METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

Similar documents
METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

METHOD 8030A ACROLEIN AND ACRYLONITRILE BY GAS CHROMATOGRAPHY

METHOD 8430 ANALYSIS OF BIS(2-CHLOROETHYL) ETHER AND HYDROLYSIS PRODUCTS BY DIRECT AQUEOUS INJECTION GC/FT-IR

METHOD 8111 HALOETHERS BY GAS CHROMATOGRAPHY

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

METHOD 5000 SAMPLE PREPARATION FOR VOLATILE ORGANIC COMPOUNDS

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

METHOD 8091 NITROAROMATICS AND CYCLIC KETONES BY GAS CHROMATOGRAPHY

METHOD 3520C CONTINUOUS LIQUID-LIQUID EXTRACTION

STANDARD OPERATING PROCEDURES SOP: 1826 PAGE: 1 of 18 REV: 0.0 DATE: 03/30/95 ANALYSIS OF METHYL PARATHION IN WIPE SAMPLES BY GC/MS

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

METHOD 7199 DETERMINATION OF HEXAVALENT CHROMIUM IN DRINKING WATER, GROUNDWATER AND INDUSTRIAL WASTEWATER EFFLUENTS BY ION CHROMATOGRAPHY

PA-DEP 3686, Rev. 1. Light Hydrocarbons in Aqueous Samples via Headspace and Gas Chromatography with Flame Ionization Detection (GC/FID)

METHOD 8000A GAS CHROMATOGRAPHY

METHOD 3520B CONTINUOUS LIQUID-LIQUID EXTRACTION

METHOD 3665 SULFURIC ACID/PERMANGANATE CLEANUP

METHOD 3510B SEPARATORY FUNNEL LIQUID-LIQUID EXTRACTION

METHOD 8082A. POLYCHLORINATED BIPHENYLS (PCBs) BY GAS CHROMATOGRAPHY

METHOD 3600C CLEANUP

METHOD 8015C NONHALOGENATED ORGANICS BY GAS CHROMATOGRAPHY

METHOD 3600B CLEANUP

METHOD 9210 POTENTIOMETRIC DETERMINATION OF NITRATE IN AQUEOUS SAMPLES WITH ION-SELECTIVE ELECTRODE

METHOD 9040B. ph ELECTROMETRIC MEASUREMENT

637. Thiamethoxam. HPLC method

ANALYTICAL METHOD DETERMINATION OF VOLATILE ALDEHYDES IN AMBIENT AIR Page 1 of 11 Air sampling and analysis

METHOD 9252A CHLORIDE (TITRIMETRIC, MERCURIC NITRATE)

METHOD 8015D NONHALOGENATED ORGANICS USING GC/FID

METHOD 8041 PHENOLS BY GAS CHROMATOGRAPHY

Perfluorinated Alkyl Acids (PFAA) in Water by LC/MS/MS - PBM

Sulfotepp impurities in Chlorpyrifos EC formulations

NCASI METHOD CI/SG/PULP-94.02

TECHNICAL BRODIFACOUM

QUALITY CONTROL CRITERIA FOR CHEMISTRY EXCEPT RADIOCHEMISTRY.

Technical Procedure for Concentration Determination of Methamphetamine in Liquids via HPLC

STANDARD OPERATING PROCEDURES

METHOD 7060A ARSENIC (ATOMIC ABSORPTION, FURNACE TECHNIQUE)

METHOD 8330A NITROAROMATICS AND NITRAMINES BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

Date 10/22/97 Page 1 Revision 4.1 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8000/8100 (MODIFIED) DIESEL RANGE ORGANICS (DRO)

GAFTI Analytical method for ISO/TS 16179:2012 Detection and Determination of Organotin Compounds in Footwear and Apparel Materials by GC-MS

GUIDELINES FOR THE DESIGN OF CHROMATOGRAPHIC ANALYTICAL METHODS INTENDED FOR CIPAC COLLABORATIVE STUDY

ALLOWAY METHOD OUTLINE

Method of determination of phtalates in spirituous beverages by gaschromatography/mass

CHAPTER 6 GAS CHROMATOGRAPHY

Hexavalent Chromium by Flow Injection Analysis (FIA)

Hach Method Total Organic Carbon in Finished Drinking Water by Catalyzed Ozone Hydroxyl Radical Oxidation Infrared Analysis

METHOD 3010A ACID DIGESTION OF AQUEOUS SAMPLES AND EXTRACTS FOR TOTAL METALS FOR ANALYSIS BY FLAA OR ICP SPECTROSCOPY

PYRIPROXYFEN TECHNICAL

Date 02/24/96 Page 1 Revison 4.0 OKLAHOMA DEPARTMENT OF ENVIRONMENTAL QUALITY METHODS 8020/8015 (MODIFIED) GASOLINE RANGE ORGANICS (GRO)

CIPAC. CIPAC Free relevant impurities methods:

TECHNICAL TEMEPHOS. 1. Specification. Full specification WHO/SIT/19.R4 Revised 10 December Description

SECTION D.2 AMMONIA NITROGEN

Determination of Volatile Substances Proof of Food Adulteration

Method 548. July 1990 EPA EMSL-Ci. J.W. Hodgeson

DETERMINATION OF DRUG RELEASE DURING DISSOLUTION OF NICORANDIL IN TABLET DOSAGE FORM BY USING REVERSE PHASE HIGH PERFORMANCE LIQUID CHROMATOGRAPHY

Method for Characterization of Gum Rosin by Capillary Gas Chromatography

Sodium Chloride - Analytical Standard

METHOD 5021 VOLATILE ORGANIC COMPOUNDS IN SOILS AND OTHER SOLID MATRICES USING EQUILIBRIUM HEADSPACE ANALYSIS

VALIDATION OF A UPLC METHOD FOR A BENZOCAINE, BUTAMBEN, AND TETRACAINE HYDROCHLORIDE TOPICAL SOLUTION

Chemistry Gas Chromatography: Separation of Volatile Organics

Technical Procedure for Blood Cannabinoid Liquid-Liquid Extraction (BCLLE) for Analysis by LC-MS/MS

METHOD 547 DETERMINATION OF GLYPHOSATE IN DRINKING WATER BY DIRECT-AQUEOUS- INJECTION HPLC, POST-COLUMN DERIVATIZATION, AND FLUORESCENCE DETECTION

Standard Test Method for Determination of Trace Thiophene in Refined Benzene by Gas Chromatography 1

Simultaneous Estimation of Residual Solvents (Isopropyl Alcohol and Dichloromethane) in Dosage Form by GC-HS-FID

H 3 CO H 3 CO S CH 3

British American Tobacco Group Research & Development. Method - Determination of phenols in mainstream cigarette smoke

STANDARD OPERATING PROCEDURES SOP: 1824 PAGE: 1 of 22 REV: 0.0 DATE: 04/21/95 ANALYSIS OF METHYL PARATHION IN SOIL SAMPLES BY GC/MS

Z/zs. tnitiats: wu\ Z-z* I A. sop-c-102. Determination of Chemical Oxygen Demand. Revision 6. Approval: Effective date: 3. Renewat date: 7lt:,:

TO-15 Checklist Determination of VOCs in Air by GC-MS

Draft Method proposal: determination of glucoheptonic acid (HGA) in fertilizers.

Colorimetric Method Method to 0.70 mg/l Ag Powder Pillows

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

NEVIRAPINE ORAL SUSPENSION Final text for addition to The International Pharmacopoeia (February 2009)

METHOD 7196A CHROMIUM, HEXAVALENT (COLORIMETRIC)

Total Kjeldahl Nitrogen by Gas Diffusion and Flow Injection Analysis (FIA)

MEDROXYPROGESTERONE INJECTION

Anethole. Gas chromatograhpy determination of trans-anethole in Spirit srinks of viti-vinicultural origin

METHOD 8271 ASSAY OF CHEMICAL AGENTS IN SOLID AND AQUEOUS SAMPLES BY GAS CHROMATOGRAPH/MASS SPECTROMETRY, ELECTRON IMPACT (GC/MS/EI)

Chloride by Flow Injection Analysis (FIA)

PCB Congener Analysis of XAD-2 Resins and GFF Filters Using GC/ECD

METHOD 509 DETERMINATION OF ETHYLENE THIOUREA (ETU) IN WATER USING GAS CHROMATOGRAPHY WITH A NITROGEN-PHOSPHORUS DETECTOR. Revision 1.0.

METHOD 8315A DETERMINATION OF CARBONYL COMPOUNDS BY HIGH PERFORMANCE LIQUID CHROMATOGRAPHY (HPLC)

METHOD 9012B TOTAL AND AMENABLE CYANIDE (AUTOMATED COLORIMETRIC, WITH OFF-LINE DISTILLATION)

Corn Syrup Analysis E-51-1 PHOSPHORUS

Method - Determination of aromatic amines in mainstream cigarette smoke

Acid Volatile Sulfides and Simultaneously Extracted Metals in Sediments

--> Buy True-PDF --> Auto-delivered in 0~10 minutes. GB Translated English of Chinese Standard: GB5009.

Method 627. The Determination of Dinitroaniline Pesticides in Municipal and Industrial Wastewater

ANALYSIS OF PERCHLORATE IN DRINKING WATER, GROUND WATER, SALINE WATER, SOIL, AND BIOTA BY LC/MSD

GB Translated English of Chinese Standard: GB NATIONAL STANDARD OF

Uncontrolled Copy. SOP 109 Ethylene Glycol Screen by Gas Chromatography/Mass Spectrometry. Table of Contents. 1. Principle of Assay...

APPENDIX A TO PART 136 METHODS FOR ORGANIC CHEMICAL ANALYSIS OF MUNICIPAL AND INDUSTRIAL WASTEWATER METHOD 609 NITROAROMATICS AND ISOPHORONE

Hach Company TNTplus 835/836 Nitrate Method Spectrophotometric Measurement of Nitrate in Water and Wastewater

Test Method: CPSC-CH-E

Standard Test Method for Anions in Water by Chemically Suppressed Ion Chromatography 1

Method Formaldehyde, Isobutyraldehyde, and Furfural by Derivatization Followed by High Performance Liquid Chromatography

Quantitative determination of cannabinoids in Hemp and Hemp products by High Performance Liquid Chromatography using a UV Detector

Hach Method Spectrophotometric Measurement of Free Chlorine (Cl 2 ) in Finished Drinking Water

LAP-013. Procedure Title: Date: 6/28/96. Author: Raymond Ruiz and Tina Ehrman ISSUE DATE: 8/26/96 SUPERSEDES: 11/01/94.

1. A gas chromatograph equipped with flame ionization detection and a computing integrator.

USGS District Laboratory, Troy, NY Dissolved Inorganic Carbon Analysis Standard Operating Procedure

Transcription:

METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION 1.0 SCOPE AND APPLICATION 1.1 Method 80 may be used to determine the concentration of acetonitrile (CAS No. 75-05-8) in aqueous matrices. 1.2 The method detection limit in water is approximately 1.7 to 2.8 picograms (1.7 to 2.8 µg/l with a 1 µl injection) at 40 and 20 volts offset, respectively. The upper limit of the range is approximately 5000 µg/l and is limited by the non-linearity of the nitrogen-phosphorus detector (NPD). The range may be extended by diluting the sample. This procedure may be applicable to other matrices where the target compound can be extracted into water. 1. This method is also applicable to aqueous condensate collected from the VOST or semi- VOST sampling trains. 1.4 This method is restricted to use by or under the supervision of analysts experienced in the use of gas chromatographs and skilled in the interpretation of gas chromatograms. Each analyst must demonstrate the ability to generate acceptable results with this method. 2.0 SUMMARY OF METHOD Water samples containing acetonitrile are introduced into a GC/NPD using cool on-column direct aqueous injection. The GC/NPD system is externally calibrated using a series of acetonitrile standards diluted in organic-free reagent water. A series of blank organic-free reagent water samples are directly injected until no peaks are observed in the chromatogram. The acetonitrile concentration is quantitated using the external calibration method..0 INTERFERENCES.1 There are no known chromatographic interferants when samples are analyzed using the optimum operating conditions. However, at column temperatures greater than 60EC, water becomes an interferant, as it coelutes with acetonitrile and produces a negative disruption of the baseline, thereby preventing accurate quantitations. This problem is eliminated by operating the column at 5EC. At this temperature, water will eventually elute and produce a positive disruption of the baseline. However, this effect can be minimized by heating the column to 120EC for 1 hour after a day's analysis..2 Glassware containing traces of chromium should not be used because chromium will form a complex with acetonitrile, thereby reducing the amount of acetonitrile available for quantitation.. Reagents, glassware, and other sample processing hardware may yield discrete artifacts and/or elevated baselines causing misinterpretation of gas chromatograms. All these materials CD-ROM 80-1 Revision 0

must be demonstrated to be free from interferences under the conditions of the analysis, by analyzing reagent blanks. 4.0 APPARATUS AND MATERIALS 4.1 Gas chromatograph 4.1.1 Gas chromatograph, analytical system complete with gas chromatograph suitable for cool on-column injections and all required accessories, including detector, analytical columns, recorder, gases, and syringes. A data system for measuring peak heights and/or peak areas is recommended. 4.1.2 Column: DB-WAX, 15 m X 0.5 mm, 1.0 µm capillary column (J&W Scientific or equivalent). 4.1. Detector: Nitrogen-Phosphorus (NPD). 4.2 Bubble flow-meter: 0-10 and 0-100 cm /min, for measuring GC gas flows. 4. Syringes: 10 µl capacity, equipped with fused silica needles. Use of a syringe with a stainless steel needle has not been evaluated. 4.4 Volumetric flasks: Class A, sizes as appropriate. 4.5 Pipets: Class A, assorted sizes. 5.0 REAGENTS 5.1 Reagent grade chemicals shall be used in all tests. Unless otherwise indicated, it is intended that all reagents shall conform to the specifications of the Committee on Analytical Reagents of the American Chemical Society, where such specifications are available. Other grades may be used, provided it is first ascertained that the reagent is of sufficiently high purity to permit its use without lessening the accuracy of the determination. 5.2 Organic-free reagent water - All references to water in this method refer to organic-free reagent water, as defined in Chapter One. 5. Gases 5..1 Helium gas - 99.9999% purity, for carrier and make-up gas. Helium is preferred as the make-up gas over nitrogen because helium produces a more stable baseline with less noise. 5..2 Air - zero grade. 5.. Nitrogen - zero grade. 5.4 Acetonitrile, CH CN - 99.9% purity. CD-ROM 80-2 Revision 0

5.5 2-Propanol, (CH ) CHOH - Pesticide quality or equivalent. 2 5.6 Stock standard solutions - Prepare from pure standard materials or purchase as certified solutions. 5.6.1 Prepare stock standard solutions by accurately weighing the appropriate amount of acetonitrile and dissolving it in organic-free reagent water to achieve a final concentration of at least 1000 mg/l. When compound purity is assayed to be 96% or greater, the weight may be used without correction to calculate the concentration of the stock standard. 5.6.2 Transfer the stock standard solutions into bottles with polytetrafluoroethylene (PTFE)-lined screw-caps. The stock standard solution is stable at 20-25EC for about months. However, stock standard solutions should be checked frequently for signs of degradation or evaporation, especially when using them to prepare calibration standards. 5.7 Working standard solutions - Prepare working standards fresh daily by diluting the stock standard solutions with organic-free reagent water. Prepare at least five different acetonitrile concentrations to bracket the expected concentration range of the samples. 6.0 SAMPLE COLLECTION, PRESERVATION, AND HANDLING 6.1 See the introductory material to this chapter, Organic Analytes, Sec. 4.1. Acetonitrile has a boiling point of approximately 81.6EC and should therefore be handled as a volatile compound. 6.2 Calibration standards are stable for at least one day. Traces of chromium in the glassware used to prepare the calibration standards may adversely affect longer term stability. 7.0 PROCEDURE 7.1 This procedure utilizes direct aqueous injection (DAI) as the sample introduction technique. Samples of condensate from a VOST or semi-vost sampling train may need to be diluted to an appropriate concentration prior to analysis. 7.2 Chromatographic Conditions (Recommended): Collector distance above the jet: 1 mm. Column temperature: 5EC. Carrier gas (He) flow rate: 7 cm /min. Make-up gas (He) flow rate: 26 cm /min. H 2 flow rate:.5 cm /min. Air flow rate: 80 cm /min. Injection temperature: Ambient temperature Injection volume: 1 µl Detector temperature: 00EC. Detector: Nitrogen-Phosphorus (NPD) operated at the conditions specified by the manufacturer. 7.2.1 Under these chromatographic conditions the retention time of acetonitrile is approximately 4.4 minutes. The offset voltage is very sensitive to H flow rate. 2 CD-ROM 80 - Revision 0

7.2.2 Maximum control of the H flow rate is obtained by using a pressure regulator at 2 the instrument. The stability of the offset voltage during the day's analysis will be enhanced by allowing the NPD to equilibrate overnight, prior to carrying out an analysis. 7. Calibration - External standard calibration should be utilized for this single analyte procedure. Refer to Method 8000 for guidance on implementation of an external standard initial and continuing calibration procedure. 7.4 Calibration standard and sample analysis 7.4.1 Allow the baseline to stabilize at the desired voltage. Inject blank samples of water (a 1.0 µl volume is recommended) until no peaks are observed in the chromatogram. An occasional positive baseline disruption (caused by the water) may occur. NOTE: The acetonitrile concentration of any standard or sample will decrease with time if left exposed to the atmosphere. Use appropriate precautions or use fresh sample for each injection to minimize this effect. 7.4.2 Inject the 5 calibration standards in order of increasing concentration. It may be necessary to condition the syringe with organic-free reagent water by rinsing it a minimum of 15 times between injections to achieve acceptable method performance. Calculate the calibration factors (CF) for each standard to determine the linearity of the calibration. 7.4. Inject the sample into the GC. If the response for the sample exceeds that of the initial calibration range, the sample must be diluted and reanalyzed. Due to the relatively short retention time of the target compound and potential instability of the detector, each sample or sample dilution may need to be analyzed in duplicate. To achieve the precision described in Section 9.0, it may be necessary to condition the syringe with organic-free reagent water by rinsing it a minimum of 15 times between injections. 7.4.4 Calibration check samples should be analyzed after every 10 samples. 7.5 Refer to Method 8000 for guidance on GC peak identification, peak confirmation, and sample concentration calculations. 8.0 QUALITY CONTROL 8.1 Refer to Chapter One and Method 8000 for specific quality control (QC) procedures. Quality control procedures to ensure the proper operation of the various sample preparation and/or sample introduction techniques can be found in Methods 500 and 5000. Each laboratory should maintain a formal quality assurance program. The laboratory should also maintain records to document the quality of the data generated. 8.2 Quality control procedures necessary to evaluate the GC system operation are found in Method 8000, Sec. 7.0 and includes evaluation of retention time windows, calibration verification and chromatographic analysis of samples. 8. Initial Demonstration of Proficiency - Each laboratory must demonstrate initial proficiency with each sample preparation and determinative method combination it utilizes, by generating data of acceptable accuracy and precision for target analytes in a clean matrix. The laboratory must also CD-ROM 80-4 Revision 0

repeat the following operations whenever new staff are trained or significant changes in instrumentation are made. See Method 8000, Sec. 8.0 for information on how to accomplish this demonstration. 8.4 Sample Quality Control for Preparation and Analysis - The laboratory must also have procedures for documenting the effect of the matrix on method performance (precision, accuracy, and detection limit). At a minimum, this includes the analysis of QC samples including a method blank, a matrix spike, a duplicate, and a laboratory control sample (LCS) in each analytical batch. 8.4.1 Documenting the effect of the matrix should include the analysis of at least one matrix spike and one duplicate unspiked sample or one matrix spike/matrix spike duplicate pair. The decision on whether to prepare and analyze duplicate samples or a matrix spike/matrix spike duplicate must be based on a knowledge of the samples in the sample batch. If samples are expected to contain target analytes, then laboratories may use one matrix spike and a duplicate analysis of an unspiked field sample. If samples are not expected to contain target analytes, laboratories should use a matrix spike and matrix spike duplicate pair. 8.4.2 A Laboratory Control Sample (LCS) should be included with each analytical batch. The LCS consists of an aliquot of a clean (control) matrix similar to the sample matrix and of the same weight or volume. The LCS is spiked with the same analytes at the same concentrations as the matrix spike. When the results of the matrix spike analysis indicate a potential problem due to the sample matrix itself, the LCS results are used to verify that the laboratory can perform the analysis in a clean matrix. 8.4. See Method 8000, Sec. 8.0 for the details on carrying out sample quality control procedures for preparation and analysis. 8.5 It is recommended that the laboratory adopt additional quality assurance practices for use with this method. The specific practices that are most productive depend upon the needs of the laboratory and the nature of the samples. Whenever possible, the laboratory should analyze standard reference materials and participate in relevant performance evaluation studies. 9.0 METHOD PERFORMANCE Using laboratory produced samples containing 0.01 ng/µl and 0. ng/µl acetonitrile in H O 2 and a 1.0 µl injection volume, the relative standard deviation was 2.0% and 1.4%, respectively. Analysis of an unknown (QC sample) showed that calibration with external standards was accurate within ± 4% of the true value. See Figure 1 for sample chromatogram of acetonitrile. 10.0 REFERENCES 1. "Draft Method for the Analysis of Acetonitrile in a Water Matrix", U.S. Environmental Protection Agency, AREAL/RTP, April, 1990. 2. Margeson, J., Memorandum: Evaluation of Method 80 (Acetonitrile by GC/NPD), U.S. Environmental Protection Agency, AREAL/Source Methods Research Branch, RTP, NC, August 14, 1991. CD-ROM 80-5 Revision 0

FIGURE 1 CHROMATOGRAM OF ACETONITRILE STANDARD (Retention Time = 4. minutes) CD-ROM 80-6 Revision 0

METHOD 80 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION CD-ROM 80-7 Revision 0