Consultancy Report Ref: 9383-R01

Similar documents
MEASUREMENTS OF SOUND ABSORPTION TIMBERCRETE

REPORT ON THE DETERMINATION OF SOUND ABSORPTION COEFFICIENTS OF WOVEN IMAGE ECHO PANEL 24MM TESTED WITH NO AIR GAP MEASURED IN A REVERBERATION ROOM.

REPORT ON THE DETERMINATION OF SOUND ABSORPTION COEFFICIENTS OF WOVEN IMAGE ECHO PANEL 7MM TESTED WITH A 20MM AIR GAP IN A REVERBERATION ROOM.

LABORATORY MEASUREMENTS OF THE SOUND ABSORPTION COEFFICIENTS OF OSCAR EVO-PANELS

Testing Procedure: AS ISO

LABORATORY MEASUREMENTS OF THE SOUND ABSORPTION OF CLASSICTONE 700 ASTM C Type E Mounting. Test Number 2

A L I O N Technical Center. Test Report

Development of a small-scale reverberation room

RIVERBANK ACOUSTICAL LABORATORIES

Technical Report. Project. The Laboratory Determination of The Random Incidence sound Absorption Coefficient of VertiQ Panels.

DELTA Test Report. DANAK TEST Reg. no Measurement of Sound Absorption Coefficient for Kvadrat Soft Cells Wall Panel Type Time

DELTA Test Report. Measurement of Sound Absorption for AqFlex ON, one Sample as a discrete Object. Performed for Flex Acoustics

A L I O N Technical Center. Test Report. Port Coquitlam, British Columbia, Canada

LABORATORY MEASUREMENTS OF THE REDUCTION OF TRANSMITTED IMPACT NOISE PANETTI OVERLAID BY CONCRETE FLOORING

Measurement of Acoustic Properties of light weight concrete SL-Deck

ISO 354 INTERNATIONAL STANDARD. Acoustics Measurement of sound absorption in a reverberation room

Measurement of sound absorption coefficient for Fraster felt SpaceCover

Test Report. RI Acoustic Lab. Measurement of Sound Absorption Coefficient for RockDelta NoiStop Noise Barrier. 20 Feb. 07

THE ACOUSTIC IMPEDANCE MEASUREMNET SYSTEM USING TWO MICROPHONES

DELTA Test Report. Measurement of sound absorption coefficient for 15 mm Fraster felt Plus acoustic panels with mounting depth 45 mm

ETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT ARL-SA6986/ARL-SA6987 Revision 0 ARL

ETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT ARL-SA6266/ARL-SA6274 Revision 0 ARL

ETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT ARL-SA6712/ARL-SA6713 Revision 0 ARL

Acoustic response in non-diffuse rooms

RIVERBANK ACOUSTICAL LABORATORIES

Laboratory for Acoustics

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

LABORATORY MEASUREMENT OF SOUND ABSORPTION OF SONOGLASS SPRAY-ON TREATMENT IN THREE CONFIGURATIONS

ROLLEASE ACMEDA ACOUSTICAL PERFORMANCE TEST REPORT

Acoustic design of lightweight cabin walls for cruise ships

Laboratory for Acoustics. Determination of the sound absorption (reverberation room method) of Airpanel, manufacturer Texdecor

TFI Report Sound Absorption Impact Sound Insulation

ETS-LINDGREN ACOUSTIC RESEARCH LABORATORY OFFICIAL LABORATORY REPORT ARL-SA6956 Revision 1 ARL

TFI Report Sound Absorption Impact Sound Insulation

Laboratory for Acoustics. Determination of the sound absorption (reverberation room method) of curtains, manufacturer Vescom BV

Calcul des coefficients d'absorption alpha sabine

TFI Report Sound Absorption Impact Sound Insulation

Fabric Dazzle 6086 Manufacturer Carnegie Fabrics

Using an ambisonic microphone for measurement of the diffuse state in a reverberant room

REPORT N CR

Renolit stretch ceilings

Curtain fabric type Ace Manufacturer Kvadrat A/S

Fabric REFLECTACOUSTIC

TFI Report Sound Absorption Impact Sound Insulation

Page 1 of 10 pages Appendix, 5 pages Report no.: P Glostrup, October 12th 2012 Project: Claimant:

Fabric Corsica. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M111250/16

Fabric ALEX. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M102794/09

Fabric Sport. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M102794/02

Fabric STREAMER pro. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M111250/04

Roller blind fabric PRIMACOUSTIC R

Fabric VELOS STRATO. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M102794/10

Report. Laboratory for Acoustics

PRODUCT DATA. Sound Intensity Calibrator Type 3541-A. Uses and Features

Fabric XPRESS + PU-foam Manufacturer Gabriel A/S

Fabric Zetacoustic. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M102794/07

Laboratory for Acoustics. Determination of the sound insulation of sliding door constructions, manufactured by Metaflex

REPORT. Client. Test object. Date of testt. Results. Rönnåsgatan 5B Ulricehamn. (7 appendices) content of the

Laboratory for Acoustics

Fabric UDINE. Measurement of sound absorption according to EN ISO 354. Test Report No. M102794/32

Fabric TWIST MELANGE Manufacturer Gabriel A/S

ISO INTERNATIONAL STANDARD

Fabric Medley Manufacturer Gabriel A/S

Fabric Umbria III R. Measurement of sound absorption in a reverberation room according to EN ISO 354. Test Report No. M102794/13

Evaluation of standards for transmission loss tests

Report. P.O. Box 70 NL 5700 AB Deurne The Netherlands Report number: A E-RA Date: 12 februari 2014 TS/TS/JW/A E-RA

Paseo Recoletos, nº Madrid TEST REPORT

An examination of some aspects of the use of ISO 140 Part 4 and 7 in field measurements.

R & D AKUSTIKA * SIA. Test Report of laboratory Sound Insulation measurements Nr. 569 / 2011 AL 8.4. CUSTOMER: AEROC AS (Estonia)

stylegreen Iceland Moss / Reindeer Moss of the company FlowerArt

Product Data. Brüel & Kjær B. Sound Intensity Calibrator Type 3541

MAC Aura XB Acoustic Test Report

EUROVENT 8/12 SOUND TEST METHOD FOR DUCTED FAN COIL UNITS. FCP 2008 testing procedures FCP acoustical test method pag. 1

The information included in the following report presents the results of sound pressure and sound power testing.

Air distribution systems. Linear whirl outlet WL...

inter.noise 2000 The 29th International Congress and Exhibition on Noise Control Engineering August 2000, Nice, FRANCE

Laser scanning vibrometry measurements on a light weight building element

Investigations on real-scale experiments for the measurement of the ISO scattering coefficient in the reverberation room

Report. Laboratory for Acoustics Determination of the sound absorption (reverberation room method) of different wired mats, manufacturer Rockwool

1. Introduction. 2. ISO 3745 and ISO standard comparison

Sound power level measurement in diffuse field for not movable sources or emitting prominent discrete tones

Available online at ScienceDirect. Energy Procedia 78 (2015 ) th International Building Physics Conference, IBPC 2015

SIMPLE APPLICATION OF STI-METHOD IN PREDICTING SPEECH TRANSMISSION IN CLASSROOMS. Jukka Keränen, Petra Larm, Valtteri Hongisto

CERTIFICATE OF CALIBRATION

Some issues in measurement of the random-incidence scattering coefficients in a reverberation room

The diagram below. to the by the. outlet into. calculation. Since TRANSMISSION VIA STRUCTURE. Vibration Via Supports Duct Breakout

HELLA. Linear slot diffuser. Ordering codes. HELLA - Linear diffuser. Linear Diffusers 1 / 18. Example of the ordering code HELLA B-0-0-AN

Yalıtım Kontrol Koruma. Insulation Control Protection

ACOUSTICS 1.01 Pantone 144 U & Pantone 5405 U

series KAT Acoustic air transfer units

Engineering Noise Control

CERTIFICATE OF CALIBRATION

Postprint.

Acoustic test report AU01+N01

Byggakustik Mätning av ljudabsorption i efterklangsrum (ISO 354:2003)

The practical application of G and C 50 in classrooms

1 BAKER HOUSE DINING: LIGHTING DIAGNOSTIC

Uncertainties associated with the use of a sound level meter

Part 2: Methods for special reverberation test rooms

CIRCO. Sound-attenuating transfer unit QUICK FACTS

Study of Wenger Audience Seating Chair Absorption. Ron Freiheit

Transcription:

Consultancy Report Ref: 9383-R01 Submitted to: Prepared by: Mr Tim Simmons The Woolley Shepherd Secret Meadow Weekmoor Milverton Taunton TA4 1QE John Fithyan Laboratory Manager Approved for issue by: Andy Varley (MIOA) Senior Consultant Sound absorption tests on a range of sheep s wool based materials February 2015 9383-R01 ISVR Consulting, University of Southampton, Highfield Campus, Southampton SO17 1BJ, United Kingdom Tel: +44 (0)23 8059 2162 Fax: +44 (0)23 8059 2728 Email: consultancy@isvr.co.uk Web: www.isvr.co.uk

Contents 1. Introduction... 3 2. Measurement Method... 5 3. Results... 6 3.1 Ambient Conditions 6 3.2 Test results 6 Tables 1 to 6: Sound Absorption Coefficients Figures 1 to 3: Photographs of the test samples inside the reverberation chamber Appendix 1 Appendix 2 Instrumentation and Calibration Acoustic Environment ISVR Consulting Report 9383-R01 / February 2015 Contents

1. Introduction ISVR Consulting were engaged by The Woolley Shepherd, to undertake measurements on a range of products to establish their sound absorption coefficient (a S ) or equivalent sound absorption area (A obj m 2 ) depending on the product being tested. The sound absorption measurements were performed in the Reverberant Suite of the Rayleigh Laboratories of the ISVR, University of Southampton. The tests were carried out on the 10 th February 2015. The measurements were carried out in accordance with BS EN ISO 354: 2003 Acoustics Measurement of sound absorption in a reverberation room. When a sound source operates in an enclosed space, the level to which reverberant sound builds up, and the subsequent decay of reverberant sound when the source is stopped, are governed by the sound-absorbing characteristics of the boundary surfaces, the air filling the space, and objects within the space. In general, the fraction of the incident sound power absorbed at a surface depends upon the angle of incidence. In order to relate the reverberation time of an auditorium, office, workshop, etc, to the noise reduction that would be effected by an absorbing treatment, knowledge of the sound-absorbing characteristics of the surfaces, usually in the form of a suitable average over all angles of incidence, is required. Since the distribution of sound waves in typical enclosures includes a wide and largely unpredictable range of angles, a uniform distribution is taken as the basic condition for the purposes of standardization. If, in addition, the sound intensity is independent of the location within the space, the sound distribution is called a diffuse sound field, and the sounds reaching a room surface are said to be at random incidence. The sound field in a properly designed reverberation room closely approximates a diffuse field. Hence, sound absorption measured in a reverberation room closely approximates the sound absorption that would be measured under the basic conditions assumed for standardization. ISVR Consulting Report 9383-R01 / February 2015 Page 3 of 20

BS EN ISO 354: 2003 Acoustics Measurement of sound absorption in a reverberation room specifies a method of measuring the sound absorption coefficient of acoustical materials used as wall or ceiling treatments, or the equivalent sound absorption area of objects, such as furniture, persons or space absorbers, in a reverberation room. The materials tested were described as follows: Standard wall panel Limpets Clouds The standard wall panels are rectangular panels designed to be mounted onto walls. They can be used individually or combined to cover larger areas of wall. The limpets are a bit like a scalloped rectangle and are designed to be fitted flush onto ceilings. They would normally be installed individually but one ceiling might have multiple limpets fitted. The clouds are similar to the limpets but they are designed to hang from ceilings. The materials were tested as either individual (discrete) absorbers or as plane absorbers depending on their application. The results obtained can be used for comparison purposes and for design calculation with respect to room acoustics and noise control. ISVR Consulting Report 9383-R01 / February 2015 Page 4 of 20

2. Measurement Method The averaged reverberation rimes (RT 60 ) in 1/3 octave bands were determined for the empty chamber in the frequency range 100 Hz to 5000 Hz, as described in the standard, using an interrupted broad band noise. Thirty two decays were measured, (eight microphone combinations at each of four sound source locations). The sample of absorbent material was introduced into the chamber and either laid out on the floor or hung from the ceiling in accordance with the guidelines given in the standard. The test arrangement was chosen depending on whether the sample was a discrete absorber or a plane absorber. In the case of the plane absorbers the edges of the sample were covered with metal plate to prevent the edges from absorbing any of the sound waves. Once the sample was positioned in the chamber, the averaged reverberation time spectrum was determined as before. The absorption in the chamber was calculated for the sample and the increase in absorption due to the absorbent material relative to the empty chamber, was found. For the plane absorbers the coefficient of absorption (a S ) in each 1/3 octave band was calculated. The results are shown as a value between 0 and 1 where 1 signifies total absorption and 0 signifies no absorption. NB. In practice the test method can give values greater than 1 which can be confusing. Where a value of greater than 1 is given for a plane absorber it has been caused by edge effects of the test sample. For the discrete absorbers the equivalent sound absorption area (A obj m 2 ) in each 1/3 octave band was calculated. The sound absorption property of discrete objects is quantified by the equivalent sound absorption area expressed as m 2 per object. This is the area of a 100% fictive absorbing surface which would absorb the same amount of sound as the object in question. ISVR Consulting Report 9383-R01 / February 2015 Page 5 of 20

3. Results 3.1 Ambient Conditions Barometric pressure 101.1 kpa Temperature 15.0 C Relative Humidity 44 % 3.2 Test results The results in the tables below are given as either the coefficient of absorption (a S ) or as the equivalent sound absorption area (A obj m 2 ) depending on the type of sample being tested. In the case of the plane absorbers where the results are given as the coefficient of absorption (a S ), BS EN ISO 11654 can be used to express the spectrum of results as a sound absorption class. The classes range from A to E where A is the top class. The standard is not applicable to single items, such as the limpets and clouds measured during these tests. NB. There are other standards, namely ASTM C4423-09a that can be used to express the sound absorption test spectrum as a single number. ASTM C423 uses the term Noise Reduction Coefficient (NRC) although this is being replaced by Sound Absorption Average (SAA). Both the NRC and SAA terms have values between 0 and 1 where 1 indicates total absorption. ASTM C423 test procedures should be used when calculating the NRC or SAA value. In this report we have used BS EN ISO 354 to determine the sound absorption coefficient. The methods in the two standards differ slightly. However for indication purposes I have used the BS EN ISO 354 values to calculate an NRC and SAA value for the materials we tested. Table 1 shows the results with 20 individual wall panels butted together to form a large rectangle. The outer edges of the rectangle were covered to prevent them absorbing more sound. ISVR Consulting Report 9383-R01 / February 2015 Page 6 of 20

Table 2 shows the results with 20 individual wall panels butted together to form a large rectangle. The outer edges of the rectangle were left uncovered and therefore absorbed more sound. Table 3 shows the results when the 20 wall panels were tested as discrete absorbers. The results are given as the equivalent sound absorption area (A obj m 2 ). Table 4 shows the results with 25 individual limpets butted together to form a large sample. The outer edges of the sample were left uncovered and therefore absorbed more sound. Table 5 shows the results when the 25 limpets were tested as discrete absorbers. The results are given as the equivalent sound absorption area (A obj m 2 ). Table 6 shows the results when the 8 clouds were tested as discrete absorbers. The results are given as the equivalent sound absorption area (A obj m 2 ). Figures 1 to 3 show photographs of the samples under test. ISVR Consulting Report 9383-R01 / February 2015 Page 7 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Standard Wall Panel All sample panels butted together. Edges covered Frequency Absorption Hz a s 100 0.28 125 0.30 160 0.42 200 0.52 250 0.75 315 0.90 400 1.00 500 1.01 630 1.01 800 1.03 1000 0.98 1250 1.01 1600 0.99 2000 0.99 2500 0.97 3150 0.97 4000 0.97 5000 0.94 Sound Absorption Coefficient a s Sound absorption coefficient 1.2 1.0 0.8 0.6 0.4 0.2 BS EN ISO 11654 Sound absorption class = A 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Using the ISO 354 test data above to calculate the ASTM C 423-09a descriptors, which is not strictly correct, you would get these values: NRC 0.95 SAA 0.93 Table 1: Standard wall panels butted together with outer edges covered ISVR Consulting Report 9383-R01 / February 2015 Page 8 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Standard Wall Panel All sample panels butted together. Edges left uncovered Frequency Absorption Hz a s 100 0.29 125 0.30 160 0.44 200 0.53 250 0.71 315 0.91 400 1.01 500 1.06 630 1.07 800 1.04 1000 1.04 1250 1.07 1600 1.08 2000 1.08 2500 1.07 3150 1.06 4000 1.08 5000 1.07 Sound Absorption Coefficient a s Sound absorption coefficient 1.2 1.0 0.8 0.6 0.4 0.2 BS EN ISO 11654 Sound absorption class = A 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Using the ISO 354 test data above to calculate the ASTM C 423-09a descriptors, which is not strictly correct, you would get these values: NRC 0.95 SAA 0.97 Table 2: Standard wall panels butted together with outer edges left uncovered ISVR Consulting Report 9383-R01 / February 2015 Page 9 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Standard Wall Panel All sample panels moved apart. Edges left uncovered Frequency Absorption Hz A obj m 2 100 0.19 125 0.25 160 0.38 200 0.47 250 0.62 315 0.77 400 0.83 500 0.89 630 0.92 800 0.91 1000 0.91 1250 0.94 1600 0.93 2000 0.94 2500 0.95 3150 0.96 4000 0.97 5000 0.95 Equivalent sound absorption per wall panel A obj m 2 Sound absorption coefficient 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Table 3: Twenty individual standard wall panels tested as discrete absorbers ISVR Consulting Report 9383-R01 / February 2015 Page 10 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Limpets All samples butted together. Edges left uncovered Frequency Absorption Hz a s 100 0.37 125 0.41 160 0.56 200 0.60 250 0.81 315 0.95 400 0.99 500 1.08 630 1.09 800 1.10 1000 1.07 1250 1.08 1600 1.09 2000 1.10 2500 1.08 3150 1.06 4000 1.06 5000 1.04 Sound Absorption Coefficient a s Sound absorption coefficient 1.2 1.0 0.8 0.6 0.4 0.2 BS EN ISO 11654 Sound absorption class = A 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Using the ISO 354 test data above to calculate the ASTM C 423-09a descriptors, which is not strictly correct, you would get these values: NRC 1.00 SAA 1.00 Table 4: Limpets butted together with outer edges left uncovered ISVR Consulting Report 9383-R01 / February 2015 Page 11 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Limpets All samples moved apart. Edges left uncovered Frequency Absorption Hz A obj m 2 100 0.23 125 0.26 160 0.36 200 0.42 250 0.54 315 0.64 400 0.69 500 0.75 630 0.75 800 0.76 1000 0.75 1250 0.77 1600 0.77 2000 0.76 2500 0.76 3150 0.76 4000 0.76 5000 0.72 Equivalent sound absorption per limpet A obj m 2 Sound absorption coefficient 1.0 0.9 0.8 0.7 0.6 0.5 0.4 0.3 0.2 0.1 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Table 5: Twenty-five individual limpets tested as discrete absorbers ISVR Consulting Report 9383-R01 / February 2015 Page 12 of 20

Airborne sound absorption measured in a reverberation chamber. BS EN ISO 354:2003 Customer: Woolly Shepherd Construction tested Clouds Frequency Absorption Hz A obj m 2 100 0.22 125 0.28 160 0.42 200 0.44 250 0.73 315 1.01 400 1.18 500 1.34 630 1.43 800 1.50 1000 1.50 1250 1.52 1600 1.49 2000 1.46 2500 1.43 3150 1.42 4000 1.36 5000 1.29 Equivalent sound absorption per cloud A obj m 2 Sound absorption coefficient 1.6 1.4 1.2 1.0 0.8 0.6 0.4 0.2 0.0 100 1000 10000 Frequency, Hz Test Standard: BS EN ISO 354 Project No: 9383 University of Southampton Tested by: J.Fithyan Highfield, Southampton Test date: 27/01/2015 Table 6: Eight individual clouds tested as discrete absorbers ISVR Consulting Report 9383-R01 / February 2015 Page 13 of 20

Figure 1: Standard wall panels ISVR Consulting Report 9383-R01 / February 2015 Page 14 of 20

Figure 2: Limpets ISVR Consulting Report 9383-R01 / February 2015 Page 15 of 20

Figure 3: Clouds ISVR Consulting Report 9383-R01 / February 2015 Page 16 of 20

Appendix 1: Instrumentation and Calibration 1.1 Measuring Equipment 1.1.1 Noise Generation Instrument Manufacturer Type Ser No Cal Due Amplifier Cambridge Audio A1 V2.0 1201-1711 Not required Loudspeaker - - - Not required 1.1.2 Noise Measurement Instrument Manufacturer Type Ser No Cal Due Microphone Brüel & Kjær 4165 1297127 August 15 Microphone Brüel & Kjær 4165 1297134 August 15 Microphone Brüel & Kjær 4165 1651334 August 15 Microphone Brüel & Kjær 4189 2566078 August 15 Microphone Brüel & Kjær 4189 2573634 August 15 Microphone Brüel & Kjær 4189 2573635 August 15 Pre-amplifier Brüel & Kjær 2669 2552964 September 15 Pre-amplifier Brüel & Kjær 2669 2552965 September 15 Pre-amplifier Brüel & Kjær 2669 2549629 September 15 Pre-amplifier Brüel & Kjær 2669 2572332 September 15 Pre-amplifier Brüel & Kjær 2669 2572333 September 15 Pre-amplifier Brüel & Kjær 2669 2572334 September 15 Freq Analyser Brüel & Kjær 3560C 2447709 October 15 Freq Analyser Brüel & Kjær 3560B 2609169 October 15 Mic Calibrator Brüel & Kjær 4231 2594478 September 15 1.2 System Calibration The noise measuring system was calibrated by applying the microphone calibrator (type 4231) to the transducer and adjusting the analyzer to the reference level at the beginning of the measurement session. The level was also checked at the end of the session to ensure that no drift or fault had occurred. The above equipment is calibrated against the transfer standards below. ISVR Consulting Report 9383-R01 / February 2015 Page 17 of 20

1.3 Traceability 1.3.1 Microphones and Calibrators Microphones and Calibrators ISVR Consulting hold two transfer standard Microphones, type 4145, serial numbers 375091 and 375617. The most recent calibration was carried out by Campbell Associates. This is fully documented in Certificates 17840 and 17841, dated 22 nd January 2015. A transfer standard Pistonphone, type 4220, serial number 1297434 and a transfer standard calibrator, type 4231, serial number 2162524 are also held. These were calibrated by Campbell Associates and are documented in certificate numbers 17844 and 17842, dated 22 nd January 2015. 1.3.2 Additional Instrumentation A Digital Voltmeter (Fluke type 8050A) and a Frequency Counter (Marconi type 2430A) are used with these Transfer Standards to calibrate the above equipment. Both instruments were calibrated by Southern Calibration Laboratories and carry certificates numbered 14111199 and 14111200 respectively and both are dated 25th November 2014. The instrumentation complies with the requirements for a type 1 instrument, as specified in BS EN 61672:2003, BS EN 60942:2003 and BS EN IEC 61260:1996. The Standards are traceable to the National Physical Laboratory, Teddington, England. ISVR Consulting Report 9383-R01 / February 2015 Page 18 of 20

Appendix 2: Acoustic Environment 2.1 Large Reverberation Chamber 2.1.1 Construction The chamber is constructed of reinforced concrete and is separated from the foundations and neighbouring walls by rubber vibration isolators. It is designed with an inclined ceiling and non-parallel walls to ensure a uniform distribution (with frequency) of the normal acoustic modes of the room. 2.1.2 Dimensions Mean edge lengths 9.15 m 6.25 m 6.10 m - height Volume 348 m 3 Surface Area 302 m 2 2.1.3 Walls and Ceiling All inside surfaces of the chamber are finished with a hard gloss paint to give a high reflection coefficient. The walls are 305 mm thick, and the ceiling, which is 460 mm thick, includes two removable sections (1.75 m x 0.86 m) which provide access for a chain hoist capable of carrying loads up to 2000 Kg, and an entry for a 4000 Watt siren driven horn. Connections to the equipment in the chamber may be made via any of five cable ports in the walls. A glazed window (305 mm x 305 mm) permits visual observation from the control area. 2.1.4 Diffusers Ten diffusers are hug from the ceiling of the chamber. These diffusing elements are plywood sheets painted with gloss paint so that they have a low sound absorption. The diffusers, which are of varying sizes are orientated at random and positioned at different heights within the chamber. The total surface area of the diffusers is 43.2 m 2. ISVR Consulting Report 9383-R01 / February 2015 Page 19 of 20

2.1.5 Floor Area The floor is 305 mm thick and has a steel vibration isolated pad (2.1 m x 3.6 m) set into it. This pad may be used for mounting test rigs or vibrators without structural vibrations being transmitted to the chamber. 2.1.6 Doors One set of double doors connects the chamber and the corridor, another set opens into the small reverberation chamber and incorporates removable panels (1.07 m x 1.07 m) for transmission loss measurements. The doorway (2.4 m x 2.0 m) may be used for testing larger panels. The doors (2.56 m x 2.26 m x 130 mm thick) are a sandwich construction of wood wool, wood and steel, and have an average transmission loss in excess of 50 db. 2.1.7 Ventilation and Lighting At each corner of the floor there is an air inlet vent, and there are four outlet vents situated high up on one wall. With all vents open the air is changed at a rate of 100 m 3 per minute. When not required the vents are covered by steel plates and these have diagonal stiffeners to reduce panel vibrations. The chamber is lit by six sodium discharge lights mounted on the wall at 3 m above the floor. ISVR Consulting Report 9383-R01 / February 2015 Page 20 of 20