U.S. EPA VOLATILE ORGANICS METHOD USING PURGE AND TRAP GC/MS

Similar documents
Validation of USEPA Method Using a Stratum PTC, AQUATek 100 Autosampler, and Perkin-Elmer Clarus 600 GC/MS

Solid Phase Micro Extraction of Flavor Compounds in Beer

Validation of USEPA Method Using a Stratum PTC and the New AQUATek 100 Autosampler

Validation of Environmental Water Methods on One System: Considerations for Sample Volume, Purge Parameters and Quality Control Parameters

Roger Bardsley, Applications Chemist; Teledyne Tekmar P a g e 1

Roger Bardsley, Applications Chemist; Teledyne Tekmar Page 1

Maximizing Sample Throughput In Purge And Trap Analysis

Validation of USEPA Method 8260C Using Teledyne Tekmar Atomx, and Perkin-Elmer Clarus 600 GC/MS

US EPA Method with the Tekmar Lumin P&T Concentrator, AQUATek LVA and Agilent 7890B GC/5977A MS

Method 8260C by Purge and Trap Gas Chromatography Mass Spectrometry using the Clarus SQ 8

US EPA Method 8260 with the Tekmar Atomx XYZ P&T System and Agilent 7890B GC/5977A MS

Exploring US EPA Method 524 Purge and Trap Variables: Water Vapor Reduction and Minimizing Cycle Time

Using Hydrogen as An Alternative Carrier Gas for US EPA 8260

Analysis. Introduction: necessary. presented. Discussion: As part of be carried. consistent and reliable data. (MoRT) to.

Analysis of Volatile Organic Compounds Using USEPA Method 8260 and the 4760 Purge and Trap and the 4100 Autosampler

Ed George and Anaïs Viven Varian, Inc.

Validation of Volatile Organic Compound by USEPA. Method 8260C. Application Note. Abstract. Introduction. Experimental-Instrument Conditions

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

Methanol Extraction of high level soil samples by USEPA Method 8260C

System and JUREK ANNE. Introduction: in an in purge and. trap sampling. Discussion: As part of. consistent and reliable data. (MoRT) to.

Analysis of Volatile Organic Compounds in Soil Samples by EPA Method 8260 with The Stratum PTC and SOLATek 72 Multi-Matrix Autosampler

A Comparison of Volatile Organic Compound Response When Using Nitrogen as a Purge Gas

Optimizing of Volatile Organic Compound Determination by Static Headspace Sampling

Analysis of Volatile Organic Compounds in Water and Soil by EPA Method 8260 with the Atomx Concentrator/Multimatrix Autosampler

Optimal VOC Method Parameters for the StratUm PTC Purge & Trap Concentrator

AUTONOMOUS, REAL TIME DETECTION OF 58 VOCS IN THE PANAMA CANAL

Application News AD Quantitative Determination of Volatile Organic Compounds in Drinking Water by EPA Method with Headspace Trap GC-MS

Performance of a Next Generation Vial Autosampler for the Analysis of VOCs in Water Matrices

Measuring Environmental Volatile Organic Compounds by U.S. EPA Method 8260B with Headspace Trap GC/MS

Optimizing. Abstract: is standardd. procedures. altered to

2017 EPA Method Update Rule and EPA Method 624.1

INNOVATIVE PRODUCTS, SUPERIOR SUPPORT. Presenter: Anne Jurek, Senior Applications Chemist, EST Analytical

Optimal Conditions for USEPA Method 8260B Analysis using the EST Analytical Sampling system and the Shimadzu GCMS-QP2010s

A Single Calibration Method for Water AND Soil Samples Performing EPA Method 8260

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: CTRA Project: 6454 T Lab Order: A 6454 Aqueous 3/1/2013.

Solid Phase Micro Extraction of Flavor Compounds in Beer

ANALYTICAL REPORT. Job Number: Job Description: Transform Complete

Determination of Volatile Organic Compounds in Air

EPA TO-17 Volatile Organic Compounds

BIO-CHEM Laboratories, Inc. Work Order Sample Summary. CLIENT: Cascade Thornapple River Assoc. Project: Water Analysis Lab Order:

Maximizing Production While Minimizing Costs

Solid Phase Microextraction of Cyanogen Chloride and Other Volatile Organic Compounds in Drinking Water with Fast Analysis by GC-TOFMS

Helium conservation in volatile organic compound analysis using U.S. EPA Method 8260C

Analysis of Low Level Volatile Organic Compounds in Air Anne Jurek

Optimizing Standard Preparation

Detection of Volatile Organic Compounds in polluted air by an Agilent mini Thermal Desorber and an Agilent 5975T LTM GC/MS

Target Compound Results Summary

U.S. EPA Method 8270 for multicomponent analyte determination

CALA Directory of Laboratories

Analytical Trap Comparison for USEPA Method 8260C

The following report includes the data for the above referenced project for sample(s) received on 5/15/2017 at Air Toxics Ltd.

National Pollutant Discharge Elimination System (NPDES); BADCT Limits for Volatile Organic Compounds

STANDARD OPERATING PROCEDURES

Optimization of 1,4-Dioxane and Ethanol Detection Using USEPA Method 8260 Application Note

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

Application Note. Application Note 081 Innovative Cryogen-Free Ambient Air Monitoring in Compliance with US EPA Method TO-15. Abstract.

Electronic Supplementary Material Experimentally Validated Mathematical Model of Analyte Uptake by Permeation Passive Samplers

Validation of New VPH GC/MS Method using Multi-Matrix Purge and Trap Sample Prep System

A Single Calibration for Waters and Soil Samples Performing EPA Method Anne Jurek Applications Chemist

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Rapid Determination of TO-15 Volatile Organic Compounds (VOCs) in Air

Determination of VOCs by USEPA Method 8260 with Extended Dynamic Range using Fast, Sensitive Capillary GC/MS

Thermal Desorption Technical Support

Associates. Project No.: October 25, 2017

UCMR 3 Low-Level VOC Analysis by Purge and Trap Concentration and GC/MS using Selective Ion Monitoring

Meeting NJ Low Level TO-15 Air Testing Method Requirements

Reduced VOC Sample Analysis Times Using a New Dual Purge-and-Trap System

ANALYTICAL REPORT. Results relate only to the items tested and the sample(s) as received by the laboratory. Page 1 of 5

Techniques for Optimizing the Analysis of Volatile Organic Compounds in Water Using Purge-and-Trap/GC/MS Application

R.E.A.C.T. Roxbury Environmental Action CoaliTion P.O. Box 244 Ledgewood, N.J Website:

METHOD MEASUREMENT OF PURGEABLE ORGANIC COMPOUNDS IN WATER BY CAPILLARY COLUMN GAS CHROMATOGRAPHY/MASS SPECTROMETRY. Revision 4.

Standard Operating Procedure for the Analysis of Purgeable Organic Compounds (VOC s) in Water by Capillary Column Gas Chromatography/Mass Spectrometry

SUMMARY REPORT OF AIR MONITORING FOR LEED CERTIFICATION PHASE 2 SWANSFIELD ELEMENTARY SCHOOL 5610 CEDAR LANE COLUMBIA, MD PREPARED FOR:

Building 280A and Building 450 Soil Sampling Results Richmond Field Station Site Berkeley Global Campus at Richmond Bay, Richmond, California

Optimization of Method Parameters for the Evaluation of USEPA Method Using a Purge and Trap with GC/MS

ORGANIC COMPOUNDS IN WATER BY GAS CHROMATOGRAPHY/MASS SPECTROMETRY USING NITROGEN PURGE GAS

ANALYTICAL RESULTS. Project: Mayflower, AR Pipeline Incident

Copyright 2009 PerkinElmer LAS and CARO Analytical Services, Inc.

Copies: Dave Favero, RACER File

Application Note. Abstract. Introduction. Experimental-Instrument Conditions. By: Anne Jurek

Volatile Organic Compounds in Water PBM

ANALYTICAL REPORT. Job Number: SDG Number: Job Description: Joint Base Cape Cod

In-Tube Extraction Sample Preparation for Gas Chromatography

Detection of Environmental Contaminants Caused by the Oil Spill in the Gulf of Mexico by GC and Hplc

317 Elm Street Milford, NH (603) Fax (603)

Method 1624 Revision C Volatile Organic Compounds by Isotope Dilution GCMS

Table 1 Soil and Groundwater Sampling Summary Table

CIA Advantage-xr. Cryogen-free automated canister and whole-air sampling system

Readiness thru Health

January 19, Dear Mr. Nightingale:


CALA Directory of Laboratories

Author. Abstract. Introduction

1. Introduction. 2. Sampling Activities

Chlorinated Compounds in Hydrocarbon Streams Using a Halogen Specific Detector (XSD)

Analysis of Residual Solvents in Pharmaceuticals (USP<467>) with Shimadzu GC-2010 Plus and HS-10 Headspace Sampler

KSA Environmental Laboratory Inc.

Residual Solvents in Pharmaceuticals by USP Chapter <467> Methodology

Analysis of USP Method <467> Residual Solvents on the Agilent 8890 GC System

EA LABORATORIES EAL-M-8010A GROUP: Volatiles ANALYTICAL METHOD Halogenated Volatile Organics by Gas Chromatography Page: 1 of: 12

Transcription:

ENVIRONMENTAL application note U.S. EPA VOLATILE ORGANICS METHOD 524.2 USING PURGE AND TRAP GC/MS Cheri Coody, Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA Michael J. Burke and Elaine A. LeMoine, PerkinElmer Instruments, 761 Main Avenue, Norwalk, CT 06897 USA Introduction Clean drinking water is of worldwide interest. Contaminants can compromise the quality of a water supply and cause short- and long-term health effects. Stringent water quality criteria help to ensure safe drinking water by requiring the detection and quantification of a wide variety of contaminants at extremely low levels. U.S. EPA Method 524.2 (1) provides the guidance necessary for identifying volatile organic pollutants at concentrations consistent with water quality objectives. The equipment and conditions presented here provide data compliant with the rigorous Method 524.2 criteria. It is applicable to a wide range of volatile organic compounds and provides the necessary sensitivity needed to meet drinking water standards. Detection limits, calibration results, precision and accuracy data are presented for the more commonly targeted compounds. Method Summary This is a Gas Chromatography/Mass Spectrometry (GC/MS) method that uses the purge and trap technique for sample introduction. An aliquot of sample is purged of its volatile components by bubbling an inert gas through the sample. These compounds are then trapped on a sorbent material, heated, and backflushed into a gas chromatography (GC) column. Individual analytes are separated using chromatographic temperature programming and eluted to a mass spectrometer (MS) for detection. Identification is performed by comparing retention times and mass spectra to those obtained from known standards under identical conditions. Quantification is performed using the internal standard technique.

Instrument Conditions The equipment and conditions for the gas chromatograph, mass spectrometer, and purge and trap unit used to generate the data presented here are listed in Tables 1, 2, and 3. The GC column is connected to the MS transfer line using an open split interface. The interface consists of a 1/16" stainless steel Swagelok "T" connector. A 0.92 x 0.12 mm i.d. x 0.1 mm 5% phenyl methylsilicone fixed restrictor is inserted into the PerkinElmer TurboMass GC/MS with the other end running straight through the connector and out the other end. The chromatography column has the fused silica transfer line at the detector end of the column removed and the "T" is connected directly to the glass column, allowing the fixed restrictor to be inserted directly inside the chromatography column. The restrictor is inserted to a distance equal to 3.5 turns of the glass column and the effluent is vented out the back of the instrument. This configuration is appropriate for low-level drinking water volatile organic analyses. Table 1: Chromatographic Conditions PerkinElmer AutoSystem XL GC Column Oven Temperature Program Coolant Liquid CO 2 Manual Pneumatic Control (PPC) Helium at 15.0 ml/min Packed Injector 100 C Table 2: Mass Spectrometer Conditions PerkinElmer TurboMass Mass Spectrometer Mass Scan Range Scan Time Inter-scan Delay Filament Delay Ion Source Temperature Transfer Line Temperature Ionization Mode 35-260 m/z 0.5 sec 0.13 sec 2 min 150 C 200 C EI Vocol 60 m x 0.75 mm, 1.5-µm film thickness 10 C for 5.00 min; 6 C/min to 75 C for 10.00 min; 15 C/min to 145 C for 5.00 min; 15 C/min to 160 C for 5.00 min Table 3: Purge and Trap Conditions Tekmar LSC 3000 Sampler Purge Flow Gas He at 40 ml/min Time and Temperature Settings: Purge 11 min ambient Desorb 4 min at 180 C Bake 5 min at 220 C Trap Vocarb 3000 Sample Size 25 ml Method Performance Tuning Criteria To test instrument performance, a 25-ng (or less) standard solution of 4-bromofluorobenzene (BFB) is analyzed and the spectra compared to the abundance criteria listed in the method. Figure 1 demonstrates a successful BFB analysis to Method 524.2 abundance requirements using the above conditions. Calibration Initially, a method calibration covering the entire analytical range for each target analyte is performed. Specific criteria must be met before samples can be analyzed. A minimum of three standards for a factor of 20 calibration range is required. A factor of 50 range requires at least four and a factor of 100 at least 5. The Figure 1.Tune report. calibration data presented here are the results of six standards, specifically 0.5, 1.0, 2.5, 5.0, 10.0, and 20.0 ppb (a factor of 40 range). Every calibration standard contains 10 ppb of the internal standard fluoroben- zene and 1 ppb of the surrogate standard p-bromofluorobenzene. Demonstration of initial calibration performance is based on a comparison of the calculated Relative Standard Deviation (RSD) of the

standard s Relative Response Factors (RRF) to a maximum threshold value. Response Factors are calculated using the formula: RF = (A x )(Q is ) (A is )(Q x ) where: RF = Response Factor, A x = integrated abundance of the analyte quantitation ion, A is = integrated abundance of the internal standard quantitation ion, Q x = quantity of analyte purged in nanograms or concentration units, and Q is = quantity of internal standard purged in nanograms concentra tion units. The RSD is then calculated using the formula: RSD =100 (SD/RF) where: RSD = Relative Standard Deviation, SD = Standard Deviation, and RF = average relative Response Factor (of initial calibration standards). The calculated %RSD for the initial calibration must be less than 20% to be considered compliant. Alternatively, a linear or second order regression calibration curve can be used by plotting: A x /A is vs. Q x Table 4: Compliant Initial Calibration Compound %RSD Compound %RSD Benzene 7.8 1,3-Dichloropropane 10.0 Bromobenzene 11.8 2,2-Dichloropropane 7.7 Bromochloromethane 8.5 1,1-Dichloropropene 3.7 Bromodichloromethane 6.7 cis-1,3-dichloropropene 5.0 Bromoform 12.9 trans-1,3-dichloropropene 4.6 Bromomethane 4.9 Ethylbenzene 9.5 n-butylbenzene 7.9 Hexachlorobutadiene 5.3 sec-butylbenzene 5.2 Isopropylbenzene 5.6 tert-butylbenzene 5.0 4-Isopropyltoluene 5.2 Carbon tetrachloride 6.3 Methylene chloride 6.5 Chlorobenzene 10.7 Naphthalene 7.2 Chloroethane 3.0 n-propylbenzene 8.5 Chloroform 9.2 Styrene 7.8 Chloromethane 15.5 1,1,1,2-Tetrachloroethane 11.1 2-Chlorotoluene 13.3 1,1,2,2-Tetrachloroethane 17.3 4-Chlorotoluene 11.5 Tetrachloroethene 4.2 Dibromochloromethane 9.8 Toluene 8.3 Dibromomethane 6.2 1,2,3-Trichlorobenzene 6.6 1,2-Dichlorobenzene 8.7 1,2,4-Trichlorobenzene 6.4 1,3-Dichlorobenzene 9.1 1,1,1-Trichloroethane 3.3 1,4-Dichlorobenzene 9.8 1,1,2-Trichloroethane 9.8 Dichlorodifluoromethane 11.8 Trichloroethene 5.9 Dichloroethane 9.0 Trichlorofluoromethane 3.3 1,2-Dichloroethane 7.9 1,2,4-Trimethylbenzene 5.8 1,1-Dichloroethene 11.2 1,3,5-Trimethylbenzene 5.6 cis-1,2-dichloroethene 8.3 Vinyl chloride 4.7 Trans-1,2-Dichloroethene 5.5 o-xylene 11.1 1,2-Dichloropropane 9.4 m&p-xylene 9.6 Compliance Limit < 20% Compliance Limit < 20% The data presented in Table 4 list the %RSDs for a compliant calibration using the RSD test. Figure 2 exemplifies a compliant linear regression curve where the coefficient of determination is greater than or equal to 0.99. Precision and Accuracy Precision and accuracy must be demonstrated initially by analyzing replicates of each analyte at a concentration of 2 5 µg/l. The measured concentrations are calculated and averaged. Accuracy is the percentage of the true value measured as % Recovery and precision is the Relative Standard Deviation of this value. Accordingly, the % Recoveries must comply with the 80-120% accuracy requirements and all Figure 2. Linear regression curve of vinyl chloride.

the %RSDs must be less than the 20% maximum to demonstrate precision. Table 5 lists the precision and accuracy data for target analytes obtained from eight (8) replicate analyses using 2.5-µg/L standards. All the compounds show compliant precision and accuracy. Detection Limits Five (5) replicates of a 1.0-µg/L standard solution were analyzed to determine the Method Detection Limits (MDLs). As stated in Method 524.2, the detection limits are calculated using the following formula: MDL = S t (n-1, alpha = 0.99) where: t (n-1, alpha = 0.99) = Student s t value for the 99% confidence level with n-1 degrees freedom, n = number of replicates, and S = the standard deviation of the replicate analyses. Table 5: Precision and Accuracy Results from Eight Replicate Analyses of a 2.5 µg/l Standard Precision and Accuracy Data Average Compound (µg/l) %Recovery %RSD Benzene 2.26 90.2 3.3 Bromobenzene 2.59 103.4 10.0 Bromochloromethane 2.18 87.3 3.2 Bromodichloromethane 2.31 92.3 3.0 Bromoform 2.20 87.9 7.3 Bromomethane 2.94 117.6 8.9 n-butylbenzene 2.57 102.7 10.0 sec-butylbenzene 2.09 83.4 9.6 tert-butylbenzene 2.57 102.6 4.7 Carbon tetrachloride 2.80 112.1 2.7 Chlorobenzene 2.37 94.8 10.0 Chloroethane 2.74 109.6 5.1 Chloroform 2.45 97.9 1.2 Chloromethane 2.28 91.2 8.3 2-Chlorotoluene 2.39 95.6 10.6 4-Chlorotoluene 2.51 100.4 10.2 Dibromochloromethane 2.57 102.8 8.5 Dibromomethane 2.62 104.6 2.6 1,2-Dichlorobenzene 2.55 102.1 10.7 1,3-Dichlorobenzene 2.58 103.0 10.1 1,4-Dichlorobenzene 2.57 103.0 10.5 Dichlorodifluoromethane 2.32 92.8 10.5 Dichloroethane 2.89 115.6 4.2 1,2-Dichloroethane 2.82 112.6 3.7 1,1-Dichloroethene 2.70 107.8 5.2 cis-1,2-dichloroethene 2.87 114.7 2.3 Trans-1,2-Dichloroethene 2.45 98.0 1.7 1,2-Dichloropropane 2.98 119.3 3.1 1,3-Dichloropropane 2.23 89.3 3.3 2,2-Dichloropropane 2.36 94.4 8.4 1,1-Dichloropropene 2.31 92.3 2.3 cis-1,3-dichloropropene 2.11 84.2 5.1 trans-1,3-dichloropropene 2.18 87.1 4.4 Ethylbenzene 2.55 101.9 7.4 Hexachlorobutadiene 2.57 102.8 14.0 Isopropylbenzene 2.08 83.3 9.6 4-Isopropyltoluene 2.66 106.2 5.9 Methylene chloride 2.26 90.4 3.3 Naphthalene 2.74 109.5 4.0 n-propylbenzene 2.46 98.4 8.7 Styrene 2.25 89.8 8.7 1,1,1,2-Tetrachloroethane 2.58 103.0 7.1 1,1,2,2-Tetrachloroethane 2.55 102.2 5.0 Tetrachloroethene 2.58 103.3 3.3 Toluene 2.82 112.9 2.5 1,2,3-Trichlorobenzene 2.53 101.3 10.2 1,2,4-Trichlorobenzene 2.69 107.6 9.9 1,1,1-Trichloroethane 2.54 101.7 2.0 1,1,2-Trichloroethane 2.54 101.7 2.7 Trichloroethene 2.57 102.7 1.1 Trichlorofluoromethane 2.67 106.6 8.9 1,2,4-Trimethylbenzene 2.25 90.0 10.6 1,3,5-Trimethylbenzene 2.11 84.5 10.0 Vinyl chloride 2.63 105.2 2.6 o-xylene 2.51 100.3 6.3 m&p-xylene* 4.89 97.8 12.5 Compliance Criteria 80 120 % < 20% * Total concentration 5.0 µg/l

Table 6 lists MDLs for all the listed target analytes obtained using the same conditions and calculations referenced here. Detection limits obtained using a wide-bore capillary column referenced in Method 524.2, range from 0.019 1.6 µg/l. The Maximum Contami- nant Levels (MCLs) cited in the National Primary Drinking Water Standards (2) are also listed for comparison. The MDLs are well below the MCLs and compare favorably with the guidence limits listed in the U.S. EPA method. Detection limits can vary tremendously depending on a number of factors such as the exact analytical conditions employed, the technique of the analyst, and the concentration of the standard used to make the determinations. Table 6: Method Detection Limits and U.S Maximum Contaminant Levels Method Detection Limits and Drinking Water Maximum Contaminant Limits Analyte MDL MCL Analyte MDL MDL (ppb) (ppb) (ppb) (ppb) Benzene 0.21 5. 1,3-Dichloropropane 0.20 Bromobenzene 0.35 2,2-Dichloropropane 0.43 Bromochloromethane 0.29 1,1-Dichloropropene 0.21 Bromodichloromethane* 0.24 100. cis-1,3-dichloropropene 0.13 Bromoform* 0.59 100. trans-1,3-dichloropropene 0.16 Bromomethane 0.49 Ethylbenzene 0.34 700. n-butylbenzene 0.29 Hexachlorobutadiene 0.38 sec-butylbenzene 0.44 Isopropylbenzene 0.47 tert-butylbenzene 0.33 4-Isopropyltoluene 0.27 Carbon tetrachloride 0.37 5. Methylene chloride 0.23 5. Chlorobenzene 0.44 100. Naphthalene 0.39 Chloroethane 0.28 n-propylbenzene 0.32 Chloroform* 0.19 100. Styrene 0.43 100. Chloromethane 0.50 1,1,1,2-Tetrachloroethane 0.41 2-Chlorotoluene 0.62 1,1,2,2-Tetrachloroethane 0.57 4-Chlorotoluene 0.41 Tetrachloroethene 0.20 5. Dibromochloromethane* 0.53 100. Toluene 0.10 100. Dibromomethane 0.36 1,2,3-Trichlorobenzene 0.40 1,2-Dichlorobenzene 0.31 600. 1,2,4-Trichlorobenzene 0.38 70. 1,3-Dichlorobenzene 0.29 1,1,1-Trichloroethane 0.25 200. 1,4-Dichlorobenzene 0.30 75. 1,1,2-Trichloroethane 0.33 5. Dichlorodifluoromethane 0.77 Trichloroethene 0.22 5. Dichloroethane 0.23 5. Trichlorofluoromethane 0.35 1,2-Dichloroethane 0.19 1,2,4-Trimethylbenzene 0.55 1,1-Dichloroethene 0.41 7. 1,3,5-Trimethylbenzene 0.51 cis-1,2-dichloroethene 0.28 70. Vinyl chloride 0.50 2. Trans-1,2-Dichloroethene 0.14 100. o-xylene 0.37 1,2-Dichloropropane 0.20 5. m&p-xylene 0.66 * Total Trihalomethanes must be less than 100 ppb. Conclusion The accurate determination of volatile organic compounds in drinking water is important for the health and safety of the population. The analysis can be challenging and requires a rugged method for optimum results. Tuning, calibration, method detection limits, precision, and accuracy data presented here are compliant with those listed in the U.S. EPA method and serve to highlight some of the more critical criteria for method validation using the TurboMass GC/MS system.

References 1. "Measurement of Purgeable Organic Compounds in Water By Capillary Column Gas Chromatography/Mass Spectrometry," Method # 524.2, Revision 4.1, Methods for the Determination of Organic Compounds in Drinking Water, Supplement III, EPA-600/R- 95/131, August 1995. 2. "Maximum Contaminant Levels for Organic Contaminants," Code of Federal Regulations, 40 CFR Ch. I (7 1 98 Edition), Part 141.61. Acknowledgement Special thanks to Sami Malone of the Public Health Laboratory, Mississippi State Health Department, Jackson, MS USA. Visit our website at www.perkinelmer.com PerkinElmer Instruments 761 Main Avenue Norwalk, CT 06859-0010 USA Tel: 800-762-4000 or (+1) 203-762-4000 Fax: (+1) 203-762-4228 PerkinElmer is a trademark of PerkinElmer, Inc. TurboMass is a trademark of PerkinElmer Instruments LLC. Swagelok is a registered trademark of Swagelok Co. 2000 PerkinElmer, Inc. Printed in U.S.A. D-6224 KG070001