Plasma Diagnostics Introduction to Langmuir Probes

Similar documents
SURFACE ENGINEERING MASS SPECTROMETERS FOR THIN FILMS, PLASMA AND SURFACE ENGINEERING

Application of the EQP and EQS mass spectrometers to time resolved studies in ion beam and plasma processes

vacuum analysis plasma diagnostics surface science gas analysis

Hiden HPR60 Molecular Beam Mass Spectrometer System

Hiden SIMS Secondary Ion Mass Spectrometers. Analysers for surface, elemental and molecular analysis

Hiden EQP Systems. High Sensistivity Mass and Energy Analyzers for Monitoring, Control and Characterization of Ions, Neutrals and Radicals in Plasma.

Pressure & temperature sampling options for Hiden gas analysis systems

Hiden CATLAB Systems Microreactor for Catalysis Studies & Thermal Analysis

Mass Spectrometer A Comparison of Positive and Negative Ion RGA Methods

Langmuir Probes as a Diagnostic to Study Plasma Parameter Dependancies, and Ion Acoustic Wave Propogation

The Q Machine. 60 cm 198 cm Oven. Plasma. 6 cm 30 cm. 50 cm. Axial. Probe. PUMP End Plate Magnet Coil. Filament Cathode. Radial. Hot Plate.

Control of gas discharge plasma properties utilizing nonlocal electron energy distribution functions DOE-OFS Plasma Science Center

THIN FILMS, PLASMA & SURFACE ENGINEERING. Mass Spectrometers. for Thin Films, Plasma & Surface Engineering

The distorting effect of the ion current on electron temperature measured by an electric probe

Low Temperature Plasma Technology Laboratory

A Kinetic Theory of Planar Plasma Sheaths Surrounding Electron Emitting Surfaces

SNMS. SNMS Applications. Combined SIMS and SNMS

A novel sputtering technique: Inductively Coupled Impulse Sputtering (ICIS)

Lecture 6 Plasmas. Chapters 10 &16 Wolf and Tauber. ECE611 / CHE611 Electronic Materials Processing Fall John Labram 1/68

Mass Spectrometers. for Surface Analysis SURFACE ANALYSIS

Floating probe for electron temperature and ion density measurement applicable to processing plasmas

Extrel Application Note

Hiden EQP Applications

A comparison of emissive probe techniques for electric potential measurements in a Hall thruster plasma

A floating potential method for measuring ion density

Hiden Analytical Ltd

Plasma parameter evolution in a periodically pulsed ICP

FIB - SIMS. Focussed Ion Beam Secondary Ion Mass Spectrometry.

Robert A. Meger Richard F. Fernster Martin Lampe W. M. Manheimer NOTICE

Repetition: Practical Aspects

DOE WEB SEMINAR,

Using a Microwave Interferometer to Measure Plasma Density Mentor: Prof. W. Gekelman. P. Pribyl (UCLA)

Low Temperature Plasma Technology Laboratory

Effect of parallel velocity shear on the excitation of electrostatic ion cyclotron waves

RESIDUAL GAS ANALYSIS MASS SPECTROMETERS FOR RESIDUAL GAS ANALYSIS - RGA

Effect of Noble Gas. Plasma Processing Laboratory University of Houston. Acknowledgements: DoE Plasma Science Center and NSF

Spectroscopic Electron Temperature Measurement in Methane/helium Plasma During Diamond-like Carbon Coating

Effect of He and Ar Addition on N 2 Glow Discharge Characteristics and Plasma Diagnostics

Plasma diagnostics of pulsed sputtering discharge

Accurately Determining the Plasma Potential Using Emissive Probes

Plasma Diagnosis for Microwave ECR Plasma Enhanced Sputtering Deposition of DLC Films

Measurement of EEDF and Distribution of Primary Electron in a Bucket Ion Source

Measurement of electron energy distribution function in an argon/copper plasma for ionized physical vapor deposition

Courtesy of ESS and TheRGA web pages part of a series of application and theory notes for public use which are provided free of charge by ESS.

Application Note GA-301E. MBMS for Preformed Ions. Extrel CMS, 575 Epsilon Drive, Pittsburgh, PA I. SAMPLING A CHEMICAL SOUP

1) Jozef Stefan Institute, Jamova 39, 1000 Ljubljana, Slovenia

Zeta Potential Analysis using Z-NTA

Study of DC Cylindrical Magnetron by Langmuir Probe

Calibration of plasma heat flux measurement

Plasma Chamber. Fortgeschrittenes Praktikum I. Supervisors: Baran Eren, Dr. Marco Wisse, Dr. Laurent Marot. Abstract

Nordson MARCH Concord, CA, USA

LANGMUIR PROBE MEASUREMENTS IN 13.56MHz DISCHARGES

Dormant Cathode Plasma Properties and Erosion Analysis in a Multiple-Cathode, High-Power, Rectangular Discharge Chamber

Chaotic-to-ordered state transition of cathode-sheath instabilities in DC glow discharge plasmas

Experimental analysis of plasma conditions in an inverted cylindrical magnetron sputtering system

Relative Sensitivity RS Measurements of Gases

Keywords. 1=magnetron sputtering, 2= rotatable cathodes, 3=substrate temperature, 4=anode. Abstract

Measurement Features and Results of TAL D-55 Plume

P. Diomede, D. J. Economou and V. M. Donnelly Plasma Processing Laboratory, University of Houston

PRINCIPLES OF PLASMA DISCHARGES AND MATERIALS PROCESSING

Contents: 1) IEC and Helicon 2) What is HIIPER? 3) Analysis of Helicon 4) Coupling of the Helicon and the IEC 5) Conclusions 6) Acknowledgments

Characterization of an Oxygen Plasma by Using a Langmuir Probe in an Inductively Coupled Plasma

MODELING OF SEASONING OF REACTORS: EFFECTS OF ION ENERGY DISTRIBUTIONS TO CHAMBER WALLS*

CHAPTER 6: Etching. Chapter 6 1

Chapter 7 Plasma Basic

Retarding Field Analyzer for Ion Energy Distribution Measurement Through a Radio-Frequency or Pulsed Biased Sheath

Electron energy distribution function measurements and plasma parameters in inductively coupled argon plasma

Particle Transport Measurements in the LHD Stochastic Magnetic Boundary Plasma using Mach Probes and Ion Sensitive Probe

EFFICIENT PLASMA PRODUCTION IN LOW BACKGROUND NEUTRAL PRESSURES WITH THE M2P2 PROTOTYPE

Plasma Chemistry Study in an Inductively Coupled Dielectric Etcher

ELECTRON-cyclotron-resonance (ECR) plasma reactors

Feature-level Compensation & Control

Supporting Online Material for

Experimental investigation of double layers in expanding plasmas

Hong Young Chang Department of Physics, Korea Advanced Institute of Science and Technology (KAIST), Republic of Korea

Recombination and Decay of Plasma Produced by Washer Stacked Plasma Gun inside a Curved Vacuum Chamber

Density Collapse in Improved Confinement Mode on Tohoku University Heliac

P. Diomede, D. J. Economou and V. M. Donnelly Plasma Processing Laboratory, University of Houston

Effect of Biasing on Electron Temperature in IR-T1 Tokamak

New 2d Far Field Beam Scanning Device at DLR s Electric Propulsion Test Facility

ALD & ALE Tutorial Speakers and Schedule

Project report. Spacecraft Charging and Mitigation Methods. By Muhammad Azam. Abstract

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

An Experimental Study to Show the Effects of Secondary Electron Emission on Plasma Properties in Hall Thrusters

Secondary Ion Mass Spectroscopy (SIMS)

Effect of Pressure and Discharge Voltage on Plasma Parameters in Air seeded Arc-plasma

Calibration of Langmuir probes against microwaves and plasma oscillation probes

APPENDIX 1 DESCRIPTION OF HOT WIRE ANEMOMETER

Measurement of electron energy distribution in low-pressure RF discharges

INTRODUCTION TO SCA\ \I\G TUNNELING MICROSCOPY

Enhancement of an IEC Device with a Helicon Ion Source for Helium-3 Fusion

ETCHING Chapter 10. Mask. Photoresist

Experiments with a Supported Dipole

Effect of wall charging on an oxygen plasma created in a helicon diffusion reactor used for silica deposition

DPP06 Meeting of The American Physical Society. Production of negative ion plasmas using perfluoromethylcyclohexane (C 7 F 14 )

CATALYSIS & THERMAL ANALYSIS

Introduction to Plasma

Measurement of Absolute Argon Excited State Populations and Electron Energy Distribution Functions in an Ar-a-Si Plasma

OVERVIEW OF ASTRIUM EROSION / CONTAMINATION MODELLING TOOL AND VALIDATION ON ONERA EXPERIMENTAL RESULTS

Linker Dependent Bond Rupture Force Measurements in Single-Molecule Junctions

Transcription:

Plasma Diagnostics Technical Information Sheet 531 Plasma Diagnostics Introduction to Langmuir Probes Introduction A Langmuir Probe is a powerful plasma diagnostic tool which capable of determining the fundamental characteristics of a plasma; namely the ion number density, electron number density, electron temperature, electron energy distribution function (EEDF), floating potential and the plasma potential. Langmuir probes are routinely used to determine the plasma parameters in areas as diverse as: low pressure plasmas for materials processing, the design of ion sources & new plasma chambers and edge plasmas in fusion devices. This information is used to characterise individual systems and processes and also to design new systems. Appendix I explains in more detail the significance of each of the aforementioned parameters. The measurement itself is accomplished by inserting a small (usually cylindrical) probe into the plasma and applying a variable voltage to the probe whilst measuring the resulting current drawn from the plasma. The current as a function of probe voltage is referred to as the I-V characteristic and it is this characteristic which is analysed to determine the various plasma parameters. The probe dimensions are chosen to cause minimum disturbance to the plasma (typically of the order of 10 mm length and 0.15 mm diameter) and provide a localised measurement of all the key parameters with very good spatial resolution. Manufactured in England by: A simplified schematic of a probe circuit and a typical I-V characteristic are shown below. HIDEN ANALYTICAL LTD 40 Europa Boulevard, Warrington, WA5 7UN, England t: +44 (0) 195 4455 f: +44 (0) 195 416518 e: info@hiden.co.uk w: www.hidenanalytical.com

I V Plasma Probe Electrode The I-V characteristic represents the current to the probe from the plasma electrons and ions as a function of probe bias. To enable analysis, the I-V characteristic is divided into three regions - the ion saturation, electron transition, and electron saturation regions. This division is useful from the point of view of understanding the shape of the I-V characteristic and more importantly in analysing the data. Simplified Langmuir Probe Circuit I - Ion Saturation The ion saturation region occurs for very negative probe biases over which the negative probe potential repels all electrons and attracts only positive ions. - Electron Saturation In the electron saturation region the probe collects electrons and repels all ions. Ion Saturation Electron Transition V -80-60 -40-0 0 0 40 60 80 Typical I-V Characteristic Langmuir Probe Circuit The simple probe system consists of an electrode, an adjustable power supply and a means to measure the current. In practice, a computer based data acquisition and control system is usually employed to automate the task of sweeping the bias voltage and recording the resulting I-V characteristic. Once acquired the I-V characteristic can then be analysed according to established theories and techniques. I-V Characteristic Electron Saturation -Transition The intermediate transition region contains the electron energy information since the increasing positive probe voltage repels electrons of progressively higher energies and hence samples the distribution of electron energies. The transition region is a composite of both electron and ion current and for proper analysis ion current and electron current must be de-coupled. This is usually accomplished by fitting a function to the ion current deep in the saturation region and extrapolating it back through the transition zone. This ion current function is then subtracted from the raw I-V data to produce the electron current characteristic. The analysis of the characteristic then Technical Information Sheet 531 Hiden Analytical

follows utilising several well known theories of interpretation. Hiden s ESPION - Advanced Langmuir Probe automates this entire process The complete process of measuring the I-V characteristic, analysing and displaying the results (as a function of time, position or other external parameter) requires little prior knowledge by the operator. Limited (OML) technique of Laframboise and the cold ion approximation of Allen, Boyd and Reynolds (ABR). Hiden s Windows ESPsoft software offers both interpretations. The OML technique is considered briefly here. An overview of the entire automated analysis routine using the standard ESPION technique follows. Step by Step Analysis Overview - Floating Potential The floating potential is the easiest parameter to determine since it is by definition the voltage at which the probe collects no current. A simple scan of the I-V characteristic will yield the floating potential: In short, a line is fitted to a plot of I versus V in the ion saturation region as shown, and the slope of this line is used to calculate the ion density according to the relation: n i 0.000001 = 1.4x10 Ion Satuation 15 M ( amu) 1 / A ( m ) p ( slope) 1 / Transition Probe Current (A) 0.001 0.000 0 5 10 15 0 5-0.001 Probe Bias (V) Floating Potential (Ion Current) 0.000000 Slope Extrapolated Square of Ion Current in Transition -100-80 -60-40 -0 0 0 40 Probe Voltage -0.00 - Ion Number Density The ion saturation zone is analysed to give the ion number density. The two main (and easily implemented) theories that can be used are the Orbital Motion - Electron Number Density, Electron Temperature & Plasma Potential The square root of the ion current fit is added to the original I-V data to obtain a characteristic which consists of an electron current component only. This is shown below. Technical Information Sheet 531 3 Hiden Analytical

0.06 0.06 0.05 0.0008 0.05 Probe Current 0.04 0.03 0.0 0.01 0.00 + Ion Current Function 0.0006 0.0004 0.000 Square Root of Fitted Line to I-V in Ion Saturation = Electron Current 0.04 0.03 0.0 0.01 0.00-100 -80-60 -40-0 0 0 40 60 80 100 Probe Voltage 0.0000-100 -50 0 50 100 Probe Voltage -100-80 -60-40 -0 0 0 40 60 80 100 Probe Voltage To determine the electron temperature, plasma potential and electron density the natural logarithm of electron current versus probe potential is plotted from this electron current characteristic as shown below. n e = 3.73x10 1 T e = Slope 13 I A ( m p est ( amps) ) T ( ev ) e current at the plasma potential, is used in the equation shown above. - Electron Energy Distribution Function (EEDF) The EEDF is determined by the Druyvestyn method which uses the second derivative of electron current with respect to voltage as calculated from the electron current characteristic. - Ln(Isat ) f E 4 ( V ) = A e p me ( V p V ) e 1/ d I ( V ) e dv Ln(I e ) -4-6 -8-10 -1-14 Slope = -1/T e Plasma Potential and electron satuation current are determined at Intersection Point Electron Saturation Line 100 80 60 EEDF -16 f(e) 40-18 V p -0 0 0 40 60 80 100 Voltage Using the ln(ie) versus V plot, the electron temperature is found by fitting a line to the transition region as shown above and is equal (in units of ev, where 1 ev = 11600K) to the negative inverse of the slope of this line. The plasma potential is the probe voltage corresponding to the intersection of the fitted electron saturation and transition lines. To find the electron density, the electron saturation current, the electron In the above analysis steps the following variable have been used: Vf Vp V A p M 0 0 0 5 10 15 0 5 30 = Floating Potential = Plasma Potential = Probe Electrode Voltage = Probe Area = Ion Mass Electron Energy (ev) Technical Information Sheet 531 4 Hiden Analytical

m e n I n e = Electron Mass = Ion Density = Electron Density I esat = Electron Saturation Current e I I e = electron charge = I ion + I electron = electron current f E (V) = electron energy distribution as a function of probe voltage Technical Information Sheet 531 5 Hiden Analytical

Appendix I - definitions - Plasma Potential The plasma potential is the voltage inside the plasma. It is always positive with respect to the most positive body with which it is in contact. In many cases, the plasma potential provides a good indication of the energy of positive ions incident on surfaces of interest. - Floating Potential This is the potential at which an object in contact with the plasma collects no current. Or in other words, the potential attained by an electrically isolated body in the plasma. The separation between the floating and plasma potential are related though the electron energy distribution and ion mass. plasma. - Electron Energy Distribution Function (EEDF) In a typical cold laboratory plasma, the electrons are usually much more energetic than the ions and the majority of the energy within the plasma is contained in the electrons. Electrons usually have a spread of energies and knowing the distribution of these energies is of key importance since the electrons are largely responsible for many of the reactions in the plasma. The figure below shows the EEDF of a Maxwellian plasma for several electron temperatures. - Electron Density The electron density is simply the number of electrons per unit volume within the plasma. f(e) 0.5 0.4 0.3 0. T=1eV T=eV - Ion Density The ion density is the number density of the ions per unit volume within the plasma. Ions can be negative or positive. For an electropositive discharge, all ions formed in the plasma have a positive charge and the ion and electron densities in the bulk of the plasma are equal. For an electronegative discharge, the ions may be either positive or negative, in which case the sum of electron and negative ion densities must equal the positive ion density. 0.1 T=5eV T=10eV 0.0 0 10 0 30 40 50 Energy (ev) - Electron Temperature The electron temperature (assuming a Maxwellian distribution of velocities) is the mean energy of electrons in the Technical Information Sheet 531 6 Hiden Analytical

Appendix II - Frequently asked questions (FAQs) The information of the preceding sections provides a general overview of the Langmuir probe technique. The following FAQs and corresponding responses attempt to deal with issues arising in specific applications by describing the features of Hiden s ESPION probe which have been specifically developed. Q1 How can the ESPION help me? A1 The ESPION provides detailed information about the key parameters in your plasma process. Knowledge of these can assist in identifying processing problems, in designing and troubleshooting new processes and in checking for deviations between chambers. Q My process uses depositing / etching gases, will the ESPION probe become contaminated and will this affect the results? A The ESPION probe has been developed with these applications in mind. ESPsoft software enables the active probe tip to be automatically toggled between acquire and clean cycles between individual scans ensuring that data is always acquired with a clean probe tip. Cleaning is achieved by ion bombardment sputtering or by joule heating when collecting an electron current. Q3 My plasma is generated with RF voltages, will this affect the measurements? A3 In an RF plasma, a significant AC voltage can develop between the plasma and probe tip. The ESPION uses a passive technique incorporating a high frequency compensation electrode together with a string of inductors immediately behind the tip, effectively a tuned LC circuit, in order to remove this AC component and its harmonics. The inductors are air cooled to prevent degradation in hot environments. Q4 What type of tip materials are used? A4 ESPION can be supplied with many different types of tip materials including Molybdenum, Tungsten, Platinum etc. The tips are easily user interchangeable. Q5 My chamber is often coated with insulating materials, does this affect the measurements? A5 ESPION employs an integral low frequency reference electrode which compensates for chambers with poor or no ground reference. The reference probe also serves to track and eliminate low frequency noise due to mains/power supply instabilities. Q6 Can I use pulsed plasmas? A6 Yes. The ESPION probe is capable of making measurements in plasmas which are modulated upto 100 khz. The gating resolution is 1 microsecond. The complete timing electronics, gating, delay etc. are contained within the ESPION control unit and controlled by ESPsoft software so that only the pulse which modulates the plasma is required as an input. Technical Information Sheet 531 7 Hiden Analytical

Q7 Can I make spatially resolved measurements? A7 Yes. The ESPION can be supplied with an automated z-drive which is controlled via ESPsoft software allowing data to be taken as a function of position within the chamber. Technical Information Sheet 531 8 Hiden Analytical