Sensitive Detection of Pyrethroids in Surface Water and Sediment

Similar documents
Application Note GCMS-01 Comparison of Ionization Techniques for the Analysis of Trace-Level Pyrethroid Insecticides by GC-MS/MS

Identification and Quantitation of PCB Aroclor Mixtures in a Single Run Using the Agilent 7000B Triple Quadrupole GC/MS

Pesticides Analysis Using the Agilent 5977A Series GC/MSD

ENVIRONMENTAL analysis

Sensitive Detection of 2-MIB and Geosmin in Drinking Water

High Sensitivity HPLC Analysis of Perchlorate in Tap Water Using an Agilent 6460 Triple Quadrupole LC/MS System

Determination of Hormones in Drinking Water by LC/MS/MS Using an Agilent InfinityLab Poroshell HPH Column (EPA 539)

Analysis of Ethanol and Isotopomers by 240 Quadrupole Ion Trap GC/MS

Analysis of Biomarkers in Crude Oil Using the Agilent 7200 GC/Q-TOF

Determination of N-Nitrosamines by USEPA Method 521 using Triple Quadrupole Gas Chromatography Mass Spectrometry

Author. Abstract. Introduction

DETERMINATION OF OCs, PCBs AND SYNTHETIC PYRETHROIDS IN ANIMAL FAT

Strategies for Analysis of Pyrethroid Insecticides in Complex Matrices

Rapid, Robust, and Sensitive Detection of 11-nor-D 9 - Tetrahydrocannabinol-9-Carboxylic Acid in Hair

Evaluation of Several Columns and Solvents for Post-Extraction Gel Permeation Chromatography (GPC) Clean-up of Fish Tissue Prior to PCB Analysis

Rapid Screening and Confirmation of Melamine Residues in Milk and Its Products by Liquid Chromatography Tandem Mass Spectrometry

Quantification of Pesticides in Food without Calibration using GC/FID with the Polyarc Reactor

Application Note. Gas Chromatography/Mass Spectrometry/Food Safety. Abstract. Authors

Determination of 17 Organotin Compounds in Beverages Using Triple Quadrupole GC-MS/MS System

Macrolides in Honey Using Agilent Bond Elut Plexa SPE, Poroshell 120, and LC/MS/MS

Enhanced Sensitivity for Biomarker Characterization in Petroleum Using Triple Quadrupole GC/MS and Backflushing

Determination of Organochlorine Pesticides and Polychlorinated Biphenyls Using GC/MS/MS Operated in the MRM Mode

The Importance of Area and Retention Time Precision in Gas Chromatography Technical Note

U.S. EPA Method 8270 for multicomponent analyte determination

Derivatization Procedure and Negative Chemical Ionization GC/MS/MS Conditions for the Analysis of Steroidal Analogs

Accelerated Solvent Extraction GC-MS Analysis and Detection of Polycyclic Aromatic Hydrocarbons in Soil

Determination of Off-Odor Compounds in Drinking Water Using an SPME Device with Gas Chromatography and Mass Spectrometry

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

Analytical determination of testosterone in human serum using an Agilent Ultivo Triple Quadrupole LC/MS

Highly Sensitive and Rugged GC/MS/MS Tool

Improved Volatiles Analysis Using Static Headspace, the Agilent 5977B GC/MSD, and a High-efficiency Source

GC Analysis of Polybrominated Flame Retardants Application

Comprehensive Pesticide Analysis in Juice Using a Combination of GC/MS and LC/MS Methods

Analysis of Trace (mg/kg) Thiophene in Benzene Using Two-Dimensional Gas Chromatography and Flame Ionization Detection Application

Analysis of Stachydrine in Leonurus japonicus Using an Agilent ZORBAX RRHD HILIC Plus Column with LC/ELSD and LC/MS/MS

Agilent 6460 Triple Quadrupole LC/MS System with an Agilent 1290 Infinity LC For Multi-Plant Growth Regulator Analysis in Grapes

Fast USEPA 8270 Semivolatiles Analysis Using the 6890/5973 inert GC/MSD with Performance Electronics Application

Determination of Total Volatile Organic Compounds in Indoor Air Using Agilent 7667A mini TD and 7820A GC

METHOD 8032A ACRYLAMIDE BY GAS CHROMATOGRAPHY

Cerno Bioscience MassWorks: Acquiring Calibration Data on Agilent GC/MSDs

Application Note. Abstract. Authors. Environmental

EPA Method 535: Detection of Degradates of Chloroacetanilides and other Acetamide Herbicides in Water by LC/MS/MS

UltiMetal Plus Advanced Chemistry for Stainless Steel Surface Deactivation

Quantitative Analysis of Water-Soluble B-Vitamins in Cereal Using Rapid Resolution LC/MS/MS. Application. Authors. Abstract.

Screening for Water Pollutants With the Agilent SureTarget GC/MSD Water Pollutants Screener, SureTarget Workflow, and Customized Reporting

Maximizing Triple Quadrupole Mass Spectrometry Productivity with the Agilent StreamSelect LC/MS System

Simultaneous Compound Identification and Quantification with Parallel Polyarc /FID and MS

Fast Screening and Quantitation of Perfluorinated Sources from Textiles using Chemical Ionization GC-MS

Determination of Chlorinated Acid Herbicides in Soil by LC/MS/MS Application Note

Analyze Hydrocarbon Impurities in 1,3-Butadiene with an Agilent J&W GS-Alumina PT Column

The Agilent 7700x ICP-MS Advantage for Drinking Water Analysis

Determination of Pesticide Residues in Oats by Automated. QuEChERS and LC/QQQ. Application Note. Abstract. Introduction

Plasma Metanephrines and 3-Methoxytyramine by LC/MS/MS Using Agilent SimpliQ WCX SPE, 1290 Infi nity LC, and 6460 Triple Quadrupole LC/MS

LC/MS/MS qua ntitation of β-estradiol 17-acetate using an Agilent 6460 Triple Quadrupole LC/MS working in ESI negative ion mode

1,2-Dibromoethane (EDB) and 1,2-dibromo-3-chloropropane (DBCP), gas chromatography, microextraction

Analysis of Opioids Using Isotope Dilution with GCMS-TQ8030 GC/MS/MS. No. GCMS No. SSI-GCMS-1401

Analysis of Low-Calorie Sweeteners by Liquid Chromatography-Tandem Mass Spectrometry

The 5975 inert MSD Benefits of Enhancements in Chemical Ionization Operation Technical Note

Analysis of Geosmin and 2-Methylisoborneol Utilizing the Stratum PTC and Aquatek 70

Analysis of Cannabinoids and Amphetamines in Serum by RRLC/Triple Quadrupole Mass Spectrometry Using a Multimode Ion Source. Application.

Quantitation of Ethyl Glucuronide and Ethyl Sulfate in Urine using LC-MS/MS

Screening of Pesticide Residues in Water by Sequential Stir Bar Sorptive Extraction-Thermal Desorption with GC/MS

Accurate analysis of neptunium 237 in a uranium matrix, using ICP-QQQ with MS/MS

Aminoglycosides in Milk Using Agilent Bond Elut Plexa SPE, Agilent Poroshell 120, and LC/Tandem MS

MS n Analysis With Fast Polarity Switching in the Agilent 1100 Series LC/MSD Trap SL. Application Note. Christine Miller Agilent Technologies

STANDARD OPERATING PROCEDURES SOP: 1828 PAGE: 1 of 14 REV: 0.0 DATE: 05/12/95 ANALYSIS OF METHYL PARATHION IN CARPET SAMPLES BY GC/MS

Determination of iopromide in environmental waters by ion chromatography-icp-ms

Food analysis. Solutions for Your Analytical Business Markets and Applications Programs

Identifying Pesticides with Full Scan, SIM, µecd, and FPD from a Single Injection Application

Agilent s New Weak Anion Exchange (WAX) Solid Phase Extraction Cartridges: SampliQ WAX

Determination of Cannabinoids (THC) in Biological Samples

Application Note 116 Monitoring VOCs in Ambient Air Using Sorbent Tubes with Analysis by TD-GC/MS in Accordance with Chinese EPA Method HJ

Accurate Mass Analysis of Hydraulic Fracturing Waters: Identification of Polyethylene Glycol Surfactants by LC/Q-TOF-MS

Fast and Reliable Method for the Analysis of Methylmalonic Acid from Human Plasma

Quantitative Determination of Dioxins in Drinking Water by Isotope Dilution using Triple Quadrupole GC-MS/MS

Advantages of Reversed Sandwich Injection for Pesticide Residue Analysis

Analysis of Extractable and Leachable (E&L) Compounds Using a Low-Energy EI-Capable High Resolution Accurate Mass GC/Q-TOF

Hints and Tips for APGC Coupled with Xevo TQ-S for Food and Environmental Analysis

Sensitive Femtogram Determination of Aflatoxins B 1 , B 2 , G 1. and G 2. in Food Matrices using Triple Quadrupole LC/MS. Authors.

Introducing the Agilent 7000A QQQ-MS for GC Sunil Kulkarni Product Specialist Agilent Technologies

PYRETHROID ANALYSIS. MARKET FOCUSED - Environmental TECHNICAL ARTICLE

Determination of Volatile Aromatic Compounds in Soil by Manual SPME and Agilent 5975T LTM GC/MSD

Application Note No Introduction. DMI Instrumentation. DMI Method. DMI for Lettuce and Pea Extracts. Lettuce results

Application Note. Author. Abstract. Introduction. Hydrocarbon Processing

METHOD 8033 ACETONITRILE BY GAS CHROMATOGRAPHY WITH NITROGEN-PHOSPHORUS DETECTION

STANDARD OPERATING PROCEDURES

A Direct 5 ms Column Performance Comparison for Active Semi-Volatile Analytes

[application note] INTRODUCTION EXPERIMENTAL. Specimens. Extraction

Determination of releasable 2,4,6-trichloroanisole in wine by cork stoppers (Resolution OIV-Oeno 296/2009)

METHOD 8100 POLYNUCLEAR AROMATIC HYDROCARBONS

Pyrethroid Concentrations in Surface Water and Bed Sediment in High Agricultural Use Regions of California. Keith Starner and K.

High-Speed Environmental Analysis Using the Agilent 7500cx with Integrated Sample Introduction System Discrete Sampling (ISIS DS)

Spectra Library Electron Capture Negative Ion (ECNI) Mass Spectra of Selected Polybrominated Diphenyl Ethers (PBDEs)

Gas Chromatography Mass Spectrometry (GCMS)

Accurate Analysis of Fuel Ethers and Oxygenates in a Single Injection without Calibration Standards using GC- Polyarc/FID. Application Note.

Analysis of Pesticides in Baby Foods Using a Triple-Quadrupole GC/MS/MS

Determination of trace elements in ultrapure semiconductor grade sulfuric acid using the Agilent 8900 ICP-QQQ in MS/MS mode

Agilent 6400 Series Triple Quadrupole Users Workshop. MH QQQ Users workshop 2/21/2014 1

Analysis of Total PCBs and PCB Congeners and Trans-nonachlor in Fish by Gas Chromatography/ Negative Chemical Ionization Single Ion Mass Spectrometry

Using UHPLC-Triple Quadrupole MS/MS to Detect the Presence of Bark Extract and Yohimbine Adulteration in Dietary Supplements and Botanicals

Transcription:

Sensitive Detection of Pyrethroids in Surface Water and Sediment Application Note Author Stephan Baumann Agilent Technologies, Inc. Santa Clara CA 91 USA Abstract A method has been developed on the Agilent 7 series Triple Quadrupole GC/MS using NCI GC/MS/MS with backflushing. It delivers estimated method detection limits (EMDLs) for pyrethroids as low as. parts per trillion (ppt) in water samples and.2 parts per billion (ppb) for sediment, enabling efficient monitoring of water sources for toxic pyrethroid levels. Introduction Pyrethroids are synthetic derivatives of pyrethrins, which are naturally occurring insecticides produced by the flowers of pyrethrum plants (Chrysanthemum cinerariiaefolium and C. coccineum). Pyrethroids were introduced 3 years ago to replace highly toxic organophosphates and DDT. Natural pyrethrins rapidly knock down flying insects, have low mammalian toxicity, and low persistence in the environment. However, they have poor efficacy in the field due to rapid degradation when exposed to light and air. The synthetic pyrethroids are substantially more resistant to degradation than pyrethrins and are now widely used in commercial pesticides, including more than 3 EPA-registered products. However, they have significantly higher mammalian toxicities and are highly toxic to fish and other aquatic organisms at extremely low concentrations (parts per billion). Pyrethroids have been found at acutely toxic levels in sediments and waterways in California, with residential water run-off being a major source of contamination due to structural and landscape pest control [1]. Monitoring for pyrethroids in water and sediments in urban and suburban water sources is therefore a high priority, and requires sensitive detection methods.

Initially, pyrethroid analysis was done using gas chromatography and an electron capture detector (GC/ECD). The use of gas chromatography/mass spectrometry (GC/MS) in electron ionization (EI) mode was later adopted as the method of choice, but it does not have sufficient sensitivity to meet all of the pyrethroids monitoring requirements. Negative Chemical Ionization (NCI) provides a much more sensitive and highly selective method for detecting pyrethroids, exploiting the electrophilic nature of halogenated compounds in the same way as electron capture detectors. This application note describes a method for high-sensitivity detection of pyrethroids in water and sediment using an efficient extraction procedure, GC backflushing with Agilent Capillary Flow Technology, and an Agilent 7 series Triple Quadrupole GC/MS in (NCI) mode [2]. Multiple Reaction Monitoring (MRM) MS/MS analysis delivers detection of pyrethroids at levels as low as.1 ppb, with a dynamic range of.1 to 1 ppb. Experimental The surrogate (used to monitor extraction and sample preparation) for water samples was dibromooctafluorobiphenyl (DBOB) from Restek Corp (Bellefonte, PA). For sediment samples the surrogate was dibutylchlorendate (DBCE), obtained from AccuStandard. GPC Calibration Standard Solution was obtained from Ultra Scientific and Florisil, 6/ mesh, PR grade was obtained from Floridin Corp (Berkeley Spring, WV). OI Analytical gel permeation chromatography (GPC) columns (J2 Scientific, Columbia, MO) were used to clean up extracts. The PFE cellulose filters were obtained from Dionex (Salt Lake City, UT). Instruments The experiments were performed on an Agilent 789A gas chromatograph equipped with a multimode inlet (MMI) and coupled to an Agilent 7B Triple Quadrupole GC/MS in negative chemical ionization (NCI) mode. Gas chromatography was performed using backflushing with a Purged Ultimate Union (Figure 1). The instrument conditions are listed in Table 1. Standards and Reagents All certified pyrethroid standards were obtained from Chem Service, Inc. (West Chester, PA). Second source standards of each analyte were obtained from AccuStandard (New Haven, CT) and Ultra Scientific (North Kingstown, RI) for verification of calibration standards. Individual stock solutions of each pyrethroid analyte were made in acetone from neat and combined, resulting in a stock mixture containing the six compounds: bifenthrin, es/fenvalerate, lambda-cyhalothrin, cyfluthrin, permethrin and permethrin. From this mixture, a matrix spiking solution used to fortify samples was made in acetone. The pyrethroid stock mixture was also used to make an instrument calibration mixture in isooctane, ranging from.1 1 ppb. Results are reported as total concentration for each analyte rather than as individual stereoisomer concentrations. External standard calibration was used to quantify the pyrethroids. Figure 1. Tandem GC/MS system configuration with mid-column back flushing. Backflushing removes higher boiling substances from the column prior to each subsequent run by flushing late eluting peaks out of the inlet split flow vent instead of driving them though the entire column and into the Agilent Triple Quadrupole GC/MS. Agilent uses a unique approach to backflushing employing Capillary Flow Technology that minimizes flow addition and subsequent decrease in sensitivity due to sub-optimal carrier gas flows into the mass spectrometer. Please refer to technical overview 99-8EN which describes mid-column backflushing in detail. 2

Table 1. Agilent 789N/7B Gas Chromatograph and Triple Quadrupole Mass Spectrometer Conditions GC run conditions Pre-column DB-XLB 1 m.2 mm,.2 µm J&W 122-1212 Analytical columns DB-XLB 1 m.2 mm,.2 µm J&W 122-1212 Injection volume 2 µl Inlet temperature 2 C Injection mode Pulsed splitless Oven program 1 C; min hold 3 C/min to 22 C; 1 minute hold C/min to 3 C; 2 minute hold Post-run oven program Time:. minutes Pre-column flow: -19.7 ml/min Analytical column flow: 2 ml/min Carrier gas Helium Transfer line temp 3 C MS conditions Tune NCI Ammonia Autotune EMV gain 2 Acquisition parameters NCI mode with multiple reaction monitoring (MRM) Reagent gas Ammonia, 3% flow Collision gas Nitrogen, constant flow,.7 ml/min Quench gas Helium, constant flow, 2.2 ml/min Solvent delay 1 minutes MS temperatures Source 1 C; Quadrupole 1 C Sample Preparation Water and sediment samples were prepared as previously described [3], with the procedure summarized in Figure 2. One liter water samples were collected in amber glass bottles, refrigerated at C and extracted as soon as possible. All water samples were fortified with pyrethroid pesticide spiking solution and DBOB surrogate directly into the sample bottles. The fortified water samples were manually agitated before extraction twice with 12 ml of dichloromethane (DCM). The extracts were dried using sodium sulfate and evaporated to approximately 2 ml using K-D glassware. The samples were then solvent exchanged by adding petroleum ether (PE) (~ ml three times). The extracts were evaporated to ml and further concentrated to less than 1 ml and brought to a final volume of 2 ml with isooctane. Sediment samples were homogenized and then extracted by automated Pressurized Fluid Extraction (PFE) with an ASE 2 Accelerated Solvent Extractor from Dionex (Salt Lake City, UT) equipped with 33 ml extraction cells and cellulose filters. Cleanup of extracts was accomplished using automated gel permeation chromatography. The mobile phase was dichloromethane (DCM) at a flow rate of ml/min. The GPC program was as follows: dump for 19 minutes, collect for 2 minutes, and then rinse for 1 minutes. The cleaned extracts were solvent exchanged from DCM to petroleum ether, then fractionated using 2 g of Florisil (6/ mesh) in an 11 mm 3 mm column. The appropriate fractions were then concentrated and reconstituted with isooctane to a final volume of 2 ml. Figure 2. A flow diagram of the sample preparative procedure shows the differences between the water and sediment procedures. PFE is an abbreviation for Pressurized Fluid Extraction and GPC is an abbreviation for Gel Permeation Chromatography. Analysis Parameters The Triple Quadrupole GC/MS parameters used are shown in Table 2. 3

Table 2. Pyrethroid Agilent 7B Triple Quadrupole GC/MS Analysis Parameters RT (min) MRM Bifenthrin 11.3 386&2 386&161 Cyhalothrin 13.1 21&2 21&3 Cyfluthrin and Cypermethrin 16.3 29&37 27&3 171&127 Deltamethrin 19.7 297&81 297&79 Fenvalerate and Es-fenvalerate 18.6 213&169 211&3 Permethrin 1.8 29&37 27&3 Results and Discussion Dwell time (ms) 7 7 7 7 7 NCI Mode Collision energy EV) 1 1 Gain Setting Minimum Detection Limits Due to the high toxicity of pyrethroids to aquatic life when present in water or sediments (<1 part per billion in some cases), very low detection limits are required of the methods used to test for these compounds. One meaningful way to set detection limits is to use the LC as a guide. The LC is the median lethal dose, which is defined as the concentration at which a toxic substance will kill half the members of a tested population. Pyrethroids are sequestered in soil sediments and break down in water, but water samples taken from aquatic sources contain dissolved solids which may be contaminated with pyrethroids. Therefore, the minimum detection limits (MDLs) can be set at the LC in water, and 1% of the LC in sediment. These limits are shown in Table 3. Table 3. Minimum Detection Limits Required for Pyrethroids in Water and Sediment Water ng/l (ppt) Sediment ng/g (ppb) LC LC Daphnia Mysidopsis LC MDL goal: Synthetic pyrethroid magnia bahia Hyalella 1% LC Bifenthrin 16. Cyfluthrin 11 1.1 Lambda Cyhalothrin.. Cypermethrtin 3.8. Es-fenvalerate 1 1 1. Permethrin 32 6 18 11 MDL: Method detection limit 2 2 7 2 2 7 Sensitive Detection Using NCI GC/MS/MS Initially, pyrethroid analysis was done using gas chromatography and an electron capture detector (GC/ECD). However, the high rate of false positives with GC/ECD led to the adoption of mass spectrometry in electron ionization (EI) mode as the method of choice. While GC/ MS/MS-EI is much more selective than GC/ECD, it is not quite sensitive enough to meet all of the MDL requirements. Negative Chemical Ionization (NCI) provides a less energetic and highly selective method for detecting and quantifying electrophilic compounds in complex matrices than electron ionization mass spectrometry. It exploits the electrophilic nature of halogenated compounds in the same way that electron capture detectors have in the past (Figure 3). The more critical type II pyrethroids such as cyfluthrin and permethrin are electronegative for NCI. However, there are at least two type I pyrethroids (resmethrin and fenpropathrin) that are not detected by traditional NCI techniques. Ammonia is preferred as the reagent gas instead of methane, because it has approximately seven times the thermalizing power of methane []. The advantage of using a tandem quadrupole mass spectrometer is that a very selective precursor to product ion transition data is generated, minimizing interferences and maximizing signal-to-noise. In addition, using MS/MS in NCI mode can significantly enhance sensitivity over the EI mode, while maintaining selectivity. Negative Chemical Ionization e (1 ev) & (buffer gas) & e (.1 ev) + M & M NCI Attributes Only electrophilic molecules are capable of capturing thermal electrons. Electron capture is extremely efficient for halogenated compounds, leading to high sensitivity. Matrix interferences typically do not capture electrons, and detection limits are generally very low due to lack of response by contaminants or matrix. Figure 3. Description of the Negative Chemical Ionization Process

EMDLs, IDLs and Quantification The GC/MS/MS-NCI method can deliver estimated method detection limits (EMDLs) well below most of the MLDs set for pyrethroids (Table ), using a fold concentration of water samples. This method provides detection of deltamethrin down to.1 ng/l, or.1 parts per trillion (ppt) in water samples. The instrument detection limits (IDLs) of standards in isooctane are also listed in Table. Detection of deltamethrin on the instrument has a dynamic range up to 1 ppb and an excellent R 2 value (.997) for the calibration (Figure ). Cypermethrin is detected down to.1 ppt in water samples. Detection on the instrument has a dynamic range up to 1 ppb and an even higher R 2 value (.9996) for the calibration (Figure ). Table. Ammonia MS/MS-NCI Instrument Detection Limit (IDL) and Estimated Method Detection Limit (EMDL)* Responses ( 1 ) 1. 1.9.8.7.6...3.2.1 Figure. Deltamethrin - 8 levels, 8 levels used, 8 points, 8 points used, QCs y = 16.22166x + 8.39123 R 2 =.9978 Type: linear, origin: ignore, weight: none 1 2 3 6 7 8 9 1 Concentration (µg/l) Calibration curve for deltamethrin at concentrations ranging from.1 to 1 µg/l, using ammonia as reagent gas in NCI GC/MS/MS. Synthetic pyrethroid IDL µg/l Estimated Method Detection Limit (EMDL) Water ng/l Bifenthrin.1..1 Cyfluthrin..1.2 Lambda Cyhalothrin.2..2 Cypermethrin..1.18 Deltamethrin..1.3 Es-fenvalerate.1.. Permethrin..1 1. Sediment ng/g The estimated EMDL values are based on -fold concentration of original water samples and 1g sediment sample to 1mL, with % moisture. The EPA defines the method detection limit as, "the minimum concentration of a substance that can be measured and reported with 99% confidence that the analyte concentration is greater than zero and is determined from analysis of a sample in a given matrix containing the analyte." Responses ( 1 ) Cyfluthrin-cypermethrin - 8 levels, 8 levels used, 8 points, 8 points used, QCs 3.2 2.8 2. 2. 1.6 1.2.8. Figure. y = 2963.891863x + 766.77211 R 2 =.9997968 Type: linear, origin: ignore, weight: none 1 2 3 6 7 8 9 1 Concentration (µg/l) Calibration curve for cypermethrin at concentrations ranging from. to 1 µg/l, using ammonia as collision gas in NCI GC/MS/MS.

Conclusion Ammonia NCI GC/MS/MS analysis of water and sediment sample extracts for pyrethroids provides increased sensitivity and selectivity over existing GC/ECD and GC/ MS/MS-EI methods. EMDLs using NCI GC/MS/MS have been established for water and sediment that are well below the toxicity of type II pyrethroids such as cyfluthrin and permethrin. Analysis of pyrethroids by this technique yields the sub-ppb detection limits that are required to ensure the safety of aquatic organisms. References 1. Weston D. P., Holmes R. W., You J., Lydy M. J. Aquatic toxicity due to residential use of pyrethroid insecticides. Environ Sci Technol. 39, 9778-978 (2). 2. Presentation by Dr. Abdou Mekebri at the September 29, 21 Bay-Delta Science Conference. 3. Mekebri, D. B. Crane, G. J. Blondina, D. R. Oros and J. L. Rocca Extraction and Analysis Methods for the Determination of Pyrethroid Insecticides in Surface Water. Sediments and Biological Tissues at Environmentally Relevant Concentrations. Bull Environ Contam Toxicol. 8, -6. (28).. Modern Practice of Gas Chromatography, R.L. Grob and E.F. Barry (Eds.), John Wiley & Sons, Hoboken, NJ, th ed., p. 382 (2). For More Information These data represent typical results. For more information on our products and services, visit our Web site at www.agilent.com/chem. www.agilent.com/chem Agilent shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material. Information, descriptions, and specifications in this publication are subject to change without notice. Agilent Technologies, Inc., 212 Printed in the USA March 29, 212 991-116EN