Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network /2017

Similar documents
TA3 Dosimetry and Instrumentation EVALUATION OF UNCERTAINTIES IN THE CALIBRATION OF RADIATION SURVEY METER POTIENS, MPA 1, SANTOS, GP 1

TITLE: Air Kerma Primary Standard: Experimental and Simulation Studies on Cs-137

Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST. Ronaldo Minniti 1 and Ladislav Czap 2

Implementing the ISO Reference Radiations at a Brazilian Calibration Laboratory

M [scale units/s] of the system

Calibration of a pencil ionization chamber with and without preamplifier

COMPARISON OF ABSORBED DOSE TO AIR CALIBRATION FACTORS FOR A PARALLEL PLATE IONIZATION CHAMBER*

Design, construction and characterization of a portable irradiator to calibrate installed ambient dose equivalent monitors

Determination of Ambient Dose Equivalent at INFLPR 7 MeV Linear Accelerator

CALIBRATION OF IONIZATION CHAMBERS FOR COMPUTED TOMOGRAPHY BEAMS IN BRAZIL: THE PRESENT REALITY*

Commissioning of the Beta Secondary Standard (BSS2)

Air Kerma Primary Standard: Experimental and. Simulation Studies on Cs-137. J. Cardoso, L. Santos, C. Oliveira

Intercomparison of 131 I and 99m Tc Activity Measurements in Brazilian Nuclear Medicine Services

Intercomparison of Ionization Chambers in Standard X-Ray Beams, Radioprotection Level

CALIBRATION OF IN VITRO BIOASSAY METHODOLOGY FOR DETERMINATION OF 131 I IN URINE

Application of patent BR in well type ionization chambers

FINAL REPORT. European co-operation for Accreditation (EA) Interlaboratory Comparison IR3: Calibration of a Radiation Protection Dosimeter

EFFICIENCY SIMULATION OF A HPGE DETECTOR FOR THE ENVIRONMENTAL RADIOACTIVITY LABORATORY/CDTN USING A MCNP-GAMMAVISION METHOD

ABSORBED DOSE TO WATER MEASUREMENTS IN HIGH ENERGY ELECTRON BEAMS USING DIFFERENT PLANE PARALLEL CHAMBERS *

Activity determination of a 201 Tl solution by 4πβ-γ and sum-peak coincidence methods

Study of the influence of phantom material and size on the calibration of ionization chambers in terms of absorbed dose to water

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the NIM, China and the BIPM in 60 Co gamma radiation

Calibration of Radioprotection Instruments and Calibrated Irradiation: Characterization of Gamma Beam of 137 Cs and 60 Co

Key comparison BIPM.RI(I)-K7 of the air-kerma standards of the CMI, Czech Republic and the BIPM in mammography x-rays

Radiation Protection Dosimetry (2011), Vol. 144, No. 1 4, pp Advance Access publication 30 November 2010

Hp(0.07) Photon and Beta Irradiations for the EURADOS Extremity Dosemeter Intercomparison 2015

Sensitivity of the IRD whole-body counter for in vivo measurements in the case of accidental intakes

Volume 1 No. 4, October 2011 ISSN International Journal of Science and Technology IJST Journal. All rights reserved

Comparison of national air kerma standards for ISO 4037 narrow spectrum series in the range 30 kv to 300 kv

Correlation Effects in the Uncertainty Estimation of Two-Pressure Humidity Generators

Analysis of drift correction in different simulated weighing schemes

Specific Accreditation Criteria Calibration ISO/IEC Annex. Ionising radiation measurements

Primary Standardization of 152 Eu by 4πβ(LS) γ (NaI) coincidence counting and CIEMAT-NIST method

Referensdosimetri. Crister Ceberg Medical Radiation Physics Lund University Sweden

CONSIDERATION ON THE H p(10) AND H*(10) SECONDARY STANDARD CHAMBER CHARACTERISTICS

City University of Hong Kong

Accuracy, Calibration, Type Testing and Traceability General. Peter Ambrosi

CHARACTERIZATION OF A RADIATION DETECTOR FOR AIRCRAFT MEASUREMENTS

Technical Protocol of the Bilateral Comparison in Primary Angular Vibration Calibration CCAUV.V-S1

KCDB-Identifiers: EURAMET.RI(I)-K1.2 / EURAMET.RI(I)-K4.2 / EURAMET.RI(I)-K5.1. Pilot laboratory: BEV (AT) Andreas Steurer 1), Wilhelm Tiefenboeck

Key comparison BIPM.RI(I)-K3 of the air-kerma standards of the BEV, Austria and the BIPM in medium-energy x-rays

ARTICLE IN PRESS. Applied Radiation and Isotopes

Study for the requalification of Inmetro's Primary Hardness Standardization Machine for Vickers HV3 scale

ABSORBED DOSE BEAM QUALITY FACTORS FOR CYLINDRICAL ION CHAMBERS: EXPERIMENTAL DETERMINATION AT 6 AND 15 MV PHOTON BEAMS

Key comparison BIPM.RI(I)-K1 of the air-kerma standards of the LNE-LNHB, France and the BIPM in 60 Co gamma radiation

Laboratoire National Henri Becquerel (CEA/LIST/LNHB), France (2) ENEA-Radiation Protection Institute, Bologna, Italy (3)

High-Energy Photon Beam Therapy Dosimetry with Ionisation Chambers

INVESTIGATIONS OF THE FRICKE GEL (FXG) DOSIMETER DEVELOPED AT IPEN IRRADIATED WITH 60 Co GAMMA RAYS

A Measuring System with Recombination Chamber for Photoneutron Dosimetry at Medical Linear Accelerators

Technical Protocol of the CIPM Key Comparison CCAUV.V-K5

Determination of the cosmic-ray-induced neutron flux and ambient dose equivalent at flight altitude

CURRENT CAPABILITIES OF THE IRD-CNEN WHOLE BODY COUNTER FOR IN VIVO MONITORING OF INTERNALLY DEPOSITED RADIONUCLIDES IN HUMAN BODY

Comparison of air kerma measurements for tungsten anode based mammography x-ray beam qualities (EURAMET.RI(I)-S4.1)

Study of the uncertainty in the determination of the absorbed dose to water during external beam radiotherapy calibration

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Trace element quality control analysis of environmental samples at the Neutron Activation Analysis Laboratory, IPEN, São Paulo, Brazil

Key comparison BIPM.RI(I)-K4 of the absorbed dose to water standards of the METAS, Switzerland and the BIPM in 60 Co gamma radiation

Analysis distribution of galactic cosmic rays particle energy with polar orbit satellite for Geant4 application

Ionizing Radiation Metrology at NMIJ/AIST and ENEA-INMRI

8/2/2012 UPDATING TG-51. When will it end? Part 1 - photon addendum. What are these updates? Photons: Electrons: More widespread revision required

Bureau International des Poids et Mesures

TECHNIQUE OF MEASURING THE FIRST IONIZATION COEFFICIENT IN GASES

Application of CFX code in Nuclear Reactor Thermal Hydraulics at the Brazilian Nuclear Technology Development Center (CDTN)

Maria Ângela de B. C. Menezes. Radojko Jaćimović. Cláubia Pereira

CERTIFICATE OF CALIBRATION

Jeancarlo Torres, CHP. Naval Surface Warfare Center Carderock Division

Retrieval Of GammaCell 220 Irradiator Isodose Curves With MCNP Simulations And Experimental Measurements

An extremity (wrist) dosemeter based on the Landauer InLight TM whole body dosemeter

(São Paulo, Brazil) Instituto de Pesquisas Energéticas e Nucleares - IPEN/CNEN-SP, Caixa Postal 11049, Pinheiros, São Paulo, Brasil 2

General Overview of Gas Filled Detectors

Half-life data a critical review of TECDOC-619 update

Dosimetry: Electron Beams

8/2/2012. Larry DeWerd has a partial interest in Standard Imaging

Gate simulation of Compton Ar-Xe gamma-camera for radionuclide imaging in nuclear medicine

Reliability and Traceability in Radiation Calibration

Bureau International des Poids et Mesures

Metrological traceability and specific needs in: - IR measurement for radiation protection (RP) - IR measurement for radiotherapy (RT)

Abstract An indirect comparison of the standards for reference air kerma rate for

Interlaboratory Comparison & Proficiency Testing Program

Larry A. DeWerd, PhD, FAAPM UW ADCL & Dept. Medical Physics University of Wisconsin

Tritium concentration analysis of groundwater samples from the environmental monitoring program at IPEN, São Paulo, Brazil

Comparison of the standards for absorbed dose to water of the ENEA-INMRI (Italy) and the BIPM for 60 Co γ rays

Results of the 2015 national indoor radon intercomparison measurements in Serbia

EVALUATION OF Ra, Th, K AND RADIUM EQUIVALENT ACTIVITY IN SAND SAMPLES FROM CAMBURI BEACH, VITÓRIA, ESPÍRITO SANTO, BRAZIL.

Interlaboratory Comparison & Proficiency Testing Program

An industrial radiography exposure device based on measurement of transmitted gamma-ray intensity

Correction for PMT temperature dependence of the LHCf calorimeters

Comparison of measurement uncertainty budgets for calibration of sound calibrators: Euromet project 576

RECOMBINATION PARAMETERS OF SOME FABRICATED IONIZATION CHAMBERS

STUDY OF THE BEST MEASUREMENT CAPABILITY IN ROCKWELL SCALE AT THE BRAZILIAN NMI INMETRO S PRIMARY HARDNESS STANDARD MACHINE

ISO INTERNATIONAL STANDARD

Measurement of Tritium in Helium

Department of Radiation Protection, Nuclear Science Research Institute, Japan Atomic Energy Agency

Precision neutron flux measurement with a neutron beam monitor

DEVELOPMENT OF A METHODOLOGY FOR LOW-ENERGY X-RAY ABSORPTION CORRECTION IN BIOLOGICAL SAMPLES USING RADIATION SCATTERING TECHNIQUES

Comparison of the standards of air kerma of the BNM-LPRI and the BIPM for 137CS y rays

Ion-Chamber Survey Meter OD-02

PHYS 3650L - Modern Physics Laboratory

Determination of absorbed dose to water for 60Co by the scaling theorem. M. Boutillon and A.-M. Perroche (1)

Interlaboratory Comparison & Proficiency Testing Program

Transcription:

Journal of Physics: Conference Series PAPER OPEN ACCESS Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 To cite this article: TS Cabral et al 2018 J. Phys.: Conf. Ser. 975 012042 Related content - Comparison of the air kerma standards for 137 Cs and 60 Co gamma-ray beams between the IAEA and the NIST Ronaldo Minniti and Ladislav Czap - Geant4 simulation of ion chambers response to 60Co spectrum of LNMRI/IRD Shepherd 81-14D Radiator P P Queiroz Filho and C N M Da Silva - New thermal neutron calibration channel at LNMRI/IRD A Astuto, K C S Patrão, E S Fonseca et al. View the article online for updates and enhancements. This content was downloaded from IP address 148.251.232.83 on 23/04/2018 at 00:57

Comparison for Air Kerma from Radiation Protection Gamma-ray Beams with Brazilian Network - 2016/2017 TS Cabral 1, CNM da Silva 1, MPA Potiens 2, CMA Soares 3, RR Silveira 4, H Khoury 5, V Saito 5, E Fernandes 6, WF Cardoso 7, HPS de Oliveira 7, MA Pires 8, AS de Amorim 9, M Balthar 9. 1 Instituto de Radioproteção e Dosimetria, Rio de Janeiro, Brazil; 2 Instituto de Pesquisas Energéticas e Nucleares, São Paulo, Brazil; 3 Centro de Desenvolvimento de Tecnologia Nuclear, Minas Gerais, Brazil; 4 Centro Regional de Ciências Nucleares do Nordeste, Pernambuco, Brazil; 5 Departamento de Energia Nuclear da UFPE, Pernambuco, Brazil; 6 Laboratório de Ciências Radiológicas da UERJ, Rio de Janeiro, Brazil; 7 Eletronuclear Eletrobrás Termonuclear S.A., Rio de Janeiro, Brazil; 8 MRA Comércio de Instrumentos Eletrônicos Ltda, São Paulo, Brazil.; 9 Centro Tecnológico do Exercito, Rio de Janeiro, Brazil. E-mail: tschirn@ird.gov.br Abstract. The results of the comparison involving 9 laboratories in Brazil are reported. The measured quantity was the air kerma in 137 Cs and 60 Co, at the level of radioprotection. The comparison was conducted by the National Laboratory Metrology of Ionizing Radiation (LNMRI/IRD) from October 2016 to March 2017. The largest deviation between the calibration coefficients was 0.8% for 137 Cs and 0.7% for 60 Co. This proficiency exercise proved the technical capacity of the Brazilian calibration network in radiation monitors and the results were used by some in the implementation of the standard ISO/IEC 17025. 1. Introduction One of the activities inherent in calibration laboratories is the periodic participation in Comparison programs where, among others, it is possible to assess the ability of the qualified technician to carry out a calibration service, in accordance with instructions laid down in procedures documented in appropriate environmental conditions and under control, getting a result consistent with the measurement accreditation as stated by the labs Participation in exercises of this nature, proficiency testing, is one of the requirements of norm ABNT ISO/IEC 17025 BR. The use of calibrated instruments for the purposes of radiological protection is a requirement that ensures the safe use of the various areas in which we work with ionizing radiation. These calibrations should be made based on measurement standards and following written procedures. The National Laboratory Metrology of Ionizing Radiation (LNMRI/IRD/CNEN) organized and conducted this Content from this work may be used under the terms of the Creative Commons Attribution 3.0 licence. Any further distribution of this work must maintain attribution to the author(s) and the title of the work, journal citation and DOI. Published under licence by Ltd 1

comparison exercise from October 2016 to March 2017. The protocol was structured according to the ISO 17043. The quantity compared is the calibration coefficient in air kerma in 137 Cs and 60 Co beams. Nine laboratories participated in the comparison exercise. These exercises are important for increasing confidence in the ability of the results of measurements carried out by the Brazilian network of radiation monitors calibration laboratories. 1.1. Participating Laboratories. Laboratório de Calibração de Monitores de Radiação LCMR/LNMRI/IRD Instituto de Pesquisas Energéticas e Nucleares - IPEN Centro de Desenvolvimento de Tecnologia Nuclear- CDTN Centro Tecnológico do Exercito - CTEx Departamento de Energia Nuclear da UFPE - DEN/UFPe Laboratório de Ciências Radiológicas da UERJ - LCR/UERJ Eletronuclear Eletrobrás Termonuclear S.A. MRA Comércio de Instrumentos Eletrônicos Ltda. Centro Regional de Ciências Nucleares do Nordeste CRCN The Calibration Laboratory of Radiation Monitors LNMRI/IRD determined the calibration factor reference. It been the mean of the calibrations performed during the exercise. 2. Objective The purpose of the comparison exercise was: a) Calculate the coefficient of calibration in air kerma and compares them; b) Determine the performance of the calibration of the qualities S-Cs and S-Co in terms of air kerma of participating laboratories; c) Identify problems and propose corrective action. 2.1. Instrument Submitted For Comparison - The Transfer Chamber The ionization chamber PTW LS01 was used as the transfer chamber is spherical graphite walls. Just as the chamber was sent, the laboratories were instructed to make their own measurements as their usual procedures. The main features of the transfer chamber are: Manufacturer: PTW Model: TN32002 Type: Ionization chamber Series Number: 490 Sensitive volume: 1,000 cm 3 Outside diameter: 140 mm Electrode diameter: 50 mm Wall thickness: 3 mm Potential of high voltage electrode: - 400 V 2

3. Determination of the calibration factor Comparison for air kerma was performed indirectly by comparing the calibration coefficients in air kerma the transfer chamber, N k. The calibration coefficient is determined using the expression: K a N k = [ Gy/C ] M e. (T, P). f. 3600 t Where: N k is the factor of chamber calibration in terms of air kerma (Gy /C). is the rate of air kerma measured with a standard chamber is the average reading of the chamber you want to calibrate divided by time; ɸ(T,P) is the correction factor for temperature and pressure; is the calibration factor of the electrometer. The LNMRI used the standard chamber to determinate the reference value of the calibration coefficient, N k,lnmri against which the results of the measurements of the calibration factor by participants, N k,part, were compared. The N k,lnmri in air kerma was determined with a standard chamber LS01 model TM32002 serial number 104 calibrated in June 2013 at the Physikalisch- Technische Bundesanstalt (PTB). The calibration coefficient for the transfer chamber was provided in terms of air kerma, in units of Gy /C in referenced standard conditions of temperature, pressure and relative humidity at T = 293.15 K, P = 101.325 kpa h = 65%. Participants did not apply any correction for ion recombination and the relative humidity should be between 30% and 75% during calibration, followed the normal routine of the laboratory with respect to the operator and the equipment used. 4. The standard ionization chambers used Participated in the comparison exercise seven ionization chambers, as shown in Table 1 All chambers are reference standards calibrated to the energies of 60 Co and 137 Cs. Table 1. The standard ionization chambers used by the participating laboratories and by LNMRI. (1) Laboratories Standard ionization chambers LNMRI PTW TN32002#104 LCR/UERJ PTW N23361#2923 LCR/UERJ PTW TN32002#489 ETN PTW N32002#86 MRA PTW 32002#309 IPEN PTW W32002#25 DEN/UFPE NE 2575#526 DEN/UFPE PTW TW32002#550 CRCN NE 2575#518 CDTN PTW TN32002#382 CTEx PTW TW32002#527 3

5. Transfer chamber calibration The transfer chamber circled among the LNMRI and participants through the planning done by LNMRI together with laboratories. The LNMRI calibrated the transfer chamber to verify the performance, before sending it to the next participating laboratory. Control measures were obtained by LNMRI thus proving the proper functioning of the transfer chamber. The following terms of reference were recommended: 1 Distance source - Chamber: 200 cm. 2 Operating Voltage: - 400 V. No deviations from these conditions were reported. Leakage currents of measurement systems, composed of electrometer and chamber were measured in accordance with the procedures of each laboratory. The percentage deviation for the calibration coefficient in terms of air kerma, and LNMRI between the participating laboratories should be equal to or less than 2.0%. 6. Results 6.1. Physical quantity to be measured The quantity chosen to be measured for comparison was the air kerma (K ar ), since all the chambers were calibrated in this quantity. The value of air kerma, K AR, was determined for the irradiation conditions with the transfer chamber in each participating laboratory. The Calibration coefficients of the transfer chamber, N k obtained by participating laboratories and by the LNMRI are in Table 2 along with their respective uncertainties expanded standard. Table 2. The calibration coefficient and uncertainties determined by the participating laboratories. Laboratories Calibration Factor N k ± U (%) Calibration Factor N k ± U (%) Quantity/Unity Cs-137 Co-60 LNMRI 2,486E+04 ± 2,7% 2,436E+04 ± 2,6% K ar /Gy.C -1 LCR/UERJ 2,488E+04 ± 4,7% K ar / Gy.C -1 LCR/UERJ 2,460E+04 ± 4,7% K ar / Gy.C -1 ETN 2,516E+04 ± 5,1% K ar / Gy.C -1 MRA 2,504E+04 ± 2,7% 2,458E+04 ± 2,7% K ar / Gy.C -1 IPEN 2,523E+04 ± 3,7% 2,426E+04 ± 4,2% K ar / Gy.C -1 DEN/UFPE 2,468E+04 ± 3,4% 2,409E+04 ± 4,6% K ar / Gy.C -1 DEN/UFPE 2,481E+04 ± 3,4% 2,438E+04 ± 4,6% K ar / Gy.C -1 CRCN 2,470E+04 ± 2,0% K ar / Gy.C -1 CDTN 2,470E+04 ± 5,5% K ar / Gy.C -1 CTEx 2,463E+04 ± 3,0% 2,415E+04 ± 2,8% K ar / Gy.C -1 OBS: LNMRI reference 4

N k (Gy/C) Nk (Gy/C) 9th Brazilian Congress on Metrology (Metrologia 2017) In Figure 1, we can see the values determined for the calibration coefficient of the LNMRI and by the participating laboratories. Error bars are the values of the standard expanded uncertainties submitted by the laboratories. 27000 CALIBRATION FACTORS Cs137 - Nk (Gy/C) 26000 Nk Cs-137 25000 24000 23000 LABS Figure 1. Calibration Factors determined for the transfer chamber by LNMRI and participating laboratories for quality S-Cs CALIBRATION FACTORS Co60 - Nk(Gy/C) 26000 25500 25000 24500 24000 23500 23000 22500 LNMRI/IRD MRA IPEN DEN/UFPE - NE 2575 LABS DEN/UFPE - PTW TW32002 CTEx Nk Co-60 Figure 2. Calibration Factors determined for the transfer chamber by LNMRI and participating laboratories for quality S-Co. 5

6.2. Normalized response - N k,part /N k, LNMRI. All values were normalized to the value of the calibration coefficient in air kerma obtained by LNMRI; the values can be seen in Table 3 and Figure 3 and 4. Table 3. Resposta normalizada - N k,part /N k,lnmri. Laboratories Normalized response R=N k,part /N k,lnmri Cs-137 Co-60 LNMRI 1,0 1,0 LCR/UERJ - PTW TN32002 0,990 LCR/UERJ PTW N23361 1,001 ETN 1,012 MRA 1,007 1,009 IPEN 1,015 0,996 DEN/UFPE PTW TW32002 0,998 1,001 DEN/UFPE NE 2575 0,993 0,989 CRCN 0,994 CDTN 0,994 CTEx 0,991 0,992 The standard deviation between the calibration coefficients is 0.8% for S-Cs quality and 0,7% for S- Co quality. The results are considered to be acceptable for the purpose of this comparison if the expanded uncertainty of the ratio N k,part /N k,lnmri covers the reference value and 0.98 R 1.02. N k,part /N k,lnmri Values of the ratio of the calibration factors N k,part / N k,lnmri for S-Cs 1,08 1,06 1,04 1,02 1 0,98 0,96 0,94 0,92 0,9 LABS Figure 3. Values of the ratio of the calibration factors N k,part / N k,lnmri for S-Cs. 6

Values of the ratio of the calibration factors N k,part / N k,lnmri for S-Co 1,060 1,040 N k,part /N k,lnmri 1,020 1,000 0,980 0,960 0,940 0,920 LNMRI/IRD MRA IPEN DEN/UFPE - NE 2575 LABS DEN/UFPE- PTW TW32002 CTEx Figure 4. Values of the ratio of the calibration factors N k,part / N k,lnmri for S-Co. Figure 5. Results by Radiation Energy 6.3. Uncertainties Uncertainties measures were calculated according to the ISO/IEC Guide 98-3 - "Guide to the expression of uncertainty in measurement". The total uncertainty was obtained by a combination of type A and B uncertainties regarding measurement multiplied by the coverage factor k = 2, which corresponds to a confidence level of 95.45%. The combined uncertainty was calculated by: 7

u = ( + u Where u A and u B are type A and type B standard uncertainties. ) 2 2 1/ 2 c u A B (2) The components of uncertainty that contributed to the combined standard uncertainty of the calibration of the transfer chamber are raised in dosimetry with standard chamber as positioning, irradiation distance, uncertainty of the calibration standard (certified standard chamber), stability, uniformity field, time factors thermometer, barometer and electrometer, energy dependence, leakage current measurement, repeatability, resolutions depending on the equipment and set up some more, and should take most of the components into account again when calibrating the chamber transfer. 7. Discussion and analysis of results The results were evaluated by the percentage difference D% using the recommended methodology in ISO/IEC 17043-1, as most reference chambers, if not all, used by the laboratories were calibrated by LNMRI. The percentage difference is calculated by the equation: D % N k, participan te N N k, LNMRI k, LNMRI.100 Where: N k,lnmri = Calibration factor N k obtained by LNMRI N k,participant = Calibration factor N k obtained by the participating laboratory The percentage difference for the calibration coefficient in terms of air kerma, and LNMRI between the participating laboratories should be equal to or less than 2.0%. None of the results exceeded the tolerance level. Table 4. Results of the comparison exercise - Percentage Difference between the reported coefficients of participating laboratories with N k,lnmri. (3) Labs D(%) S-Cs D(%) S-Co LCR/UERJ -TN32002 0,1 LCR/UERJ TN23361-1,0 ETN 1,2 MRA 0,7 0,9 IPEN 1,5-0,4 DEN/UFPE TW32002-0,2 0,1 DEN/UFPE NE 2575-0,7-1,1 CRCN -0,6 CDTN -0,6 CTEx -0,9-0,8 The results were within the acceptance limit of ±2.0%, established by LNMRI in the protocol. The acceptance limit was calculated taking into account the existence of the dependence between the 8

calibrations of the laboratory standards performed by LNMRI. The highest percentage difference was 1.5%, being below the limit of acceptance established by the LNMRI; the index of this exercise was smaller than the previous comparison, even with the inclusion of a laboratory that started to work recently, in the second half of 2016. Considering the points discussed, it is concluded that the Comparison Exercise presented an excellent result and this shows the competence of the Brazilian laboratories in performing the calibration services. Only one laboratory was informed that it needs to review its calculation of the uncertainties and their components. The Comparison Exercise tested the actual measurement capacity of the laboratories performing calibration services of radiation and irradiation monitors in Brazil, since the measurements are performed in the premises and with the laboratory procedures, which allows a greater reliability in the measurements and services Performed by the Brazilian calibration network. This work counted on the partial financial support of FINEP/SIBRATEC/METRORADI References [1] NORMA ISO/IEC 17043:2010 - Conformity assessment -- General requirements for proficiency testing. [2] VOCABULÁRIO INTERNACIONAL DE METROLOGIA/INMETRO: Vocabulários Internacional de termos Fundamentais e Gerais de Metrologia - 2012. [3] NORMA ISO-4037-2: 1997 - X and Gamma Reference Radiation for Calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 2: Dosimetry for Radiation Protection over the energy ranges 8 kev t0 1.3 MeV and 4 MeV to 9 MeV. [4] NORMA ISO-4037-3:1999 - X and Gamma Reference Radiation for Calibrating dosemeters and doserate meters and for determining their response as a function of photon energy Part 3: Calibration of area and personal dosemeters and the measurement of their response as a function of energy and angle of incidence. [5] Radioactive half-lives - Table of recommend values: http://www.nucleide.org/nucdata.htm 9