Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards

Similar documents
CHOKING PRESSURE RATIO GUIDELINES FOR SMALL CRITICAL FLOW VENTURIS AND THE EFFECTS OF DIFFUSER GEOMETRY

WGFF Guidelines for CMC Uncertainty and Calibration Report Uncertainty

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

Evaluation of the Effect of Relative Humidity of Air on the Coefficients of Critical Flow Venturi Nozzles

405 Compact Orifice Series and 1595 Conditioning Orifice Plate Flow Test Data Book and Flow Handbook

POWER UNDERSTANDING MEASUREMENT UNCERTAINTY IN DP FLOW DEVICES

TYPICAL PRESSURE MEASUREMENT UNCERTAINTY DEFINED BY AN FPG8601 FORCE BALANCED PISTON GAUGE

Vocabulary of Metrology

A MEASUREMENT ASSURANCE PROGRAM FOR LOW FLOW CALIBRATION BY THE TRANSFER METHOD CS20

Morehouse. Edward Lane, Morehouse Instrument Company 1742 Sixth Ave York, PA PH: web: sales:

NORTH SEA FLOW MEASUREMENT WORKSHOP 2004 In. St Andrews, Scotland

Compressible Gas Flow

REPORT OF SPECIAL TEST

Understanding the uncertainties associated with using the 5128A RHapid Cal portable humidity generator

Calibration Traceability Guidelines

Understanding uncertainties associated with the 5128A RHapid-Cal Humidity Generator

EA Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement

ABSTRACT I. INTRODUCTION

Pascal ET is an handheld multifunction calibrator for the measurement and simulation of the following parameters: - pressure

The Expansibility Factor Equations in ISO and ISO : Do They Deliver What They Promise?

ENERGY DISTRIBUTION ANALYSIS IN A LOW HEAD FRANCIS TURBINE DURING THERMODYNAMIC EFFICIENCY MEASUREMENTS

Standard Practices for Air Speed Calibration Testing

Typical pressure measurement uncertainty defined by an FPG8601 Force Balanced Piston Gauge

POWER QUALITY MEASUREMENT PROCEDURE. Version 4 October Power-Quality-Oct-2009-Version-4.doc Page 1 / 12

EXPERIMENT NO. 4 CALIBRATION OF AN ORIFICE PLATE FLOWMETER MECHANICAL ENGINEERING DEPARTMENT KING SAUD UNIVERSITY RIYADH

Mechanical Measurements and Metrology Prof. S. P. Venkateshan Department of Mechanical Engineering Indian Institute of Technology, Madras

Document No: TR 12 Issue No: 1

SAMPLE SHEET. Supply of PF Process Calculation Spreadsheet (Excel Forma)

CALIBRATION CERTIFICATE # 503

Introduction to the evaluation of uncertainty

Provläsningsexemplar / Preview. First edition

AN EXPLANATION OF THE DG-1000 ACCURACY SPECIFICATIONS

[Prasanna m a*et al., 5(6): July, 2016] ISSN: IC Value: 3.00 Impact Factor: 4.116

Orifice and Venturi Pipe Flow Meters

Upgrade of 5m-Bench System for Traceable Measurements of Tapes and Rules at SASO-NMCC Dimensional Laboratory

TOLERANCES AND UNCERTAINTIES IN PERFORMANCE DATA OF REFRIGERANT COMPRESSORS JANUARY 2017

R SANAS Page 1 of 7

Provläsningsexemplar / Preview INTERNATIONAL STANDARD ISO Second edition

Annubar Primary Element Flow Calculations

Review of Anemometer Calibration Standards

Draft PS 18 APPENDIX A Dynamic Spiking Procedure ( ) 1.1 This appendix to Performance Specification 18

Process Gas Flow Measurement and Performance Evaluation API TECHNICAL REPORT 2571 SECOND EDITION DRAFT, FEBRUARY 2018

Flow Management Devices, LLC An ISO 9000:2008 Certified Company

ALASTAIR MCLACHLAN Applications Cameron

Anemometer Calibration Requirements for Wind Energy Applications

Approval Block. Prepared by: Signature Date Evan Parnell 08 NOV Reviewed by: Signature Date. Approved by: Signature Date

Final Results of Bilateral Comparison between NIST and PTB for Flows of High Pressure Natural Gas

Civil aeroengines for subsonic cruise have convergent nozzles (page 83):

Orifice and Venturi Pipe Flow Meters

STATUS OF THE WIGOS DEMONSTRATION PROJECTS

Accurate Measurement of Hydrocarbon Dewpoint in Natural Gas Streams

Pneumatic fluid power -- Determination of flow-rate characteristics of components using compressible fluids -- Part 2: Alternative test methods

) where the inlet pressure ( P 1

The Fundamentals of Moisture Calibration

DM70 Hand-Held Dewpoint Meter for Spot-Checking Applications

MICHELL TECHNOLOGIES FOR MOISTURE. Calibration. General Principles & Theory, Equipment Considerations

METHODS FOR CERTIFYING MEASUREMENT EQUIPMENT. Scott Crone

Experiment (4): Flow measurement

Humidity Calibration Solutions

CHAPTER 5 MASS AND ENERGY ANALYSIS OF CONTROL VOLUMES

Part 5: Total stations

ABSTRACT 1. INTRODUCTION 1 2. THEORETICAL BACKGROUND 3 3. POSITIVE DISPLACEMENT PROVERS MEASUREMENT PRINCIPLE CALIBRATION PROCEDURE 6

The Uncertainty Principle

A61-02 CALA Guidance on Traceability Revision 1.2 October 15, 2012

Uncertainty Propagation for Force Calibration Systems

One-Dimensional Isentropic Flow

Certificate of Accreditation to ISO/IEC 17025:2005

A Method for Verifying Traceability in Effective Area for High Pressure Oil Piston-Cylinders

Multi Analyte Custom Grade Solution. Aluminum, Potassium, Magnesium, ANALYTE CERTIFIED VALUE ANALYTE CERTIFIED VALUE

High-Pressure Volumetric Analyzer

CALIBRATION REPORT FOR THERMOHYDROMETER

Diagnostic System for Venturi Meters

MCE380: Measurements and Instrumentation Lab

KNOWLEDGE

Alden Test Results On AMITY Insert Venturi

The Determination of Uncertainties in Bend Tests on Metallic Materials

Boiling Heat Transfer and Pressure Drop of R1234ze(E) inside a Small-Diameter 2.5 mm Microfin Tube

Statistical Analysis of Engineering Data The Bare Bones Edition. Precision, Bias, Accuracy, Measures of Precision, Propagation of Error

Calibration of Temperature Block Calibrators

Instruction Manual. Equipment for Engineering Education

IBV Series C Carbon Steel Inverted Bucket Vertical Steam Trap

IBV Series Z Alloy Steel Inverted Bucket Vertical Steam Trap

Experiment No.4: Flow through Venturi meter. Background and Theory

COPLEY S C I E N T I F I C. A multi-function aerosol system with aerosol generation, classification and monitoring capabilities for:

Techniques for Fast Measurements of Low Dewpoint in Portable Hygrometers

Measurement & Uncertainty - Basic concept and its application

IMPLEMENTATION OF ETHANOL HEAT PIPE AT CETIAT

2.1. Accuracy, n- how close the indication of the thermometer is to the true value.

Comparison of discharge measurements -

Calibration Models and Uncertainty Analysis of the Euroloop Liquid Hydrocarbon Flow Facility

Humidity Calibration Solutions

Euramet project 1075 Bilateral comparison

Multi Analyte Custom Grade Solution. Calcium, Iron, Potassium,

CALIBRATION. Calibration. General Principles & Theory, Equipment Considerations. Copyright Caltech 2014

Isentropic Efficiency in Engineering Thermodynamics

792A AC/DC Transfer Standard

57:020 Mechanics of Fluids and Transfer Processes OVERVIEW OF UNCERTAINTY ANALYSIS

TransCanada / Daniel Using Advances Diagnostics to Identify Measurement Uncertainties

Theory and Design for Mechanical Measurements

Transcription:

MAPAN Performance - Journal of of Metrology a Gas Flow Society Meter of Calibration India, Vol. 6, System No. 3, Utilizing 011; pp. Critical 47-54Flow Venturi Standards ORIGINAL ARTICLE Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards MICHAEL CARTER*, WILLIAM JOHANSEN and CHARLES BRITTON Colorado Engineering Experiment Station Inc. [CEESI] 54043 WCR 37 Nunn, CO 80648 USA * e-mail:: mcarter@ceesi.com [Received: 0.1.010 ; Revised: 1.03.011 ; Accepted: 30.06.011] Abstract A set of Critical Flow Venturis (CFVs) were manufactured by Flow Systems and calibrated by NIST in Gaithersburg, MD. Subsequently these flow standards along with pressure and temperature standards have been integrated into a gas flow meter calibration (GFC) system. Performance data and an uncertainty analysis is discussed here in this paper. Calibration data on these CFV standards will be used to intercompare the NIST and CEESI primary facilities. 1. Introduction Flow Systems manufactures between 300 and 400 CFVs on an annual basis. These Venturis are subsequently calibrated by CEESI. In efforts to eliminate logistical issues, improve lead times and reduce uncertainty, Flow Systems and CEESI have undertaken a joint effort to design, manufacture and commission an automated gas flow meter calibration system (GFC). The desired uncertainty for a calibration made on a meter-under-test (MUT) was to be less than ± 0.0%. This requires low uncertainty for the discharge coefficient (C d ) for the CFV, as well as the pressure and temperature instrumentation installed in the GFC.. System Design Eight CFVs with throat diameters ranging from 0.41 mm to 4.5 mm were designed and manufactured in accordance with ISO 9300 and ASME MFC-7M. Metrology Society of India, All rights reserved 011. These CFVs were designed as a binary set such that the throat areas progress by a factor of. This set of CFVs is also referred to as the "P Nozzles". These Venturis were calibrated at the NIST FMG facilities located in Gaithersburg, MD using the 677 L PVTt system [1]. The uncertainty in the discharge coefficient (C d ) versus throat Reynolds Number (Re) equation for each CFV is less than ± 0.07% at 95% confidence. Test fluids used by NIST were nitrogen and compressed air. The flowing temperature was near ambient, at approximately 1 C. Each CFV (P Nozzle) was calibrated over an inlet pressure range of 17 to 87 kpa. Using the same fluids and at the same temperature, the GFC system will service 0.00005 to 0.03 kg/s. The main test fluid for the GFC system will be compressed air that is filtered and dried to a pressure dewpoint less than -40 C. A sensor measuring the dewpoint is used as an alarm in the GFC system. The supply pressure to the system will range from 690 to 900 kpa. This supply will be diffused into several copper tubes which, along with the CFV and MUT Meter 47

Michael Carter, William Johansen and Charles Britton GAS SUPPLY Fig. 1. Layout diagram of CEESI critical flow venturi standard Runs are submerged in a water bath designed to be held at the temperature of the room (Fig. 1). Environmental conditions in the room are monitored by a separate system. This system provides trending and historical information on ambient conditions and air supply dewpoint. A computer controlled, dome loaded pressure regulator and shut off valve are located upstream 1). The regulator will set and maintain pressure at the inlet of the first CFV in series. There two locations in series where standard CFVs (P Nozzles) and the MUT may be installed (Fig. 1). The GFC system uses one CFV (P Nozzle) as the measurement standard during MUT calibrations. At each location the GFC system provides measurement of both inlet static pressure and temperature as well as exit pressure (Fig. 1). Downstream of the pressure regulator and the first CFV/MUT location are temperature stabilizing systems similar to the one previously described. Temperature at each CFV (P Nozzle)/MUT location is measured using a Hart Thermister probe and digital readout. The probes are sheathed in stainless steel and are approximately 3 mm in diameter. The uncertainty of this combination is ± 0.007 C. The GFC system uses Mensor model 610 digital output pressure transducers. The uncertainty of each is ± 0.01% of full scale at 95% confidence. There is an absolute sensor ranged to 103 kpa for measuring the barometric pressure. There is a gauge transducer at each of the two CFV/MUT inlet pressure locations (Fig. 1). There are two WIKA analog pressure transducers to measure the exit pressure at the exit of each CFV/MUT. The last CFV/MUT in series either exhausts to ambient pressure or into a vacuum receiver. In the latter case another WIKA sensor is used to measure the exit pressure of the CFV/MUT. The GFC system uses the same universal gas constant, dry air composition and NIST REFPROP 8.1 [] thermodynamic database that is used by the NIST FMG in Gaithersburg, MD. 3. Diagnostics The GFC system employs several automated diagnostics. Before the system is pressurized each of the two temperature probes are read to ensure that they agree within their estimated uncertainty of ± 0.007 C. During this check the system reads, records and adjusts the zero offset of each gauge pressure transducer. These data will be historically tracked for potential use in a measurement assurance program (MAP) for the GFC system. Since configurations of CFVs (P Nozzles)/MUTs are manually assembled and disassembled a minimum of two times during an MUT calibration, the GFC system will automatically perform pressuredecay type leak tests prior to initiating flow through the system. The operator defines the leak test volumes by selecting the CFV (P Nozzle) serial numbers, MUT, MUT test section and all required adapters. These items are cataloged in the system software along with an estimated volume determined from CAD drawings 48

Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards of each. After a cap is installed by the operator at the end of the most downstream test section, the system opens the shut off valve and the regulator slowly pressurizes the entire volume. The leak test pressure set point is 690 kpa. Once the system determines that the total leak test volume has reached equilibrium with respect to temperature, a counter is initiated and all pressures and temperatures within the system are recorded. Provided that the pressure decay is less than a predetermined limit, the test will last 60 seconds. At the end of this interval, the final pressures and temperatures are recorded along with the duration of the test. If the leak test passes, the GFC system will compare the final readings of the gauge transducers used and verify that they agree within their full scale uncertainty specifications 4. Validation Efforts The GFC system is programmed using National Instruments LabVIEW software. Data processed using this software was input into spreadsheets that perform identical calculations. GFC system calculations have been compared to spreadsheet values and were found to be less than 10 ppm. Results are archived and are revalidated with each software revision. 5. CEESI Evaluation of P Nozzles with Primary B Prior to commissioning the GFC system, the CFVs (P Nozzles) were compared to CEESI's Primary "B" system which is a gravimetric system that utilizes dry compressed air at room temperature as the test fluid [3]. Being a mass / time system it has a very short traceability chain to NIST fundamental standards of mass and time. The Primary B system has the good characteristics of low uncertainty and high precision. All of the CFVs have been evaluated by CEESI, but for space limitations, the result for P088 (which is typical of all CFVs) is shown in Fig.. 6. Evaluation of the GFC with P Nozzles The initial validation efforts at FSI were focused on mass flow correlation and reproducibility when testing two CFVs (P Nozzles) in series. These efforts yielded insight to the optimum time duration of flow from onset to data capture. Since there are two locations for installing CFVs, it is possible to perform tests with two P Nozzles in series and compare the mass flow that is calculated for each one. This method allows all the instrumentation and the discharge coefficient of both CFVs to be evaluated. Figure 3 shows the results for the P088 nozzle. The mass flow ratio is the mass flow of the P088 nozzle divided by Percent Deviation from NIST Equation 0.40 0.30 0.0 0.10 0.00-0.10-0.0-0.30-0.40 0 50000 100000 150000 00000 50000 Throat Reynolds Number, Re Primary B -- June 010 Fig.. CEESI calibration data for critical flow venturi P088 49

Michael Carter, William Johansen and Charles Britton 1.00 1.001 Mass Flow Ratio 1.000 Position 1 - P15 - July 1 Position 1 - P15 - TOPB Repro - July 1 0.999 Position - P063 - July 1 Position - P063 - TOPB Repro - July 1 Position 1 - P15 - DTD Repro - July 0.998 Position 1 - P15 - DTD & TOPB - July Position - P063 - DTD Repro - July Position - P063 - DTD & TOPB Repro - July 0.997 0 100 00 300 400 500 600 700 800 P088 --- Inlet Pressure, kpa Fig. 3. FSI critical flow venturi-p088 the mass flow of another P Nozzle. If everything were perfect, the mass flow ratio would be unity. As can be seen, most of the data falls within a range of 1.000 to 1.001. This indicates that the pressure and temperature instrumentation at each CFV location is performing in an acceptable manner and the discharge coefficient equations for each CFV is within the stated uncertainty of ± 0.07%. Due to space limitations, only the P088 data is shown. All other P Nozzles performed in a similar manner. 7. Uncertainty Analysis The uncertainty analysis for the results of these calibrations is a 3 step process. Step 1 is to assess the uncertainty of the mass flow delivered to the MUT, i.e., m MUT. This uncertainty assessment is based on the uncertainties for the measurement components that produce the mass flow, and these produce the expanded uncertainty for the mass flow through the MUT at a quoted level of confidence. Step is to assess the uncertainties for the MUT result. Step 3 is to fit the data and then assess the uncertainties associated with using the fitted results. When the MUT is a CFV, as it is in the result described here, it is handled in the conventional fashion, as described below. The uncertainty of the MUT result will therefore be comprised of both the appropriately replicated (i.e., the DTD (day-to-day) or TOPB (take-out-put-back) reproducibility associated with the different arrangements described in the System Design section) determinations for the MUT result and the combined uncertainty for the mass flow m MUT,delivered to the MUT, as produced using the GFC. The VIM [4] and the GUM [5] describe the uncertainty details for assessing and combining these component uncertainties. Using Conservation of Mass principles applied to the Connecting Volume (CV) which is the interior pipe volume between the standard (STD) and the MUT, the mass flow delivered to the MUT is, see [6]: VCV 1 P P m = MUT mstd 1 ± mstd R T CVf T and m STD = m STD (Δt) = m STD t calibration CVi (1) where, in compatible units, the subscripts denote respective quantities, R is the gas constant, and P T CVf and P T CVi () denote the spatial averages, 50

Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards respectfully, of the pressure to temperature ratios in the CV at the end and the beginning of the calibration 1 p p interval. In this way the term R T CVf T is CVi the fluid density increase in the CV during the calibration. The plus or minus term in the governing equation for m MUT depends on whether the STD is upstream or downstream from the MUT; the sign is minus when the STD is upstream of the MUT and it is plus when the MUT is upstream of the STD, see [6]. The relative uncertainty for m MUT is therefore written, see [6]: Δρ V u ( m ) = [ u ( m )] + [ u (1 )] = CV CV r MUT r STD r mstd ( pcv ) ( ) V p u + [ ur( m STD)] + [ )] mstdrtcv ur TCV CV CV r Data records are to be used to quantify the variances under the radical sign in the above equation, (3) including correlation effects, see [6]. When the STD is a P Nozzle, m STD is given by: gc Mw m = P A C C R T * * STD ostd STD dstd STD U ostd where C dstd is provided via the NIST calibration report on its calibration of the STD. The relative uncertainty for m STD is written, where it is assumed that no correlation effects are present among the component measurements that comprise m STD see [6]: u ( m ) = [ ur( PoSTD) ] + [ 0.5 ur( gc) ] + [ 0.5 ur( Mw) ] + [ 0.5 ur( RU) ] [ 0.5 u ] r( ToSTD) ur( ASTD) * + ( ) + ( ) r dsts r STD (4) + + u C u C (5) r STD * Table 1 Uncertainty Results for m MUT 1 3 4 5 6 7 8 9 10 11 1 13 Quantity/Symbol Source of Uncertainty Units Value/Range u(spec) (%) @ 68% CL Type Uncrt Prob Distrb Divisor Rel Sens Coeff u(symbl) (%) Degs Frdm % of Cmbn'd Note CFV Stag Press/P osn Mfgr Spec on Transducr F/L 17-87 kpa 0.03 B N 1 1 0.030 1.00E+06 1.8 1 Grav Const Fctr/gc Hdbk LM/Ft 9.80665 0 NA NA 1.73 0.5 0.0000 1.00E+06 0.0 Molecular Wt for Air/Mw Hdbk M 9.9 0.005 B R 1.73 0.5 0.0007 1.00E+06 0.0 3 Univers Gas Constant/R U Hdbk nm/kgk 8.1447 0.00009 B R 1.73 0.5 0.0000 1.00E+06 0.0 4 Nozzle Stag Temp/T osn Mfgr Spec on Sensor F or R 93K 0.001 B N 1 0.5 0.0006 1.00E+06 0.0 5 Nozzle Mass Flow Rate a. Throat Area NIST Cal Rept L 0.41-4.5 mm 0 NA NA 1 1 0.0000 1.00E+06 0.0 6 b. Dischrg Coefficient NIST Cal Rept Dimles 0.035 B R 1 1 0.0350 5 9.7 7 c. Crit Flo Funct (C*) NIST Cal Rept/RefProp Dimles 0.685 0.03 B N 1 1 0.031 1.00E+06 1.9 8 Connecting Vol Term/CV Estimates Dimles 1 0.049 B N 1 1 0.049 1.00E+06 44.6 9 Rel Uc(m-dot-MUT)/% Rel Cmbn'd Stnd. Uncert. (%) Norm (k=1) RSS Rows 1-9 0.064 57 100.0 10 Rel Ue(m-dot-MUT)/% Rel Expnd'd Uncertainty (%) Norm (95%: k=) xrss in Row 10 0.18 57 11 Uncertainty Characteristics Processing Uncertainties Uncertainty Results Notes: 1 17 kpa pressure measured with +/ 0.01% FS @ 95% CL 103 Kpa Baro & +/ 0.01% FS @ 95% CL 690 Kpa gauge Results are for fundamental constant with no uncertainty assigned. 3 Results come from NIST calibration report. 4 Results come from NIST calibration report. 5 Column 5 entry uses +/ 0.007C @ 95% CL as mfgr stated accuracy at ambient (0.001% @ 68% CL) 6 Uncertainty is zero as same values are used as in calibration report. 7 Results come from NIST calibration report. 8 Results come from NIST generated number. 9 Entries are based on "worst case" situation (smallest nozzle @ 17 kpa inlet to CFV. 10 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 11 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 51

Michael Carter, William Johansen and Charles Britton Typical values for these relative uncertainties are used in the tabulation for u ( m ) presented in Table 1, r see [6]. The resulting relative Expanded Uncertainty for is ± 0.18 % at 95 % confidence level. When a calibration of a P Nozzle is done using other P Nozzles, the calibration result for each nozzle is written: STD m R T A C = P P C g M * MUT U opnzl Pnzl dpnzl * opnzl Pnzl c w The relative uncertainty for the calibration result * ( ) u A C iswritten,see[4]: r Pnzl dpnzl u ( A C ) = * u ( ) r mmut urc( CPnzl ) [ u ( P )] + [ 0.5 u ( g )] + + + * r opnzl r c r Pnzl dpnzl [ 0.5 ur( Mw) ] [ 0.5 ur( RU) ] [ 0.5 u ( T )] r opnzl + + Typical values for these relative uncertainties are used in the tabulation for u r (A* Pnzl C dpnzl ) are presented (6) (7) in Table [4]. The resulting Expanded Uncertainty is ± 0.177 % at 95 % Confidence Level. Recently, the P Nozzles were calibrated using other P Nozzles calibrated by NIST and the results can be conventionally plotted, as typically shown in Fig. 4. It should be noted that the "stability" of these metering results is quantified in the legend of Fig. 4 by DTD and TOPB Reproducibilities. While these levels of scatter may be larger than "Repeatability" levels, these levels can be considered to be far more typical of what nozzle users may expect in practice using these nozzles, see [6]. When these calibration results are fitted, using the form, C d = A+B Re -05, and when the uncertainties associated with using these fits to make future flow measurements, the result for P088 is shown in Fig. 5, see also [5]. As noted on these figures, the fit form is given together with the maximum uncertainty at 95% confidence level, see [7]. It is noted that these uncertainty levels are only slightly higher than that for u r (C d P nzl ); this is the result of the extensive data records from the calibrations and from the fact the fitting form is compatible with the nozzle characteristics. Table Uncertainty Results for u r (C d...and...fit) 1 3 4 5 6 7 8 9 10 11 1 13 Quantity/Symbol Source of Uncertainty Units Value/Range u(spec) (%) @ 68% CL Type Uncrt Prob Distrb Divisor Rel Sens Coeff u(symbl) (%) Degs Frdm % of Cmbn'd Note m-dot-mut M/t a. Type A Replications (rel uncrt) Calibration M/t 0.05% 0.05 A N 1 1 0.0500 5.0E+00 3.5 1 b. Type B CFV mdot (rel uncrt) Previous Assessment M/t 0.07% 0.064 A&B N 1 1 0.064 1.1E+0 53.7 Critical Flow Funct NIST Cal Rept/RefProp Dimles 0.685 0.03 B N 1 1 0.031 1.0E+06 6.9 3 Stag Press/P o Mfgr Spec on Transducr F/L 17-87 kpa 0.03 B N 1 1 0.030 1.0E+06 6.9 4 Grav Const Fctr/g c Hdbk LM/Ft 9.80665 0 NA NA 1.73 0.5 0.0000 1.0E+06 0.0 5 Molecular Wt for Air/M w Hdbk M 9.9 0.005 B R 1.73 0.5 0.0007 1.0E+06 0.0 6 Univers Gas Constant/R U Hdbk nm/kgk 8.1447 0.00009 B R 1.73 0.5 0.0000 1.0E+06 0.0 7 Stag Temp/T o Mfgr Spec on Sensor F or R 93K 0.001 B N 1 0.5 0.0006 1.0E+06 0.0 8 Rel Uc(Cd )/% Rel Cmbn'd Stnd. Uncert. (%) N/R (k=1) RSS Rows 1-9 0.088 4 100.0 9 Rel U(Cd )/% Rel Expnd'd Uncertainty (%) N/R (95%: k=) xrss in Row 10 0.175 4 10 Uncert for Using Fitted Eqns Fitting Process Dimles < 0.03%@95%CL 0.015% A N 1 1 0.015 58.0 11 Rel Uc(Cd & Fit)/% Rel Cmbn'd Stnd. Uncert. (%) Dimles 0.089 7 100.0 1 Rel U(Cd & Using Fitted Eqn) Rel Expnd'd Uncertainty (%) Dimles 0.177 7 13 Uncertainty Characteristics Processing Uncertainties Uncertainty Results Notes: 1 Entries in this row are from "Day To Day Reproducibility" calibration conditions. Entry in Column 5 is from GFC mass flow uncertainty assessment (table 1) 3 Results come from NIST calibration report. 4 17 kpa pressure measured with +/ 0.01% FS @ 95% CL 103 Kpa Baro & +/ 0.01% FS @ 95% CL 690 Kpa gauge 5 Results are for fundamental constant with no uncertainty assigned. 6 Results come from NIST generated number. 7 Results come from NIST calibration report. 8 Column 5 entry uses +/ 0.007C @ 95% CL as mfgr stated accuracy at ambient (0.001% @ 68% CL) 9 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 10 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 11 Entry in Column 5 is an upper bound from the fit assessment process, see [7]. 1 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 13 Degree of freedom are computed using the Welch Sattersthwaite Eqn. 5

Performance of a Gas Flow Meter Calibration System Utilizing Critical Flow Venturi Standards 0.97 0.970 Disch arge Coefficient, C d 0.968 0.966 0.964 0.96 0.960 0.958 0.956 0.954 Position 1 - P15 - July 1 Position 1 - P15 - TOPB Repro - July 1 Position - P063 - July 1 Position - P063 - TOPB Repro - July 1 Position 1 - P15 - DTD Repro - July Position 1 - P15 - DTD & TOPB Repro - July Position - P063 - DTD Repro - July Position - P063 - DTD & TOPB Repro - July NIST Equation 0 50000 100000 150000 00000 50000 Throat Reynolds Number, Re Fig. 4. FSI critical flow venturi-p088 0.975 Disc harge Coefficient, C d 0.970 0.965 0.960 0.955 U(95) = 0.177% Expanded Uncertainty at 95% Confidence (Coverage Factor of ) 0.950 0.001 0.00 0.003 0.004 0.005 0.006 0.007 0.008 (Re) -0.5 Cd data Equation Fit Equation Fit +/- U Fig. 5. FSI critical flow venturi-p088 (Eq. Fit: C d = 0.98034-4.3584 (Re) -0.5 8. Conclusion The central conclusion to be drawn from the uncertainty analysis is the fitted forms can be used with a conservative uncertainty specification bounded by ± 0.18 % at 95 % confidence level. Calibration results for the P Nozzles shown in 53

Michael Carter, William Johansen and Charles Britton Table 3 above using both the GFC and CEESI's Primary B facility showed excellent agreement on both C d and reproducibility Validation tests, P Nozzle vs. P Nozzle, on the GFC showed results consistent with those achieved at NIST FMG, Gaithersburg, MD, when comparing similar artifacts. 9. Future Plans Motivation exists to scale up the flow of the GFC system using other standards that are traceable to the NIST 677 l PVTt system. The desired value is 0.5 kg/s while maintaining an MUT uncertainty not greater than ± 0.0 % at 95 % confidence. 9.1 Comparison tests and Youden Analysis to Quantify and Validate Results The GFC System will be controlled by a Measurement Assurance Program (MAP) consisting of periodic calibration of the pressure and temperature standards as well as repeating the validation tests as previously described. Initially we will conduct the validation tests on a 3 month interval and calibrate the pressure and temperature standards on a 6 month cycle. The results of the system diagnostic tests discussed in this paper will be analyzed in support of the MAP effort. These analyses may further refine and enhance the value of such diagnostics. Add pressure control at the most downstream CFV/MUT location to facilitate performing unchoking tests on a MUT(CFV) installed in the intermediate test section. This design also facilitates conducting choking pressure ratio tests when only two nozzles are in series. Future expansion of MUT types to include differential pressure flow meters. References [1] J. Wright and A. Johnson, NIST Lowers Gas Flow Uncertainties to 0.05% or Less, Measure, www.ncsli.org, (010). [] E.W. Lemmon, -M.O. McLinden and M.L. Huber, Refprop 3: Reference Fluid Thermodynamic and Transport Properties, NIST Standard Reference Database 3, Version 8.1, National Institute of Standards and Technology, Boulder, CO, (007). [3] CEESI Document Number: MA-00, Uncertainty Analysis to Support NAVLAP Accreditation of a Calibration Service based on a Gravimetric System Operating Over the 0.1-10 lbm/min Mass Flowrate Range, (009). Available via: www.ceesi.com [4] ISO - IEC Guide 99 International Vocabulary of Metrology - Basic and General Concepts and Associated Terms, Int'l Stands Organ (ISO), Geneva, Switz., (007). Available via: www.bipm.org [5] ISO - IEC Guide to the Expression of Uncertainty in Measurement, Int'l Stands Organ (ISO), Geneva, Switz., (1995). Available via: www.bipm.org [6] G.E. Mattingly, Consultant's Report: Uncertainty Analysis for a FSI Gas Flow Calibration (GFC) System Calibrating a Meter Under Test (MUT) using a Single "P" Nozzle and a Check Standard (CHK), Report Submitted to Flow Systems Inc, Berthoud, CO, (010). [7] D.C. Montgomery and E.A. Peck, Introduction to Linear Regression Analysis, nd edit.. Wiley, New York, (199). 54