may cause damage to the cell of the detector. 10. Column Inspection Column inspection method is described in Certificate of Analysis (CoA). * Assessment of the column s functional integrity prior to initial and later use by standardized comparison of the certificate of analysis is recommended. Please see CoA for the detailed analysis conditions and sample preparation. 11. Attention 1) Do not remove the end fittings of the column to prevent performance deterioration and for safety reasons. 2) Do not make a strong impact on the column: such as hitting or dropping on the floor. 6

3) Replace the solvent in the LC system with the eluent to be used before connecting the column. 4) Connect the column so that the flow direction corresponds to the arrow mark on the tag. 5) When the column is not used for two weeks or more, replace the in-column solvent with the shipping solvent, remove it from the LC system, close each end with a stopper, and store it at controlled room temperature. 6) Contact Shodex website (www.shodex.de) or Shodex partners regarding product and analysis applications. 12. Warranty (ver.3) 1) Showa Denko K. K. warrants that the Shodex Column, at the time of delivery to the user, will conform to the specification of the attached Certificate of Analysis, if the Shodex Column is used in accordance with the operating manual. The foregoing warranty is exclusive and is in lieu of all other warranties with respect to the Shodex Column, whether written, oral, implied, statutory or otherwise. No warranties by Showa Denko K. K. are implied or otherwise created, including, but not limited to, the warranty of merchantability and fitness for particular purposes. 2) Any claim of inconformity to the specification must be notified to Showa Denko K.K. within ten (10) days after delivery to the user. User s exclusive remedy and Showa Denko K.K. s exclusive liability for such claim are limited to the replacement of the Shodex Column in question. In no event is Showa Denko K.K. liable for any indirect, incidental or consequential damage arising out of in connection with the Shodex Instrument, whether or not such damage is allegedly based on breach of warranty, negligence or otherwise. 3) No warranty is made in any of the following cases: (1) If the Shodex Column is not used in accordance with the operating manual. (2) If the Shodex Column is remodeled by anyone other than person or firm designated by Showa Denko K.K. (3) If the Shodex Column is resold by the user without giving prior written notice to Showa Denko K.K. 7

(4) If the performance of the Shodex Column is not conform to the specification of the attached Certificate of Analysis due to any of the reasons below: a) Computer virus b) Impurities contained in the sample, reagent, gas air or cooling water provided by the user c) Breakdown or malfunction of equipment, apparatus or component used in combination with the Shodex Column d) Force majeure such as fire, earthquake, flood, other natural disaster, rime, riot, act of terrorism, war or radioactive contamination 4) In no event is Showa Denko K.K. liable for (i) the results of analyses or preparations using the Shodex Column or any portion of the same, including, but not limited to, the reliability, accuracy, efficacy and safety of said results, and (ii) the occupational hazard in the use of the Shodex Column, whether or not such use is made in accordance with the attached Conditions for use. 5) The Shodex instrument is for laboratory use only. It must not be used for clinical diagnosis. Showa Denko K.K. is not liable for any use of the Shodex Instrument except laboratory use. 8