Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System

Similar documents
AB SCIEX SelexION Technology Used to Improve Mass Spectral Library Searching Scores by Removal of Isobaric Interferences

Powerful Scan Modes of QTRAP System Technology

Accelerating the Metabolite Identification Process Using High Resolution Q-TOF Data and Mass-MetaSite Software

Application Note. Edgar Naegele. Abstract

Development of high speed CYP cocktail inhibition assay using UHPLC-MS/MS

A Rapid Approach to the Confirmation of Drug Metabolites in Preclinical and Clinical Bioanalysis Studies

Key Words Q Exactive, Accela, MetQuest, Mass Frontier, Drug Discovery

Analysis of a Verapamil Microsomal Incubation using Metabolite ID and Mass Frontier TM

Overview. Introduction. André Schreiber AB SCIEX Concord, Ontario (Canada)

4000 Q TRAP LC/MS/MS System. Advanced Linear Ion Trap technology at the highest level of sensitivity 4000 QTRAP. LC/MS/MS System

Analyst Software. Peptide and Protein Quantitation Tutorial

Yun W. Alelyunas, Mark D. Wrona, Russell J. Mortishire-Smith, Nick Tomczyk, and Paul D. Rainville Waters Corporation, Milford, MA, USA INTRODUCTION

Quantitative Analysis and Identification of Migrants in Food Packaging Using LC-MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

A Workflow Approach for the Identification and Structural Elucidation of Impurities of Quetiapine Hemifumarate Drug Substance

All Ions MS/MS: Targeted Screening and Quantitation Using Agilent TOF and Q-TOF LC/MS Systems

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

Rapid Quan/Qual Metabolic Stability Analysis with Online Oxidative Metabolism Synthesis

Overview. Introduction. André Schreiber 1 and Yun Yun Zou 1 1 AB SCIEX, Concord, Ontario, Canada

Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou AB SCIEX, Concord, Ontario, Canada

Stephen McDonald, Mark D. Wrona, Jeff Goshawk Waters Corporation, Milford, MA, USA INTRODUCTION

Simultaneous analysis for forensic drugs in human blood and urine using ultra-high speed LC-MS/MS

Rapid method development to study plasma stability of diverse pharmaceutical compounds using Rapid Resolution LC and triple quadrupole MS

The Quantitation and Identification of Coccidiostats in Food by LC-MS/MS using the AB SCIEX 4000 Q TRAP System

Key Features of Targeted Metabolite Screening. Using the TripleTOF System, an Accurate Mass Metabolite Spectral Library and MasterView Software

At-a-Glance. Verapamil C 27 H 38 N 2 O 4 M+H + =

Food & Environmental. Overview. Introduction. Method Details. André Schreiber and Yun Yun Zou SCIEX, Concord, Ontario, Canada

Ultrafast Analysis of Buprenorphine and Norbuprenorphine in Urine Using the Agilent RapidFire High-Throughput Mass Spectrometry System

Taking Full Advantage of UHPLC with Agilent 6460A Triple Quadrupole MS. Outline

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Metabolomics Batch Data Analysis Workflow to Characterize Differential Metabolites in Bacteria

High-sensitivity, high-throughput quantitation of catecholamine metabolites in urine by LC/MS/MS for clinical research

Measurement of Metabolic Stability Using SIM and Identification of Metabolites by Data Dependent Full-Scan MS/MS and CNL Scanning

Highly sensitive and rapid analysis of synthetic dyes in sea food by LC/MS/MS

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Improved Automated Sample Preparation for the Analysis of 25-OH-Vitamin D3 by LC/MS/MS

Analyst Software. Automatic Optimization Tutorial

+ butanal + 2-picolineborane. + acetic acid. + butanal + 2-picolineborane. + acetic acid

Metabolomic Profiling of Accurate Mass LC-MS/MS Data to Identify Unexpected Environmental Pollutants

Identification and Characterization of an Isolated Impurity Fraction: Analysis of an Unknown Degradant Found in Quetiapine Fumarate

Simplified Approaches to Impurity Identification using Accurate Mass UPLC/MS

Rapid and Accurate Forensics Analysis using High Resolution All Ions MS/MS

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Analysis and Quantification of Mycotoxins in Cereals Using MRM HR on the SCIEX X500R QTOF System with SCIEX OS

Ultrafast Analysis of Metabolic Stability Assays Using Agilent RapidFire High-resolution MS

ABI 3200 Q TRAP LC/MS/MS System

NEW TOOLS FOR FINDING AND IDENTIFYING METABOLITES IN A METABOLOMICS WORKFLOW

Guide to Peptide Quantitation. Agilent clinical research

Agilent 1290 Infinity Quaternary LC Stepwise Transfer to Methods with MS-Compatible Mobile Phases

High-Throughput Protein Quantitation Using Multiple Reaction Monitoring

Applying MRM Spectrum Mode and Library Searching for Enhanced Reporting Confidence in Routine Pesticide Residue Analysis

Application Note. David A. Weil Zoltan Timar Michael Zumwalt Edgar Naegele. Abstract

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

A Quality by Design (QbD) Based Method Development for the Determination of Impurities in a Peroxide Degraded Sample of Ziprasidone

New Dynamic MRM Mode Improves Data Quality and Triple Quad Quantification in Complex Analyses

Definitive EtG/EtS LC-MS/MS Analysis:

Information Dependent Acquisition (IDA) 1

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Rapid and Sensitive Analysis of Antibiotics in Children s Urine Using the X500R QTOF System

Routine MS Detection for USP Chromatographic Methods

PosterREPRINT COMPARISON OF PEAK PARKING VERSUS AUTOMATED FRACTION ANALYSIS OF A COMPLEX PROTEIN MIXTURE. Introduction

Overview. Introduction

The Power of LC MALDI: Identification of Proteins by LC MALDI MS/MS Using the Applied Biosystems 4700 Proteomics Analyzer with TOF/TOF Optics

OVERVIEW INTRODUCTION. Michael O Leary, Jennifer Gough, Tanya Tollifson Waters Corporation, Milford, MA USA

Quantitation of High Resolution MS Data Using UNIFI: Acquiring and Processing Full Scan or Tof-MRM (Targeted HRMS) Datasets for Quantitative Assays

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

Accurate Mass Analysis of Hydraulic Fracturing Waters

An Effective Workflow for Impurity Analysis Incorporating High Quality HRAM LCMS & MSMS with Intelligent Automated Data Mining

MassHunter TOF/QTOF Users Meeting

Computer-assisted analysis of complex natural product extracts

Profiling of Diferulates (Plant Cell Wall Cross- Linkers) Using Ultrahigh-performance Liquid. Chromatography-Tandem Mass Spectrometry

Analysis of Synthetic Cannabinoids and Metabolites: Adding New Compounds to an Existing LC-MS/MS Method

Ultrafast Analysis of Methadone and Metabolite EDDP in Urine by the Agilent RapidFire High-Throughput Mass Spectrometry System

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

A Sensitive and Reproducible Signature Peptide MRM Based Quantitation Method for Pegylated Interferon α-2b in Serum

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Applications of LC/MS in Qualitative Drug Metabolite Studies

Quadrupole Time-of-Flight Liquid Chromatograph Mass Spectrometer LCMS-9030 C146-E365

Chemical Components Identification of Cistanche Deserticola Using the X500R QTOF System

Accurate Mass Measurement for Intact Proteins using ESI-oa-TOF. Application Note. Donghui Yi and Christine Miller Agilent Technologies

Lirui Qiao, 1 Rob Lewis, 2 Alex Hooper, 2 James Morphet, 2 Xiaojie Tan, 1 Kate Yu 3

Mass Spectrometry. Hyphenated Techniques GC-MS LC-MS and MS-MS

Application Note. Author. Abstract. Pharmaceutical QA/QC. Vinayak A.K Agilent Technologies Bangalore, India

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

penta-hilic UHPLC COLUMNS

Prep 150 LC System: Considerations for Analytical to Preparative Scaling

MassHunter METLIN Metabolite PCD/PCDL Quick Start Guide

Reversed Phase Solvents

Rapid, Reliable Metabolite Ratio Evaluation for MIST Assessments in Drug Discovery and Preclinical Studies

Quantitative Analysis of EtG and EtS in Urine Using FASt ETG and LC-MS/MS

Untargeted Screening on the X500R QTOF

AppNote 2/2013. A High Throughput Automated Sample Preparation and Analysis Workfl ow for Comprehensive Toxicology Urine Screenings using LC/MS/MS

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

Analysis of Illegal Dyes in Food Matrices using Automated Online Sample Preparation with LC/MS

Analysis of Polyphenols in Saffron Petal Extracts with UHPLC/UV/MS

SEAMLESS INTEGRATION OF MASS DETECTION INTO THE UV CHROMATOGRAPHIC WORKFLOW

Transcription:

Confirmation of In Vitro Nefazodone Metabolites using the Superior Fragmentation of the QTRAP 5500 LC/MS/MS System Claire Bramwell-German, Elliott Jones and Daniel Lebre AB SCIEX, Foster City, California and Toronto, Canada Overview Recently, nefazodone (anti-depressant) has been the target compound for metabolite studies due to the quantity and diversity of biotransformations than can occur in the CYP450 enzymes. Several scientific publications have evaluated different workflows and mass spectrometry platforms for this specific application. 1-3 Herein, we report an effective and sensitive method to identify Nefazodone metabolites using the new QTRAP 5500 LC/MS/MS System. The system s significant performance improvement for Quadrupole and Linear Ion Trap sensitivity and scan speed, allow for highly effective detection and confirmation of in vitro and in vivo metabolites. The faster scan rates are a perfect match for small particle, high-pressure LC workflows for better separation and throughput. Using a pmrm approach and a rapid 10 minute chromatographic analysis, a significant number of metabolites were found and confirmed in a single injection (more than any other MS technologies used in previous literature references). This level of identification in a rapid chromatographic analysis demonstrates the abilities of the QTRAP 5500 System for high throughput and a detailed metabolite analysis, with a depth of detection not possible on other MS technologies. Furthermore, the improved sensitivity in LIT mode allows for fine structural confirmation, Figure 1. Example of pmrm transition prediction for a Oxidation and Demethylation transformation. In this case LightSight software v2.1 creates four MRMs which will cover any fragment / transformation possibility. Since MRM has better duty cycle than any other MS method, sensitivity and selectivity are very important in this mode. since a greater number of diagnostic low level fragments are detectable compared to less sensitive accurate mass systems. Introduction Metabolite profiling plays a critical role in the discovery and developmental level of analysis for lead drug candidates. Many aspects of drug activity are affected by metabolism (pharmacokinetics, bioavailability, drug distribution, toxicity and adverse drug reactions). The enhanced scan rate (4X Quadrupole mode, 5X LIT) and sensitivity (8-10X response in Quadrupole mode and 10 to 100X in LIT mode) allow for a number of unique experiments with unprecedented sensitivity. The most effective scan mode for first pass metabolic screening on the QTRAP 5500 system is pmrm (predictive MRM). Figure 1 shows an example of this method of MRM detection in which a comprehensive list of possible biotransformations along with parent drug fragment pattern are used to create a range of theoretically derived MRMs to cover all possibilities. The only major area in which pmrm might miss a metabolite is N- dealkylation and other exotic breaks in the drug core structure. In this case a second injection using a dual precursor and neutral p 1

Figure 2. Three metabolites with corresponding LIT MS/MS spectra on the QTRAP 5500 and 4000 QTRAP Systems. In each case a significant increase in interpretable fragments is exhibited on the QTRAP 5500 System. loss survey based on the parent drugs fragmentation can be employed to pick up any missed metabolites. This second injection would not be required if the molecule is unlikely to undergo an N-dealkylation transformation. The strength of the QTRAP 5500 System is that a cycle of ~300 pmrms or a dual precursor and neutral loss survey with 4 dependent Ion Trap MS/MS can be done approximately 4X faster than current 3200 and 4000 QTRAP system experiments. For example, a dual scan experiment with both a precursor and neutral loss on a standard QTRAP system, with four EPIs running at the fastest scan rate possible on the 4000 QTRAP system could take ~7.5 seconds. Since this cycle rate is too slow for most chromatographic peaks, only a single survey and one or two EPIs could be done on an LCMS time frame. However, as the QTRAP 5500 System can scan 4 to 5X faster, this same cycle would be reduced to around ~2 seconds. A two second cycle rate will allow for 3 points over even a very sharp 6 seconds chromatographic peak. The combined precursor and neutral loss is also particularly effective since few metabolic changes would affect both the neutral and charged parts of the molecule, allowing for broad metabolite detection. In the case of many metabolic forms co-eluting, a greater number of dependent EPIs is possible based on the 5X faster LIT scan rate, which will also allow for more MS/MS spectra in a single injection. Thus the dual scan survey with four EPIs can allow for greater metabolite detection without compromising scan rate. (Dynamic Background Subtraction (DBS) is another way to ensure an efficient ion selection process is made during an automated experiment and doesn t require more than one dependant EPI scan; significantly reducing the cycle time). This optimal mode is only possible on a QTRAP 5500 System. As with dual survey mode, pmrm also benefits by a 4X reduction in scan rate compared to the 4000 QTRAP System, in much the same way. Another differential feature is the LIT MS/MS sensitivity. Compared to the 4000 QTRAP System, an improvement of 10 to 100X in intensity is apparent. This has two main benefits, the first being the detection of metabolites at an unprecedented low level. Often in in vivo PK work, MRMs track metabolites at very low levels. For the first time, the QTRAP 5500 System can now allow qualitative identification at much the same levels. The second benefit is the greater number of fragments for detailed structural identification on the QTRAP 5500 System. The presence of more fragments aids in greater refinement of interpretation. Figure 2 shows an example of the QTRAP 5500 and 4000 QTRAP System in terms of the number of interpretable fragments for the same metabolites. To illustrate these points, a Nefazodone incubation was analyzed on the QTRAP 5500 System and two survey modes were used: 1) pmrm IDA and 2) dual scan precursor and neutral loss IDA. A number of highly refined MS/MS confirmations will be presented to demonstrate the level of refinement possible using the high sensitivity LIT MS/MS of the new QTRAP 5500 system. p 2

Experimental Conditions Incubations Solutions 50 mm Phosphate Buffer (ph 7.4), 5.0 mm of NADPH in phosphate buffer solution, (A) 2.0 mg/ml of rat microsome in phosphate buffer, (B) 20 μm of nefazodone solution in 5.0 mm of NADPH solution (C) Acetronitrile solvent. Control (t = 0 min) - In a 1.8 ml Eppendorf tube the following solutions were added: 200 μl (B), 400 μl (C) and 200 μl (A) - 1.0 min. vortex, left a few minutes in ice, centrifuged for 2.0 min at 10.000 rpm - Final volume 800 μl - Final substrate concentration 5.0 μm + 1.25 mm NADPH + 25 mm phosphate buffer Incubation (t = 60 min) - In a 1.8 ml Eppendorf tube the following solutions were added: 200 μl (B), 200 μl (A) - Incubation for 60 min (37 C Eppendorf apparatus, 350 rpm agitation) - Quench with 400 μl of (C) - 1.0 min. vortex, left a few minutes in ice, centrifuged for 2.0 min at 10,000 rpm - Final volume 800 μl* 20 μl of the control and 60 min incubation solutions were diluted with 200 μl 95% H2O (5.0 mm ammonium formate)/5% ACN (v/v) solution and injected - After dilution in mobile phase A, the final concentration of preincubations solution was 2.5 μm Chromatography (LC) - System: Shimadzu Prominence (Shimadzu Corp, Kyoto, Japan) - Column: Stable Bond C18 (4.6 mm ID x 50 mm) 1.8 μm (Agilent, Palo Alto, CA) - Mobile Phase: A) H2O + 0.1 % Formic Acid and B) Acetonitrile + 0.1 % Formic Acid - Column oven temperature: 45ºC - Flow Rate: 1.00 ml/min - Injection Volume: 10 μl - Gradient Runtime: 10.0 min (see table 1) Table 1. HPLC gradient Mass Spectrometry The new QTRAP 5500 System and LightSight software v2.1 were used for all LCMS experiments. A single injection pmrm approach was employed, triggering the collection of confirmatory MS/MS in an automated experiment (IDA). A second injection using dual scan parent Precursor/Neutral Loss IDA was also used, although no major metabolites were missed in the first injection. In both cases the 20,000 amu/sec scan rate was used with four dependent EPI s (enhanced product ion scan). All methods were generated automatically using the LightSight software v2.1 automatic method builder. Results A number of novel metabolites will be presented that show the identification abilities of the ultra high sensitivity LIT mode on the Figure 3. The results of a sample control comparison in LightSight software v2.1. In this example a low level N-dealkylation, oxidation and di-dehydrogenation is observed. The MS/MS comparison list in the middle right of the figure allows for quick comparison of metabolite vs. parent fragmentation. The QTRAP 5500 Systems LIT MS/MS sensitivity is excellent for this 0.1% level metabolite based on the MRM TIC. Figure 4. The fragment analysis of the same N-dealkylation, oxidation and di-dehydrogenation observed in LightSight software v2.1. ACD Fragmentor overlays structural fragment assignments on top of the MS spectra. The user can then scroll through the list deselecting any incorrect predictions. p 3

QTRAP 5500 System. Initial screening of the metabolites was accomplished by sample and control comparison using LightSight software. Figure 3 shows the XIC trace, MS/MS spectrum and other relevant information in the LightSight software interface for an N-dealkylation, oxidation and didehydrogenation metabolite. This metabolite represents 0.1% of the total TIC area and represents a good example of a low level species. Even for this trace level metabolite, 18 high quality fragments are apparent in the spectrum. LightSight software v2.1 allows for the direct import of the spectra into ACD Spectrum Manager and Fragmentor, for MS/MS automated analysis. Figure 4 shows the fragment assignment accomplished using ACD Fragmentor, in this case 80% of the fragments were assigned using the automated workflow. The metabolically created carbonyl group is likely to be on the 6, 7, 8 or 12 carbon atom. One of the other interesting metabolites, an N-dealkylations plus oxidation and dehydrogenation is found at a retention time of 4.13 minutes. Figure 5 shows the XIC and MS/MS data found by LightSight software. Examination of the MS/MS spectra reveals 19 fragments of high quality. ACD Fragmentor was able to assign 70% of these to the predicted structure with minimal manual review (Figure 6). The metabolically created ketone group is likely to be at positions 21, 20, 11 and 10 on the backbone of the structure. Two novel oxidative de-chlorinations were detected by the pmrm IDA. Figure 7 shows an example of the oxidative dechlorinations found at 4.25 minutes in the LC run. The LIT MS/MS spectrum shows 14 diagnostic fragments at excellent sensitivity. Figures 8 and 9 contain the ACD Fragmentor assignments for this metabolite. Figure 5. N-dealkylation of the chlorine containing ring on the left half of the molecule plus addition of an oxidation and dehydrogenation. The QTRAP 5500 System LIT sensitivity provides superior fragmentation detail. p 4

Figure 6. Fragmentation of an N-dealkylation of the chlorine containing ring on the left half of the molecule plus addition of an oxidation and dehydrogenation. ACD Fragmentor confirmed around 70 percent of the major fragments automatically. The ketone group matches well to positions 21, 20, 10 and 11 on the aliphatic backbone of the molecule. Figure 7. LightSight software v2.1 XIC and MS/MS of oxidative de-chlorination. p 5

Figure 8. ACD Fragmenter assigns the m/z 274 fragment to the right half the molecule. This is consistent with oxidative de-chlorination taking place on the left ring. Position 29, 30 or 31 are most likely sites for the oxidation. Figure 9. ACD Fragmenter assigns the m/z 246 fragment to the right half the molecule. This is consistent with oxidative de-chlorination taking place on the left ring. Position 29, 30 or 31 are most likely sites for the oxidation. Conclusions The QTRAP 5500 System has excellent sensitivity for both detection and confirmation of metabolites The pmrm as well as dual scan Precursor and Neutral Loss IDA methods provide comprehensive detection of in vitro metabolites in a single injection. In most cases, a single injection using pmrm alone is more then adequate for a complete analysis. The 10 to 100X LIT MS/MS sensitivity improvement allows for superior fragmentation not possible on less sensitive Trap or TOF instruments. This greater degree of fragmentation allows for a high level of structural assignment. The improved scan rate 4 or 5X over previous QTRAP Systems allows for greater IDA coverage, even with sharp peaks delivered by the most progressive small particle highpressure chromatography methods. References 1. Peterman, S. M.; Duczak, N. J.; Kalgutkar, A. S.; Lame, M. E. ; Soglia, J. R. J. Am. Soc. Mass Spectrom., 2006; 17, 363. 2. Li, A. C.; Gohdes, M. A.; Shou, W. Z. Rapid Commun. Mass Spectrom. 2007; 21, 1421. 3. Li, A. C.; Shou, W. Z.; Mai, T. T.; Jiang, X.y. Rapid Commun. Mass Spectrom. 2007; 21, 4001. For Research Use Only. Not for use in diagnostic procedures. 2010 AB SCIEX. The trademarks mentioned herein are the property of AB Sciex Pte. Ltd. or their respective owners. AB SCIEX is being used under license. Publication number: P-1035010-01 p 6