CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR

Similar documents
LAB REPORT ON XRF OF POTTERY SAMPLES By BIJOY KRISHNA HALDER Mohammad Arif Ishtiaque Shuvo Jie Hong

MT Electron microscopy Scanning electron microscopy and electron probe microanalysis

Overview of X-Ray Fluorescence Analysis

BENEFITS OF IMPROVED RESOLUTION FOR EDXRF

Determination of the activity of radionuclides

GLANCING INCIDENCE XRF FOR THE ANALYSIS OF EARLY CHINESE BRONZE MIRRORS

IMPLEMENTATION OF THE MONTE CARLO-LIBRARY LEAST- SQUARES APPROACH TO ENERGY DISPERSIVE X-RAY FLUORESCENCE ANALYSIS

MS482 Materials Characterization ( 재료분석 ) Lecture Note 4: XRF

Portable type TXRF analyzer: Ourstex 200TX

2. What is the wavelength, in nm, of light with an energy content of 550 kj/mol? a nm b nm c. 157 nm d. 217 nm e.

Small-Angle X-ray Scattering (SAXS)/X-ray Absorption Near Edge Spectroscopy (XANES).

Silicon Drift Detectors: Understanding the Advantages for EDS Microanalysis. Patrick Camus, PhD Applications Scientist March 18, 2010

1 of 5 14/10/ :21

International Journal of Scientific & Engineering Research, Volume 5, Issue 3, March-2014 ISSN

Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy. Chemistry 311: Instrumentation Analysis Topic 2: Atomic Spectroscopy

CHEM-E5225 :Electron Microscopy X-Ray Spectrometry

Summer Students lectures

X-Ray Photoelectron Spectroscopy (XPS) Auger Electron Spectroscopy (AES)

Radiation interaction with matter and energy dispersive x-ray fluorescence analysis (EDXRF)

Understanding X-rays: The electromagnetic spectrum

SEM. Chemical Analysis in the. Elastic and Inelastic scattering. Chemical analysis in the SEM. Chemical analysis in the SEM

X-ray fluorescence analysis - calibration of the X-ray energy detector

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

Multi Channel Analyzer (MCA) Analyzing a Gamma spectrum

CHEM*3440. X-Ray Energies. Bremsstrahlung Radiation. X-ray Line Spectra. Chemical Instrumentation. X-Ray Spectroscopy. Topic 13

Understanding X-rays: The electromagnetic spectrum

XRF books: Analytical Chemistry, Kellner/Mermet/Otto/etc. 3 rd year XRF Spectroscopy Dr. Alan Ryder (R222, Physical Chemistry) 2 lectures:

Quantitative XRF Analysis. algorithms and their practical use

OES - Optical Emission Spectrometer 2000

Analysis of Cadmium (Cd) in Plastic Using X-ray Fluorescence Spectroscopy

FUNDAMENTAL PARAMETER METHOD USING SCATTERING X-RAYS IN X-RAY FLUORESCENCE ANALYSIS

Elemental analysis by X-ray f luorescence. Sequential benchtop WDXRF spectrometer

print first name print last name print student id grade

Altitude influence of elemental distribution in grass from Rila mountain. Dr. E. Nikolova, Dr. A. Artinyan, S. Nikolova INRNE - BAS XRF Laboratory

THE COMPTON EFFECT Last Revised: January 5, 2007

OXEA - Online Elemental Analyzer

Detection of X-Rays. Solid state detectors Proportional counters Microcalorimeters Detector characteristics

Detecting high energy photons. Interactions of photons with matter Properties of detectors (with examples)

Peter L Warren, Pamela Y Shadforth ICI Technology, Wilton, Middlesbrough, U.K.

ADVANTAGES AND DISADVANTAGES OF BAYESIAN METHODS FOR OBTAINING XRF NET INTENSITIES

X-ray Spectroscopy. c David-Alexander Robinson & Pádraig Ó Conbhuí. 14th March 2011

hν' Φ e - Gamma spectroscopy - Prelab questions 1. What characteristics distinguish x-rays from gamma rays? Is either more intrinsically dangerous?

Standardless Analysis by XRF but I don t know what s in my sample!! Dr Colin Slater Applications Scientist, XRF Bruker UK Limited

RADIOACTIVE SAMPLE EFFECTS ON EDXRF SPECTRA

Aplicação da Tomografia Computadorizada 3D com Raios X Microfoco no exame de Peças Arqueológicas

Atomic Physics. Chapter 6 X ray. Jinniu Hu 24/12/ /20/13

Application of total reflection X-ray fluorescence spectrometry for trace elemental analysis of rainwater

X-ray Absorption Spectroscopy

X-ray spectrometry with Peltier-cooled large area avalanche photodiodes

TRACE ELEMENT ANALYSIS USING A BENCHTOP TXRF- SPECTROMETER

COMPARATIVE STUDY OF PIGE, PIXE AND NAA ANALYTICAL TECHNIQUES FOR THE DETERMINATION OF MINOR ELEMENTS IN STEELS

CASSY Lab. Manual ( )

Status of the AMoRE experiment searching for neutrinoless double beta decay of 100 Mo

B. X : in phase; Y: out of phase C. X : out of phase; Y: in phase D. X : out of phase; Y: out of phase

FUNDAMENTAL PARAMETER METHOD FOR THE LOW ENERGY REGION INCLUDING CASCADE EFFECT AND PHOTOELECTRON EXCITATION

Application of large-area avalanche photodiodes to energy-dispersive x-ray fluorescence analysis

FINDING A NEEDLE IN A HAYSTACK: PERFORMANCE EVALUATION OF PORTABLE XRF INSTRUMENTS FROM THREE MANUFACTURERS

Auger Electron Spectroscopy

PERFORMANCE OF A ROOM TEMPERATURE GAS PROPORTIONAL SCINTILLATION COUNTER IN X-RAY ANALYSIS OF METALLIC ALLOYS EXCITED WITH ALPHA PARTICLES

EDS User School. Principles of Electron Beam Microanalysis

X-Ray Fluorescence and Natural History

Calibration of the IXPE Instrument

Micro-XRF excitation in an SEM

PoS(X LASNPA)076. Titanium extraction from waste NORM. B.R. Pereira, M.A.G. Silveira, G. Fontana, P.R.S. Santos. M.A.Rizzutto, N. H.

Cauchois Johansson x-ray spectrograph for kev energy range

Exploiting the Speed of the Silicon Drift Detector to Open New Measurement Opportunities in Particle Analysis. Dale E. Newbury

Setting The motor that rotates the sample about an axis normal to the diffraction plane is called (or ).

X Rays & Crystals. Characterizing Mineral Chemistry & Structure. J.D. Price

Journal of Chemical and Pharmaceutical Research, 2012, 4(1): Research Article

MICRO-TOMOGRAPHY AND X-RAY ANALYSIS OF GEOLOGICAL SAMPLES

Micro-XRF: Principles, Methodology and Applications in CH Studies. Birgit Kanngießer

Lecture 9: Metal-semiconductor junctions

X-ray Absorption and Emission Prepared By Jose Hodak for BSAC program 2008

X-RAY SCATTERING AND MOSELEY S LAW. OBJECTIVE: To investigate Moseley s law using X-ray absorption and to observe X- ray scattering.

AXP Research group Analytical X-ray Physics

X-Ray Photoelectron Spectroscopy (XPS) Prof. Paul K. Chu

FXA UNIT G485 Module X-Rays. Candidates should be able to : I = I 0 e -μx

EDS Mapping. Ian Harvey Fall Practical Electron Microscopy

Pulse Height Analysis System (PHA) designed for W7-X Presented by Monika KUBKOWSKA

Making Lemonade: Using Graffiti to date Petroglyphs

Neutron Imaging at Spallation Neutron Sources

PART 1 Introduction to Theory of Solids

Chapter 10: Wave Properties of Particles

The first model was based only on the CANBERRA information which we. drawing. Canberra refused any other information.

Data report for elemental analysis of IMPROVE samples collected during April, May, June 2009 UC Davis Submitted June 18, 2010 SUMMARY

ISPA-Tubes with YAP:Ce Active Windows for X and Gamma Ray Imaging.

Calibration of A Small Anode Germanium Well Detector

EXTREME ULTRAVIOLET AND SOFT X-RAY LASERS

Semiconductor X-Ray Detectors. Tobias Eggert Ketek GmbH

INTRODUCTION Atomic fluorescence spectroscopy ( AFS ) depends on the measurement of the emission ( fluorescence ) emitted from gasphase analyte atoms

Developments & Limitations in GSR Analysis

SFT3000S Measurement of Sn-Cu Coating

The RX-6000 Sulfur in Oil Analyzer For the analysis of Sulfur in Oil and other elements from S to U

Geogenic versus Anthropogenic Metals and Metalloids

Principles of Neutron Imaging

In-Situ Analysis of Traces, Minor and Major Elements in Rocks and Soils with a Portable XRF Spectrometer*

- Light has properties of WAVES such as DIFFRACTION (it bends around small obstructions).

The Ones That Got Away: Neutron Leakage Spectra

DETERMINATION OF X-RAY TOTAL ATTENUATION COEFFICIENT IN Zr, Ag, In FOR ENERGY RANGE BETWEEN kev

DRAFT BRUKER XRF SPECTROSCOPY USER GUIDE: SPECTRAL INTERPRETATION AND SOURCES OF INTERFERENCE

Transcription:

CALCULATION OF THE DETECTOR-CONTRIBUTION TO ZIRCONIUM PEAKS IN EDXRF SPECTRA OBTAINED WITH A SI-DRIFT DETECTOR A. C. Neiva 1, J. N. Dron 1, L. B. Lopes 1 1 Escola Politécnica da Universidade de São Paulo Av. Prof. Luciano Gualberto, trav. 3, 38, CEP 558-4, São Paulo, Brazil {acneiva@usp.br} ABSTRACT Si-drift detectors may originate strong spurious peaks generated by the excitation of the built-in collimator. Peaks of these elements in the spectra may be attributed both to their presence in the object under analysis and to their presence in the ring, thus creating ambiguity in the interpretation and quantification of the results. Although an external collimator can be added to prevent photons to reach the ring, thus avoiding its excitation, this solution has several drawbacks. So, for most applications, the best approach is to allow this ring contribution to happen and estimate its value, so that the contribution from the object under analysis can be inferred. The intensities of Zr K-lines originated from a Zr ring, for instance, can be calculated taking into account the photoelectric absorption coefficients of Zr for each energy of the spectra above the respective absorption edge, the intensity of radiation, the probability of each line, and a geometric factor. The intensity profile of the radiation that reaches the ring can be taken directly from the spectrum, corrected for the detector efficiency and expurgated from escape and sum peaks. The geometric factor can be calculated by calibration with standard samples. This paper presents the attainment of these calibration constants and the estimation of the contribution of the collimator to the Zr peak, in a semi-portable EDXRF device. The X-Ray source was a ItalStrucutures W-tube, operated up to 6 kv. The measurements were made with a Ketek Si-drift detector with Zr collimator, with an upper measuring limit of around 29 kev. The effect of radiation above 29 kev was estimated from a) the background profile from 18 to 29 kev, and b) the estimated areas of K-peaks of elements which were detected through their L-peaks. The effect of the background was studied by using samples with no elements above Zr. The effect of excitation of Zr K- lines by measured characteristic lines was studied with samples containing Pd, Ag and Sn. INTRODUCTION Si-drift detectors may originate strong spurious peaks generated by the excitation of the built-in collimator [1, 2]. Peaks of these elements for instance Zr or Ag in the spectra may be attributed both to their presence in the object under analysis and to their presence in the ring, thus creating ambiguity in the interpretation and quantification of the results. Although an external collimator can be added to prevent photons to reach the ring, thus avoiding its excitation [2], this solution has several drawbacks: a) the sample-detector distance has to be increased, b) the count-rate will decrease, c) the analysis spot will be mainly defined by this collimator, and d) the collimator itself will generate spurious peaks. So, for most applications, the best approach is to allow this ring contribution to happen and estimate its value, so that the contribution from the object under analysis can be inferred [1]. Zr K-lines from the ring, for instance, will be excited by every photon with energy above the Zr-K absorption edge, and their intensity can be calculated taking into account the photoelectric absorption coefficients of Zr for each energy, the intensity of radiation, the probability of each line, and a geometric factor. The intensity profile of the radiation that reaches the ring can be taken directly from the spectrum, corrected for the detector efficiency and expurgated from escape and sum peaks. However, if the

detector was not configured to measure part of the spectra, this cannot be done, because every photon above the Zr-K absorption edge will excite the Zr ring. In this case, instead of calculating directly the effect, one should try to establish a mathematical relationship between the known part of the spectra and the intensity of the Zr-ring. For one side, the background average intensity above the Zr-K absorption edge is a candidate as representative of the whole continuum profile above this absorption edge. For other side, L- or M-peaks may be used to estimate characteristic peaks which exist above the upper limit of the detector. This paper presents the attainment of these relationships and the estimation of the contribution of the collimator to the Zr peak, in a semi-portable EDXRF device [2, 3]. MATERIALS AND METHODS The EDXRF spectrometer consists of an X-ray generator and an X-ray detector, and was developed at EPUSP. The generator is an Ital Structures W-tube, with voltage range from 22 to 6 kv and current from.12 to 1.5 ma. The detector is a Ketek AXAS Si-drift one, with an upper measuring limit of around 29 kev, with Peltier cooling, linked to a Ketek 4k-channel analog-digital converter. For the present measurements, a 1,3 mm-diameter Pb collimator was used in the incident beam, and the angle between the axis of the incident beam and the detector was 9 o, both at 45 o of the surface. The distance between sample and detector is 7 mm. Refrigerated water cooling was used in order to keep the detector case below 2 o C, thus assuring a peak FWHM of around 15 ev. Most measurements were made at 55 kv, for 6 s, with an adequate current to ensure a FWHM of about 15 kev for the main peaks. To establish the effect of the continuum above Zr K absorption edge on the intensity of Zr-K peaks, spectra of pure Si, Ti, Fe, Ni, Cu, W, Au and Pb were obtained at the standard configuration. To establish the effect of characteristic peaks above the absorption edge, pure Pd, pure Ag and a Sn-alloy were analysed. To check the validity of the developed model, spectra of standard alloys from BAS were obtained. The areas of the Zr peaks were measured by the PyMCA software [4]. The areas of the background above Zr-K absorption edge were calculated from the background fitting curves obtained in the software, in the mode linear polynomial with order 3. RESULTS Spectra for pure Si, Ti, Fe, Ni, Cu, Zr, Pd and Ag are shown in Figure 1. All of them present K-peaks of the constituent element and the spurious Zr K- and L-peaks. The average intensity of the background above the Zr-K absorption edge for these spectra is shown as function of the atomic number in Figure 2. The decrease of this intensity with Z allows a very good exponential fitting, with R 2 =.9879. The area of the spurious Zr-K peaks also decreases with Z for the elements below Zr, as can be seen in Figure 3 (it should be remarked that the fitting of Zr-K peaks required extreme care in the case of Ni and Cu, because of the partial overlapping of these peaks with Ni-K and Cu-K pile-up peaks). This behaviour also allows a very good exponential fitting, with R 2 =.993. The exponential factors of the curves of background intensity and Zr peak areas versus Z present close exponential factors (-2.5755 and -2.3965, respectively), which suggests that a reasonable relationship can be established directly between these values with a lower exponential factor. This is confirmed in Figure 4, which shows a good exponential fitting: A bs = 17. B.8175 (equation I) where A bs is the intensity of the Zr-K peak ( b accounts for background and s accounts for spurious ) and B is the average background intensity.

background (counts/channel mc) log(counts) (arbitrary scale) log(counts) (arbitrary scale) 18 18 16 16 14 892ac - Si 14 8717aa - Cu 12 12 1 1 8318ab - Ti 8314ad - Zr 8 8 6 6 8318aa - Fe 855aa - Pd 4 4 2 2 8318af - Ni 5 1 15 2 25 3 energy (kev) 8314ag - Ag 5 1 15 2 25 3 energy (kev) Figure 1 Spectra of pure Si, Ti, Fe,Ni, Cu, Zr, Pd and Ag. Zr-K peaks are at 16.62 and 18.62 kev. 1.6 1.4 1.2 y = 122.9x -2.5755 R 2 =.9879 1.8.6.4.2 1 15 2 25 3 35 4 45 5 atomic number Figure 2 Average intensity of the background above the Zr K-absorption edge versus Z

Zr KL3 peak area (counts/mc) Zr KL3 peak area (counts/mc) 25 2 y = 135935x -2.3965 R 2 =.993 15 1 5 12 14 16 18 2 22 24 26 28 3 atomic number Figure 3 Area of the spurious Zr-K peaks for the spectra versus Z 25 Si 2 y = 169.99x.8175 R 2 =.9968 15 1 Ni Fe Ti 5 Cu.1.3.5.7.9 1.1 1.3 1.5 1.7 background (counts/channel mc) Figure 4 Area of the spurious Zr-K peaks as function of the average intensity of the background above the Zr K- absorption edge Such a good fitting, however, could not be established for heavier elements. If they have characteristic peaks above the Zr-K absorption edge (as Pd and Ag in Figure 1, for instance), the spurious Zr-K peaks are much more intense than predicted by equation 1. The effect of these characteristic peaks can be estimated taking into account the photoelectric absorption coefficients of Zr for each energy, the intensity of radiation and the probability of each line, giving rise to a theoretical Zr-K peak area, A ts, in an arbitrary scale ( t accounts for theoretical and s for spurious ). To obtain an absolute theoretical value, however, the geometry of the spectrometer and specifically that of the detector ring should be considered. Instead of that, we obtained calibration constants. For this, the theoretical effect was calculated without any geometrical considerations, assuming an arbitrary intensity scale. These

Kg values were compared to the difference between the experimental Zr-K peak-areas and the background estimated contribution. Table 1 shows the areas of the characteristic peaks of Pd and Ag, the theoretical areas of the spurious Zr-K peaks (A ts ) which would be generated by those characteristic peaks (in an arbitrary scale), the experimental Zr-K peak area (A exp ), the average background intensity and the estimated contribution of the background from equation 1. From the difference A ep ( ep accounts for effect of peaks ), given by A ep = A exp - A bs, a calibration constant (K g = A ts / A ep ) for the effect of the characteristic peaks is obtained as a linear function of the wavelength of the exciting characteristic peak (a geometrical average of the wavelengths of K and K was taken for this): K g ( ) = -3.636 + 8.526 (equation 2) where is expressed in Angstrom. As shown in Figure 5, a good R 2 was obtained for this fitting:.9997. Table 1 Calculation of the constant K g. Areas are given in counts/mc. The average background intensity is given in counts/(channel mc). KL3 area KM3 area A ts A exp B A bs A ep K g Pd 222 433 1562 139.275 59 1249 1.25 Ag 2567 49 1563 1593.277 6 1533 1.2 Sn 493 75 18 525.713 129 396.45 1.3 1.2 1.1 1.9 y = 8.5259x - 3.6361 R 2 =.9997 Ag Pd.8.7.6 Sn.5.4.46.48.5.52.54.56.58 wavelength (Angstrom) Figure 5 Kg as function of the average wavelength of the exciting radiation Therefore, for a generic sample, the estimated total area A es of the spurious Zr-K peak is given by: A es = A bs + ( A ep (i) / K g ( ) ) (equation 3) where ( A ep (i) / K g ( ) ) is the sum of the calculated effects A ps (i) for each element i, corrected by the calibration constant. If the characteristic peaks of the sample are above the measuring energy limit of the detector, their intensities are estimated from the intensities of L- or M-peaks of the element present

log(counts) (arbitrary scale) in the spectrum. Therefore, the part of a Zr-K experimental peak that can be attributed to the true presence of Zr in the sample, A true, can be estimated by: A true = A exp A ts (equation 4) This model was applied to Zr-free copper-alloy standard samples, through a routine in the Fundamental Parameters software developed by our group. Figure 6 shows the spectra of these samples, Table 2 shows their nominal compositions and Table 3 shows the areas of K-peaks for S and Sb (the only peaks above the Zr-K absorption edge observed in these spectra), the estimated effect of them on the Zr-K spurious peak (A ep ), the corrected values (A ep /K g ) and its sum, the background and estimated effect of background (B and A bs, respectively), the total estimated intensity (A es ) and the experimental intensity (A exp ). One can see that a very good agreement between A es and A exp was obtained. 18 16 14 821al - Admiralty Brass 12 1 8 8221am - Phospor Bronze 6 4 8221an - Leaded Gunmetal 2 5 1 15 2 25 3 energy (kev) Figure 6 Spectra of three standard copper-alloy samples from BAS (Bureau of Analysed Samples, UK). Sn-K peaks are at 25.27 and 28.47 kev. Sb-K peaks are at 26.36 and 29.73 kev. Zr-K spurious peaks are partially overlapped by pile-up peaks.

Zr KL3 peak area (counts/mc) Table 2 Composition of three standard copper-alloy samples from BAS (Bureau of Analysed Samples, UK), in mass fractions 1 - Admiralty Brass 2 - Phosphor Bronze 3 - Leaded Gunmetal Mn.19.419.5 Fe.354.3 Ni.168.536.1 Cu.7436.8654.8391 Zn.2213.19.475 As.168.16.1 Sn.163.947.491 Sb.356.33.2 Pb.575.79.51 Table 3 Estimation of the spurious Zr-K peak area for the three BAS standard samples (see Table 2). Areas are given in counts/mc. The average background intensity is given in counts/(channel mc). sample / KL3 KM3 peak area area A ep (Å) K g A ep /K g sum B A bs A es A exp 1 Sn 241.92 49.97 1.65.393.45 24 Sb 45.9 9.54 1.86.376.28 6 3.17 39 69 71 2 Sn 1443.59 298.97 63.54.393.45 141 Sb 39.62 8.26 1.56.376.28 6 147.24 53 2 196 3 Sn 72.51 149.1 31.71.393.45 7 Sb.58.12.23.376.28.8 7.22 5 12 127 One should remark, however, that all this method relies on equation 1, based on samples with Z < 4, which do not present characteristic peaks above the Zr-K absorption edge. For heavier elements which also do not present these peaks (up to the maximum energy of the incident beam), however, the equation was observed not to be valid. Figure 7 shows the plot of the areas of the Zr-K peaks versus the background intensity for pure W, Au and Pb. Figure 8 shows their spectra. One observes that not only the background intensity does not follow the monotonous behaviour shown in Figure 2, but also that it cannot be used to predict the area of the spurious Zr peak. For samples with high contents of these elements, therefore, the whole spectrum should be necessary for this prediction. 5 4 Au (Z = 79) 3 2 W (Z = 74) 1 Pb (Z = 82).2.4.6.8 1 1.2 1.4 1.6 1.8 background (counts/channel mc) Figure 7 Area of the spurious Zr-K peaks as function of the average intensity of the background above the Zr K- absorption edge

log(counts) (arbitrary scale) 13 11 9 8314al - W 7 5 831aa - Au 3 1 8318ag - Pb -1 5 1 15 2 25 3 energy (kev) Figure 8 Spectra of pure W, Au and Pb. Zr-K peaks are at 16.62 and 18.62 kev. Pile-up peaks are observed above Zr-K spurious peaks. CONCLUSION Even if a complete spectrum is not available, the intensity of the spurious Zr-K peaks can be estimated from the intensity of the background and of characteristic peaks above the Zr-K absorption edge, so that the presence and the true content of Zr from the sample can be estimated. The model was successful for Zr-free copper-alloy standard samples. However, it should not give precise results for alloys containing high contents of elements above W. Acknowledgement: We are grateful for the financial support of Fapesp and CNPq for this project. References: 1. FERRETTI, M. Fluorescence from the collimator in Si-PIN and Si-drift detectors: problems and solutions for the XRF analysis of archaeological and historical materials, Nucl.Instr. and Met.in Phys. Res. B, v. 226, pp. 453-46, 24. 2. NEIVA. A. C. et al. Considering spurious peaks in the EDXRF analysis of metallic pre-columbian pieces of the Museum of Ethnology and Archaeology of the University of Sao Paulo, Heritage, Weathering and Conservation International Conference (26, Madrid). Leiden : Taylor & Francis, 26. v. 2. p. 65-611. 3. NEIVA. A. C. et al. Development of EDXRF quantitative analysis methodology for the study of ferrous alloys of historical importance, Anais do 1o Simpósio Latinoamericano de Métodos Físicos e Químicos em Arqueologia, Arte e Conservação de Patrimônio Cultural (LASMAC 27), CD-ROM, São Paulo: 28. 4. SOLÉ, V.A.; PAPILLON, E.; COTTE, M. ;SUSINI, W. J. A multiplatform code for the analysis of energy-dispersive X- ray fluorescence spectra. Spectrochem. Acta Part B 62 (27) 63-68.