Determination of Inorganic Anions in Drinking Water by Ion Chromatography

Similar documents
Determination of Inorganic Anions in Wastewater by Ion Chromatography

High-Pressure Electrolytic Carbonate Eluent Generation Devices and Their Applications in Ion Chromatography Systems

Thermo Scientific. Anion-Exchange Column. Determination of Inorganic Anions in Diverse Sample Matrices. Superior Chromatographic Performance

Determination of Tetrafluoroborate, Perchlorate, and Hexafluorophosphate in a Simulated Electrolyte Sample from Lithium Ion Battery Production

New On-Line High-Pressure Electrolytic Eluent Generators for Ion Chromatography

Dionex IonPac AS28-Fast-4µm column

Determination of trace anions in concentrated hydrofluoric acid

Keywords Haloacetic acids, Water analysis, 2-D ion chromatography, ICS-3000

Thermo Scientific Dionex Ion Chromatography Solutions. Global water safety. bromate analysis in drinking water

Thermo Scientific. Anion-Exchange Column

Using a Reagent-Free ion chromatography system to monitor trace anion contamination in the extracts of electronic components

Determinations of Inorganic Anions and Organic Acids in Beverages Using Suppressed Conductivity and Charge Detection

Determination of Cations and Amines in Hydrogen Peroxide by Ion Chromatography Using a RFIC (Reagent-Free) System

Thermo Scientific Dionex IonPac AS23 Anion-Exchange Column

columns IonPac AS17-C Anion Exchange Column

Trace Level Determination of Bromate in Ozonated Drinking Water Using Ion Chromatography

Determination of Common Inorganic Anions in Environmental Waters Using a Hydroxide Selective Column

columns IonPac AS22 Anion-Exchange Column

Determination of Bromate in Bottled Mineral Water Using the CRD 300 Carbonate Removal Device

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

columns IonPac AS18 Anion-Exchange Columns

Time Savings and Improved Reproducibility of Nitrate and Nitrite Ion Chromatography Determination in Milk Samples

anion-exchange column

Inline sample preparation for the determination of anions in sodium hydroxide

Determination of Tartaric Acid in Tolterodine Tartrate Drug Products by IC with Suppressed Conductivity Detection

columns IonPac SCS 1 Silica Cation Separator

Quantification of Trace and Major Anions in Water by Ion Chromatography in a High-Throughput Laboratory

New Developments in Capillary Ion Chromatography using 4 μm Columns and Charge Detection

Thermo Scientific Dionex IonPac AS24A Anion-Exchange Column

Separation of heat stable amine salts in methyldiethanolamine (MDEA) solutions using high-pressure IC

New Stationary Phases for Solid-Phase Extraction. Pranathi R. Perati, Rosanne Slingsby, and Carl Fisher Thermo Fisher Scientific, Sunnyvale, CA, USA

Determination of Trace Cations in Power Plant Waters Containing Morpholine

Thermo Scientific Dionex IonPac AS26 Anion-Exchange Column

Utility of the Charge Detector in Ion Chromatography Applications

Determination of an anionic fluorochemical surfactant in a semiconductor etch bath

Determination of Trace Anions in Organic Solvents

Thermo Scientific Dionex IonPac AS25 Anion-Exchange Column. Minutes

Insights Into the Nanoworld Analysis of Nanoparticles with ICP-MS

Efficient Separations Obtained by Using Smaller Particle Size Analytical Columns with a High-Pressure Ion Chromatography System

Speciation of Bromine Compounds in Ozonated Drinking Water using Ion Chromatography and Inductively Coupled Plasma Mass Spectrometry

Determination of trace anions in high-purity waters by high volume/direct injection ion chromatography

Application Note 187. Brian DeBorba and Jeff Rohrer Thermo Scientific, Sunnyvale, CA, USA

Determination of Methylamine in Drug Products

A Study of Stability, Robustness and Time Efficiency of a New HPLC and a New Tandem MS

HILIC Method Development in a Few Simple Steps

Analyzing Residual Solvents in Pharmaceutical Products Using GC Headspace with Valve-and-Loop Sampling

Determination of Inorganic Cations and Ammonium in Environmental Waters by Ion Chromatography Using the IonPac CS16 Column

Exploring the Benefits of Automated Unattended Sample Derivatization Prior to Gas Chromatography Analysis

Expanded capabilities in ammonium and amine detection

Determination of Sulfate and Chloride in Ethanol by Ion Chromatography

Determination of trace concentrations of oxyhalides and bromide in municipal and bottled waters

Active Flow Technology Understanding How the Flow Rate Profile Affects the Chromatographic Efficiency

Superior Chromatographic Performance Recommended anion-exchange column for separation of haloacetic acids prior to MS or MS/MS detection

The determination of trace anions in concentrated phosphoric acid

Thermo Scientific Dionex IonPac AS11-HC-4µm Anion-Exchange Column

Determination of Hexavalent Chromium in Drinking Water Using Ion Chromatography

Determination of urea in ultrapure water by IC-MS/MS

Characterization of Polymers and Plastics (pellets, powders and films) by the Thermo Scientific FLASH 2000 Elemental Analyzer

Increasing Speed of UHPLC-MS Analysis Using Single-stage Orbitrap Mass Spectrometer

The Determination of Trace Anions in Concentrated Phosphoric Acid

ProPac WCX-10 Columns

A Strategy for an Unknown Screening Approach on Environmental Samples using HRAM Mass Spectrometry

[ Care and Use Manual ]

Accurate and Precise Automated Dilution and In-line Conductivity Measurement Using the AS-AP Autosampler Prior to Analysis by Ion Chromatography

Method Abstract. Flow Solution Fluoride, USGS by ISE and FIA

Application of Eluent Generation for Trace Anion Analysis of Borated Waters

Determination of Perchlorate in High Ionic Strength Fertilizer Extracts by Ion Chromatography

Determination of Benzo(a)pyrene in Sausage and Preserved Ham

Standard Operating Procedure for the Analysis of Chloride, Bromide and Sulfate in Fresh Waters by Ion Chromatography CCAL 50B.2

Determination of Polyacrylic Acid in Boiler Water Using Size-Exclusion Chromatography with Charged Aerosol Detection

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Determination of trace anions in high-nitrate matrices by ion chromatography

Determination of Monochloroacetic Acid in Carbocisteine

Direct Determination of Small Organic Acids in Sea Water by IC-MS

Evaluation of a New HPLC, a New Tandem MS and a New Data Processing Software for General Clinical Use

Determination of BTEX in Cigarette Filter Fibers Using GC-MS with Automated Calibration

Waters. Ion Chromatography

Quantitation of Surfactants in Samples by High Performance Liquid Chromatography and Corona Charged Aerosol Detection

Determination of Chloride, Fluoride, Bromide, Nitrate, Sulphate and Phosphate in Water Sample...

of mass spectrometry

Simultaneous Determination of Aromatic Amines and Pyridines in Soil

Monitoring Protein PEGylation with Ion Exchange Chromatography

Frank Steiner, Michael Heidorn, and Markus M. Martin Thermo Fisher Scientific, Germering, Germany

Quantitation and Characterization of Copper Plating Bath Additives by Liquid Chromatography with Charged Aerosol Detection

LC-MS/MS Method for the Determination of Diclofenac in Human Plasma

Simultaneous, Fast Analysis of Melamine and Analogues in Pharmaceutical Components Using Q Exactive - Benchtop Orbitrap LC-MS/MS

Exploring Mixed-Mode Chromatography Column Chemistry, Properties, and Applications

A Platform Method for Pharmaceutical Counterion Analysis by HPLC

Thermo Scientific ELEMENT GD PLUS Glow Discharge Mass Spectrometer. Defining quality standards for the analysis of solid samples

Advances in Sample Preparation for Accelerated Solvent Extraction

Hydrophilic Interaction Liquid Chromatography: Method Development Approaches

Thermo Scientific Dionex IonPac CS17 Cation-Exchange Column

Determination of trace anions in concentrated glycolic acid

Plasma-free Metanephrines Quantitation with Automated Online Sample Preparation and a Liquid Chromatography-Tandem Mass Spectrometry Method

Overcoming Challenging Matrices in Ion Chromatography

Determination of silicate in high-purity water using ion chromatography and online sample preparation

California Environmental Protection Agency Air Resources Board

Comprehensive Neurochemical Profiling of Brain Tissue Samples

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Complementary Use of Raman and FT-IR Imaging for the Analysis of Multi-Layer Polymer Composites

Transcription:

Determination of Inorganic s in Drinking Water by Ion Chromatography Peter Jackson Thermo Fisher Scientific Inc. Application Note Introduction The determination of common inorganic anions in drinking water is one of the most important applications of ion chromatography (IC) worldwide. The National Primary Drinking Water Standards in the United States specify a Maximum Contaminant Level (MCL) for a number of inorganic anions, including fluoride, nitrite, and nitrate. The MCLs are specified to minimize potential health effects arising from the ingestion of these anions in drinking water. High levels of fluoride cause skeletal and dental fluorosis, and nitrite and nitrate can cause methemoglobulinemia, which can be fatal to infants. Other common anions, such as chloride and sulfate, are considered secondary contaminants. The National Secondary Drinking Water Standards in the U.S. are guidelines regarding taste, odor, color, and certain aesthetic characteristics. Although these guidelines are not federally enforced, they are recommended to all states as reasonable goals and many states adopt their own regulations governing these contaminants. 2 Ion chromatography has been approved for compliance monitoring of these common inorganic anions in U.S. drinking water since the mid-98s, as described in U.S. EPA Method.. Many other industrialized countries have similar health and environmental standards and a considerable number of regulatory IC methods have been published worldwide (e.g., in Germany, France, Italy, and Japan) for the analysis of anions in drinking water. In addition, many standards organizations (including ISO, ASTM, and AWWA) have validated IC methods for the analysis of inorganic anions in drinking water., Equipment Thermo Scientific Dionex DX-2 and Thermo Scientific Dionex DX- Ion Chromatography Systems were used for this work. The Dionex DX-2 is a dedicated ion chromatograph; the Dionex DX- is a modular system, which in this case consisted of: Thermo Scientific Dionex GP Gradient Pump Thermo Scientific Dionex CD2 Conductivity Detector Thermo Scientific Dionex LC2 Chromatography Enclosure with rear-loading injection valve Thermo Scientific Dionex AS Automated Samplers ( ml vials) and a Thermo Scientific Dionex PeakNet Chromatography Workstations were used with both systems This application note describes the determination of inorganic anions in drinking water and other environmental waters using conditions that are consistent with those in U.S. EPA Method.. The use of an optional column, the Thermo Scientific Dionex IonPac AS column, is also discussed.

2 Reagents and Standards Deionized water, Type I reagent grade, 8 MΩ-cm resistance or better.8 M Sodium carbonate/.7 M Sodium bicarbonate: Dionex IonPac ASA Eluent Concentrate. M Sodium carbonate/. M Sodium bicarbonate: Dionex IonPac AS Eluent Concentrate Sodium and Potassium salts, ACS reagent grade, for preparing anion standards Fluoride standard mg/l, ml Chloride standard mg/l, ml Sulfate standard mg/l, ml Nitrate standard mg/l, ml Phosphate standard mg/l, ml Bromide standard mg/l, ml Conditions Part A Columns: Eluent: Run Time: Flow Rate: Injection Volume: Detection: System Backpressure: Dionex IonPac AGA-SC, mm Dionex IonPac ASA-SC, 2 mm.8 mm Sodium carbonate/.7 mm Sodium bicarbonate <8 min 2. ml/min μl Suppressed conductivity, Thermo Scientific Dionex ASRS ULTRA Self-Regenerating suppressor ( mm), recycle mode, ma current ~ psi Background Conductance: ~ μs Part B Columns: Eluent: Run Time: Flow Rate: Injection Volume: Detection: System Backpressure: Dionex IonPac AG, mm Dionex IonPac AS, 2 mm. mm Sodium carbonate/. mm Sodium bicarbonate < min.2 ml/min μl Suppressed conductivity, Dionex ASRS ULTRA suppressor ( mm), recycle mode, ma current ~ psi Background Conductance: ~7 μs Preparation of Solutions and Reagents Stock Standard Solutions Stock Standard Solutions ( mg/l) For several of the analytes of interest, mg/l standard solutions are available from Thermo Scientific and other commercial sources. When commercial standards are not available, mg/l standards can be prepared by dissolving the appropriate amounts of the corresponding mass in ml of deionized water according to Table. Standards are stable for at least one month when stored at C. Table. Masses of compounds used to prepare L of mg/l anion standards. Compound Mass (g) Fluoride Sodium fluoride (NaF) 2.2 Chloride Sodium chloride (NaCl).8 Nitrite Sodium nitrite (NaNO 2 ).99 Bromide Sodium bromide (NaBr).288 Nitrate Sodium nitrate (NaNO ).7 o-phosphate Potassium phosphate monobasic (KH 2 PO ) Working Standard Solutions. Sulfate Sodium sulfate (Na 2 SO ).79 Composite working standards at lower analyte concentrations are prepared from the mg/l standards described above. Working standards containing less than mg/l anions should be prepared daily. Table 2 shows the linear concentration range investigated for each anion, as well as the concentration of the standard used to calculate the method detection limits (MDLs) and the concentration of the quality control sample (QCS) used to determine the retention time and peak area precision. Table 2. Concentration of linearity, MDL, and reproducibility standards. Seven-Point Calibration Range MDL Calculation Standard QCS Standard for RSD Calculation Fluoride..2 2 Chloride.2 2. 2 Nitrite..2 2 Bromide.. 2 Nitrate.. o-phosphate.. 2 Sulfate.2 2.

Eluent Solutions Dilute 2 ml of the appropriate eluent concentrate to 2. L with deionized water. Transfer to a 2 L eluent container and pressurize the container with helium at 8 psi. Sample Preparation The vial caps in the Thermo Scientific Dionex AS Automated Sampler contain a 2 μm filter, so no additional filtration was used in conjunction with this mode of sample introduction. If a Dionex AS is not used, filter all samples through appropriate. μm syringe filters, discarding the first μl of the effluent. The domestic wastewater sample was treated with a C 8 Sep-Pak cartridge to remove hydrophobic organic material in order to prolong column lifetimes. The C 8 cartridge was preconditioned with ml of methanol, followed by ml of deionized water. The sample ( ml) was then passed through the cartridge, with the first ml of the effluent being discarded. Aqueous soil extracts were prepared by the extraction of. g of soil in ml of deionized water in an ultrasonic bath for min, followed by filtration with a. μm filter. Figure shows a typical chromatogram of a standard containing low-ppm levels of common inorganic anions separated using an Dionex IonPac ASA-SC column as described in Part A of the Conditions section. The Dionex IonPac ASA-SC column has selectivity similar to that of the Dionex IonPac ASA column, is ethylvinylbenzene (EVB) crosslinked with % divinylbenzene (DVB), which makes the column % solvent-compatible. All the anions are well-resolved within a total run time of less than 8 min. The method linearity was determined for the inorganic anions over a seven-point calibration range. MDLs were calculated for each of the anions according to the procedure described in U.S. EPA Method.. The MDLs are estimated by injecting seven replicates of reagent water fortified at a concentration of to times the estimated instrument detection limit. The MDL is then calculated as (t) (SD) where t = Student s t value for a 99% confidence level and a standard deviation estimate with n degrees of freedom (t =. for seven replicates) and SD = standard deviation of the replicate analysis. Results and Discussion A variety of methods have been used for the analysis of inorganic anions, including traditional spectroscopic techniques such as colorimetry; wet chemical methods such as gravimetric analysis, turbidimetry, and titrimetry; and electrochemical techniques such as ion selective electrodes (ISE) and amperometric titrations. However, many of these methods are not specific and suffer from interferences or limited sensitivity, can be labor-intensive, and are often difficult to automate. Because many of the individual test procedures described above can be replaced by one chromatographic separation, IC was quickly accepted by regulatory agencies worldwide for the determination of anions in drinking water and other environmental waters. Column: Dionex IonPac AGA-SC, Dionex IonPac ASA-SC Eluent:.7 mm Sodium bicarbonate/.8 mm Sodium carbonate Flow Rate: 2. ml/min Peaks:. Fluoride 2 mg/l (ppm) 2. Chloride. Nitrite. Bromide. Nitrate. o-phosphate 7. Sulfate 2 7 Dionex IonPac ASA-SC Column U.S. EPA Method. (A) specifies the use of an Dionex IonPac ASA anion-exchange column with an eluent of.8 mm sodium carbonate/.7 mm sodium bicarbonate for the separation of common anions. The method specifies the use of an Thermo Scientific Dionex AMMS MicroMembrane suppressor operated in the chemical regeneration mode; however, an Thermo Scientific Dionex ASRS Self-Regenerating suppressor provides equivalent method performance with added convenience. Conductivity is used as a bulk property detector for the measurement of inorganic anions. 2 8 7 Figure. Separation of a low-ppm inorganic anion standard using a Dionex IonPac ASA-SC column.

Range Table shows the concentration ranges investigated, the resulting linear coefficients of determination (r 2 ), and the calculated MDL for each anion. The retention time and peak area precision (expressed as % RSD) were determined from seven replicate injections of a quality control standard, as described in Table 2. Table also shows typical retention time and peak area precision data that can be obtained for inorganic anions using the Dionex IonPac ASA-SC column with a Dionex DX-2 system. Table. Linearity, MDL, retention time, and peak area precision obtained using the Dionex IonPac ASA-SC column a. Linearity (r 2 ) Calculated MDL b (µg/l) Retention Time (% RSD c ) Area (% RSD) Fluoride..997.9.8%.7% Chloride.2 2.999 2..%.7% Nitrite..9997.7 (.8 as NO 2 -N) <.%.% Bromide..997 9.7 <.%.% Nitrate..999 o-phosphate..997.2 (. as NO -N) 7.8 (.8 as PO -P).7%.7%.%.% Sulfate.2 2.997.7 <.%.% a Dionex DX-2 system b MDL = s*t s,99 where t s,99 =. for n = 7 c RSD = Relative standard deviation, n = 7 The performance of methods used for environmental analysis are typically validated through single- and multi-operator precision and bias studies on spiked samples. Table shows typical recovery results for single-operator data obtained using the Dionex IonPac ASA-SC column for common anions spiked into drinking water, raw (unprocessed) drinking water, and other environmental water matrices. In this instance, the samples were spiked with the analytes at approximately the same levels as specified in U.S. EPA Method.. Figure 2 shows a typical chromatogram of inorganic anions in Sunnyvale, California drinking water obtained using the Dionex IonPac ASA-SC column. Figure shows a chromatogram of inorganic anions in surface water obtained from a lake in Utah. In general, Table shows that acceptable recovery data (i.e., 8 2%) was obtained for the inorganic anions in most matrices. The one exception was the domestic wastewater sample, where fluoride gave a recovery of <% under these conditions. Fluoride concentrations of <. mg/l are subject to interference from mg/l levels of small organic acids, such as formate and acetate, when using the Dionex IonPac ASA column. In this sample, the recoveries for nitrite and nitrate were also not as expected. At the time of these analyses, the sample had been stored (at C) for longer that the recommended holding time for nitrite/nitrate of 2 days. In this case, the unexpected recoveries were due to the presence of nitrifying/denitrifying microbes in the sample rather than any chromatographic resolution problems. Table. recoveries for spiked water samples obtained using the Dionex IonPac ASA-SC column. Drinking Water Raw Water Surface Water Fluoride 9.9 9. 9. Chloride 97. 2 8.2 8. Nitrite 2 9. 2 2.. Bromide 2 98.7 2 9.7 2. Nitrate 92. 9. 9.7 o-phosphate 9. 9. 7.9 Sulfate 2 97..8. Domestic Industrial Soil Extract Fluoride 7. 88. 2 99. Chloride 2 82.7 2.8.2 Nitrite 2 27.* 2 98. 2 2. Bromide 2 8. 2 92. 2 9. Nitrate.8* 9.2 9.2 o-phosphate 2. 2 98.8 2.7 Sulfate 9..9 2 9. * Sample stored for longer than recommended holding time; inappropriate recovery due to microbial action. 2 Column: Dionex IonPac AGA-SC, Dionex IonPac ASA-SC Eluent:.7 mm Sodium bicarbonate/.8 mm Sodium carbonate Flow Rate: 2. ml/min Peaks:. Fluoride. mg/l (ppm) 2. Chloride 9.. Nitrate 2.. o-phosphate.. Sulfate. 2 8 Figure 2. Determination of inorganic anions in drinking water using a Dionex IonPac ASA-SC column. 79

Column: Dionex IonPac AGA-SC, Dionex IonPac ASA-SC Eluent:.7 mm Sodium bicarbonate/.8 mm Sodium carbonate Flow Rate: 2. ml/min Peaks:. Fluoride.2 mg/l (ppm) 2 2. Chloride 2. Bromide.. Nitrate.8. o-phosphate.7. Sulfate 8.7 Column: Dionex IonPac AG, Dionex IonPac AS Eluent:. mm Sodium bicarbonate/. mm Sodium carbonate Flow Rate:.2 ml/min Peaks:. Fluoride 2 mg/l (ppm) 2. Chloride. Nitrite. Bromide. Nitrate. o-phosphate 7. Sulfate 7 2 2 8 Figure. Determination of anions in surface water using a Dionex IonPac ASA-SC column. Dionex IonPac AS Column Although U.S. EPA Method. specifies the use of an Dionex IonPac ASA column, Section.2.2. states that an optional column may be used if comparable resolution of peaks is obtained and the quality control requirements of Section 9.2 can be met. The Dionex IonPac AS column is packed with a grafted, methacrylate-based anion-exchange resin. In contrast, the Dionex IonPac ASA-SC column is packed with a functionalized latex on the surface of an EVB/DVB-based material. The Dionex IonPac AS column provides complete resolution of fluoride from formate and/or acetate, in addition to improved resolution of fluoride from the void peak. The improved selectivity and higher capacity of the Dionex IonPac AS column ( μeq/column compared to 2 μeq/column for the Dionex IonPac ASA column) also allows improved resolution of chloride and nitrite, which is important in environmental water analysis. One drawback of using the grafted, higher capacity Dionex IonPac AS column is lower peak efficiencies than those obtained using the latex-agglomerated Dionex IonPac ASA-SC column. Figure shows a typical chromatogram of a standard containing low-ppm levels of common anions separated using the Dionex IonPac AS column as described in Part B of the Conditions section. Fluoride is clearly resolved from the void volume and the overall selectivity is improved compared to the chromatogram shown in Figure, although the total run time is increased to min. 2 8 2 Figure. Separation of a low-ppm inorganic anion standard using a Dionex IonPac AS column. The method linearity using the Dionex IonPac AS column was again determined over a seven-point calibration range and the MDLs were calculated according to U.S. EPA Method.. Table shows the concentration ranges investigated, the resulting linear coefficients of determination (r 2 ), and typical calculated MDLs for each anion. The retention time and peak area precision (expressed as % RSD) were determined from seven replicate injections of the quality control sample. Table also shows typical retention time and peak area precision data that can be obtained for inorganic anions using the Dionex IonPac AS column with a Dionex DX- system. Table. Linearity, MDL, retention time, and peak area precision obtained using the Dionex IonPac AS column a. Range Linearity (r 2 ) Calculated MDL b (µg/l) 8 Retention Time (% RSD c ) Area (% RSD) Fluoride..998..2%.7% Chloride.2 2.999 2.9.%.% Nitrite..9997. (2. as NO 2 -N).%.7% Bromide..997 7.8.%.% Nitrate..997 o-phosphate..99 7.7 (.7 as NO -N).%.% 2.2 (. as PO -P).%.7% Sulfate.2 2.997 8.2.%.9% a Dionex DX- system b MDL = s*t s,99 where t s,99 =. for n = 7 c RSD = Relative standard deviation, n = 7

The performance of the Dionex IonPac AS method was also validated through single-operator precision and bias studies on spiked samples. Table shows typical recovery results for single-operator data obtained using the Dionex IonPac AS column for common anions spiked into drinking water, raw drinking water, and other environmental water matrices. Figure shows a chromatogram of inorganic anions in the same Sunnyvale, California drinking water obtained using the Dionex IonPac AS column, and Figure shows a chromatogram of inorganic anions in raw water. Table shows that acceptable recovery data (i.e., 8 2%) was obtained for the inorganic anions in all matrices when using the Dionex AS column, with the exception of nitrite and nitrate in the domestic wastewater sample. However, as discussed previously, this was due to the presence of nitrifying/denitrifying microbes rather than any chromatographic resolution problems. Table. recoveries for spiked water samples obtained using the Dionex IonPac AS column. Drinking Water Raw Water Surface Water Fluoride 9. 8.. Chloride 9. 2 8. 8. Nitrite 2. 2 92..2 Bromide 2 9. 2 9. 2 9. Nitrate 87.2 89. 9.2 o-phosphate 9.8 9.2. Sulfate 2 9... Domestic Industrial Soil Extract Fluoride 9.8 9. 2. Chloride 2 87. 2 9.7 9.7 Nitrite 2.* 2 98.2 2 89. Bromide 2 9.8 2 9.2 2 89.9 Nitrate.* 9. 92.8 o-phosphate 2 9. 2 9.9 2. Sulfate 9. 2. 2 9.7 Column: Dionex IonPac AG, Dionex IonPac AS Eluent:. mm Sodium bicarbonate/. mm Sodium carbonate Flow Rate:.2 ml/min Peaks:. Fluoride. mg/l (ppm) 2. Chloride 9.7. Nitrate 2. 2. o-phosphate.27. Sulfate 2 8 2 Figure. Determination of inorganic anions in drinking water using a Dionex IonPac AS column. Figure. Determination of anions in raw water using a Dionex IonPac AS column. Column: Dionex IonPac AG, Dionex IonPac AS Eluent:. mm Sodium bicarbonate/. mm Sodium carbonate Flow Rate:.2 ml/min Peaks:. Fluoride.2 mg/l (ppm) 2. Chloride 79.9. Bromide.. Nitrate.2 2. o-phosphate.8 2 8 2 2 * Sample stored for longer than recommended holding time; inappropriate recovery due to microbial action.

Conclusion Both the Dionex IonPac ASA-SC and Dionex IonPac AS columns provide suitable performance for the determination of inorganic anions in drinking waters as outlined in U.S. EPA Method.. The Dionex IonPac ASA-SC column is recommended for the rapid analysis of anions in low-ionic-strength, well-characterized samples such as drinking, raw, and surface water. TheDionex IonPac AS column provides improved fluoride resolution from the system void peak and complete resolution of fluoride from formate and/or acetate. The improved selectivity, along with higher capacity, makes the Dionex IonPac AS column a better choice for higher-ionicstrength water samples and more complex matrices, such as domestic and industrial wastewaters. References. Fed. Regist. 998; Vol., No. 7. 2. Fed. Regist. 99; Vol. 9, No... The Determination of Inorganic s in Water by Ion Chromatography ; U.S. Environmental Protection Agency Method.; Cincinnati, Ohio, 99.. Greenberg, A. E.; Clesceri, L. S.; Eaton, A. D., Eds.; Standard Methods for the Examination of Water and, 8th ed.; Am. Public Health Assoc.: Washington, DC, 992.. Standard Test Method for s in Water by Chemically Suppressed Ion Chromatography ; D27-97; American Society for Testing and Materials: West Conshohocken, Pennsylvania, 999; Vol.., p 2 27. Application Note. Jackson, P.E.; Haddad, P.R.; Dilli, S. The Determination of Nitrate and Nitrite in Cured Meats Using High Performance Liquid Chromatography. J. Chromatogr., 98, 29, 7 78. www.thermoscientific.com/chromatography 2, 2 Thermo Fisher Scientific Inc. All rights reserved. ISO is a trademark of the International Standards Organization. All other trademarks are the property of Thermo Fisher Scientific and its subsidiaries. This information is presented as an example of the capabilities of Thermo Fisher Scientific products. It is not intended to encourage use of these products in any manners that might infringe the intellectual property rights of others. Specifications, terms and pricing are subject to change. Not all products are available in all countries. Please consult your local sales representative for details. Thermo Fisher Scientific, Sunnyvale, CA USA is ISO 9:28 Certified. Africa + Australia + 977 Austria + 8 282 2 Belgium +2 7 2 Brazil + 7 Canada + 8 87 China 8 8 8 (free call domestic) 8 AN79-EN 7M Denmark + 7 2 2 Europe-Other + Finland +8 292 2 France + 92 8 Germany +9 8 India +9 22 72 99 Italy +9 2 9 9 Japan +8 88 2 Korea +82 2 2 8 Latin America + 88 87 Middle East + Netherlands + 7 79 New Zealand + 9 98 7 Norway + 8 8 Russia/CIS + Singapore + 289 9 Sweden + 8 8 Switzerland + 7 77 Taiwan +88 2 87 UK/Ireland + 2 2 USA + 8 2 72