GENERAL INSTRUCTIONS GENERAL PREPARATION

Similar documents
GENERAL INSTRUCTIONS GENERAL PREPARATION

The Cole-Parmer Nitrate Ion Electrodes are used to quickly, simply, accurately, and economically measure nitrate in aqueous solutions.

Use a pipet bulb when pipeting cyanide solutions, as these solutions are highly toxic.

COLE-PARMER LABORATORY NITRATE ION ELECTRODE INSTRUCTION MANUAL

C = concentration as ppm sulfide V t = volume of titrant at endpoint V s = volume of standard used (10 ml)

The Cole-Parmer Ammonium Ion Electrode is used to quickly, simply, accurately, and economically measure ammonium ions in aqueous solutions.

COLE-PARMER LABORATORY AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CALCIUM ION ELECTRODE INSTRUCTION MANUAL

1) A ph/mv meter or an ion meter, either line operated or portable.

GENERAL INSTRUCTIONS GENERAL PREPARATION

COLE-PARMER INDUSTRIAL AMMONIUM ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER LABORATORY CYANIDE ION ELECTRODE INSTRUCTION MANUAL

phoenix Electrode Company CARBON DIOXIDE GAS-SENSING ELECTRODE INSTRUCTION MANUAL 1. A ph/mv meter or an ion meter, either line operated or portable.

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER REPLACEABLE MEMBRANE CALCIUM ION ELECTRODES INSTRUCTION MANUAL

The Oakton Cadmium Ion Electrodes are used to quickly, simply, accurately, and economically measure cadmium ions in aqueous solutions.

COLE-PARMER LABORATORY CUPRIC ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL NITRATE ION ELECTRODE INSTRUCTION MANUAL

COLE-PARMER INDUSTRIAL POTASSIUM ION ELECTRODE INSTRUCTION MANUAL

NITRATE ION SELECTIVE ELECTRODE

COLE-PARMER INDUSTRIAL CALCIUM ION ELECTRODE INSTRUCTION MANUAL

The Oakton Lead Ion Electrodes are used to quickly, simply, accurately, and economically measure lead or sulfate ions in aqueous solutions.

phoenix Electrode Company Ammonia Gas-Sensing Electrode Instruction Manual

POTASSIUM ION SELECTIVE ELECTRODE

Operating Instructions for Ammonium ISE Specifications

phoenix Electrode Company Silver/Sulfide Ion Electrodes Instruction Manual 1. A ph/mv meter or an ion meter, either line operated or portable.

S ENTEK Unit 6 & 7 Crittall Court, Crittall Drive, Springwood Industrial Estate, Braintree, Essex, CM7 2SE Tel: +44 (0) Fax: +44 (0)

User Guide. Fluoroborate Ion Selective Electrode

CALCIUM SELECTIVE ELECTRODE

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

perfection Guidebook perfection Combination Iodide Electrode Successful Ion Measurement

COLE-PARMER INDUSTRIAL SILVER/SULFIDE ION ELECTRODE INSTRUCTION MANUAL

User Guide. Perchlorate Ion Selective Electrode

Sodium Sealed Electrode Sensor Bundle Product Number: ENSOD051

Fluoride Sealed Electrode Sensor Bundle Product Number: ENFLU049

LEAD ION SELECTIVE ELECTRODE

perfection Guidebook perfection Combination Cyanide Electrode Successful Ion Measurement

Ammonium Sealed Electrode Sensor Bundle Product Number: ENAMN020A

Direct Measurement ISE Method Method to 4.00 mg/l NO 3 N TISAB Solution

NO 500 DIN, NO 500 BNC NO 800 DIN, NO 800 BNC

Ion Selective Electrode Probe

Chloride, HR, Direct Measurement ISE Method Method g/l to 35 g/l Cl Powder Pillow ISA

NO 2. Orion Nitrite Electrode INSTRUCTION MANUAL. Orion 93-46, Orion 97-46, ionplus. Analyze Detect Measure Control

ph Probe Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth 1 99 C

READING A. INTRODUCTION CHE425L POTENTIOMETRY WITH K + ION-SELECTIVE ELECTRODE. Skoog, Holler and Crouch: Chapter 23 and Appendix 3.

Measurements with Ion Selective Electrodes: Determination of Fluoride in Toothpaste

Ion Selective Electrodes for the Laboratory. Presented by Chris Cushman OTCO Water Laboratory Analyst Workshop Thursday, May 14, 2015

Guide to ph Analysis.

Standard Operating Procedure for GLNPO Specific Conductance: Conductivity Bridge

[Agilent Technologies] [Electrochemistry Products] Quick Reference

Understanding ph Troubleshooting and Diagnostic information

User Guide. Cyanide Ion Selective Electrode

Ammonia Gas Sensing Combination Electrode

User Guide. High Performance Ammonia Ion Selective Electrode

User Guide. Bromide Ion Selective Electrode

User Guide. Iodide Ion Selective Electrode

perfection Guidebook perfection Combination Fluoride Electrode Successful Ion Measurement

Advanced Sensor Technologies, Inc. TEL: TOLL FREE: WOW-ASTI ( ) FAX:

INSTRUCTION MANUAL. Type: ORP500 C, ORP500 CD, ORP222 CD, ORP223 CD EN Introduction Helpful Operating Tips. 2

perfection Guidebook perfection Combination Cupric Electrode Successful Ion Measurement

Rapid RH Portable ph Meter. User's Manual

NH 3 -selective electrodes

User Guide. Carbon Dioxide Ion Selective Electrode

Ammonia Gas Sensing Combination Electrode

User Guide. Calcium Ion Selective Electrode

ph Probe Spear Tip Silver / silver chloride Reads Range / Resolution 95% in 1s Response time 100 PSI Max pressure 60m (197 ft) Max depth

User s Guide. Waterproof Palm ph Meter Model PH220. Introduction

User Manual. Thermo Scientific Orion Fluoride Ion Selective Electrode

Appendices towards the Operation Manual of Ultrasonic milk analyzer Lactoscan

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

PRETREATMENT TECHNICAL DATA SHEET A CHROME-FREE FINAL RINSE PRODUCT DESCRIPTION

Measuring Ammonia in Water and Wastewater using the Thermo Scientific Orion Dual Star ph/ise Meter

CHEM 334 Quantitative Analysis Laboratory

symphony Multisensor Probes

Experiment 7 Buffer Capacity & Buffer Preparation

STANDARD OPERATING PROCEDURE

symphony Reference Electrodes

User Guide. High Performance Ammonia Ion Selective Electrode

Determination of fluorides in aqueous samples using membrane, ion selective electrode (ISE)

PRETREATMENT TECHNICAL DATA SHEET CHROMIUM-FREE ORGANIC PASSIVATING RINSE PRODUCT DESCRIPTION

Scope and application: For water, wastewater and seawater. Distillation is required for wastewater and seawater.

MassWWP Standard Operating Procedure L-6 ph and Alkalinity Rev.0 Page 1 of 12

PHS-25CW/3BW Benchtop ph Meter Instruction Manual

PHS-25CW/3BW Benchtop ph Meter. Instruction Manual BANTE INSTRUMENTS CO., LTD

User Guide. ROSS Sodium Ion Selective Electrode

Volumetric Measurement Techniques. Technique #1 Use of a Burette. Technique #2 Use of a Pipette. Technique #3 Use of a Volumetric Flask

Hydrolysis of Salts Weak Acids and Bases

Chemistry Determination of Mixed Acids

All about ph! Is it as simple as it seems? The ph Workshop. Rick Noone Thermo Scientific Water Analysis Instruments, Orion products May 22, 2014

Electrodes for ph measurements. - when you need to be sure...

(b) Write the chemical equation for the dissolution of NaOH in water.

Titration of a Cola Product INSTRUCTOR RESOURCES

Cl - Orion Chloride Electrode INSTRUCTION MANUAL. Orion Orion ionplus. Analyze Detect Measure Control

Scientific Observations and Reaction Stoichiometry: The Qualitative Analysis and Chemical Reactivity of Five White Powders

Determination of the K a of a Weak Acid and the K b of a Weak Base from ph Measurements

GETTING THE END POINT TO APPROXIMATE. Two hours

RAPID KJELDAHL BENCHNOTES

Citric Acid Analysis L-6-1. MOISTURE (Karl Fischer)

Experiment#1 Beer s Law: Absorption Spectroscopy of Cobalt(II)

Buffer Preparation. Learning Objectives:

Transcription:

GENERAL INSTRUCTIONS Introduction The Van London-pHoenix Company Fluoroborate Ion Selective Electrode is used to quickly, simply, accurately, and economically measure Fluoroborate in aqueous solutions. Required Equipment 1. An ion meter 2. Fluoroborate Ion Sensing Electrode 3. Plastic syringe, pipets, volumetric flasks, 150 ml beakers Required Solutions 1. Deionized or distilled water for solution preparation. 2. Ionic Strength Adjuster (ISA), 2M (NH 4 ) 2 SO 4, Cat# BF4IS01. 3. Van London-pHoenix Company Fluoroborate Standard, 1000 ppm BF 4, Cat. No.BF4AS02. 4. Fluoroborate Electrode Reference Fill Solution, Cat# R001044. GENERAL PREPARATION Electrode Preparation Remove the rubber cap covering the electrode tip and the rubber insert covering the filling hole of the reference electrode. Fill the combination electrode with the filling solution shipped with the electrode to a level just below the fill hole. No preparation is required with a sealed reference electrode. Gently shake the electrode downward in the same manner as a clinical thermometer to remove any air bubbles which may be trapped behind the Fluoroborate membrane. Connect the electrode to the proper terminals of the meter as recommended by the meter manufacturer. Store the black shipping cap for later use. Electrode Slope Check (for Ion meters which display mv) 1. To a clean, dry, 150 ml beaker, add 100 ml of distilled water and 2 ml of ISA. After assuring that the meter is in the millivolt mode, lower the electrode tip into the solution. Stir moderately. Remove air bubbles on the clear plastic membrane by redipping probe. 2. Using a pipet, add 1 ml of 1000 ppm Fluoroborate standard into the solution. Stir moderately. After 1 minute, record the mv reading. 3. Using a pipet, add 10 ml of the 1000 ppm Fluoroborate standard to the beaker. Stir moderately. After 1 minute, record the mv reading. 4. Determine the difference between the two readings. The electrode is operating correctly if a slope difference of 52-59 mv is found, assuming the solution temperature is 25 o C. Slope is defined as the change in mv observed when the concentration changes by a factor of 10. Measurement using an Ion Meter (in the Concentration Mode) 1. By serial dilution of the 1000 ppm Fluoroborate standard, prepare two ammonia standards whose concentration is near the expected sample concentration. (e.g. 10 ppm and 100 ppm) For example, to make a 100 ppm standard, pipet 10 ml of the 1000 ppm standard into a 100 ml volumetric flask and dilute to volume with deionized water. Next to make a 10 ppm standard, pipet 10 ml of the newly-made 100 ppm standard into a 100 ml volumetric flask and dilute to volume with deionized water. A 1 ppm standard is made by further dilution of the 10 ppm standard. Measure out 100 ml of each standard into 2

individual 150 ml beakers. 2. Assure that the meter is in the concentration mode and set for a 2-point calibration. 3. Lower the electrode tip into the least concentrated solution. Begin stirring at a constant rate. Add 2 ml of ISA to the solution and continue stirring. 4. After 1 minute, adjust the meter to the concentration of the more dilute Fluoroborate standard and fix the value in the memory according to the meter manufacturer's instructions. 5. Rinse the electrode tip with distilled water and blot dry. 6. Lower the electrode tip into the more concentrated solution. Begin stirring at a constant rate. Add 2 ml of ISA to the solution and continue stirring. 7. After 1 minute, adjust the meter to the concentration of the more concentrated Fluoroborate standard and fix the value in the memory according to the meter manufacturer's instructions. 8. Add 100 ml of the sample and 2 ml of ISA in a 150 ml beaker. Lower the electrode tip into the solution. Begin stirring at a constant rate. Ensure that the meter is in sample mode. 9. After 1 minute, read the concentration directly from the meter display. 10. The electrode should be re-calibrated every 2-3 hours. Simply repeat Steps 2-7 above. Measuring Hints As Fluoroborate electrodes are used or stored for long periods, they will experience some deterioration in performance and slope errors will increase. By using the meter s calibration controls this error can be corrected. If an electrode is able to be calibrated and is stable and responsive, it is still a functional electrode and may be used in service even though it no longer meets new electrode specifications. All samples and standards should be at the same temperature for precise measurement. A difference of 1 o C in temperature will result in approximately a 2% error. Always rinse the electrode with distilled water and blot dry between measurements. Use a clean, dry tissue to prevent cross-contamination. Constant, but not violent, stirring is necessary for accurate measurement. Check the electrode for air bubbles adhering to the membrane surface after immersion in solution. Agitate the electrode gently to remove the air bubbles. The sensing membrane is normally subject to water uptake and might appear milky. This has no effect on performance. A slow responding electrode may be caused by interferences to the electrode. To restore proper performance, soak the electrode in distilled water for about 5 minutes to clean the membrane, rinse, and soak in standard solution for about 5 minutes. All samples must be aqueous and not contain organics which can dissolve in the membrane or extract out the liquid ion exchanger. Interferences should be absent. If they are present, use the procedures found in the Interferences section to remove them. The ph range for the Fluoroborate ion electrode is 2.51. Neutralize samples outside this range with acid or base to bring them in range. 3

Fluoroborate sample with either high or low ph must be analyzed immediately after preparation or hydrolysis of fluoroborate to BF 3 OH, BF 2 (OH) 2, and BF(OH) 3 will occur. Convert boric acid or borate ion to fluoroborate by addition of HF. ELECTRODE CHARACTERISTICS Reproducibility Electrode measurements reproducible to +2% can be obtained if the electrode is calibrated every hour. Factors such as temperature fluctuation, drift, and noise limit reproducibility. Interferences Certain anions are electrode interferences and will cause electrode malfunction, drift or measurement errors if present in high enough levels. The level of interfering common anions that will cause a 10% error at three levels of fluoroborate is given in Table 3. TABLE 3: Concentration of Possible Interferences Causing a 10% Error at Various Levels of Fluoroborate; Background Ionic Strength of 0.12M (NH 4 ) 2 SO 4. Interferences (ppm) 100 ppm 10 ppm 1 ppm -3 PO 4 7.0x10 4 7.0x10 3 7.0x10 2-2 HPO 4 7.1x10 4 7.1x10 3 7.1x10 2 H 2 PO 4 7.2x10 4 7.2x10 3 7.2x10 2 Cl 1.6x10 4 1.6x10 3 1.6x10 2-2 CO 3 1.7x10 4 1.7x10 3 1.7x10 2 HCO 3 1.7x10 4 1.7x10 3 1.7x10 2 NO 3 2.9x10 3 2.9x10 2 2.9x10 1 NO 2 4.2x10 2 4.2x10 1 4.2 Br 7.4x10 2 7.4x10 1 7.4 CN 1.2x10 2 1.2x10 1 1.2 ClO 3 4.0x10 1 4.0 4.0x10 I 6.0 6.0x10 6.0x10-2 ClO 4 5.0x10 5.0x10-2 5.0x10-3 Interferences such as phosphate, bromide, iodide and cyanide can be removed by precipitation with 0.5 grams of silver sulfate added to 100 ml of sample. Interferences such as carbonate and bicarbonate can be removed by conversion to carbon dioxide through acidification of the sample to ph 4.5 with sulfuric acid. Nitrite interference can be removed by adding 0.3 grams of sulfamic acid to 100 ml of sample. The above interference removal procedures require similar treatment of standards as well as samples. If the electrode is exposed to high levels of interfering ions which cannot be removed, the electrode reading may drift and the response may become sluggish. Restore performance by soaking in distilled water for 30 minutes followed by soaking in fluoroborate standard for 30 minutes. Temperature Influences The electrode response will shift and change slope with change in temperature. Standards and samples should be at 4

the same temperature. A 2% error results with a 1 o C temperature change for a 10 ppm solution. The electrode can be used at temperatures from 0-40 C. Room temperature measurements are recommended, since measurements at temperatures quite different from room temperature may require equilibrium times up to one hour. Electrode Response Plotting the mv potential against the Fluoroborate concentration results in a straight line with a slope of about, 56 mv between 10 ppm and 100 ppm at 25 C. For Fluoroborate concentrations above 10 ppm BF 4 the electrode exhibits good time response (95% of total mv reading in one minute or less). Response times are longer below this value. Limits of Detection Fluoroborate concentration down to 0.1 ppm Fluoroborate can be measured in neutral solutions. Since sample contamination can be a factor in low level Fluoroborate measurements, care must be taken in making determinations below 1.0 ppm. The upper limit of detection in pure sodium Fluoroborate solutions is 10,000 ppm. ph Effects The operating range of the Fluoroborate electrode is from ph 2.5 to ph 11. Electrode Life The Fluoroborate electrode will last six months in normal laboratory use. On-line measurement might shorten operational lifetime to several months. In time, the response time will increase and the calibration slope will decrease to the point calibration is difficult and electrode replacement is required. Since Fluoroborate electrodes have a limited shelf life, it is important to have a backup electrode which is in working condition when required. Electrode Storage The Fluoroborate electrode may be stored for short periods of time in 10 ppm Fluoroborate solution. For longer storage (longer than two weeks), rinse and dry the sensing tip and cover the membrane tip with any protective cap shipped with the electrode. The reference portion of the combination electrode (or the outer chamber of the reference electrode) should be drained of filling solution, if refillable, and the rubber insert placed over the filling hole. TROUBLESHOOTING HINTS Symptom Possible Causes Next Step Out of Range defective meter check meter with shorting strap (see meter Reading instruction manual) defective electrode electrodes not plugged in properly check electrode operation unplug electrodes and reseat 5

reference electrode not filled air bubble on membrane electrodes not in solution be sure reference electrode is filled remove bubble by re-dipping electrode put electrodes in solution Noisy or Unstable defective meter check meter with shorting strap Readings (readings continuously or air bubble on membrane remove bubble by re-dipping electrode rapidly changing) defective electrode replace electrode ISA not used meter or stirrer not grounded electrode exposed to interferences outer filling solution level too low use recommended ISA ground meter or stirrer soak electrode in Fluoroborate standard fill electrode to level just below the fill hole Drift (reading samples and standards allow solutions to come to room slowly changing at different temperatures temperature before measurement one direction) electrode exposed to soak electrode in Fluoroborate interferences standard incorrect reference filling solution use recommended filling solution Low Slope or standards contaminated prepare fresh standards No Slope or incorrectly made ISA not used standard used as ISA electrode exposed to interferences defective electrode air bubble on membrane use recommended ISA use ISA soak electrode in Fluoroborate standard check electrode operation remove bubble by re-dipping probe Incorrect Answer incorrect standards prepare fresh standards 6

wrong units used sample carryover apply correct conversion factor: 10-3 M = 86.8 ppm as BF 4 rinse electrodes thoroughly between samples SPECIFICATIONS Concentration Range: 10,800 ppm to 0.5 ppm ph Range: 2.5 to 11 Slope: 52 59 mv between 10 ppm and 100 ppm at 25 C Temperature Range: 0 to 40 C Interferences: ClO 4, I, CN Reproducibility: + 2% Size: 110 mm length 12 mm diameter 1 m cable length Storage: store in dilute Fluoroborate standard Samples: aqueous solutions only no organic solvents ELECTRODE THEORY Electrode Operation The Fluoroborate electrode consists of an electrode body containing a liquid internal filling solution in contact with a gelled organophilic membrane containing a Fluoroborate ion exchanger. When the membrane is in contact with a solution containing free Fluoroborate ions, an electrode potential develops across the membrane. This electrode potential is measured against a constant reference potential, using a standard ph/mv meter or an ion meter. The level of Fluoroborate ions, corresponding to the measured potential, is described by the Nernst equation: E = E' - S log X where: E = measured electrode potential E = reference potential (a constant) S = electrode slope ( 56 mv/decade) X = level of Fluoroborate ions in solution The activity, X, represents the effective concentration of the ions in solution. The total Fluoroborate ion concentration, C t, is the sum of free Fluoroborate ion, C f, and complexed or bound Fluoroborate ion, C b. The electrode is able to respond to only the free ions, whose concentration is : C f = C t - C b Since Fluoroborate ions form very few stable complexes, the free ion concentration may be equated to the total ion concentration. 7

The activity is related to the free ion concentration, C f, by the activity coefficient, γ, by: X = γ C f Activity coefficients vary, depending on total ions strength, I, defined as: I = 1/2 C x Z x 2 where: C x = concentration of ion X Z x = charge of ion X = sum of all of the types of ions in the solution In the case of high and constant ionic strength relative to the sensed ion concentration, the activity coefficient, γ, is constant and the activity, X, is directly proportional to the concentration. Rev. 6/25/10 [U:/Manuals/VL-PC Long Versions/Fluoroborate Manual.doc] 8