Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System

Similar documents
Rapid Analysis of Food and Fragrances Using High-Efficiency Capillary GC Columns. Application. Authors. Abstract. Introduction

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Fast Analysis of USP 467 Residual Solvents using the Agilent 7890A GC and Low Thermal Mass (LTM) System

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Applying the Agilent 5977A MSD to the Analysis of USP<467> Residual Solvents with the 7697A Headspace Sampler and 7890B GC

Author. Abstract. Introduction

USP<467> residual solvents

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

HIGH SPEED GC-TOFMS FOR ESSENTIAL OILS PROFILING IN ROUTINE QUALITY ASSESSMENT

New Approaches to the Development of GC/MS Selected Ion Monitoring Acquisition and Quantitation Methods Technique/Technology

Analysis of Terpenes in Cannabis Using the Agilent 7697A/7890B/5977B Headspace GC-MSD System

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Static Headspace Blood Alcohol Analysis with the G1888 Network Headspace Sampler Application

A New PEG GC Column with Improved Inertness Reliability and Column Lifetime Agilent J&W DB-WAX Ultra Inert Polyethylene Glycol Column

Application Note. Abstract. Authors. Introduction

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

Analyzing Compounds of Environmental Interest Using an LC/Q-TOF Part 1: Dyes and Pigments. Application. Introduction. Authors. Abstract.

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Fast Analysis of Aromatic Solvent with 0.18 mm ID GC column. Application. Authors. Introduction. Abstract. Gas Chromatography

Agilent J&W PoraBOND Q PT Analyzes Oxygenates in Mixed C4 Hydrocarbon Streams by GC/FID and GC/MSD

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

PAH Analyses with High Efficiency GC Columns: Column Selection and Best Practices

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Detection of Chemical Warfare Agents by Transportable GC/MS

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Application Note. Abstract. Authors. Environmental

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Improve Peak Shape and Productivity in HPLC Analysis of Pharmaceutical Compounds with Eclipse Plus C8 Columns Application

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

GC Analysis of Polybrominated Flame Retardants Application

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

Characterization and Classification of Heroin from Illicit Drug Seizures Using the Agilent 7200 GC/Q-TOF

The Analysis of Organophosphate Pesticides by LC/MS Application

The Determination of Residual Solvents in Pharmaceuticals Using the Agilent G1888 Network Headspace Sampler Application

Automated Sample Preparation of Headspace Standards Using the Agilent 7696 WorkBench

Practical Faster GC Applications with High-Efficiency GC Columns and Method Translation Software

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Application. Gas Chromatography February Introduction

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Highly Sensitive and Rugged GC/MS/MS Tool

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Introduction to Capillary GC. Page 1. Agilent Restricted February 2, 2011

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Separation of Enantiomers of Amphetamine-Related Drugs and Their Structural Isomers

Case 1:13-cv WBH Document Filed 12/30/16 Page 406 of 519. Exhibit 53

UNIQUE SOLUTION FOR FLAVOUR AND FRAGRANCE IDENTIFICATION BY MEANS OF GC-MS (TOF) / CONDENSED PHASE FTIR

Introduction to Capillary GC

Residual solvents analysis using an Agilent Intuvo 9000 GC system

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Hydrogen as a Carrier Gas in the Analysis of Polar/Non-Polar Compounds Using the Polyarc System. Application Note. Author. Introduction.

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Ultra-Inert chemistry for Trace Level Analysis

Separation and Characterization of Indian and Australian Sandalwood Oils

Agilent UltiMetal Plus Stainless Steel Deactivation for Tubing, Connectors, and Fittings

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Introduction to Capillary GC

Discovering the Scent of Celery: HS-SPME, GC-TOFMS, and Retention Indices for the Characterization of Volatiles

Enhanced Simultaneous Total ion and Selective Multi-Ion GC/MS Using a Combination of New Software and Low-Bleed Columns

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Simultaneous dual capillary column headspace GC with flame ionization confirmation and quantification according to USP <467> Application Note

Using TOF for Screening and Quantitation of Sudan Red Colorants in Food Application

Analysis of allergens found in cosmetics using MDGC-GCMS (Multi-Dimensional Gas Chromatograph Mass Spectrometer)

The Analysis of Trace Contaminants in High Purity Ethylene and Propylene Using GC/MS. Application. Agilent Technologies/Wasson ECE Monomer Analyzer

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

GAS CHROMATOGRAPHY MASS SPECTROMETRY. Pre-Lab Questions

U.S. EPA Method 8270 for multicomponent analyte determination

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Application. Gas Chromatography March 1998

Determination of Enantiomeric Excess of Metolachlor from Chiral Synthesis using the Agilent 1260 Infinity Analytical SFC System

Evaluation of Capillary Columns for General Performance Parameters

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Agilent Cary 630 FTIR Spectrometer Supporting Organic Synthesis in Academic Teaching Labs

Application. Introduction. Authors. Abstract. Hydrocarbon Processing

Agilent G2350A Atomic Emission Detector (AED)

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Analysis of Sulfur-Containing Flavor Compounds By GC/MS With A PFPD

Forensic Toxicology. Analysis of Ethanol in Blood using Master SHS Static Headspace Sampler and Master GC Gas Chromatograph APPLICATION NOTE

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Column Dimensions. GC Columns and Consumables. Mark Sinnott Application Engineer. March 12, 2010

AN INTEGRATED SYSTEM USING TEMPERATURE BASED SAMPLING FOR POLYMER CHARACTERIZATION

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

Transcription:

Analysis of Natural Oils and Extracts Using the Low Thermal Mass LTM Series II System Application Note Author Roger L. Firor Agilent Technologies, Inc. 2850 Centerville Rd Wilmington, DE 19808 Abstract The enantiomers of linalool and other important components of natural oils or natural extracts including cypress oil, geranium extract, balm leaves oil, and jasmin oil are separated on a multidimensional system using two separate GC ovens. The first oven is the conventional 7890 air bath oven equipped with a CFT deans switch and a 30 m x 0.25 mm x 0.25 µm DB5ms, and the second is a direct heated LTM II 5-inch module fitted with a 30 m x 0.25 mm x 0.25 µm CycloDex B chiral column. Use of independent temperature control for the chiral column allows for optimal separation conditions for linalool and other components in these natural materials. Full integration of LTM Series II in the Agilent 7890A GC system enables operation in constant flow mode. Both FID/MSD and FID/FID systems are described.

Introduction There are two principle use categories for Low Thermal Mass (LTM) technology. The first is simply to speed up separations and overall analysis cycle times while maintaining the desired separation. In this case, a method is usually translated to the use of a shorter and possibly smaller id column when going from an air bath oven to a LTM module. Method translation software is a powerful tool to help the method developer in this task [1]. The second category is use as a secondary oven in conjunction with the primary oven for increased flexibility to achieve a difficult separation. Linalool (3,7-dimethyl-1,6-octadien-3-ol) is an important compound in the perfume, food, and alternative medicine industries. Because it has an asymmetric carbon atom, it exists in optically active forms. The 3 (R) form has a flowery fresh scent, while the 3 (S) + form has a slightly woody, less fresh scent. Separation of the two enantiomers is accomplished using a chiral column. Linalool from synthetic sources will have a nearly equal amount of each enantiomer. In natural products, usually one isomer will predominate. The chromatographic separation of the two isomers can be used to identify origin of the material and identify possible adulteration or counterfeit products. The addition of racemic synthetics to essential oils is a known problem. This adulteration will alter their physiochemical properties with a potentially undesirable outcome. This work is based on an earlier application showing the separation of Linalool enantiomers using the original LTM system. [2] Deans switching or heat-cutting used in combination with two independent oven programs can be used to optimize separation of key compounds in both natural oils and synthetic formulations. Programming the chiral cyclodextin column to high temperatures that will result in bleed and column degradation can be avoided as well. In the configuration used in this work, it is not necessary to program the cyclodex column above 150 C, well below its maximum temperature of 200 C. The sample is injected into a DB5ms column for an initial separation. The deans switch is then used to cut specific sections of the DB5ms column to the LTM II module containing a cyclodextrin column. Linalool and other key compounds can be transferred in a single run to the LTM module for separation using its own optimized temperature program. Use of two independent ovens affords the analyst great flexibility to achieve the desired separation in optimal time. The LTM Series II incorporates several significant improvements over the original LTM hardware/firmware/software. These include: Full integration and control by the Agilent 7890A GC system Complete integration in Chemstation True constant flow mode of operation Real time display of temperatures/pressures/flows Experimental Restrictor sizing and flow rate pairing of the primary and secondary columns necessary to balance the system was determined using the Deans Switch Calculator [3]. It is worth noting that the Deans Calculator assumes all columns and restrictor are in the same oven with the same temperature program. Obviously, LTM /7890 systems to not comply with this constraint. Since the starting temperature difference between the air bath oven and LTM module is only 5 C, there is little impact on restrictor dimensions and the system will be perfectly balanced at the start of the run if restrictor size is based on air oven temperature. As the run proceeds and the temperature programs diverge (air oven and LTM), the restrictor flow and LTM flow may be different, but not enough to cause problems with the ratio of LTM column flow to primary column flow set at about 2 at initial conditions. In the system as described here, the restrictor flow starts at approximately 2.3 ml/min and decreases as the 7890 oven temperature rises to a flow of about 1.2 ml/min at 280 C. This restrictor can be sized from 0.6 m to 0.7 m for this system. 2

Firmware and Software Requirements Agilent 7890A GC system Firmware A.01.12.1 or greater Agilent GC ChemStation B.04.03 DSP1, includes LTM II software control Agilent MSD ChemStation E.02.02 LTM Series II System G6680A, 2 channel, two 5-inch format transfer line assemblies, and two power supplies. Columns ordered separately. Deans Switch G2855B or G3440A option #888 LTM Series II Column Configuration Defining a typical LTM column assembly involves specifying each segment as shown in Figure 1. Both dimension and heat source must be specified. This is necessary for the system to operate in true constant flow mode. Figure 1. IN Segment LTM Series II Column Segment OUT Segment Segment 2 GC Oven GC Oven MSD Terminology used to define an LTM Series II assembly. Agilent 7890A Options G3440A with options 112, 211, 309 System Configurations Diagram MSD System 5975C MSD Figure 2. 30 mm x 0.25 mm x 0.25 µm Cyclodex LTM II FID PCM 0.7 m x 0.100 mm restrictor Deans Switch 2 1 NO 3 Primary column 30 m x 0.24 mm x 0.25 µm DB5ms System diagram for the LTM Series II/ Deans/5975C system showing column and restrictor dimensions. Method Parameters Inlet Split/splitless, 280 C Split ratio 200:1 to 400:1 Primary column 30 m x 0.25 mm x 0.25 µm DB5ms, 1.3 ml/min (constant flow mode) 7890 Oven program 70 C (1 min) to 280 C (17 min) @ 10 C/min LTM II Program 75 C (7 min) to 130 C (5 min) @ 2 C/min Deans switch restrictor 0.7 m x 0.100 µm deactivated fused silica 5975C MSD scan 25 to 160 amu NC Split inlet Quad 175 C Source 230 C MSD transfer line 250 C LTM Module: Operated in constant flow mode, 2.5 ml/min Column 30 m x 0.25 mm x 0.25 µm CycloDex Insegment 0.5 m x.25 mm deactivated fused silica (heated by 7890 oven) Outsegment 0.8 m x.25 mm deactivated fused silica (heated by 7890 oven) Outsegment 2 0.17 m x.25 mm deactivated fused silica (heated by MSD transfer line) FID System Inlet Split/splitless, 280 C Split ratio 200:1 to 400:1 Primary column 30 m x 0.25 mm x 0.25 µm DB5ms, 1.3 ml/min (constant flow mode) 7890 Oven program 70 C (2 min) to 280 C (16 min) @ 10 C/min LTM II program 70 C (7 min) to 130 C (1 min) @ 2 C/min Deans switch restrictor 0.5 m x 0.100 µm deactivated fused silica LTM Module: Operated in constant flow mode, 2.5 ml/min Column 20 m x 0.25 mm x 0.25 µm CycloDex, constant flow @ 2.4 ml/min Insegment 0.5 m x.25 mm deactivated fused silica (heated by 7890 oven) Outsegment 0.5 m x.25 mm deactivated fused silica (heated by 7890 oven) Diagram FID System FID FID Figure 3. 20 mm x 0.25 mm x 0.25 µm Cyclodex LTM II 0.5 m x 0.100 mm restrictor PCM Deans Switch 2 1 NO 3 Primary column System diagram for the LTM Series II/Deans/FID system showing column and restrictor dimensions. NC Split inlet Samples Synthetic linalool standard, Fluka Chemical, Sigma-Aldrich Geranium extract (natural), cypress oil, jasmin extract (natural), balm leaf oil (natural), Sigma-Aldrich Detergent perfume additive, customer sample Orange oil 8X, customer sample All samples were diluted in methylene chloride as appropriate except the detergent additive, which was injected neat. 3

Results and Discussion MSD System Individual isomers of linalool or other optically active compounds present in natural oils cannot be separated on nonchiral columns such as DB5ms. This is shown in Figure 4 for analysis of geranium extract on a 30 m DB5ms using the monitor FID channel with no heart cuts to the chiral column. Linalool isomers and probably other minor constituents co-elute around 8.3 minutes. Injecting the extract directly on the chiral column is also not a sound approach as elution of all components may require a temperature exceeding the 200 C maximum of cyclodex. Bleed increases rapidly above 150 C and will shorten column life as well. 2,600,000 2,500,000 2,400,000 2,300,000 2,200,000 2,100,000 2,000,000 1,900,000 1,800,000 1,700,000 1,600,000 1,500,000 1,400,000 1,300,000 1,200,000 1,100,000 1,000,000 900,000 800,000 700,000 600,000 500,000 400,000 300,000 200,000 100,000 0 500,000 450,000 Linalool 400,000 350,000 300,000 250,000 200,000 150,000 100,000 50,000 0 7.00 7.50 8.00 8.50 9.00 9.50 10.00 10.50 2.00 4.00 6.00 8.00 10.00 12.00 14.00 16.00 18.00 20.00 22.00 24.00 26.00 28.00 Figure 4. Geranium extract without heart cutting. Sample analyzed on the 30 m DB5ms with FID. 4

A sample of pure synthetic linalool was used to establish the cut time for analysis on the cyclodex column. Nearly equal amounts of each enantiomer are seen as expected for synthetic linalool when separated on the chiral column. Figure 5 shows distribution of the R and S isomers. Using the established cut window, results for geranium extract is shown in Figure 5 where near baseline separation is obtained using a 2 C/min program rate on the LTM II chiral column. The absence of co-eluting compounds was confirmed with the MSD. Although this is a natural oil, nearly equal amounts of each linalool isomer is seen. The extraction process, perhaps hydrodistillation, may have caused racemisation to occur. Abundance TIC: 00301003.D\data.ms Figure 5. Heart cut to the chiral LTM column using the CFT deans switch. Cut times: 8.25 min to 8.38 min. Both enanitomers of linalool are separated. 5

Since there are a number of other important terpenoid compounds present in natural oils, it is possible to take advantage of the LTM II/Deans switch system for their analysis. For example, citronellol is of significant value and interest as a flavor and fragrance compound. It can be used for the synthesis of rose oxide, a powerful fragrance component with a strong fruity smell. Citronellol also exists as enantiomers. The S ( ) -B-citronellol is present in geranium oil. Multiple retention time based heart cuts in a single run allow analysis of linalool, citronellol, and a few other compounds of interest. The third cut window is wide for transfer of a few major compounds. An example of this analysis is shown in Figure 6. Linalool enantiomer distributions in other selected natural oils is given in Figures 7 through 10 for balm leaf oil, cypress oil, jasmin extract, and orange oil 8x, respectively. While linalool occurs in nature in both optical forms, all four of these oils show a prevalence of one enantiomer as is usually the case. Orange oil shows essentially enantiomerically pure linalool. Abundance 500,000 400,000 300,000 Geranium, cuts at: 8.25 to 8.38 (linalool), 9.3 to 9.5 (methone), 10.0 to 10.9 (citronellyl formate, B-citronellol, nerol) 2 3 4 6 200,000 100,000 1 5 Time --> 20.00 Abundance 100,000 50,000 0 Time --> 20.00 10.00 Figure 6. Three heart cuts in one chromatographic run used to separate linalool, methone, citronellyl formate, R ( ) -B-citronellol, and nerol on the cyclodex column. 6

TIC Cut 8.25 min to 8.38 min Linalool FID Figure 7. Balm leaf oil. FID, TIC, and zoom in on the linalool separation (upper right). TIC Terpinolene Linalool Figure 8. Linalool in cypress oil. 7

TIC Figure 9. Linalool in jasmine extract showing greater than 95% of one optical isomer. Figure 10. Orange oil 8x. TIC of linalool heart cut and FID monitor channels shown. 8

Finally, Figure 11 shows the heart cut from a complex detergent fragrance additive. Linalool appears to be of synthetic origin. FID System For more routine use, where peaks have already been identified, an FID only system can be used. In this configuration (Figure 3), a 20 m x 0.25 mm x 0.25 µm Cyclodex column was used. This allows for a faster analysis with a tradeoff of some loss of separation of the linalool enantiomers. Although in most cases, sufficient separation is still obtained to identify a material as containing natural or synthetic linalool. An example using cypress oil is shown in Figure 12 that can be compared to Figure 8 with the MSD/30 m LTM system. Elution occurs at about 14 minutes compared to 22 minutes with the 30 m cyclodex column. TIC Methyl ester Benzoic acid 2,6 dimethyl-2-octanol Linalool Tetrahydrolinalool 15.00 16.00 17.00 18.00 19.00 20.00 21.00 22.00 23.00 24.00 25.00 26.00 Figure 11. Analysis of linalool in a detergent fragrance. Linalool FID (monitor) FID (LIM) Figure 12. Analysis of linalool in geranium extract using the FID/Deans/LTM system. Cut times to the cyclodex column are 9.25 min to 10.05 min. 9

Conclusion Various constituents of essential oils and extracts have considerable value in a number of industries including foods, flavors, natural and synthetic perfumes, alternative medicine, and aromatherapy. Because adulteration is also of concern in many high value formulations, a reliable analytical method is needed to provide positive identification and separation of optical isomers. LTM series II incorporates a number of new features including full control integrated in Chemsttion, and operation in true constant flow mode that assists the analyst with development of new methods. This is especially true with more complex configurations involving the deans switch and the Agilent 5975C Series GC/MSD. The use of two independent ovens gives the analyst flexibility to optimize analysis of natural oils or synthetic formulations especially where chiral separations can be used for origin determination and authenticity. In many applications, analysis time will also be shorter since only compounds of interest are cut to the chiral column. References 1. Method Translation Software, downloadable from www.chem.agilent.com 2. Independent Column Temperature Control using an LTM Oven Module for Improved Multidimensional Separation of Chiral Compounds, Frank David and Matthew Klee, Application Note 5990-3428EN. 3. Deans Switch Calculator, available on the Agilent Capillary Flow Technology Accessories CD, included with CFT devices. For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. The systems described here can serve as valuable quality control tools for natural oils and perfumes. For routine QA, the simpler FID/FID can be considered where cost and ease of use is a priority. The combination of LTM/CFT Deans/5975C can serve as the core for a wide variety of analytical problems beyond the chiral applications highlighted here. Only a change in the column sets is required. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 2011 Printed in the USA June 8, 2011 5990-8151EN