Application Note. Pharmaceutical QA/QC. Author. Abstract. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Similar documents
Application Note. Author. Abstract. Pharmaceutical QA/QC. Siji Joseph Agilent Technologies, Inc. Bangalore, India

Application Note Pharmaceutical QA/QC. Agilent Application Solution. Authors. Abstract. Syed Salman Lateef Agilent Technologies, Inc.

Agilent 1200 Infinity Series HDR DAD Impurity Analyzer System for the Quantification of Trace Level of Genotoxic Impurity

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Performance characteristics of the Agilent 1290 Infinity Quaternary Pump

Ultrafast and sensitive analysis of sweeteners, preservatives and flavorants in nonalcoholic beverages using the Agilent 1290 Infinity LC system

Modernizing the USP Monograph for Acetaminophen

Fast and Flexible Optimization of Modifier Concentrations Using an Agilent 1290 Infinity LC System with Blend Assist

Fast methods for the determination of ibuprofen in drug products

Analysis of DNPH-derivatized Aldehydes and Ketones using the Agilent 1220 Infinity LC System with Diode Array Detector

Application. Authors. Abstract. Introduction. Hydrocarbon Processing

Quantification of growth promoters olaquindox and carbadox in animal feedstuff with the Agilent 1260 Infinity Binary LC system with UV detection

Automated Switching Between 1D-LC and Comprehensive 2D-LC Analysis

Analysis of Positional Isomers of Lapatinib with Agilent Poroshell 120 PFP Columns

Application Note. Authors. Abstract. Energy & Chemicals, Biofuels & Alternative Energy

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

High Performance Liquid Chromatography. Table 1: Allowed HPLC Adjustment of USP <621> and EP <2.2.46>

LC walk-up system using the Agilent 1200 Series LC Method Development Solution and Agilent Easy Access software. Application Note

Application Note. Author. Abstract. Xinlei Yang Agilent Technologies Co. Ltd Shanghai, China

Performance evaluation of the Agilent 1290 Infinity 2D-LC Solution for comprehensive two-dimensional liquid chromatography

High-Resolution Sampling 2D-LC for Pharmaceutical Impurity Analysis

Application Note. Agilent Application Solution Analysis of acaricides in honey. Authors. Abstract. Food

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Method Transfer between HPLC and UHPLC Instruments Equipment-related challenges and solutions

Using the Agilent 1290 Infinity II Multicolumn Thermostat with Extreme Temperatures and Flow Rates

USP Method Transfer and Routine Use Analysis of Irbesartan Tablets from HPLC to UPLC

Exceptional Selectivity of Agilent ZORBAX Eclipse Plus Phenyl-Hexyl Columns to Separate Estrogens

Determination of Beta-Blockers in Urine Using Supercritical Fluid Chromatography and Mass Spectrometry

Determination of Verapamil Hydrochloride Purity Using the Acclaim PA Column

High Resolution Fast LC

The new Agilent 1290 Infinity Quaternary LC

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Analysis of PEGylated Proteins with Agilent AdvanceBio SEC Columns

Rapid Screening and Analysis of Components in Nonalcoholic Drinks Application

Sensitive HILIC UHPLC-UV determination of steviol glycoside natural sweeteners

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

LC Technical Information

with diode array detection

Separation of Explosives in EPA 8330: Column Choices Optimize Speed, Resolution, and Solvent Use. Application. Authors. Abstract.

Developing a fast, generic method for rapid resolution LC with quadrupole MS detection. Application Note

Welcome to our E-Seminar: Choosing HPLC Columns for Faster Analysis Smaller and Faster

Accurate quantification of oleanolic acid and ursolic acid in traditional chinese medicine

Analysis of Beer by Comprehensive 2D-LC with the Agilent 1290 Infinity 2D-LC system

Achieve confident synthesis control with the Thermo Scientific ISQ EC single quadrupole mass spectrometer

Analysis of Metals, Halides, and Inorganic Ions Using Hydrophilic Interaction Chromatography

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

Routine MS Detection for USP Chromatographic Methods

Comprehensive 2D-LC Analysis of Chinese Herbal Medicine

Application Note. Author. Abstract. Food Testing & Agriculture. Edgar Naegele Agilent Technologies, Inc. Waldbronn, Germany

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Practical Applications of Method Translation Using the Agilent Method Translation Tool Rita Steed Inside Application Engineer Agilent Technologies

The Agilent InfinityLab 2D-LC Solution with Active Solvent Modulation

Figure 1. Structures for Vitamin B2 and Vitamin B1.

The Agilent 1260 Infinity Analytical SFC System with Time-of-Flight Mass Spectrometric Detection

Implementation of Methods Translation between Liquid Chromatography Instrumentation

Figure 1. Structures for Vitamin B2 and Vitamin B1.

Pharmaceutical Impurity Identification and Profiling Using Agilent Q-TOF LC/MS Combined with Advanced MassHunter Data Processing Software

Analysis of Saponins in Spine Date Seed using Agilent ZORBAX Solvent Saver HT (1.8 µm) Columns

BRIEFING. (EXC: K. Moore.) RTS C Propylparaben C 10 H 12 O Benzoic acid, 4 hydroxy, propyl ester; Propyl p hydroxybenzoate [ ].

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Value of Eclipse Plus C18 and ph Selectivity to Analyze Amine-Containing Drugs

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

SIMULTANEOUS RP HPLC DETERMINATION OF CAMYLOFIN DIHYDROCHLORIDE AND PARACETAMOL IN PHARMACEUTICAL PREPARATIONS.

Peptide and protein analysis by capillary HPLC Optimization of chromatographic and instrument parameters. Application. Angelika Gratzfeld-Huesgen

Streamlining the Analysis of Oral Contraceptives Using the ACQUITY UPLC H-Class System

Analysis of Phthalates in Body Wash using Solid-Supported Liquid-Liquid Extraction

Volume 6, Issue 2, January February 2011; Article-015

Application Note. Abstract. Authors. Pharmaceutical. William J. Long Agilent Technologies, Inc Centerville Road Wilmington, DE USA

STABILITY INDICATING METHOD OF RELATED IMPURITIES IN VENLAFAXINE HYDROCHLORIDE SUSTAINED RELEASE TABLETS

Agilent LC & LC-MS Cannabis Testing

Successfully Scaling and Transferring HPLC and UPLC Methods

The Secrets of Rapid HPLC Method Development. Choosing Columns for Rapid Method Development and Short Analysis Times

Method Development and Validation Of Prasugrel Tablets By RP- HPLC

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Bioanalytical Chem: 4590: LC-MSMS of analgesics LC-MS Experiment Liquid Chromatography Mass Spectrometry (LC/MS)

HPLC Preparative Scaleup of Calcium Channel Blocker Pharmaceuticals Application

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

Rapid quantification of Chinese medicine Zuo Jin Pill using rapid resolution liquid chromatography

Luna 2.5 µm C18(2)-HST. Advantages of 2.5 µm for increasing the speed of analysis while maintaining high efficiency

Validated HPLC Methods

Automated Sample Preparation in Quality Control of Eye-Drop Formulation

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Application Note. Agilent Application Solution Analysis of PAHs in soil according to EPA 8310 method with UV and fluorescence detection.

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

A Newly Approved ASTM Standard For Analysis of Thiophene in Benzene Using a Pulsed Flame Photometric Detector (PFPD)

SIMULTANEOUS ESTIMATION OF MEFENAMIC ACID AND PARACETAMOL IN SUSPENSION FORM BY USING UPLC

Assay Transfer from HPLC to UPLC for Higher Analysis Throughput

Application Note. Authors. Abstract. Pharmaceuticals

LC and LC/MS Column Selection Flow Chart

LC800. Smart HPLC. Until now UHPLC From now Smart HPLC

Fast Screening Methods for Steroids by HPLC with Agilent Poroshell 120 Columns

Fast Screening Methods for Analgesics and Non-Steroidal Anti-Inflammatory Drugs by HPLC

Amphetamines, Phentermine, and Designer Stimulant Quantitation Using an Agilent 6430 LC/MS/MS

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

Journal of Chemical and Pharmaceutical Research, 2013, 5(5): Research Article

Determination of underivatized aflatoxins B2, B1, G2, and G1 in ground hazelnuts by immunoaffinity solid-phase extraction with HPLC-FLD detection

Mashhour Ghanem 1 and Saleh Abu-Lafi 2 * ABSTRACT ARTICLE INFO

Development of Validated Analytical Method of Mefenamic Acid in an Emulgel (Topical Formulation)

Transcription:

Reducing analysis time and solvent consumption for isocratic USP assay methods with current and proposed USP guidelines using the Agilent 129 Infinity LC System An efficient way to reduce cost of analysis for tramadol assay Application Note Pharmaceutical QA/QC Author 4 23 Siji Joseph Agilent Technologies, Inc. Bangalore, India 4 5 1 15 Higher throughput (6 injections in 23 ) 5 1 15 Abstract In this Application Note a high pressure liquid chromatography (HPLC) isocratic method for a tramadol assay was optimized for varying column dimensions. The column dimensions were varied according to the currently existing or recently proposed Interim Revision Announcements (IRAs) of United States Pharmacopeia (USP) <621> chromatography general chapters. The intention was to achieve lower cost for assay analysis by saving analysis time and solvent. The Agilent 129 Infinity LC System was used with Agilent Poroshell and sub-2 µm columns were used to show the cost savings. The newly developed methods met USP system suitability requirements and ensured excellent performance for routine analysis without method revalidation 1. An 83% cost saving was achieved along with higher throughput and lesser solvent consumption compared to the existing USP method.

Introduction Providing accurate and efficient chromatographic results in shorter time while consug less solvent is one way to improve productivity of lab operations. The efficiency of a chromatographic separation can be evaluated by the length/particle size (L/dp) ratio. From this equation, it is clear that the efficiency is proportional to the column length and inversely proportional to the particle size. Therefore, the chromatographic efficiency can be improved by simple modifications in column dimensions. The USP method for tramadol (a bestselling analgesic drug) HPLC assay is an isocratic run on a 25-mm long column with 5-µm L1 particles 2. An Agilent ZORBAX Eclipse Plus C18, 4.6 25 mm, 5-µm column (p/n 95999-92) column was used to perform tramadol system suitability under USP conditions (Experiment 1). Official guidelines are available from USP on allowed deviations for column dimensions (Table 1) 1. Within the current guidelines, the particle size can be reduced up to 5%, to 2.5 µm. The Agilent Poroshell 1 packing has a solid core of 1.7 µm with a porous.5-µm thick outer layer and a total particle size of 2.7 µm. This column can be treated like a sub-2 µm column regarding performance. The use of an Agilent Poroshell column allows a reduction of the column length to achieve the same performance by maintaining a comparable L/dp ratio. To verify the acceptability of this modification for tramadol assay, a system suitability testing was performed using an Agilent Poroshell 1 EC-C18, 4.6 1 mm, 2.7-µm column (p/n 695975-92) (Experiment 2). USP is under discussion to relax the current restrictions on particle size for isocratic separations and incorporate a new limit based on L/dp ratio. The new proposed column dimension deviation limits for isocratic analysis is shown in Table 2 3. After the approval of the new proposed plan, the Agilent 1.8-µm particle columns can also be used for tramadol assay analysis as replacement for 5 µm. For a 25-mm column with 5-µm particle size, the L/dp ratio is 25, µm /5 µm, and gives a value of 5,. With a particle size of 1.8 µm instead 5 µm, a column length of 1 mm gives a higher L/dp value and promises better performance. When the column length is reduced by a factor of 2.5, analysis time will also reduced by this factor. The column flow can be further imized by reducing the id. The enhanced time and solvent saving as per the latest USP Interim Revision Announcement (IRA) proposal was illustrated in Experiment 3 by perforg system suitability testing of tramadol using a ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8-µm column (p/n 959758-92). The new flow rate Table 1 Currently allowed column deviations as per USP <621> recommendations. F 2 = F 1 dc 2 dp 2 ( 1 2 dc 1 dp 2 ) Column parameter USP limit for deviation Length ±7% Internal diameter No limit, but keep constant linear velocity Particle size 5% for Experiment 3 was calculated using Equation 1 mentioned in the USP IRA proposal on general chromatography chapters <621> which is currently open for discussion and will be published in USP36-NF31 1S 3. Equation 1 Where F 1 and F 2 are the flow rates for the original and modified conditions, respectively; dc 1 and dc 2 are the respective column diameters and dp 1 and dp 2 are the particle sizes. All these combinations such as reduction in column length and internal diameter along with the use of smaller particle size, can contribute to significant savings in analysis time and solvent. This results in higher throughput, less cost of analysis, and improved productivity of the lab without compromising on chromatographic performance. Column parameter USP limit for deviation Length and particle size No limit, but keep constant L/dp ratio or should be between -25% to +5% of the original L/dp ratio Internal diameter No limit, but keep constant linear velocity Flow rate ±5% Table 2 Proposed column deviations for isocratic run as per USP Interim Revision Announcements (IRAs). 2

Experimental Instruments The Agilent 129 Infinity LC System consisted of the following modules: Agilent 129 Infinity Binary Pump with integrated vacuum degasser (G42 A) and 35 µl Jet Weaver mixer. Agilent 129 Infinity High Performance Autosampler (G4226A) Agilent 129 Infinity Thermostatted Column Compartment (G1316C) Agilent 129 Infinity Diode Array Detector (G4212A) with Max-Light flow cell (1. µl dispersion volume, 1-mm path length) (G4212-68) Software The systems was controlled using the Agilent ChemStation C.1.3 software. Chromatographic parameters The mobile phase for the isocratic run was prepared as per the USP method. A premixed 3:7 solution of acetonitrile and water with.5% trifluoroacetic acid was used as the mobile phase. The detection was done at 27 nm. The detailed chromatographic method parameters for each experiment are tabulated in Table 3. Reagents and materials Commercially available USP reference standards for tramadol and corresponding related compound A were purchased from USP-India Private Ltd, Hyderabad, India. Acetonitrile was of super gradient grade and was purchased from Lab-Scan (Bangkok, Thailand). Highly purified water from a Milli Q water purification system (Millipore Elix 1 model, USA) was used for the experiment. Trifluoroacetic acid was used as additive and purchased from Aldrich (India). Procedure The system suitability solution was prepared as per USP method for tramadol described in USP 34 NF 29. System suitability solution:.5 mg/ml each of USP tramadol hydrochloride RS and USP tramadol hydrochloride related compound A (RCA) in mobile phase. Using all three experimental conditions, system suitability testing for tramadol assay method was performed. Parameter USP method (Experiment 1) Experiment 2 Experiment 3 Injection volume (µl) 8 1.7 Column Agilent ZORBAX Eclipse Plus C18, 4.6 25 mm, 5 µm (p/n 95999-92) Agilent Poroshell 1 EC-C18, 4.6 1 mm, 2.7 µm (p/n 695975-92) Flow rate (ml/) 1. 1..58 Acquisition rate (Hz) 1 1 Table 3 Detailed chromatographic parameters for all the three experiments. Agilent ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8 µm (p/n 959758-92) 3

Results and discussion System suitability results The peaks were well separated under all three experimental conditions, as shown in the chromatograms in Figure 1. The data in Figure 1 illustrates the reduction in analysis time for tramadol assay while reducing the column length from 25 mm to 1 mm. The observed back pressure with the 1.8-µm column was above 6 bars. Here, the 129 Infinity LC System can operate at higher pressures (up to 1, bar) and is best suitable for developing high throughput UHPLC methods with higher column back pressure. Six replicate injections were made under each set of experimental conditions to assess injector precision and were evaluated by calculating the relative standard deviation (RSD) for area and retention time. The system suitability results obtained under all three experimental conditions were tabulated in Table 4 and the results were found to be within the acceptance criteria. 4 DAD1 A, Sig=27, Exp 1, Agilent ZORBAX Eclipse Plus C18, 4.6 25 mm, 5 µm 3.979 4.535 5 1 15 DAD1 A, Sig=27, Exp 2, Agilent Poroshell 1EC C18, 4.6 1 mm, 2.7 µm 6 4 3 1 1.631 1.849 2 4 6 8 DAD1 A, Sig=27, Exp 3, Agilent ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8 µm.672.767.5 1 1.5 2 2.5 3 3.5 Figure 1 Separation of tramadol system suitability mix in USP method and newly developed cost saving methods. For uniformity, five times of the main peak retention time was considered as the run time. Results No. Test Limit Exp. 1 (USP) Exp. 2 Exp. 3 1 Resolution between tramadol related compound A and tramadol NLT 2. * 2.9 3.2 2. 2 RT RSD (%) NMT 2. #.9.8.7 3 Area RSD (%) NMT 2. #.14.13.34 4 RRT of related compound A.9.88.88.88 *NLT: Not less than #NMT: Not more than. Table 4 System suitability results obtained under all three experimental conditions. 4

Cost of analysis For the experiments, the cost of analysis was calculated with the cost of acetonitrile as US $ 6/L, the running cost of an instrument as US $ 8/hour, and the cost associated with waste solvent disposal as US $ 1.5/L. The approximate time required to complete one injection for tramadol USP assay method is approximately 22.7 utes and the total analysis cost is approximately US $ 3.6/injection. However, by adopting the conditions of Experiment 2, using a Poroshell 1 EC-C18, 4.6 1 mm, 2.7-µm column, the total cost of analysis can be reduced to US $ 12.5 and productivity can be increased by 2.5 times. Likewise, the total expense using a ZORBAX Eclipse Plus C18, 2.1 1 mm, 1.8-µm column under the conditions of Experiment 3 was calculated as only US $ 5.1 and the time required to complete assay analysis was just 3.84 utes. The graphical representation of analysis time, solvent consumption, and cost of analysis for all three experiments is given in Figure 2. Agilent sub-2 µm columns can be efficiently used to improve the speed of analysis. While achieving higher throughput, narrow bore sub-2 µm columns consume less solvent than conventional HPLC methods. The example shown here illustrates that by reducing the column length from 25 mm to 1 mm and the particle size from 5 µm to 2.7 µm, 59% of solvents can be saved. Reducing the particle to 1.8 µm with a narrow bore column, results in a 9% saving of solvents. 35 3 25 15 1 5 Analysis time () Run time kept as five times of main peak RT Total solvent consumption (ml) Amount of acetonitrile consumed (ml) Total cost (US$)/injection Figure 2 Solvent, time and total cost calculation for USP and newly developed cost saving methods. Experiment 1 (USP Method) Experiment 2 Experiment 3 5

Conclusion Two new methods were developed for tramadol assay by modifying existing USP recommended column dimensions. The usability of these new methods for daily assay purpose was confirmed by perforg system suitability testing with each new method. In the first method, column length and particle size were varied within the USP<621> limits that are currently in place resulting in a total cost saving of US$ 18/injection which is 59% of the USP method. In the second method, an Agilent sub-2 µm column was used as per the latest USP Interim Revision Announcements. This reduced the run time from 22.7 utes to 3.8 utes without compromising on system suitability criteria. With the time required to complete one injection under USP conditions, one can complete six injections using this method. That means, by using an Agilent 129 Infinity LC System with Agilent sub-2 µm columns for UHPLC method development a 6-fold increment in lab productivity can be achieved. When moving from a HPLC method using a 25-cm, 5-µm column to a UHPLC method with a 1-cm, 1.8-µm column, 9% of solvents and 83% of analysis time could be saved. References 1. Validation of analytical methods, Agilent Publication number 599-514EN, 11. 2. USP method for Tramadol assay, USP34 NF29 (1). 3. USP Pharmacopeial Forum (PF) 38(2) In-Process Revision: <621> chromatography (USP36-NF31 1S), Page 17 (12). 6

7